Quality Assurance System Using Statistical Process Control: An Implementation for Image Cytometry

David Chiu 1,6, Martial Guillaud 1, Dennis Cox 2, Michele Follen 3,4,5,*, Calum MacAulay 1
PMCID: PMC4611122  PMID: 15371646

Abstract

Aims: Optical technologies have shown some promise for improving the care of cervical neoplasia. We are currently evaluating fluorescence and reflectance spectroscopy and quantitative cyto‐histopathology for cervical neoplasia screening and diagnosis. Here we describe the establishment and application of a quality assurance (QA) system for detecting system malfunctions and assessing the comparability of four image cytometers used in a multicenter clinical trial. Methods: Our QA system involves three levels of evaluation based on the periodicity and complexity of the measurements. We implemented our QA system at three image cytometers at the British Columbia Cancer Agency and one at M.D. Anderson Cancer Center. The measurements or tasks were performed daily, monthly, and semi‐annually. The current and voltage of the lamp, the calibration image characteristics, and the room temperature were checked daily. Long‐term stability over time, short‐term variability over time, and spatial response field uniformity were evaluated monthly. Camera linearity was measured semi‐annually. Control charts based on statistical process control techniques were used to detect when the system did not perform optimally. Results: Daily measurements have shown good consistency in room temperature, lamp and calibration behaviour. Monthly measurements have shown small coefficients of variation between and within the four devices. There have been greater differences between sessions than within sessions. Comparability among the four systems is reasonably good. Semi‐annual measurements have shown stable camera linearity. QA events were detected using the QA system. Multiple examples of event detection leading to correction of system malfunction are described in this report. Conclusions: QA programs are critical for ensuring data integrity and therefore for the conduct of multicenter clinical trials.

Keywords: Image cytometry, quality assurance, quality control, statistical process control, histopathology, quantitative, cytology, quantitative, technology assessment, cervical intraepithelial neoplasia, lesions, squamous intraepithelial


Articles from Cellular Oncology : the Official Journal of the International Society for Cellular Oncology are provided here courtesy of Wiley

RESOURCES