Abstract
The K2 Summit camera was initially the only commercially available direct electron detection camera that was optimized for high-speed counting of primary electrons and was also the only one that implemented centroiding so that the resolution of the camera can be extended beyond the Nyquist limit set by the physical pixel size. In this study, we used well-characterized two-dimensional crystals of the membrane protein aquaporin-0 to characterize the performance of the camera below and beyond the physical Nyquist limit and to measure the influence of electron dose rate on image amplitudes and phases.
Keywords: Direct electron detection camera, electron counting, super-resolution, two- dimensional crystal
Introduction
One of the most exciting technological breakthroughs in cryo-electron microscopy (cryo-EM) in recent years has been the development and application of complementary metal-oxide-semiconductor (CMOS)-based direct electron detection device (DDD) cameras (Faruqi and McMullan, 2011). The detective quantum efficiency (DQE) of these cameras is significantly higher, at both low and high spatial frequency, than those of the more traditionally used image recording media, photographic film and scintillator-based digital cameras such as charge-coupled device (CCD) cameras (Li et al., 2013a; McMullan et al., 2009a; McMullan et al., 2009b). A number of recent studies have characterized DDD cameras and demonstrated that these cameras are superb for high-resolution cryo-EM (Bammes et al., 2012; Li et al., 2013a; McMullan et al., 2009a; Milazzo et al., 2011; Ruskin et al., 2013). The high DQE at high spatial frequency helps retain high-resolution information while the high DQE at low spatial frequency improves the image contrast needed for particle detection and alignment. The improved DQE makes it possible to record images of frozen-hydrated biological samples with sufficient contrast using a smaller defocus than previously required for imaging on photographic film or scintillator-based cameras. The high output frame rate of DDD cameras also enables recording of dose-fractionated image stacks (movies), and the correction of motion-induced image blurring (Bai et al., 2013; Campbell et al., 2012; Li et al., 2013a). The application of DDD cameras and the associated dose-fractionation movie technology has resulted in a number of high-resolution single-particle cryo-EM three-dimensional (3D) reconstructions at near-atomic resolution (e.g., Allegretti et al., 2014; Amunts et al., 2014; Campbell et al., 2012; Li et al., 2013a; Li et al., 2013b; Liao et al., 2013).
Among several commercially available DDD cameras, the K2 Summit camera from Gatan was the first one that had the capability of counting individual electron events in a practical manner (Li et al., 2013a). By identifying primary electron events, the electron counting process nearly doubles low-resolution DQE (Li et al., 2013a). It also removes Landau noise, which is generated from the statistical deposition of energy by the primary electrons. Counting also eliminates the readout noise and therefore, there is no penalty when the total dose is fractionated into multiple subframes. The signal cluster of each primary electron event can be further analyzed to identify its centroid, allowing its entry point to be assigned to a quadrant of the physical pixel. While the principle of centroiding is the same as the method used in light microscopy (Shroff et al., 2007), the exact algorithm used by the K2 Summit camera is proprietary information. However, much of the key behavior was deduced by direct analysis (Li et al., 2013b). In such super-resolution images, the effective Nyquist frequency is extended to twice the physical Nyquist frequency of the camera, thereby reducing aliasing. Counting and centroiding result in a significant improvement in image quality and resolution. For a detailed comparison of the K2 camera with other currently available DDD cameras, see Ruskin et al. (2013) and McMullan et al. (2014).
A potential drawback of a counting camera is that the usable dose rate on the camera is limited by coincidence loss, i.e., when more than one electron strikes the same pixel or neighboring pixels in the same frame, only one is counted and the additional electrons are ignored. While the K2 camera has a very high internal frame rate (400 frames/s), coincidence loss becomes significant at dose rates above ~ 8 e-/pixel/s (Li et al., 2013a). Counting images have now resulted in numerous 3D maps at near-atomic resolution including the recent map of β-galactosidase at a nominal resolution of 2.2 Å (Bartesaghi et al., 2015), but the data have largely been recorded under conditions for which the obtained resolutions are not substantially better than the physical Nyquist of the camera. Thus, it is not clear if the centroiding operation accurately preserves image amplitude and phase information.
In this work, we used two-dimensional (2D) crystals of aquaporin-0 (AQP0) as a test specimen to further characterize the influence of dose rate and centroiding on image phases and amplitudes, particularly beyond the physical Nyquist frequency. We recorded images of 2D crystals at different magnifications and different dose rates. Phase errors from merging the images recorded under different conditions provide a quantitative measure to assess the influence of dose rate and centroiding on image phases. We demonstrate that imaging with the K2 Summit camera operated in super-resolution mode does not increase the phase error of merged reflections, suggesting that the phase error introduced by centroiding and coincidence loss at low dose rates, if there is any, is sufficiently small and tolerable for high-resolution imaging. We find, however, that in images recorded at high dose rates in super-resolution mode, coincidence loss results in large amplitude errors in the low-resolution range (Li et al., 2013b). (This is likely also the case for images recorded at high dose rates in counting mode).
Centroiding quadruples the total number of physical pixels of the camera. A global motion correction algorithm can correct beam-induced motion with sub-physical pixel precision and restore image resolution (measured by amplitude) to beyond the physical Nyquist frequency (Li et al., 2013a). Although the DQE drops significantly beyond the physical Nyquist frequency, the use of super-resolution pixels could potentially further increase the efficiency of cryo-EM data acquisition. While there may be other factors that limit the resolution of single-particle cryo-EM images recorded at low magnification, we demonstrate here that images of 2D crystals recorded with the K2 Summit camera preserve phase information to near the super-resolution Nyquist frequency. Furthermore, we performed dose fractionation experiments to investigate if the loss of structural integrity of AQP0 crystals due to radiolysis under the electron beam depends on the timing of the fractional doses.
Materials and Methods
Preparation of AQP0 2D crystals
AQP0 was purified and crystallized as described previously (Gonen et al., 2004). Briefly, membranes were isolated from the core of sheep lenses (purchased from Wolverine Packing Company, Detroit, MI), washed, and solubilized with 4% (w/v) n-octyl-β,D-glucoside (OG) (Affymetrix). The insoluble fraction was removed by centrifugation at 300,000 g for 45 min at 4ºC, and the supernatant was run over a MonoQ ion-exchange column (GE Healthcare). Bound proteins were eluted with 150 mM NaCl in 1.2% (w/v) OG and 10 mM Tris-Cl, pH 8.0. Peak fractions were pooled and run over a Superose 12 column (GE Healthcare) in 1.2% (w/v) OG, 10 mM Tris-HCl, pH 8.0, and 100 mM NaCl. Purified AQP0 was first mixed with a 1:2 (mol/mol) mixture of sphingomyelin and cholesterol (Avanti Polar Lipids) at a lipid-to-protein ratio (LPR) of 0.2 or with a 4:1 (w/w) mixture of dimyristoyl phosphatidylethanolamine (DMPE) and dimyristoyl phosphatidylglycerol (DMPG) (Avanti Polar Lipids) at an LPR of 0.6. Both mixtures were dialyzed against 10 mM MES, pH 6.0, 30 mM MgCl2, 100 mM NaCl, and 0.05% NaN3 at 37ºC for one week with daily buffer exchanges.
Grid preparation and data collection
AQP0 2D crystals were prepared on molybdenum grids with 7% (w/v) trehalose solution using a modified version of the carbon sandwich method (Gyobu et al., 2004; Hite et al., 2010b), and the grids were frozen in liquid nitrogen. The grids were transferred into a Tecnai F20 electron microscope (FEI Company, Hillsborough, OR) using an Oxford CT3500 side-entry cryo-specimen holder. Data on the Tecnai F20 were collected at an acceleration voltage of 200 kV, and a nominal defocus range of −0.5 to −0.8 μm. Crystals were also imaged with a Polara electron microscope (FEI Company, Hillsborough, OR) operated at an acceleration voltage of 300 keV, using a nominal defocus range of −0.2 to −0.5 μm. Data were collected using low-dose procedures and K2 Summit cameras (Gatan Inc., Pleasanton, CA). The physical pixel size of the K2 Summit camera is 5 μm. The chip size of the camera sensor is 3838 x 3710 pixels, and images were reduced to a square size of 3710 x 3710 pixels.
Dose-fractionated image stacks were recorded in super-resolution mode following established procedures (Li et al., 2013a). To set the desired dose rate, first gain and dark references were prepared in the linear and super-resolution modes, and then the “Profile” option of the Gatan Digital Micrograph software was used to adjust the beam intensity and set up the counting rate using an area containing a 2D crystal. On the Tecnai F20, for images collected at a calibrated magnification of 40,410x (nominal magnification of 29,000x), a dose rate of 8 counts/pixel/s (5.2 counts/Å2/s) was used. Frames were read out every 150 ms and 27 frames were collected, resulting in an exposure time of 4.05 s and a total dose of 21 counts/Å2. For images collected at a calibrated magnification of 15,858x (nominal magnification of 11,500x), a dose rate of 10 counts/pixel/s (1.04 counts/Å2/s) was used. Frames were read out every 400 ms and 50 frames were collected, resulting in an exposure time of 20 s and a total dose of 20 counts/Å2. On the Polara, dose-fractionated images in super-resolution mode were recorded at liquid nitrogen temperature at a calibrated magnification of 50,926x (nominal magnification of 39,000x), and a dose rate of 8 counts/pixel/s (8.33 counts/Å2/s) was used. Frames were read out every 150 ms and 16 frames were collected, resulting in an exposure time of 2.4 s and a total dose of 20 counts/Å2.
Image stacks used to study dose-dependent radiation damage were recorded on a K2 Summit camera mounted on an FEI TF30 microscope operated at 300 kV, and using a calibrated magnification of 50,637x (nominal magnification of 39,000x), giving a super-resolution pixel size on the specimen level of 0.494 Å. All movies contained a total of 20 frames with 2.5 counts/Å2/frame, of which the final two frames were discarded. For dataset “6 parts”, the image stacks were recorded in six bursts containing three frames each. Bursts were separated by pauses of ~30 s. Image stacks for dataset “18 parts” were recorded as 18 individual frames separated by ~15 s pauses.
Image processing
The UCSF Image software was used to collect dose-fractionated image stacks and to align the frames prior to summing them up as described (Li et al., 2013a). The drift-corrected images of the AQP0 2D crystals were processed and merged using the 2dx software (Gipson et al., 2007a; Gipson et al., 2007b), which is based on the MRC image-processing package (Crowther et al., 1996). The 2D crystals have lattice parameters of a = 65.5 Å, b = 65.5 Å, and γ = 90°, and p422 plane symmetry. The crystal lattices were unbent, and the images were corrected for the contrast-transfer function (CTF), which was determined with CTFFIND3 (Mindell and Grigorieff, 2003). For all datasets, the best seven images of well-diffracting crystals were merged. The merging statistics in resolution ranges are listed in Supplementary Tables 1–11.
Image stacks recorded to assess dose-dependent radiation damage were processed as described (Li et al., 2013a). Frames of movies that included pauses were combined into single stacks for frame alignment. For each image series, six drift-corrected frame averages were calculated from groups of three frames each. Processing of the averages using the MRC image processing suite (Crowther et al., 1996) yielded lists of indexed amplitudes that were tabulated according to resolution shell and total dose received. Using only reflections with IQ values of 4 and lower (corresponding to an SNR of about 1.8 and higher), average amplitude ratios were calculated with respect to the measurement corresponding to a total dose of 7.5 counts/Å2, as well as the standard errors of the ratios. The results are plotted in Figure 7.
Results and Discussion
Imaging AQP0 2D crystal
AQP0 forms very well-ordered 2D crystals with a variety of lipids (Gonen et al., 2004; Gonen et al., 2005; Hite et al., 2010a). For this study, we used 2D crystals of AQP0 reconstituted with lipid mixtures of 1:2 (mol/mol) sphingomyelin and cholesterol, and 4:1 (w/w) phosphatidylethanolamine (PE) and phosphatidylglycerol (PG). The lattice parameters as well as the order of these 2D crystals were the same as those of AQP0 2D crystals obtained before with other lipids.
Dose-fractionated image stacks of AQP0 2D crystals were first recorded with a K2 Summit camera mounted on an FEI Tecnai F20 electron microscope operated at an acceleration voltage of 200 kV. At a calibrated magnification of 40,410x, the physical and super-resolution pixel sizes on the specimen level are 1.24 Å and 0.62 Å, corresponding to physical and super-resolution Nyquist frequencies of 1/(2.48 Å) and 1/(1.24 Å), respectively. After binning over 2 x 2 super-resolution pixels and performing motion correction (Li et al., 2013a), power spectra showed Thon rings to about 2.9 Å resolution (Figure 1A). After lattice unbending in 2dx (Gipson et al., 2007a), intensity quotient (IQ) plots showed diffraction spots with IQ values of 3 (corresponding to a peak-to-background ratio of 2.3; Henderson et al., 1986) up to a resolution of about 3.0 Å (Figure 1B). The best seven images were merged in 2dx (Gipson et al., 2007b), and the merging statistics indicate that the phase information is reliable to a resolution of 3.4 Å (90º is random) (Supplementary Table 1). The merged projection map at 3.4 Å resolution is shown in Figure 1C and D.
We also imaged AQP0 2D crystals with a K2 Summit camera mounted on an FEI Polara electron microscope operated at an acceleration voltage of 300 kV. The Polara employs an internal cartridge system, which is considerably more stable than the side-entry cryo-specimen holder used with the Tecnai F20. In addition, higher acceleration voltages (1) reduce the scattering cross-section, resulting in less multiple-scattering events, (2) give a lower Ewald sphere curvature, thus extending the breakdown limit of the central projection theorem (Zhang and Zhou, 2011), and (3) allow for a better performance of the DDD camera (Ruskin et al., 2013; Veesler et al., 2013). Dose-fractionated image stacks were recorded at liquid-nitrogen temperature at a calibrated magnification of 50,926x, giving physical and super-resolution pixel sizes of 0.98 Å and 0.49 Å on the specimen level, respectively. After 2 x 2 binning, the power spectra of motion-corrected images showed Thon rings to about 2.5 Å resolution (Figure 2A) and IQ-3 diffraction spots to about 2.3 Å resolution (Figure 2B). After lattice unbending, the best seven images were merged, and the phase residuals suggest a resolution of 2.6 Å resolution (Supplementary Table 2), close to about 3/4 of the physical Nyquist frequency of 1/(1.96 Å). The merged projection map at 2.6 Å resolution is shown in Figure 2C and D.
In comparison, electron diffraction patterns we recorded previously from the same crystals showed reflections to a resolution beyond 2.0 Å. It is well known that electron diffraction patterns of 2D crystals often show diffraction spots to a resolution higher than those seen in images of the same crystals. Therefore, the resolution of images collected from 2D crystals is not limited by the intrinsic disorder in the crystals but rather by other factors, such as specimen drift and beam-induced motions (Henderson and Glaeser, 1985; Glaeser et al., 2011). The results presented here demonstrate that the high DQE of the K2 camera together with computational motion correction make it possible to collect images of 2D crystals closer to the resolution limit imposed by the crystal order. However, even the use of a K2 camera and motion correction will not correct for the very fast initial motion that occurs when the beam first illuminates the specimen, which can be minimized by the use of thick support carbon (Glaeser et al., 2011) or simply by removing the first one or two frames of the movie. It is also encouraging that images recorded with a Tecnai F20 electron microscope operated at 200 kV could be merged to 3.4 Å resolution, suggesting that even a modest microscope can be used for near-atomic resolution cryo-EM studies. It is not entirely clear why images collected on the Polara produced better results than those taken on the Tecnai F20. However, the Polara has a more stable specimen stage, and its enclosed design may shield the sample from high-frequency vibrations resulting from acoustic noise. While the motion-correction algorithm we are currently using may suffice to correct drift-like motion caused by stage instability, the 0.15-second subframe integration time will not allow correction for most sample vibrations caused by acoustic noise.
Assessment of the signal-to-noise ratio over the exposure time
Biological specimens are sensitive to radiation damage. Exposure to the electron beam thus gradually deteriorates the structural information, especially in the high-resolution range (Breedlove and Trammell, 1970; Glaeser, 1971). As a result, the intensity of diffraction spots and their SNR decrease with increasing exposure time and electron dose (Taylor and Glaeser, 1976). As the electron-counting K2 Summit camera allows for dose fractionation, it is possible to follow changes in the power spectrum with increasing electron dose. By using the binned super-resolution dataset of the AQP0 2D crystals recorded on the Tecnai F20, we analyzed the average IQ values within resolution bins as a function of the number of averaged movie frames. The IQ value was introduced in electron crystallography to assess the SNR of reflections (Henderson et al., 1986), and changes in the average IQ value thus represent changes in the average SNR of reflections in the resolution bins.
Figure 3 shows the average IQ values plotted against the number of averaged movie frames in different resolution bins (each frame corresponds to an electron dose of 0.78 counts/Å2). In the lower resolution bins (below a resolution of 4.7 Å), the average IQ values initially decrease rapidly but then stabilize, indicating that averaging the first few frames improves the SNR whereas adding further frames does not further improve the SNR in these low-resolution bins. In the higher resolution bins (resolution between 4.7 and 3.0 Å), the average IQ values also initially decrease rapidly but keep decreasing slowly to about frame 9, from which point on they begin to slowly increase. This behavior suggests that while the SNR improves initially with the addition of frames, from frame 9 onwards beam damage deteriorates the high-resolution information and leads to a decrease in the SNR. While the effect is small and may not be of practical relevance, the same trend can be seen in all three resolution bins. The average IQ values in the highest resolution bin (resolution between 3.0 and 2.6 Å) are always close to 7, which means that the peaks equal the background level, so that the reflections in this resolution range do not contain meaningful information (Henderson et al., 1986).
Information beyond the physical Nyquist frequency in super-resolution mode
To address the question whether centroiding alters the phase information, we compared images of AQP0 2D crystals recorded at different magnifications, which thus have different pixel sizes and Nyquist frequencies. The AQP0 2D crystals used in this study diffract to high resolution and generate a sufficient number of diffraction spots for this analysis. In addition to data collected on the Tecnai F20 at a calibrated magnification of 40,410x (an IQ plot of a typical motion-corrected image after lattice unbending is shown in Figure 4A), we collected another set of dose-fractionated image stacks of AQP0 2D crystals at the lower calibrated magnification of 15,858x, giving a physical pixel size of 3.15 Å on the specimen level, corresponding to a physical Nyquist frequency of 1/(6.3 Å). All information beyond 6.3 Å will therefore be in the super-resolution range (an IQ plot of a typical motion-corrected image after lattice unbending is shown in Figure 4B). For optimal performance of the K2 Summit camera and the motion-correction algorithm, the dose rate was kept low, using 10 counts/pixel/s (1.036 counts/Å2/s; 2.5 counts/pixel/s in super resolution), which is only slightly higher than the dose rate used at higher magnification (8 counts/pixel/s; 5.203 counts/Å2/s). In order to ensure that the frames had sufficient SNR for subsequent motion correction, the frame read-out rate was adjusted to 400 ms per frame.
The dose-fractionated image stacks were recorded with a total electron dose similar to the total dose used for the high-magnification dataset. The dose-fractionated image stacks were not binned to retain the super-resolution information. After motion correction, the imaged crystals were computationally unbent, and the best seven images were merged. To determine the resolution to which the phase information is reliable, phase residuals were calculated in resolution bins for all spots with IQ values 1 to 8 (“all IQs” in the Tables) as well as phase residuals that were weighted by IQ value as implemented in the 2dx software (Gipson et al., 2007b) (“IQ-wght” in the Tables). The images collected at the lower magnification could be merged to 3.4 Å resolution (Table 1, middle; Supplementary Table 3), similar to those collected at the higher magnification after binning over 2 x 2 pixels (also 3.4 Å resolution; Table 1, left; Supplementary Table 1), demonstrating that the phase information within each group is consistent to these resolutions. When the unbinned images from the low-magnification dataset were merged with the 2 x 2 binned images from the high-magnification dataset, the phases remained consistent to 3.4 Å resolution (Table 1, right; Supplementary Table 4). This result shows that the data collected between 6.3 Å to 3.4 Å, which lie in the super-resolution range for the images collected at the lower magnification, is consistent with the information in the same resolution range that is within physical Nyquist when collected at the higher magnification. Importantly, this indicates that super-resolution yields high-quality phase data that is directly useful for structure determination.
Table 1.
40k, 2x2 binned | 16k, unbinned | combined | ||||||
---|---|---|---|---|---|---|---|---|
# | DMIN | DMAX | all IQs | IQ-wght | all IQs | IQ-wght | all IQs | IQ-wght |
1 | 1000.0 11.6 | 23.1 | 17.9 | 22.3 | 19.4 | 29.3 | 22.4 | |
1 | 323 | 323 | 329 | 329 | 652 | 652 | ||
2 | 11.6 | 8.2 | 26.1 | 21.9 | 27.3 | 24.1 | 29.0 | 24.0 |
2 | 329 | 329 | 328 | 328 | 657 | 657 | ||
3 | 8.2 | 6.7 | 27.9 | 25.0 | 21.7 | 18.8 | 25.6 | 23.1 |
3 | 346 | 346 | 347 | 347 | 693 | 693 | ||
4 | 6.7 | 5.8 | 40.9 | 36.4 | 39.5 | 34.0 | 42.2 | 37.3 |
4 | 346 | 346 | 342 | 342 | 688 | 688 | ||
|
||||||||
5 | 5.8 | 5.2 | 27.9 | 24.1 | 27.0 | 21.3 | 27.6 | 23.6 |
5 | 317 | 317 | 319 | 319 | 636 | 636 | ||
6 | 5.2 | 4.7 | 41.9 | 31.6 | 40.2 | 30.7 | 42.9 | 33.1 |
6 | 312 | 312 | 274 | 274 | 586 | 586 | ||
7 | 4.7 | 4.4 | 36.0 | 32.6 | 39.0 | 36.6 | 39.7 | 35.8 |
7 | 335 | 335 | 322 | 322 | 657 | 657 | ||
8 | 4.4 | 4.1 | 39.8 | 31.6 | 52.0 | 44.4 | 44.0 | 35.8 |
8 | 309 | 309 | 264 | 264 | 573 | 573 | ||
9 | 4.1 | 3.9 | 55.4 | 47.2 | 63.0 | 57.7 | 61.7 | 52.4 |
9 | 265 | 265 | 224 | 224 | 501 | 501 | ||
10 | 3.9 | 3.7 | 56.3 | 47.1 | 77.6 | 72.7 | 62.1 | 53.0 |
10 | 317 | 317 | 238 | 238 | 584 | 584 | ||
11 | 3.7 | 3.5 | 64.2 | 57.0 | 70.9 | 68.4 | 68.6 | 59.2 |
11 | 281 | 281 | 178 | 178 | 487 | 487 | ||
12 | 3.5 | 3.4 | 78.3 | 76.3 | 80.5 | 70.7 | 79.9 | 75.1 |
12 | 178 | 178 | 143 | 143 | 407 | 407 | ||
|
|
|
||||||
13 | 3.4 | 3.2 | 76.8 | 79.4 | 85.6 | 87.5 | 83.3 | 82.7 |
13 | 166 | 166 | 135 | 135 | 429 | 429 | ||
14 | 3.2 | 3.1 | 78.0 | 73.3 | 95.8 | 90.7 | 89.4 | 87.4 |
14 | 85 | 85 | 85 | 85 | 308 | 308 | ||
15 | 3.1 | 3.0 | 80.2 | 73.5 | 73.6 | 70.9 | 83.4 | 80.1 |
15 | 67 | 67 | 39 | 39 | 294 | 294 | ||
| ||||||||
Overall: Phase residual | 44.112 | 45.546 | 49.046 | |||||
Number of spots | 3976 | 3567 | 8152 |
The results presented here suggest that it is possible to fully utilize the super-resolution pixels for structure determination. Recording images at lower magnifications may not be desirable for single-particle cryo-EM of small molecules, due to the lower DQE at frequencies beyond the Nyquist limit. It will become useful, however, for recording single-particle cryo-EM images of very large particles, such as large icosahedral viruses, for imaging 2D crystals larger than 1 μm in size, as well as for collecting electron tomographic data.
Influence of dose rate on image quality
We have previously reported that higher dose rates result in increased coincidence loss that affects the amplitude in the low-frequency region of power spectra (Li et al., 2013a). To assess the effect of different dose rates on image quality, we collected dose-fractionated image stacks of AQP0 2D crystals in super-resolution mode at different dose rates, but keeping the total electron dose the same.
The images were collected at a magnification of 15,858x, corresponding to a physical Nyquist frequency of 1/(6.3 Å), using dose rates of 4, 8, and 20 counts/pixel/s. The dose rate was measured at the camera level after the specimen was inserted, and gain and dark references were carefully prepared after the dose rate was set. To end up with the same total electron dose of 20 counts/Å2 per image stack and to ensure that the individual frames have a sufficient SNR for subsequent motion correction, the frame read-out times were adjusted, resulting in different exposure times. For the dose rate of 4 counts/pixel/s, 120 frames were recorded at 400 ms per frame, giving an exposure time of 48 s; for the dose rate of 8 counts/pixel/s, 120 frames were recorded at 200 ms per frame, an exposure time of 24 s; and for the dose rate of 20 counts/pixel/s, 128 frames were recorded at 75 ms per frame, yielding an exposure time of 9.6 s.
Power spectra and IQ plots of motion-corrected images recorded with all three dose rates showed diffraction spots to a resolution of about 3.5 Å (Supplementary Figure 1). We first calculated one-dimensional rotational averages of the power spectra of representative images recorded at the three different dose rates at approximately the same defocus (Figure 5). While intensities for diffraction spots at a resolution below 0.2 of the physical Nyquist frequency are clearly observed for the images recorded with 4 and 8 counts/pixel/s, these intensities are missing in the image recorded with 20 counts/pixel/s (arrows in Figure 5). Thus, at a dose rate of 20 counts/pixel/s, coincidence loss results in sufficient dampening of the amplitudes in the lower resolution range that low-order reflections are no longer visible above the background (Li et al., 2013a; Ruskin et al., 2013). For each dataset, the best seven images were then merged, and the phase residuals are shown in Table 2 and Supplementary Tables 5 to 7. All datasets could be merged individually to about 3.5 Å resolution, and the behavior of the phase residuals within each dataset as a function of resolution is similar for the three datasets (Figure 6A). We note that the phase residuals in the different resolution bins of data recorded with 8 counts/pixel/s are consistently better than those of data recorded with 4 and 20 counts/pixel/s. In particular, although data recorded at a dose rate of 4 counts/pixel/s should yield better data, the phase residuals are somewhat higher, which may be caused by the longer frame read-out time that was used (400 ms instead of 150 ms). This will limit the effectiveness of motion correction, especially for data collected with an electron microscope equipped with a side-entry specimen stage such as the Tecnai F20. Still, our results indicate that for dose rates ranging from 4 to 20 counts/pixel/s, the phase information in images for each group is consistent to a resolution of about 3.5 Å. Furthermore, an analysis of the number of spots with given IQ values shows that their distribution is similar for all three datasets in all resolution ranges (Supplementary Figure 2). This result is in agreement with the previous finding that different dose rates do not affect the SNR of images recorded with the same total electron dose.
Table 2.
4 counts/pixel/s | 8 counts/pixel/s | 20 counts/pixel/s | ||||||
---|---|---|---|---|---|---|---|---|
# | DMIN | DMAX | all IQs | IQ-wght | all IQs | IQ-wght | all IQs | IQ-wght |
1 | 1000.0 13.4 | 13.5 | 11.6 | 21.9 | 18.4 | 19.4 | 16.6 | |
1 | 234 | 234 | 237 | 237 | 236 | 236 | ||
2 | 13.4 | 9.5 | 25.2 | 22.8 | 21.1 | 20.0 | 26.9 | 22.4 |
2 | 259 | 259 | 265 | 265 | 259 | 259 | ||
3 | 9.5 | 7.7 | 21.7 | 20.7 | 25.9 | 23.5 | 26.0 | 24.4 |
3 | 263 | 263 | 262 | 262 | 261 | 261 | ||
4 | 7.7 | 6.7 | 22.5 | 17.9 | 22.3 | 19.4 | 23.4 | 18.9 |
4 | 245 | 245 | 250 | 250 | 243 | 243 | ||
5 | 6.7 | 6.0 | 31.6 | 24.5 | 30.2 | 26.8 | 35.9 | 30.8 |
5 | 253 | 253 | 258 | 258 | 255 | 255 | ||
6 | 6.0 | 5.5 | 25.9 | 20.8 | 24.7 | 20.7 | 29.8 | 21.3 |
6 | 209 | 209 | 212 | 212 | 201 | 201 | ||
7 | 5.5 | 5.1 | 29.6 | 19.6 | 27.2 | 20.4 | 34.6 | 25.9 |
7 | 265 | 265 | 261 | 261 | 253 | 253 | ||
8 | 5.1 | 4.7 | 38.2 | 28.4 | 37.1 | 29.3 | 40.7 | 31.4 |
8 | 224 | 224 | 242 | 242 | 213 | 213 | ||
9 | 4.7 | 4.5 | 37.8 | 32.8 | 37.3 | 30.6 | 44.2 | 38.6 |
9 | 237 | 237 | 259 | 259 | 233 | 233 | ||
10 | 4.5 | 4.2 | 46.1 | 32.9 | 41.1 | 33.0 | 55.8 | 40.1 |
10 | 206 | 206 | 228 | 228 | 208 | 208 | ||
11 | 4.2 | 4.0 | 53.3 | 43.9 | 56.3 | 43.1 | 65.2 | 47.1 |
11 | 197 | 197 | 219 | 219 | 185 | 185 | ||
12 | 4.0 | 3.9 | 64.8 | 55.8 | 57.4 | 46.2 | 80.5 | 75.6 |
12 | 129 | 129 | 148 | 148 | 134 | 134 | ||
|
||||||||
13 | 3.9 | 3.7 | 68.6 | 60.7 | 63.8 | 50.0 | 81.5 | 74.6 |
13 | 196 | 196 | 215 | 215 | 178 | 178 | ||
14 | 3.7 | 3.6 | 78.8 | 69.4 | 74.6 | 68.6 | 86.2 | 80.9 |
14 | 141 | 141 | 155 | 155 | 138 | 138 | ||
|
||||||||
15 | 3.6 | 3.5 | 94.7 | 88.6 | 80.1 | 71.5 | 89.8 | 80.9 |
15 | 123 | 123 | 120 | 120 | 120 | 120 | ||
|
||||||||
16 | 3.5 | 3.4 | 89.7 | 87.2 | 83.7 | 77.0 | 89.5 | 79.9 |
16 | 137 | 137 | 149 | 149 | 146 | 146 | ||
17 | 3.4 | 3.3 | 99.2 | 100.1 | 92.2 | 85.1 | 102.2 | 95.0 |
17 | 133 | 133 | 142 | 142 | 129 | 129 | ||
18 | 3.3 | 3.2 | 88.4 | 82.6 | 86.3 | 85.0 | 93.3 | 88.0 |
18 | 111 | 111 | 136 | 136 | 117 | 117 | ||
| ||||||||
Overall: Phase residual | 44.787 | 43.719 | 50.079 | |||||
Number of spots | 3562 | 3758 | 3509 |
To compare the phase quality between data recorded at different dose rates, different datasets were merged with each other, i.e., data recorded at 4 counts/pixel/s with data recorded at 8 counts/pixel/s, data recorded at 4 counts/pixel/s with data recorded at 20 counts/pixel/s, data recorded at 8 counts/pixel/s with data recorded at 20 counts/pixel/s, as well as all the data. The resulting phase residual statistics show that the data recorded at different dose rates are all consistent to about 3.5 Å resolution (Table 3), and plots of phase residual against spatial frequency show similar trends for all merged datasets (Figure 6B). Thus, the phase information is consistent between datasets irrespective of what dose rate was used. Interestingly, there is a small increase in phase residual at the resolution corresponding to the physical Nyquist frequency. While this increase is small, ~15º, it was observed for all datasets.
Table 3.
4+8 counts/pixel/s | 4+20 counts/pixel/s | 8+20 counts/pixel/s | 4+8+20 counts/pixel/s | |||||||
---|---|---|---|---|---|---|---|---|---|---|
# | DMIN | DMAX | all IQs | IQ-wght | all IQs | IQ-wght | all IQs | IQ-wght | all IQs | IQ-wght |
1 | 1000.0 | 13.4 | 30.7 | 26.9 | 16.7 | 14.3 | 31.4 | 26.4 | 31.4 | 27.8 |
1 | 471 | 471 | 470 | 470 | 473 | 473 | 707 | 707 | ||
2 | 13.4 | 9.5 | 29.4 | 27.4 | 28.9 | 24.8 | 30.7 | 27.0 | 32.0 | 28.6 |
2 | 524 | 524 | 518 | 518 | 524 | 524 | 783 | 783 | ||
3 | 9.5 | 7.7 | 31.3 | 29.4 | 24.7 | 23.4 | 28.9 | 26.8 | 29.4 | 27.6 |
3 | 525 | 525 | 524 | 524 | 523 | 523 | 786 | 786 | ||
4 | 7.7 | 6.7 | 25.7 | 22.4 | 23.4 | 18.6 | 24.1 | 20.6 | 24.8 | 21.1 |
4 | 495 | 495 | 488 | 488 | 493 | 493 | 738 | 738 | ||
5 | 6.7 | 6.0 | 41.2 | 38.3 | 34.3 | 28.3 | 38.2 | 35.0 | 39.4 | 36.0 |
5 | 511 | 511 | 508 | 508 | 513 | 513 | 766 | 766 | ||
6 | 6.0 | 5.5 | 29.7 | 25.5 | 28.9 | 21.9 | 27.6 | 21.4 | 29.1 | 23.7 |
6 | 421 | 421 | 410 | 410 | 413 | 413 | 622 | 622 | ||
7 | 5.5 | 5.1 | 35.1 | 28.4 | 32.4 | 23.6 | 33.2 | 25.9 | 34.3 | 27.1 |
7 | 526 | 526 | 518 | 518 | 514 | 514 | 779 | 779 | ||
8 | 5.1 | 4.7 | 40.9 | 32.8 | 39.9 | 29.9 | 39.8 | 31.5 | 40.8 | 32.4 |
8 | 466 | 466 | 437 | 437 | 455 | 455 | 679 | 679 | ||
9 | 4.7 | 4.5 | 47.5 | 43.3 | 41.8 | 37.1 | 44.6 | 37.2 | 46.9 | 41.6 |
9 | 496 | 496 | 470 | 470 | 492 | 492 | 729 | 729 | ||
10 | 4.5 | 4.2 | 48.3 | 37.1 | 52.4 | 38.5 | 48.2 | 35.0 | 51.2 | 38.1 |
10 | 434 | 434 | 414 | 414 | 436 | 436 | 642 | 642 | ||
11 | 4.2 | 4.0 | 61.5 | 48.1 | 58.2 | 46.2 | 59.4 | 42.5 | 60.5 | 46.5 |
11 | 416 | 416 | 382 | 382 | 404 | 404 | 601 | 601 | ||
12 | 4.0 | 3.9 | 65.0 | 56.1 | 72.0 | 65.2 | 67.9 | 54.9 | 68.1 | 59.3 |
12 | 277 | 277 | 263 | 263 | 282 | 282 | 367 | 367 | ||
13 | 3.9 | 3.7 | 67.7 | 56.8 | 71.7 | 62.4 | 67.1 | 52.8 | 71.2 | 60.8 |
13 | 411 | 411 | 374 | 374 | 393 | 393 | 513 | 513 | ||
|
||||||||||
14 | 3.7 | 3.6 | 78.3 | 73.7 | 79.8 | 79.9 | 78.5 | 76.3 | 80.6 | 83.1 |
14 | 296 | 296 | 279 | 279 | 293 | 293 | 365 | 365 | ||
|
||||||||||
15 | 3.6 | 3.5 | 87.0 | 79.5 | 95.7 | 87.3 | 82.7 | 67.2 | 87.6 | 81.3 |
15 | 243 | 243 | 243 | 243 | 240 | 240 | 309 | 309 | ||
|
||||||||||
16 | 3.5 | 3.4 | 83.5 | 76.1 | 90.4 | 92.7 | 87.6 | 88.8 | 87.6 | 89.8 |
16 | 286 | 286 | 283 | 283 | 295 | 295 | 358 | 358 | ||
|
||||||||||
17 | 3.4 | 3.3 | 91.1 | 91.5 | 93.2 | 88.0 | 90.1 | 89.0 | 94.8 | 98.4 |
17 | 275 | 275 | 262 | 262 | 271 | 271 | 338 | 338 | ||
18 | 3.3 | 3.2 | 92.6 | 95.7 | 90.5 | 84.9 | 88.8 | 90.0 | 90.5 | 89.4 |
18 | 247 | 247 | 228 | 228 | 253 | 253 | 303 | 303 | ||
| ||||||||||
Overall: Phase residual | 49.469 | 47.539 | 48.459 | 48.493 | ||||||
Number of spots | 7320 | 7071 | 7267 | 10385 |
The datasets recorded at different dose rates were also used to assess the effect of dose rate on the amplitude information (Figure 6C). Comparison of the amplitude errors in resolution bins shows that the data recorded with 4 and 8 counts/pixel/sec are comparable. For data recorded with 20 counts/pixel/sec, the amplitude error is higher in the lowest resolution bin (below 62.0 Å or 0.2 of the physical Nyquist frequency), which can be explained by the higher coincidence loss at higher dose rates. However, the amplitude error is also significantly higher in the high-resolution bins (above 4.13 Å or 1.5 times the physical Nyquist frequency). A dose rate of 20 counts/pixel/sec not only increases coincidence losses (which influences mostly low resolution region), but also reduces modulation transfer function (MTF) and DQE at both high and low frequencies which can be explained by the higher coincidence loss at higher dose rates. The large amplitude errors in the high-resolution range may be attributed to the lower DQE of the K2 Summit camera in this range at higher dose rates (Li et al., 2013a; Ruskin et al., 2013).
When images recorded with different dose rates were merged (Figure 6D), the amplitude errors of all datasets including the images recorded with 20 counts/pixel/sec increased substantially in the low-resolution range (below 62.0 Å) but not as much in the high-resolution range (above 4.13 Å). In addition, all combined datasets show a peak of high amplitude error at a resolution of about 15 Å (Figure 6D, arrow), which appears less prominent in individual datasets (Figure 6C, arrow). Compared to the phase information, the amplitude information appears to be much more affected by coincidence loss that is caused by higher dose rates.
It is widely accepted in the X-ray crystallography field that phase information is far more important for the reconstruction of a 3D electron density map than the amplitude information. Moreover, image amplitudes are greatly modulated by the contrast transfer function (CTF), which complicates the accurate measurement of amplitudes. Our results indicate that the coincidence loss created by higher dose rates leads to a decrease in the measured amplitudes but does not systematically alter phases, suggesting the image reconstruction is still feasible under dose rate conditions that create greater coincidence loss.
Above all, when it is desirable to have a larger field of view, say with large 2D crystals, it is feasible to set up data collection at lower magnifications, as long as dose rates are kept in the same < 10 e−/sec regime. It may also be useful to increase the readout rate to improve motion correction, but at the expense of significantly expanded data sizes. Hence, the magnification and the subframe integration time should be carefully chosen for image data collection.
Influence of dose fractionation protocol on image deterioration
The recording of dose-fractionated image stacks to enable correction of beam-induced specimen motion has become the standard procedure for data collection (Bai et al., 2013; Campbell et al., 2012; Li et al., 2013a). A study of beam-induced motion revealed a pattern, in which motion is larger in the beginning of an exposure and then slows down (Brilot et al., 2012). The reason for this pattern is not known but may point to an annealing mechanism that relaxes strain present in the amorphous ice layer and carbon support. When the beam is turned off and back on after a pause of 60 s, a motion pattern similar to the initial exposure is observed, including initial motions of similar magnitude (Brilot et al., 2012). Only after several repeated exposures a significant reduction of motion is observed. It is therefore possible that different dose fractionation time courses may lead to different motion patterns. Using the K2 Summit camera, we recorded image stacks of AQP0 2D crystals in super-resolution mode with a total dose of 45 counts/Å2 and a dose rate of 20 counts/pixel/s. Three protocols were tested and compared: image stacks recorded without interruption (continuous), image stacks recorded in six exposures separated by ~30 s pauses (6 parts) and image stacks recorded in 18 exposures separated by ~15 s pauses (18 parts). For all exposure series, we kept the sample drift rate below 2 Å/s and chose a beam diameter that significantly exceeded the field of view. This ensured that unexposed parts of the crystal moving into the beam between exposures did not affect our measurements.
Figure 7 shows plots of amplitudes measured at different total applied electron doses. After dividing the data into different resolution ranges it can be seen that the dose sensitivity is greatest for the highest resolution and least for the lowest resolution, as observed previously (Glaeser, 1971). The plots suggest that the image amplitudes, which are indicative of the structural integrity of the crystals, decay at the same rate (within measurement error) for all three data collection protocols tested. The underlying cause of beam-induced motion remains uncertain but may include specimen charging and the build-up of internal pressure due to molecular radicals generated by radiolysis (Glaeser, 2008). Our observation that different fractionation protocols can lead to the same loss of structural integrity but different motion trajectories suggest that either radiation damage is not the sole cause of the sample motion, or the damage caused by the beam continues for a certain time after the beam has been switched off.
In conclusion, our current studies demonstrate that electron counting and centroiding used by the K2 Summit camera does not affect the phase information in the super-resolution range beyond the physical Nyquist frequency. Thus, it is possible to record images at a lower magnification to fully utilize the total number of super-resolution pixels. Furthermore, at least up to a dose rate of 20 counts/pixel/sec, the influence of coincidence loss on phase information, if there is any, may be negligible.
Supplementary Material
Acknowledgments
This work was supported in part by HHMI (to D.A.A), NSF grant DBI-0960271 (to D.A.A and Y.C.), which in part funded the development of the K2 camera in association with Gatan and Peter Denes at Lawrence Berkeley Labs, NIH grants R01 GM082893 (to Y.C.), R01 GM098672 (to Y.C.), P01 GM62580 (to N.G. and T.W.), R01 GM031627 (to D.A.A.) and P50 GM082250 (to A. Frankel). D.A.A., N.G, and T.W. are investigators with the Howard Hughes Medical Institute.
Footnotes
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
References
- Allegretti M, Mills DJ, McMullan G, Kuhlbrandt W, Vonck J. Atomic model of the F420-reducing [NiFe] hydrogenase by electron cryo-microscopy using a direct electron detector. Elife. 2014;3:e01963. doi: 10.7554/eLife.01963. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Amunts A, Brown A, Bai XC, Llacer JL, Hussain T, Emsley P, Long F, Murshudov G, Scheres SH, Ramakrishnan V. Structure of the yeast mitochondrial large ribosomal subunit. Science. 2014;343:1485–1489. doi: 10.1126/science.1249410. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bai XC, Fernandez IS, McMullan G, Scheres SH. Ribosome structures to near-atomic resolution from thirty thousand cryo-EM particles. Elife. 2013;2:e00461. doi: 10.7554/eLife.00461. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bammes BE, Rochat RH, Jakana J, Chen DH, Chiu W. Direct electron detection yields cryo-EM reconstructions at resolutions beyond 3/4 Nyquist frequency. J Struct Biol. 2012;177:589–601. doi: 10.1016/j.jsb.2012.01.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bartesaghi A, Merk A, Banerjee S, Matthies D, Wu X, Milne JL, Subramaniam S. 2.2 Å resolution cryo-EM structure of β-galactosidase in complex with a cell-permeant inhibitor. Science. 2015;348:1147–1151. doi: 10.1126/science.aab1576. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Breedlove JR, Jr, Trammell GT. Molecular microscopy: fundamental limitations. Science. 1970;170:1310–1313. doi: 10.1126/science.170.3964.1310. [DOI] [PubMed] [Google Scholar]
- Brilot AF, Chen JZ, Cheng A, Pan J, Harrison SC, Potter CS, Carragher B, Henderson R, Grigorieff N. Beam-induced motion of vitrified specimen on holey carbon film. J Struct Biol. 2012;177:630–637. doi: 10.1016/j.jsb.2012.02.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Campbell MG, Cheng A, Brilot AF, Moeller A, Lyumkis D, Veesler D, Pan J, Harrison SC, Potter CS, Carragher B, Grigorieff N. Movies of ice-embedded particles enhance resolution in electron cryo-microscopy. Structure. 2012;20:1823–1828. doi: 10.1016/j.str.2012.08.026. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Cao S, Zhang W. Characterization of an early-stage fusion intermediate of Sindbis virus using cryoelectron microscopy. Proc Natl Acad Sci U S A. 2013;110:13362–13367. doi: 10.1073/pnas.1301911110. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Crowther RA, Henderson R, Smith JM. MRC image processing programs. J Struct Biol. 1996;116:9–16. doi: 10.1006/jsbi.1996.0003. [DOI] [PubMed] [Google Scholar]
- Faruqi AR, McMullan G. Electronic detectors for electron microscopy. Q Rev Biophys. 2011;44:357–390. doi: 10.1017/S0033583511000035. [DOI] [PubMed] [Google Scholar]
- Gipson B, Zeng X, Zhang ZY, Stahlberg H. 2dx – user-friendly image processing for 2D crystals. J Struct Biol. 2007a;157:64–72. doi: 10.1016/j.jsb.2006.07.020. [DOI] [PubMed] [Google Scholar]
- Gipson B, Zeng X, Stahlberg H. 2dx_merge: data management and merging for 2D crystal images. J Struct Biol. 2007b;160:375–384. doi: 10.1016/j.jsb.2007.09.011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Glaeser RM. Limitations to significant information in biological electron microscopy as a result of radiation damage. J Ultrastruct Res. 1971;36:466–482. doi: 10.1016/s0022-5320(71)80118-1. [DOI] [PubMed] [Google Scholar]
- Glaeser RM. Retrospective: radiation damage and its associated "information limitations". J Struct Biol. 2008;163:271–276. doi: 10.1016/j.jsb.2008.06.001. [DOI] [PubMed] [Google Scholar]
- Glaeser RM, McMullan G, Faruqi AR, Henderson R. Images of paraffin monolayer crystals with perfect contrast: Minimization of beam-induced specimen motion. Ultramicroscopy. 2011;111:90–100. doi: 10.1016/j.ultramic.2010.10.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gonen T, Sliz P, Kistler J, Cheng Y, Walz T. Aquaporin-0 membrane junctions reveal the structure of a closed water pore. Nature. 2004;429:193–197. doi: 10.1038/nature02503. [DOI] [PubMed] [Google Scholar]
- Gonen T, Cheng Y, Sliz P, Hiroaki Y, Fujiyoshi Y, Harrison SC, Walz T. Lipid-protein interactions in double-layered two-dimensional AQP0 crystals. Nature. 2005;438:633–638. doi: 10.1038/nature04321. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gyobu N, Tani K, Hiroaki Y, Kamegawa A, Mitsuoka K, Fujiyoshi Y. Improved specimen preparation for cryo-electron microscopy using a symmetric carbon sandwich technique. J Struct Biol. 2004;146:325–333. doi: 10.1016/j.jsb.2004.01.012. [DOI] [PubMed] [Google Scholar]
- Henderson R, Glaeser RM. Quantitative analysis of image contrast in electron micrographs of beam-sensitive crystals. Ultramicroscopy. 1985;16:139–150. [Google Scholar]
- Henderson R, Baldwin JM, Downing KH, Lepault J, Zemlin F. Structure of purple membrane from Halobacterium halobium: Recording, measurement and evaluation of electron micrographs at 3.5 Å resolution. Ultramicroscopy. 1986;19:147–178. [Google Scholar]
- Hite RK, Li Z, Walz T. Principles of membrane protein interactions with annular lipids deduced from aquaporin-0 2D crystals. EMBO J. 2010a;29:1652–1658. doi: 10.1038/emboj.2010.68. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hite RK, Schenk AD, Li Z, Cheng Y, Walz T. Collecting electron crystallographic data of two-dimensional protein crystals. Methods Enzymol. 2010b;481:251–282. doi: 10.1016/S0076-6879(10)81011-0. [DOI] [PubMed] [Google Scholar]
- Li X, Mooney P, Zheng S, Booth CR, Braunfeld MB, Gubbens S, Agard DA, Cheng Y. Electron counting and beam-induced motion correction enable near-atomic-resolution single-particle cryo-EM. Nat Methods. 2013a;10:584–590. doi: 10.1038/nmeth.2472. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Li X, Zheng SQ, Egami K, Agard DA, Cheng Y. Influence of electron dose rate on electron counting images recorded with the K2 camera. J Struct Biol. 2013b;184:251–260. doi: 10.1016/j.jsb.2013.08.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Liao M, Cao E, Julius D, Cheng Y. Structure of the TRPV1 ion channel determined by electron cryo-microscopy. Nature. 2013;504:107–112. doi: 10.1038/nature12822. [DOI] [PMC free article] [PubMed] [Google Scholar]
- McMullan G, Clark AT, Turchetta R, Faruqi AR. Enhanced imaging in low dose electron microscopy using electron counting. Ultramicroscopy. 2009a;109:1411–1416. doi: 10.1016/j.ultramic.2009.07.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- McMullan G, Chen S, Henderson R, Faruqi AR. Detective quantum efficiency of electron area detectors in electron microscopy. Ultramicroscopy. 2009b;109:1126–1143. doi: 10.1016/j.ultramic.2009.04.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- McMullan G, Faruqi AR, Clare D, Henderson R. Comparison of optimal performance at 300keV of three direct electron detectors for use in low dose electron microscopy. Ultramicroscopy. 2014;147:156–163. doi: 10.1016/j.ultramic.2014.08.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Milazzo AC, Cheng A, Moeller A, Lyumkis D, Jacovetty E, Polukas J, Ellisman MH, Xuong NH, Carragher B, Potter CS. Initial evaluation of a direct detection device detector for single particle cryo-electron microscopy. J Struct Biol. 2011;176:404–408. doi: 10.1016/j.jsb.2011.09.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mindell JA, Grigorieff N. Accurate determination of local defocus and specimen tilt in electron microscopy. J Struct Biol. 2003;142:334–347. doi: 10.1016/s1047-8477(03)00069-8. [DOI] [PubMed] [Google Scholar]
- Ruskin RS, Yu Z, Grigorieff N. Quantitative characterization of electron detectors for transmission electron microscopy. J Struct Biol. 2013;184:385–393. doi: 10.1016/j.jsb.2013.10.016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Shroff H, Galbraith CG, Galbraith JA, White H, Gillette J, Olenych S, Davidson MW, Betzig E. Dual-color superresolution imaging of genetically expressed probes within individual adhesion complexes. Proc Natl Acad Sci U S A. 2007;104:20308–20313. doi: 10.1073/pnas.0710517105. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Taylor KA, Glaeser RM. Electron microscopy of frozen hydrated biological specimens. J Ultrastruct Res. 1976;55:448–456. doi: 10.1016/s0022-5320(76)80099-8. [DOI] [PubMed] [Google Scholar]
- Veesler D, Campbell MG, Cheng A, Fu CY, Murez Z, Johnson JE, Potter CS, Carragher B. Maximizing the potential of electron cryomicroscopy data collected using direct detectors. J Struct Biol. 2013;184:193–202. doi: 10.1016/j.jsb.2013.09.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zhang X, Zhou ZH. Limiting factors in atomic resolution cryo electron microscopy: no simple tricks. J Struct Biol. 2011;175:253–263. doi: 10.1016/j.jsb.2011.05.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.