Abstract
90Y PET/CT can be acquired after 90Y-microsphere selective radiation internal therapy (SIRT) to describe radioactivity distribution. We performed dosimetry using 90Y-microsphere PET/CT data to evaluate treatment efficacy and appropriateness of activity planning from 99mTc-MAA scan and SPECT/CT.
Twenty-three patients with liver malignancy were included in the study. 99mTc-MAA was injected during planning angiography and whole body 99mTc-MAA scan and liver SPECT/CT were acquired. After SIRT using 90Y-resin microsphere, 90Y-microsphere PET/CT was acquired. A partition model (PM) using 4 compartments (tumor, intarget normal liver, out-target normal liver, and lung) was adopted, and absorbed dose to each compartment was calculated based on measurements from 99mTc-MAA SPECT/CT and 90Y-microsphere PET/CT, respectively, to be compared with each other. Progression-free survival (PFS) was evaluated in terms of tumor absorbed doses calculated by 99mTc-MAA SPECT/CT and 90Y-microsphere PET/CT results.
Lung shunt fraction was overestimated on 99mTc-MAA scan compared with 90Y-microsphere PET/CT (0.060 ± 0.037 vs. 0.018 ± 0.026, P < 0.01). Tumor absorbed dose exhibited a close correlation between the results from 99mTc-MAA SPECT/CT and 90Y-microsphere PET/CT (r = 0.64, P < 0.01), although the result from 99mTc-MAA SPECT/CT was significantly lower than that from 90Y-microsphere PET/CT (135.4 ± 64.2 Gy vs. 185.0 ± 87.8 Gy, P < 0.01). Absorbed dose to in-target normal liver was overestimated on 99mTc-MAA SPECT/CT compared with PET/CT (62.6 ± 38.2 Gy vs. 45.2 ± 32.0 Gy, P = 0.02). Absorbed dose to out-target normal liver did not differ between 99mTc-MAA SPECT/CT and 90Y-microsphere PET/CT (P = 0.49). Patients with tumor absorbed dose >200 Gy on 90Y-microsphere PET/CT had longer PFS than those with tumor absorbed dose ≤200 Gy (286 ± 56 days vs. 92 ± 20 days, P = 0.046). Tumor absorbed dose calculated by 99mTc-MAA SPECT/CT was not a significant predictor for PFS.
Activity planning based on 99mTc-MAA scan and SPECT/CT can be effectively used as a conservative method. Post-SIRT dosimetry based on 90Y-microsphere PET/CT is an effective method to predict treatment efficacy.
INTRODUCTION
Selective internal radiation therapy (SIRT) using 90Y-microsphere is an effective treatment option for treating inoperable hepatic malignancy. Many studies have reported therapeutic efficacy of 90Y-microsphere SIRT,1–3 with various treatment response rates in hepatocellular carcinoma,4,5 and metastasis from colorectal cancer6 or neuroendocrine tumors.7 Because hepatic malignancy is supplied with blood mostly from the hepatic artery,890Y-microsphere can be selectively delivered to a target lesion by angiographic intervention. With a half-life of 2.67 days, 90Y emits β-particles with an average energy of 0.927 MeV. The average penetration range of the β-particles in tissue is 2.5 mm and 90% of the energy is absorbed within a sphere with a radius of 5.3 mm.9,10 This range is effective to deliver a large absorbed dose to a tumor while minimizing radiation hazard to the normal liver parenchyma, in case selective injection of 90Y-microspheres is successful.
Most crucial adverse effects in SIRT are related to radiation injury of the normal organs. Intrahepatic arterial shunt and consequent leakage of 90Y-microsphere to the lungs may cause radiation pneumonitis.11 Radiation injury of the normal liver may cause radioembolization-induced liver disease (REILD),12 which exhibits extensive sinusoidal congestion, hepatic atrophy, necrosis, and intimal fibrosis. To prevent these adverse effects and to determine activity requirement for treatment, planning angiography is performed before SIRT, in which 99mTc-macroaggregated albumin (MAA) is injected into the target artery. Therapeutic efficacy and REILD are closely related to the injection activity of 90Y-microsphere,12,13 and thus, several methods such as the empirical method, the body surface area (BSA) method, and the partition model (PM) method have been suggested for activity planning.10,14 Among those methods, the PM method based on 99mTc-MAA imaging is the most personalized one with a lower incidence of adverse effects,13,15 although some controversy exists regarding distribution equivalency between 99mTc-MAA and 90Y-resin microsphere.16 Recently, single photon emission computed tomography (SPECT)/computed tomography (CT) has been widely used in clinical practice, with advantage of correct localization and attenuation correction. 99mTc-MAA SPECT/CT has also been reported to be effective in activity planning for 90Y glass microsphere SIRT.17
After SIRT, the real distribution of 90Y-microsphere can be evaluated using bremsstrahlung scan or SPECT. Additionally, although 90Y emits a very small amount of positron (0.003%), positron emission tomography (PET)/CT for 90Y is also available with recent PET scanners, which are highly sensitive by adopting time-of-flight (TOF) algorithms.18–20 In this study, we performed activity planning for SIRT using 90Y-resin microsphere based on 99mTc-MAA scan and SPECT/CT. The appropriateness of the 99mTc-MAA imaging-based activity planning was evaluated in comparison with dosimetry by post-SIRT 90Y-microsphere PET/CT. Additionally, the efficacy of PET/CT-based dosimetry was also evaluated in terms of patient outcome. To the best of our knowledge, this is the first study applying 90Y-microsphere PET/CT to validate 99mTc-MAA imaging-based activity planning.
METHODS
Patients and Study Protocol
Patients who were candidates for SIRT were included in this study consecutively. Candidates visited Seoul National University Hospital, Seoul, Korea, from June 2012 to August 2014. Contrast-enhanced CT was performed in every patient for pretreatment evaluation. SIRT was considered for patients with a life expectancy of more than 3 months and a malignant hepatic tumor that was unresectable and inadequate for chemotherapy due to its size.
Patients underwent treatment-planning angiography combined with 99mTc-MAA injection. Patients with lung shunt fractions (LSF) superior to 0.2 were excluded from SIRT. Activity of 90Y-microsphere was planned based on 99mTc-MAA SPECT/CT. SIRT was performed approximately 1 week after planning angiography by the same intervention radiologist. When injection sites were significantly altered (changes above major branch of hepatic artery level) between planning angiography and 90Y-microsphere SIRT, the patient was excluded from the analysis. Treatment response to SIRT was evaluated by the response evaluation criteria in solid tumors (RECIST) version 1.1 and progression-free survival (PFS) was evaluated. Follow-up contrast-enhanced CT or magnetic resonance imaging (MRI) was obtained approximately every 2 months after SIRT and when a patient presented abnormal symptoms, signs, or serum tumor marker increase.
Planning Angiography and 99mTc-MAA Imaging
Planning angiography and injection of 99mTc-MAA (185 MBq) was performed by 1 interventional radiologist (H.C.K.), according to the previously published guidelines.2199mTc-MAA was injected into the supplying arteries split according to the approximate volume ratio of tumor proportion supplied by each artery, in case there were more than 2 tumor-feeding arteries for a tumor. After the injection of 99mTc-MAA, whole body scan and SPECT/CT were performed using a hybrid scanner combining a dual-head gamma camera and a 16-slice CT scanner (Discovery NM/CT 670, GE Healthcare, USA) equipped with low-energy high-resolution collimators. On whole body scan, conjugate anterior and posterior images were obtained over 10 minutes (table speed; 15 cm/minute) using 256 × 1024 matrices. SPECT images were acquired to cover the whole liver and the lower lung, by a step-and-shoot method (3° intervals) for 20 seconds per step. After SPECT acquisition, a helical CT scan was performed without using contrast agent and images were reconstructed into 3.75-mm slices. SPECT images were reconstructed on 128 × 128 matrices using an iterative algorithm (2 iterations, 10 subsets), including CT attenuation map-based attenuation correction, resolution recovery, and a postreconstruction Butterworth filter with a cut-off frequency 0.48 and order 5.
Selective Internal Radiation Therapy and 90Y-Microsphere PET/CT
The planned activity of 90Y-labeled resin microsphere (SIR-Spheres, Sirtex Medical, Australia) was injected through a microcatheter according to the same method as the planning angiography. Injected activity was primarily determined using the PM method, further described below. However, in some patients with high LSF (>0.10), activity was determined by the BSA method and reduced by up to 40% according to the manufacturer's package insert.22
Immediately after completion of SIRT, PET/CT images were obtained for 1 bed position including the low chest and the upper abdomen, using a large field-of-view PET/CT scanner (Biograph mCT64, Siemens Healthcare, Germany). CT images were acquired first in a spiral mode (pitch 1.2, 120 kVp, and 35 mAs). PET images were acquired for 10 minutes using a 3D mode. CT images were reconstructed using a conventional filtered back projection method, 50-cm field of view, 3.0-mm postprocessing thickness and 2.0-mm increment per slice. PET images were reconstructed on 200 × 200 matrices using an iterative method including algorithms for point spread function recovery and TOF calculation (2 iterations, 21 subsets) with CT-based attenuation correction.
Image Analysis for Activity Planning and Dosimetry
On 99mTc-MAA whole body scans, regions of interest (ROIs) were drawn for the lungs and liver, and LSF was calculated by total counts in the lungs (TClung) and the liver (TCliver) by Eq. (1)
For activity planning based on the PM method, 3 partitions were defined in the liver; tumor, in-target normal liver defined as the non-tumorous liver supplied by the target artery of SIRT, and out-target normal liver defined as the non-tumorous liver supplied by nontarget arteries of SIRT. On 99mTc-MAA SPECT/CT images, ROIs of these 3 partitions were manually drawn on every slice on CT images of fusion SPECT/CT, with reference to contrast-enhanced CT images (Figure 1) using an analysis software package (Xeleris 3, GE Healthcare, USA). Volume rendering was done to acquire volumes of interest (VOIs) of the 3 partitions.
Mean uptake counts per unit volume in the tumor (Ctm), in-target normal liver (Cin), and out-target normal liver (Cout) were measured in the VOIs, in addition to their volumes (Vtm, Vin, and Vout). Absorbed dose of each partition was basically calculated by Eq(2). 10,23,24:
where A0 is the initial radioactivity in the partition. Tissue density was assumed to be 1.0 kg/L for the liver and tumor, and 0.3 kg/L25 for the lungs. For tumor, A0 was calculated by Eq.(3):
A0 for the in-target normal or out-target normal liver was also calculated using the same method. The activity of 90Y-microsphere was planned to meet the tolerance limits of the lungs (<20 Gy), and the in-target normal liver (<70 Gy), as well as the absorbed dose requirement for tumor treatment (>120 Gy).26 In some cases whose volumes of the in-target normal liver were considerably small, tolerance limit of the out-target normal liver (<30 Gy) was considered instead of that of the in-target normal liver.27 In addition to the PM method, recommended activity of 90Y-microsphere was also calculated by using the conventional BSA method,3 for comparison with the results of PM method.
Post-SIRT 90Y-microsphere PET/CT images were analyzed using a vendor-supplied software package (Syngo.via, Siemens Healthcare, Germany) (Figure 1). According to the same method as was used for 99mTc-MAA SPECT/CT, VOIs were drawn for the lungs and 3 partitions in the liver for measuring mean uptake counts and volumes. Because PET/CT images were acquired for only 1 bed position, lung uptake was measured only in the scan-covered basal lungs, and total lung counts were presumed by multiplying lung volume and mean counts of the basal lungs included in PET/CT images. An average CT-based lung volume measurement value of 3.3 L was used. Absorbed dose to each partition was calculated from the measurements.
Statistics
Paired Student t test and χ2 test were used to compare dosimetry parameters between the results from 99mTc-MAA imaging and 90Y-microsphere PET/CT. Correlation between 2 parameters was evaluated using Pearson's correlation coefficient. For survival analysis, Kaplan–Meier survival analysis was used and PFS was compared between groups. In all statistical analyses, a commercial software package (MedCalc, Version 12.2.1.0, MedCalc Software, Belgium) was used and a P-value less than 0.05 was regarded as significant.
RESULTS
Patient Characteristics
90Y-microsphere SIRT was performed on 30 patients during the study period, among which 7 patients were excluded from the analysis because of major alterations in the injection sites between planning angiography and SIRT. Finally, 23 patients (M:F = 21:2, age 63.6 ± 12.4 years) were included in the analysis; 16 patients with hepatocellular carcinoma, 3 patients with cholangiocarcinoma, and 4 patients with metastatic liver mass from other cancers. Patient and tumor characteristics are described in Table 1. LSF was <10% in 19 patients, and 10–15% in 4 patients.
TABLE 1.
Activity Planning
Injection activity was determined by PM method in 21 patients, and by BSA method in 2 patients with high LSF. In all patients, tolerance limits for the lungs were higher than absorbed dose requirement for tumor treatment. However, tolerance limit for out-target normal liver was slightly lower than absorbed dose requirement in 1 patient, and injection activity was reduced by 14 %. In 6 patients, the full planned activity was not injected because of arterial flow stasis and reflux (n = 4) or unexpected schedule change (n = 2). PFS was able to be evaluated in 22 patients, excluding 1 patient who was missing from follow-up. In activity planning by the PM method, absorbed dose requirement for tumor absorbed dose of 120 Gy was calculated as 2.1 ± 0.9 GBq (range 0.8–3.8 GBq), whereas it was calculated as 1.7 ± 0.2 GBq (range 1.1–2.1 GBq) from the BSA method. The absorbed dose requirement by the PM method was significantly higher than that by the BSA method (P = 0.02). From the PM method, the tolerance limits for the lungs and out-target normal liver were 31.3 ± 32.0 GBq (range 8.1–146.2 GBq) and 10.5 ± 10.2 GBq (range 1.1–42.8 GBq), respectively.
SIRT and Post-SIRT Dosimetry
SIRT was performed 9 ± 5 days (range 1–15 days) after 99mTc-MAA scan and SPECT/CT. All except 1 patient underwent SIRT more than 48 hours after 99mTc-MAA injections to avoid possible embolization effect by MAA particles.28 The patient who underwent SIRT 1 day after 99mTc-MAA injection did not exhibit any flow stasis or reflux during 90Y-microsphere injection. In SIRT, 2.3 ± 1.2 GBq (range 0.3–3.9 GBq) of 90Y-microsphere was injected. With these actual injected activities, the dosimetry results were calculated by the measurements on 99mTc-MAA imaging and 90Y-microsphere PET/CT.
Absorbed dose of each partition is shown in Table 2. LSF was evaluated to be lower by 90Y-microsphere PET/CT than by 99mTc-MAA scan (P < 0.01). Absorbed dose of tumor was calculated lower by 99mTc-MAA imaging (P < 0.01), but absorbed doses of the in-target normal liver (P = 0.02) and the lungs (P < 0.01) by 99mTc-MAA imaging were higher than those by 90Y-microsphere PET/CT. There was no significant difference in absorbed dose of the out-target normal liver (P = 0.49). Significant correlations existed between the results by 99mTc-MAA imaging and 90Y-microsphere PET/CT, in absorbed doses of tumor (r = 0.64, P < 0.01), the in-target normal liver (r = 0.71, P < 0.001), and the lungs (r = 0.53, P < 0.01), but not in absorbed dose of the out-target normal liver (r = −0.18, P = 0.40) (Figure 2).
TABLE 2.
Correlation Between Tumor Outcome and Dosimetry
SIRT was successfully performed in all patients without SIRT-related complication. Overall PFS of the patients was 207 ± 22 days. When patients were classified by tumor absorbed dose evaluated by 90Y-microsphere PET/CT, the groups exhibited a significant difference in PFS. In patients who had high absorbed dose (>200 Gy, n = 14), PFS was 286 ± 56 days (range 53–383 days), whereas it was 92 ± 20 days (range 18–170 days) in patients who had low absorbed dose (≤ 200 Gy, n = 8) (P = 0.046; Figure 3A). Between these 2 groups, there were no significant differences in other clinicopathologic factors that can affect PFS 29–31; tumor burden (%, P = 0.16), previous history of chemotherapy (P = 0.47), previous history of surgery (P = 0.23), extrahepatic metastases (P = 0.84), tumor type (P = 0.13), sex (P = 0.77), abnormal pre-SIRT total bilirubin (>1.3 mg/dL, P = 0.80), abnormal pre-SIRT serum albumin (<3.5 g/dL, P = 0.69), and poor performance status (ECOG grade >0, P = 0.84). Representative cases are shown in Figure 4.
According to tumor absorbed dose evaluated by 99mTc-MAA SPECT/CT, only 3 patients received dose more than 200 Gy and statistical analysis was unavailable. When patients were classified by tumor absorbed dose of 120 Gy or 150 Gy evaluated by 99mTc-MAA SPECT/CT, the 2 groups did not exhibit a significant difference in PFS (Figure 3B, C).
DISCUSSION
In this study, we demonstrated that activity planning based on 99mTc-MAA scan and SPECT/CT is closely related to post-SIRT dosimetry using 90Y-microsphere PET/CT, although there was a significant difference in the calculated values between the 2 imaging methods. By SPECT/CT, tolerance limits for the lung and liver were evaluated to be lower, and the absorbed dose requirement for tumor treatment was evaluated to be higher than by PET/CT, which results in narrower dose margin for activity planning. Thus, 99mTc-MAA SPECT/CT can be used as a conservative activity planning method. Additionally, our study suggests that 90Y-microsphere PET/CT is an effective method for post-SIRT dosimetry and prediction of treatment efficacy.
In SIRT using 90Y-microsphere, post-SIRT dosimetry and efficacy has often been evaluated using 90Y bremsstrahlung scan or SPECT images.16,32 However, bremsstrahlung X-rays have a low count rate and a wide range of energy (50–250 keV) without an energy peak. Thus, the quality of bremsstrahlung imaging is still not satisfactory for accurate quantitative analysis, despite several attempts for optimizing reconstruction algorithm.33,34 Recently, PET imaging for 90Y has been available with PET scanners of enhanced sensitivity by adopting new reconstruction algorithms such as TOF calculation. Although the yield of positron emission from 90Y is very low, several previous studies have reported the feasibility of 90Y PET imaging and PET-based dosimetry.35,36 The activity distribution and counts on PET images are well correlated with the real activity measured in phantom studies.36,37 In the present study, we obtained 90Y-microsphere PET/CT images after SIRT and conducted dosimetry using the PM method, while previous studies performed voxel-wise dosimetry.36,38,39 Although the PM method does not provide information on heterogeneous dose distribution, it can be easily performed in clinical practice. Additionally, we defined 3 partitions in the liver by differentiating tumor, in-target normal liver, and out-target normal liver. Because radioactivity distribution is different between tumor and normal liver even if they are supplied by the same artery, this 3-partition model would provide more appropriate dosimetry results than a simple 2-partition model defining only tumor and normal liver.
In activity planning for 90Y-microsphere SIRT, simple methods such as empirical and BSA methods have a crucial limitation that individual condition of tumor size or arterial supply is not considered. PM methods are based on pretreatment 99mTc-MAA imaging and can be a practical option for individualized planning. Because 99mTc-MAA and 90Y resin microspheres theoretically do not redistribute after initial distribution, the effective half-life in a tissue is assumed identical to the physical half-life. Thus, dosimetry can easily be performed from the initial distribution of radioactivity. Recent application of SPECT/CT enables anatomical localization and more correct dosimetry compared with planar scan.
In the present study, significant differences existed between dosimetry results from pre-SIRT 99mTc-MAA imaging and post-SIRT 90Y-microsphere PET/CT, for the lungs, in-target normal liver, and tumor. Despite the correlation, absolute values of the absorbed doses were overestimated for the lungs and in-target normal liver, whereas it was underestimated for the tumor by 99mTc-MAA imaging, compared with PET/CT. There have been studies reporting discrepancy between 99mTc-MAA and 90Y-microsphere imaging.16,32,40 The discrepancy was attributed to differences in particle distribution, caused by different particle features such as size, density, and injected amount.16 It also needs to be considered that catheter location and injection sites can be different between planning angiography and SIRT, even if an operator conducts the same procedures. Moreover, image characteristics of SPECT and PET such as resolution and sensitivity may be other causes for the difference. On 99mTc-MAA scan and SPECT/CT, there is considerable spill-over effect from tumor to the lungs or adjacent normal liver. However, in spite of this limitation, 99mTc-MAA SPECT/CT can be an effective tool for activity planning, because tolerance limit and absorbed dose requirement for tumor treatment evaluated on SPECT/CT may be regarded as relatively conservative values.
Additionally, tolerance limit in SIRT appears to be different from those in external beam radiation therapy, in which absorbed dose exceeding 30 Gy may result in radiation hepatitis.27 In our study, 4 patients received excess radiation to in-target normal liver (>70 Gy), 2 patients received excess radiation to out-target normal liver (>30 Gy), and 1 patient received excess radiation both to in-target and out-target normal liver. However, there was no significant hepatic complication or hepatic enzyme elevation in these patients (data not shown). In SIRT, several factors have been suggested as risk factors for REILD, including age, tumor type, tumor volume, and delivered activity.12,13 However, tolerance limit for normal liver has been different among different studies,17,26 and no single cutoff value has been used to prevent REILD. Further studies are required to determine optimal tolerance limit in 90Y-microsphere SIRT, based on accurate activity planning and dosimetry.
We adopted 120 Gy as the requirement for tumor treatment in planning. It is based on a previous study,41 in which radioactivity of a tumor was intraoperatively measured on the tissue surface using a beta probe. However, there is a considerable difference in absorbed doses between the central and peripheral portions of a tumor.36,42 Thus, radioactivity measurement on tissue surface using a beta probe may have underestimated the actual absorbed dose. In another study, tumor absorbed dose of 205 Gy in activity planning was suggested as a cutoff value for effective treatment.17 However, although it was reported that dosimetry based on 99mTc-MAA SPECT/CT can be used to predict tumor response and survival in patients receiving 90Y-microsphere SIRT,17 tumor absorbed dose of 200 Gy measured on 90Y-microsphere PET/CT was a significant value to predict PFS, whereas tumor-absorbed dose of 120 Gy or 150 Gy in 99mTc-MAA SPECT/CT-based dosimetry was not in our study. In spite of significant correlations between tumor-absorbed doses between measurements from 99mTc-MAA SPECT/CT and 90Y-microsphere PET/CT, the dosimetry results based on 99mTc-MAA SPECT/CT were not a significant prognostic factor probably due to some patients whose results were discrepant between the 2 imaging methods. Therefore, it is suggested that post-SIRT PET/CT scan for 90Y-microsphere is necessary for evaluating treatment efficacy and predicting prognosis. Further studies including more patients are required to determine optimal absorbed dose requirement for tumor treatment in 90Y-microsphere SIRT.
There are several limitations in our study. First, the study design was not prospective, and the injection sites were not exactly controlled between planning dosimetry and SIRT. Although we excluded cases in which injection sites were significantly altered, it can be a considerable limitation to compare the dosimetry between 99mTc-MAA SPECT/CT and 90Y-microsphere PET/CT. Second, we evaluated pre-SIRT LSF on 99mTc-MAA planar scan, because it covers whole lung field and is widely used in current clinical practice. However, attenuation or spill-over effect may have affected the results. In contrast, post-SIRT LSF was evaluated on 90Y-microsphere PET/CT that covers only the basal lungs. The differences in imaging method may have been a cause for the different dose results. Third, the enrolled case number was relatively small, despite statistical significance. Based on the present study, further studies including more cases are required to determine efficacy of post-SIRT 90Y-microsphere PET/CT and optimal absorbed dose requirement in SIRT.
CONCLUSION
In this study, we demonstrated that activity planning based on 99mTc-MAA SPECT/CT is closely related to the post-SIRT dosimetry based on 90Y-microsphere PET/CT, although there were still differences in the calculated doses from the 2 imaging methods. On SPECT/CT, the tolerance limit for the lung or liver was lower, and the absorbed dose requirement for tumor treatment was higher than on PET/CT, which results in narrower activity margin for effective and safe SIRT. Thus, 99mTc-MAA SPECT/CT can be used as a conservative activity planning method. Additionally, our study suggests that post-SIRT 90Y-microsphere PET/CT is an effective dosimetry method and can be used to predict treatment efficacy.
Acknowledgments
This study was supported by a National Research Foundation of Korea grant for the Global Core Research Center (GCRC) funded by the Korea government (Ministry of Science, ICT and Future Planning) (No. 2011-0030680), and by Research Resettlement Fund for the new faculty of Seoul National University.
Footnotes
Abbreviations: BSA = body surface area, LSF = lung shunt fraction, MAA = macroaggregated albumin, PET = positron emission tomography, PFS = progression-free survival, PM = partition model, RECIST = response evaluation criteria in solid tumors, REILD = radioembolization-induced liver disease, ROI = regions of interest, SIRT = selective internal radiation therapy, SPECT = single photon emission computed tomography, TOF = time-of-flight, VOI = volumes of interest.
All human studies were approved by the local institutional review board and were therefore performed in accordance with the ethical standards laid down in the 1964 Declaration of Helsinki and its later amendments. The study was approved by the Institutional Review Board of our institution (H-1412-085-633).
The authors have no conflicts of interest to disclose.
REFERENCES
- 1.Kennedy A, Coldwell D, Sangro B, et al. Radioembolization for the treatment of liver tumors general principles. Am J Clin Oncol 2012; 35:91–99. [DOI] [PubMed] [Google Scholar]
- 2.Sangro B, Salem R, Kennedy A, et al. Radioembolization for hepatocellular carcinoma: a review of the evidence and treatment recommendations. Am J Clin Oncol 2011; 34:422–431. [DOI] [PubMed] [Google Scholar]
- 3.Murthy R, Kamat P, Nunez R, et al. Radioembolization of yttrium-90 microspheres for hepatic malignancy. Semin Intervent Radiol 2008; 25:48–57. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Carr BI. Hepatic arterial 90Yttrium glass microspheres (Therasphere) for unresectable hepatocellular carcinoma: interim safety and survival data on 65 patients. Liver Transpl 2004; 10 (2 Suppl 1):S107–S110. [DOI] [PubMed] [Google Scholar]
- 5.Dancey JE, Shepherd FA, Paul K, et al. Treatment of nonresectable hepatocellular carcinoma with intrahepatic 90Y-microspheres. J Nucl Med 2000; 41:1673–1681. [PubMed] [Google Scholar]
- 6.Jakobs TF, Hoffmann RT, Dehm K, et al. Hepatic yttrium-90 radioembolization of chemotherapy-refractory colorectal cancer liver metastases. J Vasc Interv Radiol 2008; 19:1187–1195. [DOI] [PubMed] [Google Scholar]
- 7.Kennedy AS, Dezarn WA, McNeillie P, et al. Radioembolization for unresectable neuroendocrine hepatic metastases using resin 90Y-microspheres: early results in 148 patients. Am J Clin Oncol 2008; 31:271–279. [DOI] [PubMed] [Google Scholar]
- 8.Ackerman NB, Lien WM, Kondi ES, et al. The blood supply of experimental liver metastases. I. The distribution of hepatic artery and portal vein blood to “small” and “large” tumors. Surgery 1969; 66:1067–1072. [PubMed] [Google Scholar]
- 9.Simpkin DJ, Mackie TR. EGS4 Monte Carlo determination of the beta dose kernel in water. Med Phys 1990; 17:179–186. [DOI] [PubMed] [Google Scholar]
- 10.Dezarn WA, Cessna JT, DeWerd LA, et al. Recommendations of the American Association of Physicists in Medicine on dosimetry, imaging, and quality assurance procedures for 90Y microsphere brachytherapy in the treatment of hepatic malignancies. Med Phys 2011; 38:4824–4845. [DOI] [PubMed] [Google Scholar]
- 11.Leung TW, Lau WY, Ho SK, et al. Radiation pneumonitis after selective internal radiation treatment with intraarterial 90yttrium-microspheres for inoperable hepatic tumors. Int J Radiat Oncol Biol Phys 1995; 33:919–924. [DOI] [PubMed] [Google Scholar]
- 12.Sangro B, Gil-Alzugaray B, Rodriguez J, et al. Liver disease induced by radioembolization of liver tumors: description and possible risk factors. Cancer 2008; 112:1538–1546. [DOI] [PubMed] [Google Scholar]
- 13.Kennedy AS, McNeillie P, Dezarn WA, et al. Treatment parameters and outcome in 680 treatments of internal radiation with resin 90Y-microspheres for unresectable hepatic tumors. Int J Radiat Oncol Biol Phys 2009; 74:1494–1500. [DOI] [PubMed] [Google Scholar]
- 14.Cremonesi M, Chiesa C, Strigari L, et al. Radioembolization of hepatic lesions from a radiobiology and dosimetric perspective. Front Oncol 2014; 4:210. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Kao YH, Tan EH, Ng CE, et al. Clinical implications of the body surface area method versus partition model dosimetry for yttrium-90 radioembolization using resin microspheres: a technical review. Ann Nucl Med 2011; 25:455–461. [DOI] [PubMed] [Google Scholar]
- 16.Wondergem M, Smits ML, Elschot M, et al. 99mTc-macroaggregated albumin poorly predicts the intrahepatic distribution of 90Y resin microspheres in hepatic radioembolization. J Nucl Med 2013; 54:1294–1301. [DOI] [PubMed] [Google Scholar]
- 17.Garin E, Lenoir L, Rolland Y, et al. Dosimetry based on 99mTc-macroaggregated albumin SPECT/CT accurately predicts tumor response and survival in hepatocellular carcinoma patients treated with 90Y-loaded glass microspheres: preliminary results. J Nucl Med 2012; 53:255–263. [DOI] [PubMed] [Google Scholar]
- 18.Gates VL, Esmail AA, Marshall K, et al. Internal pair production of 90Y permits hepatic localization of microspheres using routine PET: proof of concept. J Nucl Med 2011; 52:72–76. [DOI] [PubMed] [Google Scholar]
- 19.Lhommel R, Goffette P, Van den Eynde M, et al. Yttrium-90 TOF PET scan demonstrates high-resolution biodistribution after liver SIRT. Eur J Nucl Med Mol Imaging 2009; 36:1696. [DOI] [PubMed] [Google Scholar]
- 20.Walrand S, Flux GD, Konijnenberg MW, et al. Dosimetry of yttrium-labelled radiopharmaceuticals for internal therapy: 86Y or 90Y imaging? Eur J Nucl Med Mol Imaging 2011; 38 (Suppl 1):S57–S68. [DOI] [PubMed] [Google Scholar]
- 21.Salem R, Lewandowski RJ, Sato KT, et al. Technical aspects of radioembolization with 90Y microspheres. Tech Vasc Interv Radiol 2007; 10:12–29. [DOI] [PubMed] [Google Scholar]
- 22.Kwok PC, Leung KC, Cheung MT, et al. Survival benefit of radioembolization for inoperable hepatocellular carcinoma using yttrium-90 microspheres. J Gastroenterol Hepatol 2014; 29:1897–1904. [DOI] [PubMed] [Google Scholar]
- 23.Ho S, Lau WY, Leung TW, et al. Partition model for estimating radiation doses from yttrium-90 microspheres in treating hepatic tumours. Eur J Nucl Med 1996; 23:947–952. [DOI] [PubMed] [Google Scholar]
- 24.Ng SC, Lee VH, Law MW, et al. Patient dosimetry for 90Y selective internal radiation treatment based on 90Y PET imaging. J Appl Clin Med Phys 2013; 14:212–221. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Brudin LH, Rhodes CG, Valind SO, et al. Regional lung density and blood volume in nonsmoking and smoking subjects measured by PET. J Appl Physiol (1985) 1987; 63:1324–1334. [DOI] [PubMed] [Google Scholar]
- 26.Lau WY, Kennedy AS, Kim YH, et al. Patient selection and activity planning guide for selective internal radiotherapy with yttrium-90 resin microspheres. Int J Radiat Oncol Biol Phys 2012; 82:401–407. [DOI] [PubMed] [Google Scholar]
- 27.Lawrence TS, Robertson JM, Anscher MS, et al. Hepatic toxicity resulting from cancer treatment. Int J Radiat Oncol Biol Phys 1995; 31:1237–1248. [DOI] [PubMed] [Google Scholar]
- 28.Amor-Coarasa A, Milera A, Carvajal D, et al. Lyophilized kit for the preparation of the PET perfusion agent [(68)Ga]-MAA. Int J Mol Imaging 2014; 2014:269365. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Hoffmann RT, Paprottka PM, Schon A, et al. Transarterial hepatic yttrium-90 radioembolization in patients with unresectable intrahepatic cholangiocarcinoma: factors associated with prolonged survival. Cardiovasc Intervent Radiol 2012; 35:105–116. [DOI] [PubMed] [Google Scholar]
- 30.Dunfee BL, Riaz A, Lewandowski RJ, et al. Yttrium-90 radioembolization for liver malignancies: prognostic factors associated with survival. J Vasc Interv Radiol 2010; 21:90–95. [DOI] [PubMed] [Google Scholar]
- 31.Saxena A, Chua TC, Bester L, et al. Factors predicting response and survival after yttrium-90 radioembolization of unresectable neuroendocrine tumor liver metastases: a critical appraisal of 48 cases. Ann Surg 2010; 251:910–916. [DOI] [PubMed] [Google Scholar]
- 32.Fabbri C, Sarti G, Agostini M, et al. SPECT/CT 90Y-Bremsstrahlung images for dosimetry during therapy. Ecancermedicalscience 2008; 2:106. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Minarik D, Sjogreen Gleisner K, Ljungberg M. Evaluation of quantitative (90)Y SPECT based on experimental phantom studies. Phys Med Biol 2008; 53:5689–5703. [DOI] [PubMed] [Google Scholar]
- 34.Rong X, Ghaly M, Frey EC. Optimization of energy window for 90Y bremsstrahlung SPECT imaging for detection tasks using the ideal observer with model-mismatch. Med Phys 2013; 40:062502. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Lhommel R, van Elmbt L, Goffette P, et al. Feasibility of 90Y TOF PET-based dosimetry in liver metastasis therapy using SIR-Spheres. Eur J Nucl Med Mol Imaging 2010; 37:1654–1662. [DOI] [PubMed] [Google Scholar]
- 36.D’Arienzo M, Chiaramida P, Chiacchiararelli L, et al. 90Y PET-based dosimetry after selective internal radiotherapy treatments. Nucl Med Commun 2012; 33:633–640. [DOI] [PubMed] [Google Scholar]
- 37.Elschot M, Vermolen BJ, Lam MG, et al. Quantitative comparison of PET and Bremsstrahlung SPECT for imaging the in vivo yttrium-90 microsphere distribution after liver radioembolization. PLoS One 2013; 8:e55742. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Kao YH, Steinberg JD, Tay YS, et al. Post-radioembolization yttrium-90 PET/CT-part 1: diagnostic reporting. EJNMMI Res 2013; 3:56. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Kao YH, Steinberg JD, Tay YS, et al. Post-radioembolization yttrium-90 PET/CT-part 2: dose-response and tumor predictive dosimetry for resin microspheres. EJNMMI Res 2013; 3:57. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Knesaurek K, Machac J, Muzinic M, et al. Quantitative comparison of yttrium-90 (90Y)-microspheres and technetium-99m (99mTc)-macroaggregated albumin SPECT images for planning 90Y therapy of liver cancer. Technol Cancer Res Treat 2010; 9:253–262. [DOI] [PubMed] [Google Scholar]
- 41.Lau WY, Leung WT, Ho S, et al. Treatment of inoperable hepatocellular carcinoma with intrahepatic arterial yttrium-90 microspheres: a phase I and II study. Br J Cancer 1994; 70:994–999. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Sarfaraz M, Kennedy AS, Lodge MA, et al. Radiation absorbed dose distribution in a patient treated with yttrium-90 microspheres for hepatocellular carcinoma. Med Phys 2004; 31:2449–2453. [DOI] [PubMed] [Google Scholar]