Effects of temperature on Vitalograph spirometer readings.

W H Perks, T Sopwith, D Brown, C H Jones, M Green
PMCID: PMC459617  PMID: 6612650

Abstract

Spirometric variables were obtained from nine symptomless subjects, who performed forced expiratory manoeuvres with three Vitalograph spirometers at three different ambient temperatures (36.5 degrees C, 24.1 degrees C, and - 7.3 degrees C) over three days according to a Latin square design. Analysis of variance showed no significant difference between values at different ambient temperatures when measured at ATPS. Correction of the results to BTPS resulted in a significant difference in values at the three ambient temperatures for all measurements other than the Vmax25. The mean error introduced by conversion to BTPS varied from 7% (SD 5%) for the FVC at 24.1 degrees C to 30% (21%) for Vmax50 at - 7.3 degrees C. Possible explanations for these observations include the increasing compliance of the spirometer bellows with increasing temperature, relatively slow cooling of gases within the spirometer, and a combination of these effects. We conclude that there is no evidence that conversion to BTPS improves the accuracy of measurements made on a Vitalograph spirometer. Further studies on other spirometers are needed to see whether conversion to BTPS is always appropriate.

Full text

PDF
592

Images in this article

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. ABRAMOWITZ S., LEINER G. C., LEWIS W. A., SMALL M. J. VITAL CAPACITY IN THE NEGRO. Am Rev Respir Dis. 1965 Aug;92:287–292. doi: 10.1164/arrd.1965.92.2.287. [DOI] [PubMed] [Google Scholar]
  2. Gardner R. M., Hankinson J. L., West B. J. Evaluating commercially available spirometers. Am Rev Respir Dis. 1980 Jan;121(1):73–82. doi: 10.1164/arrd.1980.121.1.73. [DOI] [PubMed] [Google Scholar]
  3. Malik M. A., Moss E., Lee W. R. Prediction values for the ventilatory capacity in male West Pakistani workers in the United Kingdom. Thorax. 1972 Sep;27(5):611–619. doi: 10.1136/thx.27.5.611. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Miller G. J., Ashcroft M. T., Swan A. V., Beadnell H. M. Ethnic variation in forced expiratory volume and forced vital capacity of African and Indian adults in Guyana. Am Rev Respir Dis. 1970 Dec;102(6):979–981. doi: 10.1164/arrd.1970.102.6.979. [DOI] [PubMed] [Google Scholar]
  5. Williams D. E., Miller R. D., Taylor W. F. Pulmonary function studies in healthy Pakistani adults. Thorax. 1978 Apr;33(2):243–249. doi: 10.1136/thx.33.2.243. [DOI] [PMC free article] [PubMed] [Google Scholar]