Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2017 Feb 1.
Published in final edited form as: Magn Reson Med. 2015 Apr 4;75(2):680–687. doi: 10.1002/mrm.25627

Multi-site evaluations of a T2-Relaxation-Under-Spin-Tagging (TRUST) MRI technique to measure brain oxygenation

Peiying Liu 1, Ivan Dimitrov 1,2, Trevor Andrews 2,3, David E Crane 4, Jacinda K Dariotis 5, John Desmond 6, Julie Dumas 7, Guillaume Gilbert 2,8, Anand Kumar 9, Bradley J Maclntosh 4, Alan Tucholka 8, Shaolin Yang 9,10,11, Guanghua Xiao 1, Hanzhang Lu 1
PMCID: PMC4592780  NIHMSID: NIHMS653918  PMID: 25845468

Abstract

Purpose

Venous oxygenation (Yv) is an important index of brain physiology and may be indicative of brain diseases. A T2-relaxation-under-spin-tagging (TRUST) MRI technique was recently developed to measure Yv. A multi-site evaluation of this technique would be an important step toward broader availability and potential clinical utilizations of Yv measures.

Methods

TRUST MRI was performed on a total of 250 healthy subjects, with 125 from the developer’s site and 25 each from five other sites. All sites were equipped with a 3T MRI of the same vendor. The estimated Yv and the standard error of the estimation, εYv, were compared across sites.

Results

The averaged Yv and εYv across six sites were 61.1±1.4% and 1.3±0.2%, respectively. Multivariate regression analysis showed that the estimated Yv was dependent on age (p=0.009), but not on performance site. In contrast, the standard error of Yv estimation was site-dependent (p=0.024), but was all less than 1.5%. Further analysis revealed that εYv was positively associated with the amount of subject motion (p<0.001) but negatively associated with blood signal intensity (p<0.001).

Conclusion

This work suggests that TRUST MRI can yield equivalent results of Yv estimation across different sites.

Keywords: blood oxygenation, TRUST, brain, venous oxygenation

INTRODUCTION

Venous oxygenation (Yv) is the fraction of oxygenated hemoglobin in the venous blood. Quantification of Yv in the brain can be used to assess the brain’s oxygen homeostasis. When combined with other physiologic measures, such as arterial oxygenation and cerebral blood flow, one can estimate important physiological parameters like oxygen extraction fraction (OEF) and cerebral metabolic rate of oxygen (CMRO2) (13), which are key markers for tissue viability and brain function. Altered Yv, OEF and CMRO2 have been found in studies on normal aging (4,5), early brain development (6), and a number of diseases, such as multiple sclerosis (7), carotid artery disease (8), congenital heart disease (9,10) and drug addiction (11). Quantification of Yv is also important for understanding the mechanism of blood-oxygenation-level-dependent (BOLD) functional magnetic resonance imaging (fMRI), as it is well-known that the BOLD fMRI signal is a complex function of blood oxygenation, blood flow, and oxygen consumption in the brain (12,13). Previous studies showed that Yv can be used to normalize BOLD fMRI signal and thereby improve the characterization of neural activity (14,15).

Quantification of Yv is, however, not yet a standard procedure. Up until recently, it was considered a “niche market” of positron emission tomography (PET), which requires an onsite cyclotron, the use of 15O-labeled radiotracers, and an arterial line for dynamic blood sampling (16,17). Therefore, there is a growing interest to quantify Yv using non-invasive methods. A few techniques have been reported to measure Yv based on the susceptibility effects of deoxyhemoglobin on extravascular tissue (18,19) or intravascular blood (3,20). Previously, we have reported a T2-Relaxation-Under-Spin-Tagging (TRUST) MRI technique to obtain quantitative value of Yv in the superior sagittal sinus by measuring the T2 of venous blood (21). An advantage of TRUST MRI in comparison with previous T2-based techniques (2224) is that, TRUST utilizes the spin-tagging principle on the venous side to automatically isolate pure venous blood signal and minimize partial voluming effects from surrounding tissue, rather than relying on the post-processing operator to manually select a region of pure blood. It has been validated in humans by comparing TRUST-derived arterial Y values with those measured with the gold-standard Pulse Oximeter (25), and has also been shown to be sensitive in detecting oxygenation changes due to hypercapnia (26), hyperoxia (27), hypoxia (27), glucose (28) and caffeine (29) challenges. The current TRUST protocol has several attractive features including the absence of any exogenous agent, short scan duration of 1.2 min, and feasibility on a standard 3T scanner. Moreover, it has been shown that the single site TRUST test-retest variations was <2% (30,31), which is reliable enough potentially for use as a functional biomarker for brain diseases (47,11).

There is a growing interest in using TRUST to measure Yv under various physiological and pathophysiological conditions. As an initial effort, the TRUST sequence has been disseminated to a number of sites over the world. However, the performance and reliability of TRUST MRI across difference sites have not been assessed. Given that TRUST MRI requires special pulse sequence design and slice positioning, it is unclear whether differences in system configuration and operator experience would result in a significant variability in Yv measurement across sites.

In the present work, we evaluated the TRUST MRI technique in a multi-site setting. The aims of this study are twofold: 1) to show that TRUST can be performed at different sites without the need for special hardware or on-site training of the operators; 2) to show that TRUST measurements are comparable across sites. Altogether, six imaging centers located in North America participated in the study, with a total of 250 healthy volunteers. The measured Yv values (i.e. accuracy) as well as the standard error of Yv estimation (i.e. precision) were compared across the six sites.

METHODS

TRUST MRI technique

TRUST MRI (21) is based on the principle that T2 relaxation time of the blood has a well-established relationship with Yv (32), thus one can measure pure blood T2 and convert T2 to Yv through calibration. The technique uses an RF pulse to magnetically label the incoming venous blood, and acquires images with and without the labeling to obtain control and labeled images (Fig. 1a). The subtraction of the control and labeled images yields an image from pure venous blood signal (Fig. 1b). The labeled and control scans are repeated flow-insensitive T2-preparation pulses so as to modulate the signal with T2 weighting (Fig. 1b). The blood signal (ΔS) and the T2-preparation duration (termed effective echo time [eTE]) have the following relationship:

ΔS(eTE)=S0·eeTE·(1/T1b-1/T2b), Eq. [1]

where S0 is the pure venous blood signal with no T2 weighting, T1b is the blood T1 (assumed to be 1624ms (33), and T2b is blood T2. Therefore, the monoexponential fitting of ΔS and eTEs gives the T2 value of the venous blood (Fig. 1c). Using a calibration plot specifying the relationship between T2, Y and Hematocrit level (25), the blood T2 can then be converted to Yv. Due to relatively large flow velocities and therefore pronounced inflow of the labeled spins, TRUST measurement in large venous vessels, e.g. superior sagittal sinus, was found to be particularly robust (2,31) and is the focus of this study, which provides an estimation of the global venous oxygenation in the brain.

Fig. 1.

Fig. 1

Illustration of the positioning of TRUST MRI and representative results. (a) Imaging slice (red) and labeling slab (blue) of the TRUST MRI scan. The imaging slice was positioned to be parallel to AC-PC line and 20mm above the sinus confluence. (b) Typical data of TRUST MRI. The subtraction of the control and labeled images yields pure venous blood signal, which is subject to increasing T2 weightings. (c) Monoexponential fitting of the blood signal in superior sagittal sinus as a function of eTE results in T2 estimation. The T2 value can then be converted to venous oxygenation via a calibration plot (25).

Sites and participants

Of the six participating sites, four were located in the Unites States and two in Canada. The sites were all equipped with 3T Achieva whole body MRI systems (Philips Healthcare, Best, The Netherlands). The developer’s site, which is the site of the lead authors, scanned 125 healthy volunteers and these data served as the benchmarks. Each of the five remote sites recruited and scanned 25 healthy volunteers. As such, a total of 250 subjects were studied: 130 males and 120 females with an age range of 18 to 93 years. Table 1 lists demographic information of the participants in all sites. The ethnic composition was distributed such that there were 187 Caucasians, 46 African Americans, 15 Asians and 2 others. The Health Insurance Portability and Accountability Act (HIPAA) compliant protocol was approved by the ethics committee of each site, specifically the UT Southwestern Medical Center institutional review board (IRB), the Johns Hopkins Medicine IRB, the Sunnybrook Research Institute Research Ethics Board, the University of Illinois at Chicago IRB, the Research Ethics Board of Medical Center of the University of Montreal, and the University of Vermont IRB. Informed written consent was obtained from each participant. The TRUST data from each remote site were acquired by adding the sequence to an existing research study, thus the age ranges were not matched across the sites.

Table 1.

Demographic information of the subjects from all sites.

Site a b c d e F
Subject number 125 25 25 25 25 25
Age range (years) 18–93 21–70 31–82 22–44 18–24 20–40
Gender (M/F) 65/60 8/17 12/13 21/4 13/12 11/14

Data acquisition

The executable files of the customized TRUST pulse sequence were transferred electronically to all sites. Each remote site scanned 25 healthy volunteers and transferred the datasets to the developer’s site for centralized analysis. Before data transferring, any personal information from subjects was removed in accordance with the patient protection regulations. The developer’s site collected a relatively large sample size and age range so that the data would serve as a reference for comparison with the other sites.

Identical imaging parameters were used in all sites and the protocol was based on that optimized in a previous study (30). The imaging parameters were: voxel size 3.44×3.44×5mm3, TR=3000ms, TI=1022ms, four eTEs: 1, 40, 80 and 160ms, labeling thickness 100mm, gap 22.5mm, scan duration 1.2min.

To ensure consistent slice positioning across sites, an instruction document that included several examples (see Supporting Fig. S1) was sent to each site. Operators were instructed to position the TRUST imaging slice to be parallel to anterior-commissure posterior-commissure line with a distance of 20mm above the sinus confluence. This positioning scheme has been shown to provide sufficient labeling of venous blood in the superior sagittal sinus (SSS) in virtually all adult brains (5,31).

Data analysis

The TRUST MRI data were processed using in-house MATLAB (Mathworks, Natick, MA) scripts. The processing method followed that described previously (2,21,31). Briefly, after motion correction, pairwise subtraction between control and tag images was performed, the difference of which yields pure venous blood signal. A preliminary region of interest (ROI) was manually drawn on the difference image to include the superior sagittal sinus. Then four voxels with the highest signals in the difference images in the ROI were chosen as the final mask for spatial averaging. The model described by Eq. [1] was then fitted to the averaged venous blood signals to obtain a T2 estimate, which was in turn converted to Yv via a calibration plot obtained by in vitro blood experiments under controlled oxygenation, temperature, and hematocrit conditions (25). In the T2-Yv conversion, the hematocrit was assumed to be 0.40 for female and 0.42 for male (5). The precision of the Yv measurement, reflecting the uncertainty of the estimation, was quantified by the standard error of the estimated YvYv). Specifically, the goodness-of-fit from the T2 fitting gives a 95% confidence interval (CI) of the estimated T2, which was further converted to the confidence interval of the estimated Yv via the calibration plot. Then εYv was calculated by εYv=(upper CI of Yv − lower CI of Yv)/(2*1.96).

In addition to Yv, two other parameters were also extracted from the TRUST data, the amount of subject head motion during the TRUST scan and the blood signal intensity, since we conceived that both of these variables could affect Yv and εYv. The head motion amount was approximated by the standard deviation of the motion vector (in units of mm) obtained from the image realignment algorithm in the software Statistical Parametric Mapping (SPM) (University College London, UK). It is a measure of the subject cooperation in the scan. The blood signal intensity was calculated as the ratio between the difference (i.e. control minus labeled) signal and the control signal (which contains both blood and tissue signals) within the final mask of SSS at the first eTE. This parameter is indicative of the size of the SSS.

Statistical analysis

One-way Analysis of Variance (ANOVA) test was used to evaluate if there is a site difference in age, gender, motion, and the blood signal intensity. The statistical significance of these effects was assessed with a Bonferroni-corrected p value of 0.05. Next, to evaluate the site-effect on Yv and εYv, a multivariate regression analysis was conducted with age, gender, motion and blood signal intensity as continuous variables and site as a categorical variable. This analysis was conducted separately for Yv and εYv. If a significant site-effect was detected, each site was compared to the reference site (denoted as Site “a”) to identify which site(s) is different and whether the TRUST performance at that site is better or worse than the Site “a”. An examination of the site-effect was also performed using one-way ANOVA on the “residual” Yv and εYv (denoted by Yv¯ and εYv¯, respectively), in which the effects of age, gender, motion and blood signal intensity were regressed out. In all analyses, a Bonferroni-corrected p value of 0.05 or less was considered statistically significant.

RESULTS

General results

The transfer of the TRUST sequence and slice positioning procedures to the participating sites went smoothly. The investigators and MR technologists of each site were able to install the sequence and properly position the slice by simply following the written instructions. There was no instance of a need for onsite training.

The age and gender information from each site are shown in Table 1. The ANOVA test showed that there was a significant difference in age and gender across the six sites (p<0.001 and p=0.031, respectively), which was due to the differences in their respective parent studies. The amount of head motion and blood signal intensities are shown for each site in Fig. 2a and b, respectively. No site difference was found in these two parameters (p=0.80 and p=0.95 for head motion and blood signal intensity, respectively), indicating that subject cooperation and mean SSS size are comparable across the sites.

Fig. 2.

Fig. 2

Site-comparison of the TRUST measurements. (a) Averaged amount of motion in the scans at the six sites. (b) Averaged blood signal intensity at the six sites. (c) Averaged Yv at the six sites. (d) Averaged εYv at the six sites. Significant site-difference was found in εYv.

Evaluation of venous oxygenation (Yv) across sites

The mean Yv of all 250 subjects was 61.0±5.8%, ranging from 42.1% to 76.6%. This range of Yv in healthy subjects is consistent with previous reports (5,34). The inter-subject standard deviation, 5.8%, is also similar to earlier studies and is thought to be primarily attributed to normal variations in venous oxygenation (5,31,34). Fig. 2c shows the measured Yv categorized by sites. The inter-site standard deviation was 1.4%. Multivariate regression analysis of Yv as a function of age, gender, motion, blood signal intensity and site revealed that, the performance site has no significant effect on the measured Yv (p=0.85), indicating that there was no site dependence in Yv estimation using TRUST. As shown in Fig. 3a, there was a significant effect of age (p=0.009 from multivariate regression analysis). No significant effects of gender (p=0.99), motion (p=0.15, Fig. 3b) and blood signal fraction (p=0.25, Fig 3c) in Yv were found. When the effects of age, gender, motion, and blood signal fraction were regressed out, ANOVA test on Yv¯ still showed no significant effect of performance site (p=0.75).

Fig. 3.

Fig. 3

Scatter plots between the TRUST results and factors including subject age, motion and blood signal intensity. (a–c) Scatter plots between Yv and age, motion and blood signal intensity, respectively. (d–f) Scatter plots between εYv and age, motion and blood signal intensity, respectively.

Evaluation of the standard error of YvYv) across sites

The mean εYv of all the 250 TRUST-Yv measurements was found to be 1.4±0.6%, ranging from 0.4% to 3.9%. Fig. 2d shows the εYv categorized by sites, with an inter-site standard deviation of 0.2%. For all sites, the mean εYv were lower than 1.5%, suggesting a generally good measurement precision using TRUST. Multivariate regression analysis of εYv revealed that εYv is not dependent on age (p=0.26, Fig. 3d) or gender (p=0.29). However, significant effects of head motion (p<0.001, Fig. 3e) and blood signal intensity (p<0.001, Fig. 3f) were found. Subjects with greater head motion or lower blood signal intensity tended to have a larger εYv. The multivariate regression analysis also revealed that there was significant site-effect in εYv (p=0.024). To confirm the presence of the site-effect, one-way ANOVA test was performed on εYv¯ after regressing out the effects of age, gender, motion and blood signal intensity, and a significant site-effect remained (p=0.015). Post-hoc comparison of the εYv between the reference site and each of the participating site suggested that the site-difference was attributed to a significantly lower (i.e. better performance) εYv in Site “e” (p=0.024) and a trend of lower εYv in Site “c” (p=0.081), as the other sites showed no differences from the reference site (p>0.38).

We further conducted exploratory analysis to examine the reason underlying the site-difference in εYv. We hypothesized that the degree to which the refocusing pulse is accurate could influence εYv, since deviation in the flip angle away from the ideal 180° will result in the signal decay that does not follow a monoexponential curve. Therefore, we quantified an additional parameter, the discrepancy between the measured blood signals and the monoexponential fitting curve, by calculating the difference between the mean experimental data and the fitted values, i.e., Δm-f=1/4·eTE|mean(ΔS(eTE))-S0·eeTE·(1/T1b-1/T2b)|/S0, where eTE=1, 40, 80, and 160ms. It was found that, after regressing out Δm-f (in addition to age, gender, motion, and blood signal intensity), εYv was no longer significantly dependent on site (p=0.41), suggesting that RF pulse flip angle accuracy may be one of the main reasons for the slight differences in εYv across sites.

DISCUSSION

The present study evaluated the TRUST technique for Yv measurement in a multi-site setting. It is shown that the TRUST sequence can be implemented and performed on a standard 3T scanner at different sites. The TRUST results from the six imaging centers revealed that there was not a site difference on the estimated Yv values. The precision of the Yv estimation, measured by the standard error of Yv, are in general good at all sites (<1.5%), but showed some variability in some participating sites. To our knowledge, this work represents the first effort to conduct a multi-site evaluation of an MR-based oxygenation technique, which is a critical step toward broader availability and potential clinical utilizations of cerebral oxygenation and oxygen consumption measures.

Cerebral venous oxygenation is an important physiological parameter of the brain. Given that the arterial blood is typically fully oxygenated, the knowledge about venous oxygenation can inform how much oxygen has been extracted by the brain for its energy consumption. Indeed, when combined with blood flow measurement, venous oxygenation can be used to calculate the brain’s metabolic rate based on Fick’s principle (1). The brain’s metabolic rate is closely related to neural activity and brain health, and its abnormality has been implicated in several neurodegenerative and metabolic diseases (711,35). Therefore, a clinically practical approach to determine cerebral venous oxygenation may provide a valuable tool for diagnosis and treatment monitoring in many brain disorders. Unfortunately, up until recently, PET based measurement using 15O-labeled tracer was the only available approach for use in humans. While the PET method provides a quantitative, spatially resolved map of OEF and CMRO2, its technical complexity has limited its utility in routine clinical practice. For MR based methods, although BOLD fMRI has been used in brain mapping for more than two decades, it remains a qualitative method and its signal has no direct indication of the actual blood oxygenation in the brain. Over the past few years, several quantitative oximetry methods (3,20,21,24,3639) have been proposed, of which the T2-based approach is one method. It remains to be seen which of these methods will make it to the clinical arena, and it is possible that it will depend on the particular disease type. The present work represents an initial effort to lay preliminary foundation for a broader availability of the global oxygenation method using T2-based approach.

In this multisite setting, the transferring of the TRUST sequence was smooth. Since the imaging protocol has been streamlined, no change or modification was needed at any of the individual sites. No datasets had to be excluded due to improper positioning, suggesting that the written instruction of positioning, together with example figures of the positioning (as shown in Supporting Fig. S1), were sufficient for investigators and MR technologists to plan the TRUST scan without previous experience with this sequence. The relatively short duration (1.2 min) of the TRUST scan also allowed it be added to existing protocols without adding much burden to the participants. These features may enhance the potential of the TRUST technique to be disseminated to researchers and clinicians who are interested in non-invasive measurement of venous oxygenation in the brain.

The key advantage of TRUST is that the partial voluming effect is minimized. Specifically, partial voluming due to static tissue can be automatically removed by the control/label subtraction in date processing. Therefore, the TRUST results was shown to be independent of imaging resolution (21), and not relying on processing operator to manually select a ROI of pure blood.

The Yv values measured by TRUST are in good agreement with those measured by other techniques that have been reported in literature (3,17,24,34,37,38,40). For example, Coles et al reported an averaged Yv of 58±4% in 10 healthy subjects between 18–60 years old using the gold-standard 15O-PET (17). Recent studies using other MRI techniques reported an averaged Yv between 59.6% and 67.2% in young healthy volunteers (3,24,37,38,40).

We found that there is no site-dependence in the TRUST-measured Yv, suggesting that venous oxygenation can be measured reliably across scanners at different sites. The observation of an absence of effects of head motion and blood signal intensity on Yv indicates that the TRUST-Yv measurement requires no additional subject cooperation, compared to other physiological MR sequences, nor was the size of SSS an issue. Our multisite data showed that the measured Yv significantly decreased with age (p=0.009, Fig. 3a). This observation is consistent with previous single-site studies of normal aging (4,5). The decrease of Yv indicates that a greater fraction of the incoming oxygen is extracted in older individuals. Our previous studies have suggested that this may be due to a combination of two reasons (4,5). One is that CBF decreases with age, thus the brain has to increase the OEF to match the demand. The other reason is that the brain’s metabolic demand, i.e. CMRO2, is actually greater in older individuals, which is postulated to be compensatory and due to a lower processing efficiency in the aging brain (4,5).

Gender did not have a significant effect on Yv, when using the assumed hematocrit values of 0.40 and 0.42 in female and male, respectively. This may be expected as the optimal oxygen tension of the female and male brain is likely similar. We also tested to use other hematocrit assumptions and found that, when the assumed hematocrit difference is greater than 0.06 (i.e. 0.40 and 0.46 for female and male, respectively), a significant Yv difference would be identified. However, this assumption of large difference is not compatible with the existing literature on hematocrit in female and male individuals.

The standard error of the Yv measurement using the TRUST technique, εYv, was found to be less than 1.5% in all six sites. This amount of uncertainty is considerably less than the inter-subject normal variation of approximately 6% as well as disease-related change of 5.7% in Multiple Sclerosis (7), 45% in Stroke (41), 6.3% in Alzheimer’s disease (42). Thus, this level of precision is considered good in general. Statistical analysis showed that the εYv depends on performing site (p=0.024). Comparing the five remote sites to the developer’s site, one site showed significantly smaller εYv (p=0.024) and another site showed a trend of smaller εYv (p=0.081). No site was found to have a larger εYv than the reference site. We also examined the potential correlation between εYv and Yv, with the notion that a noisy dataset may result in a systematic bias (higher or lower Yv values). We did not observe a significant correlation (p=0.23, N=250), indicating that noise affects the precision but not the accuracy of the Yv estimation by the TRUST technique.

Regression analysis revealed that the site-difference in εYv could be explained by the difference in the measurement-fitting discrepancy (Δm-f) across sites. When the Δm-f effect was regressed out, εYv was no longer dependent on the performance site. Δm-f is expected to be mainly due to the imperfection in the 180° refocusing pulses, i.e. B1 inhomogeneity. Slight imperfection in the refocusing pulses could cause the measured blood signals to deviate from the ideal monoexponential decay curve, thereby resulting in a large Δm-f. Therefore, we speculate that the site-difference in εYv may be due to the performance of the B1 preparation on an individual scanner. However, we emphasize again that the value of εYv was in general low and the range of εYv across sites was relatively narrow (0.95%–1.48%, Fig. 2d).

The multivariate regression analysis also revealed that there are no age and gender effects on εYv. This finding suggested that the measurement precision of TRUST is comparable between young and older subjects and between females and males.

The blood signal intensity is the difference signal normalized to the control signal, and a measure of the intrinsic SNR of the TRUST data. In our data, the blood signal intensity is around 50% for all sites (Fig. 2b). As a reference, the difference signal in ASL is typically only 0.5–2% of the control signal (4345). Thus, SNR of TRUST MRI is considerably higher than that of ASL, which allows a reliable T2 fitting and measurement of Yv. The main reason that the blood signal intensity does not approach 100% is that there exists partial voluming between tissue and blood in the voxels, thus blood may not occupy the entire voxel. In this regard, the blood signal intensity provides an indirect indication of the size of the SSS. The blood signal intensity in principle is also affected by the labeling efficiency of the labeling RF pulses. In a pulsed labeling scheme, however, the efficiency is usually close to 100% (46), thus this is not expected to be a major factor in determining the blood signal intensity.

There are a few limitations in this multi-site study. One limitation is that the age and gender distributions were not exactly matched across sites, due to the study design that the TRUST sequence was added to an existing study at the remote site and the studies at the remote sites had different subject demographics. A true multi-site trial in which subjects of each site are recruited solely for the purpose of the trial would involve a substantially greater budgetary requirement and should ideally include multiple MRI vendors, which will be the goal of our future studies. To alleviate this limitation, we included the participant’s age and gender as covariates in the analysis and their effects were regressed out when evaluating the site-effect on the TRUST measurements. We also included a large sample over the adult lifespan at the reference site, to facilitate age and gender matched comparison. Another limitation is that this study has only included sites from a single MRI vendor, because of the convenience in implementing an identical pulse sequence (e.g. matching all gradient waveforms and RF pulse shapes). Therefore, it is unclear if there would be a vendor-difference in the TRUST-Yv measurements. Thus, a multi-vendor study should be performed in future. Furthermore, no participants were scanned at different sites in this study. To better assess the site-effect, future studies should consider scan the same participants at multiple locations.

CONCLUSION

In the present work, we evaluated the measurement of venous oxygenation using TRUST MRI across 6 imaging centers. The accuracy as well as the precision of the TRUST-Yv measurement was assessed. The TRUST sequence can be implemented and performed on a standard 3T scanner at remote sites with high success rate. The results revealed that the estimated Yv values are compatible across sites after age-related Yv difference is accounted for. All six sites showed good precision of the TRUST measurements, with one remote site superior to the reference site. The precision of the Yv measure was found to be affected by subject head motion and size of the blood vessel. This work suggests that TRUST MRI, as a promising MR technique to measure brain oxygenation, has the potential to be used effectively at a broad range of MRI facilities.

Supplementary Material

Supp FigureS1

Supporting Fig. S1: Positioning instructions provided to the participating sites.

Acknowledgments

Grant Sponsors: NIH R01 MH084021, NIH R01 NS067015, NIH R01 AG042753, NIH R21 NS078656, and NIH R21 NS085634.

References

  • 1.Kety SS, Schmidt CF. The Effects of Altered Arterial Tensions of Carbon Dioxide and Oxygen on Cerebral Blood Flow and Cerebral Oxygen Consumption of Normal Young Men. J Clin Invest. 1948;27(4):484–492. doi: 10.1172/JCI101995. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Xu F, Ge Y, Lu H. Non-invasive Quantification of Whole-brain Cerebral Metabolic Rate of Oxygen by MRI. Magn Reson Med. 2009;62(1):141–148. doi: 10.1002/mrm.21994. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Jain V, Langham MC, Wehrli FW. MRI estimation of global brain oxygen consumption rate. J Cereb Blood Flow Metab. 2010;30(9):1598–1607. doi: 10.1038/jcbfm.2010.49. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Lu H, Xu F, Rodrigue KM, Kennedy KM, Cheng Y, Flicker B, Hebrank AC, Uh J, Park DC. Alterations in cerebral metabolic rate and blood supply across the adult lifespan. Cereb Cortex. 2011;21(6):1426–1434. doi: 10.1093/cercor/bhq224. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Peng SL, Dumas JA, Park DC, Liu P, Filbey FM, McAdams CJ, Pinkham AE, Adinoff B, Zhang R, Lu H. Age-related increase of resting metabolic rate in the human brain. NeuroImage. 2014;98:176–183. doi: 10.1016/j.neuroimage.2014.04.078. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Liu P, Huang H, Rollins N, Chalak LF, Jeon T, Halovanic C, Lu H. Quantitative assessment of global cerebral metabolic rate of oxygen (CMRO2) in neonates using MRI. NMR in biomedicine. 2014;27(3):332–340. doi: 10.1002/nbm.3067. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Ge Y, Zhang Z, Lu H, Tang L, Jaggi H, Herbert J, Babb JS, Rusinek H, Grossman RI. Characterizing brain oxygen metabolism in patients with multiple sclerosis with T2-relaxation-under-spin-tagging MRI. J Cereb Blood Flow Metab. 2012;32(3):403–412. doi: 10.1038/jcbfm.2011.191. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Tsuchida C, Kimura H, Sadato N, Tsuchida T, Tokuriki Y, Yonekura Y. Evaluation of brain metabolism in steno-occlusive carotid artery disease by proton MR spectroscopy: a correlative study with oxygen metabolism by PET. Journal of nuclear medicine: official publication, Society of Nuclear Medicine. 2000;41(8):1357–1362. [PubMed] [Google Scholar]
  • 9.Durduran T, Zhou C, Buckley EM, Kim MN, Yu G, Choe R, Gaynor JW, Spray TL, Durning SM, Mason SE, Montenegro LM, Nicolson SC, Zimmerman RA, Putt ME, Wang J, Greenberg JH, Detre JA, Yodh AG, Licht DJ. Optical measurement of cerebral hemodynamics and oxygen metabolism in neonates with congenital heart defects. Journal of biomedical optics. 2010;15(3):037004. doi: 10.1117/1.3425884. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Jain V, Buckley EM, Licht DJ, Lynch JM, Schwab PJ, Naim MY, Lavin NA, Nicolson SC, Montenegro LM, Yodh AG, Wehrli FW. Cerebral oxygen metabolism in neonates with congenital heart disease quantified by MRI and optics. J Cereb Blood Flow Metab. 2014;34(3):380–388. doi: 10.1038/jcbfm.2013.214. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Liu P, Lu H, Filbey FM, Tamminga CA, Cao Y, Adinoff B. MRI assessment of cerebral oxygen metabolism in cocaine-addicted individuals: hypoactivity and dose dependence. NMR in biomedicine. 2014;27(6):726–732. doi: 10.1002/nbm.3114. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Davis TL, Kwong KK, Weisskoff RM, Rosen BR. Calibrated functional MRI: mapping the dynamics of oxidative metabolism. Proceedings of the National Academy of Sciences of the United States of America. 1998;95(4):1834–1839. doi: 10.1073/pnas.95.4.1834. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Hoge RD, Atkinson J, Gill B, Crelier GR, Marrett S, Pike GB. Linear coupling between cerebral blood flow and oxygen consumption in activated human cortex. Proceedings of the National Academy of Sciences of the United States of America. 1999;96(16):9403–9408. doi: 10.1073/pnas.96.16.9403. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Lu H, Zhao C, Ge Y, Lewis-Amezcua K. Baseline blood oxygenation modulates response amplitude: Physiologic basis for intersubject variations in functional MRI signals. Magn Reson Med. 2008;60(2):364–372. doi: 10.1002/mrm.21686. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Liu P, Hebrank AC, Rodrigue KM, Kennedy KM, Park DC, Lu H. A comparison of physiologic modulators of fMRI signals. Human brain mapping. 2013;34(9):2078–2088. doi: 10.1002/hbm.22053. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Mintun MA, Raichle ME, Martin WR, Herscovitch P. Brain oxygen utilization measured with O-15 radiotracers and positron emission tomography. Journal of nuclear medicine: official publication, Society of Nuclear Medicine. 1984;25(2):177–187. [PubMed] [Google Scholar]
  • 17.Coles JP, Fryer TD, Bradley PG, Nortje J, Smielewski P, Rice K, Clark JC, Pickard JD, Menon DK. Intersubject variability and reproducibility of 15O PET studies. J Cereb Blood Flow Metab. 2006;26(1):48–57. doi: 10.1038/sj.jcbfm.9600179. [DOI] [PubMed] [Google Scholar]
  • 18.An H, Lin W, Celik A, Lee YZ. Quantitative measurements of cerebral metabolic rate of oxygen utilization using MRI: a volunteer study. NMR in biomedicine. 2001;14(7–8):441–447. doi: 10.1002/nbm.717. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.He X, Yablonskiy DA. Quantitative BOLD: mapping of human cerebral deoxygenated blood volume and oxygen extraction fraction: default state. Magn Reson Med. 2007;57(1):115–126. doi: 10.1002/mrm.21108. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Haacke EM, Lai S, Reichenbach JR, Kuppusamy K, Hoogenraad FG, Takeichi H, Lin W. In vivo measurement of blood oxygen saturation using magnetic resonance imaging: a direct validation of the blood oxygen level-dependent concept in functional brain imaging. Human brain mapping. 1997;5(5):341–346. doi: 10.1002/(SICI)1097-0193(1997)5:5<341::AID-HBM2>3.0.CO;2-3. [DOI] [PubMed] [Google Scholar]
  • 21.Lu H, Ge Y. Quantitative evaluation of oxygenation in venous vessels using T2-Relaxation-Under-Spin-Tagging MRI. Magn Reson Med. 2008;60(2):357–363. doi: 10.1002/mrm.21627. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Oja JM, Gillen JS, Kauppinen RA, Kraut M, van Zijl PC. Determination of oxygen extraction ratios by magnetic resonance imaging. J Cereb Blood Flow Metab. 1999;19(12):1289–1295. doi: 10.1097/00004647-199912000-00001. [DOI] [PubMed] [Google Scholar]
  • 23.Wright GA, Hu BS, Macovski A. Estimating oxygen saturation of blood in vivo with MR imaging at 1. 5 T. J Magn Reson Imaging. 1991;1(3):275–283. doi: 10.1002/jmri.1880010303. [DOI] [PubMed] [Google Scholar]
  • 24.Qin Q, Grgac K, van Zijl PC. Determination of whole-brain oxygen extraction fractions by fast measurement of blood T(2) in the jugular vein. Magn Reson Med. 2011;65(2):471–479. doi: 10.1002/mrm.22556. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Lu H, Xu F, Grgac K, Liu P, Qin Q, van Zijl P. Calibration and validation of TRUST MRI for the estimation of cerebral blood oxygenation. Magn Reson Med. 2012;67(1):42–49. doi: 10.1002/mrm.22970. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Xu F, Uh J, Brier MR, Hart J, Jr, Yezhuvath US, Gu H, Yang Y, Lu H. The influence of carbon dioxide on brain activity and metabolism in conscious humans. J Cereb Blood Flow Metab. 2011;31(1):58–67. doi: 10.1038/jcbfm.2010.153. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Xu F, Liu P, Pascual JM, Xiao G, Lu H. Effect of hypoxia and hyperoxia on cerebral blood flow, blood oxygenation, and oxidative metabolism. J Cereb Blood Flow Metab. 2012;32(10):1909–1918. doi: 10.1038/jcbfm.2012.93. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Xu F, Liu P, Pascual JM, Xiao G, Huang H, Lu H. Acute effect of glucose on cerebral blood flow, blood oxygenation, and oxidative metabolism. Human brain mapping. 2014 doi: 10.1002/hbm.22658. In press. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Xu F, Liu P, Pekar JJ, Lu H. Does caffeine ingestion alter brain metabolism?. Proceedings of the Joint Annual Meeting ISMRM-ESMRMB; Milan, Italy. 2014; p. 4168. [Google Scholar]
  • 30.Xu F, Uh J, Liu P, Lu H. On improving the speed and reliability of T2-relaxation-under-spin-tagging (TRUST) MRI. Magn Reson Med. 2012;68(1):198–204. doi: 10.1002/mrm.23207. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Liu P, Xu F, Lu H. Test-retest reproducibility of a rapid method to measure brain oxygen metabolism. Magn Reson Med. 2013;69(3):675–681. doi: 10.1002/mrm.24295. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.van Zijl PC, Eleff SM, Ulatowski JA, Oja JM, Ulug AM, Traystman RJ, Kauppinen RA. Quantitative assessment of blood flow, blood volume and blood oxygenation effects in functional magnetic resonance imaging. Nature medicine. 1998;4(2):159–167. doi: 10.1038/nm0298-159. [DOI] [PubMed] [Google Scholar]
  • 33.Lu H, Clingman C, Golay X, van Zijl PC. Determining the longitudinal relaxation time (T1) of blood at 3. 0 Tesla. Magn Reson Med. 2004;52(3):679–682. doi: 10.1002/mrm.20178. [DOI] [PubMed] [Google Scholar]
  • 34.Ito H, Kanno I, Kato C, Sasaki T, Ishii K, Ouchi Y, Iida A, Okazawa H, Hayashida K, Tsuyuguchi N, Ishii K, Kuwabara Y, Senda M. Database of normal human cerebral blood flow, cerebral blood volume, cerebral oxygen extraction fraction and cerebral metabolic rate of oxygen measured by positron emission tomography with 15O-labelled carbon dioxide or water, carbon monoxide and oxygen: a multicentre study in Japan. European journal of nuclear medicine and molecular imaging. 2004;31(5):635–643. doi: 10.1007/s00259-003-1430-8. [DOI] [PubMed] [Google Scholar]
  • 35.Pascual JM, Liu P, Mao D, Kelly DI, Hernandez A, Sheng M, Good LB, Ma Q, Marin-Valencia I, Zhang X, Park JY, Hynan LS, Stavinoha P, Roe CR, Lu H. Triheptanoin for Glucose Transporter Type I Deficiency (G1D): Modulation of Human Ictogenesis, Cerebral Metabolic Rate, and Cognitive Indices by a Food Supplement. JAMA neurology. 2014;71(10):1255–1265. doi: 10.1001/jamaneurol.2014.1584. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Bolar DS, Rosen BR, Sorensen AG, Adalsteinsson E. QUantitative Imaging of eXtraction of oxygen and TIssue consumption (QUIXOTIC) using venular-targeted velocity-selective spin labeling. Magn Reson Med. 2011;66(6):1550–1562. doi: 10.1002/mrm.22946. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Fan AP, Benner T, Bolar DS, Rosen BR, Adalsteinsson E. Phase-based regional oxygen metabolism (PROM) using MRI. Magn Reson Med. 2012;67(3):669–678. doi: 10.1002/mrm.23050. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Guo J, Wong EC. Venous oxygenation mapping using velocity-selective excitation and arterial nulling. Magn Reson Med. 2012;68(5):1458–1471. doi: 10.1002/mrm.24145. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Wise RG, Harris AD, Stone AJ, Murphy K. Measurement of OEF and absolute CMRO2: MRI-based methods using interleaved and combined hypercapnia and hyperoxia. NeuroImage. 2013;83:135–147. doi: 10.1016/j.neuroimage.2013.06.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Krishnamurthy LC, Liu P, Ge Y, Lu H. Vessel-specific quantification of blood oxygenation with T -relaxation-under-phase-contrast MRI. Magn Reson Med. 2013 doi: 10.1002/mrm.24750. In press. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Furlan M, Marchal G, Viader F, Derlon JM, Baron JC. Spontaneous neurological recovery after stroke and the fate of the ischemic penumbra. Annals of neurology. 1996;40(2):216–226. doi: 10.1002/ana.410400213. [DOI] [PubMed] [Google Scholar]
  • 42.Thomas BP, Sheng M, Tseng B, Liu P, Martin-Cook K, Cullum M, Weiner M, Levine B, Zhang R, Lu H. Characterization of CMRO2, resting CBF, and cerebrovascular reactivity in patients with very early stage of Alzheimer’s Disease. Proceedings of the 21st Annual Meeting of ISMRM; Salt Lake City, Utah, USA. 2014; p. 619. [Google Scholar]
  • 43.Alsop DC, Detre JA. Reduced transit-time sensitivity in noninvasive magnetic resonance imaging of human cerebral blood flow. J Cereb Blood Flow Metab. 1996;16(6):1236–1249. doi: 10.1097/00004647-199611000-00019. [DOI] [PubMed] [Google Scholar]
  • 44.Liu P, Uh J, Lu H. Determination of spin compartment in arterial spin labeling MRI. Magn Reson Med. 2011;65(1):120–127. doi: 10.1002/mrm.22601. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Chen Y, Wang DJ, Detre JA. Test-retest reliability of arterial spin labeling with common labeling strategies. J Magn Reson Imaging. 2011;33(4):940–949. doi: 10.1002/jmri.22345. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Wong EC, Buxton RB, Frank LR. A theoretical and experimental comparison of continuous and pulsed arterial spin labeling techniques for quantitative perfusion imaging. Magn Reson Med. 1998;40(3):348–355. doi: 10.1002/mrm.1910400303. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supp FigureS1

Supporting Fig. S1: Positioning instructions provided to the participating sites.

RESOURCES