Abstract
Absorbable metals have been widely tested in various in vitro settings using cells to evaluate their possible suitability as an implant material. However, there exists a gap between in vivo and in vitro test results for absorbable materials. A lot of traditional in vitro assessments for permanent materials are no longer applicable to absorbable metallic implants. A key step is to identify and test the relevant microenvironment and parameters in test systems, which should be adapted according to the specific application. New test methods are necessary to reduce the difference between in vivo and in vitro test results and provide more accurate information to better understand absorbable metallic implants. In this investigative review, we strive to summarize the latest test methods for characterizing absorbable magnesium-based stent for bioabsorption/biodegradation behavior in the mimicking vascular environments. Also, this article comprehensively discusses the direction of test standardization for absorbable stents to paint a more accurate picture of the in vivo condition around implants to determine the most important parameters and their dynamic interactions.
Keywords: magnesium, absorbable stent, in vitro, in vivo, biodegradation
Introduction
With the development of regenerative biomaterials, absorbable metallic stents are being investigated as an alternative for permanent stents, which hold arteries open after percutaneous coronary intervention to maintain arterial blood flow [1–3]. Absorbable stents are expected to corrode gradually in vivo, with an appropriate host response elicited by released corrosion products, then dissolve completely upon fulfilling the mission to assist with tissue healing [4]. Potential benefits of absorbable stents include cessation or elimination of antiplatelet therapies that complicate the necessity for the treatment of other medical conditions the patients require—adding risk to the overall health [5]. Other possible benefits include the restoration of vessel properties such as vasomotion which is the ability of the vessel to expand and contract as the body attempts to modulate blood flow between times of rest and exercise. Additionally, absorbable stents would likely create the ability to more efficiently treat a previously stented site, or to enable the option of a bypass graft without the hindrance of an existing implant. Absorbable stents may also be used successfully as a way to deliver anti-restenosis drugs [6]. Absorbable stents made of Mg and its alloys have been studied for several years in clinical trials with reportedly encouraging results [6–14].
The accurate determination of bioabsorption for absorbable metals is a technologically important issue in science and biomedical applications [15]. However, there is a paucity of reports that focus on the correlation between in vitro and in vivo testing. It has been observed that corrosion rate in vivo is about four orders of magnitude slower than that in vitro with ASTM standard [16]. In vitro testing methods that accurately mimic the in vivo environment are required to improve the standardization process by which a material is deemed appropriate for more costly in vivo testing which further involves the welfare of animals.
Thorough knowledge of the comportment of Mg and its alloys in physiological environments is limited as a result of the adolescence in terms of information and research available in this area. Every medical device undergoes a common surface–environment interaction in which the device material is exposed to the biological environment that contains organic and inorganic constituents, followed by complex blood, cell and tissue response. Also, the fluid dynamics, gas diffusion, pH and temperature have led to differing impacts on the behavior of Mg at the early- and long-term implantation process. Evaluating and understanding the biological response and subsequent Mg behavior is vital in the development of absorbable metallic stents with controlled degradation rates and optimum biocompatibility. Of fundamental concern is the determination of relevant in vitro environmental setup parameters that approximate in vivo physiological parameters [17].
For the process of commercialization of absorbable Mg-based stent, it is mandatory to perform biodegradation and biological safety evaluations. Determining the most relevant in vitro bench tests as well as identifying the most important parameters to assess is a significant obstacle in spite of the international technical standards available (e.g. ASTM-G31-72 and ISO 10993 series [18]) [19, 20]. A new standard ISO/TR 37137:2014 was published about ‘Cardiovascular biological evaluation of medical devices - Guidance for absorbable implants’, which is to provide interim part-by-part guidance on potential adjustments to various test methods within the 10993 series to account for the intentional release of soluble components or degradation products from absorbable medical devices. For the success of commercialization of absorbable Mg-based stents, appropriate biodegradation testing condition still need to be improved to help screen the biocompatibility of developed Mg-based stents in vitro, which saves cost and testing periods comparing with those in vivo.
In this up-to-date review, we attempt to analyze the influence of physiological parameters with respect to the biodegradation of Mg-based stents. We would like to summarize the latest achievements as well as comment on the selection and use, test methods and the approaches to develop and produce Mg-based stents that are intended to perform clinically with an appropriate host response.
Physiological Parameters
Fluid flow and diffusion
To effectively study, model and optimize absorbable metallic stents, characterizations need to be performed in hydrodynamic conditions that are found within vessel walls. The blood fluid drag force acting on the vessel wall is mechanotransduced into biochemical and biophysical signals that result in changes in vascular behavior and microenvironment around the stent [21]. The stent is exposed to the blood fluid at the initial stage of implantation as a result of fluid convection that dominates the interaction between blood and stent. In the succeeding stage, the intima grows over the stent surface when fluid diffusion dominates in the interaction between tissue and stent. The magnitude of flow-induced shear stress on the vessel walls can be estimated in most vasculature situations by Poiseuille’s law [22] (Fig. 1A). Various arteries have different values of peak and mean wall shear stress in the human body as shown in Table 1. The shear stress ranges from 1 to 6 dyne/cm2 in the venous system and between 10 and 70 dyne/cm2 in the arterial vascular network (Fig. 1B). The presence of a deployed stent in an artery may increase the shear stress to between 70 and 100+ dyne/cm2 [23]. Thus, it is important to analyze the effects of flow conditions in a biologically relevant range of flow conditions for the successful application of absorbable metals for vascular stents [24] and other clinical applications for stents such as trachea and ureter stenosis [25, 26].
Figure 1.
(A) Cross-sectional schematic diagram of a blood vessel illustrating hemodynamic shear stress, τs, the frictional force per unit area acting on the inner vessel wall and on the luminal surface of the endothelium as a result of the flow of viscous blood. (B) Tabular diagram illustrating the range of shear stress magnitudes encountered in veins, arteries and in low-shear and high-shear pathologic states [23].
Table 1.
Peak and mean wall shear stress in typical vascular sites in healthy human subjects
Wall shear stress (dyne/cm2) | Carotid artery [27] | Coronary artery [28] | Suprarenal aorta [29] |
Infrarenal aorta [29] |
Femoral artery [30] | ||
---|---|---|---|---|---|---|---|
Anterior | Posterior | Anterior | Posterior | ||||
Peak | 30.5 ± 5.2 | 12.4 | 48.0 ± 6.0 | 54.0 ± 11.0 | 33.0 ± 2.6 | 30.0 ± 7.0 | 41.4 ± 10.4 |
Mean | 12.3 ± 2.1 | 6.8 ± 0.3 | 8.6 ± 3.2 | 10.4 ± 2.9 | 6.1 ± 2.6 | 4.7 ± 2.6 | 3.6 ± 1.6 |
Data are expressed as mean ± standard deviation. 1 dyne/cm2 = 0.1 Pa = 0.1 N/m2.
The fluid flow has a significant impact on the degradation of absorbable metallic stents, including corrosion type, corrosion rate, corrosion products and local pH change. Shear stress accelerates the overall corrosion rate (including localized, uniform, pitting and erosion corrosions) due to the increase of mass transfer and mechanical force [31, 32]. Shear stress also increases the thickness of the uniform corrosion layer, the localized corrosion coverage ratios and depth, and the removal rate of corrosion products inside the corrosion pits [32]. For instance, the volume loss ratio (31%) of AZ31 stents at a shear stress of 0.56 dyne/cm2 was nearly twice that (17%) at static condition in Dulbecco’s modified Eagle’s medium (DMEM) after 7 days degradation [32]. Furthermore, flow direction made the corrosion product layer facing the flow direction peeled off from the AZ31 stent struts (Fig. 2) [32]. Additionally, flow can assist in maintaining the pH level around a basic physiological value, resulting in continuation of normal degradation behavior [33].
Figure 2.
Scanning electron microscopy image of AZ31 stent at a shear stress of 0.56 dyne/cm2 in DMEM for 7 days. The arrow indicates the flow direction of corrosion medium [32].
To design the fluid dynamic experiment, the geometry of the test bench and the calculation of shear stress need to be considered delicately. In the case of the stent and tube samples, the wall shear stress calculated by Poiseuille’s law always represents the shear stress on the sample. Essentially, the highest shear stress on the stent occurs on the surface of struts facing the flow direction simulated using a three-dimensional (3-D) computational fluid dynamics (CFD) model. This highest shear stress value is more than twice the wall shear stress value [34]. In terms of the geometry, Poiseuille’s law is no longer applicable to the shear stress calculation of conventional plates in the lumen. Lévesque et al. [31] provided a good test bench for this case. The surface of the plate and the wall of the test channel have a smooth transition, resulting in a uniform shear stress zone on the sample surface. It avoids complex flow patterns by the interruption of the geometry of the plate (Fig. 3). Otherwise, there will be a need for a CFD model to calculate shear stress on the sample surface [32]. Additionally, the viscosity of DMEM with supplements for the in vitro test is 0.78 mPa s at 37°C [35], which is lower than that of blood (3.5 mPa s) [28].
Figure 3.
Schematic view of the dynamic test bench and close-up upside view of the test channel with specimen in place. Note: Le = length of laminar flow before reach specimen, W = width and H = height of the test channel [31].
After the intima layer covers on the stent [36], the stent is located between the intima and media. The degradation of the stent is dominated by water molecules and hydrophilic solutes and ions diffusion between vascular tissue and stent. These diffusion rates are related to the structure of the arterial wall as schematically shown in Fig. 4 [37]. The arterial wall is treated as a permeable channel which commonly consists of six layers: glycocalyx, endothelium, intima, internal elastic lamina, media and adventitia [37, 38]. Water molecules and hydrophilic solutes and ions permeate through interendothelial junctions by way of transportation utilizing fenestral pores that are embedded throughout the intima, internal elastic lamina and media sections of the arterial wall. Consequently, it is important to describe the nature of transportation of these molecules and ions through the arterial wall by way of permeation by utilizing porous media transport models. Fluid flow through porous media has been modeled by the conventional Darcy law which is a generalized linear relationship between the flow velocity () and the pressure gradient (p) across a porous medium (Fig. 4) [39]. Furthermore, mass transfer across the arterial wall occurs via two mechanisms: convection associated with pressure-driven transmural flow and mass diffusion caused by concentration gradients. The molecular diffusion within the arterial walls is driven by solute concentration gradients. Building an in vitro simulation model of this case has some challenges, due to the compilation of the structure of the arterial wall and the difficulty of the artery culture in vitro. Therefore, the data from in vivo tests are more substantial and reliable.
Figure 4.
A schematic of the anatomical structure of an arterial wall [37].
Composition of media
The overall bioactivity of Mg alloys is affected by the composition of the physiological solution. As a consequence of the complicated and adjustable in vivo environment, in vitro experiments require the analysis of various physiological corrosion media to elucidate the underlying interaction mechanism between Mg and microenvironment. It is based upon the primary interaction of inorganic ions and proteins with the Mg as they have a strong impact on cell/surface interaction. Jang et al. [40] investigated the effect of different inorganic ions on degradation (Table 2). It was reported that: (i) calcium is not present in the corrosion product layer when only Cl− and OH− anions are available; (ii) the presence of phosphate induces the formation of a densely packed amorphous magnesium phosphate corrosion product layer when and Cl− are present in solution; (iii) octacalcium phosphate and hydroxyapatite are deposited on the surface of the Mg alloy when and Ca2+ are present together in NaCl solution (this coating limits localized corrosion and increases general corrosion resistance); (iv) addition of accelerates the overall degradation rate, which increases with increasing bicarbonate concentration; (v) the degradation rate in bicarbonate solution decreases in the presence of ions due to the formation of insoluble hydroxyapatite on the surface when is present with Ca2+ and . Therefore, the inclusion of proteins in medium solutions is vital in terms of better mimicking an organic microenviroment in vitro, which will facilitate a more accurate understanding of the degradation processes. It has been reported that degradation slows down considerably when proteins are a part of the corrosion medium [41, 42]. This effect is explained as protein adsorption on the surface forming a corrosion-resistant layer. This suggests that the bioactivity of these materials in vivo will be strongly affected by the presence of biomolecules. Cell growth media such as DMEM or Eagle’s minimum essential medium are closer to physiological conditions found in vivo because of the addition of inorganic salts, amino acids and vitamins within the solution.
Table 2.
Analytical methods used in studying the solid corrosion products of Mg and its alloys in vivo and in vitro
Methods | Functions | Reference |
---|---|---|
Optical microscopy | Morphology | [63] |
Electron beam: | Morphology, distribution and composition | [63–65] |
| ||
Microscopic fourier transform infrared spectroscopy | Composition | [63, 64, 66] |
In situ attentuated total reflectance Fourier transform infrared spectroscopy | ||
X-ray diffraction | Composition | [40] |
Microtomography | Morphology and distribution | [67, 68] |
Gas diffusion
Carbon dioxide
CO2 plays a crucial role in the degradation of Mg [17]. The presence of CO2 might trigger several reactions that influence the degradation process. In terms of possible stent applications, the CO2 values in arterial and venous blood are 4.12% and 5.32%, respectively [43]. Furthermore, CO2 produces H2CO3 in aqueous environments, which dissociates into H3O+ and . The latter, hydrogen carbonate, plays a critical role in the blood buffering system and maintaining a neutral pH of the corrosion solution. In parallel, produces and H3O+ in aqueous environments that can result in a basic pH and consequently have an impact on the degradation rate of Mg.
Oxygen
O2 plays a less prominent role in the degradation of Mg [17]. This lower level of influence of O2 on the degradation behavior might be associated with the fact that O2 does not change the pH and that it most likely does not directly interact with Mg. It has been recommended that the concentration of O2 within in vitro test conditions be reduced in an effort to more accurately replicate in vivo conditions because the content of O2 in many tissues is ∼1–7%, which is less than that of ambient air (21%) [44–46]. As a consequence, O2 is not important when 5% CO2 is present within the environment.
pH
As Mg generates H2 and OH− along the progress of its degradation reaction with medium, the pH of the fluid around Mg surface will increase. Although the instability of Mg occurs at pH values <11, the formation of soluble compounds with most inorganic ions inhibits the formation of magnesium hydroxide passive films in the biological environment [47]. These reactions result in an increase in both magnesium concentration and alkalinity near the stent surface, which may affect surrounding cellular and tissue function [48]. At local pH levels over 11, the formation of the magnesium hydroxide film on the Mg surface can create a barrier with the corrosion solution that slows the degradation process [24, 49]. In biological systems, the pH is regulated by the circulation of blood which keeps environment at normal pH levels. Therefore, in the design of in vitro degradation experiments, the degradation solution must be monitored and replaced with fresh medium on a consistent basis to ensure normal pH levels, that mimic the in vivo environment, are maintained. Additionally, the calcium phosphate complex as a passivation layer can be adsorbed on the surface under a high pH microenvironment [33]. Nevertheless, flow can keep pH at around basic physiological value, due to rapid ion diffusion. Therefore, the results from the dynamic fluid tests may be closer to that in vascular environments.
Temperature
The effect of temperature in the human physiological range ∼35.8–37.2°C seems to be less important for Mg degradation [24]. It may affect the adsorption of proteins and thus the biological response.
In brief summary, these physiological parameters are separated as the first and second order relevance [17]. ‘First order relevance’ consists of parameters such as CO2, the presence of NaCl and NaHCO3 and temperature that directly influence the degradation rate. ‘Second order relevance’ consists of parameters such as O2 and proteins that influence the corrosion layer and the final corrosion products. These parameters might also influence the cell reaction in vivo.
Current Methodologies and Characterizations of Biodegradation Behavior
Corrosion rate
Mass loss (ML) in vitro experiments provide a benchmark for determining the actual amount of cumulative corrosion that has occurred for Mg alloy degradation. ML experiments typically have a simple setup in which the experiment requires only a sample, corrosion medium and an accurate microbalance. The ML procedure consists of the sample being placed in the corrosion medium for a specified period of time, after which the sample is removed and the resultant mass change is measured. Prior to measuring the final mass, a cleaning solution such as dilute chromic acid (180–200 g/l CrO3) for 5–10 min [50] is used to remove corrosion products from the surface. The corrosion rate based on this type of experiment is calculated in equation (1) [16]:
(1) |
where CR refers to the corrosion rate (mm/y), W is the weight loss of the sample, A is the exposure area, t is the immersion time, and is the standard density. Although these type of tests are simple to set up and easy to operate, there are inherent limitations such as accounting of non-uniform corrosion as well as the inability to determine the mechanism behavior of Mg alloys.
Morphology analysis provides a visualized measurement for determining the corrosion rate based on the corrosive morphology and structure of the sample. The outward appearances and cross-sectional images of the corrosive samples can be reconstructed by X-ray computed tomography (CT). Corrosion rate depending on the reduction of the metallic volume after the corrosion can be calculated from the 3-D CT data. Assuming a uniform corrosion, the reduction of the implant volume could be converted into a corrosion rate by using a modification of equation (2) [16]:
(2) |
where CR refers to the corrosion rate, ΔV is the reduction in volume that is equal to the residual alloy volume (without corrosion product volume) subtracted from the initial implant volume, A is the sample surface area exposed to corrosion and t is the exposure time. It must be noted that this method is a rough calculation without the consideration of nonuniform corrosion [32]. To obtain a more accurate analysis, the uniform and localized corrosion rates, which are the two major types of Mg corrosion, can be calculated by electron dispersive x-ray spectroscopy (EDX) and CT. The average uniform corrosion rates were calculated according to the thickness of the corrosion product layer divided by exposure time. The thickness of the corrosion product layer can be measured by EDX line analysis. The local corrosion coverage ratio of the sample at different depths can be calculated from 2-D images using Image-Pro® Plus.
Hydrogen evolution experiments are based on the amount of H2 gas that evolves from the interaction between Mg sample and the environment and this type of testing is used to artificially determine mass loss at any point in time. A typical setup is similar to that of a standard mass loss test. The sample is immersed in the corrosion medium and a collector is placed in the medium directly above the sample to collect H2 gas that is produced. A collector consists of an inverted funnel and burette filled with corrosion medium [51]. Many Mg corrosion studies employ the collection of gaseous H2 on the basis that the primary cathodic reaction (2H2O + 2e− → 2OH− +H2 [52]) is an index to the rate of anodic dissolution (where at open circuit, Ianodic = Icathodic) [15]. Thus, the evolution of 1 mol of hydrogen gas (22.4 l) directly corresponds to the dissolution of 1 mol of Mg (24.31 g). The principle behind interpreting H2 evolution is that the volume of H2 gas is deemed to be equivalent to the mass loss of the Mg [19]. However, the majority of the literature concerning hydrogen evolution has not reported the theoretical 1:1 ratio of hydrogen evolved to actual mass loss [19]. This may be attributed to some inadequacies of the hydrogen collection method, given that (i) hydrogen collection may be inefficient in cases where experiment design is not ideal, (ii) the solubility of hydrogen in water varies significantly relative to sea level and temperature, and (iii) studies rarely pre-saturate the corrosion medium with hydrogen [15]. Therefore, it is important to couple mass loss measurements with hydrogen evolution experiments as a way of vetting the results.
Electrochemical technique is crucial for determining and quantifying the corrosion mechanism of Mg alloys. A commonly used technique is the utilization of direct current polarization (DCP) test, which is preceded by a set period of exposure time where the open circuit potential (OCP) is recorded. This allows the material to ‘stabilize’ with the electrolyte and reach a near steady potential. After OCP stabilization, voltage is swept at a controlled rate (e.g. 1 mV s−1) between different pre-set potentials by regulating the current flowing between the working (Mg) and counter (inert metal) electrode. The initial voltage is nominally set to commence at values more negative than (i.e. cathodic to) the OCP, and the scan proceeds to increasingly positive values (that are anodic to the original OCP) [19]. The DCP results show as a Tafel curve, which provides thermodynamic information on corrosion potential (Ecorr), kinetic information from the corrosion current density (icorr) as well as relative anodic and cathodic reactions. In this regard, it is the only method which can reveal the relative anodic and cathodic contributions that lead to an instantaneous corrosion rate. The corrosion rate, CR, in millimeter per year can be determined from equation (3) [53]:
(3) |
where EW is the equivalent weight of the corroding species in grams and is the density of the corroding material in g/cm3. Assuming uniform corrosion, icorr is converted to a corrosion rate (in terms of penetration) and therefore the DCP results do not typically yield an absolute corrosion rate for Mg, but rather are indicative of the severity of the corrosion that is occurring at a select point in time, in terms of current density. However, uniform corrosion is seldom in the degradation of Mg alloys and thus none of the methods described in current reports provides an absolute prediction of the corrosion rate. Conversely, the corrosion rate expressed in current density is highly accurate and can be considered to have the highest resolution of all methods described herein, although the current could be originating from a number of local sites on the surface. The method, however, is short-term and destructive in nature. As such, wherease potentiodynamic polarization is indispensable in understanding the mechanistic origins of corrosion rate, it may not serve as a good index to long-term corrosion rates.
Electrochemical impedance spectroscopy (EIS) is another technique that is used to characterize the surfaces of Mg sample using the frequency response of AC polarization [54, 55]. EIS uses a range of low magnitude polarizing voltages that cycle from a peak anodic to peak cathodic voltage using a spectra of voltage frequencies. Capacitance and resistance values are obtained for each frequency and can then be used to illuminate a number of phenomena and properties of the Mg surface. An appreciated understanding can be summarized in that faithful determination of corrosion rate was possible when the EIS determined polarization resistance (Rp) parameter at the zero frequency limit was used [15]. The Rp is inversely proportional to the icorr as described by the Stern–Geary relationship [53]:
(4) |
where and are the anodic and cathodic Tafel slopes (as described and qualified further below) [56–58]. EIS determined corrosion rate can be transferred by the icorr based on equation (3). The primary benefits of conducting EIS on Mg alloys in an electrolytic solution are the identification and quantification of the formation behavior of corrosion layers produced by the corrosion process. However, EIS results have possible limitations in the fact that they can be affected by the continuation of Mg dissolution at low frequencies and the chosen equivalent circuit. Consequently, to be correctly employed, EIS requires a deep understanding of the corrosion processes that are taking place and how they might be best modeled.
Combined experiments are developed to take advantage of the relative merits of each testing method. King et al. [15] designed to permit in-situ EIS and H2 collection, along with subsequent post-mortem mass loss. The results provided an excellent correlation to mass loss, hydrogen gas collection and EIS-estimated polarization resistance. Doepke et al. [59] developed a real-time monitoring of the solution soluble corrosion products Mg2+, OH− and H2 during immersion tests. This instrumentation was also developed to record electrochemical impedance spectra simultaneously in the same solution to monitor changes in the Mg samples.
Corrosion products
Characterization of the explanted stent materials and identification of the products thereby generated are critical to the maturation of this class of absorbable implants. Various methods have been used to analyze the corrosion products of Mg and its alloys in vivo and in vitro (Table 2). The morphology, distribution and composition of corrosion products can be comprehensively analyzed. Corrosion products of Mg corrosion have become common knowledge, such as the general tendency for Mg to form an oxygen-bearing product: MgO, Mg(OH)2, a Mg carbonate or a mixture thereof [60, 61]. Bowen et al. indicated the gradual replacement of Mg by calcium and phosphorus in vivo as schematically shown in Fig. 5 [63]. Several in vivo works mentioned that calcium and phosphorus formed on the Mg surface as well [8, 11, 60, 62].
Figure 5.
Schematic showing transformations supposed to be taking place in the quasi-steady state (A), including apatitic ion exchange and reactions at the Mg–Mg/O interface, and during the terminal breakdown stage (B), dominated by dissolution of the Mg/O layer upon exposure to the arterial environment [63].
Hydrogen (H2), as a gaseous corrosion product, has always been used as an evaluation parameter of Mg corrosion. Many researchers assume that the formation of gas cavities after implantation of Mg alloys contain primarily hydrogen because it is in aqueous media. However, Kuhlmann et al. [69] reported a direct monitoring method of hydrogen concentration was developed using an amperometric hydrogen sensor and mass spectrometric measurements for Mg alloy disks in mice over a period of 10 days. The results showed that the gas cavities contained only a low concentration of hydrogen gas, even shortly after formation of the cavities, which means that hydrogen must be exchanged very quickly after implantation. Therefore, the fate of H2 in vivo is worth of attention.
The local chemical reactions on the interface are of great importance for understanding the intricate mechanisms of Mg corrosion. Lamaka et al. [70] developed ion-selective microelectrodes for mapping local activity of H+ and Mg2+ by scanning ion-selective electrode technique as well as local ionic current density measurements by scanning vibrating electrode technique over the surface of Mg alloy in aqueous chloride containing solution. The local concentration of Mg2+ ions dissolved due to corrosion from artificial defects and pH changes together with anodic and cathodic currents can be measured by combination of two selective and one vibrating microelectrodes.
Corrosion types
There are varvious corrosion types during Mg degradation process [71], including uniform corrosion [72, 73], localized corrosion [72, 73], flow-induced corrosion [31, 74], erosion corrosion [75], galvanic corrosion [76, 77], pitting corrosion [78, 79], stress corrosion [80, 81], atmospheric corrosion [82, 83], hydrogen cracking [84, 85] and intergranular corrosion [86, 87]. It is worth to note that localized corrosion is always a source of stent fracture.
To estimate the corrosion types, microscopic view of investigated surface areas is a direct method to detect changes to surface morphology. Some complementary electrochemical techniques also can be used. The prediction of corrosion activity, corrosion rate and the type of corrosion process (uniform, pitting or filiform corrosion) can be achieved, based on the corrosion current density and polarization resistance from the potentiodynamic curves. The shape of the curve gives an indication about the form of corrosion taking place at the surface, for example, a constant current range indicates the formation of a passive layer; a symmetric logarithmic curve around the corrosion potential indicates clearly the situation of uniform corrosion; pitting or crevice corrosion can be assessed based on the shape of the anodic part of the i versus E curve after the passive layer has been disrupted [88]. In EIS spectra, the inductive loop at low frequencies of Nyquist plot indicates pitting corrosion of the Mg alloy as well.
Mechanical behaviors under corrosion
When designing absorbable Mg implant devices, the implant material has to achieve certain corrosion behaviors to assure proper mechanical behaviors during the entire healing process. More specifically, the implants must be able to mechanically support the damaged tissue site for at least 3–6 months and then completely degrade and removed from the healed tissue site in 7–24 months. Therefore, the mechanical properties in vivo are vital to the success of an absorbable stent. An in vitro–in vivo correlation of mechanical analysis is proposed in Fig. 6 by Bowen et al. [89] for the physiological corrosion of pure Mg wires in the arterial walls of rats (in vivo) and in static cell culture media (in vitro). In tensile testing, the in vivo degradation was 2.2 ± 0.5, 3.1 ± 0.8, 2.3 ± 0.3 and 3.1 ± 0.7 times slower than corrosion in vitro in terms of effective tensile strength, strain to failure, sample lifetime and combined metric (defined as strength multiplied by elongation), respectively.
Figure 6.
(A) Trend observed for the combined mechanical metric for in vivo (empty circle) and in vitro (filled circle) samples, with the two relevant linear trends (in vivo and in vitro dotted and solid, respectively). (B) Visual presentation of the various mechanical in vivo–in vitro multipliers. Error bars are indicative of standard error [89].
A well-established method to analyze the potential mechanical integrity of a stent is through computational modeling. Computational modeling of the degradation process and the associated mechanical integrity degeneration are investigated in a finite element (FE) framework. A continuum damage model was developed for Mg alloy stents considering superposition of stress corrosion and uniform corrosion [90, 91]. Stress corrosion will take place at the predicted locations with high residual stress, where the stent ring will break earlier than at other locations; uniform corrosion will cover the whole exposed sample surface and attack the stent in a ‘layer-by-layer’ manner [92].
As the model depicted, the beginning of Mg alloy degradation was concentrated mainly at more deformed locations after stent expansion. The degradation proceeded with the mass loss of the outer surface of the stent, associated with the decreasing mechanical integrity of the stent. The optimized Mg alloy stent, with thicker strut than the Magic stent design [90], showed decreased maximum principal stress after recoil (163 MPa) and increased half normalized time of vessel recoil [90]. Grogan et al. [93] developed a phenomenological corrosion model in a FE framework to predict the loss of stent mechanical integrity using both pitting corrosion and uniform corrosion models. Pitting corrosion attacks led to a non-uniform breaking down of the stent geometry while the uniform model predicted the homogeneity. The experimental results were in harmony with the numerical results using the pitting corrosion model. Grogan et al. [94] further developed an Abaqus the arbitrary Lagrangian–Eulerian adaptive meshing method to model the diffusion-controlled corrosion of a 3-D absorbable metal stent geometry. The assumption of diffusion is based on observations of the formation of stable layers of corrosion products in the body [60, 63, 95] or tissue layers [96] and the known diffusion-controlled corrosion process associated with stable corrosion product layers [97–99] for the long-term device corrosion. Assuming that the corrosion rate is governed by the diffusion of Mg2+ in solution, it is predicted that the mass loss rate from the stent is inversely proportional to the square root of immersion time. It is predicted that the mass loss rate is proportional to the saturation concentration of Mg2+ in solution and is related to the diffusivity of Mg2+ in solution through a power law behavior, where doubling the diffusivity increases the mass loss rate by a factor of ∼1.48. The physical model developed here is computationally efficient and can serve as a useful accompaniment to existing phenomenological models used in the analysis and design of absorbable metal stents.
Summary
Various physiological factors can alter the biodegradation performance and subsequent testing methods in determining the feasibility of Mg-based stents for vascular applications. In this article, we strive to identify and test the relevant microenvironment and parameters in test systems as well as provide a summarization of accurate test methods with the comments on the selection and use. Examples include, understanding the degradation behavior of Mg based on the possible active interactions with a biological system, which will completely alter the environment around a stent, will eventually lead to the knowledge-based design of application-specific stent materials. Mimicking the environment of the vessel repair process is crucial in creating an effective test design and consequently there are a variety of factors that must be considered. The degradation tests should be performed in hydrodynamic conditions that include the consideration of flow conditions in the initial stage and the diffusion condition in the succeeding stage. Composition of the corrosive media, CO2 levels and pH value have a strong influence on the degradation of Mg, whereas O2 and temperature play a less prominent role.
This review highlighted the assortment of degradation assessments in an effort to expose the general development of physiologically relevant in vitro techniques that are needed to give better insight and possible prediction of the in vivo biodegradation behavior of Mg-based stents. Hydrogen collection, mass loss, direct-current polarization, electrochemical impedance spectroscopy and morphology analysis can all be used to not only get a holistic view of the degradation of Mg-based stents, but can also be used to support the accuracy of the corrosion data produced from each individual technique. Additionally, combining these experiments can also simultaneously analyze corrosion types, corrosion morphologies, corrosion products and hydrogen evolution. When designing absorbable Mg implant devices, the implant material has to achieve certain corrosion behaviors to assure proper mechanical behaviors during the entire healing process. Computational model have been developed to serve as a useful accompaniment to existing mechanical testing of absorbable metal stents. More research is being done in the exploration of improving the standardization process for in vitro testing methods, and it is necessary to establish in vitro–in vivo correlation of biodegradation behavior of Mg-based stents.
Acknowledgements
This study was supported by the National Natural Science Foundation of China (Grant No. 81330031); Engineering Research Center for Revolutionizing Metallic Biomaterials (NSF-0812348) from the US National Science Foundation; China Scholarship Council 201207000016; 2013 Cultivation Program for the Excellent Doctoral Dissertation of Southwest Jiaotong University.
Conflict of interest statement. None declared.
References
- 1.Moravej M, Mantovani D. Biodegradable metals for cardiovascular stent application: interests and new opportunities. Int J Mol Sci 2011;12:4250–70. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Mani G, Feldman MD, Patel D, et al. Coronary stents: a materials perspective. Biomaterials 2007;28:1689–710. [DOI] [PubMed] [Google Scholar]
- 3.Erne P, Schier M, Resink TJ. The road to bioabsorbable stents: reaching clinical reality? Cardiovasc Inter Rad 2006;29:11–6. [DOI] [PubMed] [Google Scholar]
- 4.Zheng YF, Gu XN, Witte F. Biodegradable metals. Mater Sci Eng R 2014;77:1–34. [Google Scholar]
- 5.Deng CZ, Radhakrishnan R, Larsen SR, et al. Magnesium alloys for bioabsorbable stents: a feasibility assessment. Magnesium Technol 2011;2011:413–8. [Google Scholar]
- 6.Haude M, Erbel R, Erne P, et al. Safety and performance of the drug-eluting absorbable metal scaffold (DREAMS) in patients with de-novo coronary lesions: 12 month results of the prospective, multicentre, first-in-man BIOSOLVE-I trial. Lancet 2013;381:836–44. [DOI] [PubMed] [Google Scholar]
- 7.Brown DA, Lee EW, Loh CT, et al. A new wave in treatment of vascular occlusive disease: biodegradable stents–clinical experience and scientific principles . J Vasc Int Radiol 2009;20:315–24. [DOI] [PubMed] [Google Scholar]
- 8.Erbel R, Di Mario C, Bartunek J, et al. Temporary scaffolding of coronary arteries with bioabsorbable magnesium stents: a prospective, non-randomised multicentre trial. Lancet 2007;369:1869–75. [DOI] [PubMed] [Google Scholar]
- 9.McMahon CJ, Oslizlok P, Walsh KP. Early restenosis following biodegradable stent implantation in an aortopulmonary collateral of a patient with pulmonary atresia and hypoplastic pulmonary arteries. Catheter Cardio Inte 2007;69:735–8. [DOI] [PubMed] [Google Scholar]
- 10.Schranz D, Zartner P, Michel-Behnke I, et al. Bioabsorbable metal stents for percutaneous treatment of critical recoarctation of the aorta in a newborn. Catheter Cardio Inte 2006;67:671–3. [DOI] [PubMed] [Google Scholar]
- 11.Waksman R, Erbel R, Di Mario C, et al. Early- and long-term intravascular ultrasound and angiographic findings after bioabsorbable magnesium stent implantation in human coronary arteries . JACC Cardio Interv 2009;2:312–20. [DOI] [PubMed] [Google Scholar]
- 12.Zartner P, Cesnjevar R, Singer H, et al. First successful implantation of a biodegradable metal stent into the left pulmonary artery of a preterm baby. Catheter Cardio Inte 2005;66:590–4. [DOI] [PubMed] [Google Scholar]
- 13.Lim GB. Interventional cardiology: DREAMS of a bioabsorbable stent coming true. Nat Rev Cardiol 2013;10:120. [DOI] [PubMed] [Google Scholar]
- 14.Muramatsu T, Garcia-Garcia HM, Serruys PW, et al. Serial 2-dimensional and 3-dimensional optical coherence tomography assessment of overhanging struts of drug-eluting absorbable metal scaffold: “DREAMS” for jailed side branch? JACC Cardio Interv 2014;7:575–6. [DOI] [PubMed] [Google Scholar]
- 15.King AD, Birbilis N, Scully JR. Accurate electrochemical measurement of magnesium corrosion rates; a combined impedance, mass-loss and hydrogen collection study. Electrochim Acta 2014;121:394–406. [Google Scholar]
- 16.Witte F, Fischer J, Nellesen J, et al. In vitro and in vivo corrosion measurements of magnesium alloys. Biomaterials 2006;27:1013–8. [DOI] [PubMed] [Google Scholar]
- 17.Willumeit R, Feyerabend F, Huber N. Magnesium degradation as determined by artificial neural networks. Acta Biomater 2013;9:8722–9. [DOI] [PubMed] [Google Scholar]
- 18.BS EN ISO 10993-5: Biological evaluation of medical devices. Tests for in vitro cytotoxicity 1999. [Google Scholar]
- 19.Kirkland NT, Birbilis N, Staiger MP. Assessing the corrosion of biodegradable magnesium implants: a critical review of current methodologies and their limitations. Acta Biomater 2012;8:925–36. [DOI] [PubMed] [Google Scholar]
- 20.Walker J, Shadanbaz S, Kirkland NT, et al. Magnesium alloys: predicting in vivo corrosion with in vitro immersion testing. J Biomed Mater Res B Appl Biomater 2012;100:1134–41. [DOI] [PubMed] [Google Scholar]
- 21.Cunningham KS, Gotlieb AI. The role of shear stress in the pathogenesis of atherosclerosis. Lab Invest 2005;85:9–23. [DOI] [PubMed] [Google Scholar]
- 22.Fung Y-c. Biomechanics: Circulation. Springer, 1997. [Google Scholar]
- 23.Malek AM, Alper SL, Izumo S. Hemodynamic shear stress and its role in atherosclerosis. JAMA 1999;282:2035–42. [DOI] [PubMed] [Google Scholar]
- 24.Witte F, Hort N, Vogt C, et al. Degradable biomaterials based on magnesium corrosion. Curr Opin Solid State Mater Sci 2008;12:63–72. [Google Scholar]
- 25.Lock JY, Wyatt E, Upadhyayula S, et al. Degradation and antibacterial properties of magnesium alloys in artificial urine for potential resorbable ureteral stent applications. J Biomed Mater Res A 2014;102:781–92. [DOI] [PubMed] [Google Scholar]
- 26.Xue BD, Lu ZH, Yuan GY, et al. Experimental study on stent implantation for congenital tracheal stenosis in infants and children.pdf. Eur Cell Mater 2013;26:62. [Google Scholar]
- 27.Samijo SK, Willigers JM, Barkhuysen R, et al. Wall shear stress in the human common carotid artery as function of age and gender. Cardiovasc Res 1998;39:515–22. [DOI] [PubMed] [Google Scholar]
- 28.Doriot P-A, Dorsaz P-A, Dorsaz L, et al. In-vivo measurements of wall shear stress in human coronary arteries. Coronary Artery Dis 2000;11:495–502. [DOI] [PubMed] [Google Scholar]
- 29.Osinnski JN, Ku DN, Mukundan S, et al. Determination of wall shear stress in the aorta with the use of MR phase velocity mapping. J Magn Reson Im 1995;5:640–7. [DOI] [PubMed] [Google Scholar]
- 30.Kornet L, Hoeks APG, Lambregts J, et al. In the femoral artery bifurcation, differences in mean wall shear stress within subjects are associated with different intima-media thicknesses. Arterioscl Throm Vas 1999;19:2933–9. [DOI] [PubMed] [Google Scholar]
- 31.Levesque J, Hermawan H, Dube D, et al. Design of a pseudo-physiological test bench specific to the development of biodegradable metallic biomaterials. Acta Biomater 2008;4:284–95. [DOI] [PubMed] [Google Scholar]
- 32.Wang J, Giridharana V, Shanov V, et al. Flow induced corrosion behavior of absorbable magnesium-based stent. Acta Biomater 2014;10:5213–23. [DOI] [PubMed] [Google Scholar]
- 33.Dorozhkin SV, Epple M. Biological and medical significance of calcium phosphates. Angew Chem Int Edit 2002;41:3130–46. [DOI] [PubMed] [Google Scholar]
- 34.LaDisa JJF, Guler I, Olson LE, et al. Three-dimensional computational fluid dynamics modeling of alterations in coronary wall shear stress produced by stent implantation . Ann Biomed Eng 2003;31:972–80. [DOI] [PubMed] [Google Scholar]
- 35.Bacabac RG, Smit TH, Cowin SC, et al. Dynamic shear stress in parallel-plate flow chambers. J Biomech 2005;38:159–67. [DOI] [PubMed] [Google Scholar]
- 36.Blutgefäß PE. und Nervennaht (nebst Mittheilung über die Verwendung eines resorbirbaren Metalles in der Chirurgie). Centralblatt für Chirurgie 1901;28. [Google Scholar]
- 37.Yang N, Vafai K. Modeling of low-density lipoprotein (LDL) transport in the artery—effects of hypertension. Int J Heat Mass Trans 2006;49:850–67. [Google Scholar]
- 38.Khakpour M, Vafai K. Critical assessment of arterial transport models. Int J Heat Mass Trans 2008;51:807–22. [Google Scholar]
- 39.Lowe GDO, Barbenel JC. Plasma and blood viscosity. Clin Blood Rheol 1988;1:11–44. [Google Scholar]
- 40.Jang Y, Collins B, Sankar J, et al. Effect of biologically relevant ions on the corrosion products formed on alloy AZ31B: an improved understanding of magnesium corrosion. Acta Biomater 2013;9:8761–70. [DOI] [PubMed] [Google Scholar]
- 41.Liu C, Xin Y, Tian X, et al. Degradation susceptibility of surgical magnesium alloy in artificial biological fluid containing albumin . J Mater Res 2007;22:1806–14. [Google Scholar]
- 42.Rettig R, Virtanen S. Time-dependent electrochemical characterization of the corrosion of a magnesium rare-earth alloy in simulated body fluids. J Biomed Mater Res A 2008;85:167–75. [DOI] [PubMed] [Google Scholar]
- 43.Xin Y, Hu T, Chu PK. In vitro studies of biomedical magnesium alloys in a simulated physiological environment: a review. Acta Biomater 2011;7:1452–9. [DOI] [PubMed] [Google Scholar]
- 44.Zhang K, Zhu L, Fan M. Oxygen, a key factor regulating cell behavior during neurogenesis and cerebral diseases. Front Mol Neurosci 2011;4:5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Valabrègue R, Aubert A, Burger J, et al. Relation between cerebral blood flow and metabolism explained by a model of oxygen exchange . J Cereb Blood Flow Metab 2003;23:536–45. [DOI] [PubMed] [Google Scholar]
- 46.Kiaer T, Grønlund J, Sørensen KH. Intraosseous Pressure and Partial Pressures of Oxygen and Carbon Dioxide in Osteoarthritis, Semin Arthritis Rheu 1989;18(4):57–60. [DOI] [PubMed] [Google Scholar]
- 47.Hiromoto S, Yamamoto A, Maruyama N, et al. Influence of pH and flow on the polarisation behaviour of pure magnesium in borate buffer solutions. Corrosion Sci 2008;50:3561–8. [Google Scholar]
- 48.Seuss F, Seuss S, Turhan MC, Fabry B, Virtanen S. Corrosion of Mg alloy AZ91D in the presence of living cells. J Biomed Mater Res, Part B 2011;99B:276–81. [DOI] [PubMed] [Google Scholar]
- 49.Bakhsheshi-Rad HR, Abdul-Kadir MR, Idris MH, et al. Relationship between the corrosion behavior and the thermal characteristics and microstructure of Mg–0.5Ca–xZn alloys. Corrosion Sci 2012;64:184–97. [Google Scholar]
- 50.Avedesian MM, Baker H. Magnesium and Magnesium Alloys. ASM International, Acta Biomaterialia, Elsevier, 1999. [Google Scholar]
- 51.Song G, Atrens A, St John DH. An hydrogen evolution method for the estimation of the corrosion rate of magnesium alloys. Tms, 2001;1:255–62. [Google Scholar]
- 52.Shaw BA, Wolfe RC. Corrosion of magnesium and magnesium-base alloys . ASM Handbook: Corrosion: Materials 2005;13:205–27. [Google Scholar]
- 53.ASTM G. Standard test method for conducting potentiodynamic polarization resistance measurements, 2009;59–97. [Google Scholar]
- 54.Macdonald JR. Impedance spectroscopy: Models, data fitting, and analysis. Solid State Ionics 2005;176:1961–9. [Google Scholar]
- 55.Ghali E. Corrosion Resistance of Aluminum and Magnesium Alloys: Understanding, Performance, and Testing. John Wiley & Sons; 2010;12. [Google Scholar]
- 56.Makar GL, Kruger J. Corrosion of magnesium. Int Mater Rev 1993;38:138–53. [Google Scholar]
- 57.Song G, Atrens A, John DS, et al. The anodic dissolution of magnesium in chloride and sulphate solutions. Corrosion Sci 1997;39:1981–2004. [Google Scholar]
- 58.Pebere N, Riera C, Dabosi F. Investigation of magnesium corrosion in aerated sodium sulfate solution by electrochemical impedance spectroscopy. Electrochimica Acta 1990;35:555–61. [Google Scholar]
- 59.Doepke A, Kuhlmann J, Guo X, et al. A system for characterizing Mg corrosion in aqueous solutions using electrochemical sensors and impedance spectroscopy. Acta Biomater 2013;9:9211–9. [DOI] [PubMed] [Google Scholar]
- 60.Wittchow E, Adden N, Riedmueller J, et al. Bioresorbable drug-eluting magnesium-alloy scaffold: design and feasibility in a porcine coronary model. EuroIntervention 2013;8:1441–50. [DOI] [PubMed] [Google Scholar]
- 61.Peng Q, Li K, Han Z, et al. Degradable magnesium-based implant materials with anti-inflammatory activity. J Biomed Mater Res A 2013;101:1898–906. [DOI] [PubMed] [Google Scholar]
- 62.Slottow TLP, Pakala R, Okabe T, et al. Optical coherence tomography and intravascular ultrasound imaging of bioabsorbable magnesium stent degradation in porcine coronary arteries. Cardiovasc Revasc Med 2008;9:248–54. [DOI] [PubMed] [Google Scholar]
- 63.Bowen PK, Drelich J, Goldman J. Magnesium in the murine artery: probing the products of corrosion. Acta Biomater 2014;10:1475–83. [DOI] [PubMed] [Google Scholar]
- 64.Willumeit R, Fischer J, Feyerabend F, et al. Chemical surface alteration of biodegradable magnesium exposed to corrosion media. Acta Biomaterialia 2011;7:2704–15. [DOI] [PubMed] [Google Scholar]
- 65.Bertazzo S, Gentleman E, Cloyd KL, et al. Nano-analytical electron microscopy reveals fundamental insights into human cardiovascular tissue calcification. Nat Mater 2013;12:576–83. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 66.Gray-Munro JE, Strong M. A study on the interfacial chemistry of magnesium hydroxide surfaces in aqueous phosphate solutions: influence of Ca2+, Cl- and protein . J Colloid Interface Sci 2013;393:421–8. [DOI] [PubMed] [Google Scholar]
- 67.Wang J, He Y, Maitz MF, et al. A surface-eroding poly (1, 3-trimethylene carbonate) coating for fully biodegradable magnesium-based stent applications: toward better biofunction, biodegradation and biocompatibility. Acta Biomater 2013;9:8678–89. [DOI] [PubMed] [Google Scholar]
- 68.Fischerauer SF, Kraus T, Wu X, et al. In vivo degradation performance of micro-arc-oxidized magnesium implants: a micro-CT study in rats. Acta Biomater 2013;9:5411–20. [DOI] [PubMed] [Google Scholar]
- 69.Kuhlmann J, Bartsch I, Willbold E, et al. Fast escape of hydrogen from gas cavities around corroding magnesium implants. Acta Biomater 2013;9:8714–21. [DOI] [PubMed] [Google Scholar]
- 70.Lamaka SV, Karavai OV, Bastos AC, et al. Monitoring local spatial distribution of Mg2+, pH and ionic currents. Electrochem Commun 2008;10:259–62. [Google Scholar]
- 71.Song G, Atrens A. Understanding magnesium corrosion—a framework for improved alloy performance. Adv Eng Mater 2003;5:837–58. [Google Scholar]
- 72.Ghali E, Dietzel W, Kainer K-U. General and localized corrosion of magnesium alloys: a critical review. J Mater Eng Perform 2004;13:7–23. [Google Scholar]
- 73.Ghali E, Dietzel W, Kainer K-U. Testing of general and localized corrosion of magnesium alloys: a critical review. J Mater Eng Perform 2004;13:517–29. [Google Scholar]
- 74.Li N, Guo C, Wu Y, et al. Comparative study on corrosion behaviour of pure Mg and WE43 alloy in static, stirring and flowing Hank’s solution. Corrosion Eng Sci Technol 2012;47:346–51. [Google Scholar]
- 75.Ballerini G, Bardi U, Bignucolo R, et al. About some corrosion mechanisms of AZ91D magnesium alloy. Corrosion Sci 2005;47:2173–84. [Google Scholar]
- 76.Avedesian MM, Baker H. ASM Specialty Handbook: Magnesium And Magnesium Alloys. ASM International, Acta Biomaterialia, Elsevier; 1999, 274. [Google Scholar]
- 77.Song G. Recent progress in corrosion and protection of magnesium alloys. Adv Eng Mater 2005;7:563–86. [Google Scholar]
- 78.Lafront AM, Zhang W, Jin S, et al. Pitting corrosion of AZ91D and AJ62x magnesium alloys in alkaline chloride medium using electrochemical techniques. Electrochim Acta 2005;51:489–501. [Google Scholar]
- 79.Arrabal R, Mingo B, Pardo A, et al. Pitting corrosion of rheocast A356 aluminium alloy in 3.5wt.% NaCl solution. Corrosion Sci 2013;73:342–55. [Google Scholar]
- 80.Winzer N, Atrens A, Song G, et al. A critical review of the stress corrosion cracking (SCC) of magnesium alloys. Adv Eng Mater 2005;7:659–93. [Google Scholar]
- 81.Bobby Kannan M, Dietzel W, Blawert C, et al. Stress corrosion cracking of rare-earth containing magnesium alloys ZE41, QE22 and Elektron 21 (EV31A) compared with AZ80. Mater SciEngi A 2008;480:529–39. [Google Scholar]
- 82.Cui Z, Li X, Xiao K, et al. Atmospheric corrosion of field-exposed AZ31 magnesium in a tropical marine environment. Corrosion Sci 2013;76:243–56. [Google Scholar]
- 83.Chen J, Wang J, Han E, et al. ESEM observation of the process of hydrogen generation around the micro-droplets forming on AZ91 magnesium alloy. Electrochem Commun 2008;10:577–81. [Google Scholar]
- 84.Jones RH. The influence of hydrogen on the stress-corrosion cracking of low-strength Al-Mg alloys. JOM 2003;55:42–6. [Google Scholar]
- 85.Song RG, Blawert C, Dietzel W, et al. A study on stress corrosion cracking and hydrogen embrittlement of AZ31 magnesium alloy. Mater Sci Eng A 2005;399:308–17. [Google Scholar]
- 86.Scamans G, Holroyd N, Tuck C. The role of magnesium segregation in the intergranular stress corrosion cracking of aluminium alloys. Corrosion Sci 1987;27:329–47. [Google Scholar]
- 87.Aperador W, Delgado A, Franco F. Improved resistance to intergranular corrosion in the AZ31B magnesium alloy by friction stir welding. Int J Electrochem Sci 2013;8:9568–77. [Google Scholar]
- 88.Mueller WD, Lucia Nascimento M, Lorenzo de Mele MF. Critical discussion of the results from different corrosion studies of Mg and Mg alloys for biomaterial applications. Acta Biomater 2010;6:1749–55. [DOI] [PubMed] [Google Scholar]
- 89.Bowen PK, Drelich J, Goldman J. A new in vitro-in vivo correlation for bioabsorbable magnesium stents from mechanical behavior. Mater Sci Eng C 2013;33:5064–70. [DOI] [PubMed] [Google Scholar]
- 90.Wu W, Gastaldi D, Yang K, et al. Finite element analyses for design evaluation of biodegradable magnesium alloy stents in arterial vessels. Mater Sci Eng B 2011;176:1733–40. [Google Scholar]
- 91.Gastaldi D, Sassi V, Petrini L, et al. Continuum damage model for bioresorbable magnesium alloy devices—Application to coronary stents. J Mech Behav Biomed Mater 2011;4:352–65. [DOI] [PubMed] [Google Scholar]
- 92.Wu W, Chen S, Gastaldi D, et al. Experimental data confirm numerical modeling of the degradation process of magnesium alloys stents. Acta Biomater 2013;9:8730–9. [DOI] [PubMed] [Google Scholar]
- 93.Grogan JA, O’Brien BJ, Leen SB, et al. A corrosion model for bioabsorbable metallic stents. Acta Biomater 2011;7:3523–33. [DOI] [PubMed] [Google Scholar]
- 94.Grogan JA, Leen SB, McHugh PE. A physical corrosion model for bioabsorbable metal stents. Acta Biomater 2014;10:2313–22. [DOI] [PubMed] [Google Scholar]
- 95.Witte F, Fischer J, Nellesen J, et al. In vivo corrosion and corrosion protection of magnesium alloy LAE442. Acta Biomater 2010;6:1792–9. [DOI] [PubMed] [Google Scholar]
- 96.Badar M, Lünsdorf H, Evertz F, et al. The formation of an organic coat and the release of corrosion microparticles from metallic magnesium implants. Acta Biomater 2013;9:7580–9. [DOI] [PubMed] [Google Scholar]
- 97.Song Y, Shan D, Chen R, et al. Biodegradable behaviors of AZ31 magnesium alloy in simulated body fluid. Mater Sci Eng C 2009;29:1039–45. [Google Scholar]
- 98.Bender S, Goellner J, Heyn A, et al. Corrosion and corrosion testing of magnesium alloys. Mater Corros 2007;58:977–82. [Google Scholar]
- 99.Feliu S, Jr, Maffiotte C, Galván JC, et al. Atmospheric corrosion of magnesium alloys AZ31 and AZ61 under continuous condensation conditions. Corrosion Sci 2011;53:1865–72. [Google Scholar]