Transmission FTIR derivative spectroscopy for estimation of furosemide in raw material and tablet dosage form

Máximo Gallignani a, Rebeca A Rondón a, José F Ovalles b,, María R Brunetto a
PMCID: PMC4629100  PMID: 26579407

Abstract

A Fourier transform infrared derivative spectroscopy (FTIR-DS) method has been developed for determining furosemide (FUR) in pharmaceutical solid dosage form. The method involves the extraction of FUR from tablets with N,N-dimethylformamide by sonication and direct measurement in liquid phase mode using a reduced path length cell. In general, the spectra were measured in transmission mode and the equipment was configured to collect a spectrum at 4 cm−1 resolution and a 13 s collection time (10 scans co-added). The spectra were collected between 1400 cm−1 and 450 cm−1. Derivative spectroscopy was used for data processing and quantitative measurement using the peak area of the second order spectrum of the major spectral band found at 1165 cm−1 (SO2 stretching of FUR) with baseline correction. The method fulfilled most validation requirements in the 2 mg/mL and 20 mg/mL range, with a 0.9998 coefficient of determination obtained by simple calibration model, and a general coefficient of variation <2%. The mean recovery for the proposed assay method resulted within the (100±3)% over the 80%–120% range of the target concentration. The results agree with a pharmacopoeial method and, therefore, could be considered interchangeable.

KEY WORDS: FTIR, Derivative spectroscopy, Furosemide, Frusemide, Pharmaceutical analysis

Abbreviations: API, active pharmaceutical ingredient; DMF, N,N-dimethylformamide; DS, derivative spectroscopy; FTIR, Fourier transform infrared; FUR, furosemide or frusemide; HPLC, high performance liquid chromatographic; MIR, mid infrared

Graphical abstract

A FTIR spectroscopic method has been developed for determining furosemide in solid dosage form. The method involves the extraction of furosemide with DMF and measurement in liquid phase mode. The second order spectrum of the major spectral band (SO2 stretching) with baseline correction was used for processing and quantitative measurement.

graphic file with name fx1.jpg

1. Introduction

Furosemide or frusemide (FUR) is a loop diuretic widely used in the treatment of congestive heart failure and edema1. As a commercial product, this active pharmaceutical ingredient (API) is administered basically as an oral tablet and an injectable solution2. FUR is a potent diuretic with some associated side effects. Additionally, FUR is included in the World Anti-Doping Agency׳s banned drug list3. Therefore, we are always concerned about the availability of a valid and simple analytical method for FUR determination in pharmaceutical dosage forms.

According to the reviews published by Espinosa-Bosch et al.4, 5, a variety of analytical methods has been proposed for determining FUR in drugs. Li et al.6 also mention various methods for determining FUR since introducing it to the pharmaceutical market. In the present decade, several methods including the resonance Rayleigh scattering technique6, spectrophotometric7, 8, 9, fluorescence10, 11, luminescence12 and liquid chromatography13, 14, 15, 16 have been proposed to quantify FUR in pharmaceutical samples.

The official methods for the determination of FUR as a bulk drug are based on titrimetry17, 18, 19, 20, 21 while the official methods for FUR determination in tablets are based on a liquid chromatographic method17 and an UV spectrophotometric method19, 22. However, until now, no direct Fourier transform infrared (FTIR) spectroscopy has been recommended for the quantitative determination of FUR in solid dosage forms.

Because of its inherent simplicity and availability in most pharmaceutical laboratories, FTIR is a technique which can be used in quality control laboratories not only for qualitative, but also for quantitative purposes. Since the wavenumber and the respective intensity of an absorption maximum depend on a particular group of the chemical structure of a substance, this measurement can be used to quantitatively determine a substance. Fig. 1 illustrates how rich the structure of FUR is in regards to diverse groups which absorb in the mid-IR (MIR) region.

Figure 1.

Figure 1

Chemical structure of 4-chloro-2-[(furan-2-ylmethyl)amino]-5-sulfamoylbenzoic acid (FUR).

For the past two decades, the transmission FTIR spectroscopy has undergone significant application for quantitative purposes, several in the pharmaceutical analysis field23, 24, 25, 26, 27, 28, 29, 30, 31, 32, 33. As suggested above, a FTIR spectrum can be used for verifying the identity and purity of API and for detecting any spectral interference originating from the excipients. A conventional calibration curve approach is used as a method for analyzing the FTIR absorption spectrum. However, it usually requires chemometric methods in order to improve the data quality whenever applicable. Derivative spectroscopy (DS) is one of the easiest pre-processing methods. It is a useful technique for finding bands hidden in broad spectral features. Virtually, all current FTIR spectrometers generate derivative spectra by mathematical means. Consequently, an additional instrument is not required. Specialized literature concerning DS has been amply described34, 35, 36.

Therefore, this study aims to develop a new method for the qualitative and quantitative determination of FUR by FTIR-DS in liquid cell-through transmission mode. The method uses the same extraction procedure of compendium pharmacopoeias. This approach represents a real alternative for not only identifying incoming raw materials, but also analyzing them quantitatively, and for assessing pharmaceutical quality of FUR tablets from multinational markets without using separating techniques. By comparison with compendial and non-compendial analytical methods, the proposed FTIR-DS method reduces solvent consumption and eliminates the use of reagents.

2. Materials and methods

2.1. Equipment

A Perkin-Elmer Spectrum 2000 FTIR spectrophotometer (Norwalk, CT, USA) equipped with a temperature stabilized deuterated triglycine sulfate detector, a KBr–Ge coated beam splitter, and a globar IR source was employed for FTIR spectra acquisition. A Wilmad Labglass FTIR liquid transmission cell (Buena, NJ, USA) was used for sampling. Its circular configuration was a general purpose liquid transmission cell. It involved ZnSe windows (32 mm diameter, 3 mm thick) and an appropriate PTFE spacer of 0.05 mm optical path length. The sample was introduced by filling ports using a 3 mL syringe.

2.2. Reagents and samples

FUR reference substance of 98.0% purity was purchased from Sigma-Aldrich (St Louis, MO, USA) and 99.05% FUR raw material purity was kindly supplied by Valmorca Lab (Mérida, Venezuela). N,N-Dimethylformamide (DMF) of 99.9% purity was purchased from Mallinckrodt (Paris, Kentucky, USA). Commercial pharmaceutical 20 mg/Tab and 40 mg/Tab FUR tablets were purchased from local market (Mérida, Venezuela) but were manufactured by different laboratories: Sanofi Aventis (Caracas, Venezuela), Proula (Lagunillas, Venezuela), PluxAndex (El Vigía, Venezuela), Meyer Daliol (Caracas, Venezuela) and Genéricos Rx (Caracas, Venezuela).

2.3. Preparation of a standard solution

A 40.0 mg/mL FUR stock solution was prepared in DMF and a 20.0 mg/mL diluted FUR stock solution was also prepared in DMF by diluting 1:1 (v/v). The stock solution was kept protected from light using amber colored flasks and refrigerated. After being diluted with DMF, increasing volumes of FUR stock solutions were quantitatively transferred to a set of 10 mL volumetric flasks to provide solutions within the 2.0–20.0 mg/mL concentration range. Final standard solutions were kept protected from light.

2.4. Preparation of a sample solution

Twenty tablets were accurately weighed, finely pulverized, and thoroughly mixed. An accurately weighed portion of pulverized tablets equivalent to 100 mg of FUR was transferred to a 10-mL volumetric flask, 9 mL of diluting solvent (DMF) was added, and was sonicated for 30 min. After cooling, the diluting solvent was added to bring the volume to the mark. Finally, it was mixed, centrifuged at 2500 rpm for 5 min, and filtered, discarding the first 2.5 mL of the filtrate. All sample solutions were filtered through 0.22 μm analytical Nylon filters (New Jersey, USA). Samples for the standard addition technique were obtained by increasing volumes from 0 mL to 9 mL of the 20 mg/mL FUR stock solution to a set of 10 mL volumetric flasks containing 1.0 mL of sample solution equivalent to 20 mg/mL FUR. They were then quantitatively diluted to obtain solutions within the nominal concentration range.

2.5. Method validation

Validation of the analytical method and procedures was conducted, unless otherwise specified herein, following a tutorial from LabCompliance, a private organization devoted to validation and qualification in analytical laboratories37.

3. Results and discussion

3.1. Spectral characterization of FUR in DMF

FUR is insoluble in water or in chloroform. It is freely soluble in acetone, DMF and alkali-metal hydroxide solutions. We selected DMF taking into account FUR solubility, volatility and IR transparency. This is considered a green chemical solvent due to its amenable properties, such as low toxicity, low water miscibility and low volatility38.

As can be expected, in order to use DMF as a solvent in the IR spectroscopy, a reduced optical path length was required. An optical path length of 0.05 mm resulted adequate for the purpose of finding transparency zones in the MIR region. A guide to path length selection and cell thickness, depending on analyte concentration in the sample solution, has been described by Stuart39. Using DMF as the background, transparency zones were found in the IR fingerprint range (ν˜/cm−1) as follows: 1370–1270, 1240–1120 and 1050–500 (Fig. 2).

Figure 2.

Figure 2

Liquid phase FTIR spectrum of DMF (spectrum line 1), FUR standard 10 mg/mL (spectrum line 2), and FUR pharmaceutical sample (spectrum line 3). The nominal concentration of the pharmaceutical product is equivalent to 10 mg/mL. Spectra are the average of 10 accumulated scans obtained with a nominal resolution of 4 cm−1 using DMF as background correction.

The FTIR spectrum of FUR obtained in the solid phase, KBr pellets, shows well defined bands with relatively high intensity in the 1800–400 cm−1 wavenumber range. Using the KBr pellet, comparison of the FUR analyte FTIR spectrum vs. the FUR reference spectrum from the International Pharmacopoeia website21 showed an excellent frequency correlation. The chemical sulfur group (Fig. 1) gives specificity to the FUR chemical structure. Indeed, SO2 and SO sulfur compound groups produced strong IR bands in the 1400–1000 cm−1 range. The expected S=O stretching bands for sulfonamides are as follows: 1390–1290 cm−1 range for SO2 asymmetric stretching; 1190–1120 cm−1 range for SO2 symmetric stretching; and 1060–1020 cm−1 range for S=O stretching39.

Therefore, taking into account the above information, the FUR FTIR spectrum, using DMF as background, showed two notable spectral bands in the MIR with maximum wavenumbers located at 1335 cm−1 and 1165 cm−1 (Fig. 2). The spectra also showed that the presence of excipients, at the stated concentrations in the formulation, did not have any additional IR bands interfering with the mentioned IR absorption bands belonging to FUR.

3.2. Selection of the analytical spectral band

The two FUR spectral bands in the MIR, referred to above, showed a slightly spectral baseline shift for FUR in the pharmaceutical formulation. Under these circumstances, we decided to use a pre-processing method. When analyzing spectral data, our research group commonly used the concept of derivatizing spectral data, one of the most used pre-processing methods in spectroscopy. It allows for minimizing unwanted variations such as baseline shift, enhancement of spectral significant differences, and eliminating interferences by broad band constituents34, 35.

The DS revealed that first-order spectrum, a plot of dA/dλ vs 1/λ, allowed eliminating baseline shifts. However, apparently the FUR standard spectrum did not highly resemble the FUR sample (figure not shown). Additional spectroscopic processing, using the second-order spectrum, d2A/d(1/λ)2 vs 1/λ, showed quite similar spectral bands for the analyte in the two analyzed solutions, sample and standard (Figure 2, Figure 3).

Figure 3.

Figure 3

The zeroth (A) and second (B) derivative spectra of FUR standard and FUR sample. Second order derivative of FUR as a function of the concentration (C).

Briefly, the negative bands of the second order derivative with minimum at the same maximum on the zero-order band were quite similar, sample and standard, coming from the two additional positive satellite bands, the first at 1174 cm−1 and the second at 1158 cm−1. This behavior was also observed for the other possible analytical band whose maximum was found at 1335 cm−1. Both signal to noise ratio and baseline shift corrections were the preponderant factors for discriminating between the two last bands. In this sense, the band at 1165 cm−1 provided the best precision, sensitivity, and quantitative features.

3.3. Selection of the analytical measurement criterion

Various measurement criteria were evaluated using the second-order derivative of the spectral band in the selected range (Fig. 3B). Taking into account the main features of a linear regression curve, we decided to select the area under the main second-order band denoted as DZL (1170–1158 cm−1) in Table 1.

Table 1.

Analytical figures obtained for FUR determination by FTIR-DS using different measurement modes.

Band parameters (cm−1)a
Linear regressionb Analytical parametersc
Criterion Baseline mode Peak location Y=a+b [X]
r LOQ RSD
a (×10−5) b (×10−3)
Peak area DSL 1165 15.4 2.52 0.9996 2.00 0.50
DSZ 1165 15.6 2.01 0.9995 2.12 0.72
DZL 1165 −10.6 1.68 0.9999 1.03 0.32
DZ 1165 −6.0 1.42 0.9999 1.14 0.60
Peak height DS 1174 1.61 0.105 0.9998 1.46 1.38
DB 1165 −0.41 0.347 0.9998 1.44 1.45
DL 1158 2.63 0.067 0.9997 1.56 1.55
DZ 1165 −2.11 0.254 0.9999 1.05 1.90
a

Fig. 3B illustrates the different possibilities to obtain the intensity of the derivative spectral band; DL is the long wavenumber peak satellite, DS is the short wavenumber peak satellite, and DB and DZ are the peak tangent baseline and the peak-derivative zero, respectively.

b

Y=2nd derivative signal; X=FUR concentration ranging from 2.0 mg/mL to 20.0 mg/mL (n=10).

c

LOQ, limit of quantification in mg/mL calculated as 10σ/slope; RSD, relative standard deviation established at the target concentration; r=correlation coefficient.

3.4. Effect of FTIR instrument conditions

A monoparametric study was performed by monitoring the selected analytical signal for each variable by triplicate. As a whole, a nominal resolution of 4 cm−1, selected from 2 cm−1 to 32 cm−1 range, by accumulating 10 scans, selected from 1 scan to 50 scans, was chosen mainly as a compromise between precision of <1.0% RSD and spectral acquisition time <60 s. On the other hand, as a compromise with respect to sensitivity, precision, and band shape, the selected derivative window for analytical purposes was 9 points, selected from a 5 points to 37 points range, with a related precision of 0.27% RSD, n=3.

3.5. FTIR-DS system analytical figures of merit

3.5.1. Linearity

Data from the simple regression line described in Table 2 demonstrated acceptable FUR linearity over 20%–200% range of the target concentration (10 mg/mL) with a quantification limit of 1.0 mg/mL. The Y-intercept was negative but not significantly different from zero, P=0.05. Furthermore, the y-intercept variability resulted less than 1% of the response obtained at the target level which is usually an acceptable criterion40. The statistical significance of the regression line (r2=0.9998) also revealed that the slope is not zero, P<0.05, which in turn, suggested that changes in the concentration variable are highly associated with changes in the response variable (Table 2).

Table 2.

Analytical performance characteristics and system suitability of the proposed FTIR-DS method for determining FUR.

Parameter Result Observation
Dynamic linear range (mg/mL) 1.0–40.0 20–40 mg (Tab)
Working linear range (mg/mL) 2.0–20.0
Detector linearity (r2) 0.9998 Criterion: >0.9990
External linear regression Y=−1.1×10−5+1.68×10−3 [X] Y=D2−peak area;
X=concentration
SD of the response 0.00017 σ
Limit of quantification (mg/mL) 1.03 (10σ)/slope
Correlation coefficient 0.99988 Criterion: ≥0.999
Statistical significance of Y-intercept −0.90<2.36 µα=0;
(P<0.05) tExp<tTab; tn−2
Statistical significance of slope 175>2.36 µβ=0;
(P<0.05) tExp<tTab; tn−2
Instrument precision data from 10 replicate determinations 0.23% Acceptance criteria: ≤1.0% (RSD)
Instrument precision data from 6 replicate determinations 0.31% Acceptance criteria: ≤2.0% (RSD)
(pharmaceutical sample)
Intra-assay precision by analyzing aliquots of a homogeneous sample, independently prepared 0.45% for 8 mg/mL (n=3) Acceptance criteria: ≤2.0% (RSD)
0.30% for 10 mg/mL (n=3)
0.38% for 12 mg/mL (n=3)
Intermediate precision: Acceptance criteria: 2.0–5.0% (RSD)
Operator 1, day 1 and day 2 0.83%
Operator 2, day 1 and day 2 1.6%
Operators, day 1 1.3%
Operators, day 2 1.2%
Analytical frequency (sample/h) 25 Manual sampling
Stability of analytical solutions (coefficient of variation) 1.2% (n=13) Cell removed after each measurement
Test period: 10–120 min.
Time extraction during sonication depending on FUR concentration in sample to be prepared <10 min 1 mg/mL
<15 min 10 mg/mL
<70 min 40 mg/mL
Sample carry over by filling the cell with a high standard concentration 0.16% Acceptance criteria: <2.0% (RSD)
(followed by a blank solvent injection)

Standard deviation, SD. Relative standard deviation, RSD. Symbol “σ” represents SD(Y/X). Tablets, Tab. Second order derivative, D2.

3.5.2. Precision

As can be seen in Table 2, data for instrument precision and repeatability met the requirements for a method to be precise37.

3.5.3. Selectivity and specificity

The fingerprint region is often the most complex and confusing region to interpret. However, its utility is that many bands provide a molecule pattern. Fortunately, FUR has an inorganic group, S=O, which absorbs in the IR fingerprint region giving intense vibrational bands, fewer in number and broader than those observed for organic materials. These features generate a characteristic IR spectrum. By appropriately selecting one of these large bands, the method provided some grade of specificity. The use of DS allowed demonstrating that the selected spectral region for analytical purposes only belonged to the API (Fig. 2). Nonspecific background interference and baseline shift were automatically corrected by the conversion of the zero-order derivative spectrum into the second-order derivative spectrum, thus improving accuracy of quantification.

The standard addition technique did not evidence spectral interference due to concomitant absorbing species coming from excipients. It is important to remember that the analyte is extracted from the excipient using an organic solvent, where FUR is very soluble. This cannot be said for the known common excipients such as maize starch, talc, magnesium stearate, sodium starch glycollate, and colloidal silicon dioxide, except for lactose, which is very soluble in DMF2. Preliminary studies showed a lactose transmission FTIR spectrum with absorption intensity values in the 1150–950 cm−1 region, where solvent absorption saturation was observed. The proximity of lactose absorption, characteristic IR absorption of carbohydrate moieties, to FUR absorption only caused displacement of the baseline (1158–1000 cm–1). Depending on the pharmaceutical brand type, displacement was caused in one more than the other (Fig. 2). This spectral interference was corrected later on using second order DS.

Furthermore, as can be seen in Figure 3, Figure 4, supplementary studies demonstrated the viability of the intended objective of the analytical procedure. The visual spectral evaluation showed similar band shapes for all the commercial samples. Also, they were quite similar to the standards. Certainly, at the level of the satellite bands of the second derivative spectra, some spectral differences were observed from sample to sample. Nevertheless, they were attributed to small variations in concentration (Fig. 3C) more than any spectral interference from the excipients (Fig. 4).

Figure 4.

Figure 4

Representative FUR spectra in various commercial brands and a FUR standard as reference. FUR standard, 8 mg/mL (spectrum line 1), generic drug 1 (spectrum line 2), innovator drug (spectrum line 3), brand drug 1 (spectrum line 4), and generic drug 2 (spectrum line 5). Nominal concentration of pharmaceuticals represents 10 mg/mL FUR.

3.5.4. System suitability

Generalizing, Table 2 shows the analytical performance characteristics of the FTIR-DS method for determining FUR. In this particular case, the table also shows some required analysis factors used in order to guarantee system suitability, such as stability of the sample solutions and percentage of extraction by sonication. This last procedure, using DMF, is official in most compendium pharmacopoeias. Although, in a lengthy procedure, several samples can be extracted simultaneously, even during the pre-conditioning of the analytical equipment. Variation coefficients within an acceptable range demonstrated that factors, which could affect robustness of the method, had a negligible effect on the analytical response (Table 2). Other parameters were determined via robustness (Section 3.5.5).

3.5.5. Robustness

In principle, the proposed method could be considered as very simple. Consequently, it involved few parameters that could affect the method performance. FTIR analytical procedures are usually protected from changes in humidity and temperature. Therefore, they were kept controlled throughout the methodological development. The sample was introduced via filling ports. In this regard, signal response variation due to cuvette manipulation was kept to a minimum by prolonging sample application on the entrance port. That is, the sample solution was passed through the cell interior for at least 30 s at 0.5 mL/min. Additionally, when the cell was filled with a high standard concentration, followed by a blank solvent injection, the variation coefficient due to sample carry-over was very low (Table 2). Parameters such as pH and composition of the dissolvent were not involved in the method development. Possible DMF water content was corrected by the background and the selected mode of quantification, i.e. by DS. In conclusion, the analytical conditions can be adequately controlled and small changes in method parameters will show negligible effects on the analytical response.

3.5.6. Accuracy

Samples spiked with known quantities of the API were analyzed. The standard addition calibration test of significance for the innovator product showed that the estimated y-intercept was different from zero (P<0.05) using H0: µα=0 (tExp<tTab), the result being 27>2.36. A similar result was observed for the y-intercept standard addition calibration for one of the representative generic products, 20>2.36. On the other hand, the standard addition slope belonging to the innovator product was within the confidence limits of the slope of reference (1.68×10−3±2.3×10−4) at the 0.05 level of significance. An identical result was observed for the standard addition slope of one representative generic product (Table 3). The relative percentage deviation between the reference slope (b) belonging to the external calibration and the other two slopes, (b1) and (b2), was less than 1.2%. This indicated that the sensitivity of the analyte was only slightly affected by the presence of any interference coming from the inactive substance formulated alongside the API.

Table 3.

Results of assessment of the linearity of the FTIR-DS system for the assay method of FUR employing the addition standard technique.

Curve type Linear regression: Y=a+b [X]a
a±SD; (×10−3) b±SD; (×10−3) r SD(y/x)
External (Reference standard) −0.11±0.11 1.68±0.0096 0.99988 0.00017
Standard addition 1 (Innovator) 3.78±0.14 1.70±0.014 0.99976 0.00024
Standard addition 2 (Generic) 3.98±0.20 1.70±0.021 0.99949 0.00034
a

Calibration, where Y=peak area of the 2nd derivative signal; X=FUR concentration in mg/mL (n=9); SD, standard deviation; r=correlation coefficient.

Furthermore, assay percentage recovery of known added amounts of API in the pharmaceutical sample was also explored (Table 4). The mean recovery for the proposed assay method resulted within (100±3)% over the 80%–120% range of the target concentration for the innovator and generic pharmaceutical products. Considering that the expected recovery depends on the sample matrix, the sample processing procedure and the analyte concentration, the accuracy of the proposed method can be considered suitable for determining FUR content in solid dosage pharmaceutical products.

Table 4.

Recovery of FUR from pharmaceutical samples with known concentration using the proposed FTIR-DS method.

Type of brand Sample No. Concentration of FUR (mg/mL); n=3
Recovery (%) Recovery (%)
Nominal Added Found Mean Acceptable a
Innovator 1 2.21 4.0 6.40 104.8 102.3±1.9 97–103
2 8.0 10.41 102.6
3 10.0 12.59 102.8
4 12.0 14.16 99.6
5 18.0 20.50 101.6



Generic 1 2.39 4.0 6.58 104.8 101.2±2.4 97–103
2 8.0 10.50 101.4
3 10.0 12.26 98.8
4 12.0 14.28 99.1
5 18.0 20.72 101.9
a

The acceptable recovery is for the analysis of 1% (w/v) active ingredient37. The accuracy criteria for an assay method (FDA) is that the mean recovery will be (100±2)% at each concentration over the range of 80%–120% of the target concentration40.

Additionally, an official reference assay method was used to estimate the accuracy of the proposed method. In the global F test from the ANOVA table, we obtained a F-ratio equal to 0.036 and a P value greater than 0.05. Therefore, we concluded that there was insufficient evidence to assume any bias between the methods at the 95% confidence level (Table 5).

Table 5.

Analytical application of the proposed FTIR-DS method for determining FUR in solid dosage pharmaceutical products.

Type of branded drug Concentration (mg/Tab)±SD; n=6
Labeled Found Reference HPLC methoda
FTIR-DS method Found Accepted range
Innovator 1 20 20.9±0.30 19.0±0.10 18–22
Generic 1 40 39.9±0.31 40.2±0.56 36–44
Brand 1 40 41.3±0.10 41.6±0.60 36–44
Brand 2 40 40.8±0.20 39.69±0.77 36–44
Innovator 2 40 41.9±0.70 36–44
Generic 2 40 44.9±0.31 36–44
Generic 3 40 39.6±0.20 36–44
Bulkb 99.05 99.0±0.26b 98–101
a

USP17. The symbol (–) means that it was not analyzed by the method of reference.

b

It represents dried substance or raw material powder. USP acceptance range in % w/w17.

4. Conclusions

The developed method for determining FUR based on the application of transmission FTIR in the liquid phase fulfilled most validation requirements in a range of concentrations suitable for quality control of both incoming raw materials and solid dosage forms (Table 2, Table 5). Therefore, this is a pharmaceutical industry quality control method that could be considered interchangeable with other pharmacopoeial methods. As referenced in practically all compendium pharmacopoeias, the major use of FTIR in the transmission mode is for the identification of API in raw materials and finished products. Regardless of the many advantages described in the last two decades28, FTIR equipment in the pharmaceutical field for qualitative issues is widely available but still undervalued for quantitative purposes. We believe that FTIR instruments in conjunction with DS are ideal tools for pharmaceutical quality control. By comparison with compendial and non-compendial analytical methods, the proposed FTIR-DS method reduces solvent consumption and eliminates the use of reagents. Additionally, the proposed FTIR-DS method allows off-site re-analysis of spectra for other chemicals not originally targeted. Finally, by selecting different measurement spectral bands, this study opens up the possibility of applying the proposed FTIR-DS method to quantify FUR when combined with other API in the same dosage form.

Acknowledgments

The authors gratefully acknowledge the financial support of the Council of Scientific, Humanistic, Technological and Artistic Development (CDCHTA) of the University of Los Andes from Venezuela and the National Foundation for Science and Technology (FONACIT) also from Venezuela for providing financial support throughout Projects FA-371-06-08-B and G-2005000641, respectively. The authors would also like to thank Michele Lee, Director of the Venezuelan American Center of Merida (CEVAM), for refining the use of English in our manuscript.

Footnotes

Peer review under responsibility of Institute of Materia Medica, Chinese Academy of Medical Sciences and Chinese Pharmaceutical Association.

References

  • 1.Darandale SS, Vavia PR. Design of a gastroretentive mucoadhesive dosage form of furosemide for controlled release. Acta Pharm Sin B. 2012;2:509–517. [Google Scholar]
  • 2.Drugs.com. Furosemide. Auckland, New Zealand: Drugsite Trust; 2014. Available from: 〈http://www.drugs.com/monograph/furosemide.html〉.
  • 3.Thevis M, Kuuranne T, Geyer H, Schänzer W. Annual banned-substance review: analytical approaches in human sports drug testing. Drug Test Anal. 2014;6:164–184. doi: 10.1002/dta.1591. [DOI] [PubMed] [Google Scholar]
  • 4.Espinosa Bosch M, Ruiz Sánchez AJ, Sánchez Rojas F, Bosch Ojeda C. Recent developments in analytical determination of furosemide. J Pharm Biomed Anal. 2008;48:519–532. doi: 10.1016/j.jpba.2008.07.003. [DOI] [PubMed] [Google Scholar]
  • 5.Espinosa Bosch M, Ruiz Sánchez AJ, Sánchez Rojas F, Bosch Ojeda C. Analytical determination of furosemide: the last researches. Int J Pharm Biol Sci. 2013;3:168–181. [Google Scholar]
  • 6.Li CX, Liu SP, Liu ZF, Hu XL. The interaction between furosemide-palladium (II) chelate and basic triphenylmethane dyes by resonance Rayleigh scattering spectra and resonance non-linear scattering spectra and their analytical applications. Sci China Chem. 2010;53:1767–1777. [Google Scholar]
  • 7.Tharpa K, Basavaiah K, Basavaiah Vinay K. Use of a diazocoupling reaction for sensitive and selective spectrophotometric determination of furosemide in spiked human urine and pharmaceuticals. Chem Pap. 2010;64:415–423. [Google Scholar]
  • 8.Mehrotra A, Malviya G, Maheshwari RK. Application of mixed hydrotropy in spectrophotometric analysis of frusemide in different formulations. Bull Pharm Res. 2011;1:15–21. [Google Scholar]
  • 9.Patel H, Solanki S. Development and validation of spectrophotometric methods for simultaneous estimation of furosemide and spironolactone in combined tablet dosage form. Int J Pharm Pharm Sci. 2012;4:383–386. [Google Scholar]
  • 10.Buitrago A, Calderón L, León A, Brunetto R, Gallignani M. Desarrollo y validación de un método espectrofluorométrico para la determinación de furosemida en formas farmacéuticas sólidas. Avances en Química. 2010;5:15–25. 〈http://www.saber.ula.ve/bitstream/123456789/31054/1/articulo2.pdf〉 Available from. [Google Scholar]
  • 11.Peralta CM, Fernández LP, Masi AN. A novel application of immobilization on membranes for the separation and spectrofluorimetric quantification of amiloride and furosemide in pharmaceutical samples. Anal Chim Acta. 2010;661:85–90. doi: 10.1016/j.aca.2009.12.015. [DOI] [PubMed] [Google Scholar]
  • 12.Liu Y, Wang H, Wang J, Li Y. A simple and sensitive spectrofluorimetric method for the determination of furosemide using zinc(II)-1,4-bis(imidazol-1-ylmethyl)benzene complexes. Luminescence. 2013;28:882–887. doi: 10.1002/bio.2451. [DOI] [PubMed] [Google Scholar]
  • 13.Nagori BP, Solanki R. RP-HPLC method for simultaneous estimation of frusemide and amiloride hydrochloride in tablet formulation. Indian J Pharm Sci. 2010;72:384–387. doi: 10.4103/0250-474X.70491. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Ram VR, Dave PN, Joshi HS. Development and validation of a stability-indicating HPLC assay method for simultaneous determination of spironolactone and furosemide in tablet formulation. J Chromatogr Sci. 2012;50:721–726. doi: 10.1093/chromsci/bms062. [DOI] [PubMed] [Google Scholar]
  • 15.Youm I, Youan B-BC. Validated reverse-phase high-performance liquid chromatography for quantification of furosemide in tablets and nanoparticles. J Anal Methods Chem. 2013;2013:1–9. doi: 10.1155/2013/207028. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Kaynak MS, Sahin S. Development and validation of a RP-HPLC method for determination of solubility of furosemide. Turk J Pharm Sci. 2013;10:25–34. [Google Scholar]
  • 17.United States Pharmacopeial Convention, (Furosemide). United States Pharmacopeia, 32nd ed., 2009, United States Pharmacopeial Convention; Rockville, MD, USA 〈http://www.uspbpep.com/search.asp〉, (Available from)
  • 18.European Pharmacopoeia Commission, (Furosemide)European Pharmacopoeia, 6th ed., 2007, European Pharmacopoeia Commission; Strasbourg, France 〈http://www.uspbpep.com/search.asp〉, (Available from).
  • 19.Society of Japanese Pharmacopoeia, (Furosemide)Japanese Pharmacopoeia, 16th ed., 2011, Society of Japanese Pharmacopoeia; Japan 〈http://jpdb.nihs.go.jp/jp16e/〉, (Available from).
  • 20.The Indian Pharmacopoeia Commission, (Furosemide)Indian Pharmacopoeia 2, 2011, The Indian Pharmacopoeia Commission; Ghaziabad, India 〈http://ajprd.com/downloadebooks_pdf/9.pdf〉, (Available from).
  • 21.WHO Department of Essential Medicines and Pharmaceutical Policies, (Furosemide)The International Pharmacopoeia, 4th ed., 2011, WHO Department of Essential Medicines and Pharmaceutical Policies; New Zealand 〈http://apps.who.int/phint/en/p/docf/anchor,get-started.html〉, (Retrieved February 10, 2014. Available from).
  • 22.British Pharmacopoeia Commission. The Stationery Office, (Furosemide)British Pharmacopoeia III, 2011, British Pharmacopoeia Commission. The Stationery Office; London, UK 〈http://bp2012.infostar.com.cn/Bp2012.aspx〉, (Available from).
  • 23.Schindler R, Lendl B. FTIR spectroscopy as detection principle in aqueous flow analysis. Anal Commun. 1999;36:123–126. [Google Scholar]
  • 24.Gallignani M, Brunetto MR. Infrared detection in flow analysis – developments and trends (review) Talanta. 2004;64:1127–1146. doi: 10.1016/j.talanta.2004.06.035. [DOI] [PubMed] [Google Scholar]
  • 25.Moros J, Garrigues S, de la Guardia M. Quality control Fourier transform infrared determination of diazepam in pharmaceuticals. J Pharm Biomed Anal. 2007;43:1277–1282. doi: 10.1016/j.jpba.2006.10.036. [DOI] [PubMed] [Google Scholar]
  • 26.Ovalles F, Gallignani M, Rondón R, Brunetto MR, Luna R. Determination of sulphate for measuring magnesium sulphate in pharmaceuticals by flow analysis-Fourier transforms infrared spectroscopy. Lat Am J Pharm. 2009;28:173–182. [Google Scholar]
  • 27.Ayala C, Brunetto MR, Ovalles F, Gallignani M. Determinación de atenolol en productos farmacéuticos por espectrometría infrarroja con transformada de Fourier (FTIR) Rev Tec Ing Univ Zulia. 2009;32:238–248. [Google Scholar]
  • 28.Cadet F, Garrigues S, de la Guardia M. Quantitative analysis, infrared. In: Meyers RA, editor. Encyclopedia of analytical chemistry. John Wiley & Sons; Chichester: 2012. [Google Scholar]
  • 29.Konoz E, Mohsen Sarrafi AH, Samadizadeh M, Boreiri S. Quantitative analysis of lorazepam in pharmaceutical formulation through FTIR spectroscopy. E-J Chem. 2012;9:2232–2238. [Google Scholar]
  • 30.Mallah M, Sherazi S, Mahesar S, Khaskheli A. Simultaneous quantification of ibuprofen and paracetamol in tablet formulations using transmission Fourier transform infrared spectroscopy. Am J Anal Chem. 2012;3:503–511. [Google Scholar]
  • 31.Ovalles JF, Gallignani M, Rondón R, Ayala C, Brunetto MR. Proposal for determining sulfate counter ion in amikacin sulfate formulations by Fourier-transform infrared derivative spectroscopy. Curr Pharm Anal. 2013;9:20–30. [Google Scholar]
  • 32.Ovalles JF, Gallignani M, Brunetto MR, Rondón RA, Ayala C. Reagent-free determination of amikacin content in amikacin sulfate injections by FTIR derivative spectroscopy in a continuous flow system. J Pharma Anal. 2014;4:125–131. doi: 10.1016/j.jpha.2013.08.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Robaina NF, de Paula CER, Brum DM, de la Guardia M, Garrigues S, Cassella RJ. Novel approach for the determination of azithromycin in pharmaceutical formulations by Fourier transform infrared spectroscopy in film-through transmission mode. Microchem J. 2013;110:301–307. [Google Scholar]
  • 34.El-Sayed AA, El-Salem NA. Recent developments of derivative spectrophotometry and their analytical applications. Anal Sci. 2005;21:595–614. doi: 10.2116/analsci.21.595. [DOI] [PubMed] [Google Scholar]
  • 35.Bosch Ojeda C, Sanchez Rojas F. Recent applications in derivative ultraviolet/visible absorption spectrophotometry: 2009–2011: a review. Microchem J. 2013;106:1–16. [Google Scholar]
  • 36.Moros J, Garrigues S, de la Guardia M. Vibrational spectroscopy provides a green tool for multi-component analysis. Trends Anal Chem. 2010;29:578–591. [Google Scholar]
  • 37.Huber L. Lab-Compliance tutorials; Obere Alm/Oberkirch, Germany: 2007. Validation of analytical methods and procedures.〈http://www.labcompliance.com/tutorial/methods/default.aspx〉 Available from. [Google Scholar]
  • 38.Valavanidis A, Vlachogianni T. Synchrona Themata; Athens: 2012. Green chemistry and green engineering. [Google Scholar]
  • 39.Stuart B. John Wiley & Sons; Chichester: 2004. Infrared spectroscopy: fundamentals and applications. [Google Scholar]
  • 40.Shabir GA. A practical approach to validation of HPLC methods under current good manufacturing practices. J Valid Technol 2004:10:210–8. Available from. 〈http://www.ivtnetwork.com/article/hplc-resources〉

Articles from Acta Pharmaceutica Sinica. B are provided here courtesy of Elsevier

RESOURCES