Ten-year pathological study of Beall mitral valve prostheses.

K S Chang, J Fernandez, S D Cha, S S Yang
PMCID: PMC459265  PMID: 6210973

Abstract

Detailed pathological studies are reported on 20 cloth-covered Beall mitral valve prostheses, which had to be removed for valve failure nine months to 10 years after implantation. Results in all patients except one indicated that there was a progressive and persistent granulomatous reaction that continued up to at least 10 years. Although the rate of tissue ingrowth into the valve decreased with time, the total volume of tissue formation increased. Cloth wear and disruption and deformity of the teflon discs was frequently seen. In these 20 valves the desired result--namely, the complete and smooth endothelialisation of the prosthesis--was certainly not achieved.

Full text

PDF
137

Images in this article

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Adams D. O. The granulomatous inflammatory response. A review. Am J Pathol. 1976 Jul;84(1):164–192. [PMC free article] [PubMed] [Google Scholar]
  2. Braunwald N. S., Morrow A. G. Tissue ingrowth and the rigid heart valve. Review of clinical and experimental experience during the past year. J Thorac Cardiovasc Surg. 1968 Sep;56(3):307–322. [PubMed] [Google Scholar]
  3. Davila J. C., Amongero F., Sethi R. S., Rincon N. L., Palmer T. E., Lautsch E. V. The prevention of thrombosis in artificial cardiac valves. Ann Thorac Surg. 1966 Sep;2(5):714–741. doi: 10.1016/s0003-4975(10)66643-x. [DOI] [PubMed] [Google Scholar]
  4. Fermandez J., Morse D., Spagna P., Lemole G., Gooch A., Yang S. S., Maranhao V. Results of mitral valve replacement with the Beall prosthesis in 209 patients. J Thorac Cardiovasc Surg. 1976 Feb;71(2):218–225. [PubMed] [Google Scholar]
  5. Goldner R. D., Adams D. O. The structure of mononuclear phagocytes differentiating in vivo. III. The effect of particulate foreign substances. Am J Pathol. 1977 Nov;89(2):335–350. [PMC free article] [PubMed] [Google Scholar]
  6. Nichols H. T., Fernandez J., Morse D., Gooch A. S. Improved results in 336 patients with the isolated mitral Beall valve replacement. Chest. 1972 Sep;62(3):266–272. doi: 10.1378/chest.62.3.266. [DOI] [PubMed] [Google Scholar]
  7. O'Connell T. X., Fee H. J., Golding A. Sarcoma associated with dacron prosthetic material: case report and review of the literature. J Thorac Cardiovasc Surg. 1976 Jul;72(1):94–96. [PubMed] [Google Scholar]
  8. Page R. C., Davies P., Allison A. C. Participation of mononuclear phagocytes in chronic inflammatory diseases. J Reticuloendothel Soc. 1974 May;15(5):413–438. [PubMed] [Google Scholar]
  9. Silver M. D., Wilson G. J. The pathology of wear in the Beall model 104 heart valve prosthesis. Circulation. 1977 Oct;56(4 Pt 1):617–622. doi: 10.1161/01.cir.56.4.617. [DOI] [PubMed] [Google Scholar]
  10. Spector W. G. The granulomatous inflammatory exudate. Int Rev Exp Pathol. 1969;8:1–55. [PubMed] [Google Scholar]
  11. Szilagyi D. E., Smith R. F., Elliott J. P., Allen H. M. Long-term behavior of a dacron arterial substitute: clinical, roentgenologic and histologic correlations. Ann Surg. 1965 Sep;162(3):453–477. doi: 10.1097/00000658-196509000-00015. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Thorax are provided here courtesy of BMJ Publishing Group

RESOURCES