Uncertainty Calculation for Spectral-Responsivity Measurements

John H Lehman 1, CM Wang 1, Marla L Dowell 1, Joshua A Hadler 1
PMCID: PMC4646579  PMID: 27504228

Abstract

This paper discusses a procedure for measuring the absolute spectral responsivity of optical-fiber power meters and computation of the calibration uncertainty. The procedure reconciles measurement results associated with a monochromator-based measurement system with those obtained with laser sources coupled with optical fiber. Relative expanded uncertainties based on the methods from the Guide to the Expression of Uncertainty in Measurement and from Supplement 1 to the “Guide to the Expression of Uncertainty in Measurement”-Propagation of Distributions using a Monte Carlo Method are derived and compared.

An example is used to illustrate the procedures and calculation of uncertainties.

Keywords: laser, monochromator, optical fiber, optical power, spectral responsivity, uncertainty calculation

1. Introduction

The National Institute of Standards and Technology (NIST) provides a calibration service for absolute spectral-responsivity of optical-fiber power meters at wavelengths between 400 nm and 1800 nm [1]. The service is unique because it characterizes the optical-fiber power meter, not just an optical detector, which is normally coupled to a laser with an optical fiber. We rely on two measurement systems for a single calibration; one system to provide complete wavelength coverage, the second to duplicate the systematic behavior of coherent light exiting the fiber. In the first instance, we rely on a broadband source and a monochromator (the monochromator system) to cover the entire wavelength range because a continuously variable laser source is not currently available. In the second instance, several laser-diode sources, independently coupled to optical fiber (the fiber system), provide the optical input for measurements at discrete wavelengths. The reference detector for the monochromator system is a pyroelectric wedge-trap detector and the reference detector for the fiber system is an electrically calibrated pyroelectric described elsewhere [2].

Figure 1 displays typical measurement results from the monochromator system and the fiber system. Measurements from the monochromator system consist of the meter’s relative responsivity (A/W) at 10 nm increments (transmitted through air) over the entire wavelength range of the meter. Measurements from the fiber system represent the absolute responsivity at several laser wavelengths, typically at 850 nm, 1310 nm, and 1550 nm. In each calibration procedure, measurements from the monochromator system are normalized to higher accuracy measurements made with the fiber system. That is, the results from the mono-chromator-based measurement are adjusted (multiplied by a constant) so that they agree as closely as possible with the fiber-based measurements. The uncertainty of the meter’s calibration depends on the disagreement between the fiber-based and the adjusted mono-chromator-based results, as well as the uncertainties of the independent measurements. Independent uncertainty analysis of the fiber-based and monochromator-based measurements, including various type A and type B evaluations of uncertainties due to the working standard and optical sources, is given in [1]. In the present work, we describe procedures for calculating the normalized spectral responsivity and its uncertainty at each wavelength increment. The procedures consist of a least-squares estimation of the adjustment and a ratio estimation of the spectral responsivity. We compare two methods for calculating the uncertainty of the normalized spectral responsivity. We use an example to illustrate the procedures.

Fig. 1.

Fig. 1

Spectral-responsivity measurements of a commercially available optical fiber power meter based on a 5 mm diameter, germanium photodiode. The empty circles are obtained from the monochromator system. The solid circles are results from the fiber system. The solid line represents the meter’s spectral responsivity calibration.

2. Measurement Adjustment

Determination of the adjustment is based on the common spectral-responsivity measurements made at selected laser wavelengths. Let k be the number of laser wavelengths used, and (xi, yi), i = 1, …, k, be the fiber-based (xi) and monochromator-based (yi) measurements at these laser wavelengths. Since the measurement errors in xi are negligible relative to the measurement errors in yi, an adjustment factor c can be obtained by minimizing the sum of squares

i=1k(yicxi)2

and is given by

c=i=1kxiyii=1kxi2.

Once c is obtained, the monochromator-based measurements yj are normalized by

yj=yj/c. (1)

Also, based on the standard regression analysis, the square of the standard uncertainty of c is given by

u2(c)=i=1k(yicxi)2(k1)i=1kxi2,

with νc = k − 1 degrees of freedom.

3. Uncertainty Analysis

When one calculates the uncertainty of a ratio estimate like j in (1), two distinct scenarios must be recognized. The first involves dependent measurements where the quantities in the numerator and denominator are measured in the course of a single measurement experiment. In this scenario, it is expected that common sources of error exist that contribute to errors in estimating both quantities. Hannig et al. [3] calculated the uncertainty of the ratio estimate of dependent measurements based on the method proposed in the ISO Guide to the Expression of Uncertainty in Measurement (GUM) [4] and used it to construct confidence intervals. They compared the resulting GUM intervals with intervals obtained from an exact method, known as Fieller’s method [5], and concluded that the GUM interval is very similar to the exact Fieller interval and can be recommended in metrological applications.

For measurement experiments where the quantities in the numerator and denominator are uncorrelated, as in the case of j in this application, there is no exact interval available. (The reason they are not correlated is that c is determined from the measurements at laser wavelengths, while the yj used in calculating j are the monochromator-based measurements at non-laser wavelengths.) In this paper we first consider the GUM approach and calculate the uncertainty of j as

u2(yj)=u2(yj)c2+yj2u2(c)c4,

where u(yj) is the combined standard uncertainty of the monochromator-based measurement yj, which is discussed in detail in [1]. To obtain the expanded uncertainty we need to obtain the effective degrees of freedom associated with u (j). The GUM recommends the use of the Welch-Satterthwaite formula to evaluate the degrees of freedom, which is given by

ν=(u2(yj)c2+yj2u2(c)c4)2u4(yj)c4νb+yj4u4(c)c8νc,

where νb is the degrees of freedom corresponding to u(yj).

In spectral-responsivity measurements, a single relative uncertainty is calculated and reported for the entire measurement curve due to the multiplicative nature of the measurement equation [1]. That is, if we use the symbol ur(yj) to denote the relative uncertainty of yj, then

ur(yj)=u(yj)yj=b,

where b is a constant and does not depend on index j. Similarly, the relative uncertainty of c is

ur(c)=u(c)c.

As a consequence, we obtain the relative uncertainty of the normalized measurement j as

ur2(yj)=ur2(yj)+ur2(c), (2)

and its associated effective degrees of freedom,

ν=(ur2(yj)+ur2(c))2ur4(yj)/νb+ur4(c)/νc. (3)

The relative expanded uncertainty of j is then given by

Ur=t0.975,νur2(yj)+ur2(c), (4)

where t0.975, ν is the 0.975 quantile of the t distribution with ν degrees of freedom.

The Welch-Satterthwaite approximation in (3) is known to be effective when the two degrees of freedom νb and νc are of about the same order of magnitude [6]. Since in our calibration experiments, the number of laser wavelengths used is typically equal to three, i.e., νc = 2, and νb is much larger than 2, we also consider two alternative methods for calculating the relative expanded uncertainty of j. The first alternative method we consider is a fiducial procedure [7]-[11], and the second is the method from Supplement 1 to the GUM [12]. For the current application, these two alternative methods produce identical results, so we will discuss only the Supplement 1 method here.

The Supplement 1 method obtains a probability density function (pdf) for the measurand by propagating the pdf’s of the input quantities appearing in the measurement equation.

The resulting pdf describes one’s knowledge of the measurand given the observed data and assumptions made in assigning the joint pdf of the input quantities used in propagation. Once the pdf is obtained, a 95 % uncertainty interval for the measurand can be constructed by finding two limits such that the area under the pdf between these limits is 95 %. For example, the limits can be the 0.025 and 0.975 quantiles of the distribution. The measurand for this problem is the true spectral responsivity at each wavelength. Using the Supplement 1 approach, it can be shown that the pdf of the measur-and is equivalent to the distribution of

Yj=yju(yj)Tνb(1)cu(c)Tνc(2),

where Tνb(1) and Tνc(2) are independent random variables from the t distributions with νb and νc degrees of freedom, respectively. We can also express Yj using the relative uncertainties of yj and c as

Yj=1ur(yj)Tνb(1)1ur(c)Tνc(2)yj. (5)

Let z1 and z2 be the 0.025 and 0.975 quantiles of the distribution of Yj; then (z1, z2) is a 95 % uncertainty interval for the true spectral responsivity at wavelength j. Let

Qr=1ur(yj)Tνb(1)1ur(c)Tνc(2); (6)

then

Yj=Qryj.

Since ur(yj) = b, Qr is free of index j. Consequently, (z1, z2) = (j q1, j q2), where q1 and q2 are the 0.025 and 0.975 quantiles of the distribution of Qr. Note that q1 and q2 are the same for the entire wavelengths. To obtain the expanded uncertainty from the uncertainty interval (j q1, j q2), we note that in a 95 % symmetric interval of the form x ± U(x), the expanded uncertainty U (x) is half of the interval width. We write

(yjq1,yjq2)=(yj(1q1)yj,yj+(q21)yj). (7)

Since the interval (7) is not symmetric about j we use the maximum of (1 − q1) j and (q2 − 1) j as the expanded uncertainty of j. As a consequence, the relative expanded uncertainty of j based on the Supplement 1 method is given by

Ur*=max(1q1,q21). (8)

The quantiles q1 and q2 are most conveniently estimated by use of a Monte Carlo approach. This involves generating a large number of realizations from the distribution of Qr and determining q1 and q2 empirically. A single realization may be generated as follows:

  1. Generate a realization of Tνb(1) of a t random variable with νb degrees of freedom.

  2. Generate a realization of Tνc(2) of a t random variable with degrees of freedom, independent of Tνb(1).

  3. Calculate Qr as in (6).

We conduct a simulation study to compare the coverage probabilities of the uncertainty intervals constructed using Ur and Ur*. The coverage probabilities depend on the values of νb and νc, as well as the true relative uncertainties of the monochromator-based spectralresponsivity measurements (denoted by σrm) and of the adjustment factor (denoted by σrc). Table 1 displays the coverage probabilities of the GUM and Supplement 1 intervals for various combinations of σrm, σrc, νb and νc. The simulation parameters used here are closely related to those we observed in our calibration experiments. The standard error in each entry of Table 1, based on the assumption of binomial distribution and 10000 simulation runs, is 0.95(10.95)/10000=0.002.

Table 1.

Coverage probabilities of nominally 95 % GUM and Supplement 1 intervals

σrm (%) σrc (%) νb νc GUM Sup 1
 0.5 0.1  10 2 0.9515 0.9667
 1 0.1  10 2 0.9471 0.9539
 2 0.1  10 2 0.9500 0.9530
 0.5 0.5  10 2 0.9509 0.9801
 1 0.5  10 2 0.9567 0.9816
 2 0.5  10 2 0.9527 0.9694
 0.5 0.1  30 2 0.9497 0.9639
 1 0.1  30 2 0.9501 0.9564
 2 0.1  30 2 0.9489 0.9515
 0.5 0.5  30 2 0.9447 0.9753
 1 0.5  30 2 0.9493 0.9747
 2 0.5  30 2 0.9510 0.9666
 0.5 0.1 2 0.9493 0.9628
 1 0.1 2 0.9522 0.9579
 2 0.1 2 0.9499 0.9529
 0.5 0.5 2 0.9451 0.9749
 1 0.5 2 0.9487 0.9751
 2 0.5 2 0.9504 0.9665

The above study indicates that the GUM intervals perform well even for the extreme cases where νb = ∞ and νc = 2. The Supplement 1 intervals are more conservative, which implies that Ur* is greater than Ur for the cases considered here. This is due to the symmetrization of the Supplement 1 intervals that expands the size of the intervals. As a consequence, we use Ur as the relative expanded uncertainty for the absolute responsivity in the final calibration report.

4. An Example

We use the data shown in Fig. 1 to illustrate the procedures described in Sec. 3. Measurements were made at 10 nm increments over wavelengths from 750 nm to 1800 nm by use of a monochromator-based system. A detailed uncertainty analysis for the system [1] yielded a relative standard uncertainty of 0.62 % with practically infinite degrees of freedom. Measurements were also made at wavelengths 850 nm, 1310 nm, and 1550 nm by use of a fiber-based system with a relative uncertainty of 0.25 %. The three pairs of spectral-responsivity measurements (xi, yi) (in A/W) at these wavelengths are (0.233, 0.2307), (0.7125, 0.7119), and (0.8073, 0.7986). The adjustment factor based on these three pairs of measurements is found to be c = 0.9934193, with the standard uncertainty u (c) = 0.00344334 or ur(c) = 0.346615 %. The line in Fig. 1 shows the normalized spectral responsivity.

To calculate the uncertainty of the normalized responsivity, we first calculate the effective degrees of freedom in (3) as

ν=(0.622+0.3466152)2/(0.624/+0.3466154/2)35.

The relative expanded uncertainty is then given by

Ur=t0.975,350.00622+0.003466152=1.44%.

For comparison, the relative expanded uncertainty found by use of the Supplement 1 method is Ur*=1.90% based on 500000 Monte Carlo samples.

5. Conclusion

In this paper we described procedures for calculating the normalized spectral responsivity and its relative expanded uncertainty. The normalization is based on the common spectral responsivity measurements made at some selected laser wavelengths. We compared two methods based on the GUM and Supplement 1 to the GUM for calculation of uncertainty. We used a simulation study to demonstrate that the uncertainty intervals constructed using the expanded uncertainty obtained from the GUM approach maintain the nominal level of 95 % for all the parameters we encountered in our experiments. Thus, the GUM-based expanded uncertainty is given in the calibration report.

Acknowledgments

This work is a contribution of the National Institute of Standards and Technology and is not subject to copyright in the United States.

Biography

About the authors: Jack Wang is a Mathematical Statistician in the Statistical Engineering Division of the Information Technology Lahoratory at NIST in Boulder. John Lehman is Project Leader for Laser Radiometry. Marla Dowell is Group Leader for Sources, Detectors and Displays. Josh Hadler is a Physicist responsihle for calihrations of spectral responsivity of optical fiher power meters. John, Marla and Josh are part of the Optoelectronics Division at NIST. The National Institute of Standards and Technology is an agency of the U.S. Department of Commerce.

Contributor Information

John H Lehman, Email: john.lehman@nist.gov.

CM Wang, Email: chih-ming.wang@nist.gov.

Marla L Dowell, Email: marla.dowell@nist.gov.

Joshua A Hadler, Email: joshua.hadler@nist.gov.

6. References

  • 1.Lehman John H. Calibration Service for Spectral Responsivity of Laser and Optical-Fiber Power Meters at Wavelengths Between 0.4 μm and 1.8 μm. National Inst. of Standards and Technology Special Publication; 1999. pp. 250–53. [Google Scholar]
  • 2.Lehman John H. Pyroelectric trap detector for spectral responsivity measurements. Applied Optics. 1997;36:9117–9118. [Google Scholar]
  • 3.Hannig Jan, Wang CM, Iyer Hari K. Uncertainty calculation for the ratio of dependent measurements. Metrologia. 2003;40:177–83. [Google Scholar]
  • 4.BIPM, IEC, IFCC, ISO, IUPAC, IUPAP, OIML Evaluation of measurement data—Guide to the expression of uncertainty in measurement, First edition 2008. http://www.bipm.org/utils/common/documents/jcgm/JCGM_100_2008_E.pdf.
  • 5.Fieller EC. Some problems in interval estimation, Journal of the Royal Statistical Society, Series B. 1954;16:175–185. [Google Scholar]
  • 6.Wang CM. On estimating approximate degrees of freedom of chi-squared approximations. Communications in Statistics—Simulation and Computation. 1994;23:769–88. [Google Scholar]
  • 7.Wang CM, Iyer Hari K. Propagation of uncertainties in measurements using generalized inference. Metrologia. 2005;42:145–53. [Google Scholar]
  • 8.Wang CM, Iyer Hari K. A generalized confidence interval for a measurand in the presence of type-A and type-B uncertainties. Measurement. 2006;39:856–63. [Google Scholar]
  • 9.Wang CM, Iyer Hari K. Uncertainty analysis for vector measurands using fiducial inference. Metrologia. 2006;43:486–94. [Google Scholar]
  • 10.Hannig Jan, Iyer Hari K, Wang CM. Fiducial approach to uncertainty assessment accounting for error due to instrument resolution. Metrologia. 2007;44:476–83. [Google Scholar]
  • 11.Wang CM, Iyer Hari K. Fiducial approach for assessing agreement between two instruments. Metrologia. 2008;45:415–21. [Google Scholar]
  • 12.BIPM, IEC, IFCC, ISO, IUPAC, IUPAP and OIML Evaluation of measurement data—Supplement 1 to the Guide to the expression of uncertainty in measurement—Propagation of distributions using a Monte Carlo method. (First edition) 2008 http://www.bipm.org/utils/common/documents/jcgm/JCGM_101_2008_E.pdf.