Abstract
Digital cardiovascular angiography accounts for a major portion of the radiation dose among the examinations performed at cardiovascular centres. However, dose-related information is neither monitored nor recorded systemically. This report concerns the construction of a radiation dose monitoring system based on digital imaging and communications in medicine (DICOM) data and its use at the cardiovascular centre of the University Hospitals in Korea. The dose information was analysed according to DICOM standards for a series of procedures, and the formulation of diagnostic reference levels (DRLs) at our cardiovascular centre represents the first of its kind in Korea. We determined a dose area product (DAP) DRL for coronary angiography of 75.6 Gy cm2 and a fluoroscopic time DRL of 318.0 s. The DAP DRL for percutaneous transluminal coronary intervention was 213.3 Gy cm2, and the DRL for fluoroscopic time was 1207.5 s.
Keywords: DICOM MPPS, Radiation dose monitoring, Cardiovascular centre, Diagnostic reference levels, Order communication system
Introduction
Digital radiographic imaging technology can reduce the potential risk of radiation exposure in patients and increase the utility of images. However, concerns have been raised regarding unnecessary radiation exposure that may result from operator inexperience or misinformation. The International Commission on Radiological Protection (ICRP) indicated in Publication 93 that the ease of obtaining and deleting images during digital angiography was accompanied by a tendency among clinical users to take greater numbers of images. Consequently, concern about the likelihood of radiation exposure exceeding the necessary dose has also been raised [1]. In light of this, the Ministry of Food and Drug Safety (MFDS) has implemented a project as a 2011 outsourced task in Korea to establish a recommended diagnostic reference level (DRL) for exposure doses in patients undergoing interventional radiological procedures.
Diagnostic reference levels were first recommended by the ICRP in 1990 [2] and subsequently revised in 1996 [3]. Reference levels are typically set at the 75th percentile of the dose distribution based on a survey conducted across a broad user base (i.e. large and small facilities, public and private, hospital and out-patient) using a specified dose measurement protocol and phantom [4].
Digital cardiovascular angiography and interventional radiological procedures have become essential diagnostic and treatment tools in cases of patients with critical cardiovascular diseases, and the frequency of those procedures continues to increase. The interventional radiological procedures are accompanied by the second largest radiation dose after computed tomography (CT) in diagnostic radiology. In the USA, the Food and Drug Administration (FDA) has issued a regulation requiring manufacturers of fluoroscopic equipment produced since 2006 to attach a patient radiation dose-monitoring device [5]. This radiation-monitoring device records information during patient examinations, including the dose area product (DAP), the total radio-fluoroscopic time and the cumulative radiation dose, which is available within the digital imaging and communications in medicine (DICOM) images. In particular, the DAP value can provide an important index for predicting radiation injury and stochastic radiation effects in the patient [6].
Korea has previously been without a separate system for dose monitoring in interventional radiological procedures, and data regarding radiation doses in patients undergoing cardiovascular interventional procedures have not been reported. The purpose of the present study was to develop DRLs for cardiovascular interventional procedures in Korea and to introduce the first real-time radiation dose monitoring system at a Korean cardiovascular centre.
Medical Radiation Injury
The ICRP has previously issued warnings about the hazards of interventional radiological procedures (ICRP Publication 85). To minimize the radiation hazard for patients and providers, the ICRP has also issued recommendations for preventing skin injury that may be incurred by interventional radiological procedures (ICRP Publication 103). A number of previous studies have reported radiation doses incurred during coronary interventional procedures and during radiological cine-scanning that have exceeded the acceptable limits for skin doses, which are 2–6 Gy for skin edema, 3 Gy for alopecia and 18 Gy for tissue necrosis [7, 8]. These types of radiation injuries are closely associated with DAP values recorded during interventional radiological procedures. Therefore, recording total fluoroscopic time and DAP during such procedures is very important for the assessment of radiation injuries.
Medical radiation may have both stochastic effects and deterministic effects in humans. Koenig [6] has published the details of 13 cases of radiation injury incurred during cardiac angiography and coronary intervention procedures. The report includes detailed descriptions of patient demographics and of the examinations, locations of skin injuries and associated clinical findings [9].
Radiation Dose Monitoring
Radiation exposures incurred by patients during interventional radiological procedures at a cardiovascular centre are classified by type and are measured in terms of total fluoroscopy time, peak skin dose, reference dose and DAP. These measurement methods present values of different profiles. Total fluoroscopy time is defined as the total length of time of fluoroscopic radiation exposure as measured during the course of an interventional radiological procedure. The US FDA regulations have also stated that fluoroscopic times should be recorded by all types of angiography equipment [6]. Furthermore, the US FDA requires that all angiographic equipment that has been sold in the USA after June 2006 must measure, indicate and record reference doses including air kerma rate and cumulative air kerma [10]. The cumulative dose is an especially useful index for evaluating risk of deterministic radiation effect in patients [11]. The cumulative data represent the total entrance dose from the angiography equipment to patients. The DAP is measured through an ionization chamber located in the front of the collimator, and the measured values are transferred via DICOM. In Europe, The DAP is considered the reference value and is compared with the measured radiation exposures of patients undergoing interventional radiological procedures [12].
Digital Imaging and Communications in Medicine Modality Performed Procedure Step
In order to resolve the problem of hospital workflow, DICOM 3.0 was introduced by the American College of Radiology (ACR) and the National Electrical Manufacturer Association (NEMA). DICOM 3.0 is currently used in conjunction with the contents of the DICOM modality work list (DICOM MWL, 1995) and the DICOM modality performed procedure step (DICOM MPPS, 1998). The DICOM MPPS is a protocol for transferring data from radiologic examinations to the picture archiving and communication system (PACS) or the radiology information system (RIS) [13].
DICOM MPPS contains various data including patient information, the order information for the IR procedure, MPPS information, result information and radiation exposure information [14]. Patient information includes the name and age of the patient. The order information for the IR procedure contains an access number and the study instance unique identifiers (UID). The MPPS information includes location performed, starting date and time performed, ending date and time performed, name of procedure, status, and reason for discontinuation. Lastly, the radiation exposure result contains the anatomic region exposed, exposure time and exposure dose. However, many Korean hospitals use an order communication system (OCS) that is tailored to the needs of each hospital, rather than an RIS, which is regulated by the Integrating the Healthcare Enterprise (IHE) organisation, and therefore, most hospitals in Korea do not record or store DICOM MPPS data in particular.
Order Communication System
As noted, most hospitals in Korea use an OCS. The system has been installed from 1990 and is customized to each hospital [15]. The OCS was established before the introduction of DICOM MWL and DICOM MPPS. Thus, it does not follow any specific standard. Furthermore, as OCS focuses on the delivery of physician orders and application of medical insurance fees, it is not utilized for or does not store information such as MPPS. Therefore, so far, there are no reports concerning the use of MPPS through OCS.
Workflow of Interventional Radiological Procedures
In the past, most cardiovascular centres in Korea did not prescribe before interventional procedures but entered the order information into the OCS after the procedure was completed. This custom in Korean hospitals has been a major hindrance to the implementation of a DICOM MPPS that includes the characteristics of interventional procedures, and the OCS did not support DICOM MWL. However, when the cardiovascular information system (CVIS) and cardiovascular PACS (CV PACS) were simultaneously introduced into cardiovascular centres in Korea, the method of the order information was changed to comply with the IHE regulation. However, PACS and electronic medical records (EMR) that are used widely in Korean hospitals cannot record or store the DICOM MPPS information defined in DICOM 3.0. Because most hospitals in Korea use OCS, a standard communication protocol is essential for incorporating DICOM MPPS according to the workflow recommended by IHE. Therefore, a Health Level 7 (HL7) converter has been introduced to allow interlocking of the order information database from OCS with the order information database from CVIS in this study. HL7 was designated as the standard for interlocking information by the American National Standards Institute (ANSI) in 1994. Whether or not an interventional procedure is prescribed depends on the result of diagnostic cardiovascular angiography. Thus, it is difficult to transmit DICOM MWL to radiologic equipment before entering the order information in OCS. Also, as there were many tasks that had to be performed manually, such as entering patient information and matching order information to the images, the digitalization of hospital information has been completed slowly (Figs. 1 and 2).
Fig. 1.
Legacy workflow for cardiovascular angiography
Fig. 2.
DICOM MPPS workflow for cardiovascular angiography
Methods
System Configuration
Although PACS was used in the cardiovascular centre described in this study, images related to cardiac angiographic procedures were previously not stored or recorded at storage because of problems with transmission speed and the storage device. In 2010, this cardiovascular centre was appointed as a regional cardiovascular centre and was provided with a CVIS and a cardiac PACS. However, the connection between the DICOM MWL and the CVIS equipment remained unavailable because of the OCS interface until the EMR system was constructed in 2011. The CVIS connection for DICOM MPPS became available after that, and using this system, a monitoring system for radiation dose (radiation exposure monitoring system) could be established based on MPPS (Centricity CARDDAS, GE Healthcare, USA). Figure 3 shows the configuration of the cardiac information system that was constructed.
Fig. 3.
Configuration of the dose quality monitoring system
An Intel Xeon 20-GHz processor and 4 GB of CPU memory were used for construction of the CVIS. A dual hard disk drive (HDD) system was chosen to minimize the risk of data loss due to physical loss of one HDD. The database used was Gupta SQL 9.01 (Unify Corporation, California, USA), and a Microsoft Server 2008 R2 standard operating system was used (Microsoft Corporation, USA). In addition, in order to apply the OCS, Smart Gap was used to convert order information to HL7 on a Microsoft Windows Server 2008 R2 standard system (Microsoft Corporation, USA) that had an Intel Xeon 2.1 3-GHz processor with 2 GB of CPU memory. A modality for generation of MPPS has been installed and has been in operation with INOVA2000 (GE Healthcare, Milwaukee, WI, USA) angiography equipment since 2006. On this system, the MPPS information about the examination could be transmitted to the CVIS when cardiovascular angiography was performed, and after transmission, the data were recorded on the database and could be checked by the viewer software shown in Fig. 4.
Fig. 4.
Radiation dose quality monitoring software
The system that was constructed can provide information such as total fluoroscopic time and sum of DAP values from acquisition of images, exposure time of each run, total number of frames, direction of exposure, tube voltage, tube current, angle of projection and field of view.
Data Analysis
Data related to patient radiation exposures during coronary angiography from 875 interventional procedures conducted at a cardiovascular centre located in the Gyeongsang-do area of Korea were collected from March 1 to December 31, 2012. These data were used in analysis of frequency by SPSS v20.0 (IBM Corporation, New York, USA), and third quartile values were obtained.
This retrospective study complied with the Health Insurance and Portability and Accountability Act and was approved by our respective institutional review boards; the need for informed consent was waived.
Results
The data included information about the procedures such as total fluoroscopic time, total DAP, DAP for acquisition of radiographic images, DAP for fluoroscopy, number of frames, frame rate, total number of runs, and kilovolts, milliampere, milliamperes, projection angle and FOV (field of view) size for each run. The 875 procedures were divided into coronary angiography (CAG, n = 361, 41.3 %) and coronary angiography with percutaneous transluminal coronary intervention (PCI; n = 514, 58.7 %) (Fig. 5).
Fig. 5.
The sex distribution of patients according to procedure
Patient age range was from 18 to 87 years in the CAG group and 35 to 97 years in the PCI group. The average age of all patients was 63.3 years and the standard deviation was 11.9 years. The age distribution is shown in Fig. 6. The maximum DAP values for CAG and DPI were 853.9 and 1384.9 Gy cm2, respectively. The range of fluoroscopy times was 54 to 3948 s, average of 283.3 s, for CAG and 94 s to 5520 s, average of 984.3 s, for PCI (Fig. 5).
Fig. 6.
Age distribution of patients
In frequency analysis for CAG (Table 1), total fluoroscopy time for the third quartile diagnostic reference level (DRL) was 318 s, total DAP was 75.6 Gy cm2, X-ray exam DAP was 39.9 Gy cm2, X-ray exam fluoro DAP was 36.3 Gy cm2 and number of X-ray frames was 731.
Table 1.
CAG results
| Fluoro time | Total exam DAP (Gy cm2) | Exam exposure DAP (Gy cm2) | Exam fluoro DAP (Gy cm2) | X-ray (runs) | X-ray (frames) | X-ray (frames sec−1) | X-ray (kV) | X-ray (mA) | X-ray (mAs) | X-ray (SID cm) | X-ray (FOV cm) | Age (years) | |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| N | 361.00 | 361.00 | 361.00 | 361.00 | 361.00 | 361.00 | 361.00 | 361.00 | 361.00 | 361.00 | 361.00 | 361.00 | 361.00 |
| Mean | 283.33 | 67.60 | 34.64 | 32.96 | 9.91 | 664.96 | 15.04 | 95.94 | 575.80 | 4.53 | 110.41 | 14.17 | 61.72 |
| Median | 198.00 | 54.67 | 32.32 | 20.85 | 9.00 | 637.00 | 15.00 | 94.00 | 559.38 | 4.43 | 108.31 | 13.67 | 63.00 |
| Mode | 114.00 | 43.21 | 24.97a | 4.80a | 8.00 | 628.00 | 15.00 | 97.00000000a | 450.5555556a | 4.38 | 110.00 | 13.25 | 63.00 |
| SD | 309.52 | 59.36 | 13.81 | 51.28 | 4.05 | 228.04 | 0.87 | 47.20 | 212.51 | 1.90 | 47.04 | 5.85 | 12.56 |
| S2 | 95803.03 | 3523.56 | 190.64 | 2629.96 | 16.39 | 52001.72 | 0.75 | 2227.79 | 45162.10 | 3.62 | 2212.34 | 34.18 | 157.67 |
| Min | 54.00 | 13.81 | 0.18 | 3.75 | 1.00 | 9.00 | 5.00 | 34.00 | 234.20 | 1.80 | 41.53 | 5.47 | 18.00 |
| Max | 3942.00 | 853.95 | 115.05 | 738.90 | 49.00 | 1813.00 | 19.09 | 973.00 | 4410.00 | 40.00 | 998.00 | 123.00 | 87.00 |
DAP dose area product, SID X-ray source to image receptor distance, FOV field of view
aSmallest median values
The DRL values for PCI (Table 2) were total fluoroscopy time of 1207.5 s, total DAP of 213.3 Gy cm2, X-ray DAP of 74.3 Gy cm2, X-ray fluoro DAP of 135.6 Gy cm2 and number of X-ray frames of 1489. The DRL values for all patients were total fluoroscopy time of 834.0 s, total DAP of 160.2 Gy cm2, X-ray DAP of 63.8 Gy cm2, X-ray fluoro DAP of 100.3 Gy cm2 and number of X-ray frames of 1244. Most exams were performed at 15 frames s−1. The average tube voltage for all exams was 94.4 kV, average tube current was 4.5 mAs, average shooting distance was 109.6 cm and the average field of view (FOV) was 13.4 cm2 (Figs. 7 and 8).
Table 2.
PCI results
| Fluoro time | Total exam DAP (Gy cm2) | Exam exposure DAP (Gy cm2) | Exam fluoro DAP (Gy cm2) | X-ray (runs) | X-ray (frames) | X-ray (frames sec−1) | X-ray (kV) | X-ray (mA) | X-ray (mAs) | X-ray (SID cm) | X-ray (FOV cm) | Age (years) | |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| N | 514.00 | 514.00 | 514.00 | 514.00 | 514.00 | 514.00 | 514.00 | 514.00 | 514.00 | 514.00 | 514.00 | 514.00 | 514.00 |
| Mean | 984.26 | 187.63 | 64.84 | 122.79 | 28.18 | 1289.68 | 15.03 | 100.15 | 524.39 | 4.30 | 109.02 | 12.85 | 64.37 |
| Median | 720.00 | 143.51 | 57.71 | 83.07 | 25.00 | 1164.00 | 15.00 | 100.55 | 512.60 | 4.25 | 108.57 | 12.63 | 65.00 |
| Mode | 768.00 | 102.95a | 63.16 | 58.37a | 19a | 950.00 | 15.00 | 80.10526316a | 518.6666667a | 4.00 | 107.5a | 12.50 | 68.00 |
| SD | 855.01 | 157.86 | 35.53 | 127.83 | 12.59 | 660.49 | 0.16 | 13.31 | 73.26 | 0.45 | 10.44 | 1.44 | 11.43 |
| S2 | 731046.39 | 24919.27 | 1262.14 | 16341.20 | 158.56 | 436251.43 | 0.03 | 177.09 | 5367.75 | 0.20 | 108.99 | 2.07 | 130.69 |
| Min | 96.00 | 25.12 | 13.92 | 9.34 | 5.00 | 244.00 | 15.00 | 74.75 | 376.91 | 3.50 | 99.72 | 12.00 | 35.00 |
| Max | 5220.00 | 1384.94 | 389.47 | 995.47 | 116.00 | 8478.00 | 17.00 | 323.53 | 1505.53 | 12.47 | 332.67 | 42.13 | 97.00 |
DAP dose area product, SID X-ray source to image receptor distance, FOV field of view
aSmallest median values
Fig. 7.
Distribution of total fluoroscopy times for CAG (a) and PCI (b)
Fig. 8.
Distribution of total DAP values in CAG (a) and PCI (b)
Discussion
Coronary angiography and intervention procedures are examinations in which a specific quantity of radiation is delivered to the same area, and injury to this area by radiation exposure has been well documented. While it is essential to monitor and record patient exposure doses to digital angiographic equipment in real time, it is also very difficult to evaluate a patient’s risk of radiation exposure during the procedure. Direct measurement by equipment such as a thermoluminescent dosimeter (TLD) requires post-processing in order to evaluate the real dose; therefore, it is not adequate for monitoring at the clinical site. Furthermore, methods that measure doses directly are inappropriate because the chamber can appear in images as a foreign body. Indirect measurement tools such as a dose-area product meter are the preferred method for producing a DRL in order to prevent radiation injury to patients.
In the present study, the indicated DRLs for DAP were 75.6 Gy cm2 for CAG and 213.3 Gy cm2 for PCI. These values were higher than those reported in 2013 (57 Gy cm2 for CAG and 94 Gy cm2 for PCI) Neofotistou et al. [16]. The DRL for CAG in the present study was lower than the first action level of 125 Gy cm2 reported by Bogaert et al. [17], but in PCI, the DRL was close to the second action level of 250 Gy cm2. A DRL of 125 Gy cm2 corresponds to an exposure dose of 2 Gy, and appearance of erythema corresponds to maximum skin dose of 3 Gy, emphasizing that caution is needed to prevent radiation injuries to the skin [4]. In the present study, 6.9 % of CAG patient doses exceeded the first action level and 18.1 % exceeded the second action level.
In patients undergoing coronary intervention, the DAP should be checked with the monitor of the angiographic device during the procedure, and the angles of the X-ray tube should be controlled in order to avoid continuous exposure at the same area. However, the DAP value can overestimate the actual skin dose. When a patient completes an exam with a DRL exceeding an action level, in order to monitor for changes such as skin erythema, it is necessary to register the patient as a high-dose exposure patient and to follow up with trace investigation after 2 weeks. If meaningful changes appear, proper medical treatment should be provided and continuous trace investigation with hospital consultation is needed. If a patient who has exceeded action levels receives additional radiation exposure within 60 days, the total dose of all exposures should be evaluated [18]. Particular caution is needed in coronary angiography when additional exams to check prognosis are performed because of the additional exposure of the same area.
In the present study, the DRL for fluoroscopic time was 318 s (5.3 min) in CAG and 1207 s (20.1 min) in PCI. These values are quite high, although they are lower than the 30-min range for patients in the first action level in exams using fluoroscopy [11]. This can also come close to the total value in cases of two consecutive exposures, and it is desirable to establish the DRL, record the level of patient radiation exposure and monitor for radiation injury. DRL values established by studies similar to the present study are presented in Table 3. Also see Fig. 9.
Table 3.
Mean, median and third quartile DAP values for CAG and PCI
| Study | No. of patients | DAP or KAP (Gy cm2) | ||||
|---|---|---|---|---|---|---|
| Mean | Median | Third quartile | Range or upper limit | |||
| CAG | Present study 2012 | 361 | 67.6 | 54.7 | 75.6 | 13.8 to 853.9 |
| Sapinn et al. 2004 [19] | 176 | 48.6 | 37 | 59.6 | 6.3 to 452.8 | |
| Van de Putte et al. 2000 [20] | 62 | 60.64 | 56.82 | 80.58 | 144.23 | |
| Betsou et al. 1999 [21] | 29 | 30.4 | ||||
| D’Helft et al. 2009 [22] | 967 | 37.86 | 30.64 | 41.71 | 1.36 to 231.01 | |
| PCI | Present study 2014 | 514 | 187 | 143 | 213 | 96 to 5220 |
| Sapinn et al. 2004 [19] | 70 | 153 | 103 | 189.5 | 18.8 to 655.0 | |
| Van de Putte et al. 2000 [20] | 10 | 165.95 | 131.61 | 185.83 | 2.0 to 345.72 | |
| Betsou et al. 1999 [21] | 7 | 70.7 | ||||
| D’Helft et al. 2009 [22] | 463 | 78.3 | 58.05 | 83.56 | 47.5 to 410.38 | |
DAP dose area product, CAG coronary angiography, PCI percutaneous transluminal coronary intervention
Fig. 9.
Diagnostic reference levels for each procedure
Conclusion
Cardiovascular intervention is commonly performed and can be associated with high levels of skin exposure to radiation. Therefore, establishment of DRLs by accurate monitoring of real-time radiation exposure is absolutely necessary to reduce the risk to patients. In the present study, DRL values determined for fluoroscopic times were considerably higher than those reported in previous studies. This was a result of increased diagnostic and interventional procedure times, prior to invasive treatment, in patients with multi-vessel disease. Moreover, this is evidence that the effort to monitor exposure was inadequate and that images may be taken because of the convenience of on-site digital image processing.
The use of diagnostic reference levels has been shown to reduce the overall dose and the range of doses utilized in clinical practice. For example, UK national dose surveys demonstrated a 30 % decrease in typical radiographic doses from 1984 to 1995 and an average drop of about 50 % between 1985 and 2000 [23]. This study is important in that it is the first to identify DRLs to optimize radiation exposure of patients at Korean cardiovascular centres. Continuous monitoring using a real-time radiation dose monitoring system and optimizing DRLs in order to reduce radiation exposure in patients in Korea is necessary. Quality control process from real-time monitoring of exposure condition and dose information such as exposure DAP, fluoroscopic DAP, fluoroscopic time, frame rate and detector focus distance will reduce overall patient dose more.
Contributor Information
Jungsu Kim, Phone: 82-2-940-2755, Email: rtkjs@korea.ac.kr.
Deoknam Seo, Phone: 82-2-940-2755, Email: duck9298@korea.ac.kr.
Inseok Choi, Phone: 82-2-940-2755, Email: joajoa1982@korea.ac.kr.
Sora Nam, Phone: 82-2-940-2755, Email: srnam@korea.ac.kr.
Yongsu Yoon, Phone: 81-92-642-6675, Email: doublewhys@korea.ac.kr.
Hyunji Kim, Phone: 82-2-940-2755, Email: taminylove@gmail.com.
Jae Her, Phone: 82-2-940-2755, Email: enisi2001@korea.ac.kr.
Seonggyu Han, Phone: 82-2-940-2755, Email: goodcoat@korea.ac.kr.
Soonmu Kwon, Phone: 82-53-320-1311, Email: kwoncine2@hanmail.net.
Hunsik Park, Phone: 82-53-200-6591, Email: hspark@knu.ac.kr.
Dongheon Yang, Phone: 82-53-200-6591, Email: ddhyang@knu.ac.kr.
Jungmin Kim, Phone: 82-2-940-2755, Email: minbogun@korea.ac.kr.
References
- 1.International Commission on Radiological Protection Managing patient dose in digital radiology. ICRP publication 93. Ann ICRP. 2004;34:1–74. doi: 10.1016/j.icrp.2004.02.001. [DOI] [PubMed] [Google Scholar]
- 2.International Commission on Radiological Protection. 1990 Recommendations of the International Commission on Radiological Protection (Report 60). Annals of the ICRP. 1991; 21(1-3). (1991) [PubMed]
- 3.International Commission on Radiological Protection. Radiological Protection and Safety in Medicine (Report 73). Annals of the ICRP. 1996; 26(2):1-31. (1996). [PubMed]
- 4.Cynthia H. McCollough. Diagnostic Reference Levels. Available at http://www.imagewisely.org/imaging-modalities/computed-tomography/medical-physicists/articles/diagnostic-reference-levels. Accessed 16 December 2014
- 5.Conference of Radiation Control Directors, Inc. Technical white paper: Monitoring and tracking of fluoroscopic dose. CRCPD Publication #E-10-7 (2010)
- 6.Jaco JW, Miller DL. Measuring and monitoring radiation dose during fluoroscopically guided procedures. Tech Vasc Interv Radiol. 2010;13:188–193. doi: 10.1053/j.tvir.2010.03.009. [DOI] [PubMed] [Google Scholar]
- 7.International Commission on Radiological Protection. Avoidance of radiation injuries from medical interventional procedures. ICRP Publication 85 (2000) [DOI] [PubMed]
- 8.International Atomic Energy Agency. Applying radiation safety standards in diagnostic radiology and interventional procedures using x ray. Safety Report Series No. 39, Vienna (2006)
- 9.Koenig TR, Wolff D, Mettler FA, Wagner L. Skin injuries from fluoroscopically guided procedures: part 1, characteristics of radiation injury. AJR. 2001;177:3–11. doi: 10.2214/ajr.177.1.1770003. [DOI] [PubMed] [Google Scholar]
- 10.Electronic products; performance standard for diagnostic X-ray systems and their major components: final rule. Fed. Regist. 70. 33998-34042 (2005) [PubMed]
- 11.Stecker MS, Balter S, Towbin RB. Guidelines for patient radiation dose management. J Vasc Interv Radiol. 2009;20:S263–S273. doi: 10.1016/j.jvir.2009.04.037. [DOI] [PubMed] [Google Scholar]
- 12.Vano E, Järvinen H, Kosunen A. Patient dose in interventional radiology: a European survey. Radiat Prot Dosim. 2008;129:39–45. doi: 10.1093/rpd/ncn024. [DOI] [PubMed] [Google Scholar]
- 13.Noumeir R. Benefits of the DICOM modality performed procedure step. J Digit Imaging. 2005;18:261. doi: 10.1007/s10278-005-6702-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Vano, E., Padovani, R., Neofotistou, V., Tsapaki, V., Kottou, S., Ten, J. I., Fernandez, J. M., Faulkner, K. Improving patient dose management using DICOM header information. The European SENTINEL experience. Proceedings of the International Special Topic Conference on Information Technology in Biomedicine. Available at http://medlab.cs.uoi.gr/itab2006/proceedings/Medical%20Imaging/149.pdf. Accessed 19 July 2010
- 15.Korean Society of PACS Technology and Korean Society of Medical Imaging Technology. Textbook of PACS and Digital Image. 2nd edition. Page 199 (2007)
- 16.Neofistou V, Vano E, Padovani R, Kotre J, Dowling A, Toivonen M, Kottou S, Tsapaki V, Willis S, Bernadi G, Faulkner K. Preliminary reference levels in interventional cardiology. Eur Radiol. 2003;13:2259–2263. doi: 10.1007/s00330-003-1831-x. [DOI] [PubMed] [Google Scholar]
- 17.Bogaert E, Bacher K, Lemmens K, Carlier M, Desmet W, Dewagter X, Djian D, Hanet C, Heyndrickx G, Legrand V, Taeymans Y, Thierens H. A large-scale multicentre study of patient skin doses in interventional cardiology: dose-area product action levels and dose reference levels. Br J Radiol. 2009;82:303–312. doi: 10.1259/bjr/29449648. [DOI] [PubMed] [Google Scholar]
- 18.Stecker MS. Guidelines for patient radiation dose management. J Vasc Interv Radiol. 2009;20:S263–S273. doi: 10.1016/j.jvir.2009.04.037. [DOI] [PubMed] [Google Scholar]
- 19.Sapiin B, Ng KH, Abdullah BJJ. Radiation dose to patients undergoing interventional radiological procedure in selected hospital in Malaysia: retrospective study. JHK Coll Radiol. 2004;7:129–136. [Google Scholar]
- 20.Van de Putte S, Verhaegen F, Taeymans Y, Thierens H. Correlation of patient skin doses in cardiac interventional radiology with dose area product. Br J Radiol. 2000;73:504–513. doi: 10.1259/bjr.73.869.10884747. [DOI] [PubMed] [Google Scholar]
- 21.Bestou S, Efstathopoulos EP, Katritsis D, Faulkner K, Panayiotakis G. Patient radiation doses during cardiac catheterization procedures. Br J Radiol. 1998;71:634–639. doi: 10.1259/bjr.71.846.9849387. [DOI] [PubMed] [Google Scholar]
- 22.D’helft, C. J., Brennan, P. C., Mcgee, A. M., Mcfadden, S. L., Hughes, C. M., Winder, R. J. 3, and Rainford, L. A. Potential Irish dose reference levels for cardiac interventional examinations. Br. J. Radiol. 296-302 (2009) [DOI] [PubMed]
- 23.Hart D, Wall BF. U.K. population dose from medical X-ray examinations. Eur J Radiol. 2004;50(3):285–291. doi: 10.1016/S0720-048X(03)00178-5. [DOI] [PubMed] [Google Scholar]









