Abstract:
Patient blood volume impacts most facets of perfusion care, including volume management, transfusion practices, and pharmacologic interventions Unfortunately, there is a wide variability in individual blood volumes, and experimental measurement is not practical in the clinical environment. The purpose of this study was to evaluate a mathematical algorithm for estimating individual blood volume. After institutional review board approval, volumetric and transfusion data were prospectively collected for 165 patients and applied to a series of calculations. The resultant blood volume estimate (BVE) was used to predict the first and last bypass hematocrit. The estimated hematocrits using both BVE and 65 mL/kg were compared with measured hematocrits using the Pearson moment correlation coefficient and the Bland Altman measures of accuracy and precision. There was a wide range of BVE (minimum, 35 mL/kg; mean ± SD, 64 ± 22 mL/kg; maximum, 129 mL/kg). Using BVE, the estimated hematocrit was similar to the measured first (24.7 ± 6.4% vs. 24.5 ± 6.2%, r = 0.9884, p > .05) and last (24.5 ± 5.9% vs. 25.1 ± 5.7%, r = 0.9001, p > .05) bypass hematocrit. Using 65 mL/kg resulted in a larger difference between estimated and measured hematocrits for the first (25.6 ± 4.5% vs. 24.5 ± 6.2%, r = 0.6885, p = .030) and last (23.8 ± 3.6% vs. 25.1 ± 5.7%, r = 0.5990, p = .001) bypass hematocrits. Compared with using 65 mL/kg for blood volume, the BVE allowed for a more precise estimated hematocrit during CPB.
Keywords: blood volume, cardiopulmonary bypass, hematocrit
Patient blood volume impacts most facets of perfusion care, including volume management, hemodynamic interventions, transfusion practices, and pharmacologic interventions. Unfortunately, there is a wide variability in individual blood volumes, and extremes (30–130 mL/kg) will be encountered in clinical practice (1–3). If an extremely low blood volume is encountered without identification, weight-based pharmaceutic concentrations will be higher than expected, and the effects of hemodilution will be more pronounced. An extremely high blood volume would be expected to have the opposite effect, potentially leading to under-heparinization and inappropriate transfusion (i.e., low hemoglobin treated with transfusion instead of volume removal through autologous priming).
Indeed, early knowledge of a patient’s blood volume would empower the clinician to make more informed clinical decisions. Unfortunately, experimental measurements are not applicable to the clinical environment. The methods require either radioactive labeling or the measurement of plasma volume by injecting indocyanine green, subsequently measured by a specialty spectrophotometer, and measurement of red cell volume by labeling autologous red cells with sodium fluorescein and subsequent measurement by flow cytometry (3). Neither option could be consider viable in the clinical arena or justifiable for the information obtained. Less demanding techniques using new technology are being developed, but they are invasive and only moderately accurate and precise (4).
However, it is possible to estimate blood volume from the change in hematocrit during carefully controlled acute, normovolemic hemodilution (2). The estimated blood volume obtained by the Jacobs method correlated well with the measured values (y = 0.912x + 560; r = 0.80) (2). The Jacobs mathematical model was modified significantly and includes additional variables to allow a more dynamic, less time-dependent blood volume estimate. The purpose of this study was to evaluate the capacity of the blood volume estimate to predict hematocrit changes.
MATERIALS AND METHODS
An estimate for the post-induction blood volume [blood volume estimate (BVE); baseline (BVE(B))] was developed by applying a conversion factor (F) to a reference, or ideal, blood volume (BVi):
Other groups have used BVi (ideal blood volume), and it is calculated based on an ideal body surface area (2).
and
To calculate BVE(B), a conversion factor (CF) is applied to the BVi. The conversion factor is calculated based on the change in hematocrit at two time points in relation to what the change would be with the BVi:
where
HCTi(Pre)A = Ideal Hematocrit Before Change A;
HCTi(Post)A = Ideal Hematocrit After Change A;
HCTm(Pre)A = Measured Hematocrit Before Change A;
HCTm(Post)A = Measured Hematocrit After Change A;
Change A is characterized by a blood volume change (ΔBVA) and a red cell volume change (ΔRCVA):
where
and
and
then
where
BVE(B) can now be calculated:
or
or
The BVE(B) value gives the static blood volume of the patient at the first hematocrit sample; for this study, the sample was drawn immediately after the induction of anesthesia.
The BVE(B) was used to predict the first and last CPB hematocrit. Two additional periods were defined, change B (beginning when change A ended and ending when the first CPB hematocrit was taken) and change C (beginning when change A ended and ending when the last CPB was taken).
The first CPB hematocrit (Hcte(first)) and the last CPB hematocrit (Hcte(last)) were estimated by:
and
Both ΔRCVB and ΔRCVC were calculated in the same fashion as ΔRCVA, and ΔBVB and ΔBVC were calculated in a similar fashion as ΔBVA. For purposes of comparison, the same calculations were performed by substituting 65 mL/kg for the BVE(B).
All formulas were entered into a Microsoft Excel Workbook, and the variable data were collected in a prospective fashion on 165 adult (18–89 years old) patients undergoing cardiopulmonary bypass. The Hcte(first) and Hcte(last) were calculated using both BVE(B) and 65 mL/kg. The results using BVE(B) were compared with the results using 65 mL/kg using the Pearson moment correlation and the Bland Altman bias and limits of agreement. Significance was accepted at p < .05.
RESULTS
There was a wide range of BVE (minimum, 35 mL/kg; mean ± SD, 64 ± 22 mL/kg; maximum, 129 mL/kg). Using BVE, the estimated hematocrit was similar to the measured first (24.7 ± 6.4% vs. 24.5 ± 6.2%, r = 0.9884, p > .05) and last (24.5 ± 5.9% vs. 25.1 ± 5.7%, r = 0.9001,p > .05) bypass hematocrits (Figure 1). Using 65 mL/kg resulted in larger differences and less correlation between estimated and measured hematocrit for the first (25.6 ± 4.5% vs. 24.5 ± 6.2%, r = 0.6855,p = .030) and last (23.8 ± 3.6% vs. 25.1 ± 5.7%, r = 0.5990, p = .001) bypass hematocrits (Figure 1). The BVE also provided more precise prediction of first CPB hematocrit (limits of agreement: 1.0% vs. 3.4%) and last CPB hematocrit (limits of agreement: 2.4% vs. 4.0%) than the use of 65 mL/kg. Both blood volume estimates provided similar accuracy at the first CPB hematocrit (bias: −0.4% vs. 0.4%) and last CPB hematocrit (bias: 1.2% vs. 1.9%)
Figure 1.
Estimate vs. measured hematocrit.
DISCUSSION
The BVE was used to calculate changes in hematocrit during CPB. However, the use of this value extends beyond its predictive capacity for hematocrit changes. For example, most pharmacologic interventions, including anti-coagulation, are based on volume. In the absence of firm BVEs, weight is used as a reflection of blood volume. Unfortunately, weight seems to be a poor reflection of blood volume, because the estimated blood volumes (BVE) ranged from 35 to 129 mL/kg in this study. These were not unique findings, and other groups that have experimentally measured blood volumes have found similar diversity (1–3). Unfortunately, in the absence of valid estimates of blood volume, weight continues to be used in the majority of these calculations. Perhaps the principle hazard of this strategy is best illustrated by example: in this case anti-coagulation with heparin for a 70-kg patient. The standard heparin-dosing guidelines for CPB are 3 units heparin/mL of circulating blood volume. Using 65 mL/kg and 1700-mL prime volume, a target dose of 18,750 units would be calculated. However, if this patient was a high blood volume patient (120 mL/kg), the circulating heparin concentration would only be 1.9 units/mL.
An accurate BVE could be useful beyond pharmacologic interventions. Discrepancies between measured and predicted hematocrit values could alert the perfusionist to undetected blood loss, and excessive extravasation of fluid, and identify the need for volume replacement interventions. Likewise, dilutional coagulopathy could be more readily discriminated, allowing earlier detection and intervention.
Estimating blood volume by this method has two categories of limitations. The first group could be considered logistic or external limitations. For example, the formulas used to derive BVE and predict CPB hematocrits do not lend themselves to hand calculation or written charting methods. In the absence of automated data collection and calculation, the algorithms used would be too cumbersome and time consuming to be of clinical value. Furthermore, accurate capture of volume management values is essential to yield predictive results. The second category of limitations is internal and is a reflection of using a modified one compartment model in a biological system. The mobilization of capillary-based plasma layers and extravasation of fluid are not measured, but the effects are expected to be differential between individual patients. Conditions that alter vascular permeability or tone are not used in the calculations presented here but would be expected to diminish the functional capacity of them.
However, with modern computer technology and the national shift toward automated charting, a valid BVE may become a valuable clinical tool, used to guide volume replacement strategies, transfusion decisions, and pharmacologic interventions. The ability of the BVE to predict CPB hematocrit suggests that it is an accurate reflection of the patient’s actual blood volume. The BVE did not require additional equipment, invasive techniques, or added laboratory samples. Rather, the estimate was derived from existing test data and patient information.
In addition, standard corrections were used for cell saver efficiency and distribution volumes. The other internal limitations seem to be systematic and did not impact the precision of predicted hematocrits. Indeed, aberhent results may indicate drastic deviation from normal conditions, such as excessive vascular permeability or unexplained blood loss. For example, if the predicted bypass hematocrit value is dramatically lower than the measured value, this may be evidence of large volumes accumulating in the third physiologic space.
This study was limited by design. It was designed to determine whether the range of variables currently collected could be used to predict BV-dependent changes in hematocrit. The method should be validated with double radiolabel measurement techniques. However, compared with using 65 mL/kg, using the blood volume resulted in more precise bypass hematocrit prediction that correlated better with measured results.
REFERENCES
- 1.Pearson TC, Guthrie DL, Simpson J, et al. Interpretation of measured red cell mass and plasma volume in adults: Expert Panel on Radionuclides of the International Council for Standardization in Haematology. Br J Haematol. 1995;89:748–56. [DOI] [PubMed] [Google Scholar]
- 2.Jacob M, Bruegger D, Conzen P, et al. Development and validation of a mathematical algorithm for quantifying preoperative blood volume by means of the decrease in hematocrit resulting from acute normovolemic hemodilution. Transfusion. 2005;45:562–71. [DOI] [PubMed] [Google Scholar]
- 3.Haller M, Dressel H, Theil M, et al. Determination of red-cell volume with a nonradioactive marker (sodium fluorescein) in humans using flow cytometry. Infusions Ther Transfusion Med. 1997;24:132–7. [Google Scholar]
- 4.Marx G, Schuerholz T, Pedder S, et al. Blood volume measurements using an integrated fiberoptic monitoring system in a porcine septic shock model. Crit Care Med. 2006;34:1483–8. [DOI] [PubMed] [Google Scholar]