Skip to main content
Dentomaxillofacial Radiology logoLink to Dentomaxillofacial Radiology
. 2015 Feb 11;44(5):20140322. doi: 10.1259/dmfr.20140322

Comparison of mandibular bone microarchitecture between micro-CT and CBCT images

S Panmekiate 1, N Ngonphloy 1, T Charoenkarn 1, T Faruangsaeng 1, R Pauwels 1,2,
PMCID: PMC4628495  PMID: 25564887

Abstract

Objectives:

To compare microarchitecture parameters of bone samples scanned using micro-CT (µCT) to those obtained by using CBCT.

Methods:

A bone biopsy trephine bur (3 × 10 mm) was used to remove 20 cylindrical bone samples from 20 dry hemimandibles. Samples were scanned using µCT (µCT 35; SCANCO Medical, Brüttisellen, Switzerland) with a voxel size of 20 µm and CBCT (3D Accuitomo 170; J. Morita, Kyoto, Japan) with a voxel size of 80 µm. All corresponding sample scans were aligned and cropped. Image analysis was carried out using BoneJ, including the following parameters: skeleton analysis, bone surface per total volume (BS/TV), bone volume per total volume (BV/TV), connectivity density, anisotropy, trabecular thickness and spacing, structure model index, plateness and fractal dimension. Pearson and Spearman correlation coefficients (R) were calculated. CBCT values were then calibrated using the slope of the linear fit with the µCT values. The mean error after calibration was calculated and normalized to the standard deviation of the µCT values.

Results:

R-values ranged between 0.05 (plateness) and 0.83 (BS/TV). Correlation was significant for both Spearman and Pearson’s R for 8 out of 16 parameters. After calibration, the smallest normalized error was found for BV/TV (0.48). For other parameters, the error range was 0.58–2.10.

Conclusions:

Despite the overall correlation, this study demonstrates the uncertainty associated with using bone microarchitecture parameters on CBCT images. Although clinically relevant parameter ranges are not available, the errors found in this study may be too high for some parameters to be considered for clinical application.

Keywords: cone-beam computed tomography, microcomputed tomography, bone, bone quality, computer-assisted image processing

Introduction

The success rate of oral implants has been proposed to pertain to bone quality, among other factors.15 The concept of bone quality has recently evolved from a density-based approach towards a structural approach, as it has been shown that trabecular bone microarchitecture affects implant stability.6 The pre-operative analysis of trabecular structure could therefore predict patient outcome and be used to select (un)favourable implant sites.

Conventional histomorphometry has been used as the gold standard for analysing bone microarchitecture, such as the percentage of bone tissue.7 However, this method is irreversible, takes a long time and is costly. In addition, it provides just two dimensions of the three-dimensional specimen. Micro-CT (µCT) has been validated8 and is presently implemented as the new gold standard method for quantifying bone microarchitecture.9 Nevertheless, this method is limited to the investigation of small animal (in vivo) or biopsy (ex vivo) specimens and cannot be applied for clinical imaging.

Currently, CBCT systems have evolved to offer images with higher resolution than do early generation CBCTs.10 High-resolution CBCTs use voxel sizes as small as 0.08 mm and are being used to visualize small anatomical details and pathologies in the oral region. This has raised the question whether CBCT images can be used to quantify trabecular microarchitecture on patient images. While it can be expected that CBCT and µCT images do not yield identical values for bone quality parameters because the latter has superior image sharpness, it might be possible to obtain stable bone microarchitecture measurements for CBCT images and to relate these values to clinical bone quality and thus, implant outcome or bone healing.

The aim of this study was to compare microarchitecture parameters of bone samples scanned using µCT to those obtained by using CBCT.

Methods and materials

The study protocol was approved by the human research ethics committee of the Faculty of Dentistry, Chulalongkorn University, Bangkok, Thailand (study code HREC-DCU-2013-058).

Sample preparation

20 cylindrical bone samples were removed from 20 cadaver hemimandibles that had edentulous ridges posterior to the mental foramen. A bone biopsy trephine bur (3 × 10 mm) was used for the removal. Samples were placed in Styrofoam holders (Dow Chemical Co., Midland, MI) for scanning by CBCT and µCT, in order to mimic the orientation of an implant site.

CBCT and micro-CT scanning

Samples were scanned using µCT (µCT 35; SCANCO Medical, Brüttisellen, Switzerland) with radiographic parameters of 70 kVp, 114 µA, 0.5-mm aluminium filter, 360° rotation, 450 basis projections and an isotropic voxel size of 20 µm. On the same day, samples were scanned by CBCT (3D Accuitomo 170; J. Morita, Kyoto, Japan) using a field of view of 4 × 4 cm, 90 kVp, 62 mAs, 2.5-mm aluminium filter, 360° rotation and 970 basis projections with an isotropic voxel size of 80 µm. For the CBCT, the mA used was 60% lower than that used for standard patients to avoid overexposure of the detector. Reconstructed images were exported as multifile image data and converted into single-file tag image files for analysis. Image analysis comprised four steps: alignment, cropping, thresholding and bone analysis (Figure 1).

Figure 1.

Figure 1

Workflow of image analysis. First, CBCT images were aligned along the z-axis by rotating the coronal and sagittal planes. Second, axial micro-CT (µCT) images were rotated using corresponding CBCT images as visual templates. Third, all images were cropped to 3.6 × 3.6 mm in the axial plane, and to corresponding sizes (according to the length of the bone sample) along the z-axis using visual matching (VM). Finally, all images were thresholded and underwent bone analysis.

Alignment of micro-CT and CBCT images

The samples were manually aligned in the coronal, sagittal and axial planes to ensure that the µCT and CBCT images were in the same orientation using ImageJ (US National Institutes of Health, Bethesda, MD). First, samples were aligned vertically in both the coronal and sagittal planes, ensuring that their long axis is parallel to these planes. For the axial plane, CBCT and µCT images were matched, that is, µCT images of each sample were rotated using the corresponding CBCT image as a visual template. Axial images after matching are shown in Figure 2. Alignment in all three planes was performed by two observers in order to minimize observer bias. When differences of >3° were found, the alignment was repeated until the deviation was <3°. The average rotation angles from the two observers, discarding angles differing >3°, were then selected for final alignment.

Figure 2.

Figure 2

Original and automatically thresholded images from micro-CT (µCT) and CBCT, aligned in the axial plane.

After alignment, all µCT and CBCT scans were tightly cropped to limit the amount of air space around the bone to a minimum. It was ensured that the total volume for each corresponding µCT and CBCT scans was identical. In the axial plane, this was carried out by manually cropping each image to a size of 3.6 × 3.6 mm (i.e. 180 × 180 voxels for µCT, 45 × 45 voxels for CBCT), ensuring that the cropped images fully encompassed the bone sample. In the coronal/sagittal planes (i.e. z-axis cropping) each sample was cropped according to its exact length, matching the cropped length for each corresponding CBCT and µCT image by use of the voxel size (e.g. 500 voxels for µCT, 125 voxels for CBCT = 10 mm in both images), and visually matching the start and end points of the cropped image.

Thresholding

To segment the bone from the images, they were thresholded and converted into a binary image (Figure 2). Various thresholding algorithms available in ImageJ were visually evaluated. The stack-based “Moments” method was selected, as it provided the most consistent bone segmentation.11 Figure 3 shows a binary CBCT sample image at various manually set threshold values as a comparison; it can be seen that it was not possible to manually obtain a better segmentation than with the automatic thresholding shown in Figure 2.

Figure 3.

Figure 3

CBCT image of Sample 20 (also shown in Figure 2, bottom row) at various threshold values.

Analysis of bone microarchitecture

Image analysis was carried out using ImageJ by means of the BoneJ plugin (http://www.bonej.org).12 The following parameters were included:

  • • bone surface per total volume (BS/TV)13

  • • bone volume per total volume (BV/TV)

  • • fractal dimension14

  • • connectivity density15,16

  • • anisotropy17,18

  • • trabecular thickness (Tb.Th) and spacing (Tb.Sp)19,20

  • • structure model index (SMI);21 both the overall SMI and its positive and negative components were calculated

  • • plateness, an alternative for SMI. Plateness is expressed as the ratio of eigenvalues (ev) along the longest (ev1), middle (ev2) and shortest (ev3) axis of the trabeculae

  • • skeleton analysis;22 the following parameters associated with the skeletonized image of the samples were included: number of branches, number of junctions, branch length and triple points.

Analysis of results and statistics

Statistical analysis was performed using Prism 5.01 (GraphPad Software, San Diego, CA). Interobserver agreement was evaluated by comparing the angles used to align the samples. The originally determined angles, including those for which deviations between observers was >3°, were used for this comparison. Spearman correlation was used, as the data did not pass the D'Agostino and Pearson omnibus normality test (p < 0.01). In addition, the Wilcoxon signed rank test (i.e. a non-parametric paired test) was used at a significance level α of 0.05.

To correlate µCT to CBCT values for each parameter, Pearson (RP) and Spearman (RS) correlation coefficients were calculated, and p-values corresponding to a linear regression were determined, with p < 0.05 considered to be statistically significant. The rationale of using both Pearson’s and Spearman’s R is that the former corresponds to a linear fit, whereas the latter is valid for any monotonous relationship between two variables. While the former is conventionally used for normally distributed data and the latter as a non-parametric alternative, normality of the data was not relevant for in this study.

Paired t-tests, or Wilcoxon signed rank tests in case of non-normally distributed data, were performed on the µCT and CBCT values. p < 0.05 was considered as statistically significant.

Next, CBCT values were calibrated to the µCT values in order to estimate the applicability of calibration coefficients for each bone parameter. These calibration coefficients corresponded to the slope of a linear fit between the CBCT and µCT values, passing through (x = 0, y = 0), with the formula:

y=a×x (1)

with y the calibrated CBCT value, a the calibration coefficient and x the original CBCT value. The mean error after calibration (i.e. the average difference between calibrated CBCT value and µCT value for the 20 samples) was calculated for each bone parameter and normalized to the standard deviation of the µCT values for that particular parameter. This normalization was carried out in order to take the “normal spread” of the values into account, as some of the investigated parameters can have a wide range of values (e.g. BV/TV), whereas others are limited to a narrow range (e.g. fractal dimension).

Results

Correlation, expressed as RS, between alignment angles used by the two observers was 0.56. Wilcoxon signed rank test revealed no significant difference between the angles (p = 0.74).

Correlation between CBCT and micro-CT

Table 1 shows correlation coefficients (R) for CBCT vs µCT values for all bone parameters. The highest R-value was seen for BV/TV (RP = 0.80, RS = 0.73) and BS/TV (RP = 0.67, RS = 0.83). Six other parameters showed a statistically significant correlation for both RP and RS (p < 0.05): branch length, anisotropy, Tb.Th, trabecular spacing, fractal dimension and the negative component of SMI. Connectivity density, SMI and positive SMI showed a significant Spearman but not Pearson correlation. Branches, junctions, triple points and the plateness parameters ev2/ev1 and ev3/ev1 showed no significant correlation. For BS/TV, Tb.Th and SMI, RS was notably higher than RP (+0.14 or more), which may indicate that a linear fit may not be the best option for these parameters; the applicability of different types of non-linear fits was not further explored in this study.

Table 1.

Correlation coefficient, calibration coefficient and normalized error for each bone parameter

Bone parameter R, CBCT–µCT
Calibration coefficient Error normalized to SD of µCT values
Pearson Spearman Mean Max
Skeleton analysis
 Branches 0.41 0.37 40.025 1.04 3.06
 Junctions 0.42 0.35 37.062 1.04 3.42
 Branch length 0.49a 0.52a 0.253 0.58 2.48
 Triple points 0.40 0.33 34.755 1.10 3.70
BS and volume
 BS/TV 0.67a 0.83a 1.995 0.64 1.67
 BV/TV 0.80a 0.73a 0.767 0.48 1.26
Trabecular structure
 Anisotropy 0.57a 0.54a 0.718 0.65 1.63
 Conn. Dens. 0.44 0.47a 39.889 1.05 3.19
 Tb.Th 0.52a 0.66a 0.348 0.66 2.29
 Tb.Sp 0.55a 0.64a 0.774 0.82 2.53
 Fractal dimension 0.68a 0.61a 1.071 0.85 2.11
Plateness
 SMI 0.37 0.55a 0.540 0.75 2.63
 Positive SMI 0.44 0.47a 0.815 0.71 1.84
 Negative SMI 0.68a 0.76a 1.091 0.61 1.68
 ev2/ev1 −0.12 −0.17 1.160 2.10 5.54
 ev3/ev1 0.05 0.10 1.049 1.00 2.81

µCT, micro-CT; BS, bone surface; BS/TV, bone surface per total volume; BV/TV, bone volume per total volume; Conn. Dens., connectivity density; ev1/ev2/ev3, eigenvalues for longest/middle/shortest axis; Max, maximum; R, correlation coefficient, SD, standard deviation; SMI, structure model index; Tb.Sp, trabecular spacing; Tb.Th, trabecular thickness.

a

Significant correlation (p < 0.05).

Figures 46 depict scatter plots with linear regression lines. Although several parameters show a notable correlation between µCT and CBCT values, several data points with large deviations can be spotted. Furthermore, in most cases, the linear fit does not traverse through (0, 0) making it somewhat unrealistic.

Figure 4.

Figure 4

Scatter plots with linear fit (dashed line) for skeleton analysis and bone surface per total volume (BS/TV) and bone volume per total volume (BV/TV) bone parameters. x- and y-axis scales were adapted to the data distribution. µCT, micro-CT.

Figure 6.

Figure 6

Scatter plots with linear fit (dashed line) for structure model index (SMI), plateness and fractal dimension. x- and y-axis scales were adapted to the data distribution. ev1/ev2/ev3, eigenvalues for the longest, middle and shortest axis of the trabeculae, respectively; µCT, micro-CT

Figure 5.

Figure 5

Scatter plots with linear fit (dashed line) for anisotropy, connectivity density (Conn. Dens.), trabecular thickness (Tb.Th) and trabecular spacing (Tb.Sp). x- and y-axis scales were adapted to the data distribution. µCT, micro-CT.

Calibration of CBCT values

Table 1 contains calibration coefficients used to convert CBCT values for error estimations. A calibration coefficient equal to 1 implies that, on average, the numerical values for CBCT and µCT corresponded exactly, a value >1 corresponds to CBCT underestimation relative to µCT and <1 to overestimation. CBCT showed severe underestimation for branches, junctions and connectivity density, and an underestimation by a factor 2 for BS/TV. Considerable overestimation was found for branch length and Tb.Th. For other parameters, calibration coefficients were between 0.54 and 1.16. Paired t-tests and Wilcoxon signed rank tests showed highly significant (p < 0.001) differences for all parameters, including those for which the calibration coefficient was close to 1.

Error after calibration

The error after calibration, that is the average of the difference between recalibrated CBCT values and corresponding µCT values, divided by the standard deviation of the µCT values, is shown in Table 1. Lower values correspond to smaller errors. The smallest error was seen for BV/TV, followed by branch length, negative SMI, BS/TV, anisotropy and Tb.Th. Despite showing significant correlation, Tb.Sp and fractal dimension had a larger error than did SMI and positive did SMI, for which RP was not significant, illustrating the effect of forcing the calibration lines through (0, 0). Errors of other parameters were equal or higher than the standard deviation of the µCT values (i.e. ≥1).

Discussion

In this study, the applicability of bone microarchitecture parameters on CBCT was assessed using µCT as a reference. Out of 16 evaluated bone parameters, 8 showed significant Pearson and Spearman correlation. However, from the scatter plots (Figures 46) and error values (Table 1), it was seen that there is considerable uncertainty regarding the stability of these parameters.

A number of previous studies have investigated the use of one or more of these parameters in CBCT. Fractal dimension, and its relation with bone architecture, has been investigated on in vitro, animal and clinical CBCT images.2326 Various other studies have measured morphometric and other parameters, correlating values from CBCT with those from µCT and dual-energy X-ray absorptiometry.2731 Correlation coefficients were typically interpreted as “high”. However, when comparing the present results with previous studies, it is important to note that a correlation coefficient has limited value when evaluating the potential clinical applications of these bone parameters. The interpretation of a correlation coefficient is mostly subjective and open for interpretation. As seen in Table 1, all coefficients with a value of 0.64 or higher were statistically significant. However, there is a large gap between statistical and clinical significance. Secondly, a high correlation does not imply that the numerical values correspond between the two variables, and large differences between bone parameter values of CBCT and other modalities have been revealed.2731 As seen in Table 1, correlation coefficients could be reasonably high even if the values for CBCT were severely under- or overestimated (as shown by the calibration coefficients). In other words, even if a bone parameter is found to have a high correlation coefficient, the value obtained from CBCT images may be far off the true value. In addition, correlation coefficients derived from a “best fit” are often unrealistic, that is, do not intercept the x- and y-axes at or close to (0, 0), limiting the clinical applicability of these regression lines. In our study, it is shown that after calibration using a regression line going through (0, 0) CBCT values can show notable errors even if the correlation coefficient was relatively high.

Another difference between the present study and previous research is the relatively large sample size (n = 20) in this study, which increases the reliability of the calculated correlation coefficients. Also, this study involved a small volume of interest, equal to the size of the extracted bone samples. When evaluating larger volumes, it may be possible to find a more consistent relationship between bone parameters from CBCT and µCT, but clinical applicability of such results are limited as bone evaluation prior to implant surgery will be limited to small volumes at a potential implant site.

To apply these bone parameters on clinical CBCT images, error margins should be as low as possible to get an accurate assessment of the patient's bone architecture. As even the highest correlation coefficient in our study (i.e. 0.80) shows large deviations from the regression line when looking at the scatter plot, it can be judged that a very high correlation may be required to consider the error margin as acceptable, although an exact cut-off between an acceptable and unacceptable correlation cannot be made at this point. In vitro and clinical studies, possibly retrospective, on the relationship between these bone parameters and implant stability and outcome may aid in the determination of new classification systems. However, before the step to clinical practice is made, further validation on the reliability of these parameters on CBCT data is needed. A bone parameter can only be applied clinically if its values are stable between different CBCT devices, although device-specific bone classifications are not entirely unfeasible. The values should also be independent of exposure parameters such as kVp, mAs and voxel size. It can be expected that values for each bone parameter will be different if the spatial resolution of the image varies, but the stability of each parameter for these variations remains to be found out. Two studies by Ibrahim et al32,33 have investigated the effect of object location and scan parameters, showing varying stability. More evidence is needed to verify the nature of this instability and the possibility for defining correction factors (e.g. as a function of voxel size) to normalize values obtained from images obtained with different exposure settings.

The present study has a few limitations related to its translation to a clinical situation. First of all, it involved the scanning of small bone samples, with no simulation of soft tissue or jawbones. When using clinical exposure parameters, this may lead to an overestimation of image quality; however, in this study, the mAs was lower than that used for adult patients, which compensates for the lack of soft- and hard-tissue simulation. A second limitation that is applicable to all non-clinical studies is that the sample was stationary. As CBCT acquisitions are relatively slow (i.e. typically 10–40 s, 31 s in this study), movement of patients will always induce image blurring, even for very slight motion. It should therefore be kept in mind that image quality on clinical scans will be somewhat worse than that obtained during non-clinical experiments. A final limitation is that this study involved a single CBCT model. It has been shown that there is a large variability in spatial and contrast resolution between CBCTs, with the currently used model found at the high end of the image quality range.10 Therefore, no conclusions can be made regarding CBCT as a whole; as mentioned above, differences between CBCTs as a result of varying exposure and reconstruction factors should be further investigated.

In conclusion, despite the overall correlation, this study demonstrates the large uncertainty associated with using bone microarchitecture parameters on CBCT images. Although clinically relevant parameter ranges are not available, the errors found in this study may be too high for most parameters to be considered for clinical application. Further evidence on the stability of these parameters in varying exposure conditions is warranted.

References

  • 1.Adell R, Lekholm U, Rockler B, Brånemark PI. A 15-year study of osseointegrated implants in the treatment of the edentulous jaw. Int J Oral Surg 1981; 10: 387–416. [DOI] [PubMed] [Google Scholar]
  • 2.Engquist B, Bergendal T, Kallus T, Linden U. A retrospective multicenter evaluation of osseointegrated implants supporting overdentures. Int J Oral Maxillofac Implants 1988; 3: 129–34. [PubMed] [Google Scholar]
  • 3.Jaffin RA, Berman CL. The excessive loss of branemark fixtures in type IV bone: a 5-year analysis. J Periodontol 1991; 62: 2–4. [DOI] [PubMed] [Google Scholar]
  • 4.Drago CJ. Rates of osseointegration of dental implants with regard to anatomical location. J Prosthodont 1992; 1: 29–31. [DOI] [PubMed] [Google Scholar]
  • 5.Jemt T, Lekholm U. Oral implant treatment in posterior partially edentulous jaws: a 5-year follow-up report. Int J Oral Maxillofac Implants 1993; 8: 635–40. [PubMed] [Google Scholar]
  • 6.Wirth AJ, Goldhahn J, Flaig C, Arbenz P, Müller R, van Lenthe GH. Implant stability is affected by local bone microstructural quality. Bone 2011; 49: 473–8. doi: 10.1016/j.bone.2011.05.001 [DOI] [PubMed] [Google Scholar]
  • 7.González-García R, Monje F. Is micro-computed tomography reliable to determine the microstructure of the maxillary alveolar bone? Clin Oral Implants Res 2013; 24: 730–7. doi: 10.1111/j.1600-0501.2012.02478.x [DOI] [PubMed] [Google Scholar]
  • 8.Müller R, Van Campenhout H, Van Damme B, Van Der Perre G, Dequeker J, Hildebrand T, et al. Morphometric analysis of human bone biopsies: a quantitative structural comparison of histological sections and micro-computed tomography. Bone 1998; 23: 59–66. [DOI] [PubMed] [Google Scholar]
  • 9.Swain MV, Xue J. State of the art of micro-CT applications in dental research. Int J Oral Sci 2009; 1: 177–88. doi: 10.4248/IJOS09031 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Pauwels R, Beinsberger J, Stamatakis H, Tsiklakis K, Walker A, Bosmans H, et al. ; SEDENTEXCT Project Consortium. Comparison of spatial and contrast resolution for cone-beam computed tomography scanners. Oral Surg Oral Med Oral Pathol Oral Radiol 2012; 114: 127–35. [DOI] [PubMed] [Google Scholar]
  • 11.Tsai WH, Yu SS. Moment-preserving thresholding: a new approach. Comput Vis Graph 1985; 29: 377–93. [Google Scholar]
  • 12.Doube M, Klosowski MM, Arganda-Carreras I, Cordelières FP, Dougherty RP, Jackson JS, et al. BoneJ: free and extensible bone image analysis in ImageJ. Bone 2010; 47: 1076–9. doi: 10.1016/j.bone.2010.08.023 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Lorensen WE, Cline HE. Marching cubes: a high resolution 3D surface construction algorithm. Proceedings of the 14th Annual Conference on Computer Graphics and Interactive Techniques. July 1987; Association for Computing Machinery; 1987; pp. 163–9.
  • 14.Fazzalari NL, Parkinson IH. Fractal dimension and architecture of trabecular bone. J Pathol 1996; 178: 100–5. [DOI] [PubMed] [Google Scholar]
  • 15.Odgaard A, Gundersen HJ. Quantification of connectivity in cancellous bone, with special emphasis on 3-D reconstructions. Bone 1993; 14: 173–82. [DOI] [PubMed] [Google Scholar]
  • 16.Toriwaki J, Yonekura T. Euler number and connectivity indexes of a three dimensional digital picture. Forma 2002; 17: 183–209. [Google Scholar]
  • 17.Odgaard A. Three-dimensional methods for quantification of cancellous bone architecture. Bone 1997; 20: 315–28. [DOI] [PubMed] [Google Scholar]
  • 18.Harrigan TP, Mann RW. Characterization of microstructural anisotropy in orthotropic materials using a second rank tensor. J Mater Sci 1984; 19: 761–7. [Google Scholar]
  • 19.Dougherty RP, Kunzelmann K-H. Computing local thickness of 3D structures with ImageJ. Microsc Microanal 2007; 13: 1678–9. [Google Scholar]
  • 20.Hildebrand T, Rüegsegger P. A new method for the model-independent assessment of thickness in three-dimensional images. J Microsc 1997; 185: 67–75. [Google Scholar]
  • 21.Hildebrand T, Rüegsegger P. Quantification of bone microarchitecture with the structure model index. Comput Methods Biomech Biomed Engin 1997; 1: 15–23. [DOI] [PubMed] [Google Scholar]
  • 22.Arganda-Carreras I, Fernández-González R, Muñoz-Barrutia A, Ortiz-De-Solorzano C. 3D reconstruction of histological sections: application to mammary gland tissue. Microsc Res Tech 2010; 73: 1019–29. doi: 10.1002/jemt.20829 [DOI] [PubMed] [Google Scholar]
  • 23.Hua Y, Nackaerts O, Duyck J, Maes F, Jacobs R. Bone quality assessment based on cone beam computed tomography imaging. Clin Oral Implants Res 2009; 20: 767–71. doi: 10.1111/j.1600-0501.2008.01677.x [DOI] [PubMed] [Google Scholar]
  • 24.González-Martín O, Lee EA, Veltri M. CBCT fractal dimension changes at the apex of immediate implants placed using undersized drilling. Clin Oral Implants Res 2012; 23: 954–7. doi: 10.1111/j.1600-0501.2011.02246.x [DOI] [PubMed] [Google Scholar]
  • 25.Huang Y, Van Dessel J, Liang X, Depypere M, Zhong W, Ma G, et al. Effects of immediate and delayed loading on peri-implant trabecular structures: a cone beam CT evaluation. Clin Implant Dent Relat Res 2014; 16: 873–83. doi: 10.1111/cid.12063 [DOI] [PubMed] [Google Scholar]
  • 26.Torres SR, Chen CS, Leroux BG, Lee PP, Hollender LG, Schubert MM. Fractal dimension evaluation of cone beam computed tomography in patients with bisphosphonate-associated osteonecrosis. Dentomaxillofac Radiol 2011; 40: 501–5. doi: 10.1259/dmfr/14636637 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Van Dessel J, Huang Y, Depypere M, Rubira-Bullen I, Maes F, Jacobs R. A comparative evaluation of cone beam CT and micro-CT on trabecular bone structures in the human mandible. Dentomaxillofac Radiol 2013; 42: 20130145. doi: 10.1259/dmfr.20130145 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Hsu JT, Wang SP, Huang HL, Chen YJ, Wu J, Tsai MT. The assessment of trabecular bone parameters and cortical bone strength: a comparison of micro-CT and dental cone-beam CT. J Biomech 2013; 46: 2611–18. doi: 10.1016/j.jbiomech.2013.08.004 [DOI] [PubMed] [Google Scholar]
  • 29.Parsa A, Ibrahim N, Hassan B, van der Stelt P, Wismeijer D. Bone quality evaluation at dental implant site using multislice CT, micro-CT, and cone beam CT. Clin Oral Implants Res 2015; 26: e1–7. doi: 10.1111/clr.12315 [DOI] [PubMed] [Google Scholar]
  • 30.Ibrahim N, Parsa A, Hassan B, van der Stelt P, Aartman IH, Wismeijer D. Accuracy of trabecular bone microstructural measurement at planned dental implant sites using cone-beam CT datasets. Clin Oral Implants Res 2014; 25: 941–5. doi: 10.1111/clr.12163 [DOI] [PubMed] [Google Scholar]
  • 31.Ho JT, Wu J, Huang HL, Chen MY, Fuh LJ, Hsu JT. Trabecular bone structural parameters evaluated using dental cone-beam computed tomography: cellular synthetic bones. Biomed Eng Online 2013; 12: 115. doi: 10.1186/1475-925X-12-115 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Ibrahim N, Parsa A, Hassan B, van der Stelt P, Aartman IH, Wismeijer D. The effect of scan parameters on cone beam CT trabecular bone microstructural measurements of human mandible. Dentomaxillofac Radiol 2013; 42: 20130206. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Ibrahim N, Parsa A, Hassan B, van der Stelt P, Aartman IH, Nambiar P. Influence of object location in different FOVs on trabecular bone microstructure measurements of human mandible: a cone beam CT study. Dentomaxillofac Radiol 2014; 43: 20130329. doi: 10.1259/dmfr.20130329 [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Dentomaxillofacial Radiology are provided here courtesy of Oxford University Press

RESOURCES