Abstract
Purpose
The aim of this work was to compare and validate various computed tomography (CT) number calibration techniques with respect to cone beam CT (CBCT) dose calculation accuracy.
Methods
CBCT dose calculation accuracy was assessed for pelvic, lung, and head and neck (H&N) treatment sites for two approaches: (1) physics-based scatter correction methods (CBCTr); (2) density override approaches including assigning water density to the entire CBCT (W), assignment of either water or bone density (WB), and assignment of either water or lung density (WL). Methods for CBCT density assignment within a commercially available treatment planning system (RSauto), where CBCT voxels are binned into six density levels, were assessed and validated. Dose-difference maps and dose-volume statistics were used to compare the CBCT dose distributions with the ground truth of a planning CT acquired the same day as the CBCT.
Results
For pelvic cases, all CTN calibration methods resulted in average dose-volume deviations below 1.5 %. RSauto provided larger than average errors for pelvic treatments for patients with large amounts of adipose tissue. For H&N cases, all CTN calibration methods resulted in average dose-volume differences below 1.0 % with CBCTr (0.5 %) and RSauto (0.6 %) performing best. For lung cases, WL and RSauto methods generated dose distributions most similar to the ground truth.
Conclusion
The RSauto density override approach is an attractive option for CTN adjustments for a variety of anatomical sites. RSauto methods were validated, resulting in dose calculations that were consistent with those calculated on diagnostic-quality CT images, for CBCT images acquired of the lung, for patients receiving pelvic RT in cases without excess adipose tissue, and for H&N cases.
Keywords: Cone beam computed tomography; Adaptive radiation therapy, ART; Dose calculation; Hounsfield units; Density
Abstract
Ziel
Ziel dieser Arbeit ist der Vergleich und die Validierung mehrerer CT-Kalibrierungsmethoden zur Dosisberechnung auf der Grundlage von Kegelstrahlcomputertomographie(CBCT)-Aufnahmen.
Methoden
Bei 4 Becken-, 3 Lungen- und 4 Kopf-Hals-Patienten wurde die Genauigkeit der Dosisberechnung auf der Basis von CBCT-Aufnahmen für die folgenden Ansätze untersucht: einerseits Rekonstruktion der CBCT-Aufnahmen mithilfe eines Streukorrekturalgorithmus (CBCTr) und andererseits 3 verschiedene Methoden zur expliziten Zuweisung der Elektronendichten auf Basis des CBCT-Datensatzes (W: Zuordnung von Wasserdichte auf dem gesamten CBCT, WB: Zuordnung von entweder Wasser- oder Knochendichte, WL: Zuordnung von entweder Wasser- oder Lungendichte) sowie der in einem kommerziellen Planungssystem implementierten Methode der automatischen Zuordnung von 6 Dichtestufen auf Basis der CBCT-Hounsfield-Werte (RSauto). Als Grundlage zur Evaluierung der Methoden diente die Dosisverteilung, welche anhand einer am gleichen Tag wie die CBCT aufgenommenen Planungs-CT berechnet wurde. Die Genauigkeit der einzelnen Ansätze wurde anhand von Dosis-Volumen-Statistiken und lokalen Dosisunterschieden beurteilt.
Ergebnisse
Bei den Datensätzen mit Tumoren im Beckenbereich ist die mittlere Dosisabweichung für alle Kalibrierungsmethoden unter 1,5 %, wobei RSauto in einer überdurchschnittlichen Abweichung für Patienten mit einem höheren Anteil von Fettgewebe resultiert. Die mittlere Abweichung für Kopf-Hals-Patienten beträgt unter 1,0 %, wobei CBCTr (0,5 %) und RSauto (0,6 %) am besten abschneiden. WL und RSauto resultieren für die Patienten mit Lungentumoren in Dosisverteilungen, welche der Referenzdosisverteilung entsprechen.
Schlussfolgerung
RSauto zur Kalibrierung von CBCT-Aufnahmen zur Dosisberechnung ist eine aussichtsreiche Methode für die untersuchten Indikationen. Es wurde gezeigt, dass mithilfe von RSauto die Dosisberechnung auf der Basis von CBCT-Aufnahmen von Lungen- und Kopf-Hals-Erkrankungen sowie für Tumoren im Beckenbereich bei Patienten ohne ein Übermaß an Fettgewebe in einer akkuraten Dosisverteilung resultiert.
Schlüsselwörter: Cone-Beam-CT, Adaptive Strahlentherapie, Dosisberechnung, Hounsfield-Einheiten, Dichte
Since its introduction, cone beam CT (CBCT) has been mainly utilized for image-guided radiotherapy (IGRT) [1–4]. CBCT images generated prior to treatment can also be used for adaptive radiotherapy (ART) [5, 6] in various ways, ranging from using the CBCT to estimate the dosimetric effect of patient weight loss, CBCT assessment in order to choose a “plan-of-the-day,” to dose recalculation and treatment plan generation on CBCT images [7–11]. Commercially available treatment planning systems (TPSs) are now able to implement ART strategies.
CBCT image quality is dependent on acquisition parameters and, compared with diagnostic-quality planning CT images, exhibit increased artifacts and poorer image contrast owing to increased scatter, which itself is dependent on the size of the scanned object/patient [12–14]. CBCT images (CBCTs) generated using XVI on Elekta treatment units are not in true CT numbers (CTNs) and, therefore, in their original form, such CBCTs cannot be used directly for accurate dose computation [12]. In order to be able to perform CBCT-based ART, it is important that dose can be calculated accurately (i.e., consistent with that calculated on diagnostic CT images) on the CBCT image; various methods have been suggested for CTN adjustment in order to achieve this [12, 15–21].
Strategies for CTN adjustment include calibration of the CBCT image voxel values with physical density following CBCT acquisitions of phantoms with inserts of differing densities [15–19]. Richter et al. [17] proposed a method based on average CBCT values for separate treatment sites in order to generate population-specific conversion curves for brain, head and neck (H&N), thorax, and prostate treatment sites. Apart from taking a large amount of time to define, conversion curves are subject to errors caused by image artifacts and patient variability, resulting in differences of 5 % and larger between doses calculated on CBCTs compared with CT-based calculation [12, 17]. Patient-specific conversion curves have also been investigated [20]. However, these methods are still prone to dosimetric errors resulting from CBCT artifacts.
Poludniowski et al. have shown that differences, between the doses on CBCT and CT, of less than 2.5 % can be achieved when the CBCT has been reconstructed after scatter correction of the individual projections [20, 22]. The implementation of such an approach can be slow and difficult to introduce into a clinical workflow. Commercially available systems that generate CBCTs in CTNs by using sophisticated scatter-correction algorithms are starting to become commercially available in recent software releases. However, the accuracy of dose calculation on such images has yet to be validated.
Relatively unsophisticated methods, such as density overrides of regions of interest, can be used to populate CBCTs with density values that result in acceptable dose calculation accuracy [12, 20, 21]. Such strategies can be fast to implement and can easily be adapted into clinical workflows. These techniques are not as sensitive to the problems, such as image artifacts, that the more sophisticated methods struggle to cope with.
Unlike other studies presented for density assignment methods [12, 21] and other approaches [15–20, 23], we use radiotherapy planning CTs (PCTCBCT) acquired on the same day as the CBCT as the ground truth for dose calculation. Thus, we have selected patients with minimal anatomical differences between CBCTs and planning CTs, eliminating the need for, and uncertainties associated with, deformable image registration (DIR) [24–27]. Using the original planning CT (PCTorig), which may have been captured weeks before the CBCT, as the ground truth for dose calculation is unsuitable in the context of ART because establishing an estimate of the difference in dose between that calculated on the PCTorig and CBCT is itself a subject of importance. We have investigated three anatomical sites: H&N, pelvis (prostate and bladder), and lung.
The aim of this work was to assess the CBCT dose calculation accuracy for density override approaches and to compare this with the physics-based scatter correction of Poludniowski [20, 22]. We evaluated the accuracy of the RayStation TPS (V3.99, RaySearch Laboratories, Stockholm, Sweden) density override approach and compared it with more simple density override methods investigated in the literature [12]. We aimed to establish, for different anatomical locations, the methods that were best in terms of CBCT dose calculation accuracy. The methods for CTN adjustment investigated in this work are transferable to dose calculation on magnetic resonance (MR) images and could, therefore, be of interest to MR-based ART [28].
Patients and methods
We selected patients who had a CBCT and diagnostic CT (both with the same immobilization and/or support devices) on the same day. The study consisted of four H&N cases, four pelvis cases (two prostate and two bladder treatment sites), and three lung cases (two left upper lobe GTVs, one right upper lobe GTV). The RayStation TPS was used for dose calculations, using a collapsed cone algorithm. The median time difference between the CBCT and PCTCBCT was 59 min (range: 17–215 min).
CBCT dose calculation
The CBCTs were acquired with Elekta XVI-V4.5 (XVI Elekta, Stockholm, Sweden); the scanning parameters are displayed in Table 1. To calculate dose on the CBCT images, an external patient contour was defined. First, an external contour of the CBCT image was defined using the TPS automated algorithm that uses a thresholding technique based on the image data. Small holes were then removed from the threshold-generated volume and finally contours not associated with the largest object were deleted. Next, a field-of-view (FOV) region of interest (ROI) was defined for the CBCT image set (see Table 1 for FOV parameters). Finally, with a priori knowledge of the planning CT scan, an automated algorithm (available in the TPS) updated the external contour of the CBCT (where necessary) to extend beyond the CBCT FOV. A missing tissue ROI (with density = 1 g/cm3) is created and is defined as the newly generated external ROI exterior to the FOV ROI. Figure 1 (top left) displays the FOV, external, and missing tissue ROIs for an H&N CBCT image as a red contour, green contour, and gray color wash, respectively.
Table 1.
Protocol name | KeV | mA | mS | FOV (cm) | Scan length (cm) | Filter | Nominal dose CTDI (mGy) |
---|---|---|---|---|---|---|---|
Chest S20 | 120 | 25 | 40 | 26 | 26 | F0 (none) | 9.00 |
Head and neck S20 | 100 | 10 | 10 | 26 | 26 | F0 (none) | 0.45 |
Pelvis M10 | 120 | 100 | 40 | 40 | 13 | F1 (bowtie) | 27.00 |
CTDI computed tomography dose index.
aThe H&N and lung patients were scanned with a field of view (FOV) of 26 cm allowing a partial gantry rotation for acquisition, but limiting the full patient contour (i.e., missing the shoulders on the H&N patients and only seeing the treated sides of the lung patients), whereas acquisitions for the pelvis patients were made with the 40-cm FOV.
CTN adjustment using CBCT reconstruction
Following Poludniowski [20, 22, 29], CBCT reconstructions (CBCTr) were made after scatter correction. The software developed in-house explicitly simulates and corrects for scatter in the projection images using the patients PCTorig as a prior [20, 22]. A patient-independent look-up table was applied to the reconstructed images in order to convert the reconstructed CTNs into physical density [29].
CTN adjustment using density overrides
The “water only” (W) method, where all tissue was assigned as water (1 g/cm3), was investigated in all three treatment sites. For pelvis and lung cases, “water-and-bone” (WB) and “water-and-lung” (WL) techniques were also investigated, respectively. For WB, the CBCT voxels were assigned as either water or bone. Bone ROIs were generated for each CBCT image using a segmentation algorithm based on gray value thresholds on the PCTorig data and copying the resulting ROIs onto the CBCT image. The bone ROIs were edited if necessary, to ensure they were appropriate with reference to the CBCT image. The PCTorig bone ROIs were additionally used to establish an appropriate density to assign to the CBCT bone ROI. The average CTN of bone ROIs on PCTorig was 1,400, which was similar to previously reported data [30] and equivalent to 1.35 g/cm3. This density was used for all cases included in this study. In the WL technique, the patient was assumed to be made up of either water or lung tissue (0.2 g/cm3), where the density of lung was calculated using similar methods as those used to determine the pelvic bone density.
More sophisticated density override techniques were investigated using the CBCT density-assignment tools available in the TPS where six different set densities (air, lung, adipose tissue, connective tissue, cartilage/bone, and higher density for prosthesis) are assigned to the CBCT image. This is achieved by binning the CBCT image histogram into six density levels. The CBCT image thresholds are individual to each patient whereas the assigned density values are fixed (Fig. 1). An automatic algorithm (RSauto) extracts the various density thresholds by approximating the CBCT image histogram with two normal distributions which were then interrogated to determine the threshold values. The physical densities applied by the TPS to the various tissue types were: air = 0.00121 g/cm3, lung = 0.26 g/cm3, adipose = 0.95 g/cm3, connective tissue = 1.05 g/cm3, cartilage/bone = 1.6 g/cm3, and other (such as prosthetic hip) = 3 g/cm3. The consideration of the other override approaches (W, WL, WB) provides a useful benchmark and understanding for the more complex RSauto methodology. Table 2 outlines the CTN adjustment techniques investigated in this study.
Table 2.
Method name | Method description | Sites used in study |
---|---|---|
CBCTr | CBCT reconstruction after scatter correction Poludniowski | Pelvis, H&N, lung |
W | All tissue was assumed to be water (1 g/cm3) | Pelvis, H&N, lung |
WB | CBCT voxels assigned as either water or bone | Pelvis |
WL | CBCT voxels assigned as either water or lung | Lung |
RSauto | CBCT voxels automatically binned into six density levels | Pelvis, H&N, lung |
CBCT r physics-based scatter correction methods, W assignment of water density to the entire CBCT, WB assignment of either water or bone density, WL assignment of either water or lung density, RS auto RayStation TPS, H&N head and neck.
Limited FOV for lung CBCTs
At our institution, the FOV for CBCT acquisitions for lung patients does not encompass the entire patient contour (Table 1, Fig. 1). When performing CTN adjustment using CBCTr or RSauto, it may, therefore, be necessary to perform an additional density override to the contralateral lung. As well as applying a density override of 1 g/cm3 to the missing tissue ROI (region shaded turquoise in Fig. 1), an additional density override of 0.2 g/cm3 was applied to the left lung outside of the FOV (shaded orange in Fig. 1). If CBCT scans of the lung were to be routinely used for dose calculation, we would recommend using either a medium or large FOV to ensure the entire patient contour at the superior/inferior extent of the tumor was encompassed. This contralateral lung density issue has only a small effect on treatments where all beams enter ipsilaterally. However, for treatments where one (or more) beams entered contralaterally, the lung ROI from the planning CT scan was copied to the CBCT image and a density override was applied (Fig. 1).
Data analysis
Dose statistics and dose-difference maps were generated using the TPS. Dose–volume histogram (DVH) analysis was performed on ROIs that were propagated to the CBCT from the PCTCBCT scan.
Results
Differences in dose-volume statistics between that calculated on the CBCT and the ground truth, for all three treatment sites, are shown in Table 3 and Fig. 2.
Table 3.
CBCTr mean%diff (% range) |
W mean%diff (% range) |
WB mean%diff (% range) |
WL mean%diff (% range) |
RSauto
mean%diff (% range) |
||
---|---|---|---|---|---|---|
Pelvic treatments | ||||||
CTV | Dmedian | 0.8 (0.5–1.2) | 0.3 (− 1.3–1.7) | − 1.0 (− 1.6–− 0.1) | − 1.3 (− 2.6–− 0.1) | |
D95 % | 0.8 (0.0–2.0) | 0.0 (− 2.4–1.4) | − 1.5 (− 2.7–− 0.4) | − 1.5 (− 3.7–− 0.2) | ||
D98 % | 0.8 (0.0–2.0) | − 0.1 (− 2.6–1.4) | − 1.5 (− 2.8–− 0.4) | − 1.7 (− 3.9–− 0.4) | ||
D2 % | 1.2 (0.7–1.5) | 0.7 (− 0.2–1.9) | − 0.7 (− 1.4–− 0.2) | − 0.8 (− 1.2–0.1) | ||
Femoral heads | D50 % | 2.4 (1.6–3.1) | 0.9 (− 1.2–3.9) | − 1.1 (− 1.8–− 0.5) | − 0.7 (− 1.5–− 0.1) | |
Rectum | D50 % | 1.4 (− 0.2–4.2) | 0.2 (− 1.1–1.1) | − 1.0 (− 1.8–0.5) | 0.7 (− 1.3–0.6) | |
Bladder | Dmean | 0.4 (0.0–0.8) | 0.1 (− 0.3–0.4) | − 1.0 (− 1.2–− 0.8) | − 1.5 (− 1.6–− 1.4) | |
Average a | 1.4 (− 0.3–4.2) | 0.0 (− 2.6–3.9) | − 1.0 (− 2.8–0.5) | − 1.1 (− 3.9–2.0) | ||
Absolute average a | 1.4 (0.0–4.2) | 0.7 (0.0–3.9) | 1.0 (0.1–2.8) | 1.2 (0.0–3.9) | ||
Head and neck treatments | ||||||
CTV | Dmedian | 0.3 (− 0.1–1.0) | 0.4 (0.0–1.3) | − 0.1 (− 0.6–0.1) | ||
D95 % | 0.5 (0.1–0.9) | 1.9 (0.2–4.7) | 0.2 (− 0.3–1.0) | |||
D98 % | 0.4 (0.0–0.9) | 0.4 (0.0–0.8) | − 0.1 (− 0.3–0.0) | |||
D2 % | 0.3 (− 0.2–1.0) | 0.8 (0.1–1.4) | − 0.1 (− 0.4–0.2) | |||
Spinal cord | D2 % | 0.7 (0.0–1.7) | 1.4 (1.0–2.5) | 0.5 (− 0.1–1.0) | ||
Brain stem | D2 % | 0.5 (− 0.6–1.4) | 0.5 (− 1.7–1.6) | − 0.5 (− 2.5–0.6) | ||
Parotids | Dmean | 0.5 (0.0–1.1) | 0.4 (− 0.5–1.4) | − 0.3 (1.5 − 0.4) | ||
Average b | 0.4 (− 0.6–1.7) | 0.8 (− 1.7–4.7) | − 0.1 (− 2.5–1.0) | |||
Absolute average b | 0.5 (0.0–1.7) | 1.0 (0.0–4.7) | 0.6 (0.1–2.5) | |||
Lung treatments | ||||||
CTV | Dmedian | 2.4 (1.9–2.9) | − 6.8 (− 7.7–− 6.1) | 0.4 (− 0.6–0.9) | − 1.3 (− 2.1–− 0.8) | |
D95 % | 0.0 (− 1.7–1.7) | − 7.0 (− 7.4–− 6.5) | 0.5 (− 2.0–1.8) | − 1.1 (− 1.7–− 0.7) | ||
D98 % | − 1.3 (− 4.8–2.2) | − 6.9 (− 7.1–− 6.6) | 0.3 (− 2.7–2.2) | − 1.0 (− 1.5–− 0.7) | ||
D2 % | 3.0 (1.9–4.1) | − 6.2 (− 6.6–− 5.9) | 0.8 (− 0.1–1.4) | − 0.9 (− 2.3–0.5) | ||
Heart | Dmean | − 0.2 (− 3.4–2.9) | − 3.4 (− 8.7–1.7) | − 0.9 (− 3.3–1.2) | − 2.2 (− 4.0–− 0.8) | |
Lungs | Dmean | 2.9 (− 1.0–8.0) | − 7.8 (− 12.7–4.6) | 0.4 (− 1.6–1.6) | − 1.8 (− 3.9–− 0.3) | |
Spinal cord | D2 % | 3.7 (3.0–4.4) | − 2.4 (− 5.7–1.4) | 1.1 (0.0–1.7) | − 0.8 (− 1.9–0.5) | |
Average c | 1.5 (− 4.8–8.0) | − 5.9 (− 12.7–1.7) | 0.2 (− 3.3–1.8) | − 1.4 (− 4.0–0.5) | ||
Absolute average c | 2.8 (1.0–8.0) | 6.1 (1.4–12.7) | 1.3 (0.0–3.3) | 1.4 (0.3–4.0) |
W water, WB water or bone, WL water or lung, CTV clinical target volume.
aAverage values are the mean of the dose difference statistics for all CTVs (D98 %, D95 %, mean, D50 %, D 2 %), D50 % for femoral heads, Dmean for bladder, D50 % and D2 % for rectum.
bThe average values are the mean of the dose difference statistics for all CTVs (D95 %, mean, D50 %, D 2 %), D50 % for parotids, D2 % for brain stem and spinal cord.
cThe average values are the mean of the dose difference statistics for the CTV (D98 %, D95 %, mean, D50 %, D 2 %), mean dose to heart and healthy lung, D2 % for spinal cord.
Internal anatomy variability of a prostate case is illustrated in Fig. 3, demonstrating that the use of PCTCBCT as the ground truth is, in some instances, essential for meaningful DVH dosimetric comparison for CBCT dose calculation accuracy. Previous studies [12, 15–21, 23] of CBCT dose calculation accuracy have either used rigid (which may have been acquired weeks prior to the CBCT and therefore exhibit substantial anatomical differences) or deformable image registrations (and their associated uncertainty) between the original planning CT and CBCT.
Pelvis cases
RSauto tended to underestimate the proportion of lower-density tissue present in larger patients (Fig. 4), corresponding to an underestimation of the dose. This underestimation has been observed in a larger cohort of pelvic cases that were not used in this dose validation study as they did not have a CBCT and CT acquired on the same day. However, for patients with anterior-posterior distance (DAP) < 25 cm, RSauto resulted in the most accurate dose distributions of all CTN adjustment methods with an average of 0.7 % absolute difference in dose when compared with the ground truth. Furthermore, for patients with DAP < 25 cm, the WB method worked well with the assignment of water (1 g/cm3) to all non-bone tissue, being adequate in such cases as these patients commonly have similar amounts of lower-density (0.95 g/cm3) and connective (1.05 g/cm3) tissue. However, for patients with higher proportions of adipose tissue, the WB method did not perform as well. The W method performed better than WB for patients with DAP > 25 cm, with the additional adipose tissue being balanced out by the higher-density bone. The CBCTr method was unable to accurately reconstruct higher-density material for the larger patient, resulting in a similar effect to that observed with the W method for patients with DAP < 25 cm. For this treatment site, the RSauto method produced the best results when compared with the ground truth, with an average absolute difference of 0.7 % (range: 0.1–2.5 %) when only considering patients with DAP < 25 cm. For CBCT images acquired around the pelvis of patients with DAP < 25 cm, we therefore recommend using the RSauto method for CTN adjustment. However, the WB and W methods worked with sufficient accuracy [20] for patients with DAP < 25 cm and > 25 cm, respectively.
H&N cases
All three methods investigated for CTN adjustment for H&N cases resulted in acceptable [20] average absolute dose deviations of ≤ 1.0 % when compared with the ground truth, with CBCTr (0.5 %) and RSauto (0.6 %) performing best. For CBCT images acquired of the H&N, on the basis of the results presented in Table 3, its integration within a TPS, and its ease and speed of implementation, we recommend using the RSauto method for CTN adjustment. Larger dose differences were observed in the inferior part of the target at the level of the shoulders. Similar to the methods employed for the contralateral lung ROIs (Fig. 1), this dose disagreement may be improved by generating additional bone ROIs for the shoulders to enable more accurate dose calculation on the CBCT data set.
Lung cases
Assuming that density overrides external to the FOV ROI were applied as per Fig. 1, the WL and RSauto methods generated dose distributions most similar to PCTCBCT calculations, with average absolute dose differences of 1.3 and 1.4 %, respectively (Table 3). When the RSauto or CBCTr methods were applied to VMAT treatments without applying additional contralateral lung density overrides, the average absolute dose difference increased to over 3.0 %. The W method performed poorly for lung cases as, unlike the other methods investigated for this site, it did not account for the lower-density lung tissue present. On the basis of the results presented in Table 3, for CBCT images acquired of the lung, we recommend using either the RSauto or WL method for CTN adjustment. This study did not investigate the dependence of CBCT-based dose calculations on lung intra-fraction motion.
Discussion
For the sites investigated in this study, the differences between the doses calculated on the CBCT compared with the diagnostic CT, when using RSauto, were similar or better than other reported results. Fotina et al. demonstrated that density override techniques applied to CBCT scans of the prostate resulted in dosimetric differences below 2 %, whereas the use of conversion curves to calculate dose on CBCT images has been shown to result in differences of up to 5 % [12]. However, Richter et al. demonstrated that conversion curves specific to anatomical sites can result in comparable dose calculation accuracy with differences below 2 % [17].
Owing to weight loss during radiotherapy for H&N cancer [7, 24, 31, 32], for the H&N cases it was necessary to use the PCTCBCT as the input to the algorithm within the TPS used to define the CBCT body contour. When the PCTorig was used, the large differences in the body contour between the PCTCBCT and CBCT meant dose differences between that calculated on the CBCT and the ground truth could not be assessed with confidence. The use of the PCTCBCT data as a priori knowledge for this validation study reduced the uncertainty of the shape of the CBCT body contour, resulting in more accurate estimations of CBCT dose. If CBCT scans of the H&N were to be routinely used for dose calculation, we would recommend using either a medium or large FOV to ensure the entire patient contour at the superior/inferior extent of the tumor was imaged in order to generate the entire patient contour on the CBCT image. Cupping artifacts can have a large effect on CBCT dose calculations, and the degree to which they affect the dose depends on the CTN adjustment method used. CBCT-density look-up table techniques are the most sensitive in this context. However, CBCT dose distributions generated using density override methods, such as those discussed in this study, are also affected by image quality with increasing sensitivity with increasing number of density bins. However, the results indicate that binning the CBCT voxels into six densities is an appropriate methodology that results in dose distributions that are sufficiently accurate. As well as automatically defining the density threshold levels, the TPS allows them to be manually defined by the user. Especially important for larger-sized patients, this approach could be used to ensure an appropriate proportion is assigned as lower-density adipose tissue. However, tissue assignment as a manual process is subject to error, which was not investigated in this study but will be explored in future work. Although this study involved a small number of cases in each of the three treatment sites, we believe that we have identified patients with minimal anatomic differences between the CBCT and PCTCBCT, enabling the accuracy of CBCT dose calculation to be assessed and for the RSauto method to be validated.
Conclusion
We investigated various methods, including RSauto that has not been assessed before, for CTN adjustment for CBCT dose calculations. RSauto methods were validated, resulting in CBCT dose calculations for lung patients, for patients treated in the pelvis with DAP < 25 cm, and for H&N cases, which were similar to those calculated on diagnostic CT images. Its implementation into a commercially available TPS makes the RSauto approach the fastest and easiest-to-use of all methods investigated in this study and it can easily be performed as part of a typical clinical workflow. The differences reported in this study between the doses calculated on the CBCT compared with the ground truth, when using density overrides, were similar to other reported results [12] and better than those that use CBCT-density look-up table methods [12]. However, in this study we used a planning CT acquired on the same day as the CBCT as the ground truth in order to minimize the uncertainties often present in studies of this kind. Although the RayStation TPS was used for this work, the methods described could be replicated in other treatment planning systems. Furthermore, the density override methods for CTN adjustment that were investigated in this work are applicable and transferable to dose calculation on segmented MR images.
Acknowledgments
This work was undertaken in The Royal Marsden NHS Foundation Trust, which received a proportion of its funding from the NHS Executive; the views expressed in this publication are those of the authors and not necessarily those of the NHS Executive. This work was supported by Cancer Research UK Programme Grants C7224/A13407 and C33589/A19727. The authors also acknowledge the support of the National Institute for Health Research Royal Marsden and Institute of Cancer Research Biomedical Research Centre.
Compliance with ethical guidelines
Conflict of interest
A. Dunlop reports grants from NIHR Biomedical Research Centre and grants from Cancer Research UK C7224/A13407. D. McQuaid reports grants from NIHR Biomedical Research Centre and grants from Cancer Research UK C7224/A13407. J. Murray states that there are no conflicts of interest. K. Newbold has received honoraria for advisory roles for Genzyme, Astra Zeneca, and Eisai. S. Nill reports grants from NIHR Biomedical Research Centre, grants from Cancer Research UK Programme Grant C33589/A19727, during the conduct of the study. V.N. Hansen states that there are no conflicts of interest. C. Nutting reports grants from Cancer Research UK C7224/A13407. G. Poludniowski states that there are no conflicts of interest. K. Harrington reports grants from NIHR Biomedical Research Centre, grants from Cancer Research UK C7224/A13407, during the study, and outside the submitted work, grants from Merck, from Cellgene, from Oncolytics Biotech, and from Genelux Corporation. S. Bhide states that there are no conflicts of interest. U. Oelfke reports grants from NIHR Biomedical Research Centre, grants from Cancer Research UK Programme Grant C33589/A19727, and grants from EPSRC Platform Grant EP/H046526/1, during the study.
All patients in this study had previously taken part in separate protocols which were approved by the institutional review board and research ethics committee, for which informed consent had been obtained for use of data for research purposes.
References
- 1.Smitsmans MH, de Bois J, Sonke JJ, et al. Automatic prostate localization on cone-beam CT scans for high precision image-guided radiotherapy. Int J Radiat Oncol Biol Phys. 2005;63:975–984. doi: 10.1016/j.ijrobp.2005.07.973. [DOI] [PubMed] [Google Scholar]
- 2.Purdie TG, Bissonnette JP, Franks K, et al. Cone-beam computed tomography for on-line image guidance of lung stereotactic radiotherapy: localization, verification, and intrafraction tumor position. Int J Radiat Oncol Biol Phys. 2007;68:243–252. doi: 10.1016/j.ijrobp.2006.12.022. [DOI] [PubMed] [Google Scholar]
- 3.Moseley DJ, White EA, Wiltshire KL, et al. Comparison of localization performance with implanted fiducial markers and cone-beam computed tomography for on-line image-guided radiotherapy of the prostate. Int J Radiat Oncol Biol Phys. 2007;67:942–953. doi: 10.1016/j.ijrobp.2006.10.039. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Jaffray D, Siewerdsen J, Wong J, et al. Flat-panel cone-beam computed tomography for image-guided radiation therapy. Int J Radiat Oncol Biol Phys. 2001;53:1337–1349. doi: 10.1016/S0360-3016(02)02884-5. [DOI] [PubMed] [Google Scholar]
- 5.Vestergaars A, Muren L, Sondergaard J, et al. Adaptive plan selection vs. Re-optimisation in radiotherapy for bladder cancer: a dose accumulation comparison. Radiother Oncol. 2013;109:457–462. doi: 10.1016/j.radonc.2013.08.045. [DOI] [PubMed] [Google Scholar]
- 6.Pos F, Hulshof M, Lebesque J, et al. Adaptive radiotherapy for invasive bladder cancer: a feasibility study. Int J Radiat Oncol Biol Phys. 2006;64:862–868. doi: 10.1016/j.ijrobp.2005.07.976. [DOI] [PubMed] [Google Scholar]
- 7.Hansen E, Bucci M, Quivey J, et al. Repeat CT imaging and replanning during the course of IMRT for head-and-neck cancer. Int J Radiat Oncol Biol Phys. 2006;64:355–362. doi: 10.1016/j.ijrobp.2005.07.957. [DOI] [PubMed] [Google Scholar]
- 8.Schwartz D, Dong L (2011) Adaptive radiation therapy for head and neck cancer—can an old goal evolve into a new standard? J Oncol 2011. doi:10.1155/2011/690595. (Article ID 690595, 13 pages) [DOI] [PMC free article] [PubMed]
- 9.Chai X, van Herk M, Betgen A, et al. Automatic bladder segmentation on CBCT for multiple plan ART of bladder cancer using a patient-specific bladder model. Phys Med Biol. 2012;57:3945–3962. doi: 10.1088/0031-9155/57/12/3945. [DOI] [PubMed] [Google Scholar]
- 10.Létourneau D, Wong R, Moseley D, et al. Online planning and delivery technique for radiotherapy of spinal metastases using cone-beam CT: image quality and system performance. Int J Radiat Oncol Biol Phys. 2007;67:1229–1237. doi: 10.1016/j.ijrobp.2006.09.058. [DOI] [PubMed] [Google Scholar]
- 11.Létourneau D, Kaus M, Wong R, et al. Semiautomatic vertebrae visualization, detection, and identification for online palliative radiotherapy of bone metastases of the spine. Med Phys. 2008;35:367–376. doi: 10.1118/1.2820631. [DOI] [PubMed] [Google Scholar]
- 12.Fotina I, Hopfgartner J, Stock M, et al. Feasibility of CBCT-based dose calculation: comparative analysis of HU adjustment techniques. Radiother Oncol. 2012;104:249–256. doi: 10.1016/j.radonc.2012.06.007. [DOI] [PubMed] [Google Scholar]
- 13.Kim S, Yoo S, Yin FF, et al. Kilovoltage cone-beam CT: comparative dose and image quality evaluations in partial and full-angle scan protocols. Med Phys. 2010;37:3648–3459. doi: 10.1118/1.3438478. [DOI] [PubMed] [Google Scholar]
- 14.Stock M, Pasler M, Birkfellner W, et al. Image quality and stability of image- guided radiotherapy (IGRT) devices: a comparative study. Radiother Oncol. 2009;93:1–7. doi: 10.1016/j.radonc.2009.07.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Yang Y, Schrwibmann E, Wang C, et al. Evaluation of on-board kV cone beam CT (CBCT)-based dose calculation. Phys Med Biol. 2007;52:685–705. doi: 10.1088/0031-9155/52/3/011. [DOI] [PubMed] [Google Scholar]
- 16.Yoo S, Yin F-F. Dosimetric feasibility of cone-beam CT-based treatment planning compared to CT-based treatment planning. Int J Radiat Oncol Biol Phys. 2006;66:1553–1561. doi: 10.1016/j.ijrobp.2006.08.031. [DOI] [PubMed] [Google Scholar]
- 17.Richter A, Hu Q, Steglich D, et al. Investigation of the usability of conebeam CT data sets for dose calculation. Radiat Oncol. 2008;3:42–55. doi: 10.1186/1748-717X-3-42. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Hatton J, McCurdy B, Greer P. Cone beam computerized tomography: the effect of calibration of the Hounsfield unit to electron density on dose calculation accuracy for adaptive radiotherapy. Phys Med Biol. 2009;54:329–346. doi: 10.1088/0031-9155/54/15/N01. [DOI] [PubMed] [Google Scholar]
- 19.Hatton J, Greer P, Tang C, et al. Does the planning dose-volume histogram represent treatment doses in image-guided prostate radiation therapy? Assessment with cone-beam computerised tomography scans. Radiother Oncol. 2011;98:162–168. doi: 10.1016/j.radonc.2011.01.006. [DOI] [PubMed] [Google Scholar]
- 20.Poludniowski G, Evans P, Webb S. Cone beam computed tomography number errors and consequences for radiotherapy planning: an investigation of correction methods. Int J Radiat Oncol Biol Phys. 2012;84:109–114. doi: 10.1016/j.ijrobp.2012.02.019. [DOI] [PubMed] [Google Scholar]
- 21.Onozato Y, Kadoya N, Fujita Y, et al. Evaluation of on-board cone beam computed tomography-based dose calculation with deformable image registration using Hounsfield unit modifications. Int J Radiat Oncol Biol Phys. 2014;89:416–423. doi: 10.1016/j.ijrobp.2014.02.007. [DOI] [PubMed] [Google Scholar]
- 22.Poludniowski G, Evans P, Hansen VN, et al. An efficient Monte-Carlo-based algorithm for scatter correction in keV cone-beam CT. Phys Med Biol. 2009;54:3847–3864. doi: 10.1088/0031-9155/54/12/016. [DOI] [PubMed] [Google Scholar]
- 23.Wang J, Hu W, Cai G, et al. Using corrected cone-beam CT image for accelerated partial breast irradiation treatment dose verification: the preliminary experience. Radiat Oncol. 2013;8:214–222. doi: 10.1186/1748-717X-8-214. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Veiga C, McClelland J, Moinuddin S, et al. Toward adaptive radiotherapy for head and neck patients: feasibility study on CT-to-CBCT deformable registration for “dose of the day” calculations. Med Phys. 2014;41:031703. doi: 10.1118/1.4864240. [DOI] [PubMed] [Google Scholar]
- 25.Fortunati V, Verhaart R, Angeloni F, et al. Feasibility of multimodal deformable registration for head and neck tumor treatment planning. Int J Radiat Oncol Biol Phys. 2014;90:85–93. doi: 10.1016/j.ijrobp.2014.05.027. [DOI] [PubMed] [Google Scholar]
- 26.Brock K. Results of a multi-institution deformable registration accuracy study (MIDRAS) Int J Radiat Oncol Biol Phys. 2010;76:583–596. doi: 10.1016/j.ijrobp.2009.06.031. [DOI] [PubMed] [Google Scholar]
- 27.Mencarelli A, van Kranen S, Hamming-Vrieze O, et al. Deformable image registration for adaptive radiation therapy of head and neck cancer: accuracy and precision in the presence of tumor changes. Int J Radiat Oncol Biol Phys. 2014;90:680–687. doi: 10.1016/j.ijrobp.2014.06.045. [DOI] [PubMed] [Google Scholar]
- 28.Lagendijk L, Raaymakers B, Raaijmakers AJ, et al. MRI/linac integration. Radiother Oncol. 2008;86:25–29. doi: 10.1016/j.radonc.2007.10.034. [DOI] [PubMed] [Google Scholar]
- 29.Poludniowski G, Evans P, Kavanagh A, et al. Removal and effects of scatter-glare in cone-beam CT with an amorphous-silicon flat-panel detector. Phys Med Biol. 2011;56:1837–1851. doi: 10.1088/0031-9155/56/6/019. [DOI] [PubMed] [Google Scholar]
- 30.Lee Y, Bollet M, Charles-Edwards G, et al. Radiotherapy treatment planning of prostate cancer using magnetic resonance imaging alone. Radiother Oncol. 2003;66:203–216. doi: 10.1016/S0167-8140(02)00440-1. [DOI] [PubMed] [Google Scholar]
- 31.Barker J, Garden A, Ang KK, et al. Quantification of volumetric and geometric changes occuring during fractionated radiotherapy for head and neck cancer using an integrated CT/linear accelerator system. Int J Radiat Oncol Biol Phys. 2004;59:960–970. doi: 10.1016/j.ijrobp.2003.12.024. [DOI] [PubMed] [Google Scholar]
- 32.Marzi S, Pinnaro P, D D, et al. Anatomical and dose changes of gross tumour volume and parotid glands for head and neck cancer patients during intensity-modulated radiotherapy: effect on the probability of xerostomia incidence. Clin Oncol (R Coll Radiol) 2012;24:e54–e62. doi: 10.1016/j.clon.2011.11.006. [DOI] [PubMed] [Google Scholar]