Evaluation of the Ohmeda 3700 pulse oximeter.

A R Warley 1, J H Mitchell 1, J R Stradling 1
PMCID: PMC461017  PMID: 3424271

Abstract

Arterial oxygen saturation values (Sao2) from 60% to 98% were measured by the Ohmeda 3700 pulse oximeter with the three types of probe available and compared with values of oxygen saturation estimated from direct arterial sampling (arterial oxygen and carbon dioxide tensions and pH) on 65 occasions. The response time of the oximeter was measured after a sudden rise in inspired oxygen concentration. Artefact rejection was assessed by arterial compression proximal to the probe site, and by simultaneous recordings of overnight Sao2 on opposite hands. The ability to recreate patterns of oscillating Sao2 from the data stored in the oximeter was also investigated. With the best probe system the oximeter measured Sao2, relative to arterial values estimated from Pao2, with a mean (SD) difference of -0.4% (1.8%). The response time was comparable with those of previous oximeters. It was not possible to generate artefactual dips in excess of 2% Sao2, and the dual overnight recordings rarely showed even small dips on one tracing alone. The stored data can recreate oscillating Sao2 signals with wavelengths down to about 35 seconds, but not below. The Ohmeda 3700 pulse oximeter appears to be suitable for unattended overnight recordings of Sao2.

Full text

PDF
892

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Chapman K. R., D'Urzo A., Rebuck A. S. The accuracy and response characteristics of a simplified ear oximeter. Chest. 1983 Jun;83(6):860–864. doi: 10.1378/chest.83.6.860. [DOI] [PubMed] [Google Scholar]
  2. Chapman K. R., Liu F. L., Watson R. M., Rebuck A. S. Range of accuracy of two wavelength oximetry. Chest. 1986 Apr;89(4):540–542. doi: 10.1378/chest.89.4.540. [DOI] [PubMed] [Google Scholar]
  3. Douglas N. J., Brash H. M., Wraith P. K., Calverley P. M., Leggett R. J., McElderry L., Flenley D. C. Accuracy sensitivity to carboxyhemoglobin, and speed of response of the Hewlett-Packard 47201A ear oximeter. Am Rev Respir Dis. 1979 Feb;119(2):311–313. doi: 10.1164/arrd.1979.119.2.311. [DOI] [PubMed] [Google Scholar]
  4. Rebuck A. S., Campbell E. J. A clinical method for assessing the ventilatory response to hypoxia. Am Rev Respir Dis. 1974 Mar;109(3):345–350. doi: 10.1164/arrd.1974.109.3.345. [DOI] [PubMed] [Google Scholar]
  5. Tweeddale P. M., Douglas N. J. Evaluation of Biox IIA ear oximeter. Thorax. 1985 Nov;40(11):825–827. doi: 10.1136/thx.40.11.825. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. Von Békésy G. Taste theories and the chemical stimulation of single papillae. J Appl Physiol. 1966 Jan;21(1):1–9. doi: 10.1152/jappl.1966.21.1.1. [DOI] [PubMed] [Google Scholar]

Articles from Thorax are provided here courtesy of BMJ Publishing Group

RESOURCES