Skip to main content
The British Journal of Radiology logoLink to The British Journal of Radiology
. 2015 May 28;88(1051):20140829. doi: 10.1259/bjr.20140829

Intracavitary moderator balloon combined with 252Cf brachytherapy and boron neutron capture therapy, improving dosimetry in brain tumour and infiltrations

S F Brandão 1, T P R Campos 1,
PMCID: PMC4628521  PMID: 25927876

Abstract

Objective:

This article proposes a combination of californium-252 (252Cf) brachytherapy, boron neutron capture therapy (BNCT) and an intracavitary moderator balloon catheter applied to brain tumour and infiltrations.

Methods:

Dosimetric evaluations were performed on three protocol set-ups: 252Cf brachytherapy combined with BNCT (Cf-BNCT); Cf-BNCT with a balloon catheter filled with light water (LWB) and the same set-up with heavy water (HWB).

Results:

Cf-BNCT-HWB has presented dosimetric advantages to Cf-BNCT-LWB and Cf-BNCT in infiltrations at 2.0–5.0 cm from the balloon surface. However, Cf-BNCT-LWB has shown superior dosimetry up to 2.0 cm from the balloon surface.

Conclusion:

Cf-BNCT-HWB and Cf-BNCT-LWB protocols provide a selective dose distribution for brain tumour and infiltrations, mainly further from the 252Cf source, sparing the normal brain tissue.

Advances in knowledge:

Malignant brain tumours grow rapidly and often spread to adjacent brain tissues, leading to death. Improvements in brain radiation protocols have been continuously achieved; however, brain tumour recurrence is observed in most cases. Cf-BNCT-LWB and Cf-BNCT-HWB represent new modalities for selectively combating brain tumour infiltrations and metastasis.


Malignant brain tumours grow rapidly and often spread to the adjacent tissues. One of the most malignant and aggressive of brain tumours is glioblastoma multiforme (GBM),1 which generally affects both brain hemispheres and, in some cases, presents multifocal growth.2

The tumour histology and anatomical location are significant factors in the choice of treatment. Tumour resection is often the first attempt, although tumour resection is not possible in some cases owing to the risk of inducing more damage to brain function. Additionally, in the case of a glioblastoma that is highly infiltrative, even with complete tumour resection, microscopic neoplasm extensions still remain in the surrounding tissues and cannot be removed.3 The main purpose of surgery is therefore the reduction of tumour cell population. Additionally, surgery is followed by radiation therapy and chemotherapy. Tumour resection has often produced an empty space occupied early by the tumour mass, resulting in a collapsed brain. The average patient's survival is usually less than 2 years; thus, treatment is considered palliative owing to the frequent local recurrence.4

Boron neutron capture therapy (BNCT) is an experimental radiotherapy for patients with glioblastoma. In this case, a borate compound infusion that concentrates on the tumour is applied, and later, the entire brain is exposed to an epithermal neutron beam. The neutrons are captured by the boron in 10B(n,α)7Li reactions, producing particles with high linear energy transfer and a mean range of 10 µm (approximately the cell diameter). Thus, the dose deposition resulting from the nuclear reaction with boron remains in the tumour. Additionally, owing to the higher cancer cell metabolism, the boron concentration is often lower in the normal tissue and blood but is higher in the tumour, promoting a selective absorbed dose effect.5,6

BNCT is still considered experimental but has been under continuous development over the past few decades.57 Clinical trials have been developed in various international centres, strengthening the applicability of this technique in medicine and confirming the safety of the method.811

One of the major difficulties in BNCT application is associated with neutron source generation required for brain irradiation. Usually, these neutrons are produced by nuclear reactors, which require complex installations, resulting in a very expensive protocol.12,13 According to Wagner et al,14 there are less than ten BNCT clinical facilities for glioblastoma treatment in the world, and the number of treated patients is still low, <1000 patients since BNCT started.

A needle of californium-252 (252Cf) can be an alternative neutron source for BNCT application in radiotherapy centres, and this technique can be combined with brachytherapy.15,16 The application of 252Cf as a neutron source in BNCT has already been investigated.5,16,17 252Cf is an isotope that decays by spontaneous fission, emitting fast neutrons with an average energy of 2.1 MeV, followed by gamma radiation. Its half-life is 2.7 years, which is satisfactory for brain tumour brachytherapy.1719

Fast neutrons emitted by a 252Cf source slow down in the human brain along their pathways because of nuclear interactions, especially with hydrogen nuclei. Thus, the neutrons reach the tumour with a suitable thermal spectrum, and their nuclear interaction with boron nuclei is 103 times higher than that with constituent elements of the human tissue. This article proposes a combination of 252Cf brachytherapy, BNCT and an intracavitary balloon catheter filled with a moderator solution that should slow down the neutrons emitted by the discretely sealed 252Cf source.

METHODS AND MATERIALS

A computational phantom of a human head was used in the simulations. This computational phantom represents a voxel model that was built based on images of the Visible Human Project20 according to dimensions and materials of the BNCT protocol.21 A simulated brain tumour was established in the left temporal lobe of the phantom.16,22,23 Two hypothetical infiltrations were also incorporated in the computational phantom, each one in a side of the temporal lobe. The tumour, and infiltrations in the left (I1) and right (I2) temporal lobes had volumes of 13.7, 0.43 and 0.27 cm3, respectively. The model was constructed and modified using SISCODES (stochastic model-based computational dosimetric system)24 that exports the brain model to MCNP525 (Monte Carlo N-Particle Transport Code, version 5) in a readable format by this code.

The following boron concentrations were established in the computational phantom for the BNCT simulations: 52.5 µg g−1 in tumours and infiltrations, and 15 µg g−1 in the adjacent normal tissue, such as white and grey matter and the limbic system, resulting in a realistic tumour concentration ratio of 1 : 3.5; these values were based on the literature.10,11

The 252Cf source was a cylinder filled with a Pd-Cf2O3 cermet wire with an active length of 7.6 mm and a diameter of 0.7 mm. The cylinder case was 0.2-mm thick and was made of stainless steel. The neutron energy spectrum of the 252Cf source was assumed to be a Watt distribution, with an average energy of 2.1 MeV, as recommended by the MCNP manual.25 The moderator balloon was defined as a sphere of 4 cm in diameter, filled with heavy water (D2O) or light water (H2O). The centre of the moderator balloon was matched with the centre of the 252Cf source and with the centre of the tumour, at 4.3 cm below the surface of the skin. The MCNP5 simulation geometry represents a real partial resection of the tumour and the occupation of its original space by the water balloon.

The neutron and gamma spectra from the 252Cf-emitting source were considered to perform dosimetry. The chemical composition of a human brain tissue was adopted. All possible nuclear and atomic interactions were considered. The microscopic cross-sections of each chemical element present were taken at ENDF-B7 (evaluated nuclear data files-type B, version 7). The dose rate of the BNCT combined with 252Cf brachytherapy addressed the following components: DH, DG, Dγ, DN and DB. The variable DH represents the dose component owing to the elastic interaction of fast neutrons with hydrogen, 1H(n,n)1H. The variables DG and Dγ address the dose components related to the γ-rays emitted by the source and by thermal neutrons captured by hydrogen, 1H(n,γ)2H, respectively. DN is the dose component that represents thermal neutrons captured by nitrogen, 14N(n,p)14C. Finally, DB represents the dose owing to thermal neutrons captured by boron, 10B(n,α)7Li.

The DH, DG and Dγ components were evaluated by MCNP5, which conducted the average energy deposition (MeV g−1) in the cerebral tissues and transformed it to the absorbed dose rate (Gy h−1 p−1 s) through a conversion factor (CF). Then, it was multiplied by 109 p s−1 to consider the differences in neutron and photon particle 252Cf emissions for a specific Cf source activity. This unit means that a dose rate of a unitary gray per hour is absorbed in brain tissue if exposed to a fluency of 109 neutron particles emitted per second by the Cf source.

The version of the code used in this study does not track protons and ions of 14C [generated in 14N(n,p)14C] or α particles and 7Li nuclei [generated in 10B(n,α)7Li reactions]. Therefore, the DN and DB components were calculated from the manipulation of fluency values obtained in the MCNP5 simulation, which represented the neutron fluency in the cerebral tissues. Similar to boron, the sum of the product of nitrogen cross-sections and neutron fluency over the neutron spectra was obtained from MCNP5 and multiplied by the elemental atomic density in each voxel. Thus, the reaction rates for the 14N(n,p)14C and 10B(n,α)7Li were evaluated.26 Using the reaction rates and the exothermic energy values of the reactions, the average energy deposited was calculated and divided by the voxel weight. This absorbed dose was multiplied by CF. Thus, the absorbed dose rates from the two components were determined by the voxels of the model. In all MCNP5 simulations, the cerebral tissues were irradiated by 1 × 108 particles emitted by the source.

The absorbed dose rate for each component, in units of Gy h−1, was multiplied by a weighting factor. The biological weighting factors were considered a guide to achieve dose prescription and used to compare doses across various institutions. The diverse BNCT dose components have their own radiation quality or their own relative biological effectiveness (RBE). Thus, the weighting factors were established to the dose components addressed by the particle emissions of the Cf source. These weighting factors consider the diversity of biological effects on a specific tissue composition owing to a mixed field of high- and low-linear energy transfer (LET) radiation, such as the radiation provided by neutrons, alpha particles and γ-rays on tumour, brain, skin or skull. The DNA breaks caused by the gamma component of the dose undergo repair after 252Cf irradiation. A low-dose rate implies a dose reduction factor, which was incorporated in the weighting factor for γ-rays, wγ. The choice of the remaining three weighting factors considered their respective RBEs.21

The weighting factor for BNCT based on 252Cf emission is presented in Table 1 and is compared with parameters taken from the literature.911,27,28 Table 1 also shows the weight percentage of nitrogen, hydrogen and boron concentrations on the main tissues of the model.

Table 1.

Weighting factor (W) for conventional boron neutron capture therapy (BNCT)9,25,26 and californium-252 coupled to BNCT,1,10 followed by weight percentage of nitrogen (N) and hydrogen (H) in main tissues

Main tissues Weighting factors
W (%)
wG wH wγ wN wB N H
Tumour and metastasis 0.5 3.2 0.5 3.2 3.8 2.2 10.6
Normal brain 0.5 3.2 0.5 3.2 1.3 2.2 10.6
Skin 0.5 3.2 0.5 3.2 2.5 4.2 10.0

The weighting factors wG and wγ for the components of dose rates DG and Dγ, respectively, were lower than the general values used for photons owing to reduced dose rates produced by these components. We included a reduction factor in the coefficients for calculating the probability of induction of deterministic biological effects owing to dose rates <0.19 Gy h−1 of low LET radiation, as in the case of gamma radiation.

The energy per unit of mass generated in the MCNP5 simulation on all voxels of the model was exported to SISCODES. The absorbed dose rates were multiplied by the respective biologic weighting factors. Thus, the SISCODES generated the isodose surfaces superposing them to the computational phantom. These regions represent the total biologically weighted dose rate (Dw) in cerebral tissues.

Dosimetric evaluations were performed with a moderator balloon filled with light or heavy water (H2O or D2O) by applying the same methodology described above. The same evaluations were performed for the 252Cf source without the moderator balloon to have parameters for comparison. To distinguish the proposed radiation therapy protocols in this article, BNCT combined with 252Cf brachytherapy is denoted Cf-BNCT; BNCT combined with 252Cf brachytherapy and moderator balloon filled with heavy water is denoted Cf-BNCT-HWB; and BNCT combined with 252Cf brachytherapy and moderator balloon filled with light water is denoted Cf-BNCT-LWB.

RESULTS

The head voxel model provided a human brain equivalent environment for the simulations. Figure 1 presents the section of the computational human head phantom wherein the centres of 252Cf source and moderator balloon can be identified.

Figure 1.

Figure 1.

A section of the computational human head phantom containing the hypothetical tumour and infiltrations. The sphere represents the moderator balloon with the californium-252 source placed at the central position.

The dosimetric analyses performed in this study were based on the dose deposition in the tumour tissue (tumour and infiltrations), in the adjacent normal tissue (at surface balloon or infiltrations) and in the overall normal tissue, in positions following a linear axis crossing the tumour or balloon centre.

The total biologically weighted dose rates of Cf-BNCT, Cf-BNCT-HWB and Cf-BNCT-LWB are presented in Figure 2. The hypothetical non-resectable tumour was 3.0–5.1 cm below the surface of the skin in the case of the Cf-BNCT protocol. The Cf-BNCT-HWB and Cf-BNCT-LWB protocols received a moderator balloon at a depth of 2.25–6.25 cm in the supposed tumour bed. Tumour infiltrations I1 and I2 were at 6.8–7.7 and 11.0–12.5 cm depth, respectively.

Figure 2.

Figure 2.

Total biologically weighted dose rate for californium-252 brachytherapy combined with boron neutron capture therapy (Cf-BNCT), Cf-BNCT combined with a moderator balloon filled with heavy water (Cf-BNCT-HWB) and Cf-BNCT with balloon filled with light water (Cf-BNCT-LWB) protocols at 432-μg 252Cf source.

In the Cf-BNCT-LWB protocol, the balloon filled with light water generated an overall dosimetric advantage in normal tissue because the dose depositions in the overall voxel's positions in normal tissues were lower than in the Cf-BNCT and Cf-BNCT-HWB protocols, per unit of time and source yield. However, a higher normal-to-tumour dose ratio was found only near the balloon surface at the I1 infiltration. Apparently, the balloon filled with light water did not generate overall dosimetric advantage in distant infiltrations. Indeed, the dose deposition for Cf-BNCT-LWB was lower than that of Cf-BNCT at I2 infiltration, and the dose rate found in adjacent normal tissue and I2 were equivalent for both techniques.

In the Cf-BNCT-HWB protocol, although the dose deposition was only slightly larger than that of the Cf-BNCT, in voxels filled with tumour tissue, the higher values occurred in the two infiltrations distant from the 252Cf source.

Near the balloon surface, there were some interesting values in the normal tissue and infiltrations. The remarkable findings of the dose distribution from the three protocols near the moderator balloon and infiltrations can be better analysed in Figures 35.

Figure 3.

Figure 3.

Total biologically weighted dose rate in the adjacent normal tissue near moderator balloon, which had its radial surface located at 2.25–6.25 cm depth at 432 μg californium-252 (252Cf) source. Cf-BNCT, 252Cf brachytherapy combined with boron neutron capture therapy; Cf-BNCT-HWB, Cf-BNCT combined with a moderator balloon filled with heavy water; Cf-BNCT-LWB, Cf-BNCT with balloon filled with light water.

Figure 5.

Figure 5.

Total biologically weighted dose rate in infiltrations in the right temporal lobe (I2), located at 11.15–12.35 cm depth, and in the adjacent normal tissue, at 432-μg californium-252 (252Cf) source. Cf-BNCT, 252Cf brachytherapy combined with boron neutron capture therapy; Cf-BNCT-HWB, Cf-BNCT combined with a moderator balloon filled with heavy water; Cf-BNCT-LWB, Cf-BNCT with balloon filled with light water.

Figure 3 compares the total biologically weighted dose rate in the adjacent normal tissue to the surface's balloon. The total biologically weighted dose rates owing to Cf-BNCT-HWB and Cf-BNCT-LWB were, respectively, 37% and 50% lower in the left side of the balloon and 11% and 29% lower in the right side than were the dose rates in Cf-BNCT. The lower doses near the balloon surface in the moderator balloon protocols, compared with Cf-BNCT, occurred owing to the improved slowing of the source's neutrons, providing a lower fast neutron dose in the normal tissue. This result is relevant because the balloon protocols provide lower doses in the normal tissue near the source.

Figure 4 depicts the dose values in the region of I1. Dose values in the voxels that represents the I1 infiltration and the normal surrounding tissues were determined. Both the heavy water- and the light water-filled balloon protocols showed dose advantages promoting a larger average total biologically weighted dose rate in tumour tissue and lower dose in the adjacent normal tissues than did the Cf-BNCT. One can observe the inversion of values at 6.45 and 6.95 cm in the Cf-BNCT-LWB protocol, for example. Indeed, Cf-BNCT-LWB provided an average tumour dose in I1, 194% larger than in the normal tissue in the same region, whereas Cf-BNCT-HWB and Cf-BNCT provided 121% and 84% larger doses, respectively. At the volume up to 19 mm from the balloon surface (6.25 up to 8.15 cm—studied region), there was a lower dose in the normal tissue in all protocols in relation to the tumour tissue in the same region.

Figure 4.

Figure 4.

Total biologically weighted dose rate in infiltrations in the left temporal lobe (I1), located at 6.95–7.55 cm depth, and in I1 adjacent normal tissue, at 432 μg californium-252 (252Cf) source. Cf-BNCT, 252Cf brachytherapy combined with boron neutron capture therapy; Cf-BNCT-HWB, Cf-BNCT combined with a moderator balloon filled with heavy water; Cf-BNCT-LWB, Cf-BNCT with balloon filled with light water.

Figure 5 presents the total biologically weighted dose rate in the region of I2. It clearly shows a larger dose in the tumour tissue in the Cf-BNCT-HWB and Cf-BNCT protocols than in the Cf-BNCT-LWB protocol, with an average additional dose of 104% and 79%, respectively. The heavy water balloon provided an additional 16.7% total biologically weighted dose rate in comparison with Cf-BNCT. In the adjacent normal tissue, the average biologically weighted dose was the same in both Cf-BNCT and Cf-BNCT-HWB protocols. The light water balloon showed a dose reduction in I2. In the adjacent normal tissue, 58.3% and 54.2% of the dose produced in Cf-BNCT was observed in both sides of the infiltration, respectively.

Thus, Cf-BNCT-HWB demonstrated better dosimetric advantages than those of Cf-BNCT-LWB and Cf-BNCT at infiltrations 5.0 cm from the balloon surface.

Figure 6 presents the spatial isodose distribution on a section of the computational phantom, representing the total biologically weighted dose rate (Dw) in the cerebral tissues owing to Cf-BNCT-HWB.

Figure 6.

Figure 6.

Spatial isodose distribution of the total biologically weighted dose rate in an axial plane of the computational phantom produced in californium (Cf)-boron neutron capture therapy combined with a moderator balloon filled with heavy water with a balloon 4 cm in diameter. The scale represent the percentage of the maximum dose rate of 12.0 Gy h−1 at 432 μg 252Cf source. I1, infiltrations in the left temporal lobe; I2, infiltrations in the right temporal lobe.

As shown in Figure 6, the dose values in the periphery of the balloon, up to 7.7 cm from the surface, presented 25.0–49.9% of the maximal total biologically weighted dose. The regions of infiltrations were highlighted, showing the dose selectivity at these tumour regions. In I1, the average dose was 77% of the maximum dose, whereas I2 received an average dose of 23% of the maximum dose. Additionally, a dose >75% of the maximum dose can also be seen on the right and above the I1, in a small region comprising residual tumour tissue. In the interior of the nasal cavity, the dose between 25% and 49.9% was owing to 14N(n,p)14C reactions owing to the air composition in the voxels filled with this material. This larger dose in air is not relevant to these analyses.

The advantage of the 252Cf brachytherapy associated with BNCT and combined with a heavy or light water balloon compared with Cf-BNCT was also confirmed by analysing the neutron fluency in the brain, particularly fast neutrons and thermal neutrons. In comparing the average fluency in the Cf-BNCT and Cf-BNCT-HWB protocols (Figure 7), a 25% reduction of fast neutrons and 20% increase of thermal neutrons were observed with the presence of a moderator balloon. The Cf-BNCT-LWB protocol provided more thermal neutrons near the balloon surface up to 1.5 cm from the surface with lower fast neutrons; however, the Cf-BNCT-HWB protocol provided more thermal neutrons at 1.5 cm and further from the surface.

Figure 7.

Figure 7.

Normalized thermal and fast neutron fluency in function of depth from right balloon surface starting at 6.25 cm depth. Infiltrations I1 and I2 are at 6.8–7.7 cm depth and 11.0–12.5 cm depth, respectively. Cf, californium-252; Cf-BNCT, Cf brachytherapy combined with boron neutron capture therapy; Cf-BNCT-HWB, Cf-BNCT combined with a moderator balloon filled with heavy water; Cf-BNCT-LWB, Cf-BNCT with balloon filled with light water.

Table 2 shows the possible doses and exposure times to achieve a prescribed biologically weighted dose of 60 Gy in the tumour and in the adjacent normal tissue near the balloon. A surgency of 1.66 108 n s−1 to the 252Cf source was considered for this analysis, referring to 72 µg of the californium mass, which is in agreement with the dimensions of the simulated source applied in this study. Table 2 also shows that it took 1.0 h to reach a biologically weighted dose of 60 Gy in the tumour tissue, adjacent to the 252Cf source, in the Cf-BNCT protocol. In this case, the prescribed dose of 60 Gy was achieved at 0.25 cm from the source. The time can be scaled to 9.4 h to produce 60 Gy in the adjacent normal tissue to the tumour (the edge of the tumour, at 0.95 cm from the source). Thus, the dose at 0.25 cm should be 554 Gy and 15.1 and 4.5 Gy at infiltrations I1 and I2, respectively.

Table 2.

Dose vs exposure time to achieve a prescribed biologically weighted dose of 60 Gy in the tumour (T) and balloon adjacent tumour bed tissue (TB) and normal tissue (N), with 72 µg of californium mass

Therapy 252Cf plus Time (h) Dose (Gy) (relative biological effectiveness)
T
TB
I1
I2
Na
0.25b 2.15 2.65 6.85  
BNCTc 1.0 60d 6.5 1.6 0.5 2.4
BNCTc HWB 54.0 e 60 91.5 30.0 20.7
BNCTc LWB 68.2 e 60 126.8 19.3 23.0

BNCT, boron neutron capture therapy; Cf, californium-252 brachytherapy; HWB, balloon catheter filled with heavy water; LWD, balloon catheter filled with light water.

a

Higher dose at normal tissue at a relative distance referent to 1.65 cm depth (surface balloon) in case of HWB and LWB, and 5.85 cm right to Cf-BNCT.

b

Relative distance between the centre of the Cf source and the voxel position in which the dose was evaluated, in centimetres.

c

Boron concentrations of 52.5 µg g−1 in the tumour and infiltrations, and 15 µg g−1 in the adjacent normal tissue (concentration ratio of 1:3.5).

d

Value taken at a voxel near the surface of the californium-252 source placed at a non-resectable tumour represented on the model.

e

Resected tumour; volume filled with moderator balloon.

The exposure times of 54.0 and 68.2 h in the Cf-BNCT-HWB and Cf-BNCT-LWB protocols were required to reach 60 Gy in the adjacent tissue to the balloon (at tumour bed), respectively. In the Cf-BNCT-HWB or Cf-BNCT-LWB protocols, the prescribed dose was assumed at the right balloon surface at 2.15 cm from the source. In the Cf-BNCT-HWB protocol, the doses in the infiltrations were approximately 60 times larger than those in the Cf-BNCT, whereas in the Cf-BNCT-LWB protocol, the doses in I1 and I2 were, respectively, 79 and 39 times larger than those in the Cf-BNCT protocol.

Table 3 presents the estimated values of dose components followed by achievable total biologically weighted dose in the brain and tumour in the Cf-BNCT-LWB and Cf-BNCT-HWB protocols and similar estimated values at Massachusetts Institute of Technology (MIT)9,29 and Petten30 institutions were taken from BNCT facilities, considering an irradiation field directed towards the tumour at the same depth (d) as infiltrations I1 (6.95–7.55 cm) and I2 (11.15–12.35 cm). Boron concentrations in tumour and brain tissue were found in the literature.9,29,30

Table 3.

Typical estimated values of biologically weighted dose components and their achievable total values for californium-252 (252Cf) brachytherapy combined with boron neutron capture therapy (Cf-BNCT) with a balloon catheter filled with light water (LWB) and Cf-BNCT with a balloon catheter filled with heavy water (HWB) to a 432-μg 252Cf, and for conventional boron neutron capture therapy (BNCT) at MIT, Petten institutions taken on BNCT facilities, considering an irradiation field directed towards the tumour at the same depth (d) referent to infiltrations I1 and I2 positions, boron concentration on tumour (CBT) and brain (CBb)

BNCT type d CBT/CBb (ppm) Biologically weighted dose components (Gy) and weighting factors
Biologically weighted total dose (Gy)
Exposure time (min)
DB
DγwF; DGwF= 0.5 Dnwn 3.2 DNwN 3.2 Brain Tumour
wBb 1.3 wBT 3.8
M67-MITa I1 –/15 0.5   1.8 0.25 0.1 2.65   100
I2 0.15 1.2 0.15 0.03 1.53
MITb I1 65/18 0.8 4.2 0.55; 0.06 0.03 0.21 1.65 5.05 1
I2
Pettenc I1 30 (blood) 2.9   3.15 0.12 0.1 6.27   74.77
I2 0.6 1.8 0.05 0.05 2.50
Cf-LWBd I1 52.5/15 0.54 7.04 0.09; 0.11e 1.73 0.49 2.96 9.45 60
I2 0.22 1.04 0.02; 0.01e 0.20 0.07 0.52 1.34
Cf-HWBf I1 52.5/15 0.54 5.96 0.04; 0.10e 2.86 0.41 4.21 9.63 60
I2 0.22 2.18 0.02; 0.01e 0.42 0.15 0.84 2.80

Cf, 252Cf brachytherapy.

a

M67 epithermal neutron beam at BNCT facilities in MIT Nuclear Reactor Laboratory (MITR-II) in Boston, MA.9

b

MIT fission converter operating at 83 kW.29

c

Petten, calculated with the Idaho National Engineering and Environmental Laboratory treatment planning software.30

d

Weighted dose from Cf-BNCT-HWB; emitting 109 p s−1 evaluated per neutron emitted by source.

e

Adjusted by the difference of gamma and neutron emitting.

f

Weighted dose from Cf-BNCT-LWB; emitting 109 p s−1 evaluated per neutron emitted by source.

Dose values in Table 3 were found by the M67 epithermal neutron beam at the MIT Nuclear Reactor Laboratory in a basement medical treatment room. We measured physical dose rates in an ellipsoidal phantom assuming 1.84% nitrogen by mass, 15 μg g−1 boron in the brain and a reactor power of 5 MW. We used units of centigray per minute and scaled using weighting factors.9 In the fission converter facility, a weighted dose depth profile with a 160-mm diameter filled size with fission converter operating at 83 kW was used, assuming boron concentrations typical for boronophenylalanine (BPA) of 65 and 18 μg g−1 for tumour and normal brain tissue, respectively (1.0 for photon, 3.2 for neutrons; 1.35 for boron in brain and 3.8 in tumour), in the ellipsoidal water phantom and BPA uptake. In this case, in I1, dose components were estimated by percentage values for DB of 49%; Dγ (n,γ), 34%; DG (source), 4%; Dn, 2%; and DN, 13%.29

Dose values from Petten (Table 3) were estimated from MCNP depth dose calculations represented by the four dose component curves in which a single BNCT beam towards a patient head was applied [irradiation time of 74.77 min; 10B-blood concentration (cb) of 30 ppm], evaluated by the Idaho National Engineering and Environmental Laboratory treatment planning software (DB, boron absorbed dose by 7Li and 4He particles; Dg, γ-ray absorbed dose; Dn, fast neutron absorbed dose; and DN, nitrogen absorbed dose).30

In the Cf-BNCT-HWB and Cf-BNCT-LWB protocols, the emission of 109 p s−1 was considered, in which p was evaluated per neutron emitted by source. The prompt spectral γ-ray yield from the reaction 252Cf(sf) was adopted as 8.30.31 The average number of neutrons for 252Cf was 3.7509 neutrons per fission.32 Therefore, the ratio of γ-ray and neutron emitting was 2.2128. Thus, the dose produced per γ-ray particle emitted by the source was previously multiplied by 2.2128 to be added to the dose values normalized by neutron particles emitted by 252Cf. A neutron was considered to emit 2.31434 × 1012 n s−1 g−1 per 252Cf mass. Thus, in the Cf-BNCT-LWB and Cf-BNCT-HWB in Table 3, a source mass of 432 μg of 252Cf was assumed to generate 109 n s−1.

DISCUSSION

GBM presents a high degree of malignancy and tends to spread quickly, invading the surrounding normal brain tissue. Investments in new technologies have continuously occurred with the purpose of maximizing tumour control and minimizing their side effects.5,16,33 However, the cure for glioblastoma still represents a challenge and involves the overall elimination of neoplastic cellular residues and infiltrations in the periphery of the resection region at the tumour bed.3436 Currently, the standard protocol for early diagnosed glioblastoma comprises tumour resection, followed by radiotherapy with concomitant or adjuvant chemotherapy, mainly with the chemotherapeutic agent known as temozolomide (TMZ). Nevertheless, recurrence is observed in most cases.2,4

Brachytherapy is often prescribed following external beam radiation, especially in cases of recurrence. However, recent studies have investigated the use of brachytherapy as an initial treatment at the time of diagnosis.37,38,39 Usually, 125I and 192Ir are the isotopes employed in brachytherapy for brain tumour treatment,40,41 but 252Cf has also been used in brain tumour therapy with some hopeful results.18,37,41

During the past few years, the 125I-brachytherapy modality has been developed. It consists of a device that includes a balloon and a catheter for infusing a radioactive fluid. During the irradiation period, the balloon is filled with 125I radioactive solution, emitting photons with an average energy of 28 keV.4,42 The device is inserted into the patient during the surgical procedure for tumour resection so that the space originally occupied by the tumour is filled by the balloon.33,42 Early investigations42,43 showed that the mean survival of patients treated with such a technique was 4.6–14 months and 13.7 months, respectively. 252Cf brachytherapy combined with BNCT delivered a selective irradiation to the target tumour and to infiltration zones, whereas intracavitary balloon catheter brachytherapy with 125I delivered negligible doses in the tumour infiltration zones. However, this modality was addressed because 125I-brachytherapy uses a tumour resectable cavity and a catheter in which its tip has a balloon that can be inflated.

The cavity formation owing to the excised tumour in the brain may also be explored to improve BNCT. In the 1980s and 1990s, the Japanese had often performed a craniotomy on the patient and explored an air-empty cavity to overcome the thermal neutron depth dose deficit in BNCT.44 Sakurai et al25 investigated the effect of a cavity filled with air on the depth-dose distribution in BNCT for malignant brain tumours. According to Sakurai et al, the cerebrospinal fluid in the tumour-resectable cavity was drained out, and air was infused to the space providing a void in the brain. This air space improves the epithermal and thermal external neutron beam penetrations to the brain. Indeed, the thermal neutron flux at a particular depth increases with this technique.45 The present Cf-BNCT-LWB and Cf-BNCT-HWB modalities also explore cavity formation owing to tumour resection; we introduced a water-inflated balloon with the goal of improving the slowing of fast neutrons provided by a segment of 252Cf-fission source placed at the balloon centre. Although spatial distribution of neutrons is quite different because one is provided by a teletherapy and the other by brachytherapy, both techniques may improve thermal neutrons at the tumour bed. However, the reduction in fast neutrons by a slow-down moderator is only possible in a water-filled cavity, such as in the Cf-BNCT-LWB or Cf-BNCT-HWB modalities.

BNCT is a technique that is being continuously developed and presents encouraging results. Clinical results have shown a median survival from 7 to 21.9 months in a total of 190 patients with glioblastoma treated with BNCT between the years 1994 and 2008.6 Among these 53 patients, only 9.4% achieved a 2-year survival. Recently, the treatment of 23 patients with BNCT has been reported with a median survival of 19.5 months and 26.1%, 17.4% and 5.8% of patients had a 2-, 3- and 5-year survival, respectively.46 Kageji et al46 also suggested that the combination of BNCT with TMZ could produce better results.

A boron carrier to the tumour is still a challenge for the success of the BNCT technique. Currently, the only compounds in clinical use are BPA and sodium borocaptate, but none is considered ideal in relation to its uptake by the tumour and concentration ratio of tumour : normal tissue.6,13,47 Recently, several studies have addressed boron incorporation into the tumour-targeting molecules, such as peptides, proteins, antibodies, nucleosides, sugars, porphyrins, liposomes and nanoparticles, showing promising results in animal models and in vitro.6 Another drawback in BNCT application is related to the complexity and the high cost of a reactor as a neutron source for tumour irradiation, as previously mentioned. However, reports on clinical attempts based on reactor technology have been continuously published in the past two decades, with encouraging results.5,12,48,49

This article proposed a combination of 252Cf brachytherapy, BNCT and an intracavitary balloon catheter, filled with a solution that slowed the neutrons emitted by a sealed source for the treatment of GBM. The results showed a dosimetric advantage in Cf-BNCT-HWB compared with Cf-BNCT because the dose in the infiltrations was approximately 60 times larger with HWB. Infiltration at 3.0 up to 7.0 cm from the balloon surface receives a higher dose in Cf-BNCT-HWB than that of LWB; however, a better dose is observed in the LWB protocol in infiltrations up to 2.0 cm from the balloon surface. However, overall normal tissue received lower doses in the Cf-BNCT-LWB protocol.

252Cf neutron brachytherapy is a fast neutron therapy in the entire brain. In a previous study that coupled this modality with BNCT (Cf-BNCT),22 there was a small advantage at infiltrations I1 and I2 owing to a larger contribution to the boron component to the total biologically weighted dose in relation to the fast neutron component at 4 and 7 cm from the 252Cf source. At the distance of 0–4 cm, the fast neutron dose was superior to other components. It was also shown that the dependence of the boron ratio on tumour and normal brain dictates the success of Cf-BNCT, which depends on the development of new drugs with higher boron ratios, such as 1 : 10. Additionally, fast neutron fluency in Cf-BNCT near the source is larger than thermal fluency. Thus, the Cf-BNCT modality could be a fast neutron brachytherapy, and its weight dose component may be large with few exceptions only at positions with a larger boron ratio close to 1 : 10.

In the Cf-BNCT-HWB or Cf-BNCT-LWB protocols, the normalized thermal and fast neutron fluencies in relation to depth, from the right side of the balloon starting at its surface, are shown in Figure 7. The starting evaluated position was 6.25 cm, whereas the Cf source was at 4.3 cm inside the water balloon. The results show that the fast neutron fluency was lower than the thermal fluency past 6.5 cm from the balloon surface. Indeed, the boron component was higher at I1 and I2 than the scattering neutron component values. Therefore, the biological weighted dose components generated by the thermal fluency was higher than the component generated by the fast neutron fluency. Thus, Cf-BNCT-HWB and Cf-BNCT-LWB are not fast brachytherapies. Figure 4 presents an inversion of the total biologically weighted dose from Cf-BNCT-LWB and Cf-BNCT-HWB in relation to the values from Cf-BNCT at the infiltration positions, demonstrating suitable dose selectivity in the regions in which boron concentration ratios were equal to or higher than 1 : 3.5 owing to the presence of differential uptake of cancerous cells.

Table 3 presents a dosimetric comparison of reputable centres that perform conventional BNCT and Cf-BNCT-LWB and Cf-BNCT-HWB. This comparison is not absolute because the parameters from conventional BNCT were not equal to our Cf-BNCT-balloon simulations, but they were representative of a clinical or a calibration situation described in the literature.9,29,30 Additionally, it was limited to the I1 and I2 infiltration positions. However, it was shown that, by applying a 432-μg 252Cf source, the total biological dose at Cf-HWB and Cf-LWB, found in I1 and I2, were similar to those in conventional BNCT. Additionally, the exposure time with a larger mass source of 252Cf may be clinically acceptable and in the same order of magnitude of those addressed by Petten and M67-MIT.

The Cf-BNCT-HWB, with a smaller activity with a 72-μg 252Cf source, required 54.0 h to reach a biologically weighted dose of 60 Gy in the adjacent tissue to the balloon. If 60 Gy was assumed in the infiltration I1, the achievable dose in the adjacent normal tissue to the balloon would be 39.4 Gy. In this case, the total irradiation time would be reduced to 35.4 h, which means six fractions of 5 h 54 min. It is relevant to mention that the dose in the infiltration I1 was 52% higher than in the adjacent tissue to the balloon, although it was near the 252Cf source.

The protocol for Cf-BNCT-HWB or Cf-BNCT-LWB may consist of a daily fractionate exposure with a 72-μg 252Cf source, as described in Table 2, in which the source is inserted, positioned and inflated by an automated brachytherapy device. Considering a prescribed biologically weighted dose of 60 Gy in the adjacent tissue to the balloon and a 252Cf source of 72 μg mass, the treatment could be performed in 9 fractions of 6 h or in 18 fractions of 3 h of exposure in the Cf-BNCT-HWB protocol. In the Cf-BNCT-LWB protocol, the treatment could be performed in 9 fractions of 7 h 35 min or in 18 fractions of 3 h 47 min of exposure. These exposure times could be further reduced by employing a 252Cf source with a larger 252Cf mass. Compared with teletherapy, in which the patient is irradiated with a dose of 2.0 Gy in 30 fractions, there is significant time saving, regardless of the advantages in the spatial dose distribution.

Although it is not possible to identify small infiltrations at the time of the diagnosis, or if there are multiple metastases developing in the brain, or even if there are infiltrations remaining after tumour resection, these residual cells may be capable of taking up boron compounds in large concentrations, which differentiates them from normal cells. Thus, the therapy could address these regions without previous identification.

Furthermore, progress in selective boron carriers would improve the present proposed technique's selectivity for neoplasia disseminations. Additionally, the 252Cf source is affordable, and this type of neutron source is independent of reactor systems. 252Cf brachytherapy in association with BNCT, improved by a moderator balloon, may represent an alternative method for establishing a BNCT protocol in various radiotherapy centres around the world. The side effects produced by Cf-BNCT-LWB or Cf-BNCT-HWB should be evaluated, primarily through experimental studies in animals and then in clinical trials. New boron drugs that bring higher cancerous cell capitation will make the present study's protocols more attractive in radiation therapy.

CONCLUSION

Considering the highly infiltrative feature of glioblastoma, Cf-BNCT-HWB and Cf-BNCT-LWB protocols provide a selective dose for brain tumour and infiltrations, far from the 252Cf source, sparing the surrounding normal brain tissue. Cf-BNCT-HWB demonstrated better dosimetric advantages than Cf-BNCT-LWB and Cf-BNCT at infiltrations from 2.0 to 5.0 cm from the balloon surface. Moreover, Cf-BNCT-LWB provides dosimetric advantages up to 2.0 cm near balloon surface compared with the other protocols.

Acknowledgments

ACKNOWLEDGMENTS

The authors are thankful to the scholarship to one of these authors, from CNPq, CAPES and FAPEMIG.

Contributor Information

S F Brandão, Email: samiabrandao@gmail.com.

T P R Campos, Email: tprcampos@yahoo.com.br.

REFERENCES

  • 1.Lamers LM, Stupp R, van den Bent MJ, Al MJ, Gorlia T, Wasserfallen J, et al. ; EORTC 26981/22981 NCI-C CE3 Intergroup Study. Cost-effectiveness of temozolomide for the treatment of newly diagnosed glioblastoma multiforme: a report from the EORTC 26981/22981 NCI-C CE3 intergroup study. Cancer 2008; 112: 1337–44. doi: 10.1002/cncr.23297 [DOI] [PubMed] [Google Scholar]
  • 2.Ferreira NF, Barbosa M, Amaral LL, Mendonça RA, Lima SS. Studies based on magnetic resonance image of 67 glioblastoma multiform cases and metastases occurences. [in Spanish.] Arq Neuropsiquiatr 2004; 62: 695–700. doi: 10.1590/S0004-282X2004000400024 [DOI] [PubMed] [Google Scholar]
  • 3.Thumma SR, Elaimy AL, Daines N, Mackay AR, Lamoreaux WT, Fairbanks RK, et al. Long-term survival after gamma knife radiosurgery in a case of recurrent glioblastoma multiforme: a case report and review of the literature. Case Rep Med 2012; 2012: 545492. doi: 10.1155/2012/545492 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Amelio D, Amichetti M. Radiation therapy for the treatment of recurrent glioblastoma: an overview. Cancer (Basel) 2012; 4: 257–80. doi: 10.3390/cancers4010257 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Savolainen S, Kortesniemi M, Timonen M, Reijonen V, Kuusela L, Uusi-Simola J, et al. Boron neutron capture therapy (BNCT) in Finland: technological and physical prospects after 20 years of experiences. Phys Med 2013; 29: 233–48. doi: 10.1016/j.ejmp.2012.04.008 [DOI] [PubMed] [Google Scholar]
  • 6.Barth RF, Vicente MG, Harling OK, Kiger WS, 3rd, Riley KJ, Binns PJ, et al. Current status of boron neutron capture therapy of high grade gliomas and recurrent head and neck cancer. Radiat Oncol 2012; 7: 146. doi: 10.1186/1748-717X-7-146 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Barth RF, Solloway AH, Fairchild RG. Boron neutron capture therapy of cancer. Cancer Res 1990; 50: 1061–70. [PubMed] [Google Scholar]
  • 8.Hatanaka H. A revised boron-neutron capture therapy for malignant brain tumors. II. Interim clinical result with the patients excluding previous treatments. J Neurol 1975; 209: 81–94. doi: 10.1007/BF00314601 [DOI] [PubMed] [Google Scholar]
  • 9.Palmer MR, Goorley JT, Kiger WS, Busse PM, Riley KJ, Harling OK, et al. Treatment planning and dosimetry for the Harvard-MIT phase I clinical trial of cranial neutron capture therapy. Int J Radiat Oncol Biol Phys 2002; 53: 1361–79. doi: 10.1016/s0360-3016(02)02862-6 [DOI] [PubMed] [Google Scholar]
  • 10.Binns PJ, Riley KJ, Harling OK, Kiger WS, 3rd, Munck af Rosenschöld PM, Giusti V, et al. An international dosimetry exchange for boron neutron capture therapy, part I: absorbed dose measurements. Med Phys 2005; 32: 3729–36. doi: 10.1118/1.2132572 [DOI] [PubMed] [Google Scholar]
  • 11.Riley KJ, Binns PJ, Harling OK, Albritton JR, Kiger WS, 3rd, Rezaei A, et al. An international dosimetry exchange for BNCT, part II: computational dosimetry normalizations. Med Phys 2008; 35: 5419–25. doi: 10.1118/1.3005480 [DOI] [PubMed] [Google Scholar]
  • 12.Yamamoto T, Nakai K, Nariai T, Kumada H, Okumura T, Mizumoto M, et al. The status of Tsukuba BNCT trial: BPA-based boron neutron capture therapy combined with X-ray irradiation. Appl Radiat Isot 2011; 69: 1817–18. doi: 10.1016/j.apradiso.2011.02.013 [DOI] [PubMed] [Google Scholar]
  • 13.Cruickshank GS, Ngoga D, Detta A, Green S, James ND, Wojnecki C, et al. A cancer research UK pharmacokinetic study of BPA-mannitol in patients with high grade glioma to optimise uptake parameters for clinical trials of BNCT. Appl Radiat Isot 2009; 67(Suppl. 7–8): S31–3. doi: 10.1016/j.apradiso.2009.03.016 [DOI] [PubMed] [Google Scholar]
  • 14.Wagner FM, Loeper-Kabasakal B, Breitkreutz H. Neutron medical treatment of tumours—a survey of facilities. J Inst 2012; 7: 102–9. doi: 10.1088/1748-0221/7/03/c03041 [DOI] [Google Scholar]
  • 15.Campos TPR. Consideration about boron neutron capture therapy. [In Portugese.] Rev Bras Cancerologia 2000; 46: 283–92. [Google Scholar]
  • 16.Brandão SF, Campos TP. Brain tumor and infiltrations dosimetry of boron neutron capture therapy combined with 252Cf brachytherapy. Radiat Prot Dosimetry 2012; 149: 289–96. doi: 10.1093/rpd/ncr250 [DOI] [PubMed] [Google Scholar]
  • 17.Chaves IF, Campos TPR. Radiotherapy coupling boron neutron capture therapy (BNCT) and 252Cf brachytherapy. Eur J Cancer 1995; 31: 44. doi: 10.1016/0959-8049(95)95454-e [DOI] [Google Scholar]
  • 18.Patchell RA, Yaes RJ, Beach L, Kryscio RJ, Daves DG, Tibbs PA, et al. A phase I trial of neutron brachytherapy for the treatment of malignant gliomas. Br J Radiol 1997; 70: 1162–8. doi: 10.1259/bjr.70.839.9536908 [DOI] [PubMed] [Google Scholar]
  • 19.Al-Saihati I, Naqvi AA. Neutron and gamma ray doses from a 252Cf brachytherapy source in a water phantom. J Radioanal Nucl Chem 2013; 296: 963–6. doi: 10.1007/s10967-012-2172-5 [DOI] [Google Scholar]
  • 20.VHP. Visible Human Project. Bethesda, MD: United States National Library of Medicine: [Updated 4 March 2010; cited 12 December 2014]. Available from: http://www.nlm.nih.gov/research/visible/applications.html [Google Scholar]
  • 21.Current status of neutron capture therapy. IAEA-TECDOC-1223. Chapter 6, Radiobiology. Vienna, Austria: International Atomic Energy Agency; 2001. pp. 32–4.
  • 22.Brandão SF, Campos TP. Dosimetria comparativa da braquiterapia por balão de I-125 e por Cf-252 associada à BNCT para tumores cerebrais. RB Radiol Bras (Impresso) 2013; 46: 221–6. [Google Scholar]
  • 23.Brooks RA, Dichiro G, Keller MR. Explanation of cerebral white—gray contrast in computed tomography. J Comput Assist Tomogr 1980; 4: 489–91. doi: 10.1097/00004728-198008000-00016 [DOI] [PubMed] [Google Scholar]
  • 24.Trindade BM, de Campos TPR. Stochastic method-based computational system for neutron/photon dosimetry applied to radiotherapy and radiology. RB Radiol Bras (Impresso) 2011; 44: 109–16. doi: 10.1590/S0100-39842011000200011 [DOI] [Google Scholar]
  • 25.X-5 Monte Carlo Team. MCNP—a general Monte Carlo N-Particle Transport Code, version 5. Los Alamos, NM: Los Alamos National Laboratory; 2003. [Google Scholar]
  • 26.Photon, electron, proton and neutron interaction data for body tissues. ICRU Report 46. Bethesda, MD: International Commission on Radiation Units and Measurements; 1992. [Google Scholar]
  • 27.Sakurai Y, Ono K, Miyatake S, Maruhashi A. Improvement effect on the depth-dose distribution by CSF drainage and air infusion of a tumour-removed cavity in boron neutron capture therapy for malignant brain tumours. Phys Med Biol 2006; 51: 1173–83. doi: 10.1088/0031-9155/51/5/009 [DOI] [PubMed] [Google Scholar]
  • 28.Coderre JA, Morris GM. The radiation biology of boron neutron capture therapy. Radiat Res 1999; 151: 1–18. doi: 10.2307/3579742 [DOI] [PubMed] [Google Scholar]
  • 29.Riley KJ, Binns PJ, Harling OK. Performance characteristics of the MIT fission converter based epithermal neutron beam. Phys Med Biol 2003; 48: 943–58. doi: 10.1088/0031-9155/48/7/310 [DOI] [PubMed] [Google Scholar]
  • 30.Rassow J, Stecher-Rasmussen F, Voorbraak W, Moss R, Vroegindeweij C, Hideghéty K, et al. Comparison of quality assurance for performance and safety characteristics of the facility for Boron Neutron Capture therapy in Petten/NL with medical electron accelerators. Radiother Oncol 2001; 59: 99–108. doi: 10.1016/s0167-8140(00)00298-x [DOI] [PubMed] [Google Scholar]
  • 31.Billnert R, Hambsch FJ, Oberstedt A, Oberstedt S. New prompt spectral γ-ray data from the reaction 252Cf(sf) and its implication on present evaluated nuclear data files. Phys Rev C 2012; 87: 024601. doi: 10.1103/PhysRevC.87.024601 [DOI] [Google Scholar]
  • 32.Axton EJ, Bardell AG. Neutron yield from the spontaneous fission of 252Cf. Metrologia 1985; 21: 59. doi: 10.1088/0026-1394/21/2/003 [DOI] [Google Scholar]
  • 33.Noda SE, Lautenschlaeger T, Siedow MR, Patel DR, El-Jawahri A, Suzuki Y, et al. Technological advances in radiation oncology for central nervous system tumors. Semin Radiat Oncol 2009; 19: 179–86. doi: 10.1016/j.semradonc.2009.02.006 [DOI] [PubMed] [Google Scholar]
  • 34.Amelio D, Lorentini S, Schwarz M, Amichetti M. Intensity-modulated radiation therapy in newly diagnosed glioblastoma: a systematic review on clinical and technical issues. Radiother Oncol 2010; 97: 361–9. doi: 10.1016/j.radonc.2010.08.018 [DOI] [PubMed] [Google Scholar]
  • 35.Sadeghi M, Enferadi M, Shirazi A. External and internal radiation therapy: past and future directions. J Cancer Res Ther 2010; 6: 239–48. doi: 10.4103/0973-1482.73324 [DOI] [PubMed] [Google Scholar]
  • 36.Cuneo KC, Vredenburgh JJ, Sampson JH, Reardon DA, Desjardins A, Peters KB, et al. Safety and efficacy of stereotactic radiosurgery and adjuvant bevacizumab in patients with recurrent malignant gliomas. Int J Radiat Oncol Biol Phys 2012; 82: 2018–24. doi: 10.1016/j.ijrobp.2010.12.074 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Vitaz TW, Warnke PC, Tabar V, Gutin PH. Brachytherapy for brain tumors. J Neuro Oncol 2005; 73: 71–86. doi: 10.1007/s11060-004-2352-4 [DOI] [PubMed] [Google Scholar]
  • 38.Schwarz SB, Thon N, Nikolajek K, Niyaz M, Tonn JC, Belka C, et al. Iodine-125 brachytherapy for brain tumours—a review. Radiat Oncol 2012; 7: 30. doi: 10.1186/1748-717x-7-30 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Liu Y, Diaz AZ, Papanikolaou N. Dose calculation for brain using a Cf-252 neutron brachytherapy source based on human anatomy Monte Carlo model and compared to standard photon IMRT planning. Int J Radiat Oncol Biol Phys 2008; 72: S664. doi: 10.1016/j.ijrobp.2008.06.358 [DOI] [Google Scholar]
  • 40.Tselis N, Kolotas C, Birn G, Röddiger S, Filipowicz I, Kontova M, et al. CT-guided interstitial HDR brachytherapy for recurrent glioblastoma multiforme. Strahlenther Onkol 2007; 10: 563–70. doi: 10.1007/s00066-007-1721-2 [DOI] [PubMed] [Google Scholar]
  • 41.Liu BL, Cheng JX, Zhang X, Zhang W. Controversies concerning the application of brachytherapy in central nervous system tumors. J Cancer Res Clin Oncol 2010; 136: 173–85. doi: 10.1007/s00432-009-0741-y [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Gobitti C, Borsatti E, Arcicasa M, Roncadin M, Franchin G, Minatel E, et al. Treatment of recurrent high-grade gliomas with GliaSite brachytherapy: a prospective mono-institutional Italian experience. Tumori 2011; 97: 614–19. doi: 10.1700/989.10721 [DOI] [PubMed] [Google Scholar]
  • 43.Wernicke AG, Sherr DL, Schwartz TH, Pannullo SC, Stieg PE, Boockvar JA, et al. The role of dose escalation with intracavitary brachytherapy in the treatment of localized CNS malignancies: outcomes and toxicities of a prospective study. Brachytherapy 2010; 9: 91–9. doi: 10.1016/j.brachy.2009.06.005 [DOI] [PubMed] [Google Scholar]
  • 44.Hatanaka H, Nakagawa Y. Clinical results of long-surviving brain tumor patients who underwent boron neutron capture therapy. Int J Radiat Oncol Biol Phys 1994; 28: 1061–6. doi: 10.1016/0360-3016(94)90479-0 [DOI] [PubMed] [Google Scholar]
  • 45.ICRP. 1990 recommendations of the International Commission on Radiological Protection. ICRP Publication 60. Ann ICRP 1991; 21. [PubMed] [Google Scholar]
  • 46.Kageji T, Mizobuchi Y, Nagahiro S, Nakagawa Y, Kumada H. Clinical results of boron neutron capture therapy (BNCT) for glioblastoma. Appl Radiat Isot 2011; 69: 1823–5. doi: 10.1016/j.apradiso.2011.05.029 [DOI] [PubMed] [Google Scholar]
  • 47.Wittig A, Stecher-Rasmussen F, Hilger RA, Rassow J, Mauri P, Sauerwein W. Sodium mercaptoundecahydro-closo-dodecaborate (BSH), a boron carrier that merits more attention. Appl Radiat Isot 2011; 69: 1760–4. doi: 10.1016/j.apradiso.2011.02.046 [DOI] [PubMed] [Google Scholar]
  • 48.Aiyama H, Nakai K, Yamamoto T, Nariai T, Kumada H, Ishikawa E, et al. A clinical trial protocol for second line treatment of malignant brain tumors with BNCT at University of Tsukuba. Appl Radiat Isot 2011; 69: 1819–22. doi: 10.1016/j.apradiso.2011.04.031 [DOI] [PubMed] [Google Scholar]
  • 49.Nakai K, Yamamoto T, Aiyama H, Takada T, Yoshida F, Kageji T, et al. Boron neutron capture therapy combined with fractionated photon irradiation for glioblastoma: a recursive partitioning analysis of BNCT patients. Appl Radiat Isot 2011; 69: 1790–2. doi: 10.1016/j.apradiso.2011.03.049 [DOI] [PubMed] [Google Scholar]

Articles from The British Journal of Radiology are provided here courtesy of Oxford University Press

RESOURCES