Determination of Cefoperazone Sodium in Presence of Related Impurities by Linear Support Vector Regression and Partial Least Squares Chemometric Models

Ibrahim A Naguib 1,2,*, Eglal A Abdelaleem 2, Hala E Zaazaa 3, Essraa A Hussein 2
PMCID: PMC4668319  PMID: 26664764

Abstract

A comparison between partial least squares regression and support vector regression chemometric models is introduced in this study. The two models are implemented to analyze cefoperazone sodium in presence of its reported impurities, 7-aminocephalosporanic acid and 5-mercapto-1-methyl-tetrazole, in pure powders and in pharmaceutical formulations through processing UV spectroscopic data. For best results, a 3-factor 4-level experimental design was used, resulting in a training set of 16 mixtures containing different ratios of interfering moieties. For method validation, an independent test set consisting of 9 mixtures was used to test predictive ability of established models. The introduced results show the capability of the two proposed models to analyze cefoperazone in presence of its impurities 7-aminocephalosporanic acid and 5-mercapto-1-methyl-tetrazole with high trueness and selectivity (101.87 ± 0.708 and 101.43 ± 0.536 for PLSR and linear SVR, resp.). Analysis results of drug products were statistically compared to a reported HPLC method showing no significant difference in trueness and precision, indicating the capability of the suggested multivariate calibration models to be reliable and adequate for routine quality control analysis of drug product. SVR offers more accurate results with lower prediction error compared to PLSR model; however, PLSR is easy to handle and fast to optimize.

1. Introduction

Cefoperazone sodium (CEF), (6R,7R)-7-[[(2R)-2-[[(4-ethyl-2,3-dioxopiperazin-1-yl)carbonyl]amino]-2-(4-hydroxyphenyl)acetyl]amino]-3-[[(1-methyl-1H-tetrazol-5-yl) sulphanyl]methyl]-8-oxo-5-thia-1-azabicyclo[4.2.0]oct-2-ene-2-carboxylate (Figure 1(a)) [1], is a third-generation cephalosporin that acts by inhibiting bacterial cell wall biosynthesis [2]. According to B. P. [3], 7-aminocephalosporanic acid (7-ACA) (Figure 1(b)), (6R,7R)-3-[(acetyloxy)methyl]-7-amino-8-oxo-5-thia-1-azabicyclo[4.2.0]oct-2-ene-2-carboxylic acid, and 5-mercapto-1-methyl-tetrazole (5-MER) (Figure 1(c)), 1-methyl-1H-tetrazole-5-thiol, are deemed as potential impurities for CEF. 7-ACA is a significant and key intermediate needed for the synthesis of semisynthetic cephalosporin antibiotics in pharmaceutical industries [4]. Additionally, 7-ACA is the core chemical structure of cephalosporins. Chemical compounds containing this core are known to be relatively stable to hydrolysis and tolerant to β-lactamase, the enzyme that hydrolyzes cephalosporins. Accordingly, 7-ACA is a very important moiety for the activity of all cephalosporins. 5-MER is one of tetrazole derivatives that are used as intermediate of cephalosporin side chains.

Figure 1.

Figure 1

The chemical structure of CEF (a) and its reported impurities 7-ACA (b) and 5-MER (c).

Literature review presents several analytical methods for assay of CEF in its pharmaceutical formulation including spectrophotometric methods for determination of CEF [5, 6], near-infrared reflectance spectroscopy [7], and derivative UV spectrophotometry for determination of CEF in combination with sulbactam [8]. Chromatographic methods were applied for analysis of CEF and sulbactam [9, 10]; besides, an HPLC method with β-cyclodextrin stationary phase for determination of CEF, ampicillin, and sulbactam was reported [11]. CEF and sulbactam were determined in plasma also by LC-MS/MS method [12]. Additionally, electrochemical behavior and voltammetric determination of CEF [13, 14] were reported.

There are two main aims for the presented study. Firstly, the presented chemometric models show the ability of multivariate models to analyze selectively CEF in ternary mixtures with its two reported impurities using cost-effective and available instruments like UV spectrophotometer. Second, the presented study aims to compare two methods of multivariate calibration, PLSR and linear SVR chemometric models, through assay of CEF, 7-ACA, and 5-MER mixtures indicating the advantages and limitations of each model. The selected models offer better trueness and precision for quantitative determination of CEF in pharmaceutical formulation compared to the reported HPLC method [15].

2. Experimental

2.1. Instruments

A double beam UV-VIS spectrophotometer (SHIMADZU, Japan) model UV-1601 PC equipped with a quartz cell of 1 cm width and connected to IBM compatible computer was used. The software used was UVPC personal spectroscopy software version 3.7.

2.2. Material and Reagents

2.2.1. Pure Standard

CEF was provided by Pharco Pharmaceuticals Co., Egypt. Both 7-ACA (CAS number 957-68-6) and 5-MER (CAS number 13183-79-4) were bought from Sigma-Aldrich through the Egyptian International Center for import and export (EIC, Egypt).

2.2.2. Pharmaceutical Formulation

Cefobid 0.5 gm vials (batch number (2203)) were produced by Pfizer Pharmaceutical Industries Co. and cefoperazone 1 gm vials (batch number (1240395)) were manufactured by Sigmatec Pharmaceutical Industries Co. There are 3 concentrations available in the market for vials: 0.5 gm, 1 gm, and 1.5 gm.

2.2.3. Chemicals and Reagents

All chemicals and solvents used in this study were of analytical grade, including methanol (E. Merck, Germany), water for injection B. P. 2003 (Egypt Otsuka Pharmaceutical Co., S.A.E., 10th of Ramadan City, ARE), and dipotassium hydrogen phosphate (K2HPO4) (El-Nasr Pharmaceutical Chemicals Co., Abu-Zaabal, Cairo, Egypt).

2.2.4. Standard Solutions

(a) Stock standard solutions of 1 mg mL−1 for each of CEF, 7-ACA, and 5-MER were prepared in 3 mL of 0.05 M K2HPO4 solution and volume was completed with pure methanol. 7-ACA is soluble only in slightly alkaline solvent; accordingly the solubility was achieved by addition of fixed small volume of 0.05 M K2HPO4 solution before adding methanol. This step was carried out with all stock solutions of CEF, 7-ACA, 5-MER, and pharmaceutical formulation as well.

(b) Working standard solution of 100 μg mL−1 of CEF was prepared in methanol. Two working standard solutions were prepared for each impurity. Working standard solution (1) of 100 μg mL−1 and working standard solution (2) of 10 μg mL−1 for both 7-ACA and 5-MER were prepared to allow preparation of final mixtures with very small concentrations of the impurities, up to 3% of CEF calculated on molar basis.

2.3. Linearity

UV spectra of a set of standards of CEF from 1 to 70 μg mL−1 were recorded from 210 to 300 nm. CEF exhibited linearity between 5 and 50 μg mL−1 at its λ max is 229 nm. The superimposed spectra of 10 μg mL−1 of CEF, 7-ACA, and 5-MER are shown in Figure 2.

Figure 2.

Figure 2

Zero order absorption spectra of 10 μg mL−1 of CEF (—), 7-ACA (-  -  -), and 5-MER (…….) using methanol as blank.

2.4. Experimental Design

2.4.1. Calibration Set

A 4-level, 3-factor calibration design was implemented using 4 concentration levels coded as +2, +1, −1, and +1, where (−1) is the central level for each of the components to be analyzed including the main drug (CEF) and its two impurities (7-ACA and 5-MER). The design aims to span the mixture space appropriately, where 4 mixtures for every compound at every concentration level exist, ending up with 16 mixtures for the training set [16]. The central level selected for the design was 20 μg mL−1 for CEF and the concentration of each level for CEF depended on its calibration range. Concentration levels of impurities were based on involving impurities in an amount up to 3% of CEF calculated on molar basis to span most of the probabilities in future analyses. Table 1 represents the design matrix for concentrations. 2D scores plot of the first two PCs of the concentration matrix was drawn to affirm the orthogonality, rotatability, and symmetry of the training set mixtures (presented as circles) as shown in Figure 3. Mean centering of all types of data was the best preprocessing method to provide best results.

Table 1.

Concentration design matrices in µg mL−1 for the 4-level 3-factor experimental design, showing 16 training set mixtures together with the 9 test set mixtures.

Training set Test set
CEF 7-ACA 5-MER CEF 7-ACA 5-MER
18 0.13 0.065 25 0.17 0.09
18 0.15 0.07 19 0.14 0.07
20 0.15 0.09 22 0.18 0.075
20 0.2 0.07 21 0.14 0.08
26 0.15 0.065 23 0.15 0.065
20 0.13 0.085 25 0.2 0.085
18 0.18 0.085 19 0.13 0.08
24 0.18 0.07 22 0.16 0.07
24 0.15 0.085 21 0.16 0.08
20 0.18 0.065
24 0.13 0.09
18 0.2 0.09
26 0.2 0.085
26 0.18 0.09
24 0.2 0.065
26 0.13 0.07
Figure 3.

Figure 3

Scatter plot for scores of mean centered 16 training set samples and the 9 test set samples concentration matrices of the 4-level 3-component experimental design.

2.4.2. Test Set

To examine the validity and prediction capabilities of the introduced chemometric models, the independent test set mixtures were obtained by preparation of nine independent mixtures other than the training set mixtures but within the concentration space of the design as indicated in Table 1. The well position of the mixtures of both training set and test set mixtures is shown in Figure 3.

2.4.3. Analysis of Cefobid and Cefoperazone Vials

For each of the two dosage forms, accurately weighed aliquot equivalent to 100 mg of CEF was transferred into 100 mL volumetric flask. To prepare stock solution, 3 mL of 0.05 M K2HPO4 solution was added and volume was completed using pure methanol. The solution was then diluted to prepare 100 μg mL−1 working solution using methanol as solvent. Lastly, 2 mL portion of the working solution was diluted to 10 mL with methanol. The average of three corresponding spectra was recorded. This experiment was replicated six times and the produced spectra were analyzed by the suggested models.

2.5. Software

Codes for PLSR (PLS1 algorithm [17], bootstrap, and grid search were written using Matlab 7.5.0 (R2007b)). The codes for SVR algorithm were obtained from the internet website http://onlinesvr.altervista.org/.

3. Chemometric Methods

Multivariate calibration models are chemometric tools that set a relation between the spectra in data matrix X and the concentrations in data vector c. Multiple linear regression (MLR), principal component regression (PCR), and partial least squares regression (PLSR) are among the common models used for pharmaceutical analysis. PCR and PLSR methods are more preferable than MLR, because MLR needs further variable selection steps to perform optimally and to avoid multicollinearity. Additionally, PCR and PLSR can cope with a large number of spectral variables by decomposing the X data matrix into a relatively small number of scores. The scores matrix T then replaces the original X data matrix in the further steps. PLSR is more developed than PCR, where the c data vector shares in construction of the scores as well [17, 18]. Furthermore, the compression to a small number of scores works as a useful filter for noise [19]. Hence, PLSR is implemented in our presented analysis.

3.1. Partial Least Squares Regression (PLSR)

The PLSR model depends on the theory of existence of a linear relation between the X data matrix and the independent variables in concentration vector c [20]. The data matrix X and the response vector c are decomposed using a given number of PLS components (latent variable LVs) [2124] as follows:

X=T·P+E,c=T·q+f, (1)

where T and P are the scores and loadings for X and q is the loading vector for c. E and f are the residuals for X and c, respectively. PLSR is considered one of the best in multivariate calibration because it is reported to perform better than MLR and PCR in several pharmaceutical applications [20].

Optimization of Number of PLS Components for PLSR Model. For prediction of optimum number of PLS components, bootstrap technique [25, 26] was used. This technique is based on dividing the original training set to two-thirds (bootstrap training set) and one-third (bootstrap test set). The PLSR model is then applied through building a model with the bootstrap training set to predict concentrations in the bootstrap test set and calculating the prediction error through the following equation:

RMSEP=1Nn=1Ncnc^nA2, (2)

where N is the number of bootstrap test set samples, c n is the known concentration of sample n, and c^nA is the corresponding predicted concentration at the defined number of PLS components. Equation (2) represents just one iteration out of 1000 iterations that were implemented in the presented study. The higher the number of iterations, the higher the probability of selecting all samples in both data sets (training set and test set). Finally a plot was established of the average of the 1000 root mean square error of prediction (RMSEP) values for different number of PLS components against the corresponding number of PLS components to choose the optimum number of optimum PLS components. The bootstrap training set was mean centered with every iteration.

3.2. Support Vector Regression (SVR)

For a data set X  (I × J) of an output vector c, SVR models aim to find a multivariate regression function f(x) that depends on X to predict a desired response (e.g., concentration of a chemical compound) from an object (e.g., a spectrum). SVR model equations are illustrated in literature [2628] and the summary equation can be given as follows:

fx=i,j=1Nαiαixi·xi+b, (3)

where α i and α i are the Lagrange multipliers that fit to the constraint 0 ≤ α i, α i C. C is known as penalty error or regularization constant which determines the trade-off between model simplicity and training error. The parameter b is the offset of regression function f(x). Further illustrations of (3) and the parameters α and C are found in literature [27, 29, 30]. ϵ-insensitive loss function is an additional factor commonly applied for SVR and will be used and optimized in this study [31, 32]. SVR method can be applied for both linear and nonlinear data. Linear SVR model is used in this study, where the used spectral data exhibit linearity guaranteed through the well planned experimental design.

Finally, in prediction step, unknown c^ value can be calculated as given below [33]:

c^=i=1Iαiαixxj+b. (4)

Optimization of Number of the Linear SVR Model Parameters. Optimum ϵ and C values were calculated by using a grid search that depends on 4-fold cross validation to give the lowest root mean square error of cross validation (RMSECV). Primary range of values was set for ϵ (0.01–1) and C (30–1000). For each set of SVR parameters, 4 samples (N = 4) were taken out; the linear established SVR model was applied on the remaining 12 (IN) samples. Further, RMSECV was predicted for the N samples and finally, the average of RMSECV after removal of all samples was calculated as follows:

RMSECV=1Ii=1Icic^i2, (5)

where c i is the true concentration for sample n and c^i is the corresponding predicted concentration.

4. Results and Discussion

4.1. Parameters' Optimization Results

Bootstrap technique was applied to choose optimum number of PLS components to build the best calibration model based on the training set. The optimum number was “three” as shown in Figure 4. Concerning SVR, the grid search that gave the lowest RMSECV (5) resulted in the values (ϵ = 0.02 and C = 280).

Figure 4.

Figure 4

Choice of optimum number of PLS components (latent variables (LVs)) through plotting number of PLS components against the corresponding root mean square error of prediction (RMSEP) by using the bootstrap method.

4.2. Data Analysis Results

This study aims to introduce a comparative study between two chemometric methods known as PLSR and linear SVR via analysis of CEF in presence of its reported impurities: 7-ACA and 5-MER. The two multivariate models could handle the UV data and overcome the overlapping difficulty of the components' spectra shown in Figure 2. Both models were successfully able to determine the concentrations of CEF in training set and test set manifested by high recovery % with low SD as presented in Table 2. The RMSEP is a parameter used to evaluate the prediction abilities of the two models, Table 2. RMSEP comparative plot between PLSR and linear SVR for prediction of test set samples is presented in Figure 5.

Table 2.

Assay results for prediction of training set (autoprediction) and independent test set of CEF by PLSR and linear SVR chemometric models.

Training set PLSR Linear SVR
Taken (µg mL−1) Found (µg mL−1) R% Found (µg mL−1) R%
18 18.577 103.21 18.629 103.50
18 18.029 100.16 18.136 100.76
20 19.850 99.25 19.920 99.60
20 19.889 99.44 19.980 99.90
26 26.114 100.44 26.110 100.42
20 19.899 99.49 19.971 99.85
18 17.897 99.43 17.980 99.89
24 23.737 98.90 23.754 98.97
24 23.997 99.99 23.980 99.92
20 19.727 98.63 19.775 98.87
24 24.005 100.02 23.980 99.92
18 18.089 100.50 18.116 100.65
26 26.236 100.91 26.143 100.55
26 25.919 99.69 25.980 99.92
24 24.118 100.49 24.181 100.76
26 25.917 99.68 25.980 99.92
Mean (%) 100.01 100.21
SD 1.050 1.036
RMSEC 0.2306 0.1960

Test set PLSR Linear SVR
Taken (µg mL−1) Found (µg mL−1) R% Found (µg mL−1) R%

25 25.443 101.77 25.320 101.28
19 19.069 100.36 19.067 100.35
22 22.357 101.62 22.278 101.27
21 21.403 101.92 21.330 101.57
23 23.515 102.24 23.397 101.73
25 25.630 102.52 25.440 101.76
19 19.294 101.55 19.238 101.25
22 22.432 101.96 22.296 101.35
21 21.603 102.87 21.493 102.35
Mean (%) 101.87 101.43
SD 0.7082 0.5359
RMSEP 0.4457 0.3386

Figure 5.

Figure 5

RMSEP bar plots for prediction of independent test set samples for CEF using 1: PLSR and 2: linear SVR models.

There are many ways of comparison included in our study such as root mean square error of calibration (RMSEC), root mean square error of prediction (RMSEP), and calculation intensity and optimization steps. First, it was observed that linear SVR gives the least RMSEC (0.1960) compared to PLSR (0.2306) indicating better trueness and the corresponding standard deviation is smaller also indicating better precision. Second, the comparative bar plot in Figure 5 indicates that linear SVR gives the least RMSEP (0.3386) compared to PLSR (0.4457) indicating higher ability to process future samples and better generalization ability of linear SVR when compared to PLSR.

Concerning calculation intensity, PLSR is computationally simpler than SVR which is deemed to be more intense in calculation and time-consuming because of intense calculations for optimization. Choice of optimum parameters' values for SVR model could be misleading, where parameters that give the lowest RMSECV can be used (considered as overfitting as in PLSR) but still give better RMSEP or vice versa. In this study, we apply 4-fold cross validation to optimize SVR parameters to evade overfitting through prediction of small subsets of data instead of single sample (as used in leave-one-out cross validation technique), so improving the robustness of model and its generalization ability. The model that is more subject to overfitting is usually less robust. Accordingly, with PLSR, selecting too many components will lead to less robust model, that is, with less ability to predict future samples that have unknown signals. SVR parameters are calculated by using k-fold cross validation to evade overfitting, showing better robustness and higher prediction ability for future samples and hence being a more general model [34]. The probability of adherence of SVR to overfitting is less common than PLSR [35, 36]. The implemented linear SVR model in this study has another merit of optimizing only two parameters, unlike the nonlinear SVR models that use kernels and need optimization of more parameters and hence are more time-consuming in optimization process.

In conclusion, PLSR model is considered as one of the best in multivariate calibration and usually used in quality control routine applications and industry. It proved to be simple in optimization and computation and gives comparable results to reference HPLC methods in spite of processing simple UV data. However, SVR is still considered as a more general model with higher predictive ability for future samples.

4.3. Application of the Proposed Methods to the Pharmaceutical Formulation

The proposed chemometric methods were implemented for analysis of CEF in Cefobid 0.5 gr vial and cefoperazone 1 gr vial and satisfactory results with good recoveries were obtained. These results were statistically compared to the results obtained by applying the reported HPLC method [15] using t- and F-tests. The calculated t- and F-values are less than the tabulated ones showing no significant difference between the two introduced models and the reference HPLC method with regarding both trueness and precision, Table 3.

Table 3.

Statistical comparison of the two proposed PLSR and linear SVR chemometric methods against the reported HPLC method for analysis of CEF in pharmaceutical formulations.

(a) Cefobid vial

Parameters PLSR Linear SVR Reported HPLC method∗∗
Mean 102.77 100.30 102.85
SD 0.898 1.031 1.424
Variance 0.807 1.063 2.027
n 6 6 6
Student's t-test (2.228) 0.907 0.006
F-test (5.050) 2.513 1.907

(b) Cefoperazone vial

Parameters PLSR Linear SVR Reported HPLC method∗∗
Mean 98.17 99.33 99.40
SD 1.078 0.993 1.317
Variance 1.161 0.985 1.736
n 6 6 6
Student's t-test (2.228) 0.108 0.914
F-test (5.050) 1.223 1.327

The values between parenthesis are corresponding to the theoretical values of t and F (P = 0.05).

∗∗Reference method is HPLC [15].

5. Conclusion

In general, the goals of this paper are presenting two multivariate chemometric models, PLSR and linear SVR, for analysis of CEF in presence of its reported impurities and making a modest comparison between the two models highlighting the advantages and limitations of each. Concerning predictive ability, the linear SVR proved to be better than PLSR according to RMSEP values indicating better generalization ability. However, PLSR is simpler and fast to optimize.

The two chemometric methods were also applied for the pharmaceutical formulations and statistically compared to reference HPLC method [15]. The calculated t- and F-values were found to be less than tabulated ones showing no significant difference in respect to both trueness and precision. The proposed methods showed high selectivity and trueness. The presented advantages of the proposed models suggest their use for routine quality control analysis without interference of normally encountered pharmaceutical additives or impurities that could be present in minor ratios.

Additionally, the obtained results affirm the possibility of using modern chemometric approaches, especially linear SVR, for assay of different pharmaceutical dosage forms using accessible cheap and simple instruments like UV spectrophotometer even in presence of large number of interfering components with extremely overlapped spectra.

Acknowledgment

The authors would like to acknowledge financial support for this work from the Deanship of Scientific Research (DSR), University of Tabuk, Tabuk, Saudi Arabia, under Grant no. 0054-1436-s.

Disclosure

Parts of Experimental (Section 2) and Chemometric Methods (Section 3) are reproduced from our previous studies on other mixtures [3537]. The authors would like to declare the presence of another study entitled “Determination of Cefoperazone Sodium in Presence of Related Impurities by Improved Classical Least Squares Chemometric Methods: A Comparative Study” which uses the same data sets as a case study to establish a comparison among a set of improved CLS models (not published to date).

Conflict of Interests

The authors declare that there is no conflict of interests regarding the publication of this paper.

References

  • 1.O’Neil M. J. The Merck Index, an Encyclopedia of Chemicals, Drugs and Biologicals. Whitehouse Station, NJ, USA: Merck and Co. Inc.; 2001. [Google Scholar]
  • 2.Reynolds J. E. F. Martindale, The Extra Pharmacopoeia. 31st. London, UK: Royal Pharmaceutica Society; 1996. [Google Scholar]
  • 3.Her Majesty’s Stationary office. British Pharmacopoeia. London, UK: Her Majesty’s Stationary Office; 1980. [Google Scholar]
  • 4.Tan Q., Song Q., Wei D. Single-pot conversion of cephalosporin C to 7-aminocephalosporanic acid using cell-bound and support-bound enzymes. Enzyme and Microbial Technology. 2006;39(5):1166–1172. doi: 10.1016/j.enzmictec.2006.02.028. [DOI] [Google Scholar]
  • 5.Senthilraja M., Sanjaypai P. N. Spectrophotometric determination of cefoperazone sodium in pharmaceutical formulations. Indian Journal of Pharmaceutical Sciences. 2006;68(3):384–385. [Google Scholar]
  • 6.Chen L., Sha O., Zhai W.-J., Yang J.-H. Fading spectrophotometric determination of cefoperazone sodium in injections by Fe3+-sulfosalicylic acid system. Asian Journal of Chemistry. 2013;25(14):7918–7920. doi: 10.14233/ajchem.2013.14705. [DOI] [Google Scholar]
  • 7.Pang H.-H., Feng Y.-C., Hu C.-Q., Xiang B.-R. Construction of universal quantitative models for determination of cefoperazone sodium for injection from different manufacturers using near infrared reflectance spectroscopy. Guang Pu Xue Yu Guang Pu Fen Xi. 2006;26(12):2214–2218. [PubMed] [Google Scholar]
  • 8.Parra A., Garcia-Villanova J., Rodenas V., Gomez M. D. First and second derivative spectrophotometric determination of cefoperazone and sulbactam in injections. Journal of Pharmaceutical and Biomedical Analysis. 1994;12(5):653–657. doi: 10.1016/0731-7085(93)E0022-F. [DOI] [PubMed] [Google Scholar]
  • 9.Li F. S., Xu Z. X., Xiao H. B., Liang X. M. Simultaneous determination of sulbactam sodium and cefoperazone sodium in sulperazone by high-performance liquid chromatography. Se Pu. 2000;18(6):525–526. [PubMed] [Google Scholar]
  • 10.Vinod Kumar K., Dharuman J., Karuna Sree A. RP-HPLC method development and validation for simultaneous estimation of sulbactam and cefoperazone in dosage form and in plasma. International Journal of Pharma and Bio Sciences. 2010;1(4) [Google Scholar]
  • 11.Tsou T.-L., Chen S.-H., Tsou T.-L., et al. Simultaneous determination of ampicillin, cefoperazone and sulbactam in pharmaceutical formulations by HPLC with β-cyclodextrin stationary phase. Journal of Separation Science. 2007;30(15):2407–2413. doi: 10.1002/jssc.200600529. [DOI] [PubMed] [Google Scholar]
  • 12.Zhou Y., Zhang J., Guo B., et al. Liquid chromatography/tandem mass spectrometry assay for the simultaneous determination of cefoperazone and sulbactam in plasma and its application to a pharmacokinetic study. Journal of Chromatography B. 2010;878(30):3119–3124. doi: 10.1016/j.jchromb.2010.09.021. [DOI] [PubMed] [Google Scholar]
  • 13.Hammam E., El-Attar M. A., Beltagi A. M. Voltammetric studies on the antibiotic drug cefoperazone: quantification and pharmacokinetic studies. Journal of Pharmaceutical and Biomedical Analysis. 2006;42(4):523–527. doi: 10.1016/j.jpba.2006.05.002. [DOI] [PubMed] [Google Scholar]
  • 14.Dogan B., Golcu A., Dolaz M., Ozkan S. A. Electrochemical behaviour of the bactericidal cefoperazone and its selective voltammetric determination in pharmaceutical dosage forms and human serum. Current Pharmaceutical Analysis. 2009;5(2):179–189. doi: 10.2174/157341209788172924. [DOI] [Google Scholar]
  • 15.The United States Pharmacopoeial Convention. The United States Pharmacopoeia (USP30) (2007) National Formulary (NF25) Rockville, Md, USA: The United States Pharmacopoeial Convention; 2007. [Google Scholar]
  • 16.Brereton R. G. Multilevel multifactor designs for multivariate calibration. Analyst. 1997;122(12):1521–1529. doi: 10.1039/a703654j. [DOI] [Google Scholar]
  • 17.Gasteiger J. Handbook of Chemometrics. Vol. 3. Weinheim, Germany: Wiley-VCH, KGaA; 2003. [Google Scholar]
  • 18.Wold S., Ruhe A., Wold H., Dunn W. J., III The collinearity problem in linear regression. The partial least squares (PLS) approach to generalised inverses. SIAM Journal on Scientific and Statistical Computing. 1984;5:735–743. [Google Scholar]
  • 19.Faber N. M., Rajkó R. How to avoid over-fitting in multivariate calibration—the conventional validation approach and an alternative. Analytica Chimica Acta. 2007;595(1-2):98–106. doi: 10.1016/j.aca.2007.05.030. [DOI] [PubMed] [Google Scholar]
  • 20.Darwish H. W., Naguib I. A. Improved partial least squares models for stability-indicating analysis of mebeverine and sulpiride mixtures in pharmaceutical preparation: a comparative study. Drug Testing and Analysis. 2013;5(5):325–333. doi: 10.1002/dta.320. [DOI] [PubMed] [Google Scholar]
  • 21.Gerlach R. W., Kowalski B. R., Wold H. O. A. Partial least-squares path modelling with latent variables. Analytica Chimica Acta. 1979;112(4):417–421. doi: 10.1016/S0003-2670(01)85039-X. [DOI] [Google Scholar]
  • 22.Martens H., Naes T. Multivariate Calibration. Chichester, UK: John Wiley & Sons; 1989. [Google Scholar]
  • 23.Brereton R. G. Introduction to multivariate calibration in analytical chemistry. Analyst. 2000;125(11):2125–2154. doi: 10.1039/b003805i. [DOI] [Google Scholar]
  • 24.Geladi P., Kowalski B. R. Partial least-squares regression: a tutorial. Analytica Chimica Acta. 1986;185:1–17. doi: 10.1016/0003-2670(86)80028-9. [DOI] [Google Scholar]
  • 25.Efron B., Tibshirani R. J. An Introduction to the Bootstrap. Chapman & Hall; 1993. [DOI] [Google Scholar]
  • 26.Hjorth J. S. U. Computer Intensive Statistical Methods: Validation, Model Selection and Bootstrap. New York, NY, USA: Chapman & Hall; 1994. [Google Scholar]
  • 27.Suykens J. A. K., Gestel T. V., De Brabanter J., De Moor B., Vandewalle J. Least Squares Support Vector Machines. Singapore: World Scientific; 1999. [Google Scholar]
  • 28.Cristianini N., Shawe-Taylor J. An Introduction to Support Vector Machines and Other Kernel-Based Learning Methods. Cambridge, UK: Cambridge University Press; 2000. [DOI] [Google Scholar]
  • 29.Xu Y., Zomer S., Brereton R. G. Support vector machines: a recent method for classification in chemometrics. Critical Reviews in Analytical Chemistry. 2006;36(3-4):177–188. doi: 10.1080/10408340600969486. [DOI] [Google Scholar]
  • 30.Schölkopf B., Smola A. J. Learning with Kernels. Cambridge, UK: The MIT Press; 2002. [Google Scholar]
  • 31.Gunn S. R. Southampton, UK: University of Southampton; 1998. Support vector machines for classification and regression. http://users.ecs.soton.ac.uk/srg/publications/pdf/SVM.pdf. [Google Scholar]
  • 32.Parrella F. Online Support Vector Regression, Thesis in Information Science, University of Genoa, Italy, 2007, http://onlinesvr.altervista.org/
  • 33.Thissena U., Pepersb M., Üstüna B., Melssena W. J., Buydens L. M. C. Comparing support vector machines to PLS for spectral regression application. Chemometrics and Intelligent Laboratory Systems. 2004;73:169–179. [Google Scholar]
  • 34.Üstün B., Melssen W. J., Oudenhuijzen M., Buydens L. M. C. Determination of optimal support vector regression parameters by genetic algorithms and simplex optimization. Analytica Chimica Acta. 2005;544(1-2):292–305. doi: 10.1016/j.aca.2004.12.024. [DOI] [Google Scholar]
  • 35.Naguib I. A. Stability indicating analysis of bisacodyl by partial least squares regression, spectral residual augmented classical least squares and support vector regression chemometric models: a comparative study. Bulletin of Faculty of Pharmacy—Cairo University. 2011;49(2):91–100. doi: 10.1016/j.bfopcu.2011.10.001. [DOI] [Google Scholar]
  • 36.Naguib I. A., Abdelaleem E. A., Draz M. E., Zaazaa H. E. Linear support vector regression and partial least squares chemometric models for determination of Hydrochlorothiazide and Benazepril hydrochloride in presence of related impurities: a comparative study. Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy. 2014;130:350–356. doi: 10.1016/j.saa.2014.04.024. [DOI] [PubMed] [Google Scholar]
  • 37.Abdelaleem E. A., Naguib I. A., Zaazaa H. E., Hussein E. A. Development and validation of HPLC and HPTLC methods for determination of cefoperazone and its related impurities. Journal of Chromatographic Science. 2015 doi: 10.1093/chromsci/bmv125. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Analytical Methods in Chemistry are provided here courtesy of Wiley