Abstract
The new BD Max GC real-time PCR assay showed high clinical and analytical sensitivity and specificity. It can be an effective and accurate supplementary test for the BD ProbeTec GC Qx amplified DNA assay, which had suboptimal specificity, and might also be used for initial detection of Neisseria gonorrhoeae.
TEXT
Neisseria gonorrhoeae is estimated to cause 106 million gonorrhea cases every year, resulting in substantial morbidity and economic costs globally (1). The need for highly sensitive and specific laboratory diagnostics and subsequent effective therapy is crucial. Nucleic acid amplification tests (NAATs) have increasingly replaced culture diagnostics internationally, due to their higher sensitivity, speed, automation, and use of noninvasive specimens (2–7). However, due to suboptimal specificity, especially of several early generation gonococcal NAATs, in several regions (e.g., Europe and Australia), it is recommended to verify positive samples with another NAAT targeting a different genetic sequence, particularly in low-prevalence populations and in pharyngeal infections (6, 8–18). The psychological, social, and legal consequences of false-positive gonococcal test results can be substantial.
The BD Viper System with XTR technology (BD Diagnostics, Sparks, MD) is a third-generation platform that when operating in “extraction mode” provides automated DNA extraction using ferric oxide and strand displacement amplification (19). The BD ProbeTec GC Qx amplified DNA assay, targeting a pilin-inverting gene, is used on the BD Viper system to detect N. gonorrhoeae. However, suboptimal (particularly analytical) specificity and cross-reaction with commensal Neisseria species have been described for this assay as well as most other gonococcal NAATs (6, 8, 13, 20). Recently, the BD Max GC real-time PCR assay, targeting the gonococcal opcA gene, was developed to be run on the BD Max system, which provides automated DNA extraction and real-time PCR.
We evaluated the performance of the new BD Max GC real-time PCR assay by examining clinical specimens positive in the BD ProbeTec GC Qx amplified DNA assay and samples spiked with isolates of gonococci, nongonococcal Neisseria species, and other closely related bacteria.
During July to October 2014, 23,815 individuals (14,846 females and 8,969 males representing asymptomatic individuals presenting for screening and symptomatic patients) were tested with the BD ProbeTec GC Qx amplified DNA assay in a single replicate according to routine diagnostic protocol. All positive clinical specimens were subsequently stored in the primary tube (including BD transportation medium) prior to analysis (DNA extraction and real-time PCR) with the BD Max GC real-time PCR assay, which was performed within 1 to 12 h. Specimens negative in the BD Max GC real-time PCR assay were further tested with the Aptima Combo 2 assay (Hologic, Bedford, MA) and a gonococcal dual-target real-time PCR targeting the porA pseudogene and opa genes (21).
To challenge the analytical sensitivity and specificity of the BD Max GC real-time PCR assay, 460 bacterial isolates were examined. These isolates comprised gonococci (n = 189), nongonococcal Neisseria species (n = 261), and closely related bacteria (n = 10) (Table 1). Species was determined using routine phenotypic methods, including the sugar utilization test, the PhadeBact GC monoclonal test (Mkl Diagnostics AB, Stockholm, Sweden), and matrix-assisted laser desorption–ionization time of flight mass spectrometry (MALDI-TOF MS) (Microflex LT; Bruker Daltonics, Bremen, Germany), and genotypic methods (Aptima Combo 2 and Aptima GC [Hologic], a gonococcal dual-target real-time PCR [21], and 16S rRNA gene sequencing). Cultures of gonococcal and nongonococcal isolates were suspended in BD ProbeTec CT/GC Qx specimen collection tubes, and 500 μl was resuspended in a BD Max UVE sample buffer tube to concentrations of approximately 4 colonies/ml and 20 colonies/ml, respectively. All false-positive or false-negative analytical samples were retested from both the original dilution and fresh dilution using new culture from frozen stock. The retesting was also performed on different dilutions after repeated species verification, according to previously described algorithms (22). All testing using commercially available tests was performed in accordance with the manufacturer's instructions.
TABLE 1.
Bacterial species | No. of isolates: |
|
---|---|---|
By species | Positive by BD Max GC real-time PCR | |
Gonococcal | ||
N. gonorrhoeaea | 189 | 189 |
Nongonococcal Neisseria | ||
N. meningitidisb | 150 | 0 |
N. lactamica | 20 | 0 |
N. flavescens | 20 | 0 |
N. subflava | 13 | 0 |
N. mucosa | 12 | 0 |
N. sicca | 11 | 0 |
N. cinerea | 10 | 1 |
N. perflava | 8 | 0 |
N. gonorrhoeae subsp. kochii | 4 | 0 |
N. flava | 2 | 0 |
N. sicca/subflava | 2 | 0 |
N. polysaccharea | 1 | 0 |
N. animalis | 1 | 0 |
N. weaver | 1 | 0 |
N. caviae | 1 | 0 |
Other Neisseria species | 5 | 0 |
Total | 261 | 1 |
Closely related species | ||
Kingella denitrificans | 3 | 0 |
Moraxella catarrhalis | 3 | 0 |
Moraxella osloensis | 1 | 0 |
Other Moraxella species | 3 | 0 |
Total | 10 | 0 |
N. gonorrhoeae isolates reflecting temporally (isolated from 1973 to 2013), geographically (mainly a global panel), and genetically different strains.
N. meningitidis isolates reflecting temporally, geographically, and genetically different strains.
Of 23,815 patients tested with the BD ProbeTec GC Qx amplified DNA assay, 85 (0.6%) females and 259 (2.9%) males were positive. Of these 344 positive specimens, 322 (94%) contained sufficient material for testing with the BD Max GC real-time PCR assay. Two-hundred fifty-two (78%) and 70 (22%) specimens were positive and negative, respectively. All 70 negative specimens were negative also in the Aptima Combo 2 NAAT, and 69 of them were repeatedly negative in the gonococcal dual-target PCR. These 69 false-positive specimens were obtained from pharynx (50.0%), urine (33.0%), vagina (10%), rectum (4.3%), and cervix (1.4%) (Table 2).
TABLE 2.
No. of samples BD Viper positive | No. of samples positive/negative by BD Max GC real-time PCR confirmatory testing |
|||||||
---|---|---|---|---|---|---|---|---|
Positive | Negative (% of BD Viper positive)a |
|||||||
Male |
Female |
|||||||
Pharynx | Urine | Rectum | Pharynx | Vaginal | Urine | Cervix | ||
322 | 252 | 27 (33) | 19 (14) | 3 (15) | 8 (57) | 8 (24) | 4 (50) | 1 (33) |
All 70 negative samples were negative also in the Aptima Combo 2 NAAT (Hologic); however, one of the vaginal samples was repeatedly positive in the gonococcal dual-target PCR (21).
In the analytical examination of the BD Max GC real-time PCR assay, all gonococcal isolates were positive, and all but one of the nongonococcal isolates (99.4%) were negative (Table 1). The cross-reacting Neisseria cinerea strain was also positive in retesting from the initial dilution and new culture from frozen stock, and 16S rRNA gene sequencing, MALDI-TOF, growth characteristics (colistin susceptibility and growth on tryptic soy agar and Mueller-Hinton agar [23]), and the sugar utilization test confirmed the N. cinerea species. The detection limit of this N. cinerea strain was 10 genome equivalents per reaction, which indicates clinical relevance. A sample spiked with this N. cinerea strain also cross-reacted in the BD ProbeTec GC Qx amplified DNA assay.
Validated, quality-assured, sensitive and specific gonococcal NAATs are crucial for accurate diagnostics and subsequent appropriate treatment of gonorrhea. However, due to suboptimal specificity of several gonococcal NAATs, in several regions (e.g., Europe and Australia), it is recommended to verify positive samples with another NAAT targeting a different genetic sequence (6, 8–18). This supplementary testing for verification of screening NAAT-positive specimens is particularly important in low-prevalence populations, if the positive predictive value (PPV) does not exceed 90%, and in pharyngeal infection. However, as shown in the present study, the number of false-positive urogenital specimens with the BD ProbeTec GC Qx amplified DNA assay can also be alarmingly high. Accordingly, supplementary testing of N. gonorrhoeae NAAT-positive specimens regardless of sample type can be essential, particularly in low-prevalence populations. The BD Max GC real-time PCR assay showed high clinical and analytical sensitivity and specificity, with only one cross-reacting N. cinerea strain, which was also false positive in the BD ProbeTec GC Qx amplified DNA assay. The performance of the BD Max GC real-time PCR was also simple and rapid (2 h and 45 min for automated DNA extraction and real-time PCR) with minimal hands-on time (20 to 30 min per 24 samples), including all appropriate controls (positive, negative, and the sample processing control).
In conclusion, the new BD Max GC real-time PCR assay showed high clinical and analytical sensitivity and specificity. However, the BD ProbeTec GC Qx amplified DNA assay had suboptimal specificity for both urogenital and extragenital clinical specimens. Accordingly, the new BD Max GC real-time PCR assay can be exceedingly valuable as a supplementary test for the BD ProbeTec GC Qx amplified DNA assay, particularly in low-prevalence populations, and might also be used for initial detection of N. gonorrhoeae. However, additional studies are needed to prove that the BD Max GC real-time PCR assay is sufficiently sensitive and specific for initial detection of N. gonorrhoeae and to investigate if the assay can be used for verification of specimens positive on NAAT platforms from other manufacturers. This study further highlights the need for supplementary testing when NAATs are used for detection of N. gonorrhoeae.
ACKNOWLEDGMENTS
We are very grateful to BD Diagnostics for providing the BD Max GC real-time PCR tests.
The present work was funded by grants from the Örebro County Council Research Committee and the Foundation for Medical Research at Örebro University Hospital, Sweden.
REFERENCES
- 1.World Health Organization. 2012. Global incidence and prevalence of selected curable sexually transmitted infections—2008. World Health Organization, Geneva, Switzerland: http://www.who.int/reproductivehealth/publications/rtis/2008_STI_estimates.pdf. [Google Scholar]
- 2.Bachmann LH, Johnson RE, Cheng H, Markowitz L, Papp JR, Palella FJ Jr, Hook EW III. 2010. Nucleic acid amplification tests for diagnosis of Neisseria gonorrhoeae and Chlamydia trachomatis rectal infections. J Clin Microbiol 48:1827–1832. doi: 10.1128/JCM.02398-09. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Schachter J, Moncada J, Liska S, Shayevich C, Klausner JD. 2008. Nucleic acid amplification tests in the diagnosis of chlamydial and gonococcal infections of the oropharynx and rectum in men who have sex with men. Sex Transm Dis 35:637–642. doi: 10.1097/OLQ.0b013e31817bdd7e. [DOI] [PubMed] [Google Scholar]
- 4.Alexander S. 2009. The challenges of detecting gonorrhoea and chlamydia in rectal and pharyngeal sites: could we, should we, be doing more? Sex Transm Infect 85:159-160. doi: 10.1136/sti.2008.035469. [DOI] [PubMed] [Google Scholar]
- 5.Hook EW III, Handsfield HH. 2008. Gonococcal infections in the adult, p 627–646. In Holmes KK, Sparling PF, Stamm WE, Piot P, Wasserheit JN, Corey L, Cohen MS, Watts DH (ed), Sexually transmitted diseases, 4th ed McGraw-Hill, New York, NY. [Google Scholar]
- 6.Unemo M, Ison C. 2013. Gonorrhoea, p 21–55. In Unemo M, Ballard R, Ison C, Lewis D, Ndowa F, Peeling R (ed), Laboratory diagnosis of sexually transmitted infections, including human immunodeficiency virus. WHO Document Production Services, Geneva, Switzerland. [Google Scholar]
- 7.Dize L, Agreda P, Quinn N, Barnes MR, Hsieh YH, Gaydos CA. 2012. Comparison of self-obtained penile-meatal swabs to urine for the detection of C. trachomatis, N. gonorrhoeae and T vaginalis. Sex Transm Infect 89:305–307. doi: 10.1136/sextrans-2012-050686. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Tabrizi SN, Unemo M, Limnios AE, Hogan TR, Hjelmevoll SO, Garland SM, Tapsall J. 2011. Evaluation of six commercial nucleic acid amplification tests for detection of Neisseria gonorrhoeae and other Neisseria species. J Clin Microbiol 49:3610–3615. doi: 10.1128/JCM.01217-11. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Bignell C, Unemo M, European Guidelines Editorial Board STI . 2013. 2012 European guideline on the diagnosis and treatment of gonorrhoea in adults. Int J STD AIDS 24:85–92. doi: 10.1177/0956462412472837. [DOI] [PubMed] [Google Scholar]
- 10.Golparian D, Johansson E, Unemo M. 2012. Clinical Neisseria gonorrhoeae isolate with a N. meningitidis porA gene and no prolyliminopeptidase activity, Sweden, 2011: danger of false negative genetic and culture diagnostic results. Euro Surveill 17:20102. [PubMed] [Google Scholar]
- 11.Whiley DM, Limnios A, Moon NJ, Gehrig N, Goire N, Hogan T, Lam A, Jacob K, Lambert SB, Nissen MD, Sloots TP. 2011. False-negative results using Neisseria gonorrhoeae porA pseudogene PCR—a clinical gonococcal isolate with an N. meningitidis porA sequence, Australia, March 2011. Euro Surveill 16:19874. [PubMed] [Google Scholar]
- 12.Ison CA, Golparian D, Saunders P, Chisholm S, Unemo M. 2013. Evolution of Neisseria gonorrhoeae is a continuing challenge for molecular detection of gonorrhoea: false negative gonococcal porA mutants are spreading internationally. Sex Transm Infect 89:197–201. doi: 10.1136/sextrans-2012-050829. [DOI] [PubMed] [Google Scholar]
- 13.Palmer HM, Mallinson H, Wood RL, Herring AJ. 2003. Evaluation of the specificities of five DNA amplification methods for the detection of Neisseria gonorrhoeae. J Clin Microbiol 41:835–837. doi: 10.1128/JCM.41.2.835-837.2003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Chow EPF, Fehler G, Read TRH, Tabrizi SN, Hocking JS, Denham I, Bradshaw CS, Chen MY, Fairley CK. 2015. Gonorrhoea notifications and nucleic acid amplification testing in a very low-prevalence Australian female population. Med J Aust 202:321–323. doi: 10.5694/mja14.00780. [DOI] [PubMed] [Google Scholar]
- 15.Smith DW, Tapsall JW, Lum G. 2005. Guidelines for the use and interpretation of nucleic acid detection tests for Neisseria gonorrhoeae in Australia: a position paper on behalf of the Public Health Laboratory Network. Commun Dis Intell Q Rep 29:358–365. [PubMed] [Google Scholar]
- 16.Bignell C, Fitzgerald M, Guideline Development Group, British Association for Sexual Health and HIV UK . 2011. UK national guideline for the management of gonorrhoea in adults, 2011. Int J STD AIDS 22:541–547. doi: 10.1258/ijsa.2011.011267. [DOI] [PubMed] [Google Scholar]
- 17.Katz AR, Effler PV, Ohye RG, Brouillet B, Lee MV, Whiticar PM. 2004. False positive gonorrhea test results with a nucleic acid amplification test: the impact of low prevalence on positive predictive value. Clin Infect Dis 38:814–819. [DOI] [PubMed] [Google Scholar]
- 18.Fifer H, Ison CA. 2014. Nucleic acid amplification tests for the diagnosis of Neisseria gonorrhoeae in low-prevalence settings: a review of the evidence. Sex Transm Infect 90:577–579. doi: 10.1136/sextrans-2014-051588. [DOI] [PubMed] [Google Scholar]
- 19.Taylor SN, Van Der Pol B, Lillis R, Hook EW III, Lebar W, Davis T, Fuller D, Mena L, Fine P, Gaydos CA, Martin DH. 2011. Clinical evaluation of the BD ProbeTec Chlamydia trachomatis Qx amplified DNA assay on the BD Viper system with XTR technology. Sex Transm Dis 38:603–609. doi: 10.1097/OLQ.0b013e31820a94d2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Pope CF, Hay P, Alexander S, Capaldi K, Dave J, Sadiq ST, Ison CA, Planche T. 2010. Positive predictive value of the Becton Dickinson VIPER system and the ProbeTec GC Qx assay, in extracted mode, for detection of Neisseria gonorrhoeae. Sex Transm Infect 86:465–469. doi: 10.1136/sti.2010.044065. [DOI] [PubMed] [Google Scholar]
- 21.Goire N, Nissen MD, LeCornec GM, Sloots TP, Whiley DM. 2008. A duplex Neisseria gonorrhoeae real-time polymerase chain reaction assay targeting the gonococcal porA pseudogene and multicopy opa genes. Diagn Microbiol Infect Dis 61:6–12. doi: 10.1016/j.diagmicrobio.2007.12.007. [DOI] [PubMed] [Google Scholar]
- 22.Golparian D, Tabrizi SN, Unemo M. 2013. Analytical specificity and sensitivity of the APTIMA Combo 2 and APTIMA GC assays for detection of commensal Neisseria species and Neisseria gonorrhoeae on the Gen-Probe Panther instrument. Sex Transm Dis 40:175–178. doi: 10.1097/OLQ.0b013e3182787e45. [DOI] [PubMed] [Google Scholar]
- 23.Knapp JS, Totten PA, Mulks MH, Minshew BH. 1984. Characterization of Neisseria cinerea, a nonpathogenic species isolated on Martin-Lewis medium selective for pathogenic Neisseria spp. J Clin Microbiol 19:63–67. [DOI] [PMC free article] [PubMed] [Google Scholar]