Abstract
A solution containing 68Ge in equilibrium with its daughter, 68Ga, has been standardized for the first time at the National Institute of Standards and Technology (NIST) using 3 liquid scintillation-based techniques: live-timed 4πβ -γ anticoincidence (LTAC) counting, the Triple-to-Double Coincidence Ratio (TDCR) method, and 3H-standard efficiency tracing with the CIEMAT1/NIST (CNET) method. The LTAC technique is much less dependent on level scheme data and model-dependent parameters and was thus able to provide a reference activity concentration value for the master solution with a combined standard uncertainty of about 0.3 %. The other two methods gave activity concentration values with respective differences from the reference value of +1.2 % and −1.5 %, which were still within the experimental uncertainties. Measurements made on the NIST “4π”γ secondary standard ionization chamber allowed for the determination of calibration factors for that instrument, allowing future calibrations to be made for 68Ge/68Ga without the need for a primary measurement. The ability to produce standardized solutions of 68Ge presents opportunities for the development of a number of NIST-traceable calibration sources with very low (<1 %) relative standard uncertainties that can be used in diagnostic medical imaging.
Keywords: anticoincidence counting, CIEMAT/NIST method, germanium-68, liquid scintillation counting, positron emitter, standardization, TDCR method
1. Introduction
The use of Positron Emission Tomography (PET) as a tool for diagnosing diseases, particularly cancer, continues to rise at a rate of about 20 % per year [1], partially due to technological advances that allow for more quantitative data to be collected. The ability to consistently acquire truly quantitative imaging data depends on the use of radioactivity standards traceable to the National Institute of Standards and Technology (NIST).
Currently the most widely used radionuclide in PET imaging is 18F. Although NIST has previously standardized 18F [2,3], its short half-life (1.83 h) makes it nearly impossible to prepare and distribute Standard Reference Materials to most users. As a potential solution, 68Ge, in equilibrium with its daughter, 68Ga, has been proposed as a surrogate.
A simplified scheme for the decays of 68Ge and its 68Ga daughter is shown in Fig. 1. Germanium-68 decays by pure electron capture (EC) to the ground state of 68Ga with a half-life of 270.95(16) d [4]. Gallium-68 in turn decays with a half-life of 67.71(9) min by a combination of EC and positron emission primarily to the ground state of 68Zn, but also with a branch to an excited state at 1077 keV with a probability of about 3 % and a number of higher excited states with a combined probability of under 0.4 %.
The nature of the decay scheme of 68Ge/68Ga makes it amenable to a variety of different standardization techniques. In 1994, Schönfeld, et al. [5] reported on the results of measurements made with liquid scintillation (LS) counting using the CIEMAT/NIST 3H standard efficiency tracing method (CNET) [6,7], 4πβ - γ coincidence, and a calibrated ionization chamber (IC). The data showed good agreement between all three methods, giving activity concentration values within the respective experimental uncertainties (nominally 1 % relative standard uncertainty).
More recently, Grigorescu, et al. [8], reported on the results of measurements using 4πβ - γ coincidence counting. As with Schönfeld, et al., the coincidence spectrometer consisted of a proportional counter and NaI(Tl) detector for the β and γ detection channels, respectively. Because this experimental arrangement requires the use of dried sources, corrections for the loss of 68Ge due to chemical volatility were necessary in both studies. This effect is reported by Grigorescu to be on the order of 20 % to 26 %. Nonetheless, they were able to obtain a measurement result with about a 1 % relative standard uncertainty.
Liquid scintillation (LS) counting has been the method of choice in our laboratory for the measurement of β-emitting radionuclides, primarily due to the high LS detection efficiency and the relative ease of sample preparation. Methods based on LS counting have another advantage in the context of measuring 68Ge because the sample is introduced into the cocktail while still in solution, thereby eliminating the need to prepare dry sources. Seeking to take advantage of this, we have measured a single solution of 68Ge/68Ga using three LS-based methods: live-timed 4πβ - γ anticoincidence (LTAC) using LS as the β counting channel, the Triple-to-Double Coincidence Ratio Method (TDCR) [9,10], and CNET [6,7].
2. Experimental
All evaluation of measurement uncertainties throughout this work follow accepted conventions used by the NIST Radioactivity Group and are in accordance with those recommended by the principal metrology organizations [11]. All individual uncertainty components are given as estimated experimental standard deviations (or standard deviations of the mean, if appropriate), or quantities assumed to correspond to standard deviations regardless of the method used to evaluate their magnitude. Unless explicitly stated, all uncertainties cited in this paper are “standard uncertainties,” corresponding to one uncertainty interval. One particular exception is the uncertainty reported for the activity concentration of the calibrated 68Ge solution, which is given as an “expanded combined standard uncertainty.” In accordance with NIST policy [12], the combined standard uncertainty (calculated by combining the individual uncertainty components in quadrature) is multiplied by a “coverage factor” of k = 2 to obtain an “expanded uncertainty” assumed to give an uncertainty interval having a confidence level of 90 % to 95 %.
2.1 Initial Solution Preparation
The master solution used in these experiments contained nominally 125 MBq 68Ge in 5 mL of 0.5 mol · L−1 HCl and was prepared by International Isotopes Idaho, Inc. (Idaho Falls, ID)2 using 68Ge produced at the 100 MeV Isotope Production Facility at Los Alamos National Laboratory using the nat.Ga(p,2n)68Ge reaction.
A generalized scheme for the preparation of the counting samples is shown in Fig. 2. The first step involved transfer of the master solution out of the shipping vial into a NIST standard 5 mL flame-sealed ampoule while at the same time performing the first of three serial dilutions that would be needed in order to bring the activity level in one of the ampoules down to that suitable for LS counting. The ampoule used for this study, labeled A1, was prepared by volumetrically adding 1 mL of the stock solution to an ampoule containing 4 g of gravimetrically added carrier solution having nominally 45 μg each of nonradioactive Ge+4 and Ga+3 per gram of solution using 0.5 mol · L−1 HCl as the solvent. The ampoule was weighed again after the addition of the 68Ge to determine the mass of added radioactive solution. Ampoule A2, shown in Fig. 2, was held in reserve for future experiments.
Ampoule A1 was measured in the NIST-maintained radionuclide activity calibrators (“dose calibrators”) and the NIST “4π”γ IC [13] to allow for the determination of calibration factors in this specific measurement geometry. The solution in A1 was then diluted by a factor of about 200 through two serial gravimetric dilutions, giving two additional ampoules, A1D1 and A1D2. As an additional check of the dilution factor between A1 and A1D1, the latter was also measured in the NIST IC.
2.2 Liquid Scintillation Source Preparation
All counting sources for the three counting techniques were prepared using solution A1D2. A total of 18 LS cocktails containing 68Ge were prepared for these studies. For the LTAC experiments, 4 mL of HiSafe-3 (Perkin Elmer, Waltham, MA) or PCS (GE Healthcare Biosciences, Piscataway, NJ) scintillant were added to each of two 3 cm diameter, glass pseudo-hemispheres. Nominally 0.04 g of 68Ge solution were gravimetrically added and the hemispheres were sealed using epoxy.
For the TDCR experiments, two cocktails each of HiSafe-3 and PCS were prepared by dispensing 10 mL of the scintillant into four 22 mL borosilicate glass LS vials (two per scintillant), followed by the gravimetric addition of nominally 0.04 g of solution from A1D2. Similarly, 10 cocktails were prepared for the CNET experiments (five vials per scintillant). To vary the counting efficiency of the CNET cocktails, between 2 drops and 18 drops of a 10:1 (by volume) dilution of nitromethane in ethanol were added as a quenching agent to the CNET cocktails in addition to the scintillant and radioactive solution. In order to perform the efficiency tracing, a separate set of 10 LS vials having identical composition to the 68Ge cocktails were prepared using a dilution of a NIST tritiated water Standard Reference Material 4927F [14] in place of the 68Ge. In order to make the 68Ge and 3H cocktails as close in composition as possible, nominally 1 mL of the Ge4+/Ga3+ carrier was added to each of the cocktails.
Two background blanks (one for each scintillant) were prepared for the TDCR and LTAC measurements in their appropriate vials. In order to properly mimic the composition of the radioactive cocktails, an equivalent mass of nonradioactive carrier solution was added to each blank. For the CNET measurements, four blanks were prepared so as to have the identical sample compositions of the least- and most-quenched of the radioactive 68Ge cocktails. As with the TDCR and LTAC blanks, nonradioactive Ge/Ga carrier was substituted for the 68Ge solution.
2.3 4πβ - γ Anticoincidence Counting (LTAC)
The system constructed at NIST uses an LS source optically coupled to an appropriate photomultiplier tube for the beta channel and a thallium-doped sodium iodide [NaI(Tl)] detector for the γ-ray channel, as described previously [15,16]. The LS-based system is well suited to this case since the source solution does not have to be dried, and therefore, the large (20 % to 26 %) correction for Ge loss reported by Grigorescu [8] is avoided.
The four active sources were each measured for between 2 and 4 cycles and the blank sources up to 3 cycles during the period from 24 April to 5 May 2007. Each counting cycle consisted of measurements at between 8 to 12 threshold levels on the LS detector for between 200 and 1000 seconds. A minimum of 5 · 106 LS and 1 · 105 anti-coincident NaI detector counts were recorded for each non-blank measurement. The LS signal-to-background ratio for the lowest threshold (highest background) data points was about 950:1, while the signal-to-background ratio for the NaI detector was about 350:1. Further systematic tests demonstrated that the background variability during the span of the experiment, the variation of extending dead-time, and the presence or absence of the aluminum absorber did not affect the measurement results.
The positron decay of the 68Ga was detected in the LS channel, with count rate Nβ, while electron capture events from both 68Ga and 68Ge were avoided by constraining the lower level discriminator (LLD) on the amplified signal to be above about 20 keV beta energy. In this way, the determined activity value was independent of any atomic transitions (all below 11 keV), and directly proportional to the total positron emission probability. The LS positron efficiency, εβ, was varied between about 0.5 and 0.95 using the LLD and extrapolated to 1.0. The NaI(Tl) detector was gated on the 511 keV region using a single channel analyzer and the total γ-ray (Nγ) and anticoincidence (NAC) count rates were recorded. The extrapolation parameter used was Y ≡ NAC / Nγ ≈ (1 − εβ). Most of the γ-ray counts were due to positron-annihilation decays, detected with efficiency,
(1) |
where N0 is the activity and (b1 + b2) is the total positron emission probability. There was an additional, approximately 0.2 %, contribution from Compton scattering of 1077 keV γ-rays, detected with efficiency εγ1077. Since some of these 1077 keV γ-rays correspond to electron capture events, and not positron emission, a small (0.2 %) correction to the intercept was necessary. The modified extrapolation formula is,
(2) |
where the numerator and denominator of the correction term correspond to total γ-rays and anticoincident γ-rays, respectively, and the branching probabilities b1, b2, and b3 are illustrated and enumerated in Fig. 1. Note that the extrapolation is linear in Y, and the Y = 0 (εβ = 1) intercept is given by,
(3) |
Equation (3) is similar to Eq. (4) in [8], with a significant difference that here we account for the fact that some of the 1077 keV γ-rays do correspond to a positron branch, and thus do not need to be subtracted. This effect is accommodated by the presence of b2 in the numerator of the correction term (final term in Eq. (3)). Corrections due to the LS efficiency for γ-rays and for coincidences due to those events were not necessary, as described below.
An estimate of εγ1077 was obtained during a separate set of measurements with a 60Co point-source and εann was estimated using Eq. (1). The correction term was checked by exaggerating the effect during additional measurements made with various γ-ray energy gates. Gates G1, G2, and G3 corresponded to a gate around the Compton region of the annihilation spectrum, the annihilation photopeak (511 keV) and the annihilation sum peak (1022 keV) respectively. The sum peak was unresolved from the 1077 keV peak in the NaI(Tl) detector events. As designed, the data from gates G1 and G3 required large corrections due to reduced εann and enhanced εγ1077 respectively. The uncorrected and corrected N0 values are shown in Table 1. The fact that the large corrections for G1 and G3 gave consistent results with G2, supports the use of this method for the small (0.2 %) correction to the final N0 value, based on G2 alone.
Table 1.
γ-ray gate | Uncorrected N0 (Bq) | Corrected N0 (Bq) |
---|---|---|
G1 - Comptons | 1.020 ± 0.002 | 1.002 ± 0.004 |
G2 - photopeak | 1.0021 ± 0.001 | 1.0000 ± 0.0004 |
G3 - sum peak | 1.033 ± 0.002 | 1.000 ± 0.007 |
No correction in the final result was made for the gamma efficiency of the LS detector, or for coincidences between such events and the NaI(Tl) detector. For gates G1 and G2, these effects tend to cancel out due to the two-photon annihilation process. If one photon is detected in the LS detector, that efficiency can be monitored by the other photon interacting in the NaI detector [15]. If such an effect were present it would lead to a non-linear efficiency extrapolation. A typical G2 data set and residuals from a linear least-squares fit are shown in Fig. 3 and it is evident that a linear fit is satisfactory. Yet, a quadratic extrapolation was needed to fit the entire range of G3 (sum peak) data due to the unmonitored LS efficiency for γ-rays in that configuration. Thus, for G3, a smaller εβ range (0.9 – 0.95) was employed such that linear and quadratic fits gave consistent results. This value was only used for the εγ1077 sensitivity test.
Another possible cause for a non-linear extrapolation would be if both the LS and NaI(Tl) efficiencies differed depending on whether the positron was stopped in the LS hemisphere, or escaped before annihilating. This effect was mitigated by three factors: (1) the fact that most positrons annihilated within the hemisphere, (2) the high LS efficiency, and (3) the well-type geometry of the NaI(Tl) detector. The sensitivity of the result to this effect was tested by placing an approximately 0.5 cm thick aluminum foil over the hemisphere and comparing the resulting activity determination. No change in the goodness of the linear fit was detected and the ratio of the intercept with to without the foil was 1.000 ± 0.001, where the uncertainty is a standard (k = 1) uncertainty on the linear fit coefficients.
2.4 Liquid Scintillation Counting Using the Triple-to-Double Coincidence Ratio (TDCR) Method
Each counting source was counted in the NIST TDCR system [17] on at least two separate occasions over the course of 27 days. Counting times were typically 1200 s, which allowed for the accumulation of at least 106 counts in each of the three doubles counting channels. For each counting experiment, data were acquired at a minimum of 4 efficiency points, which were varied through the use of a set of grey filters that were fitted over the LS vials. Data were acquired in triplicate at each efficiency point. The experimental efficiencies for the logical sum of double photon coincidence events, εLSD, ranged from 0.89 to 1.14.
The counting data were analyzed using a program developed in-house for use with the Mathematica [18] symbolic mathematics package. Details of the program and the computation strategy will be published separately. However, it should be noted that the program calculates the total detection efficiency for the case of decay of 68Ge in equilibrium with its 68Ga daughter. To do this, the program was required to solve the TDCR equations [9,10,19,20] for the EC branch of the 68Ge parent, as well as both the EC and positron decay of the 68Ga daughter. A relatively simplistic model, considering twelve possible decay pathways, was adopted to describe the atomic transitions encountered in the EC decay of 68Ge and 68Ga. These are depicted in Fig. 4. The values of the various nuclear and atomic input data were taken from the evaluation of the Decay Data Evaluation Project (DDEP) [4].
The analysis program calculates the individual phototube efficiencies, thereby allowing for correction due to asymmetry in the counting rates in each of the-doubles counting channels. The contribution to the detection efficiency due to detection of the 511 keV anihillation photons was taken into account by using the positron spectrum calculated by the program SPEBETA [21] as input for the Monte Carlo simulation package PENELOPE [22] using the techniques described in [23]. The resulting spectrum of energy (positrons+annihilation photons) absorbed in the LS cocktail was then used as input data for the TDCR analysis code.
The stopping power, dE/dx, for electrons in the LS cocktail was calculated by fitting a function of the form
(4) |
(E is the value of the midpoint energy for each bin of the calculated beta spectrum and a, b, c, d, and e are fitting parameters) to data from the NIST ESTAR [24] database using previously published LS cocktail compositions [25].
A separate program, assuming equal phototube efficiencies, was developed for evaluating the effects of varying different input and model parameters. Calculations of εLSD were made as a function of the TDCR for kB values between 0.009 cm · MeV−1 and 0.018 cm · MeV−1 and the resulting εLSD values were found to be insensitive to the value of kB. For consistency with previous measurements made in this laboratory [26], the value of kB for all analyses was taken to be 0.012 cm · MeV−1. A plot of the theoretical εLSD values as a function of TDCR at kB = 0.012 cm · MeV−1 is shown in Fig. 5.
2.5 Liquid Scintillation Counting Using the CIEMAT-NIST 3H-Standard Efficiency Tracing (CNET) Method
Each LS cocktail was sequentially counted for 10 cycles of 25 min per source in a Packard (Perkin Elmer, Waltham, MA) 2500TR LS spectrometer. Samples were then removed from the counter, agitated and sequentially counted for 10 cycles of 30 min per source in a Beckman LS6500 (Beckman Coulter, Fullerton, CA) spectrometer.
Efficiency tracing involves calculating a relationship between the measured 3H LS efficiencies and the LS efficiencies expected for 68Ge, in equilibrium with its daughter 68Ga, over a range of experimental quench indicating parameters [7]. The efficiency tracing computer program CN2004 [27] was used in the analysis of the LS data after changing the default input file to include the nuclear and atomic data found in the DDEP evaluation [4]. The average calculated 68Ge/68Ga efficiency was nominally 138 % in the Packard LS counter and 147 % in the Wallac LS counter using a kB value of 0.0075 cm · Mev−1 and assuming that the cocktail had the composition of Ultima Gold as specified in the default CN2004 input files. Plots of the calculated theoretical 68Ge, 68Ga, and total efficiencies as a function of 3H tritium are given in Fig. 6.
2.6 Ionization Chamber Measurements
For the NIST IC measurements, both A1 and A1D1 were measured 40 times each, in four groups of 10 measurements, alternating with 5 groups of 10 measurements of either radium (226Ra) reference source RRS100 or RRS500b. Results are analyzed as a ratio of the response of the ampoule to the response of the RRS. After correction for background, the resulting ratio is used to derive a calibration factor, or K-value, defined as the activity of a given radionuclide that would produce the same response as the RRS. The relative values of the RRS100 and RRS500b are well characterized. By determining the K-value using the activity derived from different ampoules of different activity levels, it is also possible to verify the gravimetric dilution factor. The dilution factor from A1 to A1D1 was verified by this method to within 0.022 %. The LTAC activity values and the mass dispensed into A1 were used to determine K-values that can be used for future measurements of 68Ge in the NIST ampoule geometry.
2.7 Gamma Ray Spectrometry
The solution that remained in A1D2 after making the LS cocktails was analyzed for possible photon-emitting radionuclidic impurities using two calibrated High-Purity Germanium (HPGe) photon spectrometers at two different counting distances each. In addition, the data provided an additional, confirmatory measurement of the activity concentration using the 1078 keV gamma ray from the decay of 68Ga. Characteristics of the detectors used in this study are given in Table 2.
Table 2.
Detector parameter | X-detector | B-detector |
---|---|---|
Detector diameter | 43.6 ± 0.1 mm | 54.9 ± 0.1 mm |
Detector length | 36.2 ± 0.1 mm | 54.2 ± 0.05 mm |
End cap window material | Beryllium | Beryllium |
Window thickness | 0.5 ± 0.05 mm | 0.5 ± 0.05 mm |
Crystal-window distance | 3 ± 0.5 mm | 3 ± 0.5 mm |
Crystal top dead zone thickness | 0.3 ± 0.03 μm | 0.3 ± 0.03 μm |
Crystal material | Germanium | Germanium |
Crystal hole depth | 32.6 ± 0.5 mm | 47.2 ± 0.5 mm |
Crystal hole diameter | 10.4 ± 0.5 mm | 12 ± 0.5 mm |
Detector side cap thickness | 1.3 ± 0.1 mm | 1.3 ± 0.1 mm |
Detector side cap diameter | 70 ± 1 mm | 63.5 ± 0.5 mm |
Detector side cap material | Aluminum | Magnesium |
Detector type | n-type | n-type |
Calibration Geometries (distances are source-to-detector) | Ampoule; side-mount, end-on 24 cm | Ampoule; 24 cm, 35 cm |
Data were collected using the GammaVision-32 (Ortec, Oak Ridge, TN) software package and analyzed using both GammaVision-32 and Genie 2000 (Canberra, Meriden, CT). Detection efficiencies were calculated from efficiency-energy relationships determined using solutions previously calibrated at NIST and measured in the 5 mL NIST ampoule geometry.
3. Results and Discussion
3.1 Impurity Analyses
No photon-emitting radionuclidic impurities were detected in solution A1 to within the following limits (at the reference time) of the massic photon emission rate:
where E is the gamma-ray energy.
3.2 Activity Measurements Results
The results of the massic activity determinations for the solution contained in A1 as of the reference time are given in Table 3. The values in Table 3 take into account the dilution factor of 204.903231 between the solution in A1 and that used in the assays, A1D2. The uncertainties given in the table are expanded (k = 2) uncertainties based on the components given in Tables 4–7.
Table 3.
Technique | CA, 106 Bq · g−1 |
---|---|
4πβ - γ anticoincidence counting (LTAC) | 3.104(18) |
LS counting with the Triple-to-Double Ratio (TDCR) method | 3.141(25) |
LS counting with the CIEMAT/NIST 3H-standard efficiencytracing method (CNET) | 3.058(44) |
Gamma-ray spectrophotometry with High Purity Germanium (HPGe) detectors | 3.2(9) |
Table 4.
Component, ui | Comment | Evaluation type | % |
---|---|---|---|
Measurement variability | Standard deviation of the mean on the determination of CA for 8 trials encompassing 4 samples, 3 backgrounds, 2 extending dead-times and absorber/no absorber- over a 2-week period | A | 0.03 |
Background variability | Additional variation estimate based on 3 trials and within-run variation | A | 0.03 |
Additional variability in background | Additional variability not embodied by “random sources” | B | 0.05 |
Half-life | Standard uncertainty in half-life. (0.059 %) over the measurement decay interval | B | 0.0008 |
Livetime | Estimated standard uncertainty on CA due to uncertainty in counting livetime. | B | 0.1 |
Branching ratio | Estimated standard uncertainty due to uncertainty in published decay branching ratios | B | 0.13 |
Extrapolation | Estimated standard uncertainty due to extrapolation to zero non-detection efficiency; based on sensitivity tests, previous measurements and models | B | 0.2 |
Correction due to detection of 1077 keV photons | Estimated standard uncertainty on CA due to efficiency of detecting 1077 keV photons based on sensitivity tests | B | 0.1 |
Mass determinations | Estimated standard uncertainty of mass for any single LS cocktail | B | 0.05 |
Dilution factor | Uncertainty in CA of solution Ge1A1 due to uncertainty in gravimetrically-determined dilution factor between solutions in ampoules Ge1A1 and Ge1A1D2 | B | 0.04 |
| |||
Combined | 0.29 | ||
| |||
Expanded (Uc = uc · k; k = 2) | 0.58 |
Table 5.
Component, ui | Comment | Evaluation type | % |
---|---|---|---|
Sample repeatability | Standard deviation of the mean on the determination of massic activity for a single LS cocktail (n = 3 – 5 determinations of CA per source) | A | 0.06 |
LS cocktail composition variability | Standard deviation on the determination of CA for three LS cocktail compositions (n = 6 – 19 determinations of CA per composition) | A | 0.09 |
Efficiency dependence | Median difference between maximum and minimum value of CA determined for a single source at between 3 and 4 efficiency values, varied by use of grey filters (n = 12 independent measurements) | B | 0.20 |
Effect of 68Ga beta endpoint energy, Eβ,max on efficiency calculations | Standard uncertainty in efficiency calculation due to standard uncertainties on positron endpoint energies of 68Ga | B | 0.18 |
Effect of other atomic and nuclear input data | Standard uncertainty due to uncertainties on data used as input to the TDCR analysis code as determined by Monte Carlo methods. A total of 20 data sets were generated from normal distributions defined by the published nuclear and atomic data and their associated standard uncertainties, which were taken as the standard deviation of the respective distributions. Each data set was used to calculate CA using a single experimental data set | A | 0.26 |
Half-life | Standard uncertainty in half-life (0.059 %) over the measurement decay interval | B | 6·10−3 |
Mass determinations | Estimated standard uncertainty of mass for any single LS cocktail | B | 0.05 |
Livetime | Standard uncertainty arising from an estimated uncertainty of 0.007 % on the determination of the live time | B | 7 · 10−3 |
Background | Standard deviation on the determination of CA determined via Monte Carlo simulation. A total of 5 background data sets were constructed from random data arising from normal distributions defined by the average and standard deviation of experimental backgrounds at 4 efficiency points having 3 repetitions each; calculations were carried out with all 5 background data sets for a single experimental data set | A | 5 · 10−3 |
Dilution factor | Uncertainty in CA of solution Ge1A1 due to uncertainty in gravimetrically-determined dilution factor between solutions in ampoules Ge1A1 and Ge1A1D2 | B | 0.04 |
| |||
Combined | 0.39 | ||
| |||
Expanded (Uc = uc · k; k = 2) | 0.79 |
Table 6.
Component, ui | Comment | Evaluation type | % |
---|---|---|---|
Sample repeatability | Standard deviation of the mean on the determination of massic activity for a single LS cocktail (n = 10 determinations of CA per source) | A | 0.05 |
LS measurement reproducibility | Standard deviation on the determination of CA for 10 cocktails of 2 compositions | A | 0.24 |
Mass determinations | Estimated standard uncertainty of 68Ge mass for any single LS cocktail | B | 0.05 |
Dilution factor | Uncertainty in CA of solution Ge1A1 due to uncertainty in gravimetrically-determined dilution factor between solutions in ampoules Ge1A1 and Ge1A1D2 | B | 0.04 |
68Ge decay corrections | Standard uncertainty in half-life (0.059 %) over the measurement decay interval | B | 0.001 |
68Ge efficiency | Estimated uncertainty in CA due to step size in CN2004 calculations | B | 0.65 |
Livetime determinations | Estimated uncertainty in the correction to the LS counting interval | B | 0.05 (and PE)3 |
Background | Estimated uncertainty due to an average 4 % uncertainty in background determination | B | 0.004 |
Activity of 3H standard | Estimated uncertainty due to 0.36 % uncertainty in 3H standard activity | B | 0.18 |
Branching ratios | Estimated uncertainty due to uncertainty in branching ratios | B | 0.08 |
| |||
Combined | 0.73 | ||
| |||
Expanded (Uc = uc · k; k = 2) | 1.45 |
The relative uncertainty for this component is partially embodied (PE) in the relative standard uncertainties of the repeatability and reproducibility components.
Table 7.
Component, ui | Comment | Evaluation type | % |
---|---|---|---|
Measurement repeatability | Standard deviation on determination of CA for 3 repeated measurements of a single source at a single geometry | A | 0.98 |
Efficiency curve | Standard deviation of the mean on determination of detection efficiency for 4 sample geometries | B | 0.12 |
Sample geometry | Typical uncertainty due to change of sample geometry (detector and source-to-detector distance) for a single counting source | B | 0.33 |
Decay correction | Standard uncertainty in half-life (0.059 %) over the measurement decay interval | B | 2.6 · 10−3 |
Decay data | Standard uncertainty (0.93 %) on emission probablity of 1078 keV gamma-ray in the decay of 68Ge | B | 0.93 |
Dilution factor | Estimated standard uncertainty in CA of solution Ge1A1 due to uncertainty in gravimetrically-determined dilution factor between solutions in ampoules Ge1A1 and Ge1A1D2 | B | 0.04 |
| |||
Combined | 1.4 | ||
| |||
Expanded (Uc = uc · k; k = 2) | 2.8 |
Of the different techniques used in this study to determine the activity concentration of the 68Ge solution, the LTAC technique is much less dependent on level scheme data and model parameters not directly measured in the experiment. For this particular measurement, the only input parameter significantly impacting the activity calculation that was not directly measured in the experiment was the positron branching ratio. The other branching ratios only contributed to the minor (0.2 %) correction for the leakage of 1077 keV γ−rays into the annihilation γ-ray gate. And even this small correction was checked experimentally by modifying the experimental design to exaggerate the effect and then verifying that the corrected activity agreed with the original value.
On the other hand, our implementations of the TDCR and CNET efficiency tracing methods are unable to separate the positron and EC decay signals and must therefore account for all possible decay paths, including atomic rearrangements following electron capture. From a practical standpoint, a compromise between treating all possible paths and reasonable computation times must be made. While this certainly introduces some small amount of uncertainty, it is not expected that the weak contributions due to paths not considered in Fig. 4 would be significant, at least for the TDCR method. Instead, as seen in Table 5, the uncertainties on the input data play a very significant role.
Because of the more direct nature of the measurement in the LTAC technique, the LTAC activity value for the solution in A1 was adopted as the reference value for this study and was used in the calculation of the K-value for the NIST IC. The fact that the LTAC and TDCR measurements agree to within their respective experimental uncertainties is encouraging, given the complexity of the TDCR efficiency calculation. Nonetheless, one would hope that improvements in the NIST TDCR spectrometer would lead to higher EC detection efficiencies, thereby providing better results in the measurement of radionuclides that decay by this mode. The CNET results indicate that some improvements in the method are still needed to be able to reliably measure nuclides that undergo EC decay.
3.3 Determination of K-Value for NIST IC
In order to avoid the need to perform a primary standardization every time a NIST-calibrated solution of 68Ge/68Ga is required, we determined a calibration factor (K-value) for the NIST IC. This K-value is not to be confused with the coverage factor, k, applied to uncertainty evaluations. Using the LTAC reference activity value and the measured responses in the IC against radium reference sources (RR) 500B and 100, the K-values were found to be 2.695 × 107 ± 1.7 × 105 Bq and 5.032 × 106 ± 3.2 × 104 Bq, respectively. The uncertainties on the K-values are expanded (k = 2) uncertainty and include relative standard uncertainty components due to the original primary standardization (0.29 %), repeatability on 40 measurements in the IC (0.015 %), source mass (0.05 %), decay correction (0.002 %), and source positioning (0.1 %).
4. Conclusion
A solution containing 68Ge in equilibrium with its decay daughter 68Ga has been standardized for the first time at NIST, with a combined standard uncertainty of 0.29 % using LTAC. Measurements made with two other LS techniques, TDCR and CNET, confirmed the LTAC result to within 1.1 % and 1.5 %, respectively. The differences between results obtained with the latter two methods and the LTAC technique indicate that improvements in the models and/or their applications are needed, particularly for EC nuclides.
Data collected on the NIST 4πγ ionization chamber allowed for the determination of calibration factors for that chamber in the 5 mL NIST ampoule geometry, thereby enabling future calibrations of solutions having the same solution composition without the need for the measurements to be made by a primary method.
Acknowledgments
The authors wish to thank Ms. Michelle Hammond and Dr. Leticia Pibida for performing the HPGe measurements.
A portion of this work was supported by RadQual, LLC.
Biography
About the authors: Dr. B. E. Zimmerman is a research chemist in the Radioactivity Group (Ionizing Radiation Division) of the NIST Physics Laboratory. Mr. J. T. Cessna and Dr. R. Fitzgerald are physicists in NIST Radioactivity Group. The National Institute of Standards and Technology is an agency of the U.S. Department of Commerce.
Footnotes
CIEMAT is an acronym for Centro de Investigaciones Energéticas, Medioambientales y Tecnológicas, which is the National Metrology Institute of Spain.
Certain commercial equipment, instruments, or materials are identified in this paper to foster understanding. Such identification does not imply recommendation or endorsement by the National Institute of Standards and Technology, nor does it imply that the materials or equipment identified are necessarily the best available for the purpose.
Contributor Information
B. E. Zimmerman, Email: bez@nist.gov.
J. T. Cessna, Email: jeffrey.cessna@nist.gov.
R. Fitzgerald, Email: ryan.fitzgerald@nist.gov.
5. References
- 1.Wiley G. PET/CT: Market & Business Models for Ownership. Imaging Economics. 2006;6 [Google Scholar]
- 2.Coursey BM, Hoppes DD, Unterweger MP, GrauMalonda A, Margolin RA, Kesslerr RM, Manning R. Standardization of 18F for use in Positron-Emission Tomography. Int J Appl Radiat Isot. 1983;34:1181–1189. [Google Scholar]
- 3.Zimmerman BE, Kubicek GJ, Cessna JT, Plascjak PS, Eckelman WC. Radioassays and experimental evaluation of dose calibrator settings for 18FDG. Appl Radiat Isot. 2001;54:113–122. doi: 10.1016/s0969-8043(99)00260-2. [DOI] [PubMed] [Google Scholar]
- 4.DDEP Decay Data Evaluation Project Data. 2008 http://www.nucleide.org/DDEP_WG/DDEPdata.htm.
- 5.Schönfeld E, Schötzig U, Günther E, Schrader H. Standardization and decay data of 68Ge/68Ga. Appl Radiat Isot. 1994;45:955–961. doi: 10.1016/s0969-8043(00)00362-6. [DOI] [PubMed] [Google Scholar]
- 6.Coursey BM, Mann WB, GrauMalonda A, Garcìa-Toraño E, Los Arcos JM, Gibson JAB, Reher D. Standardization of carbon-14 by 4πβ liquid scintillation efficiency tracing with hydrogen-3. Appl Radiat Isot. 1986;5:403–408. [Google Scholar]
- 7.Zimmerman BE, Collé R. Standardization of 63Ni by 4πβ liquid scintillation spectrometry with 3H-standard efficiency tracing. J Res Natl Inst Stand Technol. 1997;102:455–477. doi: 10.6028/jres.102.031. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Grigorescu EL, Negut CD, Luca A, Razdolescu AC, Tanase M. Standardization of 68(Ge + Ga) Appl Radiat Isot. 2004;60:429–432. doi: 10.1016/j.apradiso.2003.11.054. [DOI] [PubMed] [Google Scholar]
- 9.Broda R, Pochwalski K. The enhanced triple to double coincidence ratio (ETDCR) method for standardization of radionuclides by liquid scintillation counting. Nucl Inst Meth Phys Res. 1992;A312:85–89. [Google Scholar]
- 10.Pochwalski K, Broda R, Radoszewski T. Standardization of pure beta emitters by liquid scintillation counting. Appl Radiat Isot. 1988;39:165–172. [Google Scholar]
- 11.International Organization for Standardization (ISO) Guide to the Expression of Uncertainty in Measurement. ISO; Geneva: 1993. [Google Scholar]
- 12.Taylor BN, Kuyatt CE. Guidelines for evaluating and expressing the uncertainty of NIST measurement results. National Institute of Standards and Technology; Gaithersburg, MD: 1994. (NIST Technical Note 1297). [Google Scholar]
- 13.Calhoun JM. NBS Special Publication 250–10. National Institute of Standards and Technology; Gaithersburg, MD: 1987. NBS Special Publication 250-10: Radioactivity calibrations with the ‘4π’ gamma ionization chamber and other radioactivity calibration capabilities. [Google Scholar]
- 14.National Institute of Standards and Technology . Standard Reference Material 4927F, Radioactivity Standard, Hydrogen-3. National Institute of Standards and Technology; Gaithersburg, MD: 2000. [Google Scholar]
- 15.Fitzgerald R, Schultz MK. Liquid scintillation-based anticoincidence counting of Co-60 and Pb-210. Appl Radiat Isot. 2008;66:937–940. doi: 10.1016/j.apradiso.2008.02.037. [DOI] [PubMed] [Google Scholar]
- 16.Lucas LL. Calibration of the massic activity of a solution of Tc-99. Appl Radit Isot. 1998;49:1061–1064. [Google Scholar]
- 17.Zimmerman BE, Collé R, Cessna JT. Construction and implementation of the NIST triple-to-double coincidence ratio (TDCR) spectrometer. Appl Radiat Isot. 2004;60:433–438. doi: 10.1016/j.apradiso.2003.11.055. [DOI] [PubMed] [Google Scholar]
- 18.Wolfram Research Mathematica, V.4.0. 2001. [Google Scholar]
- 19.Broda R. A review of the triple-to-double coincidence ratio (TDCR) method for standardizing radionuclides. Appl Radiat Isot. 2003;58:585–594. doi: 10.1016/s0969-8043(03)00056-3. [DOI] [PubMed] [Google Scholar]
- 20.Cassette P, Bé MM, Jaubert F, Lépy MC. Measurement of a 103Pd solution using the TDCR method by LSC. Appl Radiat Isot. 2004;60:439–446. doi: 10.1016/j.apradiso.2003.11.056. [DOI] [PubMed] [Google Scholar]
- 21.Cassette P. SPEBETA programme de calcul du spectre en énergie des électrons émis par des radionucléides émetteurs beta. Commissariat à l’Énergie Atomique (CEA); Saclay, France: 1992. (CEA Technical Note). [Google Scholar]
- 22.Sempau J, Acosta E, Fernandez-Varea JM, Salvat F. An algorithm for Monte Carlo simulation of the coupled electron-photon transport. Nucl Inst Meth. 1997;B132:377–390. [Google Scholar]
- 23.Zimmerman BE. Monte Carlo calculations of spectra and interaction probabilities for photons in liquid scintillators for use in the standardization of radionuclides. Appl Radiat Isot. 2006;64:1492–1498. doi: 10.1016/j.apradiso.2006.02.063. [DOI] [PubMed] [Google Scholar]
- 24.National Institute of Standards and Technology Electron Stopping Power and Range Tables. 2008 http://physics.nist.gov/STAR.
- 25.Labortoire National Henri Becquerel LS Sources Data. 2008 http://www.nucleide.org/ICRM_LSC_WG/icrmlsdata.htm.
- 26.Collé R, Zimmerman BE, Cassette P, Laureano-Perez L. 63Ni, its half-life and standardization: revisited. Appl Radiat Isot. 2008;66:60–68. doi: 10.1016/j.apradiso.2007.07.017. [DOI] [PubMed] [Google Scholar]
- 27.Günther E. CIENIST 2004. 2004. private communication. [Google Scholar]