A system to measure the Optokinetic and Optomotor response in mice

Friedrich Kretschmer 1,, Szilard Sajgo 1, Viola Kretschmer 1, Tudor C Badea 1,
PMCID: PMC4651724  NIHMSID: NIHMS715755  PMID: 26279344

Abstract

Background

Visually evoked compensatory head movements (Optomotor responses) or eye movements (Optokinetic responses) are extensively used in experimental mouse models for developmental defects, pathological conditions, and testing the efficacy of therapeutic manipulations.

New Method

We present an automated system to measure Optomotor and Optokinetic responses under identical stimulation conditions, enabling a direct comparison of the two reflexes. A semi-automated calibration procedure and a commercial eye tracker are used to record angular eye velocity in the restrained animal. Novel video tracking algorithms determine the location of the mouse head in real time and allow repositioning of the stimulus relative to the mouse head.

Results

Optomotor and Optokinetic responses yield comparable results with respect to determining visual acuity in mice. Our new head tracking algorithms enable a far more accurate analysis of head angle determination, and reveal individual head retractions, analogous to saccadic eye movements observed during Optokinetic Nystagmus.

Comparison with Existing Methods

To our knowledge this is the first apparatus allowing the direct comparison of Optomotor and Optokinetic responses in mice. Our tracking algorithms, which allow an objective determination of head movements are a significant increment over existing systems which rely on subjective human observation. The increased accuracy of the novel algorithms increases the robustness of automated Optomotor response determinations and reveals novel aspects of this reflex.

Conclusions

We provide the blueprints for inexpensive hardware, and release open source software for our system, and describe an accurate and accessible method for Optomotor and Optokinetic response determination in mice.

Keywords: Mouse, Vision, Retina, Behaviour, Optomotor response, Head movements, Optokinetic reflex, Eye movements, Video tracking Visual field

1. Introduction

The mouse is the most commonly used mammalian model organism due to its fast reproduction cycle, relatively small size and the vast amount of genetic strategies available for gene manipulations, functional manipulations, pathogenetic and therapeutic studies. Assays addressing the visual system of mice in basic and clinical research applications employ a variety of anatomic and physiological techniques. However the number of established paradigms for assessing overall visual function of the whole animal are limited. Tests can be broadly classified in (i) learning/discrimination assays, in which reward or penalty cues are used to train animals to discriminate visual stimuli, and (ii) tests addressing discrete visual reflexes, such as circadian photoentrainment, pupil constriction or vestibulo- ocular coordination. The reflex responses do not require training and can be easily quantified resulting in objective measures of visual function. During locomotion, animals integrate vestibular, proprioceptive and visual information to minimize apparent image motion across the retina. Compensatory eye (Optokinetic responses, OKR) or head/body movements (Optomotor responses, OMR) can be elicited when vestibular input is dissociated from visual information, by simulating a global movement of the visual surround. The measurement of these responses is a widely established method to assess various parameters of the visual system in the context of impaired visual function caused by disease models or developmental defects [1, 2, 3]. In this study we present a largely automated system to evoke, record and objectively quantify Optokinetic and Optomotor responses in mice.

1.1. Evoking the Optomotor and Optokinetic reflex in mice

To evoke Optokinetic and Optomotor reflexes in mice the animal is typically placed in the center of a cylinder or sphere that is covered with a repetitive pattern, such as stripes [4, 5], sinusoidal gratings [1], or random dots [6]. The pattern is rotated around the animal at a constant velocity [1] in one [1, 5] or alternating directions [7]. Alternatively the velocity can be changed sinusoidally [8, 9]. These patterns can be painted or projected onto the inner surface of card board, plastic or wood cylinders as previously described for OKR or OMR experiments in mice [4, 10, 11, 12] and other species [13, 14], generating a complete 360° image at a homogeneous distance from the animal. Centering in the horizontal plane can be achieved by restraining the animal [15, 11, 10] when measuring OKR. When measuring OMR, the freely moving animal can alter its distance to the walls of the physical cylinder, and therefore the perceived spatial frequency of the objects presented on the wall. Thus the tested spatial frequencies vary during the experiment. Digital presentation techniques (with projectors [16, 9] or monitors [1, 17, 18, 7]) significantly speed up experiments by allowing the presentation of a large array of stimuli in quick succession. By simulating a virtual cylinder a constant grating can be presented in the field of view of the animal. The actual physical distance to the stimulus has to be readjusted dynamically based on the location of the animal, either manually, by the observer [1], or automatically through software control [18, 7].

Perspective errors in the vertical dimension, as perceived at the top or bottom ends of the cylinder are not accounted for in case of the OMR but can be compensated by using a virtual sphere during OKR measurements [8].

1.2. The measurement of the Optomotor response in mice

During OMR measurements, stimuli consisting of striped patterns with distinct luminosities, contrasts, spatial or temporal frequencies can be presented in a sequential [12, 7] or step-wise manner [1, 19, 18]. The ability of mice to detect and react to these stimuli by head/body movements is judged for each individual condition by a human observer [1, 19] resulting in determination of threshold criteria [4, 1]. Typically these experiments are performed using a two-alternative, forced-choice protocol. The observer has to decide whether the mouse performed OMR [1] or has to determine the direction in which OMR occurred [19]. The determined visual acuity thresholds therefore depend on the subjective judgement and experience of the observer, resulting in an error-prone and time intensive procedure. Thus, automatic head/body detection is desirable. Body movements in mice were previously quantified using video-tracking by determining the speed of movement of the center of mass and the angular body velocity derived through a linear regression of the whole body [12]. A different approach used an automated approximation of the head movement (based on the snout location and the center of mass) to determine whether or not an animal saw a stimulus [18]. Based on this binary “yes” or “no” decision the spatial tuning of the visual system was derived by determining a contrast sensitivity curve. We previously presented a method for automatic head movement measurement during OMR experiments [7]. By computing the fraction of time tracked during 1 minute, spatial frequency dependency curves can be derived directly from the recorded angular head movements. The actual velocity of the head during OMR has never been analyzed in mice. The existing methodologies quantify all stimulus correlated behavior within a certain velocity range. This range inherently depends on the quality of the video tracking algorithm and does not necessarily rely on the actual velocity of the body or head during OMR. A linear regression through the thresholded pixel area of the mouse body was used to determine the body axis in [12]. The authors used a velocity threshold of ≤ 50 °/s to discard methodological artifacts. A different study [18] used commercial tracking software (Viewer3, Biobserve GmbH, Germany) and approximated the head location based on the snout location and the center of mass. The authors used a velocity threshold of ≤ 2.5 ×stimulus velocity to discard irregular movements. In another previous study [7] tracking behaviour was defined as head movements in the correct direction at an angular velocity of less than 9 °/s (0.8 × stimulus velocity at 12°/s). Thus, existing video-tracking algorithms overcome the subjectivity of approaches based on human observers, however are limited by reduced accuracy and low signal to noise ratios. In addition, a clear quantification of actual head angle and velocity, are not possible.

1.3. The measurement of the Optokinetic reflex in mice

Visual function in mice can also be analyzed using OKR measurements, in which the head of the mouse is restrained, and the eye position is determined using either magnetic coils or video tracking of pupil position. Since the eye has to stay in focus throughout such recordings the measurement typically requires a restraint of the animal. While the animal can be restrained in a non-invasive way using a restrainer box [10, 20] the quality of the recordings is not good enough to quantify responses in an automated way and derive measures like the gain from the data. Instead these studies rely on human observation. A much better signal to noise ratio can be achieved by fixing the animal by glueing [21, 22] or screwing [15, 23] a holder to its skull.

When measuring OKR the performance of an animal's visual system is traditionally assessed by quantifying the gain of the eye movement [15, 17]. The velocity at which the eye follows and the derived gain are parameters which can only be calculated by estimating the rotation center of the pupil. This data can either be approximated by assuming a general eye model [21] or it can be derived for each individual animal by calibrating the system [15]. The calibration procedure is performed to estimate the angle the eye moved based on the pixel coordinates determined by the eye tracker. This task is more difficult in rodents, since they are avofeate and cannot be asked or trained to focus on a defined point in space. To circumvent this limitation, typically several images of the eye are taken from various angles. The distance of the pupil from the center of corneal curvature is indirectly derived from these images based on the pixel distance of the pupil to one or more corneal reflections generated by a stationary light source [15]. This procedure is time demanding and requires precise control of the position of the camera to set it to a specific rotation and distance from the animal. Thus OKR determinations are an accurate, but more demanding approach compared to OMR analysis.

1.4. Motivation

Measurements of OMR in mice are far more common than measurements of the OKR. Beside the fact that these measurements are non-invasive and do not require a complicated calibration procedure, the most important reason might be the commercial availability of a device to measure OMR (Optomotry by CerebralMechanics Inc.). While the OMR serves as a common and useful tool to measure parameters of the visual system, the actual reflex, i.e. the movement of the mouse head in response to large scale visual stimuli, was never systematically analyzed in mice. A reason for this might be the lack of a suitable video tracking algorithm to determine the head and body angle precisely enough during an experiment. Commercially available tracking algorithms typically determine the center of gravity of the animal and the location of the snout and tail. These coordinates can only give a rough estimate of the head angle. For example it can be calculated as the difference between the angle of the vector from the tail to the snout and the angle of the vector from the center of gravity to the snout. This estimate is sufficient to determine whether OMR was induced or not [18], but is not precise enough for a quantitative analysis of the occurring head movements.

Given the technical peculiarities of the two reflexes, it is unclear how the OMR and OKR relate to each other and it is hard to directly compare threshold parameters determined through either methodologies because of the varying stimulus conditions and different setups that were used in previous publications. Beside the direct comparison of these two reflexes, our main goal was to develop a system that combines the advantages of previously published methodologies to measure OMR and OKR sequentially to enable routine measurements of both reflexes in mice for the study of disease models, developmental defects, genetic manipulations and drug treatments using the highest possible degree of automation.

2. Materials and Methods

2.1. Hardware

We have built a setup to evoke optokinetic and optomotor responses in mice. All components are mounted inside a custom built box 652 mm × 652 mm × 404 mm build from 30 mm × 30 mm extruded aluminium profiles (Bosch Rexroth, Germany). The box was designed using FreeCAD (the design files can freely be downloaded, link included in supplementary materials). The box accommodates four 23 inch LCD screens that surround the animal. The setup can quickly be reconfigured to measure either the Optomotor responses or the Optokinetic reflex.

2.1.1. Stimulus presentation

Stimuli are presented on four computer screens surrounding the animal covering the whole field of view with a texture projected onto the surface of a virtual sphere (see figure 1, c). The sphere is positioned in such a way that the animal is always located at its center. Hence the distance to all areas of the stimulus is held constant. OpenGL was used for the perspective projection of the sphere. Additionally a sphere with a radius one pixel smaller than the main sphere is rendered inside the larger sphere. This secondary sphere is transparent by default but can be covered with a solid or semitransparent texture to mask parts of the stimulus. Both spheres can be independently rotated about arbitrary axes and hence can be used to present any kind of horizontal, vertical or diagonal movement. The radius of the virtual sphere was set to 288 mm -this corresponds to ≈0.55× the width of the computer screens. This radius minimizes the difference between the real distance of the animal to the screens and the virtual distance of the sphere [24]. The software can also be used to present a virtual cylinder instead of a sphere.

Figure 1.

Figure 1

The setup can easily be converted between condition (a) to measure the OMR in the freely behaving animal and condition (b) to measure OKR in the head fixed animal. In both cases the animal is located inside a virtual sphere (c) that is presented on the four computer screens surrounding the animal. This sphere is adjusted in such a way that the animal is always located in its center (inset in (c)). (d) The lid, mirror and platforms can easily be removed.

We used four S23A700A 23-Inch (Samsung, South Korea) screens for stimulation. These screens were chosen because of their small panel frame border. The panels 120 Hz LCD module is illuminated by an LED back light that produces relatively little heat. The panels (523 mm × 302 mm) were removed from their plastic casing and were slid into plastic rails on each side of the box. The heat sinks on the back of the panel circuit boards touch the outer aluminium plates to passively transfer the heat to the outside. An EAH6950 graphic card (ASUSTeK Computer Inc., Taiwan) was used to drive the four screens that were connected through DisplayPort DVI-D Dual-Link Adapters (Accell B087B-002B). A fifth screen, placed outside the box, is used to display the image recorded by the camera and the graphic user interface and control the experiment. It was connected to the internal graphics card (ATI Radeon HD 4250) on the Motherboard trough DVI. Additional acrylic anti-glare sheets (TAP plastics, CA, USA) were placed in front of each panel leaving a small space in between to accommodate neutral density filters. These acrylic sheets also simplify cleaning of the arena. To reduce the luminosity levels of the LCD screens to scotopic light conditions we used gel sheet 211 .9 neutral density filters (Lee Filters, UK), cut to size and placed in Polyester Film sleeves (TAP plastics, CA, USA) in front of the screens. Each filter reduces the light intensity by ≈ 1 log unit (transmittance: 12.5%). By stacking two filters, the light intensity can be reduced by ≈2 log units. The three color channels of the four screens result in a combined light intensity of 9 × 1010 Q·s/cm2. This intensity was measured at the center of the platform using a 0.2cyc/° grating. The Michelson contrast ( CM=LmaxLminLmax+Lmin, for the maximum and minimum luminances) of the gratings was 100% (at the maximum color values).

A PC was used to drive the stimulation and perform the head tracking. We used a system with a AMD FX-8150 CPU running on a M5A88-V EVO AM3+ board (ASUSTeK Computer Inc., Taiwan), with 8 GB of RAM. A 128 GB SSD (510 series, Intel, CA, USA) was used during an experiments for recordings. An additional 2TB HDD (Deskstar 5K3000, Hitachi, Japan) was used to store the data.

Measurement of the Optomotor response

In the OMR measurement configuration (Figure 1, a) the bottom of the box is covered with a mirror (532 mm × 532 mm) with a 50.8 mm hole in its center. It is held in place by plastic rails which allow us to quickly remove it for cleaning.

A round platform (50.8 mm) is positioned in the center of the arena (at a distance of ≈262 mm using a short vertical aluminium profile (184 mm) attached to a horizontal profile running below the mirror. A second mirror (same size as the bottom mirror) is mounted to the top lid. The mirror is held in place by light-tight 8 mm glazing strips. The lid itself is mounted to the box through sliding door guides. Its edges are covered with rubber seals to allow a light-tight closure. All small parts were acquired from the Bosch aluminium framing catalogue (Bosch Rexroth, Germany). A Pro 9000 camera (Logitech, Switzerland) is mounted at the center of the top lid to record head movements and is connected to the main PC through USB. The PCB of the camera was removed from its casing and was directly screwed onto an acrylic sheet, which was attached to the mirror.

Measurement of the Optokinetic response

For OKR measurements an acrylic holder replaces the platform (see figure 1). The animal is fixed to this holder by its head implant. A hot mirror is located next to the animal's eye to observe the eye from above under infrared illumination using an ETL-200 (ISCAN, Burlington, MA, USA) video camera. The hot mirror appears transparent to the animal and hence the stimulus can be presented in the entire field of view. The hot mirror is mounted on a 70mm rotary stage (Edmund optics, NJ, USA) to allow a precise adjustment of the angle, required for the calibration of the system (see section 2.2.2). The camera can be translated smoothly in either the horizontal plane or vertical plane (for focal adjustments) using sliding carriages (OpenBuilds Part Store, NJ, USA) mounted on rails. Two electronic calipers (LCD Digital Vernier caliper 150 mm, CE Compass, CA, USA) were attached to the horizontal rail and one of the vertical side beams, allowing precise determination of the camera position (see section 2.2.2). We used the DQW Acquisition software (ISCAN, Burlington, MA, USA) to record the pixel coordinates of the pupil and reflection. The recording was triggered remotely by the okr arena software module through RS-232. The recorded data was transferred automatically to the main PC through the same interface after each trial and stored on the hard drive.

2.1.2. Infrared light conditions

In order to record under infrared light conditions we removed the infrared cut-off filter from the camera by unmounting the circuit board from its original casing, and used an injection needle to break out the glass filter. The infrared filter is located inside the autofocus unit, which has to be unsoldered in order to access the filter on this model. We determined the thickness of the filter under a microscope (300 μm) and glued two 150 μm coverslips at it's original position in order to maintain the auto-focus function. The circuit board was then screwed on to a polyethylene sheet which was centered and glued into the hole of the top mirror. We added 950nm IR strips (Elemental LED, CA, USA) in all four edges of the arena to provide general illumination (see figure 2a). The led strips were connected in series and their intensity/voltage could be controlled through an external DC power supply (R.S.R. Electronics Inc., NJ, USA) that allows us to alter the level of general illumination. To keep the lighting conditions as constant as possible under all stimulus conditions, an additional 25mm infrared bandpass filter was added in front of the camera (RT-830, Edmund optics, NJ, USA) to balance the luminance.

Figure 2.

Figure 2

(a) Infrared illumination is provided by infrared LED strips attached in all four edges inside the setup. (b) Masks to cover the left binocular field (green), right binocular field (magenta), left monocular field (gray) and right monocular field (blue). The dashed lines represent the visual field that is covered by the four screens. Areas based on [25].

2.2. Software

The software consists of several interconnected modules. A main program written in MATLAB (The MathWorks Inc., MA, USA) connects the various components (Figure 3) and provides a centralized graphical user interface to monitor and control experiments (Figure 4). In brief, stimuli are generated by the patterGen module. These stimuli are used by okrarena to render the virtual sphere. Head tracking is performed by the omr monitor module. The commercial ETL200 system is used for eye tracking and communicates with the MATLAB program.

Figure 3.

Figure 3

Software architecture of the system. The stimulus is presented on four screens by the program okr arena. This program receives images that are used as textures on the virtual sphere and protocols that describe the rotation of the sphere. These files were previously generated through the MATLAB routine patternGen. Head movements (OMR) are recorded with a camera at 25 Hz, which feeds the program omr monitor. This program records the video and determines the animal's head position. The head position is stored and communicated back to okr arena, which readjusts the position of the virtual sphere accordingly. To record eye movements (OKR) the ETL200 Eye tracker is remotely triggered by the stimulation software. The location of the pupil and cornea are read out at 120 Hz and are continuously buffered for storage. The experiment is controlled with a centralized graphical user interface.

Figure 4.

Figure 4

The graphical user interface. New stimuli can be generated in the right panel of the window. A stimulus consists of an image that is used as a texture on the virtual sphere, a mask that can cover areas of the sphere and a protocol that describes the movement of the sphere and the mask over time. Stimuli can be added to a list, which can then be run without human interaction. The left panel shows a live view of the inside of the arena with the platform and a set of user interface elements to set the parameters of the tracking algorithm. This is also where the user can set the recording path and start and stop an experiment.

Stimulus generation

Stimuli were designed using a custom MATLAB program patternGen, an improved version of the program described in [7]. This program is used to generate gratings, masks and rotation protocols. Gratings are mapped as textures on to the virtual sphere. We implemented routines to generate random dot patterns, saw-tooth, sinusoidal or square-wave gratings at arbitrary spatial frequencies. All patterns are initially generated as grayscale images. In a second step these pattern can be “colorized” to set a specific contrast or color schema. To allow a linear color and contrast adjustment all color values are color corrected based on the opto-electrical transfer function of the LCD panels we used. For this purpose we measured the luminance at various intensities for the three color channels of the screens using a calibrated spectrometer (USB-4000, Oceanoptics, Inc. Dunedin, FL, USA). All patterns are internally stored as PNG files. In case of vertical gratings the patterns are stored as one individual line of pixels, in case of horizontal gratings as one pixel column. This significantly reduces space and loading time of these patterns. Additionally, pattern masks can be generated from within the program to cover parts of the visual field of the animal. We implemented various generators for the masks. The mask can be used to obscure steradians of arbitrary sizes within the field of view of the animal. Based on previous literature [25] we also designed masks specific for the visual field of the mouse that cover only the monocular or binocular field of one or both eyes (see figure 2). Masks are also stored as PNG files. Transparency is realized through the alpha channel this format offers.

The rotation of the stimulus was realized through rotation protocols. These protocols contain a series of values for each refresh of the LCD monitors that describe the rotation of the sphere. Internally these protocols are stored as text files containing comma separated values, as required for the use with okrarena. We implemented rotations at arbitrary velocities and amplitudes as well as rotations at sinusoidally changing velocities (pendular movements).

A combination of a grating, a mask and a rotation protocol can be added to a list. A list contains a sequence of stimuli that can be presented consecutively during an experiment without human interaction. The order of the stimuli within a list can be randomized.

Stimulus presentation

The stimulation on the four screens was realized through an improved version of the standalone program okrarena ([7], GPLv3, freely downloadable at openetho.com) which uses OpenGL and the Simple Direct-Media Layer. We used this program to render the main sphere and the masking sphere on the four screens. okrarena can remotely be controlled through TCP/IP. In the described setup all programs (except for the eye tracking software) were running on the same PC and communication hence only occurred on a local socket. TCP/IP was realized from within MATLAB using JTCP (Kevin Bartlett, MATLAB Central). Once a set of stimuli is created (see section 2.2) and the experiment is started, the main MATLAB routine calls the separate program okrarena that receives all previously generated files as startup parameters. The MATLAB program then connects to okrarena to update the cylinder position based on the position of the animals head determined by the head tracking module (see section 2.2.1).

2.2.1. Head movements

Head movements are recorded by the top camera in the lid. We track the animal's head movements during an experiment and use the calculated head angle to automatically quantify the Optomotor response for each stimulus condition.

Head tracking algorithm

Two tracking algorithms were used to determine the location of the head:

A slightly modified version of the previously described algorithm [7] was used during the experiment to determine head position. This location is used to translate the virtual cylinder to keep the distance to the cylinder wall constant.

A novel tracking algorithm was used offline, after the experiments, to accurately determine the head angle and quantify tracking. The previously published version performs reliably under bright illumination [7], however it is less effective for some coat colors, under infrared illumination and/or dim light conditions.

The new version overcomes these issues, yields a better signal to noise ratio and runs much faster. It can compute real time data at up to 50 fps on an AMD FX-8150 CPU and thus is fast enough to be used for online visual stimulus repositioning. The algorithm currently runs within a standalone executable program and was implemented in C using the Open Source Computer Vision library (OpenCV, Willowgarage/Intel). The novel approach uses the contours in the image as a basis for all computation. It consists of : a) image thresholding, b) mouse contour determination, c) identification of the snout position, and d) determination of the head angle.

a) A range threshold is applied on all three color channels to segment the image.

The color values of the image I = (ik,l) ∈ ℝw × h with the height h and the width w are ik, l ∈ [0, 255] for all k,l. With the help of two threshold values s1, s2 ∈ [0, 255] a binarized image BI = (bik, l) ∈ {0, 1}w × h can be calculated:

bik,l={0,ifik,ls1ik,ls21,ifs1ik,ls2. (1)

b) The contour and the areas of all binary objects are derived from the binary image through the FindContours function of OpenCV [26]. Instead of using the largest object within a region of interest [7] we introduced more specific geometric constraints that are applied consecutively (see figure 5, e) to discriminate the mouse shape from other contours:

  1. Contours outside of a certain area range (10000 px ≤ area ≤ 40000 px) are discarded.

  2. Contours that do not cross at least three of the four quarters of the image (E.g. animals snout in upper left part of image, the hind part of the body in the bottom right and bottom left part of the image) are discarded.

  3. The contour of the animal is then defined as the contour with the closest center of mass to the center of the image.

Figure 5.

Figure 5

The new tracking algorithm allows us to determine the head orientation of animals of different coat color under various lighting conditions. To keep the lighting conditions more constant under all stimulus conditions regardless of stimulus type, an additional 25 mm infrared bandpass filter can be placed in front of the camera. a) A mouse with agouti coat color at photopic light condition without the additional infrared filter. b) A black coated animal at photopic light condition with infrared filter. c) An albino animal at scotopic light condition with infrared filter. d) A black coated animal at scotopic light condition with infrared filter. Colors in a-d: Green line: detected contour of the mouse, green dot: center of gravity, yellow dot: corrected center of mass, magenta dot: “tail” location, blue dot: “snout” location, blue circles: two radii around the snout location, cyan and yellow: the contour within the two radii, red dots: the two points that determine the head location e) Geometrical constraints to reliably separate the contour of the mouse from the background: 1. The detected contour needs to be within area limits. 2. The contour needs to cross three or more quadrants of the image (green area). 3. The contour needs to be the closest contour to the center of the image. f) Illustration of the “snout stabilization” method. The snout (green dot) is detected as the point farthest away from the center of gravity (yellow dot). 2. Incorrect snout detection: when the mouse changes it's posture the center of gravity shifts (blue) and a different location is detected as the snout (green dot) since it is further from the center of mass (new green segment) In the example the location jumps to the other side of the head. 2′, Corrected snout detection method: We rotate the center of gravity (blue dots) around the snout location from the frame 1 (black dot). The point farthest away from this rotated center of gravity lies on the same side as the location determined in 1 (green dot).

Based on this contour ℳt, defined as the set of pixel positions Mt, at time point t we calculate the reference locations on the body. The hindquarters are calculated as the center of gravity, which lies towards the back of the animal. The center of gravity Cogt of the set ℳt at timepoint t is calculated as:

Cogt=1|Mt|MtMtMt,with|Mt|the number of elements (2)

The point Ct ∈ ℳt farthest away from the body center Cogt is used to determine the rough location of the head:

CtargmaxMtMtMtCogt (3)

c) Snout identification (Figure 5): We introduce a novel feature which we named “snout stabilization” that specifically addresses two particular cases, which impaired the quality of tracking in a previous implementation [7]. In some circumstances, the center of mass of the animal does not lie in it's hind quarter but is located more frontally. This happens often with very thin animals or animals that stretch their body during the experiment. Hence the detection of the snout fails and the tail is detected instead. Additionally, throughout all recordings animals tend to rotate their head along the long axis of the head (roll rotation) or the interaural axis (pitch rotation) resulting in only a very small area of the snout being visible (or not visible at all). In these situations the snout is mistakenly detected on either the right or the left part of the head, e.g. at the insertion of the ear lobe into the head (compare figure 5, c with 5, f). Both issues not only reduced the quality of tracking but led to the necessity of manually going through all recordings to exclude such periods. These issues can be solved for tracking periods during which pitch and roll axes are relatively constant by the following procedure. The algorithm remembers the locations of the head from the previous frame and rotates the calculated center of mass from the current frame around the snout location of the previous frame in increments of 1°, to minimize the euclidean distance between the previous head location and the current head location (see figure 5, f). In practice the center of gravity is rotated in a range of -30° to +30° from its original position around the snout location from the previous frame. For each iteration the algorithm calculates the point farthest away from each rotated center of gravity. The one closest to the location of the previous frame is picked as the new snout location.

Hence we rotate the center of gravity Cogt around the point Ct − 1 (from equation 3) of the previous frame recorded at t − 1 by the angle θ by a 2D rotation:

With (x,y), the coordinates of Cogt and (xCt − 1, yCt − 1) the coordinates of the snout in the last frame, the rotated coordinates (x′, y′) can be calculated as:

x=cos(θ)(xxCt1)sin(θ)(yyCt1)+xCt1 (4)
y=sin(θ)(xxCt1)+cos(θ)(yyCt1)+yCt1 (5)

We perform this operation in a range from -30° to +30° and create a set Inline graphict of rotated center of gravities Cogt. We then calculate the points of ℳt farthest away from each of the rotated center of gravities.

CogtAtCtargmaxMtMtMtCogt (6)

For the new set Inline graphict of all determined snout locations Ct we determine the point closest to Ct − 1 by calculating the euclidean distance between Ct and Ct − 1, which defines the new snout location Ct.

CtargmaxCtCtCtCt1 (7)

d) Head orientation: (see figure 5) Typically, there is a fair amount of pixel noise occuring right at the edge of the shape. This noise is especially pronounced in the snout region due to the whiskers and the inhomogeneous color caused by the interrupted fur coverage in this area. These considerations make the direct use of the snout position as a reference point for head angle determination impractical. We therefore determine the head orientation using two additional reference points. (see figure 5, f), which more robustly define the head orientation. These points are defined as follows.

  1. The retrieved snout location derived in c) serves as the center for two circular regions of interest, the diameter of which should be equal or smaller than the length of the mouse head. The value of both radii r1 and r2 is set empirically so that the two locations reflect the anatomy and size of the measured animal.

  2. These two radii r1 and r2 are used to define two additional contours (cyan and yellow outlines in figure 5, a,b,c,d). The sets corresponding to pixel of the mouse head within the two regions are then:
    UC,r1={MtMt|MtCt<r1}. (8)
    UC,r2={MtMt|MtCt<r2}. (9)
  3. For each of the contours, two additional centers of gravity Cogr1 of UC,r1 and Cogr2 of UC, r2 are calculated. The center of gravity within the first, smaller radius corresponds to a point on the bridge of the snout. The second center of gravity within the larger radius determines the point of rotation of the head. Due to the symmetry of the head these locations are located on the sagittal plane of the head.

  4. The head-orientation g⃗ can be computed as the vector between the two centres of gravity.
    g=Cogr1Cogr2. (10)

A requirement is, that the locations of the head are reliably detected in the first frame that is recorded. This is usually the case since the mouse calmly rests on the platform at this time. If the tail is detected instead of the snout the user can correct this by clicking on the tail with the mouse cursor once only at the beginning of the recording.

Based on these additional refinements the tracking algorithm performs robustly on animals of all ages, sizes and coat colors (including albinos) under all light conditions investigated (Figure 5).

Determining the optimal dual threshold criterion

There is little information on the ability of mice to match the visual stimulus velocity with their head movements (OMR head gain). During typical recording sessions, mice can occasionally be engaged in exploratory movements, sniffing and changing position on the platform, in a visual stimulus independent manner. In addition, head and body angle changes result in pixel jitter, and therefore contour and head angle artifacts. These various sources of apparent head movements result in a wide range of head angular velocities in the same direction or opposite direction to the stimulus (Figure 6a). Thus the head movements potentially triggered by the visual stimulus are embedded in a significant amount of unrelated head angle changes. We reasoned that, in the absence of given visual cues, the random head movements would occur with relatively equal distributions in both clockwise and counterclockwise directions, and that visually driven responses, if occurring with enough frequency would bias this distribution in the stimulus (correct) direction. Given these considerations, we decided to compare globally the amount of head motions in the correct and incorrect directions. For this purpose we collected a data set of 10 animals (C57B/6) which we recorded under the reported optimal stimulus conditions (maximum contrast, spatial frequency 0.2 cyc/°, stimulus velocity 12°/s [1, 19, 7]) for 10 trials each. We then determined the number of frames containing head movements in the correct stimulus direction (nCorrect) or opposing direction (nWrong) with an angular head velocity window faster than Vmin and slower than Vmax. The amount of tracking was then computed as the ratio nCorrect/nWrong. In order to determine the window of velocities to be included in our analysis we systematically varied the upper and lower velocity thresholds VMin and Vmax (VMin: 0°/s to 12°/s, Vmax: 12°/s to 40°/s) and recorded the nCorrect/nWrong ratios for each condition (Figure 6b). Low velocity thresholds (Vmin < 2°/s) clearly reduced the ratio (Figure Figure 6b). The upper velocity threshold has less impact but a slight decrease of the ratio can still be observed at values of Vmax >20/s. Under optimal stimulus conditions the tracking behavior and hence the nCorrect/nWrong ratio should be maximal, if the ideal threshold criteria were set. At the speed of 12°/s stimulus, chosen in our experiments, movements in a range of 2-14°/s seem to correlate well with stimulus direction.

Figure 6.

Figure 6

Quantifying head movements during OMR experiments. (a) Illustration of OMR quantification. The histogram plots the number of frames for which the animal's head was moving with a given velocity in either stimulus direction (“correct”, positive angle value) or against stimulus direction (“incorrect”, negative value). Windows of head velocities of amplitudes and orientations similar to the stimulus velocity (correct, green), are compared to the windows of identical amplitudes but reversed orientation (“incorrect”, magenta). The ratio between the sums in the correct and incorrect windows is defined as the Optomotor response. The histograms were calculated for all trials of one animal at the control condition (Left, 0.2 cyc/°, stimulus not moving), 0.05 cyc/° (Center) and 0.2 cyc/° (Right) (b) The effects of varying the thresholds for the windows defined in (a) on the determined value of the OMR. The x and y axes represent the slow and fast motion criteria, expressed in °/s. The heatmap color represents the amount of visually driven responses for each threshold criteria, expressed as number of tracking events in the correct vs. incorrect direction, illustrating the influence the thresholds have on the outcome of the analysis. The stippled lines determine the lower and upper bounds of the head velocity windows that were used for successive analysis. Data were collected from 10 animals, measured 10 times at each condition.

Quantifying head movements

We can count the frames where the movement occurred in stimulus direction as:

Tυheadυstim=fFRAMES:υstim10°/sυhead<υstim+2°/s (11)

where υheadυstim (parallel velocities)

and the movements against stimulus direction as:

Tυheadυstim=fFRAMES:υstim10°/sυhead<υstim+2°/s (12)

where υhead ‖ −υstim (antiparallel velocities)

The ratio of the two sums is then used to quantify the response as:

OMR=TυheadυstimTυheadυstim (13)

Figure 6a illustrates the procedure. The boundaries are indicated by the cyan and green areas on the histogram for a stimulus running at 12°/s (blue line).

Further quality control

All recorded head and eye movement data were manually checked in a first step. For this purpose we developed two MATLAB tools to quickly browse through the traces of all recordings. In case of the head movements we implemented a program (BES) to browse through the recordings. The program allows us to discard segments in which the mouse was jumping from the platform consecutive times throughout an experiment. It also enables the user to retrack segments where the automated head tracking algorithm did not perform optimally due to a suboptimal setting of the threshold values (e.g. due to uneven or wet fur). For this purpose the user can set a new color value and retrack the selected segment using the described tracking algorithm.

2.2.2. Eye movements

The eye tracker derives the x and y coordinates of the mouse eye (see section 2.1.1). The recorded traces are the basis for an automatic detection of slow and fast components of the response. These data were used to quantify the Optokinetic reflex for each stimulus condition.

Detecting eye movements

We implemented the MATLAB program eyeMovementMarker to automatically detect and mark all slow and fast components within a recording. The detection is performed based on the following steps:

The trace (recorded at 120 Hz) is filtered with a moving average filter with a window of 10 samples. All continuous movements in stimulus direction with a minimum length of 20 samples are marked as slow components. If a movement in the opposite direction (minimal duration 10 samples) occurs directly after such a period, this event is considered to be the fast component of the OKR. The results are then displayed to the user, who can delete or merge periods manually. The pixel coordinates are converted into vertical and horizontal eye angles (azimuth and elevation) using the radius of pupil rotation (rp) previously derived through our calibration procedure (transformation described in Stahl, 2000).

Calibration of the system

The eye tracking system determines the x and y position of the pupil and one corneal reflection on the 512 px × 256 px CCD sensor at 120 Hz. To derive the azimuth and elevation angles of the eye from this data, the system needs to be calibrated. Since the mouse cannot be asked or trained to orient it's eye to a defined point in space the 3D center of the corneal curvature and the radius around which the pupil rotates are instead estimated from a series of images taken from defined angles by rotating the camera around the animal [15]. Instead of physically rotating the camera around the animal one can also translate the camera along an axis and then rotate it about its axis of rotation [27]. We adapted this methodology to our system. To avoid the camera being anywhere in the field of view of the animal we used a hot mirror (Edmund optics, NJ, USA Optics) to monitor the eye from above and performed such pseudo rotations by rotating the mirror instead of the camera. Figure 7 shows this geometrical relationship. In order to determine the distance from the sensor of the camera to the mirror we performed a camera calibration procedure using a chequerboard pattern that was attached to the mirror. We used the Camera calibration toolbox for MATLAB [28]. We followed all the calibration steps as described in [27] and implemented a MATLAB script that navigates the user through the procedure. Since our rails were not motorized the user had to translate the camera and mirror by hand and rotate the mirror. This was facilitated by the two electronic callipers that registered the position in real time. The circuitry from the callipers was connected to an Arduino Uno microcontroller board (Arduino LLC and Smart Projects S.R.L., Italy) through a custom circuitry to adjust their voltage for read-out. The Arduino was connected to the Main PC through USB so that the location of the calipers was available from within the MATLAB script.

Figure 7.

Figure 7

(a) A mouse restrained in the acrylic holder. To derive the azimuth and elevation of the eye, the system needs to be calibrated. (b) Two callipers enable a precise translation of the camera. A laser pointer facilitates alignment of the camera and the hot mirror. (c) By translating the camera and rotating the mirror we can take multiple images of the eye from different virtual angles to calibrate the system. (d,e) Validation of the calibration procedure. Adjusted and measured angles across a range of 0-10° in the horizontal (d) and vertical (e) plane. The mean error was calculated as the mean difference between the actual angle and the measured angle.

Validating the calibration procedure

We used a glass bead (radius 4 mm) to simulate a mouse eye. A circle was painted on the bead to represent the pupil. The pupil was rotated to a specific angle by using a 70mm rotary stage (Edmund optics, NJ, USA). We performed two separate measurements to determine the precision and accuracy of the angle determination in the horizontal and vertical plane by mounting the rotary stage respectively. We performed the calibration procedure 5 times and determined the gaze angles at 20 different elevations and azimuth. The results are shown in figures 7d and 7e. In a range of 0-20° our calibration yields a mean error <1°. The error is higher at large excursions (≈2.5° at -20° and +20°) The precision (determined by calculating the mean standard deviation was 0.43° for azimuth and 0.72° for elevation.

Quantifying eye movements

The total number, Gain, length, duration and amplitude of all movements are calculated for each individual trace based on the found onsets and offsets of the slow and fast components. The mean eye velocity to determine the gain is calculated using a linear regression over each detected time window.

2.2.3. Further Analysis

All further steps in the analysis are automated, require no human interaction and are performed in the same way for head- and eye movements using our analysis framework okranalyzer.

During an experiment, the recording software automatically generates file names that contain all relevant stimulus parameters (e.g. spatial frequency, contrast, stimulus velocity, stimulus amplitude, mask used etc.) for an experiment, complemented by a user defined prefix and suffix that typically indicates the ear tag of the animal and its age and/or genotype.

These parameters are parsed from the file names and allow us to automatize the analysis to some extent for any given condition. This meta data and data like the date and time of the experiment are stored in a structure that is maintained throughout all steps of analysis. A graphical user interface is provided to access this structure and easily select any subset of recordings and plot arbitrary dependency curves (e.g. median tracking performance of all 20 week old animals throughout all trials where the stimulus velocity was 12°/s and the stimulus amplitude was 60 in relation to the spatial frequency). The provided tool chain can cover any upcoming experimental condition and is not limited to the current listed parameters.

2.3. Mouse Handling, Surgeries and Experimental Condition

C57BL/6 mice were kept in plexiglas cages (46 × 26 × 16 cm [L × W × H]) and housed at room temperature with a 12 hr light/dark cycle. Food and water was available ad libitum. The mice were ear punched to recognize individuals. Experiments were conducted between 8 a.m. and 11 a.m. over a period of one week. We recorded Optomotor and Optokinetic responses of C57BL/6 mice using three different stimuli.

  1. A control stimulus consisting of a sinusoidal grating with a spatial frequency of 0.2 cyc/° which was not moving.

  2. A stimulus with a low spatial frequency of 0.05 cyc/° rotating at a velocity of 12°/s.

  3. A stimulus with a spatial frequency of 0.2 cyc/° (reported optimum) rotating at a velocity of 12°/s.

The moving stimuli alternated their direction every 5 s. We recorded 3 trials of 1 minute for each condition. For the determination of the dual threshold criterion for the analysis of OMR experiments 5 male and 5 female animals were used. For the comparison of OKR and OMR we used 3 female and 2 male animals. All animals were 8-16 weeks old. All mouse handling procedures used in this study were approved by the National Eye Institute Animal Care and Use Committee (ACUC) under protocols NEI 640, NEI 651 and NEI 652. All NIH rodent surgery guidelines were followed.

2.3.1. Headpost Surgery

All instruments were sterilized using a germ terminator dry glass bead sterilizer. A sterile environment was prepared and mice were anesthesized using an intraperitoneal Ketamine (80-120 mg/kg) and Xylazine (1-10 mg/kg) injection. Ophthalmic ointment was applied to the eyes to prevent corneal drying. Once the mouse was areflexive to paw pinch, the skin over the dorsal skull was opened with a 10 mm incision and the skull surface was washed with 3% hydrogen peroxide for about 30s and then dried with a cotton swab. The skull was then washed several times with saline. Holes for four 1/16 inch diameter screws were drilled into the bone; the screws were inserted, and dental cement was used to fill the space between the screws. The metal head-post was then inserted into the mound to secure the animal during testing. More dental cement was used stabilize the head post. After the surgery that lasted ≈ 15 m the mice were placed on a 37°C heating pad. The mice typically awoke 40 to 60m after surgery and were monitored for feeding, grooming, and gait. Buprenorephine (0.5 mg/kg) was delivered as needed for pain or discomfort. Eye movements were recorded not earlier than 48 hours after surgery.

3. Results

3.1. Exemplary head and eye movements

In case of the measurement of head movements the animal is freely moving on the platform. The animal does not necessarily continuously participate in the experiment but instead may explore the platform or perform sniffing or grooming behaviour. Figure 8a shows three typical traces that were recorded at three different spatial frequencies. We quantify Optomotor responses as OMR=TυheadυstimTυheadυstim (see section 2.2.1). The total time the mouse was moving its head in the correct direction (stimulus direction) was more than 1.5 × more than the time the mouse was moving in the incorrect direction at the reported optimal spatial frequency of 0.2cyc/° [7, 1]. The ratio decreases at suboptimal values (see 0.05 cyc/°) and is close to one when the stimulus is not turning (Nullcondition). The mouse hence moves its head as much in the correct as in the incorrect direction at Nullcondition. Figure 10 shows the calculated Optomotor response at all stimulus conditions of all trials.

Figure 8.

Figure 8

Exemplary trace of recorded head (OMR) and eye movements (OKR) at three different stimulus conditions. The Nullcondition (0.2 cyc/°, stimulus not moving), and two spatial frequencies 0.05 cyc/° and 0.2 cyc/° presented with a stimulus rotating at 12°/s. Note the different scaling on the y-axes in the inserts.

Figure 10.

Figure 10

Comparison of OKR and OMR in five C57BL/6 mice at three stimulus conditions. i) Nullcondition, s stationary grating of 0.2 cyc/°, ii) a very low spatial frequency of 0.05 cyc/° which is turning at 12 °/s and iii) the reported optimal spatial frequency of 0.2 cyc/° turning at 12°/s. OMR was quantified as described in section 2.2.1. OKR is represented by the Gain (eye velocity / stimulus velocity). Three trials were measured for each animal. In both OMR and OKR experiments, the individual trials for each of five mice, are represented in separate columns at each stimulus condition. The exemplary trials depicted for mouse 3 in figures 8a and 8b are marked with a cross. Note that random head movements under Nullcondition result in an Optomotor response index centered around 1. No eye movements are detected under the Nullcondition (OKR Gain = 0). Both OMR and OKR increase at optimal spatial frequency (0.2 cyc/°) compared to a lower special frequency (0.05 cyc/°).

Optokinetic eye movements typically consist of two distinct components. A slow phase during which the eye moves in stimulus direction and a fast phase that follows the slow phase and occurs in the opposite direction [23]. We occasionally observe a similar behaviour based on the head movements. Figure 9b shows the x and y coordinates of the two landmarks on the head (see section 2.2.1). The isolated x and y translation of the head reveal distinct tracking periods that consist of slow periods in stimulus direction lasting up two 2 s and faster very short periods in the opposite direction. These fast periods can also be automatically detected as movements of the head along the axis between the two landmarks (figure 9a bottom trace). To our knowledge, this is the first report of head saccade movements during Optomotor recordings in mice. Surprisingly, the slow component of these head tracking movements consists of either pure translation (figure 9c, t0 t2) or combines rotation and translation (figure 9c, t3 t5).

Figure 9.

Figure 9

The mouse head does not perform a homogeneous rotation. (a) The movement of the head is instead composed of a rotation and translation. To illustrate this phenomenon the x and y pixel coordinates of the two tracked locations of the forehead are plotted separately. (b) The animal performs head saccades that consist of a fast retraction of the head rather than a rotation opposite to stimulus direction as described in OKR. This retraction can be visualized by calculating the translation of the mouse head along the axis of gaze (bottom trace in (a)). (c) The time course of six individual slow phases (green), followed by fast phases (magenta). These periods occurred consecutively at time points t0 to t5. Traces are shown for both markers (Top: Marker 1, Bottom: Marker 2).

Three exemplary traces of recorded eye movements are shown in figure 8b. The mouse is restrained and rarely performs spontaneous eye movements. Optokinetic responses occur during the entire minute. At low spatial frequencies (0.005 cyc/° the eye continuously follows the stimulus but does not perform resetting saccades (fast phase). At the optimal spatial frequency (0.2 cyc/°) both the slow and the fast component of the response can be observed. We automatically detect the slow phases in the recordings.

Traditionally the gain (eye velocity / stimulus velocity) is used as a measure to determine the tuning of the visual system to the spatial and temporal aspects of the stimulus. Since we calibrated the system (see section 2.2.2) we can calculate the velocity at which the eye followed and hence the gain (Azim. eye velocity / Stimulus velocity). Figure 10 illustrates the gain we measured in five animals.

4. Discussion

We present here a hardware apparatus and software suite designed to detect and analyse Optomotor and Optokinetic responses in mice. Our integrated stimulus display and data collection pipeline allows for the presentation of a large variety of stimuli coupled with the automated recording and analysis of head or eye movements, resulting in significantly improved OMR visual threshold determinations, and the first quantitative analysis of visually driven head movements in mice.

4.1. Stimulus Presentation

Our setup can display visual stimuli on the inner surface of a freely rotating virtual sphere, ensuring that the mouse visual system perceives them under the same angle and at the same spatial resolution. A significant aspect of our stimulation conditions is that the visual stimulus can be centered in real time on the mouse head horizontal plane position, using feedback from our improved video tracking algorithm (see below). This feature is particularly useful when trying to assess the spatial resolution detected by the mouse, and facilitates the recording procedure, reducing the subjectivity of human participation in the experiment. The ability of masking allows us to cover portions of the visual field that correspond to defined areas like the binocular or monocular field of the animal. The ability to concomitantly translate and/or rotate the visual stimulus will allow the use of our setup in the design of other behavioral paradigms, including for instance looming behaviour [29].

4.2. Quantifying OMR in mice

Our improved tracking algorithm (section 2.1.1) can reliably identify animals of different coat color under photopic and scotopic light conditions. We determined the velocity range of stimulus correlated head movements (figure 6b) and found stimulus correlated head movements in a velocity range of 2° /s to 14° /s. This information is crucial to objectively quantify visually driven behaviour in future. We introduce here a new global metric to quantify Optomotor responses in mice. We had previously found that the fraction of time at which stimulus correlated behaviour occurs [7] is not ideal in situations where an animal moves very little in general but still participates well in the experiment or when an animal is very agitated. By taking the ratio of all movements in the correct and incorrect directions we avoid this problem and can more directly compare the performance of different animals regardless of their overall activity status. Our tracking algorithm, enabling the determination of the actual trajectory of landmarks on the mouse head allows for a more precise description of the occurring Optomotor responses (Figure 9a). We are now able to identify individual head retractions that follow a tracking period similarly to the fast component of the Optokinetic response. These movements do not seem to consist of a simple rotation of the head against stimulus direction but are more complex and seem to involve a strong translational component. For this type of head movements, one could derive the actual head gain of the slow component and use head saccades to quantify the response.

Compared to other methodologies our approach does not rely on human observation [1] requiring trained personnel. Currently available setups require optimally several people performing the experiments to assure that the measurements are not biased by observer visual perception. The user is not only required to assess the animals visual performance but is also responsible for adjusting the position of the stimulus to keep the grating constant. While an automated readjustment has been described before [18, 7] our system allows us to directly measure spatio- temporal properties of the OMR, rather than relying on a binary decision criterion [18, 1], under all stimulus conditions. Due to the lack of additional illumination and a limitations of the tracking algorithm the previously published approach [7] was limited to photopic conditions.

A direct assessment of the reflex enables a measurement of parameters that would otherwise only be indirectly possible by varying a secondary parameter (e.g. contrast to measure contrast sensitivities at various spatial frequencies). This significantly speeds up experiments. This was previously possible [7] but the improved tracking algorithm allowed us to apply a much more stringent criteria and enable a far more detailed analysis of OMR. We are for the first time able to directly validate and compare OMR and OKR which allows for a more detailed analysis of compensatory reflexes under the exact same conditions.

4.2.1. Calibration of the eye tracking system

Our calibration procedure combines the advantages of several previously published methodologies: In order to maintain a complete 360° surrounding stimulus, our camera is not placed anywhere in the field of view of the animal ([27], [15]) but is instead located above the animal (similar to [8]). The image is recorded through an infra red reflective mirror to see the eye. To calibrate the system under this arrangement we used a semi-automated calibration procedure involving electronic callipers to give a precise control of the elevation and translation of the camera. By combining the translation of the camera with a rotation of the mirror we are able to determine the position of the pupil and corneal reflections from various virtual viewpoints. A previous approach simulated rotations around the animal, by rotating and translating the camera ([27]) but this required the camera to be in the animals field of view. All rotations and translations of the camera and mirror in our setup could be automatized by the use of stepper motors (similar to [27]). Since the Eye Tracker also determines the pupil diameter one could easily integrate the measurement of the pupillary reflex into the existing system.

4.2.2. Comparison of OMR and OKR

One of the advantages of our setup is the ability to derive quantitative measures of gain for both reflexes in response to the same set of stimuli. Thus, the OKR gain we measured for five animals varied from ≈0.7 to ≈0.9. Similar values have been described before [17] at comparable stimulus conditions. While the OKR Gain can directly be compared to absolute values from previous studies, this is not possible in case of OMR due to the different nature of analysis. Qualitatively the results match previous findings very well: Mice seem to perform OMR [1, 7] and OKR [17, 30] best at a spatial frequency of ≈0.1-0.2cyc/°. Reduced responses have been reported at lower spatial frequencies like 0.05 cyc/° for both OMR[1, 7] and OKR [17, 30].

The OMR performance of the animal correlates well with the OKR performance. Animals that perform better than other animals under one condition also outperform their litter mates in the other condition (figure 10).

A more detailed parallel of the two reflexes will be described in a follow-up study.

4.3. Software design

Our integrated software suite, facilitates recording and analysis in several ways. All components are interfaced through a GUI. The program design is modular, and hence individual components, such as the stimulus display (okrarena) and tracking algorithm (omrmonitor) could also be used in entirely different behavioural paradigms in mice or other species. Stimulus generation is straight-forward and flexible, and the entire stimulus, recording and analysis workflow is highly integrated. The software could run simultaneously on several machines controlled from one central computer through TCP/IP could potentially enable high throughput measurements. The tracking algorithm okrarena is free software under GNU GPLv3 license, is compatible with all major operating systems (GNU/Linux, MAC, Windows) and can be downloaded at openetho.com. In addition, all software components are available on request from either of the corresponding authors (the level of documentation for individual components varies and the software is provided as is).

4.4. Implication and future plans

Whereas OMR measurements have mostly benefited the clinical research community, in particular with regard to the study of disease and therapy models, the OKR analysis in mice has been typically related to the study of neuronal circuits, specifically vestibulo-ocular integration. The measurement of OMR responses in the freely behaving mouse allows us to quickly and objectively assess its vision. Such measurements can be performed periodically over a long time period and the methodology is hence suitable for screening studies on a large number of animals involving genetic manipulations, drug treatments, ageing or retinal degenerations. The improved objectivity and ease of use of our setup makes it particularly useful for these purposes. The ability to precisely describe the trajectory of mouse head movements, allows their quantification and evaluation of relevant parameters, such as amplitude and gain. This will allow us a direct comparison with eye movements elicited by visual stimulation during OKR in the same setup. This comparison will be interesting from the clinical investigation perspective, since it will better relate OMR data to the wealth of visual circuit knowledge available to the eye motion field. It will also be interesting to the visual circuit community, since it will allow for the direct comparison of OMR and OKR contribution to gaze stabilization. Given the flexibility of our setup, it will be interesting to explore other visually driven head and eye movements, and their perturbations in normal and pathologic conditions.

Highlights.

  • A novel system for automated detection of Optomotor and Optokinetic responses under identical stimulation conditions reveals parallels between the two reflexes.

  • Novel tracking algorithms allow the accurate representation of complex visually evoked head movements.

  • We provide the blueprints for inexpensive hardware, and release open source software for our system, making it easy to replicate.

Acknowledgments

Hugh Cahill has provided guidance with implementing the head surgeries and designing the head fixation apparatus. Felix Ighovie Onojafe has provided technical help with head surgery optimization. Haohua Qiang has helped with calibration and measurement of screen light intensities under scotopic and photopic conditions. Lance Optican has provided helpful suggestions and critical reading of the manuscript. Funding for this work was provided by the Intramural Research Program of the National Eye Institute.

Footnotes

URL: openetho.com (Friedrich Kretschmer)

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

Contributor Information

Friedrich Kretschmer, Email: friedrich@openetho.com.

Tudor C. Badea, Email: tudor.badea@nih.gov.

References

  • 1.Prusky GT, Alam NM, Beekman S, Douglas RM. Rapid quantification of adult and developing mouse spatial vision using a virtual optomotor system. Invest Ophthalmol Vis Sci. 2004;45(12):4611–4616. doi: 10.1167/iovs.04-0541. [DOI] [PubMed] [Google Scholar]
  • 2.Anstis S, Hutahajan P, Cavanagh P. Optomotor test for wavelength sensitivity in guppyfish (poecilia reticulata) Vision Res. 1998;38(1):45–53. doi: 10.1016/s0042-6989(97)00159-4. [DOI] [PubMed] [Google Scholar]
  • 3.Tappeiner C, Gerber S, Enzmann V, Balmer J, Jazwinska A, Tschopp M. Visual acuity and contrast sensitivity of adult zebrafish. Front Zool. 2012;9(1):10. doi: 10.1186/1742-9994-9-10. URL http://dx.doi.org/10.1186/1742-9994-9-10. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Thaung C, Arnold K, Jackson IJ, Coffey PJ. Presence of visual head tracking differentiates normal sighted from retinal degenerate mice. Neurosci Lett. 2002;325(1):21–24. doi: 10.1016/s0304-3940(02)00223-9. [DOI] [PubMed] [Google Scholar]
  • 5.Abdeljalil J, Hamid M, Abdel-Mouttalib O, Stphane R, Raymond R, Johan A, Jos S, Pierre C, Serge P. The optomotor response: a robust first-line visual screening method for mice. Vision Res. 2005;45(11):1439–1446. doi: 10.1016/j.visres.2004.12.015. [DOI] [PubMed] [Google Scholar]
  • 6.Dubois MF, Collewijn H. Optokinetic reactions in man elicited by localized retinal motion stimuli. Vision Res. 1979;19(10):1105–1115. doi: 10.1016/0042-6989(79)90005-1. [DOI] [PubMed] [Google Scholar]
  • 7.Kretschmer F, Kretschmer V, Kunze VP, Kretzberg J. Omr-arena: automated measurement and stimulation system to determine mouse visual thresholds based on optomotor responses. PLoS One. 2013;8(11):e78058. doi: 10.1371/journal.pone.0078058. URL http://dx.doi.org/10.1371/journal.pone.0078058. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.van Alphen B, Winkelman BHJ, Frens MA. Three-dimensional optokinetic eye movements in the c57bl/6j mouse. Invest Ophthalmol Vis Sci. 2010;51(1):623–630. doi: 10.1167/iovs.09-4072. [DOI] [PubMed] [Google Scholar]
  • 9.Kretschmer F, Ahlers MT, Ammermüller J, Kretzberg J. Automated measurement of spectral sensitivity of motion vision during optokinetic behavior. Neurocomputing. 2012;84:39–46. [Google Scholar]
  • 10.Mitchiner JC, Pinto LH, Vanable JW. Visually evoked eye movements in the mouse (mus musculus) Vision Res. 1976;16(10):1169–1171. doi: 10.1016/0042-6989(76)90258-3. [DOI] [PubMed] [Google Scholar]
  • 11.Mangini NJ, Vanable JW, Williams MA, Pinto LH. The optokinetic nystagmus and ocular pigmentation of hypopigmented mouse mutants. J Comp Neurol. 1985;241(2):191–209. doi: 10.1002/cne.902410207. [DOI] [PubMed] [Google Scholar]
  • 12.Schmucker C, Seeliger M, Humphries P, Biel M, Schaeffel F. Grating acuity at different luminances in wild-type mice and in mice lacking rod or cone function. Invest Ophthalmol Vis Sci. 2005;46(1):398–407. doi: 10.1167/iovs.04-0959. [DOI] [PubMed] [Google Scholar]
  • 13.Easter S, Jr, Nicola GN. The development of vision in the zebrafish (danio rerio) Dev Biol. 1996;180(2):646–663. doi: 10.1006/dbio.1996.0335. [DOI] [PubMed] [Google Scholar]
  • 14.Beck JC, Gilland E, Tank DW, Baker R. Quantifying the ontogeny of optokinetic and vestibuloocular behaviors in zebrafish, medaka, and goldfish. J Neurophysiol. 2004;92(6):3546–3561. doi: 10.1152/jn.00311.2004. doi:10.1152/jn.00311. 2004. [DOI] [PubMed] [Google Scholar]
  • 15.Stahl JS, van Alphen AM, Zeeuw CID. A comparison of video and magnetic search coil recordings of mouse eye movements. J Neurosci Methods. 2000;99(1-2):101–110. doi: 10.1016/s0165-0270(00)00218-1. [DOI] [PubMed] [Google Scholar]
  • 16.Rinner O, Rick JM, Neuhauss SCF. Contrast sensitivity, spatial and temporal tuning of the larval zebrafish optokinetic response. Invest Ophthalmol Vis Sci. 2005;46(1):137–142. doi: 10.1167/iovs.04-0682. [DOI] [PubMed] [Google Scholar]
  • 17.van Alphen B, Winkelman BHJ, Frens MA. Age- and sex-related differences in contrast sensitivity in c57bl/6 mice. Invest Ophthalmol Vis Sci. 2009;50(5):2451–2458. doi: 10.1167/iovs.08-2594. URL http://dx.doi.org/10.1167/iovs.08-2594. [DOI] [PubMed] [Google Scholar]
  • 18.Benkner B, Mutter M, Ecke G, Münch TA. Characterizing visual performance in mice: an objective and automated system based on the optokinetic reflex. Behav Neurosci. 2013;127(5):788–796. doi: 10.1037/a0033944. doi:10.1037/ a0033944. URL http://dx.doi.org/10.1037/a0033944. [DOI] [PubMed] [Google Scholar]
  • 19.Umino Y, Solessio E, Barlow RB. Speed, spatial, and temporal tuning of rod and cone vision in mouse. J Neurosci. 2008;28(1):189–198. doi: 10.1523/JNEUROSCI.3551-07.2008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Sinex DG, Burdette LJ, Pearlman AL. A psychophysical investigation of spatial vision in the normal and reeler mutant mouse. Vision Res. 1979;19(8):853–857. doi: 10.1016/0042-6989(79)90018-x. [DOI] [PubMed] [Google Scholar]
  • 21.Iwashita M, Kanai R, Funabiki K, Matsuda K, Hirano T. Dynamic properties, interactions and adaptive modifications of vestibulo-ocular reflex and optokinetic response in mice. Neurosci Res. 2001;39(3):299–311. doi: 10.1016/s0168-0102(00)00228-5. [DOI] [PubMed] [Google Scholar]
  • 22.Sakatani T, Isa T. Pc-based high-speed video-oculography for measuring rapid eye movements in mice. Neurosci Res. 2004;49(1):123–131. doi: 10.1016/j.neures.2004.02.002. [DOI] [PubMed] [Google Scholar]
  • 23.Cahill H, Nathans J. The optokinetic reflex as a tool for quantitative analyses of nervous system function in mice: Application to genetic and drug-induced variation. PLoS ONE. 2008;3(4):e2055. doi: 10.1371/journal.pone.0002055. doi:10.1371/ journal.pone.0002055. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Kretschmer F. Ph D thesis. Fakultät für Informatik, Wirtschafts- und Rechtswissenschaften der Carl von Ossietzky Universität Oldenburg; 2012. Automatisiertes tracking von kopf- und augenbewegungen bei 360°-stimulation zur charakterisierung des visuellen systems kleiner wirbeltiere. [Google Scholar]
  • 25.Dräger UC. Observations on monocular deprivation in mice. J Neuro-physiol. 1978;41(1):28–42. doi: 10.1152/jn.1978.41.1.28. [DOI] [PubMed] [Google Scholar]
  • 26.Suzuki S, et al. Topological structural analysis of digitized binary images by border following. Computer Vision, Graphics, and Image Processing. 1985;30(1):32–46. [Google Scholar]
  • 27.Zoccolan D, Graham BJ, Cox DD. A self-calibrating, camera-based eye tracker for the recording of rodent eye movements. Front Neurosci. 2010;4:193. doi: 10.3389/fnins.2010.00193. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Bouguet JY. Camera calibration toolbox for matlab. World Wide Web electronic publication; 2008. URL http://www.vision.caltech.edu/bouguetj/calib_doc/ [Google Scholar]
  • 29.Yilmaz M, Meister M. Rapid innate defensive responses of mice to looming visual stimuli. Curr Biol. 2013;23(20):2011–2015. doi: 10.1016/j.cub.2013.08.015. doi:10.1016/j.cub. 2013.08.015. URL http://dx.doi.org/10.1016/j.cub.2013.08.015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Tabata H, Shimizu N, Wada Y, Miura K, Kawano K. Initiation of the optokinetic response (okr) in mice. J Vis. 10(1) doi: 10.1167/10.1.13. [DOI] [PubMed] [Google Scholar]