Skip to main content
PLOS ONE logoLink to PLOS ONE
. 2015 Oct 20;10(10):e0139707. doi: 10.1371/journal.pone.0139707

Pulse Oximetry: A Non-Invasive, Novel Marker for the Quality of Chest Compressions in Porcine Models of Cardiac Arrest

Jun Xu 1,#, Chen Li 2,#, Liangliang Zheng 3,#, Fei Han 4, Yan Li 1, Joseph Walline 5, Yangyang Fu 1, Dongqi Yao 1, Xiaocui Zhang 4, Hui Zhang 1, Huadong Zhu 1, Shubin Guo 1, Zhong Wang 6, Xuezhong Yu 1,*
Editor: Chiara Lazzeri7
PMCID: PMC4613139  PMID: 26485651

Abstract

Objective

Pulse oximetry, which noninvasively detects the blood flow of peripheral tissue, has achieved widespread clinical use. We have noticed that the better the quality of cardiopulmonary resuscitation (CPR), the better the appearance of pulse oximetry plethysmographic waveform (POP). We investigated whether the area under the curve (AUC) and/or the amplitude (Amp) of POP could be used to monitor the quality of CPR.

Design

Prospective, randomized controlled study.

Setting

Animal experimental center in Peking Union Medical Collage Hospital, Beijing, China.

Subjects

Healthy 3-month-old male domestic swine.

Interventions

34 local pigs were enrolled in this study. After 4 minutes of untreated ventricular fibrillation, animals were randomly assigned into two resuscitation groups: a “low quality” group (with a compression depth of 3cm) and a “high quality” group (with a depth of 5cm). All treatments between the two groups were identical except for the depth of chest compressions. Hemodynamic parameters [coronary perfusion pressure (CPP), partial pressure of end-tidal carbon dioxide (PETCO2)] as well as AUC and Amp of POP were all collected and analyzed.

Measurements and Findings

There were statistical differences between the “high quality” group and the “low quality” group in AUC, Amp, CPP and PETCO2 during CPR (P<0.05). AUC, Amp and CPP were positively correlated with PETCO2, respectively (P<0.01). There was no statistical difference between the heart rate calculated according to the POP (FCPR) and the frequency of mechanical CPR at the 3rd minute of CPR. The F CPR was lower than the frequency of mechanical CPR at the 6th and the 9th minute of CPR.

Conclusions

Both the AUC and Amp of POP correlated well with CPP and PETCO2 in animal models. The frequency of POP closely matched the CPR heart rate. AUC and Amp of POP might be potential noninvasive quality monitoring markers for CPR.

Introduction

Both the American Heart Association (AHA) and the European Resuscitation Council (ERC) 2010 guidelines for cardiopulmonary resuscitation (CPR) emphasize the need for high quality CPR, including compressions to a depth of at least 5cm and a rate of at least 100 min-1[1, 2]. Such “high-quality” CPR is currently felt to be the best way to achieve a successful return of spontaneous circulation (ROSC), and help patients survive cardiac arrest[1]. An inexpensive, noninvasive and quantitative hemodynamic monitor that can help rescuers adjust their performance to maximize artificial circulation to patients suffering cardiac arrest is urgently needed. Quantitative waveform capnography and invasive arterial pressure monitoring are recommended for physiologic monitoring of the effectiveness of chest compressions in the 2010 guidelines[3]. However, both of these methods are expensive, invasive or depending on advanced airway, and are not available in most CPR situations[4].

Pulse oximetry, which is noninvasive and inexpensive, is widely used in most clinical situations to monitor the oxygenation of peripheral blood. Several types of probes are commonly used, such as a finger clip, nasal clip, or forehead sticker. The operating principles of which are similar – a light detector and a light source with two waves which differ in their absorbance by oxyhemoglobin and reduced hemoglobin. Most of the light is absorbed by the fixed tissue such as the skin, muscle, bone and venous blood in a constant amount. The absorbance of the arterial blood varies with each beat, and this variance is transformed into an electric signal which is then transferred from the detector as a periodical waveform with each beat, namely the pulse oximetry plethysmographic waveform (POP) that reflects the peripheral tissue perfusion[57]. In patients suffering cardiac arrest, no blood flow is detected in the peripheral tissue, and, as a result, no waveform can be transported (or detected).

In our clinical practice, we found that when CPR was performed, POP regularly showed with each compression, and we also noticed that the “better” (i.e. deeper) the compression was performed, the “better” (i.e. larger) the waveform seemed. Recently, more and more studies had shown that POP could be used as an early indicator of changes in circulation[5, 8, 9]. Thus we wondered if it could be feasible to seek circulatory information from the POP during CPR.

We hypothesized that POP could reliably reflect the quality of the CPR beat by beat—both the frequency of chest compressions and their hemodynamic effect. In this study, we analyzed differences in coronary perfusion pressure (CPP), partial pressure of end-tidal carbon dioxide (PETCO2), and the area under the curve (AUC) and amplitude (Amp) of POP between two different quality CPR programs and studied the relationship between the POP values, CPP and PETCO2 to get some information about whether these parameters could be used to assess the quality of CPR. Finally, we analyzed the relationship between the frequency of POP and chest compressions to determine whether the frequency of POP could accurately reflect the frequency of chest compressions.

Materials and Methods

Preparation

This experimental protocol was approved by the Animal Care and Use Committee at Peking Union Medical College Hospital. A total of 34 domestic pigs (weight, 28±2 kg) received Pentobarbital Sodium 30 mg/kg IM (Merck, 719F034, Germany) followed by inhalational 4% isoflurane (ABBOTT, H20059911, USA) via a snout mask with 100% oxygen via an anesthesia apparatus (Veterinary Anesthesia Ventilator, Midmark Corporation, USA). After endotracheal intubation and mechanical ventilation initiation, anesthesia was maintained with intravenous propofol (Corden Pharma S.P.A., H20100645, Italy) (5~10mg/min). Each pig was mechanically ventilated in volume control mode without PEEP (VT = 8~10ml/kg, Rate = 12/min) (Esprit Ventilato, V1000, Germany), and the tidal volume was adjusted to maintain PETCO2 at approximately 35~45 mmHg. An arterial blood gas sample was analyzed to verify the baseline conditions.

Pigs were placed in the supine position and fastened in a special “U” shape frame to avoid the pigs moving during the chest compressions. Right femoral arteries were cannulated with 6 Fr single-lumen arterial catheters with their distal tips in the thoracic aorta to monitor blood pressure. Central venous catheters (18Ga, 20 cm, Arrow, USA) were inserted from the right internal jugular veins to the right atria for right atrial pressure (RAP) monitoring. Another catheter was inserted into the left external jugular vein to place an electrode catheter in order to induce VF. Electrocardiogram (ECG) and PETCO2 (Capnostat 5, Respironics, CT, USA) were continuously monitored. A pulse oximetry sensor (Mindray Biological Medical Electronic Co, Ltd, Shenzhen, China) was fixed around the tails of the models to obtain the POP. The ECG, invasive arterial blood pressure (ABP), RAP, PETCO2 and POP were monitored and recorded by a T8 Mindray monitor (Mindray Biological Medical Electronic Co, Ltd, Shenzhen, China).

Protocol Description

After baseline data was collected, a pacing electrode was positioned in the right ventricle. Ventricular fibrillation (VF) was induced by 24V/50Hz AC current applied to the right ventricular endocardium for 1 second. Once VF was confirmed by the ECG waveform and a precipitous decline in aortic pressure, the ventilator was disconnected from the endotracheal tube temporarily. After 4 minutes of untreated VF (mimicking a bystander recognizing cardiac arrest and calling for help), resuscitation was started. Once chest compressions were started, the ventilator was reconnected to the endotracheal tube and ventilated to the original parameters. All the animals were randomly allocated into two groups using random number table when grouping the animals (Fig 1): (1) a low quality resuscitation group (LQ group)–which was provided continuous chest compressions to a depth of 3cm at a rate of 105 compressions per minute; (2) a high quality resuscitation group (HQ group)–which was provided continuous chest compressions to a depth of 5cm at a rate of 105 compressions per minute. Each group had 17 pigs and received different depth chest-compressions via a mechanical CPR device (WISH-SL-FS-A, Wuhan, China). During the resuscitation, hemodynamic parameters, as well as PETCO2 and POP were continuously monitored. 10 minutes after the end of data collection, the animals were euthanized with potassium chloride. To avoid the influence of resuscitation drugs (such as epinephrine, vasopression, etc.), no other medication was given and all treatment between the two groups was the same except for the depth of chest compressions.

Fig 1. Experimental Protocol.

Fig 1

Data Acquisition

The following parameters were monitored: heart rate (HR), systolic arterial pressure (SAP), diastolic arterial pressure (DAP), mean arterial pressure (MAP), right atrial pressure (RAP) and PETCO2. CPP during CPR was calculated by subtracting right atrial relaxation (“diastolic”) pressure from aortic relaxation (“diastolic”) pressure[1012].

The AUC, Amp and frequency of POP were acquired by the T8 monitor and stored in Compact Flash cards. The data was analyzed by Mindray POP viewer V8.0 (Mindray Research Center for Monitoring and Life Support) using MATLAB software V7.10.0 (MathWorks, Natick, Massachusetts, USA). In order to reduce the influence of noise, the pulse oximetry signal was limited to a frequency of 11 Hz and only the near-infrared signal was analyzed. Pulse oximetry signals include the fluctuant component (S AC) and the constant component (S DC). Since the S AC is related to pulsating blood volume, and the S DC is related to non-pulsating tissues such as muscle and bone, we focused on the fluctuant component S AC by filtering out the constant component S DC in the original signals with FIR/IIR filtering technology. All the POP parameters were calculated based on the fluctuant component S AC.

The frequency of mechanical chest compressions was compared to the frequency of the fluctuant component (f SAC) waveform monitored by POP (F CPR).

In order to get the stable value of the amplitude, the mean square root method was used to extract the absolute amplitude value (Amp CPR) of each single pulse wave in the fluctuant component. The formula was as follows:

AmpCPR=n=0N1SAC2(n)N

A point-by-point integral method was used to calculate the absolute area parameter. The formula was as follows:

AreaCPR=n0N1SAC(n)

[Note: S AC (n) represented the nth sampling data point of a single pulse wave, and N represented the total sampling point count of a single pulse wave. Amp CPR represented the absolute amplitude value of a single pulse wave.]

Statistical Analysis

Statistical analysis was performed with SPSS 16.0 for Windows (SPSS, Inc.). The normality of the distribution of each variable was assessed using the Kolmogorov-Smirnov test. The One-Sample test was used to assess the frequency variation in samples compared to the theoretical value. F-test and Student’s T-test were applied to the normal distribution data and Mann-Whitney U test was applied to the non-normal distribution data to determine differences between the two groups. P value < 0.05 was considered to be statistically significant.

We used the Pearson correlation coefficient for linear correlation analysis. P value < 0.01 was considered to be statistically significant according to Bonferroni correction.

Normal distribution data was shown as mean±SD and non-normal distribution data was shown as median (25%-75%).

Results

In HQ group, there was an animal with extreme values. We removed that data to reduce error.

No ROSC was achieved during the initial 9 minutes of CPR. Hemodynamic parameters, PETCO2, and the POP parameters were similar in the two groups at baseline, as shown in Table 1. However, when VF was induced, all the parameters changed not only during VF but also during chest compressions.

Table 1. Physiological parameters at baseline.

Group HR (bpm) CPP (mmHg) PETCO2 (mmHg) Amp (PVA) AUC (PVPG)
LQ group (n = 17) 95±26 102±25 40±3 477±264 4268±563
HQ group (n = 16) 96±22 102±24 39±3 445±300 4138±590

Note: Amp and AUC were calculated as the average of the data collected 30s before the detect time.

No statically significant differences were detected in any parameters between the two groups.

LQ group: low quality resuscitation group; HQ group: high quality resuscitation group; HR: heart rate; CPP: coronary perfusion pressure; PETCO2: pressure of end-tidal carbon dioxide; AUC: area of under the curve; Amp: amplitude of POP; PVA: Pulse Oximeter Voltage Amplitude; PVPG: Pulse Oximeter Voltage Plehtysmography.

When VF was induced, both the regular aortic blood pressure waveform and the POP disappeared, as shown in Fig 2. Correspondingly, the CPP, HR, AUC and Amp of POP all declined to zero suddenly. Similarly, PETCO2 also decreased to zero after 3 minutes of untreated VF.

Fig 2. The arterial waveform and POP during different stages of CPR A.

Fig 2

The arterial waveform and POP during spontaneous circulation at baseline; B. Both the arterial waveform and POP disappeared during VF; C. The arterial waveform and POP during low quality resuscitation; D. The arterial waveform and POP during high quality resuscitation.

Not surprisingly, when resuscitation began, both the regular aortic blood pressure waveform and the POP appeared, as shown in Fig 2. All the hemodynamic parameters, PETCO2, and the POP parameters increased significantly compared to during VF (P<0.05) in both groups, but all the parameters were lower than those of the baseline before VF (P<0.05) (Table 2).

Table 2. Physiological parameters during resuscitation.

LQ group (n = 17) HQ group (n = 16) t/U P
HR (bpm) 3min 104±6 104±5 0.251 0.803
6min 103±3 103±3 0.245 0.808
9min 103±3 103±2 0.532 0.599
3min 12±4 19±4 -4.830 <0.001
PETCO2 (mmHg) 6min 13±4 19±4 -3.948 <0.001
9min 12±4 18±4 -4.690 <0.001
3min 14.6±9.8 25.0±17.5 -2.146 0.040
3min 14±10 21±5 -2.301 0.028
CPP (mmHg) 6min 15±10 21±5 -2.221 0.034
9min 12±7 23±6 -4.872 <0.001
3min 70±62 189±129 -3.317 0.003
Amp (PVA) 6min 71±63 194±132 -3.379 0.003
9min 79±81 188±119 -3.081 0.005
3min 2215±852 3191±556 -3.872 0.001
AUC (PVPG) 6min 2211±781 3193±517 -4.231 <0.001
9min 2022±665 3067±522 -5.001 <0.001

Note: Amp and AUC were calculated as the average of the data collected 30s before the detect time.

PVA: Pulse Oximeter Votage Amplitude; PVPG: Pulse Oximeter Voltage Plehtysmography

There were statistically significant differences in CPP, PETCO2, AUC and Amp between the high quality group and the low quality group at the 3rd, 6th and 9th minutes of chest compressions. When the quality of chest compressions increased, these parameters increased correspondingly (P<0.05) (Table 2).

AUC was not correlated with CPP in this study (r = 0.348, 0.281 and 0.396, P = 0.047, 0.114, and 0.023 at the 3rd, 6th and 9th minute respectively) (Fig 3A). Amp was positively correlated with CPP at the 3rd minute (r = 0.450, P = 0.009) but no correlation was detected at the 6th minute (r = 0.378, P = 0.030) and 9th minute (r = 0.441, P = 0.010) (Fig 3B). CPP was positively correlated with PETCO2 at the 3th minute (r = 0.543, P = 0.001) and 9th minute (r = 0.728, P<0.001), but no correlation was detected at the 6rd minute (r = 0.429, P = 0.013) (Fig 3C). There was positive correlation between AUC and PETCO2 (r = 0.725, 0.648, and 0.644 at the 3rd, 6th and 9th minute respectively, P<0.001) (Fig 4A), the same between Amp and PETCO2 (r = 0.529, 0.493 and 0.480, P = 0.002, 0.004, and 0.005 at the 3rd, 6th and 9th minute respectively) (Fig 4B).

Fig 3. A. The correlation between AUC and CPP during different stages of CPR a. AUC was not correlated with CPP at the 3rd minute of CPR (r = 0.348, P = 0.047). b. At the 6th minute of CPR, AUC was not correlated with CPP (r = 0.281, P = 0.114). c. A correlation couldn’t be detected between AUC and CPP at the 9th minute of CPR (r = 0.396, P = 0.023). B. The correlation between Amp and CPP during different stages of CPR a. Amp was positively correlated with CPP at the 3rd minute of CPR (r = 0.450, P = 0.009). b. At the 6th minute of CPR, Amp was not correlated with CPP (r = 0.378, P = 0.030). c. A positive correlation could be detected between Amp and CPP at the 9th minute of CPR (r = 0.441, P = 0.010). C. The correlation between PETCO2 and CPP during different stages of CPR a. PETCO2 was positively correlated with CPP at the 3rd minute of CPR (r = 0.543, P = 0.001). b. At the 6th minute of CPR, PETCO2 was not correlated with CPP (r = 0.429, P = 0.013). c. PETCO2 was positively correlated with CPP at the 9th minute of CPR (r = 0.728, P<0.001).

Fig 3

Fig 4. A.The correlation between AUC and PETCO2 during different stages of CPR a. AUC was positively correlated with PETCO2 at the 3rd minute of CPR (r = 0.725, P<0.001). b. At the 6th minute of CPR, AUC was positively correlated with PETCO2 (r = 0.648, P<0.001). c. AUC was positively correlated with PETCO2 at the 9th minute of CPR (r = 0.644, P<0.001). B. The correlation between Amp and PETCO2 during different stages of CPR a. Amp was positively correlated with PETCO2 at the 3rd minute of CPR (r = 0.529, P = 0.002). b. At the 6th minute of CPR, Amp was positively correlated with PETCO2 (r = 0.493, P = 0.004). c. At the 9th minute of CPR, Amp was positively correlated with PETCO2 (r = 0.480, P = 0.005).

Fig 4

During resuscitation, the F CPR seen on POP was compared to the frequency of mechanical CPR (105 compressions per min) in both groups regardless of compression depth. There was no statistical difference between the F CPR and the frequency of mechanical CPR at the 3th minute of CPR (104±5, P = 0.209). The F CPR was lower than the frequency of mechanical CPR at the 6th (103±3, P<0.001) and the 9th (103±3, P<0.001) minute of CPR. The POP reflected the heart beat or effective chest compressions beat-by-beat, as shown in Fig 2.

Discussion

High quality chest compressions are the cornerstone of cardiopulmonary resuscitation. When rescuers compress to a depth of <38 mm, survival-to-discharge rates after out-of-hospital arrest are reduced by 30%[13]. Similarly, when rescuers compress too slowly, ROSC after in-hospital cardiac arrest falls from 72% to 42%[14]. But the actual quality of chest compressions including compression depth and frequency during both outside-hospital and in-hospital CPR is frequently sub-optimal. The 2010 AHA Guidelines for CPR strongly emphasized that delivering high quality chest compressions is essential, which includes pushing to a depth of at least 5 cm at a rate of at least 100 compressions per minute, allowing full chest recoil, and minimizing interruptions in chest compressions[1]. In order to improve the quality of CPR further, a consensus statement “CPR Quality: Improving Cardiac Resuscitation Outcomes Both Inside and Outside the Hospital” from the American Heart Association was released again in June 2013[15]. Real-time CPR monitoring and feedback to rescuers is critical.

As far as we know, there are two types of CPR quality monitoring systems: physiological monitors (for monitoring hemodynamic effects) and CPR performance metrics (for monitoring the work of the rescuers), both of which can provide real-time feedback. CPP, a surrogate for myocardial perfusion, defined as the aortic diastolic pressure minus the right atrial diastolic pressure, is an example of one the first types. Higher CPP is associated with and predictive of ROSC during CPR both in cardiac arrest patients and in animal models[1619]. Although CPP is the gold standard for monitoring CPR quality, the use of CPP is still limited because measuring both the aortic diastolic pressure and the right atrial diastolic pressure are invasive and time-consuming, and rarely feasible in resuscitation situations. Pulse oximetry, which reflects the peripheral tissue perfusion status, is similar in appearance to an arterial blood pressure waveform, but what makes it different is that it is noninvasive, inexpensive and nearly ubiquitous in hospitals. As a result, the extraction of circulatory information from POP has been a popular subject of contemporary research.

Anecdotal clinical evidence told us that the higher the quality of CPR was performed, the better the perfusion of peripheral tissue was, and, as a result, the better the POP waveform appeared[20]. In this study, the CPP, AUC and Amp of POP dropped to zero in VF and increased when CPR was performed, even though it was still lower than the initial stage. It was worth mentioning that all these parameters were statistically higher in the high-quality CPR group than in the low quality CPR group. Also, the relationship of AUC and Amp with CPP was analyzed and not exactly correlation between the two parameters was detected, which because CPP monitoring the coronary perfusion, while AUC and Amp of POP the peripheral one. All these results revealed the fact that not only CPP, but also AUC and Amp might be used as a potential marker of high quality CPR.

The 2010 AHA Guidelines for CPR strongly recommended to monitor the quality of the CPR and identify ROSC, and indeed PETCO2 has more clinical operability than CPP. Although the optimal target for PETCO2 during CPR had not been established[21], a value of less than 10mmHg had been associated with a failure to achieve ROSC and might indicate that the quality of chest compressions should be improved[22]. Studies also showed that PETCO2 directly correlated with cardiac output in the cardiac arrest patient[23]. Here, we also found that PETCO2 positively correlated with CPP which was consistent with what had been previously reported[24]. In this study, the AUC and Amp of POP were positively correlated with PETCO2, which also suggested that the AUC and Amp were a potential clinical way to monitor the quality of CPR. In clinical practice, capnography also had some limitations: it needed advanced airway and calibration which were time-consuming, and it could be easily affected by airway secreta and drugs such as vasoactive agents and sodium bicarbonate administration[25, 26]. Parameters of POP could be used as additive information to PETCO2 in the quality monitoring of CPR. And these parameters could be used earlier than PETCO2 to monitor the quality of CPR before the construction of the advanced airway.

The impact of chest compressions might influence the stability of the Amp of POP but not the AUC, which reflected the perfusion of peripheral tissue of the whole CPR process instead of the peak point of the process. This result suggested that the AUC might be a more stable parameter than the Amp of POP even though both of them could be used to monitor the quality of CPR non-invasively.

As for the second CPR metric (performance), there were some devices available, such as Q-CPR (HeartStart MRx, Philips). These kinds of devices provided real-time feedback of sternal force and displacement, and an audible signal was produced as an aid to maintaining the desired compression rate and duration, and, as a result, were capable of measuring, recording, and displaying multiple physical quantities during CPR[27]. However, these devices did not correct for any compression of an underlying mattress which might lead to significant under-compression of the chest during CPR[28, 29]. As a system for monitoring the pulsatile volume of blood in tissue instead of measuring the physical movement mechanically, the POP could reflect the quality of CPR without worrying about any such direct effects of the location (e.g. mattress thickness).

Besides the depth of chest compressions, minimizing interruptions and maintaining a high compression rate (100-120bpm) were two other recommended components of high-quality CPR[30]. In this study, pauses in effective compressions showed-up as the absence of waveforms. And the compression rate calculated according to the POP was very similar to the rate of the mechanical compression device. The rate seen on POP decreased over time during CPR, which showed that not all effective peripheral hemoperfusion could be made by chest compression when CPR lasting. So the rate of waveforms could reflect the rate of effective compressions. Therefore, the POP could also be used to monitor the frequency and interruption of CPR in real time, and as a result, the chest compression fraction of CPR could be calculated.

Even though the present findings about POP were interesting, there were also several limitations. First, pulse oximetry itself had some environmental limitations. Strong lights in the room, nail polish, movement of the testing part and other factors could affect the accuracy of the test. Also, the calculation of AUC and Amp of POP was not real-time and needed to be analyzed via separate computer software. However, real-time monitoring software of POP is under development. Third, the perfusion of the peripheral tissue which could affect the parameters of POP could be affected by the neural and humoral factors. However, when cardiac arrest happened, all the regulatory mechanisms were impaired and weakened, and as a result, the perfusion of the peripheral tissue of all patients might go to a trend of homogeneity and could be only affected by the chest compression.

Conclusions

In summary, the AUC and Amp of POP increased during CPR in cardiac arrested animal models and were higher in the high-quality CPR group than in the low-quality group. They correlated well with existing CPR quality monitoring parameters such as CPP and PETCO2, and the AUC was much more stable than the Amp. The frequency of POP reflected the performed chest compressions. AUC and Amp of POP might be used as potential noninvasive quality monitoring parameters for CPR.

Supporting Information

S1 Minimal Data Set. This file includes all the data in the Manuscript.

Line 29, ID 27, marked red, was removed because of the extreme values.

(XLS)

Data Availability

All relevant data are within the paper and its Supporting Information files.

Funding Statement

This study received funding from the Capital Health Research and Development Fund (2011-4001-04), http://www.bjhbkj.com/, JX received this funding; the National Health and Family Planning Commission of the People's Republic of China Special Research Fund (201502019), http://www.moh.gov.cn/qjjys/s3577/201401/e9f3635e7acb47778225ccb729ffec62.shtml, XY received this funding. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript. The Institute of Life Monitoring, Mindray Corporation provided support in the form of salaries for authors FH and XZ, but did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript. The specific roles of these authors are articulated in the "author contributions" section.

References

  • 1. Field JM, Hazinski MF, Sayre MR, Chameides L, Schexnayder SM, Hemphill R, et al. Part 1: Executive Summary 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2010;122(18):S640–S56. 10.1161/Circulationaha.110.970889 . [DOI] [PubMed] [Google Scholar]
  • 2. Nolan JP, Soar J, Zideman DA, Biarent D, Bossaert LL, Deakin C, et al. European Resuscitation Council Guidelines for Resuscitation 2010 Section 1. Executive summary. Resuscitation. 2010;81(10):1219–76. Epub 2010/10/20. 10.1016/j.resuscitation.2010.08.021 . [DOI] [PubMed] [Google Scholar]
  • 3. Field JM, Hazinski MF, Sayre MR, Chameides L, Schexnayder SM, Hemphill R, et al. Part 1: executive summary: 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2010;122(18 Suppl 3):S640–56. 10.1161/CIRCULATIONAHA.110.970889 . [DOI] [PubMed] [Google Scholar]
  • 4. Deakin CD, Nolan JP, Soar J, Sunde K, Koster RW, Smith GB, et al. European Resuscitation Council Guidelines for Resuscitation 2010 Section 4. Adult advanced life support. Resuscitation. 2010;81(10):1305–52. 10.1016/j.resuscitation.2010.08.017 . [DOI] [PubMed] [Google Scholar]
  • 5. Arntz HR, Bossaert LL, Danchin N, Nikolaou NI. European Resuscitation Council Guidelines for Resuscitation 2010 Section 5. Initial management of acute coronary syndromes. Resuscitation. 2010;81(10):1353–63. 10.1016/j.resuscitation.2010.08.016 . [DOI] [PubMed] [Google Scholar]
  • 6. Biarent D, Bingham R, Eich C, Lopez-Herce J, Maconochie I, Rodriguez-Nunez A, et al. European Resuscitation Council Guidelines for Resuscitation 2010 Section 6. Paediatric life support. Resuscitation. 2010;81(10):1364–88. 10.1016/j.resuscitation.2010.08.012 . [DOI] [PubMed] [Google Scholar]
  • 7. Richmond S, Wyllie J. European Resuscitation Council Guidelines for Resuscitation 2010 Section 7. Resuscitation of babies at birth. Resuscitation. 2010;81(10):1389–99. 10.1016/j.resuscitation.2010.08.018 . [DOI] [PubMed] [Google Scholar]
  • 8. Reisner A, Shaltis PA, McCombie D, Asada HH. Utility of the photoplethysmogram in circulatory monitoring. Anesthesiology. 2008;108(5):950–8. Epub 2008/04/24. 10.1097/ALN.0b013e31816c89e1 . [DOI] [PubMed] [Google Scholar]
  • 9. Pizov R, Eden A, Bystritski D, Kalina E, Tamir A, Gelman S. Arterial and plethysmographic waveform analysis in anesthetized patients with hypovolemia. Anesthesiology. 2010;113(1):83–91. Epub 2010/06/08. 10.1097/ALN.0b013e3181da839f . [DOI] [PubMed] [Google Scholar]
  • 10. Soar J, Perkins GD, Abbas G, Alfonzo A, Barelli A, Bierens JJ, et al. European Resuscitation Council Guidelines for Resuscitation 2010 Section 8. Cardiac arrest in special circumstances: Electrolyte abnormalities, poisoning, drowning, accidental hypothermia, hyperthermia, asthma, anaphylaxis, cardiac surgery, trauma, pregnancy, electrocution. Resuscitation. 2010;81(10):1400–33. 10.1016/j.resuscitation.2010.08.015 . [DOI] [PubMed] [Google Scholar]
  • 11. Paradis NA, Martin GB, Rivers EP, Goetting MG, Appleton TJ, Feingold M, et al. Coronary perfusion pressure and the return of spontaneous circulation in human cardiopulmonary resuscitation. JAMA. 1990;263(8):1106–13. . [PubMed] [Google Scholar]
  • 12. Reynolds JC, Salcido DD, Menegazzi JJ. Coronary perfusion pressure and return of spontaneous circulation after prolonged cardiac arrest. Prehospital emergency care: official journal of the National Association of EMS Physicians and the National Association of State EMS Directors. 2010;14(1):78–84. 10.3109/10903120903349796 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Soar J, Monsieurs KG, Ballance JH, Barelli A, Biarent D, Greif R, et al. European Resuscitation Council Guidelines for Resuscitation 2010 Section 9. Principles of education in resuscitation. Resuscitation. 2010;81(10):1434–44. 10.1016/j.resuscitation.2010.08.014 . [DOI] [PubMed] [Google Scholar]
  • 14. Lippert FK, Raffay V, Georgiou M, Steen PA, Bossaert L. European Resuscitation Council Guidelines for Resuscitation 2010 Section 10. The ethics of resuscitation and end-of-life decisions. Resuscitation. 2010;81(10):1445–51. 10.1016/j.resuscitation.2010.08.013 . [DOI] [PubMed] [Google Scholar]
  • 15. Soar J, Edelson DP, Perkins GD. Delivering high-quality cardiopulmonary resuscitation in-hospital. Curr Opin Crit Care. 2011;17(3):225–30. 10.1097/MCC.0b013e3283468b5c . [DOI] [PubMed] [Google Scholar]
  • 16. Edelson DP, Yuen TC, Mancini ME, Davis DP, Hunt EA, Miller JA, et al. Hospital cardiac arrest resuscitation practice in the United States: A nationally representative survey. Journal of hospital medicine: an official publication of the Society of Hospital Medicine. 2014;9(6):353–7. 10.1002/jhm.2174 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. Reynolds JC, Salcido DD, Menegazzi JJ. Coronary Perfusion Pressure and Return of Spontaneous Circulation after Prolonged Cardiac Arrest. Prehospital Emergency Care. 2010;14(1):78–84. 10.3109/10903120903349796 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Sawyer KN, Kurz MC. Caution when defining prolonged downtime in out of hospital cardiac arrest as extracorporeal cardiopulmonary resuscitation becomes accessible and feasible. Resuscitation. 2014. 10.1016/j.resuscitation.2014.05.018 . [DOI] [PubMed] [Google Scholar]
  • 19. Hutchens MP, Kosaka Y, Zhang W, Fujiyoshi T, Murphy S, Alkayed N, et al. Estrogen-Mediated Renoprotection following Cardiac Arrest and Cardiopulmonary Resuscitation Is Robust to GPR30 Gene Deletion. PLoS One. 2014;9(6):e99910 10.1371/journal.pone.0099910 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Lee JH, Park SH, Song IA. Cardiopulmonary resuscitation with assisted extracorporeal life support during cardiac arrest caused by drug-eluting stent thrombosis: a case report. Korean journal of anesthesiology. 2014;66(5):383–7. 10.4097/kjae.2014.66.5.383 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Gupta T, Kolte D, Khera S, Aronow WS, Palaniswamy C, Mujib M, et al. Relation of Smoking Status to Outcomes After Cardiopulmonary Resuscitation for In-Hospital Cardiac Arrest. Am J Cardiol. 2014. 10.1016/j.amjcard.2014.04.021 . [DOI] [PubMed] [Google Scholar]
  • 22. Hutchens MP, Fujiyoshi T, Koerner IP, Herson PS. Extracranial hypothermia during cardiac arrest and cardiopulmonary resuscitation is neuroprotective in vivo. Therapeutic hypothermia and temperature management. 2014;4(2):79–87. 10.1089/ther.2014.0003 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23. Hsin T, Chun FW, Tao HL. Ultra-long cardiopulmonary resuscitation with thrombolytic therapy for a sudden cardiac arrest patient with pulmonary embolism. Am J Emerg Med. 2014. 10.1016/j.ajem.2014.04.035 . [DOI] [PubMed] [Google Scholar]
  • 24. Lewis LM, Stothert J, Standeven J, Chandel B, Kurtz M, Fortney J. Correlation of End-Tidal Co2 to Cerebral Perfusion during Cpr. Annals of Emergency Medicine. 1992;21(9):1131–4. 10.1016/S0196-0644(05)80658-4 . [DOI] [PubMed] [Google Scholar]
  • 25. Callaham M, Barton C, Matthay M. Effect of epinephrine on the ability of end-tidal carbon dioxide readings to predict initial resuscitation from cardiac arrest. Crit Care Med. 1992;20(3):337–43. . [DOI] [PubMed] [Google Scholar]
  • 26. Koster RW, Baubin MA, Bossaert LL, Caballero A, Cassan P, Castren M, et al. European Resuscitation Council Guidelines for Resuscitation 2010 Section 2. Adult basic life support and use of automated external defibrillators. Resuscitation. 2010;81(10):1277–92. 10.1016/j.resuscitation.2010.08.009 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27. Sutton RM, Friess SH, Maltese MR, Naim MY, Bratinov G, Weiland TR, et al. Hemodynamic-directed cardiopulmonary resuscitation during in-hospital cardiac arrest. Resuscitation. 2014. 10.1016/j.resuscitation.2014.04.015 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28. Lopez-de-Sa E, Lopez-Sendon J. Cardiac arrest survivors before they reach hospital. Beyond cardiopulmonary resuscitation. Revista espanola de cardiologia. 2013;66(8):606–8. 10.1016/j.rec.2013.04.002 . [DOI] [PubMed] [Google Scholar]
  • 29. Kantamineni P, Emani V, Saini A, Rai H, Duggal A. Cardiopulmonary Resuscitation in the Hospitalized Patient: Impact of System-Based Variables on Outcomes in Cardiac Arrest. The American journal of the medical sciences. 2014. 10.1097/MAJ.0000000000000290 . [DOI] [PubMed] [Google Scholar]
  • 30. Idris AH, Guffey D, Aufderheide TP, Brown S, Morrison LJ, Nichols P, et al. Relationship between chest compression rates and outcomes from cardiac arrest. Circulation. 2012;125(24):3004–12. 10.1161/CIRCULATIONAHA.111.059535 [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

S1 Minimal Data Set. This file includes all the data in the Manuscript.

Line 29, ID 27, marked red, was removed because of the extreme values.

(XLS)

Data Availability Statement

All relevant data are within the paper and its Supporting Information files.


Articles from PLoS ONE are provided here courtesy of PLOS

RESOURCES