Abstract
Large format (i.e., > 25 cm2) cranioplasty is a challenging procedure not only from a cosmesis standpoint, but also in terms of ensuring that the patient's brain will be well-protected from direct trauma. Until recently, when a patient's own cranial flap was unavailable, these goals were unattainable. Recent advances in implant Computer Aided Design and 3-D printing are leveraging other advances in regenerative medicine. It is now possible to 3-D-print patient-specific implants from a variety of polymer, ceramic, or metal components. A skull template may be used to design the external shape of an implant that will become well integrated in the skull, while also providing beneficial distribution of mechanical force distribution in the event of trauma. Furthermore, an internal pore geometry can be utilized to facilitate the seeding of banked allograft cells. Implants may be cultured in a bioreactor along with recombinant growth factors to produce implants coated with bone progenitor cells and extracellular matrix that appear to the body as a graft, albeit a tissue-engineered graft. The growth factors would be left behind in the bioreactor and the graft would resorb as new host bone invades the space and is remodeled into strong bone. As we describe in this review, such advancements will lead to optimal replacement of cranial defects that are both patient-specific and regenerative.
Keywords: Cranioplasty, 3-D printing, Additive Manufacturing, Computer Aided Design (CAD), Cranial Implant, Regenerative Medicine, Tissue Engineering
I Introduction
The neurosurgical reconstruction of cranial defects has witnessed a tremendous evolution over recorded history. In discussing this evolution, it is important to elaborate upon the following: cranial defect size (i.e., critical and non-critical size defects); implant material choice (e.g., autologous or alloplastic grafts); and the use of tissue engineering strategies (i.e., resorbable implants alone or the addition of cells and/or growth factors to facilitate regeneration). Other than cell-based therapy, these aspects of the surgical approach to cranioplasty are reported in Sanan and Haines' 1997 signature report1. Significant progress in the field since their landmark paper has prompted us to provide an updated review of the surgical approach to cranioplasty. We review recent advancements in: a) bone substitute materials (Section II); b) cranial implant fabrication through Computer Aided Design (CAD) and Computer Aided Manufacture (CAM) (Section III); c) Tissue Engineering therapies for the regeneration of critical size or larger bone defects (Section IV); and, d) advances in the neurosurgical approach to cranioplasty (Section V). We conclude with expected future improvements to cranial defect treatment options.
II Advances in Alloplastic Bone Substitutes for Cranial Repair
The Role of the Cranial Vault as a Skeletal Element
The skull is a unique boney element with structural and functional idiosyncrasies. Unlike other bones, the skull does not transmit weight or bear loads during ambulation or manipulation of the environment. The skull shields the brain from both trauma and infection. Current treatment options for large cranial defects are less effective at protecting the underlying brain than the unmodified adult skull. The implant-bearing skull may suffer abnormal strain concentrations that threaten its protective capability and an increased susceptibility to fatal head trauma2. The gap between the skull and an alloplastic implant may serve as a route for infection3.
Critical Size Defect
The surgical demands of procedures involving trauma, vascular disease, infection, cancer, or congenital deformity often require neurological surgeons to osteotomize large areas of the cranium. Skull resection and entry sites of greater than 25 cm2 are common in adults4, and defects of 100 cm2 are the average size seen by manufacturers of cranial prosthetic plates5. In an adult, 100 cm2 is roughly a quarter of the total cranial surface area. A “critical size” defect is any that is too large for the patient to heal unaided6. If revascularization fails or is incomplete following bone flap engraftment, then resorption, necrosis, and/or infection commonly follows7. Furthermore, post-surgical healing of large bone flaps requires osteogenic activity that may tax the patient's regenerative abilities which are known to decline with age8. Re-incorporation of a large bone flap depends on the local healing response and vascular supply. The blood supply for cranial bone comes primarily from the underlying middle meningeal artery with a much smaller contribution from the overlying scalp9.
Cranial defect repair following cranioplasty in children is of a different nature than in adults. Although there is a higher risk of resorption when flap replacement must be delayed10, and there is a significant risk of inflammation or infection from delayed cranioplasty in the pediatric patient11, the greater regenerative abilities and variety of sources of autologous bone in children increases the likelihood of healthy regeneration of the cranial vault, when compared to adults12.
Autografts and Allografts
First, patients commonly receive bone grafts from elsewhere in their body (i.e., autograft tibia, rib, scapula, sternum, ilium, fibula, or cranium). Autograft therapy of cranial defects is the standard to which all other strategies are compared.13 However, there are several limitations to its use: the supply of graft material is limited by donor site size; autograft harvesting risks long-term donor site pain and adjacent bone morbidity14; and the grafted bone may fail to vascularize15. Complex autografts require advanced technical abilities of the surgeon. For example, fibular or iliac bone may be difficult to shape for the complex geometry of some cranial regions. As a whole, this process is time-consuming, cumbersome, and may result in long-term pain at the donor site1. If graft integration is insufficient or fails, the resulting necrotic bone is at high risk for non-union, mal-union, or infection. Another outcome for necrotic cranial bone is resorption. Insufficiently vascularized regions of a cranial flap may also shield antibiotics and mute immune response. On rare occasions, extra-dural infection occurs.
A second option for large cranial defect repair, cadaver or live bone donation (i.e., allograft), is often equally difficult to shape and carries risk of immunological rejection and pathogen transmission16. The use of bone and cartilage tissue from cadavers was an extension of the success of autograft transplantation that was popularized during WWI7. Allografts provide beneficial properties of structural integrity and facilitate remodelling7 without the complications of donor site morbidity. However, significant risk of disease transmission, infection, and immunologic reactivity has impeded the use of allografts. Of note, advances in gene sequencing, sterilization, and graft harvesting and storage have minimized these risks to a significant extent.
Alloplastic (Synthetic) Bone Substitutes: Metal
If a patient loses their own cranial bone flap to trauma, infection, or cancer resection, there are primarily three classes of artificial augmentation materials: metal, polymer, and ceramic. The use of metallic bone substitutes (i.e., alloplastic materials) has a long history. The most common material still in use is grade 5 surgical titanium (i.e., Ti-6Al-4V), usually referred to as “titanium” or Ti-6-417. Titanium is used in the cranium for fixation devices (e.g., plates and screws), mesh, or solid plates (Figure 1), or in combination with other materials such as inert plastic or ceramic components.4 It is generally accepted that titanium is corrosion-resistant, although there is evidence that trace minerals are released over time,18 perhaps eliciting a subtle immunologic response.19 Insufficient perfusion of the overlying scalp flap and mechanical irritation from underlying alloplastic materials or fixation devices (e.g., Ti micro-plating and screws), may lead to inflammation, seroma, infection, or implant fenestration through the scalp.20
Figure 1.
Titanium mesh cranial implant (courtesy: Medtronic, Minneapolis, MN).
The skull has elastic properties that are similar to the cortical bone found in long bones.21 Ti-6Al-4V has an elastic modulus that ranges from 14.5-38.5 GPa22 (Giga-Pascal) for porous implants to 110 GPa23 for bulk specimens. This compares to an elastic modulus of approximately 10.4-19.6 GPa, depending on the location and quality of bone (e.g., radionecrosis or osteoporosis can reduce bone quality), in the adult skull. 21 While titanium implants are less likely to fail during traumatic loading than other alloplastic materials, they may potentially break the surrounding bone or suffer a fixation screw pull-out, especially in regions where strain is concentrated (e.g., microplate fixation).24 An alternative strategy would be to use an overlapping margin to maximize contact between the implant and the patient's skull.25 With abutting, overlapping, or both types of margins it is unlikely that the implant-bearing skull would dissipate strain as effectively (i.e., without failure) as a normal unaffected skull.26 Finally, it is important to note that surface-textured or bioactively functionalized28 titanium implants can strongly incorporate soft tissue such as dura and periosteum; this can make revision difficult.27
Synthetic Bone Substitutes: Polymers
Poly(methyl methacrylate) (PMMA) is perhaps the most frequently used cranial bone graft substitute (Figure 2). It is often provided as a powder and liquid (e.g., Cranioplastic™, DePuy Synthes, Raynham, MA), which are combined to form a malleable paste that is shaped intra-operatively to fit the contours of the patient's cranial defect during surgery. PMMA offers the advantages of a protective, defect-filling replacement that lacks post-operative inflammation. However, PMMA is not without its disadvantages. The polymerization process is significantly exothermic and generates large amounts of heat energy (78-120 °C)29 during the period when it can be manually molded into place. Inadequate cooling of the material can thus damage surrounding brain tissue or dura. PMMA is associated with significant post-placement shrinkage. Other complications of PMMA cranioplasty include infection30 and pulmonary embolism (due to PMMA extravasation into venous circulation).31 Most importantly, PMMA lacks the osseointegrative and osteoinductive properties that would provide complete regeneration of reliable bone in a large cranial defect. However, the lower elastic modulus of PMMA (approximately 3 GPa) compared to surgical titanium (approximately 110 GPa) conveys a lesser discontinuity at the implant-host interface, and may result in less stress-shielding and less loosening of fixation devices over time. Nonetheless, a discontinuity remains. Therefore, traumatic injury directly to a PMMA plate can force it through the adherent screws and drive it directly into the brain, grossly endangering the patient.32 The safety of one particular product line, Hard Tissue Replacement by Lorenz Surgical, Inc. (BIOMET, Warsaw, IN), has been called into question for patients undergoing decompressive craniectomy.33
Figure 2.
PMMA implant (courtesy: Stryker, Kalamazoo, MI).
Another polymer, polyethylene, has long been in use as a material for cranial implants. It is also known to carry a risk of infection.34 The leading vendor has been MEDPOR® (Figure 3), now a subsidiary of Stryker (Kalamazoo, MI), which provides a highly porous form of this material that is flexible, can be easily cut with a scissors, and has recently been combined with titanium mesh in a new product.35
Figure 3.
MEDPOR® implant (courtesy: Stryker).
Several other polymers have been investigated for large format cranial implants: PEEK (poly-ether-ether ketone),36 Polypropylene Polyester Knitwear,37 and PEKK (Poly-ether-ketone ketone) (Figure 4).38 There is also an FDA-approved PEKK custom cranial implant product (Oxford Performance Materials, South Windsor, CT). These materials have been used for the prefabrication of cranial implants (see Section IV). Vendors of these polymeric materials claim that they are more similar in their material properties to the surrounding bone, especially in regard to elastic modulus. Oxford Performance Materials also claims that PEKK is osteoconductive.38
Figure 4.
PEEK implant (courtesy: A: Stryker; B: KLS Martin, (City, St of HQ); C: KLS Martin).
Synthetic Bone Substitutes: Ceramics
Calcium phosphate (CaP) has received mixed reviews as a bone substitute. Some consider CaP a tissue engineering scaffold and have observed limited resorption and/or support of new cranial bone growth and remodeling, perhaps in children;39 others, however, have seen no such resorption.4 In either case, these malleable ceramic alloplastic materials are extremely brittle and are largely used to fill minor gaps in the skull. Stryker's CaP product, Bone Source™, for example, is FDA-approved for repair of cranial fracture gaps of less than 5 mm.40 CaP paste hydrolyzes into solid HA (hydroxyapatite) inside the body. Bone grows epitaxially onto solid HA or into the margins of adequately porous HA.41, 42 HA, hydrolyzed in situ, is approximately one order of magnitude more brittle than calvarial bone43 and exhibits approximately 60% of the compressive strength of PMMA (much less than 60% under tension or bending).44-46 Self-setting CaP paste is commonly troweled into large cranial defects over a titanium mesh to provide additional stiffness1, 4 or pre-operatively prepared via the use of skull models.47 The fragility of HA, however, precludes its isolated use for large deficits.
III Advances in Pre-operative Cranial Implant Fabrication
Pre-fabrication of a well-fitting implant saves operating room procedure time, may render the procedure less invasive, and reduces cost and the risk of infection. Until recently, preoperative fabrication was achieved manually, using topographical information lacking precision.48, 49 Although infection remains the most cited reason for implant revision or removal,46, 47 poor fit nonetheless represents a crucial area for improvement. Fortunately, the recent revolution in 3-D printing has made its way to cranioplasty graft design, thereby vastly improving the “fit” of premade cranial grafts. Barnatt50 discusses the various types of 3-D printing technologies, now preferably referred to as Additive Manufacturing (AM)51 (formerly known as Rapid Prototyping or Solid Free-Form) that have been utilized to produce both inert implants for patients and tissue-engineered implants for preclinical tissue engineering research. In regard to cranial implants, one of the first AM technologies, stereolithography, utilized models of the skull obtained from CTscans to manually design cranial implants that were then recast in implantable materials.
There are not many non-toxic, resorbable materials that can be used in these printers at this time. All AM technologies build an implant layer by layer from a Computer Aided Design (CAD) file. Table 1 presents the three basic mechanisms by which current 3-D printers render implants. The first method is light-based polymerization, i.e., photocrosslinking, of liquid polymer resin. Two types of devices use this first method, photocrosslinking; they are stereolithography52, 53 and continuous Digital Light Processing.54 A second method for 3-D printing implants use liquid or heat to bind particles. One of the earliest technologies to do so, referred to as 3-D Printing (3DP™, Therics Inc., Princeton, NJ), involves the injection of a liquid binder (which may contain cells and/or growth factors) into powder55. Selective Laser Sintering (SLS) uses a laser to bind powder, layer by layer (3D Systems, Valencia, CA). The third AM modality, Fused Deposition Modeling (Stratasys Inc., Eden Prairie, MN), uses an inkjet to extrude material that is heated just above melting temperature.56 A variation on this method prints an implant in wax (Sanders Prototype, Milton, NH) for later replacement with resorptive materials.57
Table 1.
Three basic additive manufacturing technologies for binding materials together during a 3-D printing process and the materials which can be used. There are not many implantable materials that can be 3-D printed at this time. However, at a minimum, these technologies can be used to 3-D print a model of a shape designed on a computer. That part can then be the positive for a mold that is then used to cast an implantable material.
| Modalities | Light-Based | Liquid or Heat-Based | Mesh/Coat |
|---|---|---|---|
| Stereolithography | Polymer, Ceramic | ||
| DLP Projection (cDLP) | Polymer, Ceramic | ||
| Inkjet/Cell | polymer, cells | ||
| 3D Printing | Polymer | ||
| Extrusion | Polymer | ||
| Laser Sintering & Selective Laser Melting, Electron Beam Deposition, Electron Beam Melting | Polymer, Metal | ||
| Electrospinning | Polymer |
Cranial Implant CAD Software
Computer aided design of cranial implants can involve the use of templates, such as that of an average skull image or of a right-left mirrored image of the patient (only useful if the defect does not cross the midline).58-60 The template is then appropriately warped to the region of the skull defect. The defect-filling portion of the template is then “cookie-cut” so that thickness can be added to create a tapered edge that abuts and/or overlaps the surrounding skull (Figure 5). During this process of ensuring a good fit, it is also important to: (1) ensure there is no intersection of the adjacent dural sac and brain, and (2) ensure there is sufficient scalp to cover the implant or that galeal scoring and/or tissue expansion may be necessary prior to placement of the implant..5, 59, 62, 63
Figure 5.

Cranial template used to design five (a-e) patient-specific implants, all with tapered edge fit. Red surface on taper is in contact with surrounding skull (Figure 4 from: Kyoung-june Min and David Dean. Highly Accurate CAD Tools for Cranial Implants. In (R.E. Ellis and T.M. Peters, Eds.): MICCAI 2003, LNCS 2878, pp. 99-107, 2003. © Springer-Verlag Berlin Heidelberg 2003).
One area that deserves special attention with either autologous or alloplastic repair is the upper orbital rim and brow, also referred to as the orbital bandeau. Where there is risk to the frontal sinus mucous membranes it is often obliterated; these risks include either (1) a lack of containment, risking mucocoele, or (2) adjacency to alloplastic materials, risking deterioration and erosion due to poor integration which may then be followed by infection. While congenital malformation repair may require orbital advancement, care must be taken not to encroach the orbital roof, apex, or medial wall. Encroachment in these areas can lead to neurovascular compromise, especially compromising visual acuity or causing external ophthalmoplegia. It is also important in these cases that revision around the nasion or glabellar regions not cause problems with nasal angulation or asymmetry.61
Additive Manufacture: Metallic Cranial Implants
Carr et al. (1997) and Eufinger et al. (1995)64 have demonstrated a system for milling of titanium (inert) prosthetic cranial plates. These techniques continue to be in practice today.65 While new Ti alloys are being developed, the most commonly used material continues to be Ti-6Al-4V, which can be rendered via SLS,66 Selective Laser Melting (SLM),67 Electron Beam Melting.68, 69 These technologies may be able to create porous spaces in the 150-1000 μm range that are mostly likely to promote adjacent bone ingrowth.67, 69 It remains to be shown whether surface ingrowth will create a stress-strain gradient that can successfully translate strain during a traumatic impact. One group70 has noted CAD/CAM-rendered ceramic-implants to be more expensive than PMMA- or Ti-implants prepared in the same manner. Further, they find Ti-implants are the least likely to be removed and that they are a useful alternative when PMMA-implants need to be removed.
Additive Manufacture: Polymeric Cranial Implants
The utilization of synthetic polymers for additive manufacturing (AM) of cranial implants is currently the most widely utilized methodology. Given the difficulties of working with PMMA, most of the companies that offer AM of patient-specific polymer implants do so with PEEK via SLS. The design of these implants can be modified to match the material properties to those of the surrounding skull.71 Lethaus et al.36 concluded that the material properties of PEEK cranial implants are preferable to those of titanium cranial implants. Jaekel et al. report that PEEK material properties are sensitive to the crystallinity of the material that is used.72 Carbon-fiber reinforcement of PEEK, which can be rendered by stereolithography,73 has been offered as a means of increasing the material's fatigue resistance in areas of the skeleton that carry a constant load.74
Additive Manufacture: Ceramic Cranial Implants
Ceramics have been demonstrated to be a good bone scaffolding material in sub-critical size defects, although their material properties preclude their use for larger deficits (see section 2.4). Attention has recently turned to stereolithography to render large format ceramic cranial implants.42 A French group, led by Moreau, has generated ceramic graft implants via an AM technique in which polymer is used as a binder during 3-D printing and then sintered away. Their results indicated the potential to use ceramics for large scale defect filling. Also under investigation is a new composite material composed of 50% Beta-tricalcium phosphate (B-TCP) and 50% poly(D, L)-lactide (PDLLA) that is synthesized via SLM.75
IV Tissue Engineering
Tissue engineering involves the use of resorbable scaffolds, growth factors, cells, and sometimes bioreactors. There are thus no FDA-approved standard of care approaches to tissue engineered repair of large (> 25 cm2)4 cranial defects. We limit discussion here to tissue engineering cranial deficit therapies that are in the clinic or clinical trials.
Scaffolds
In addition to the previously noted use of ostensibly resorbable ceramics, patients in several studies involving large resorbable Poly(α-hydroxy ester)-based prosthetics have encountered problems of excess lactic acid due to normal implant degradation. While solid PLGA has shown osteoconductive properties in humans,76-79 large PLGA prosthetics undergoing mechanical strain have been observed to develop a rapid decrease in molecular weight, and loss of strength followed by bulk degradation, releasing dangerous levels of lactic acid (Bos et al., 1991).80 These polymer chains are soluble in the extracellular fluid when their molecular weight falls below 7,000 Daltons,81 at which point the polymer has little strength and begins to fragment due to local mechanical stresses. When particles are approximately 20 μm they can be phagocytosed by multinucleated cells or macrophages. Those particles that are not phagocytosed continue to degrade and decrease in size. As lactic and glycolic acid are released, there is danger of a local drop in pH, regardless of peri-implant vascularity.82, 83 Many workers concur that these byproducts prevent bone ingrowth beyond a few millimeters and suggest that PLGA or PLA suture, meshes, solid plates, and screws are best for narrow gap repair of 5 mm or less in bone and/or overlying membranes.84-87
Polycaprolactone (PCL) can be 3-D printed via Fused Deposition Modeling. It has been used in a number of preclinical bone tissue engineering studies. Probst et al.88 report the implantation of a 3-D printed PCL and tricalcium phosphate cranial scaffold. They note observing CT-based evidence of bone consolidation as well as good integration of the implant after 6 months. In a previous study, they concluded that the in vivo resorption of these scaffolds would be due to molecular weight, crystallinity, hydrophilicity, and design.89
Growth Factors
Bone tissue engineering became a large scale clinical reality with Medtronic's (Minneapolis, MN) Infuse® (collagen sponge, BMP-2, and titanium cage) for intervertebral fusion. Despite concern about the safety and efficacy of this approach,90, 91 off-label use has extended to osteonecrotic mandible92 and upper93 and lower alveolar repair94 with mixed results. Questions about dose and release rate have led to studies of new implant release strategies.95
Cells
Research is underway to characterize96 the potency and safety of using embryonic, postnatally induced pluripotent or isolated mesenchymal stem cells (MSCs) to seed a scaffold with bone progenitor cells for culture in a bioreactor97, 98 or inside the body (Figure 6).99 It is expected that allografted bone marrow-derived MSCs will not cause an immune response, and if quickly replaced by autologous bone progenitor cells they may provide a banked source of cells to produce osteoconductive and osteoinductive scaffolds.
Figure 6.

3-D printed Poly(propylene fumarate) (resorbable polymer) scaffolds seeded with human mesenchymal stem cells at (A) 6 hours, (B) 18 hours, (C) 30 hours, and (D) 48 hours. (Figure 8 from: Wallace, Jonathan, Martha O. Wang, Paul Thompson, Mallory Busso, Vaijayantee Belle, Nicole Mammoser, Kyobum Kim, John P. Fisher, Ali Siblani, Yueshuo Xu, Jean F Welter, Donald P. Lennon, Jiayang Sun, Arnold I Caplan, and David Dean. “Validating continuous digital light processing (cDLP) additive manufacturing accuracy and tissue engineering utility of a dye-initiator package.” Biofabrication 6, no. 1 (2014): 015003).
V Neurosurgical techniques and clinical implications
Clinically, the nature of cranial repair is dictated largely by circumstance. Defect size, location, and laterality limit which material (or fabrication method) can be used. The cost and availability of resources, as well as the timing/emergent nature of the case, present similar constraints. Only in elective surgery can the full spectra of cranioplasty resources be considered (Figure 7).
Figure 7.
(A) Mesh cranioplasty after removal of large infiltrative space-occupying intracranial lesions may lead to accumulation of subgaleal/epidural hematoma through the pores of the mesh graft (white arrow) – one disadvantage of using mesh-cranioplasty in such cases. (B) Trauma to the cranioplasty graft in such conditions can precipitate subdural hemorrhage (red arrow) as well as impact brain injury leading to parenchymal bleeds. (C, D) In patients with abnormally thick skulls or hyperostosis, a synthetic graft that matches the skull thickness can produce catastrophic intracranial injury upon subsequent direct impact over the graft. In such cases, a thicker plate with embedded re-engineered bone would be superior. (E, F) Post-operative cranial osteomyelitis involving a craniotomy flap across the superior sagittal sinus. (G) CT head non-contrast axial section following cranioplasty removal and wound wash-out. (H) 3-D reconstruction of cranial defect crossing superior sagittal sinus (indicated in blue) showing the extent of proposed customized delayed non-mesh cranioplasty.
Most frequently, as in the case of decompressive hemicraniectomy, the autologous flap remains available for use. In these instances, flaps may be frozen (typically at -80°C) or implanted subcutaneously in the abdominal wall for later implantation (weeks to months after the initial surgery, although the timing is controversial100, 101). As such, evidence maintains the superiority of the autograft. Importantly, however, autologous bone may be subject to aseptic necrosis, infection, and may not survive freezing.102 The frequency and predictability of these scenarios is unclear. However, we recommend the use of autologous flaps when possible.
When autologous bone is unavailable or unsuitable for cranioplasty, the choice of specific allograft material is left to the discretion of the neurosurgeon. However, it is critical that a highly protective material like allograft or titanium be used in an area where brain herniation, infection, and/or CSF leak is present or imminent. Indeed, given the possibility that time may be needed for autologous bone grafts to incorporate, it may best to use inert material, such as titanium, for reliable protection where the risk is high.
At present, the evidence for adoption of a specific synthetic graft material is lacking.103 Hospital resources and policy, cost, and technical experience tend to dictate which material will be used, and clinical history seems to predict outcome more so than choice of material. In our experience, the availability and long-term success rate of titanium (i.e., surgical grade 5 titanium, Ti-6Al-4V), specifically with regard to its resistance to infection, has contributed to its dominance in recent decades. However, in a review of the literature Yadia et al. report the infection rate following cranioplasty averages 7.9% and that the difference in infection of autologous versus alloplastic grafts is not statistically significant.104
VI Future Directions in Cranioplasty
Since Sanan and Haines'1 review of the historical and current approaches to cranioplasty, a revolution has occurred in fabrication technologies (e.g., CAD software and 3-D printing) and regenerative medicine (e.g., tissue engineering). Now the two fields are coming together. Regenerative medicine can solve the problems of good integration of implants and protection of the underlying brain from infection. Patient-specific fabrication of non-resorbable materials that serve specific local needs should be able to replicate, if not improve upon, the distribution of traumatic mechanical forces that might threaten the brain following cranioplasty.
Acknowledgments
This research was partially supported by the Research Foundation of the Department of Neurological Surgery, Case Western Reserve University (CWRU), Cleveland, OH, and NIH grant R01-DE013740. We would like to thank Dr. Clare Rimnac (Department of Mechanical and Aerospace Engineering, CWRU) for her helpful comments on the biomechanical passages in this paper. This paper was improved by the comments of five reviewers.
Footnotes
Disclosures: DD is an inventor on patents and patent applications that include material in this manuscript. DD has received compensation from, and has an ownership stake in, Osteoplastics, LLC (Shaker Heights, OH), an entity with an interest in these patents. DD also has a Research Collaboration Agreement (RCA) with EnvisionTEC, Inc. (Dearborn, MI), a 3D printer manufacturer. The other authors have no personal, financial, or institutional interest in any of the drugs, materials, or devices described in this article.
References
- 1.Sanan A, Haines SJ. Repairing holes in the head: a history of cranioplasty. Neurosurgery. 1997;40(3):588–603. doi: 10.1097/00006123-199703000-00033. [DOI] [PubMed] [Google Scholar]
- 2.Laure B, Tranquart F, Geais L, Goga D. Evaluation of skull strength following parietal bone graft harvest. Plastic and Reconstructive Surgery. 2010 Nov;126(5):1492–1499. doi: 10.1097/PRS.0b013e3181ef8bb1. [DOI] [PubMed] [Google Scholar]
- 3.Yadla S, Campbell PG, Chitale R, Maltenfort MG, Jabbour P, Sharan AD. Effect of early surgery, material, and method of flap preservation on cranioplasty infections: a systematic review. Neurosurgery. 2011 Apr;68(4):1124–1129. doi: 10.1227/NEU.0b013e31820a5470. discussion 1130. [DOI] [PubMed] [Google Scholar]
- 4.Durham SR, McComb JG, Levy ML. Correction of large (>25 cm(2)) cranial defects with “reinforced” hydroxyapatite cement: technique and complications. Neurosurgery. 2003 Apr;52(4):842–845. doi: 10.1227/01.neu.0000054220.01290.8e. discussion 845. [DOI] [PubMed] [Google Scholar]
- 5.Dean D, Min KJ, Bond A. Computer aided design of large-format prefabricated cranial plates. The Journal of craniofacial surgery. 2003;14(6):819–832. doi: 10.1097/00001665-200311000-00002. [DOI] [PubMed] [Google Scholar]
- 6.Schmitz JP, Hollinger JO. The critical size defect as an experimental model for craniomandibulofacial nonunions. Clinical orthopaedics and related research. 1986 Apr;(205):299–308. [PubMed] [Google Scholar]
- 7.Prolo DJ, Gutierrez RV, DeVine JS, Oklund SA. Clinical utility of allogeneic skull discs in human craniotomy. Neurosurgery. 1984 Feb;14(2):183–186. doi: 10.1227/00006123-198402000-00011. [DOI] [PubMed] [Google Scholar]
- 8.Quarto R, Thomas D, Liang CT. Bone progenitor cell deficits and the age-associated decline in bone repair capacity. Calcif Tissue Int. 1995 Feb;56(2):123–129. doi: 10.1007/BF00296343. [DOI] [PubMed] [Google Scholar]
- 9.Hobar PC, Masson JA, Wilson R, Zerwekh J. The importance of the dura in craniofacial surgery. Plastic and Reconstructive Surgery. 1996 Aug;98(2):217–225. doi: 10.1097/00006534-199608000-00003. [DOI] [PubMed] [Google Scholar]
- 10.Bowers CA, Riva-Cambrin J, Hertzler DA, 2nd, Walker ML. Risk factors and rates of bone flap resorption in pediatric patients after decompressive craniectomy for traumatic brain injury. J Neurosurg Pediatr. 2013 May;11(5):526–532. doi: 10.3171/2013.1.PEDS12483. [DOI] [PubMed] [Google Scholar]
- 11.Bhaskar IP, Inglis TJ, Lee GY. Clinical, Radiological, and Microbiological Profile of Patients with Autogenous Cranioplasty Infections. World neurosurgery. 2013 doi: 10.1016/j.wneu.2013.01.013. Journal Article. [DOI] [PubMed] [Google Scholar]
- 12.Rocque BG, Amancherla K, Lew SM, Lam S. Outcomes of cranioplasty following decompressive craniectomy in the pediatric population. J Neurosurg Pediatr. 2013 Aug;12(2):120–125. doi: 10.3171/2013.4.PEDS12605. [DOI] [PubMed] [Google Scholar]
- 13.Prolo DJ, Oklund SA. The use of bone grafts and alloplastic materials in cranioplasty. Clinical orthopaedics and related research. 1991;(268):270–278. 268. [PubMed] [Google Scholar]
- 14.St John TA, Vaccaro AR, Sah AP, et al. Physical and monetary costs associated with autogenous bone graft harvesting. Am J Orthop (Belle Mead NJ) 2003 Jan;32(1):18–23. [PubMed] [Google Scholar]
- 15.Habal MB, Reddi AH. Bone grafts and bone induction substitutes. Clin Plast Surg. 1994 Oct;21(4):525–542. [PubMed] [Google Scholar]
- 16.Vastel L, Meunier A, Siney H, Sedel L, Courpied JP. Effect of different sterilization processing methods on the mechanical properties of human cancellous bone allografts. Biomaterials. 2004 May;25(11):2105–2110. doi: 10.1016/j.biomaterials.2003.08.067. [DOI] [PubMed] [Google Scholar]
- 17.Andani MT, Moghaddam NS, Haberland C, Dean D, Miller MJ, Elahinia M. Metals for bone implants. Part 1. Powder metallurgy and implant rendering. Acta biomaterialia. 2014 Jun 20; doi: 10.1016/j.actbio.2014.06.025. [DOI] [PubMed] [Google Scholar]
- 18.Jorgenson DS, Mayer MH, Ellenbogen RG, et al. Detection of titanium in human tissues after craniofacial surgery. Plastic and Reconstructive Surgery. 1997 Apr;99(4):976–979. doi: 10.1097/00006534-199704000-00006. discussion 980-971. [DOI] [PubMed] [Google Scholar]
- 19.Bianco PD, Ducheyne P, Cuckler JM. Titanium serum and urine levels in rabbits with a titanium implant in the absence of wear. Biomaterials. 1996 Oct;17(20):1937–1942. doi: 10.1016/0142-9612(96)00023-3. [DOI] [PubMed] [Google Scholar]
- 20.Adams K, Couch L, Cierny G, Calhoun J, Mader JT. In vitro and in vivo evaluation of antibiotic diffusion from antibiotic-impregnated polymethylmethacrylate beads. Clinical orthopaedics and related research. 1992 May;(278):244–252. [PubMed] [Google Scholar]
- 21.Peterson J, Dechow PC. Material properties of the human cranial vault and zygoma. Anat Rec A Discov Mol Cell Evol Biol. 2003 Sep;274(1):785–797. doi: 10.1002/ar.a.10096. [DOI] [PubMed] [Google Scholar]
- 22.Li X, Feng YF, Wang CT, et al. Evaluation of biological properties of electron beam melted Ti6Al4V implant with biomimetic coating in vitro and in vivo. PLoS One. 2012;7(12):e52049. doi: 10.1371/journal.pone.0052049. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Majumdar P, Singh SB, Chakraborty M. Elastic modulus of biomedical titanium alloys by nano-indentation and ultrasonic techniques - A comparative study. Mat Sci Eng a-Struct. 2008 Aug 20;489(1-2):419–425. [Google Scholar]
- 24.Tsouknidas A, Maropoulos S, Savvakis S, Michailidis N. FEM assisted evaluation of PMMA and Ti6Al4V as materials for cranioplasty resulting mechanical behaviour and the neurocranial protection. Bio-Med Mater Eng. 2011;21(3):139–147. doi: 10.3233/BME-2011-0663. [DOI] [PubMed] [Google Scholar]
- 25.Scholz M, Wehmoller M, Lehmbrock J, et al. Reconstruction of the temporal contour for traumatic tissue loss using a CAD/CAM-prefabricated titanium implant-case report. J Cranio Maxill Surg. 2007 Dec;35(8):388–392. doi: 10.1016/j.jcms.2007.06.006. [DOI] [PubMed] [Google Scholar]
- 26.Yoganandan N, Pintar FA. Biomechanics of temporo-parietal skull fracture. Clin Biomech (Bristol, Avon) 2004 Mar;19(3):225–239. doi: 10.1016/j.clinbiomech.2003.12.014. [DOI] [PubMed] [Google Scholar]
- 27.Sayama CM, Sorour M, Schmidt RH. Dural adhesion to porous cranioplastic implant: A potential safety concern. Surg Neurol Int. 2014;5:19. doi: 10.4103/2152-7806.127377. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Zile MA, Puckett S, Webster TJ. Nanostructured titanium promotes keratinocyte density. J Biomed Mater Res A. 2011;97(1):59–65. doi: 10.1002/jbm.a.33028. [DOI] [PubMed] [Google Scholar]
- 29.Golz T, Graham CR, Busch LC, Wulf J, Winder RJ. Temperature elevation during simulated polymethylmethacrylate (PMMA) cranioplasty in a cadaver model. Journal of clinical neuroscience : official journal of the Neurosurgical Society of Australasia. 2010 May;17(5):617–622. doi: 10.1016/j.jocn.2009.09.005. [DOI] [PubMed] [Google Scholar]
- 30.Klinger DR, Madden C, Beshay J, White J, Gambrell K, Rickert K. Autologous and Acrylic Cranioplasty: A Review of 10 Years and 258 Cases. World neurosurgery. 2013 Sep 13; doi: 10.1016/j.wneu.2013.08.005. [DOI] [PubMed] [Google Scholar]
- 31.Kaur R, Younas F, Rajput FA, Kumar S, Afonso L. Polymethyl Methacrylate Induced Pulmonary Embolism. Heart, lung & circulation. 2013 doi: 10.1016/j.hlc.2013.05.646. Journal Article. [DOI] [PubMed] [Google Scholar]
- 32.Ko AL, Nerva JD, Chang JJ, Chesnut RM. Traumatic fracture of polymethylmethacrylate patient-specific cranioplasty implant. World neurosurgery. 2013 Sep 18; doi: 10.1016/j.wneu.2013.09.025. [DOI] [PubMed] [Google Scholar]
- 33.Nassiri N, Cleary DR, Ueeck BA. Is Cranial Reconstruction With a Hard-Tissue Replacement Patient-Matched Implant as Safe as Previously Reported? A 3-Year Experience and Review of the Literature. J Oral Maxil Surg. 2009 Feb;67(2):323–327. doi: 10.1016/j.joms.2008.08.032. [DOI] [PubMed] [Google Scholar]
- 34.Kim H, Sung SO, Kim SJ, Kim SR, Park IS, Jo KW. Analysis of the factors affecting graft infection after cranioplasty. Acta Neurochirurgica. 2013 Nov;155(11):2171–2176. doi: 10.1007/s00701-013-1877-8. [DOI] [PubMed] [Google Scholar]
- 35.Boghani Z, Choudhry OJ, Schmidt RF, Jyung RW, Liu JK. Reconstruction of Cranial Base Defects Using the Medpor Titan Implant: Cranioplasty Applications in Acoustic Neuroma Surgery. Laryngoscope. 2013 Jun;123(6):1361–1368. doi: 10.1002/lary.23840. [DOI] [PubMed] [Google Scholar]
- 36.Lethaus B, Safi Y, ter Laak-Poort M, et al. Cranioplasty with customized titanium and PEEK implants in a mechanical stress model. Journal of neurotrauma. 2012;29(6):1077–1083. doi: 10.1089/neu.2011.1794. [DOI] [PubMed] [Google Scholar]
- 37.Kasprzak P, Tomaszewski G, Kotwica Z, Kwinta B, Zwolinski J. Reconstruction of Cranial Defects with Individually Formed Cranial Prostheses Made of Polypropylene Polyester Knitwear: An Analysis of 48 Consecutive Patients. Journal of neurotrauma. 2012 Apr;29(6):1084–1089. doi: 10.1089/neu.2011.2247. [DOI] [PubMed] [Google Scholar]
- 38.Molitch-Hou M. The First 3D Printed Polymer Implant to Receive FDA Approval. 2013 http://3dprintingindustry.com/2013/02/22/the-first-3d-printed-polymer-implant-to-receive-fda-approval/
- 39.Engstrand T, Kihlstrom L, Neovius E, et al. Development of a bioactive implant for repair and potential healing of cranial defects. Journal of neurosurgery. 2014 Jan;120(1):273–277. doi: 10.3171/2013.6.JNS1360. [DOI] [PubMed] [Google Scholar]
- 40.Schmitz JP, Hollinger JO, Milam SB. Reconstruction of bone using calcium phosphate bone cements: A critical review. J Oral Maxil Surg. 1999 Sep;57(9):1122–1126. doi: 10.1016/s0278-2391(99)90338-5. [DOI] [PubMed] [Google Scholar]
- 41.Bagambisa FB, Joos U, Schilli W. Mechanisms and Structure of the Bond between Bone and Hydroxyapatite Ceramics. Journal of Biomedical Materials Research. 1993 Aug;27(8):1047–1055. doi: 10.1002/jbm.820270810. [DOI] [PubMed] [Google Scholar]
- 42.Brie J, Chartier T, Chaput C, et al. A new custom made bioceramic implant for the repair of large and complex craniofacial bone defects. J Cranio Maxill Surg. 2013 Jul;41(5):403–407. doi: 10.1016/j.jcms.2012.11.005. [DOI] [PubMed] [Google Scholar]
- 43.Xu HHK, Eichmiller FC, Barndt PR. Effects of fiber length and volume fraction on the reinforcement of calcium phosphate cement. J Mater Sci-Mater M. 2001 Jan;12(1):57–65. doi: 10.1023/a:1026753020208. [DOI] [PubMed] [Google Scholar]
- 44.Fritsch EW. Static and fatigue properties of two new low-viscosity PMMA bone cements improved by vacuum mixing. Journal of Biomedical Materials Research. 1996 Aug;31(4):451–456. doi: 10.1002/(SICI)1097-4636(199608)31:4<451::AID-JBM3>3.0.CO;2-K. [DOI] [PubMed] [Google Scholar]
- 45.Pascual B, Vazquez B, Gurruchaga M, et al. New aspects of the effect of size and size distribution on the setting parameters and mechanical properties of acrylic bone cements. Biomaterials. 1996 Mar;17(5):509–516. doi: 10.1016/0142-9612(96)82725-6. [DOI] [PubMed] [Google Scholar]
- 46.Takahashi A, Koshino T. Compressive and Bone-Bonding Strength of Hydroxyapatite Thermal-Decomposition Product Implanted in the Femur of Rabbit as a Bioactive Ceramic Bone-Cement. Biomaterials. 1995 Aug;16(12):937–943. doi: 10.1016/0142-9612(95)93119-x. [DOI] [PubMed] [Google Scholar]
- 47.Staffa G, Barbanera A, Faiola A, et al. Custom made bioceramic implants in complex and large cranial reconstruction: a two-year follow-up. J Craniomaxillofac Surg. 2012 Apr;40(3):e65–70. doi: 10.1016/j.jcms.2011.04.014. [DOI] [PubMed] [Google Scholar]
- 48.Gooch MR, Gin GE, Kenning TJ, German JW. Complications of cranioplasty following decompressive craniectomy: analysis of 62 cases. Neurosurgical focus. 2009 Jun;26(6):E9. doi: 10.3171/2009.3.FOCUS0962. [DOI] [PubMed] [Google Scholar]
- 49.Stephens FL, Mossop CM, Bell RS, et al. Cranioplasty complications following wartime decompressive craniectomy. Neurosurgical focus. 2010 May;28(5):E3. doi: 10.3171/2010.2.FOCUS1026. [DOI] [PubMed] [Google Scholar]
- 50.Barnatt C. 3D Printing: The Next Industrial Revolution: CreateSpace Independent Publishing Platform. 2013 [Google Scholar]
- 51.Stucker BDC, Tuck C, Philips CJ, Boivie KM, Jurrens KK, Lenz J, Vermeulen M. Committee F42 on Additive Manufacturing Technologies. Paper presented at: American Society for Testing and Materials International (ASTM); Conshohocken, PA. [Google Scholar]
- 52.Cooke MN, Fisher JP, Dean D, Rimnac C, Mikos AG. Use of stereolithography to manufacture critical-sized 3D biodegradable scaffolds for bone ingrowth. Journal of biomedical materials research Part B, Applied biomaterials. 2003;64(2):65–69. doi: 10.1002/jbm.b.10485. [DOI] [PubMed] [Google Scholar]
- 53.Dhariwala B, Hunt E, Boland T. Rapid prototyping of tissue-engineering constructs, using photopolymerizable hydrogels and stereolithography. Tissue engineering. 2004 Sep;10(9-10):1316–1322. doi: 10.1089/ten.2004.10.1316. [DOI] [PubMed] [Google Scholar]
- 54.Dean D, Jonathan W, Siblani A, et al. Continuous Digital Light Processing (cDLP): Highly Accurate Additive Manufacturing of Tissue Engineered Bone Scaffolds. Virtual and physical prototyping. 2012;7(1):13–24. doi: 10.1080/17452759.2012.673152. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55.Giordano RA, Wu BM, Borland SW, Cima LG, Sachs EM, Cima MJ. Mechanical properties of dense polylactic acid structures fabricated by three dimensional printing. J Biomater Sci Polym Ed. 1996;8(1):63–75. doi: 10.1163/156856297x00588. [DOI] [PubMed] [Google Scholar]
- 56.Zein I, Hutmacher DW, Tan KC, Teoh SH. Fused deposition modeling of novel scaffold architectures for tissue engineering applications. Biomaterials. 2002 Feb;23(4):1169–1185. doi: 10.1016/s0142-9612(01)00232-0. [DOI] [PubMed] [Google Scholar]
- 57.Wilson CE, de Bruijn JD, van Blitterswijk CA, Verbout AJ, Dhert WJA. Design and fabrication of standardized hydroxyapatite scaffolds with a defined macro-architecture by rapid prototyping for bone-tissue-engineering research. Journal of Biomedical Materials Research Part A. 2004 Jan 1;68A(1):123–132. doi: 10.1002/jbm.a.20015. [DOI] [PubMed] [Google Scholar]
- 58.Dean D, Bookstein FL, Koneru S, et al. Average African American three-dimensional computed tomography skull images: The potential clinical importance of ethnicity and sex. J Craniofac Surg. 1998 Jul;9(4):348–358. doi: 10.1097/00001665-199807000-00011. [DOI] [PubMed] [Google Scholar]
- 59.Dean D, Min KJ. Deformable templates for preoperative computer-aided design and fabrication of large cranial implants. Int Congr Ser. 2003;1256:710–715. [Google Scholar]
- 60.Subramanya K, Dean D. A procedure to average 3D anatomical structures. Med Image Anal. 2000 Dec;4(4):317–334. doi: 10.1016/s1361-8415(00)00031-1. [DOI] [PubMed] [Google Scholar]
- 61.Chepla KJ, Alleyne BJ, Gosain AK. Primary correction of nasal asymmetry in patients with unilateral coronal synostosis. Plast Reconstr Surg. 2014;134(2):294–300. doi: 10.1097/PRS.0000000000000368. [DOI] [PubMed] [Google Scholar]
- 62.Min KJ, Dean D. Highly accurate CAD tools for cranial implants. Lect Notes Comput Sc. 2003;2878:99–107. [Google Scholar]
- 63.Min KJ, Dean D. Surface smoothing and template partitioning for cranial implant CAD. P Soc Photo-Opt Ins. 2005;5744:492–503. [Google Scholar]
- 64.Eufinger H, Wehmoller M, Falk A, Schneider T, Gellrich NC. Helical CT and freeform surfaces geometric design for the prefabrication of individual cranial prostheses. Computer Assisted Radiology. 1995:1229–1234. [Google Scholar]
- 65.Schebesch KM, Hohne J, Gassner HG, Brawanski A. Preformed titanium cranioplasty after resection of skull base meningiomas - a technical note. J Craniomaxillofac Surg. 2013 Dec;41(8):803–807. doi: 10.1016/j.jcms.2013.01.030. [DOI] [PubMed] [Google Scholar]
- 66.Zong G, Wu Y, Marcus HL, Bourell DL, Beaman JJ. Direct Solid Freeform Fabrication of High-Temperature Materials Using Selective Laser Sintering. Processing and Fabrication of Advanced Materials for High Temperature Applications Ii. 1993:503–518. [Google Scholar]
- 67.Sallica-Leva E, Jardini AL, Fogagnolo JB. Microstructure and mechanical behavior of porous Ti-6Al-4V parts obtained by selective laser melting. J Mech Behav Biomed Mater. 2013 Oct;26:98–108. doi: 10.1016/j.jmbbm.2013.05.011. [DOI] [PubMed] [Google Scholar]
- 68.Li X, Luo Y, Wang C, Zhang W, Li Y. Fabrication and in vivo evaluation of Ti6Al4V implants with controlled porous structure and complex shape. Frontiers of Mechanical Engineering. 2012;7(1):66–71. [Google Scholar]
- 69.Ponader S, von Wilmowsky C, Widenmayer M, et al. In vivo performance of selective electron beam-melted Ti-6Al-4V structures. Journal of Biomedical Materials Research Part A. 2010 Jan;92A(1):56–62. doi: 10.1002/jbm.a.32337. [DOI] [PubMed] [Google Scholar]
- 70.Cabraja M, Klein M, Lehmann TN. Long-term results following titanium cranioplasty of large skull defects. Neurosurgical focus. 2009 Jun;26(6):E10. doi: 10.3171/2009.3.FOCUS091. [DOI] [PubMed] [Google Scholar]
- 71.El Halabi F, Rodriguez JF, Rebolledo L, Hurtos E, Doblare M. Mechanical characterization and numerical simulation of polyether-ether-ketone (PEEK) cranial implants. J Mech Behav Biomed Mater. 2011 Nov;4(8):1819–1832. doi: 10.1016/j.jmbbm.2011.05.039. [DOI] [PubMed] [Google Scholar]
- 72.Jaekel DJ, MacDonald DW, Kurtz SM. Characterization of PEEK biomaterials using the small punch test. J Mech Behav Biomed Mater. 2011 Oct;4(7):1275–1282. doi: 10.1016/j.jmbbm.2011.04.014. [DOI] [PubMed] [Google Scholar]
- 73.Wurm G, Tomancok B, Holl K, Trenkler J. Prospective study on cranioplasty with individual carbon fiber reinforced polymere (CFRP) implants produced by means of stereolithography. Surgical neurology. 2004 Dec;62(6):510–521. doi: 10.1016/j.surneu.2004.01.025. [DOI] [PubMed] [Google Scholar]
- 74.Saringer W, Nobauer-Huhmann I, Knosp E. Cranioplasty with individual carbon fibre reinforced polymere (CFRP) medical grade implants based on CAD/CAM technique. Acta Neurochirurgica. 2002;144(11):1193–1203. doi: 10.1007/s00701-002-0995-5. [DOI] [PubMed] [Google Scholar]
- 75.Lindner M, Hoeges S, Meiners W, et al. Manufacturing of individual biodegradable bone substitute implants using selective laser melting technique. Journal of Biomedical Materials Research Part A. 2011 Jun;97A(4):466–471. doi: 10.1002/jbm.a.33058. [DOI] [PubMed] [Google Scholar]
- 76.Hollinger J, Mark DE, Bach DE, Reddi AH, Seyfer AE. Calvarial Bone Regeneration Using Osteogenin. J Oral Maxil Surg. 1989 Nov;47(11):1182–1186. doi: 10.1016/0278-2391(89)90009-8. [DOI] [PubMed] [Google Scholar]
- 77.Ashammakhi N, Rokkanen P. Absorbable polyglycolide devices in trauma and bone surgery. Biomaterials. 1997 Jan;18(1):3–9. doi: 10.1016/s0142-9612(96)00107-x. [DOI] [PubMed] [Google Scholar]
- 78.Ashammakhi NA. Neomembranes: A concept review with special reference to self-reinforced polyglycolide membranes. Journal of Biomedical Materials Research. 1996 Win;33(4):297–303. doi: 10.1002/(SICI)1097-4636(199624)33:4<297::AID-JBM10>3.0.CO;2-C. [DOI] [PubMed] [Google Scholar]
- 79.Al Salamah L, Babay N. Guided bone regeneration using resorbable and nonresorbable membranes: A histologic evaluation in beagle dogs. J Dent Res. 2000 May;79(5):1279–1279. [Google Scholar]
- 80.Bos RRM, Rozema FR, Boering G, et al. Degradation of and Tissue Reaction to Biodegradable Poly(L-Lactide) for Use as Internal-Fixation of Fractures - a Study in Rats. Biomaterials. 1991 Jan;12(1):32–36. doi: 10.1016/0142-9612(91)90128-w. [DOI] [PubMed] [Google Scholar]
- 81.Agrawal CM, Kennedy ME, Micallef DM. The Effects of Ultrasound Irradiation on a Biodegradable 50-50-Percent Copolymer of Polylactic and Polyglycolic Acids. Journal of Biomedical Materials Research. 1994 Aug;28(8):851–859. doi: 10.1002/jbm.820280803. [DOI] [PubMed] [Google Scholar]
- 82.Peter SJ, Miller MJ, Yasko AW, Yaszemski MJ, Mikos AG. Polymer concepts in tissue engineering. Journal of Biomedical Materials Research. 1998 Win;43(4):422–427. doi: 10.1002/(sici)1097-4636(199824)43:4<422::aid-jbm9>3.0.co;2-1. [DOI] [PubMed] [Google Scholar]
- 83.Vonrecum HA, Cleek RL, Eskin SG, Mikos AG. Degradation of Polydispersed Poly(L-Lactic Acid) to Modulate Lactic-Acid Release. Biomaterials. 1995 Apr;16(6):441–447. doi: 10.1016/0142-9612(95)98816-w. [DOI] [PubMed] [Google Scholar]
- 84.Dawes E, Rushton N. Response of macrophages to poly(L-lactide) particulates which have undergone various degrees of artificial degradation. Biomaterials. 1997 Dec;18(24):1615–1623. doi: 10.1016/s0142-9612(97)00095-1. [DOI] [PubMed] [Google Scholar]
- 85.Gosain AK, Song LS, Corrao MA, Pintar FA. Biomechanical evaluation of titanium, biodegradable plate and screw, and cyanoacrylate glue fixation systems in craniofacial surgery. Plastic and Reconstructive Surgery. 1998 Mar;101(3):582–591. doi: 10.1097/00006534-199803000-00004. [DOI] [PubMed] [Google Scholar]
- 86.Lundgren AK, Lundgren D, Sennerby L, Taylor A, Gottlow J, Nyman S. Augmentation of skull bone using a bioresorbable barrier supported by autologous bone grafts - An intra-individual study in the rabbit. Clin Oral Implan Res. 1997 Apr;8(2):90–95. doi: 10.1034/j.1600-0501.1997.080203.x. [DOI] [PubMed] [Google Scholar]
- 87.Peltoniemi H, Ashammakhi N, Kontio R, et al. The use of bioabsorbable osteofixation devices in craniomaxillofacial surgery. Oral Surg Oral Med O. 2002 Jul;94(1):5–14. doi: 10.1067/moe.2002.122160. [DOI] [PubMed] [Google Scholar]
- 88.Probst FA, Hutmacher DW, Muller DF, Machens HG, Schantz JT. Calvarial Reconstruction by Customized Bioactive Implant. Handchir Mikrochir P. 2010 Dec;42(6):369–373. doi: 10.1055/s-0030-1248310. [DOI] [PubMed] [Google Scholar]
- 89.Lam CXF, Savalani MM, Teoh SH, Hutmacher DW. Dynamics of in vitro polymer degradation of polycaprolactone-based scaffolds: accelerated versus simulated physiological conditions. Biomed Mater. 2008 Sep;3(3) doi: 10.1088/1748-6041/3/3/034108. [DOI] [PubMed] [Google Scholar]
- 90.Carragee EJ, Hurwitz EL, Weiner BK. A critical review of recombinant human bone morphogenetic protein-2 trials in spinal surgery: emerging safety concerns and lessons learned. Spine J. 2011 Jun;11(6):471–491. doi: 10.1016/j.spinee.2011.04.023. [DOI] [PubMed] [Google Scholar]
- 91.Cray J, Jr, Henderson SE, Smith DM, et al. BMP-2-Regenerated Calvarial Bone: A Biomechanical Appraisal in a Large Animal Model. Ann Plast Surg. 2013 May 6; doi: 10.1097/SAP.0b013e31827f4c82. [DOI] [PubMed] [Google Scholar]
- 92.Sweeny L, Lancaster WP, Dean NR, et al. Use of Recombinant Bone Morphogenetic Protein 2 in Free Flap Reconstruction for Osteonecrosis of the Mandible. J Oral Maxil Surg. 2012 Aug;70(8):1991–1996. doi: 10.1016/j.joms.2011.08.037. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 93.Boyne PJ, Marx RE, Nevins M, et al. A feasibility study evaluating rhBMP-2/absorbable collagen sponge for maxillary sinus floor augmentation. Int J Periodont Rest. 1997 Feb;17(1):10–25. [PubMed] [Google Scholar]
- 94.Cochran DI, Jones AA, Lilly LC, Fiorellini JP, Howell H. Evaluation of recombinant human bone morphogenetic protein-2 in oral applications including the use of endosseous implants: 3-year results of a pilot study in humans. J Periodontol. 2000 Aug;71(8):1241–1257. doi: 10.1902/jop.2000.71.8.1241. [DOI] [PubMed] [Google Scholar]
- 95.Suarez-Gonzalez D, Barnhart K, Migneco F, Flanagan C, Hollister SJ, Murphy WL. Controllable mineral coatings on PCL scaffolds as carriers for growth factor release. Biomaterials. 2012 Jan;33(2):713–721. doi: 10.1016/j.biomaterials.2011.09.095. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 96.Rajala K, Lindroos B, Hussein SM, et al. A Defined and Xeno-Free Culture Method Enabling the Establishment of Clinical-Grade Human Embryonic, Induced Pluripotent and Adipose Stem Cells. PLoS One. 2010 Apr 19;5(4) doi: 10.1371/journal.pone.0010246. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 97.Thibault RA, Mikos AG, Kasper FK. Protein and Mineral Composition of Osteogenic Extracellular Matrix Constructs Generated with a Flow Perfusion Bioreactor. Biomacromolecules. 2011 Dec;12(12):4204–4212. doi: 10.1021/bm200975a. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 98.Wallace JWM, Thompson P, Busso M, Belle V, Mammoser N, Kim K, Fisher JP, Siblani A, Xu Y, Welter JF, Lennon DP, Sun J, Caplan AI, Dean D. Validating Continuous Digital Light Processing (cDLP) Additive Manufacturing Accuracy and Tissue Engineering Utility of a Dye-Initiator Package. Biofabrication. doi: 10.1088/1758-5082/6/1/015003. [DOI] [PubMed] [Google Scholar]
- 99.Thesleff T, Lehtimaki K, Niskakangas T, et al. Cranioplasty With Adipose-Derived Stem Cells and Biomaterial: A Novel Method for Cranial Reconstruction. Neurosurgery. 2011 Jun;68(6):1535–1540. doi: 10.1227/NEU.0b013e31820ee24e. [DOI] [PubMed] [Google Scholar]
- 100.Dunisch P, Walter J, Sakr Y, Kalff R, Waschke A, Ewald C. Risk factors of aseptic bone resorption: a study after autologous bone flap reinsertion due to decompressive craniotomy. Journal of neurosurgery. 2013 May;118(5):1141–1147. doi: 10.3171/2013.1.JNS12860. [DOI] [PubMed] [Google Scholar]
- 101.Barone CM, Jimenez DF, Boschert MT. Temporalis muscle resuspension using titanium miniplates and screws: technical note. Neurosurgery. 2001 Feb;48(2):450–451. doi: 10.1097/00006123-200102000-00049. [DOI] [PubMed] [Google Scholar]
- 102.Harris DA, Fong AJ, Buchanan EP, Monson L, Khechoyan D, Lam S. History of synthetic materials in alloplastic cranioplasty. Neurosurgical focus. 2014 Apr;36(4):E20. doi: 10.3171/2014.2.FOCUS13560. [DOI] [PubMed] [Google Scholar]
- 103.Vercler CJ, Sugg KB, Buchman SR. Split cranial bone grafting in children younger than 3 years old: debunking a surgical myth. Plastic and Reconstructive Surgery. 2014 Jun;133(6):822e–827e. doi: 10.1097/PRS.0000000000000222. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 104.Yadla S, Campbell PG, Chitale R, Maltenfort MG, Jabbour P, Sharan AD. Effect of early surgery, material, and method of flap preservation on cranioplasty infections: a systematic review. Neurosurgery. 2011;68(4):1124–9. doi: 10.1227/NEU.0b013e31820a5470. discussion 1130. [DOI] [PubMed] [Google Scholar]