Skip to main content
The Journal of ExtraCorporeal Technology logoLink to The Journal of ExtraCorporeal Technology
letter
. 2010 Sep;42(3):242.

Letters to the Editor

PMCID: PMC4679967  PMID: 21114230
letter J Extra Corpor Technol. 2010 Sep;42(3):242.

Evaluation of Quadrox-i® Adult Hollow Fiber Oxygenator with Integrated Arterial Filter

Antoine P Simons 1, Patrick W Weerwind 1

To the Editor,

In your June issue of the Journal of ExtraCorporeal Technology we read the interesting article of Guan et al. in which they investigated the air handling capacity of two different oxygenators used in cardiopulmonary bypass (1).

We congratulate the authors for their results, showing that the Quadrox-i adult oxygenator had better gaseous micro-emboli handling than the standard Quadrox adult oxygenator, while both artificial lungs had similar trans-membrane pressure drops. Nevertheless, we want to comment on the data presented by the authors. Table 2 shows both emboli count and volume. When comparing baseline values, the reader notices that both venous emboli count and volume differ significantly. To be more specific, the baseline count values for the Quadrox are nearly three times higher and the baseline volume values are nearly twice those measured in the Quadrox-i group. As the venous bubble measurement is located before the pump and the oxygenator, one expects the measured total venous bubble volume to be equal for all groups at the three flow rates, respectively. Moreover, intergroup difference in baseline is expected to bias the post-oxygenator emboli count and might explain the study outcome, which is in favor of the Quadrox-i.

Applying their protocol, the authors were able to separately investigate the impact of the integrated arterial filter on gaseous emboli removal. However, since the authors state both the standalone Quart filter and the integrated filter to be similar in terms of the filter component, would simply investigating the emboli removal capabilities of a standalone Quart filter in a mock circulation not be the same?

Part of the volumetric data on bubbles presented in Table 2 contains measured volumes larger than .5 mL, which is beyond the range of the bubble counting device. The discussion section hypothesizes that the centrifugal pump induces bubble break-up which aggravates the creation of smaller bubbles at higher flows and rpm. As a result, bubbles may become too small to detect, and the presented number of bubbles counted may be misleading. Indeed, the potential of centrifugal pumps to break up air emboli into even smaller bubbles has been reported in literature (2). Table 3 presents bubbles in three size ranges of which one corresponds to bubbles of 0–20 microns, thus including the bubbles that cannot be detected since they are too small to be measured. Would this not weaken the conclusion of the study?

In the literature currently available and also in this study, air handling of cardiopulmonary bypass components is tested by infusing air into mock circulations. In an ongoing clinical study at our hospital, we found that microembolization occurs when kinetic venous assisted drainage as used in minimized bypass circuits is applied, and found a positive correlation between sub-atmospheric pressure and the number of gaseous emboli counted. Compared to the bubble data presented by the authors, we found similar diameters of gaseous bubbles. This implies that air injection into mock circulations can be a reliable method to assess air handling capacity of bypass components under comparable conditions as found in clinical routine.

REFERENCES

  • 1.Guan Y, Su X, McCoach R, Wise R, Kunselman A, Ündar A.. Evaluation of Quadrox-i adult hollow fiber oxygenator with integrated arterial filter. J Extra Corpor Technol. 2010;42:134–8. [PMC free article] [PubMed] [Google Scholar]
  • 2.Rider SP, Simon LV, Rice BJ, Poulton CC.. Assisted venous drainage, venous air, and gaseous microemboli transmission into the arterial line: An in-vitro study. J Extra Corpor Technol. 1998;30:160–5. [PubMed] [Google Scholar]
J Extra Corpor Technol. 2010 Sep;42(3):243.

Response: Evaluation of Quadrox-i® Adult Hollow Fiber Oxygenator with Integrated Arterial Filter

Yulong Guan 1, Xiaowei Su 1, Robert McCoach 1, Robert Wise 1, Allen Kunselman 2, Akif Ündar 3

To the Editor,

We greatly appreciate Drs. Simons and Weerwind’s interest and comments on our recent article entitled Evaluation of Quadrox-i ® Adult Hollow Fiber Oxygenator with Integrated Arterial Filter (1).

Regarding comments on different emboli counts and volume at the venous site between the two groups, we would clarify that the duration of bolus air injection was 15 seconds and the duration of data collection was 5 minutes. As we stated in the discussion section of the article, in this particular in-vitro design, we specifically selected the amount of air and duration of data collection based on our previous experiences with this novel device (24). Once we injected the bolus air in the venous site, only a fraction of air can be captured at the first pass through the circuit via a membrane oxygenator, tubing, and the pseudo patient. It takes about 3–4 minutes to clear all the air after several passes through the closed circuit. This is the rationale behind the delay of 5 minutes prior to subsequent bolus air injection. Since we have identical circuit components and conditions (pressures, flows, etc.), it is logical to have less emboli and volume at the venous site when the purge line of the integrated arterial filter is open because of emboli capture with the open purge line during 5 minutes of circulation. Not all the bolus air can be captured during the first pass after injection.

We have not investigated the impact of a separate arterial filter on gaseous microemboli. To make that particular comparison, experimental conditions (flow rate, circuit pressure, etc.) must be identical but this is not possible due to significant diverted flow (stolen blood flow) via the arterial filter purge line. Even though the stolen blood flow will be less in the adult model compared to the neonatal circuit, this will be a significant limitation across groups because one of the major factors for microemboli delivery is the flow rate (5).

The Emboli Detection and Classification Quantifier is the only FDA approved device in the United States that can measure and classify air emboli as small as 10 microns compared to 40 microns with traditional transcranial Doppler devices (6). We have also compared roller pumps versus centrifugal pumps in addition to evaluating several oxygenators in terms of microemboli capture under cardiopulmonary bypass conditions (3, 4).

We are pleased to learn that Drs. Simons and Weerwind have similar clinical experiences in terms of measured microemboli diameter compared to our study. Recently we also published our clinical results (7) and more clinical data are under preparation for publication.

Once again, we thank Drs. Simons and Weerwind for their comments and we hope that we were adequately able to respond to their comments.

REFERENCES

  • 1.Guan Y, Su X, McCoach R, Wise R, Kunselman A, Ündar A.. Evaluation of Quadrox-i adult hollow fiber oxygenator with integrated arterial filter. J Extra Corpor Technol. 2010;42:134–8. [PMC free article] [PubMed] [Google Scholar]
  • 2.Wang S, Ündar A.. Vacuum-assisted venous drainage and gaseous microemboli in cardiopulmonary bypass. J Extra Corpor Technol. 2008;40:249–56. [PMC free article] [PubMed] [Google Scholar]
  • 3.Guan Y, Palanzo D, Kunselman AR, Ündar A.. Evaluation of membrane oxygenators and reservoirs in terms of capturing gaseous microemboli and pressure drops. Artif Organs. 2009;33:1037–43. [DOI] [PubMed] [Google Scholar]
  • 4.Yee S, Qiu F, Su XW, et al. Evaluation of HL-20 roller pump and Rotaflow centrifugal pump on perfusion quality and gaseous micro-emboli delivery. Artif Organs. 2010; (in press). [DOI] [PubMed] [Google Scholar]
  • 5.Wang S, Miller A, Myers JL, Ündar A.. “Stolen” blood flow:The effects of an open arterial filter purge line in a simulated neonatal CPB model. ASAIO J. 2008;54:432–5. [DOI] [PubMed] [Google Scholar]
  • 6.Lynch JE, Riley JB.. Microemboli detection on extracorporeal bypass circuits. Perfusion. 2008;23:23–32. [DOI] [PubMed] [Google Scholar]
  • 7.Wang S, Woitas K, Clark JB, Myers JL, Ündar A.. Clinical real-time monitoring of gaseous microemboli in pediatric cardiopulmonary bypass. Artif Organs. 2009;33:1026–30. [DOI] [PubMed] [Google Scholar]

Articles from The Journal of Extra-corporeal Technology are provided here courtesy of EDP Sciences

RESOURCES