Background and Aims
Indirect calorimetry (IC) allows accurate assessment of resting energy expenditure (REE) from measured oxygen consumption (VO2) and carbon dioxide elimination (VCO2) (1). IC is recommended in critically ill patients for accurate measurement of REE and to guide energy prescriptions (2, 3). However, IC requires specialized equipment and trained personnel, factors that limit its availability. In multicenter studies, IC was reportedly used in only a minority of centers worldwide (4, 5). Estimates of REE and metabolic state utilizing prediction equations do not agree with measured values (6, 7). Inaccurate energy prescriptions can lead to unintended underfeeding or overfeeding in the critically ill patient (8-10). Suboptimal feeding can lead to adverse consequences such as malnutrition, longer length of stay, longer time to wean from ventilator and increased complications. In a recent multicenter study, we observed significantly higher mortality in mechanically ventilated children that received inadequate (less than 66% prescribed) energy intake during their stay in the PICU (5). A simplified and more readily available method to assess energy expenditure in the pediatric intensive care unit (PICU) to facilitate optimal energy prescription is desirable.
We aimed to examine the accuracy of REE values and metabolic classification of patients by a simplified metabolic equation that uses VCO2 measurement alone. We hypothesized that a VCO2-based REE equation would provide a clinically reliable estimate of REE and would be more accurate compared to standard equations that are currently used to estimate REE in this cohort. Independent VCO2 measurement capability is now available in most PICUs as stand-alone monitors or devices integrated into the mechanical ventilator. If the VCO2-REE equation is indeed reliable, then the benefits of metabolic assessment could be extended to a larger population in whom IC is currently not available, and for continuous metabolic measurement during mechanical ventilation.
Methods
Subjects
Children less than 18 years of age who were mechanically ventilated in the multidisciplinary tertiary PICU at 2 academic centers, and in whom IC was performed according to institutional guidelines, were included in the study. Subjects were either on enteral or parenteral nutrition. Continuous enteral feeds or parenteral nutrition was not held for the test. Steady state gas exchange measurements from consecutive IC tests at Boston Children's Hospital (Boston, USA) obtained with the Vmax® Encore (Viasys Healthcare, Loma Linda, CA) were used to derive the simplified metabolic equation that included VCO2 values only. IC measurements from Sophia Children's Hospital (Rotterdam, Netherlands) obtained with the Deltatrac II® (Datex-Ohmeda, Finland) were used as to test the accuracy of REE obtained by the simplified equation in comparison to the estimated energy expenditure by the Schofield equation. IC tests with respiratory quotient (RQ) outside the physiologic range (>1.3 or <0.67) were excluded. Steady state was defined as a period of at least 5 minutes with less than 10% fluctuation in VO2 and VCO2, and less than 5% fluctuation in respiratory quotient (RQ, which is the ratio of VCO2: VO2) (11).
Equation Derivation
The modified Weir equation was used to generate the simplified equation. The mean RQ from the derivation dataset was determined. The VO2 in the modified Weir equation was then replaced with VCO2/RQ, to derive the simplified equation (VCO2-REE) that included only the VCO2 value (12).
Energy Expenditure and Metabolic State Determination by the new equation
Using gas exchange measurements from the validation dataset we compared the accuracy of the simplified equation (VCO2-REE) with the estimated energy expenditure (EEE) by Schofield equation in predicting the actual measured REE (MREE) by IC. Further, in order to clearly illustrate the limitations of the RQ assumption and its impact on accuracy of the VCO2-REE equation we computed the expected error for a range of RQ values.
To examine the effect of dietary composition information on predicting the MREE, RQmacro (based on the ratio of carbohydrate to fat in the diet) was determined for each subject in the validation set (13). REE was now calculated by substituting RQmacro for RQ in the modified Weir equation and utilizing VCO2 measurements (VCO2-RQmacro). Finally, we examined the difference between RQmacro and actual RQ in relation to energy balance.
Metabolic state for each subject in the validation set was determined using the ratio of MREE measured by IC to EEE by the Schofield equation, which incorporates body weight, gender and age in the estimate (14, 15). Subjects were classified as hypometabolic when MREE:EEE <0.9, normometabolic when MREE:EEE = 0.9-1.1, or hypermetabolic when MREE:EEE >1.1. We then examined the accuracy of metabolic classification of each subject in this dataset, using the VCO2-REE:EEE ratio (i.e., substituting the derived REE from VCO2 for the measured REE by IC).
Statistical Analysis
Statistical analyses were performed with Prism (v 5.04, GraphPad® Software Inc. La Jolla, CA). Values are expressed as mean ± standard deviation (SD) or as median and interquartile range (IQR) according to distribution. The Mann-Whitney test was used to test for differences in patient demographics between the derivation and validation datasets. Bland-Altman analysis was used to assess the agreement between estimated (VCO2-REE, VCO2-RQmacro and Schofield) and IC measured REE values. Two-by-two contingency tables were constructed for each of hypo-metabolism and hyper-metabolism diagnostic tests. Fisher's exact test, sensitivity, specificity, positive and negative likelihood ratios and odds ratios were calculated in order to assess the performance of the metabolic classification of subjects using VCO2-REE.
Results
Derivation dataset
Steady state data from 79 cases were included in the analyses. Mean RQ (±SD) of the derivation cohort was 0.89±0.09. Using this mean RQ value, a simplified equation was derived;
-
Modified Weir equation (12)
Assume RQ = 0.89 (from the derivation dataset) and rearrange for VO2 = VCO2/0.89
The VCO2 and RQmacro values were added to the modified Weir equation to derive REE:
Validation dataset
Steady state data from 94 mechanically ventilated children were included in the validation dataset. Daily median (IQR) energy intake was 54.5 (31.9, 68.5) kcal/kg. Patient demographics and gas exchange values from the 2 datasets are depicted in Table 1. The validation and derivation sets included subjects with significantly different characteristics. Subjects in the validation set were younger, smaller and expended more energy per body weight. The RQ was similar between the groups. The test for agreement (Figure 1) between MREE obtained by the IC using the modified Weir equation (utilizing both VCO2 and VO2 values) and the VCO2-REE revealed a mean bias (limits of agreement) of – 0.65% (-14.4 to 13.1%).
Table 1.
| Derivation | Validation | P | |||
|---|---|---|---|---|---|
| Age (y) | 2.34 | (7.83) | 0.46 | (2.13) | <0.05 |
| Height (cm) | 74 | (48) | 61 | (36) | <0.05 |
| Weight (kg) | 10.9 | (21.6) | 5.8 | (8.5) | <0.05 |
| EEE (kcal/day) | 656 | (677) | 313 | (520) | <0.05 |
| MREE (kcal/day) | 518 | (502) | 316 | (461) | <0.05 |
| modREE (kcal/day) | 505 | (484) | 305 | (378) | <0.05 |
| MREE (kcal/kg/day) | 41 | (24) | 55 | (13) | <0.05 |
| modREE (kcal/kg/day) | 43 | (4) | 51 | (13) | <0.05 |
| VCO2 (ml/kg/min) | 5.4 | (3.4) | 6.4 | (1.6) | <0.05 |
| VO2 (ml/kg/min) | 5.9 | (3.6) | 7.1 | (1.6) | <0.05 |
| RQ | 0.89 | (0.09) | 0.89 | (0.10) | ns |
Values are expressed as median (interquartile range). Since the data were not normally distributed, the Mann-Whitney U test was utilized to determine difference between the derivation and validation datasets.
RQ = respiratory quotient, VO2 = volumetric oxygen consumption, VCO2 = volumetric carbon dioxide elimination, EEE = estimated energy expenditure, MREE = measured energy expenditure, modREE = modified REE from the VCO2 based equation.
Figure 1.
Figure 2 shows the expected error introduced by the VCO2-REE equation over a range of actual RQ values. There was no error in predicting MREE when the actual RQ was 0.89 (i.e., matched with the assumed RQ for the equation). For an RQ range of 0.7 to 1.2, the percentage expected error ranged from -20 to +20% respectively. The proportion of patients who fell within this RQ range was 99.9% in the validation cohort. Further, for an RQ range of 0.79 to 1.02, the expected error ranges from -10 to +10%. This RQ range represents 80.9% of validation subjects.
Figure 2.

Median (IQR) value for RQmacro, obtained from dietary information, in the validation dataset was 0.9 (0.85, 0.96). Using the individual RQmacro values for each subject in the modified Weir equation to derive REE, the agreement between MREE and VCO2-RQmacro revealed a mean bias (limits) of -2.0% (-42.9 to 38.9%). The difference between actual RQ and RQmacro was inversely related to the ratio of energy intake to measured energy expenditure. (Figure 3)
Figure 3.

The agreement between MREE and EEE (Schofield) revealed a mean bias (limits) of 0.09% (-40.5 to 40.7%). Figure 1 shows the Bland Altman plots for agreement between the different methods. The mean bias (limits) for agreement between MREE and a) VCO2-REE, b) VCO2-RQmacro and c) EEE by Schofield equation were, -0.6 (-14.4 to 13.1), -2.0 (-42.9 to 38.8) and 0.1 (-40.5 to 40.6) respectively. Using VCO2-REE (in place of MREE), the overall diagnostic accuracy for metabolic classification in the validation set was 86.2% for hypometabolic and 81.9% for hypermetabolic subjects. Sensitivity, specificity and other results are depicted in Table 2.
Table 2.
Diagnostic accuracy of metabolic state determination by the VCO2-based REE equation.
| Hypometabolism | Hypermetabolism | |
|---|---|---|
| Diagnostic accuracy | 0.862 | 0.819 |
| Sensitivity | 1.000 | 0.622 |
| Specificity | 0.833 | 1.000 |
| Positive Predictive Value | 0.55 | 1.00 |
| Negative Predictive Value | 1.00 | 0.74 |
| Positive Likelihood Ratio | 6.00 | - |
| Negative Likelihood Ratio | 0.00 | 0.38 |
VCO2 – volumetric carbon dioxide elimination; REE resting energy expenditure
Discussion
We have described a VCO2-based equation, which was able to predict the measured REE and metabolic state of critically ill children undergoing mechanical ventilation in the PICU. This simplified equation predicted measured REE in the validation dataset with much higher accuracy (narrow limits of agreement) compared to the Schofield equation. Thus, VCO2 based REE prediction may be a superior alternative to standard equations that are currently used to estimate energy expenditure and are frequently inaccurate. Due to a fixed RQ assumed for the simplified equation, there is inherent error in the prediction when actual RQ is higher or lower than 0.89. However, in over 80% of the subjects in the validation cohort, the difference between actual and predicted REE was +/- 10%. Furthermore, VCO2-REE helped classify patients according to their metabolic state with high sensitivity and specificity. Bedside VCO2 monitoring is increasingly available in the PICU as devices measuring VCO2 are incorporated into mechanical ventilators or as stand-alone monitors (16). Therefore, the results of our study indicate a unique opportunity for continuous metabolic assessment in the PICU population.
A reliable VCO2 based technique could make metabolic assessment available for a large number of mechanically ventilated patients who are currently deprived of this testing in centers without an IC. VCO2 has been previously used to derive REE in critically ill children with the use of a Douglas bag to collect expired air for VCO2 measurements (17). This method however is not practical and is typically only feasible in a research setting. The results of our study suggest the possibility of a bedside technique for metabolic assessment based on VCO2 measurements. With the advent of stand-alone VCO2 monitors employing infrared technology or modern ventilators incorporating VCO2 modules, minute-to-minute VCO2 values will be increasingly available in most PICUs. In steady state conditions, VCO2 reflects CO2 production at the cellular level and is dependent on the metabolic activity of the tissues and the composition of the nutrient mixture delivered and utilized for energy metabolism. Any change in minute ventilation significantly alters VCO2 until a new steady state is reached. For measured VCO2 to reflect the metabolic production of CO2, enough time should be allowed for the body pool of CO2 to reach equilibrium. Future studies will need to define steady state criteria for VCO2 values obtained by stand-alone monitors.
An assumed RQ value was essential when attempting to derive REE from VCO2 alone. It allowed a simplification of the Weir equation where we substituted the VO2 value by the assumed RQ and measured VCO2. Table 3 shows RQ values reported in critically ill infants and children with a variety of illnesses. The mean RQ for this group of studies is 0.87 and a majority of patients had values within a narrow range. Hence, the assumed RQ utilized in the equation reflects the value reported in literature. Furthermore, the mean RQ at the 2 institutions in our study were nearly identical, 0.89. Hence, we used a universal RQ value of 0.89 to derive our simplified equation. Clearly this is an oversimplification and ignores variation in RQ between individual patients. RQ reflects substrate handling, which may be altered during critical illness and actual values may be distinct from the assumed 0.89 in some patients. Figure 2 shows the degree of the error in predicting the MREE of individual patient, based on the deviation of the actual RQ from the assumed RQ. For a majority of the patients in the validation set with RQ ranging from 0.79 to 1.02, this error was approximately 10%.
Table 3. Respiratory quotient (RQ) values reported in critically ill infants and children (Total n = 638).
| Author | Year | RQ | SD | Setting | n |
|---|---|---|---|---|---|
| Mehta (3) | 2011 | 0.94 | Range = 0.79 –1.28 | PICU | 29 |
| Hulst (22) | 2005 | NA | NA | PICU | 98 |
| Briassoulis (23) | 2000 | NA | NA | PICU | 37 |
| Smallwood (24) | 2012 | 0.9 | 0.11 | PICU | 34 |
| Briassoulis (25) | 2010 | 0.89 | 0.01 | PICU | 77 |
| Verhoeven (26) | 1998 | 0.89 | SEM = 0.01 | Ventilator | 50 |
| Joosten (27) | 1999 | 0.87 | 0.07 | PICU | 36 |
| White (28) | 2000 | 0.85 | 0.07 | Ventilator | 100 |
| Vanderkuip (21) | 2004 | 0.88 | Range = 0.71 –1.1 | Ventilator | 14 |
| Framson (29) | 2007 | 0.83 | 0.12 | PICU | 41 |
| Framson (29) | 2007 | 0.83 | 0.07 | PICU | 36 |
| Framson(29) | 2007 | 0.88 | 0.16 | PICU | 17 |
| Vazquez-Martinez (30) | 2004 | 0.78 | 0.01 | PICU | 43 |
| Mehta (10) | 2012 | 0.85 | 0.06 | CICU | 26 |
Mean RQ for all reported values = 0.87
Since RQ value is dependent on macronutrient composition (see Table 4), knowledge of the fuel partitioning (ratio of CHO to Fat) in the diet might help obtain a value of RQ that is closer to the actual value. This is referred to as the RQmacro or the food quotient (FQ). If we used the RQmacro in the equation and not the preset RQ of 0.89 we found a mean bias for agreement between MREE and VCO2-REE of -2.0%, but with wide limits. Subjects with disagreement between the 2 methods demonstrated intake to MREE ratios very low (<0.6) or very high (>1.2), which limits the utility of RQmacro as a surrogate for true RQ. Indeed, RQmacro may not always be accurate in predicting actual RQ as it is impacted by a complex interplay between fuel oxidation and other clinical factors (18). In a previous study, we have shown that RQ and RQmacro were approximated in less than half the subjects (13). The difference between RQmacro and RQ in the validation set was related to the energy balance (intake compared to MREE) with a higher difference seen in cases with a positive energy balance. Any method that utilizes isolated VCO2 values to derive REE must also account for the energy equivalents of CO2, (energy released/L CO2 produced; EeqCO2) (19). The EeqCO2 may vary by up to 20% in patients on artificial nutrition. Table 4 shows RQ and energy equivalents values of O2 and CO2 for different macronutrients (20). The impact of accurate energy balance and macronutrient composition on REE derived from VCO2 values must be taken into account and examined in future prospective studies.
TABLE 4. Respiratory quotient (RQ) and energy equivalents (O2 and CO2) for individual macronutrients.
| RQ | Energy equivalents (kcal/L) | ||
|---|---|---|---|
| O2 | CO2 | ||
| Fat | 0.71 | 4.68 | 4.60 |
| Protein | 0.83 | 4.66 | 5.57 |
| Carbohydrate | 1.00 | 5.04 | 5.05 |
| Alcohol | 0.67 | 4.86 | 7.23 |
Future studies must validate the VCO2-REE equation by using a stand-alone VCO2 monitor against IC-derived REE or a stable isotope technique for REE. Different IC devices were used at the 2 sites in our study. Vmax Encore® and the Deltatrac® have been extensively used for clinical metabolic testing in children. Both centers have a long history of clinical application and published research with these devices. Furthermore, uniform procedures were used for metabolic testing, steady state criteria, data handling and classification of metabolic states. The mean age of subjects in the 2 cohorts was significantly different. However, the application of the equation derived from one dataset and its successful applucation in another dataset with distinct characteristics is reassuring and speaks to the generalizability of our results.
Conclusion
Accurate assessment of REE and metabolic state is essential to optimize nutrient intake in critically ill patients. A new metabolic equation, using real-time, bedside, steady state VCO2 measurements, allowed assessment of energy expenditure in mechanically ventilated children. REE derived by this simplified method is superior to standard equations and permits accurate metabolic classification of patients. This equation could be used in mechanically ventilated patients at centers where IC is not available, or as a screening test in centers with limited IC resources. The availability of 24-hour continuous VCO2-based assessment would be a significant advantage over the snapshot assessment by IC. Stand-alone VCO2 monitors are routinely available in most PICUs and can be obtained and maintained at a fraction of the cost of an IC. This method may be appropriate for resource-limited environments, extending the benefits of metabolic testing while limiting expenses and burden on personnel. Further evaluation of this equation with other CO2 monitoring devices should be conducted.
Acknowledgments
Funding Sources: None
Abbreviations
- VCO2
volumetric carbon dioxide elimination
- VO2
volumetric oxygen consumption
- RQ
respiratory quotient
- REE
resting energy expenditure
Footnotes
Statement of Authorship: “NM conceived and designed the study. NM, KJ and JH collected the data. NM and CS carried out the initial data analyses and drafted the manuscript. All authors participated in the design of the study, the statistical analyses and the drafting and approval of the manuscript”.
Conflict of Interest: None
References
- 1.Brandi LS, Bertolini R, Santini L, Cavani S. Effects of ventilator resetting on indirect calorimetry measurement in the critically ill surgical patient. Critical care medicine. 1999 Mar;27(3):531–9. doi: 10.1097/00003246-199903000-00030. Epub 1999/04/13. eng. [DOI] [PubMed] [Google Scholar]
- 2.Mehta NM, Compher C. A.S.P.E.N. Clinical Guidelines: nutrition support of the critically ill child. JPEN J Parenter Enteral Nutr. 2009 May-Jun;33(3):260–76. doi: 10.1177/0148607109333114. Epub 2009/04/29. eng. [DOI] [PubMed] [Google Scholar]
- 3.Martindale RG, McClave SA, Vanek VW, McCarthy M, Roberts P, Taylor B, et al. Guidelines for the provision and assessment of nutrition support therapy in the adult critically ill patient: Society of Critical Care Medicine and American Society for Parenteral and Enteral Nutrition: Executive Summary. Critical care medicine. 2009 May;37(5):1757–61. doi: 10.1097/CCM.0b013e3181a40116. Epub 2009/04/18. eng. [DOI] [PubMed] [Google Scholar]
- 4.van der Kuip M, Oosterveld MJ, van Bokhorst-de van der Schueren MA, de Meer K, Lafeber HN, Gemke RJ. Nutritional support in 111 pediatric intensive care units: a European survey. Intensive Care Med. 2004 Sep;30(9):1807–13. doi: 10.1007/s00134-004-2356-8. eng. [DOI] [PubMed] [Google Scholar]
- 5.Mehta NM, Bechard LJ, Cahill N, Wang M, Day A, Duggan CP, et al. Nutritional practices and their relationship to clinical outcomes in critically ill children--an international multicenter cohort study*. Critical care medicine. 2012 Jul;40(7):2204–11. doi: 10.1097/CCM.0b013e31824e18a8. Epub 2012/05/09. eng. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Suman OE, Mlcak RP, Chinkes DL, Herndon DN. Resting energy expenditure in severely burned children: analysis of agreement between indirect calorimetry and prediction equations using the Bland-Altman method. Burns. 2006 May;32(3):335–42. doi: 10.1016/j.burns.2005.10.023. [DOI] [PubMed] [Google Scholar]
- 7.Taylor RM, Cheeseman P, Preedy V, Baker AJ, Grimble G. Can energy expenditure be predicted in critically ill children? Pediatr Crit Care Med. 2003 Apr;4(2):176–80. doi: 10.1097/01.PCC.0000059425.09149.D8. eng. [DOI] [PubMed] [Google Scholar]
- 8.Mehta NM, Bechard LJ, Dolan M, Ariagno K, Jiang H, Duggan C. Energy imbalance and the risk of overfeeding in critically ill children. Pediatr Crit Care Med. 2011 Oct 21; doi: 10.1097/PCC.0b013e3181fe279c. Epub 2010/10/27. Eng. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Mehta NM, Bechard LJ, Leavitt K, Duggan C. Cumulative energy imbalance in the pediatric intensive care unit: role of targeted indirect calorimetry. JPEN J Parenter Enteral Nutr. 2009 May-Jun;33(3):336–44. doi: 10.1177/0148607108325249. Epub 2009/01/08. eng. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Hulst JM, Joosten KF, Tibboel D, van Goudoever JB. Causes and consequences of inadequate substrate supply to pediatric ICU patients. Curr Opin Clin Nutr Metab Care. 2006 May;9(3):297–303. doi: 10.1097/01.mco.0000222115.91783.71. Epub 2006/04/12. eng. [DOI] [PubMed] [Google Scholar]
- 11.McClave SA, Spain DA, Skolnick JL, Lowen CC, Kieber MJ, Wickerham PS, et al. Achievement of steady state optimizes results when performing indirect calorimetry. JPEN J Parenter Enteral Nutr. 2003 Jan-Feb;27(1):16–20. doi: 10.1177/014860710302700116. Epub 2003/01/29. eng. [DOI] [PubMed] [Google Scholar]
- 12.Weir JB. New methods for calculating metabolic rate with special reference to protein metabolism. 1949. Nutrition. 1990 May-Jun;6(3):213–21. Epub 1990/05/01. eng. [PubMed] [Google Scholar]
- 13.Joosten KF, Verhoeven JJ, Hazelzet JA. Energy expenditure and substrate utilization in mechanically ventilated children. Nutrition. 1999 Jun;15(6):444–8. doi: 10.1016/s0899-9007(99)00081-7. Epub 1999/06/23. eng. [DOI] [PubMed] [Google Scholar]
- 14.Schofield WN. Predicting basal metabolic rate, new standards and review of previous work. Hum Nutr Clin Nutr. 1985;39(Suppl 1):5–41. [PubMed] [Google Scholar]
- 15.Coss-Bu JA, Klish WJ, Walding D, Stein F, Smith EO, Jefferson LS. Energy metabolism, nitrogen balance, and substrate utilization in critically ill children. Am J Clin Nutr. 2001 Nov;74(5):664–9. doi: 10.1093/ajcn/74.5.664. [DOI] [PubMed] [Google Scholar]
- 16.Walsh BK, Crotwell DN, Restrepo RD. Capnography/Capnometry during mechanical ventilation: 2011. Respir Care. 2011 Apr;56(4):503–9. doi: 10.4187/respcare.01175. Epub 2011/01/25. eng. [DOI] [PubMed] [Google Scholar]
- 17.van der Kuip M, de Meer K, Oosterveld MJ, Lafeber HN, Gemke RJ. Simple and accurate assessment of energy expenditure in ventilated paediatric intensive care patients. Clinical nutrition (Edinburgh, Scotland) 2004 Aug;23(4):657–63. doi: 10.1016/j.clnu.2003.11.001. eng. [DOI] [PubMed] [Google Scholar]
- 18.Westerterp KR. Food quotient, respiratory quotient, and energy balance. Am J Clin Nutr. 1993 May;57(5 Suppl):759S–64S. doi: 10.1093/ajcn/57.5.759S. discussion 64S-65S. Epub 1993/05/01. eng. [DOI] [PubMed] [Google Scholar]
- 19.Elia M. Energy equivalents of CO2 and their importance in assessing energy expenditure when using tracer techniques. Am J Physiol. 1991 Jan;260(1 Pt 1):E75–88. doi: 10.1152/ajpendo.1991.260.1.E75. Epub 1991/01/01. eng. [DOI] [PubMed] [Google Scholar]
- 20.Livesey G, Elia M. Estimation of energy expenditure, net carbohydrate utilization, and net fat oxidation and synthesis by indirect calorimetry: evaluation of errors with special reference to the detailed composition of fuels. Am J Clin Nutr. 1988 Apr;47(4):608–28. doi: 10.1093/ajcn/47.4.608. Epub 1988/04/01. eng. [DOI] [PubMed] [Google Scholar]