. 2015 Jul 17;4(5):764–773. doi: 10.1002/mbo3.277

Table 3.

Estimated 6-log10 reduction times and approximate standard errors for spores from Bacillus subtilis, Bacillus anthracis, and Clostridium sporogenes using the three different disinfectants

Species Disinfectant Heat-shock 6 log10 reduction time (min)
Estimate SE
Bacillus anthracis CIDEX™ 4.37 0.16
+ 4.14 0.19
STERIPLEX™ HC 4.04 0.18
+ 3.50 0.14
STERIPLEX™ Ultra 0.72 0.04
+ 0.57 0.03
Bacillus subtilis CIDEX™ 214.04 51.65
+ 187.57 24.32
STERIPLEX™ HC 0.86 0.06
+ 0.98 0.07
STERIPLEX™ Ultra 0.22 0.01
+ 0.26 0.01
Clostridium sporogenes CIDEX™ 23.22 1.40
+ 23.20 1.43
STERIPLEX™ HC 0.21 0.01
+ 0.21 0.01
STERIPLEX™ Ultra 0.21 0.01
+ 0.21 0.01