Abstract
A radiotracer technique has been used to assess aerosol delivery from a pressurised metered dose inhaler, used both with and without a 10 cm cylindrical spacer attachment (Syncroner), which has an open section in its upper surface. The radionuclide technetium-99m (99mTc) was added to sodium cromoglycate in a canister (Intal inhaler; 1 mg/puff); in vitro studies with a multistage liquid impinger showed that the radiolabel acted as a marker for the presence of drug over a wide range of particle sizes. Ten healthy volunteers were studied after they had inhaled from (1) a metered dose inhaler alone (slow inhaled flow rate, about 25 l/min); (2) metered dose inhaler plus spacer (slow flow rate); and (3) metered dose inhaler plus spacer (fast inhaled flow rate, about 100 l/min). Inhalation was coordinated with firing the spray and was followed by 10 seconds' breath holding. With the metered dose inhaler alone a mean 11.0% (SEM 1.4%) of the dose reached the lungs, compared with significantly higher doses for slow (16.1% (2.2%] and fast (13.3% (1.7%] inhalations through the spacer. The distribution pattern within the lungs was significantly more peripheral after slow inhalation. Oropharyngeal deposition was halved by the spacer. The open spacer should teach patients good coordination and delivers more aerosol to the lungs than a correctly used metered dose inhaler.
Full text
PDF




Images in this article
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Coady T. J., Stewart C. J., Davies H. J. Synchronization of bronchodilator release. Practitioner. 1976 Aug;217(1298):273–275. [PubMed] [Google Scholar]
- Crompton G. K. Inhalation devices. Eur J Respir Dis. 1982 Nov;63(6):489–492. [PubMed] [Google Scholar]
- Crompton G. K. Problems patients have using pressurized aerosol inhalers. Eur J Respir Dis Suppl. 1982;119:101–104. [PubMed] [Google Scholar]
- Dolovich M., Ruffin R. E., Roberts R., Newhouse M. T. Optimal delivery of aerosols from metered dose inhalers. Chest. 1981 Dec;80(6 Suppl):911–915. [PubMed] [Google Scholar]
- Dolovich M., Ruffin R., Corr D., Newhouse M. T. Clinical evaluation of a simple demand inhalation MDI aerosol delivery device. Chest. 1983 Jul;84(1):36–41. doi: 10.1378/chest.84.1.36. [DOI] [PubMed] [Google Scholar]
- Epstein S. W., Manning C. P., Ashley M. J., Corey P. N. Survey of the clinical use of pressurized aerosol inhalers. Can Med Assoc J. 1979 Apr 7;120(7):813–816. [PMC free article] [PubMed] [Google Scholar]
- Köhler D., Fleischer W., Matthys H. New method for easy labeling of beta-2-agonists in the metered dose inhaler with technetium 99m. Respiration. 1988;53(2):65–73. doi: 10.1159/000195399. [DOI] [PubMed] [Google Scholar]
- Lawford P., McKenzie D. Pressurized aerosol inhaler technique: how important are inhalation from residual volume, inspiratory flow rate and the time interval between puffs? Br J Dis Chest. 1983 Jul;77(3):276–281. [PubMed] [Google Scholar]
- Lindgren S., Bake B., Larsson S. Clinical consequences of inadequate inhalation technique in asthma therapy. Eur J Respir Dis. 1987 Feb;70(2):93–98. [PubMed] [Google Scholar]
- Lippmann M., Albert R. E. The effect of particle size on the regional deposition of inhaled aerosols in the human respiratory tract. Am Ind Hyg Assoc J. 1969 May-Jun;30(3):257–275. doi: 10.1080/00028896909343120. [DOI] [PubMed] [Google Scholar]
- Morén F., Andersson J., Stiksa G. Mode of inhalation for trained and untrained asthmatics using a pressurized aerosol. Respiration. 1984;45(1):56–60. doi: 10.1159/000194598. [DOI] [PubMed] [Google Scholar]
- Newman S. P., Millar A. B., Lennard-Jones T. R., Morén F., Clarke S. W. Improvement of pressurised aerosol deposition with Nebuhaler spacer device. Thorax. 1984 Dec;39(12):935–941. doi: 10.1136/thx.39.12.935. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Newman S. P., Morén F., Pavia D., Little F., Clarke S. W. Deposition of pressurized suspension aerosols inhaled through extension devices. Am Rev Respir Dis. 1981 Sep;124(3):317–320. doi: 10.1164/arrd.1981.124.3.317. [DOI] [PubMed] [Google Scholar]
- Newman S. P., Pavia D., Clarke S. W. How should a pressurized beta-adrenergic bronchodilator be inhaled? Eur J Respir Dis. 1981 Feb;62(1):3–21. [PubMed] [Google Scholar]
- Newman S. P., Pavia D., Garland N., Clarke S. W. Effects of various inhalation modes on the deposition of radioactive pressurized aerosols. Eur J Respir Dis Suppl. 1982;119:57–65. [PubMed] [Google Scholar]
- Newman S. P., Woodman G., Clarke S. W. Deposition of carbenicillin aerosols in cystic fibrosis: effects of nebuliser system and breathing pattern. Thorax. 1988 Apr;43(4):318–322. doi: 10.1136/thx.43.4.318. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Newman S. P., Woodman G., Clarke S. W., Sackner M. A. Effect of InspirEase on the deposition of metered-dose aerosols in the human respiratory tract. Chest. 1986 Apr;89(4):551–556. doi: 10.1378/chest.89.4.551. [DOI] [PubMed] [Google Scholar]
- Pavia D., Thomson M. L., Clarke S. W., Shannon H. S. Effect of lung function and mode of inhalation on penetration of aerosol into the human lung. Thorax. 1977 Apr;32(2):194–197. doi: 10.1136/thx.32.2.194. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Pedersen S., Frost L., Arnfred T. Errors in inhalation technique and efficiency in inhaler use in asthmatic children. Allergy. 1986 Feb;41(2):118–124. doi: 10.1111/j.1398-9995.1986.tb00287.x. [DOI] [PubMed] [Google Scholar]
- Spiro S. G., Singh C. A., Tolfree S. E., Partridge M. R., Short M. D. Direct labelling of ipratropium bromide aerosol and its deposition pattern in normal subjects and patients with chronic bronchitis. Thorax. 1984 Jun;39(6):432–435. doi: 10.1136/thx.39.6.432. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Woodcock A. Training aid for pressurized inhalers. Br J Dis Chest. 1980 Oct;74(4):395–397. [PubMed] [Google Scholar]