Abstract
Novel MCO high-flux membranes for hemodialysis have been developed with optimized permeability, allowing for filtration close to that of the natural kidney. A comprehensive in vitro characterization of the membrane properties by dextran filtration is presented. The sieving profile of pristine membranes, as well as that of membranes exposed to blood for 40 minutes, are described. The effective pore size (Stokes-Einstein radius) was estimated from filtration experiments before and after blood exposure, and results were compared to hydrodynamic radii of middle and large uremic toxins and essential proteins. The results indicate that the tailored pore sizes of the MCO membranes promote removal of large toxins while ensuring the retention of albumin.
Patients with end stage renal disease on hemodialysis have an increased mortality risk compared to the general population, which indicates that current renal replacement therapy requires improvements that translate into better patients’ outcomes. One of the unmet needs in hemodialysis is the adequate removal of uremic toxins over a broad molecular weight range. As synthetic membranes are less selective than the glomerular membrane, current hemodialysis membranes do not remove higher molecular weight toxins appropriately1. As a consequence, patients on hemodialysis have higher levels of middle and large molecular solutes in plasma. Additionally, hemodialysis membranes for blood purification are used generally during 4-hours-treatments, thrice weekly, while the healthy kidney operates continuously.
Membrane innovation was initially dedicated to improve biocompatibility, and is currently directed towards enhanced removal of uremic toxins and increased membrane permeability1,2. The progressive shift towards increased membrane permeability might be influenced, as mentioned, by the current gap between synthetic membranes and the natural kidney. Additionally, during the last decade some experience has been gathered with high permeability membranes such as the high cut-off membranes. Such membranes have been used in small pilot trials for controlled time periods (up to 12 weeks), and the results indicate that the expanded toxin removal might benefit the patients by decreasing the general inflammatory state3.
The MCO membranes were designed to deliver such expanded toxin removal as observed with the high cut-off membranes, while retaining albumin so that they are appropriate for regular use in conventional treatment schedules and treatment mode, i.e. 4-hours-treatments, thrice weekly in Europe. Here, the characterization of four experimental prototypes of the novel MCO membranes by dextran filtration is presented. Additionally, the sieving properties of the membranes before and after blood contact are reported, and the pore size during operation (i.e., hemodialysis treatment) is compared to the size of uremic toxins and vital proteins.
Results
Characterization of the MCO high-flux membranes by dextran sieving profiles in aqueous solution
The sieving profile of the MCO high-flux membranes as determined by dextran filtration are representatively presented in Fig. 1, together with that of one conventional high-flux membranes and one high cut-off membrane. The sieving curves for the MCO membranes are located at molecular weights between that of the conventional high-flux membrane and the high cut-off membrane. The MCO sieving curves are similar to the ficoll sieving curve for the glomerular membrane.
The sieving profile of the glomerular barrier determined with ficoll is presented for the sake of comparison. The differences between dextran, ficoll and globular proteins are well documented in the literature4,5. Ficoll and dextran molecules have different molecular shapes in solution, so that the hydrodynamic radius of the molecules are different at similar molecular weight. Additionally, ficoll molecules are more symmetric, while dextrans are more flexible. In consequence, the glomerular barrier shows different selectivity towards the two types of molecules. Furthermore, the behavior of proteins in solutions and their sieving coefficient as a function of the molecular Stokes-Einstein Radius is also different. In general, the glomerular barrier is more permeable to dextrans than to ficolls, and more permeable to ficolls than to globular proteins4.
Table 1 depicts the values for MWRO, MWCO and pore radius (i.e., the effective Stokes-Einstein radius calculated from MWCO) for the four MCO membranes, as well as for conventional high-flux and high cut-off membranes for comparison. The differences between the MCO membranes are evident from 1 to 4, showing increased MWRO and MWCO values, which indicate increased permeability.
Table 1. Characterization of MCO hemodialysis membranes, conventional high-flux and high cut-off membranes, based on dextran sieving experiments before and after blood exposure.
Before blood exposure |
After blood exposure |
|||||
---|---|---|---|---|---|---|
Membrane | MWRO [kDa] | MWCO [kDa] | Pore radius [nm] | MWRO [kDa] | MWCO [kDa] | Pore radius [nm] |
Revaclear | 5.7 ± 0.5 | 32 ± 3 | 3.9 ± 0.1 | 4.4 ± 0.3 | 14.2 ± 0.2 | 2.68 ± 0.02 |
MCO 1 | 9.4 ± 0.2 | 56 ± 3 | 5.0 ± 0.1 | 5.0 ± 0.4 | 18.1 ± 0.8 | 3.0 ± 0.1 |
MCO 2 | 10.0 ± 0.6 | 64 ± 3 | 5.4 ± 0.1 | 5.84 ± 0.09 | 18.2 ± 0.3 | 3.0 ± 0.1 |
MCO 3 | 11.3 ± 0.4 | 81 ± 9 | 6.0 ± 0.3 | 6.32 ± 0.06 | 22.7 ± 0.8 | 3.3 ± 0.1 |
MCO 4a | 12.1 ± 0.7 | 99 ± 7 | 6.5 ± 0.2 | 6.77 ± 0.06 | 25 ± 5 | 3.5 ± 0.3 |
Theralite | 15 ± 1 | 300 ± 100 | 10 ± 2 | 8.1 ± 0.8 | 40 ± 8 | 4.3 ± 0.4 |
Values are average ± standard deviation for n = 3.
aExperiments for MCO 4 included also one dextran fraction of 150 kDa.
Furthermore, Table 2 lists the parameters describing the pore size distribution calculated from the sieving profiles for the same membranes. The presented values are average and standard deviations from the mean and variance of the log-normal distributions, assessing therefore the mean and broadness of the pore size distribution, respectively. The values of the mean of the distribution increase with membrane permeability. The variance of the respective distribution is larger as the pore sizes increase.
Table 2. Mean (pore radius) and variance of the log-normal pore size distribution for the four MCO prototype membranes, conventional high-flux and high cut-off membranes before and after blood exposure.
Membrane | Before blood exposure |
After blood exposure |
||
---|---|---|---|---|
Mean [nm] | Variance [nm] | Mean [nm] | Variance [nm] | |
Revaclear | 3.0 ± 0.3 | 2 ± 1 | 2.42 ± 0.08 | 0.587 ± 0.003 |
MCO 1 | 4.1 ± 0.2 | 4.7 ± 0.8 | 2.5 ± 0.1 | 0.6 ± 0.2 |
MCO 2 | 4.0 ± 0.2 | 4.6 ± 0.4 | 2.55 ± 0.07 | 0.7 ± 0.1 |
MCO 3 | 4.40 ± 0.03 | 6.4 ± 0.1 | 2.9 ± 0.1 | 1.3 ± 0.4 |
MCO 4 | 4.8 ± 0.2 | 8.8 ± 0.4 | 3.3 ± 0.6 | 3 ± 3 |
Theralite | 5.1 ± 0.3 | 11 ± 2 | 3.4 ± 0.2 | 2 ± 1 |
Values are average ± standard deviation for n = 3.
Membrane characteristics after blood contact
Figure 2 shows the sieving curves before and after exposing the MCO membranes to blood for 40 min. The untreated MCO membranes, i.e., before blood contact, are slightly more open than the glomerular membrane for molecules above 20–30 kDa. After blood exposure, the sieving curves are shifted towards lower molecular weights, and the sieving profiles indicate that the MCO membranes are less permeable than that of the glomerular membrane.
The molecular weight retention onset and molecular weight cut-off values after blood contact are also listed in Table 1 for the different MCO membranes, as well as for one conventional high-flux membrane and one high cut-off membrane. Additionally, the effective Stokes-Einstein pore radiuses calculated from those MWCO values are presented for reference. Table 2 includes the calculated pore sizes from the respective log-normal distributions after blood contact for the different membranes.
Update on the classification of blood purification membranes
We have modified the landscape of blood purification membranes previously published6 to include the MCO high-flux membranes (Fig. 3). The square used to denote the known high-flux dialyzers (continuous line) has been expanded to include the MWRO vs. MWCO values obtained for the MCO membranes (broken line).
Discussion
Characterization of the MCO high-flux membranes by dextran sieving profiles in aqueous solution
The objective of the membrane development initiative was to develop novel high-flux membranes with toxin removal capabilities similar to that of high cut-off membranes and that are also able to retain albumin adequately. The challenge resides in the membrane manufacturing process, where increasing the pore sizes usually leads to an increase in the broadness of the pore size distribution. Inevitably, the broad pore size distribution at large mean pore sizes would cause undesirable albumin permeation. Controlled membrane manufacture allows some improvement in this direction. As can be seen, the MCO 4 membrane shows similar mean pore size to that of Theralite, while having a 20% smaller variance. The less permeable versions, MCO 1 and 2, show mean pore size of around 4 nm with less than half of the variance of Theralite. This indicates that the MCO membranes offer enhanced selectivity compared to the high cut-off membranes. As expected, when compared to conventional high-flux membranes, the mean pore size and variance of the pristine membranes is larger for the MCO 1 and 2 membranes than that of Revaclear.
Membrane characteristics after blood contact
The natural formation of the protein layer on top of the synthetic membrane during hemodialysis affects gradually the solute removal during the first 40 min of treatment. This phenomenon is illustrated by the comparison of the sieving characteristics before and after blood contact. While the pristine MCO membrane allows the passage of molecules above 70 kDa to some extent, the sieving profile of the MCO membranes (as that of every artificial membrane) shifts towards lower molecular weights during operation. This circumstance is a compromise in order to deliver a tailored removal after the inevitable membrane fouling. It has been demonstrated that the protein layer on hemodialysis membranes is completely formed after 40 min of blood contact7. As the hemodialysis treatment time is around 4 h in Europe, this means that during more than 80% of the treatment blood purification is accomplished by a membrane the performance of which is governed by protein fouling. The MCO membranes show sieving profiles close to that of the natural kidney after the formation of the protein layer, thereby maintaining the required performance along the treatment.
The pore size distribution for all membranes is narrower after blood contact, indicating that the selectivity of synthetic membranes improves during the operation. The correlation between pore size and broadness of the distribution is still present, although less pronounced. The advantages of the novel MCO membranes are more evident when compared to conventional high-flux membranes. For example, MCO 1 and 2 show mean pore sizes slightly larger than those for Revaclear, meanwhile their distribution variance did not increase with the mean pore size. For those membranes with larger mean pore sizes, i.e., MCO 3 and 4, the distribution variance correlates more with the pore size, increasing as the mean pore size increases.
The effective pore radius is a better indication than the mean pore size of the distribution for the biggest molecule that will pass through the membranes. The hydrodynamic radii (Rh) for albumin and some uremic toxins are summarized in Table 3. The effective pore size of the MCO membranes is between 3.0 and 3.5 nm after the formation of the protein layer, indicating that the membranes retain albumin during treatment. Additionally, the least permeable MCO membrane has an effective pore radius of 3.0 nm during treatment, which should allow adequate removal of large uremic toxins, up to λ-FLCs.
Table 3. Hydrodynamic radius (Rh) for albumin and some representative middle and large uremic toxins.
Molecule | Rh [nm] | Comments | Ref. | |
---|---|---|---|---|
β2 microglobulin | 1.7 | calculated from the diffusion coefficient in free solution | 16 | |
Tumor necrosis factor (TNFα) | 1.9 − 2.3 | depending on its aggregation state, influenced by concentration and pH | 17 | |
Free light chains (FLC) monomeric state (mostly κ-FLC) | 2.3 | Stokes’ radius determined by chromatography | 18 | |
Free light chains (FLC) dimeric form (mostly λ-FLC) | 2.8 | Stokes’ radius determined by chromatography | 18 | |
Albumin | 3.51 | calculated from the intrinsic viscosity (agrees with Stokes’ radii from diffusion and sedimentation coefficient) | 19 |
Based on the data here presented it can be presumed that some albumin permeation takes place even after the formation of the protein layer for the most permeable membrane MCO 4. This effect, if properly controlled, is not necessarily detrimental to the patients. Hemodialysis treatment as renal replacement therapy aims to accomplish blood purification during treatment sessions (varying in time and frequency between 3 to 8 hours per session, 3 to 6 sessions per week). In contrast, a healthy kidney performs the task continuously. Therefore, renal replacement therapy is inevitably underperforming in treatment time. This fact could be partially compensated by increased removal. Albumin loss is tolerated to some extent, as demonstrated in peritoneal dialysis patients where weekly albumin losses of 21–42 g/1.73 m2 are accepted and not linked to outcome detriment8,9. It has also been speculated that a small and controlled albumin loss might trigger albumin synthesis in patients, which is associated with a better general state of health. For example, Krieter and Canaud have argued that since the loss of albumin was associated with the removal of, among others, advanced glycation end‐products or advanced oxidation protein products, a higher membrane permeability might even be beneficial10. Currently, the MCO membrane performance is being evaluated in clinical studies. Besides removal rate and clearance of middle and large uremic toxins, the albumin removal per treatment is being investigated.
Update on the classification of blood purification membranes
The inclusion of the novel MCO membranes in the landscape of membranes for blood purification from dextran characterization evidences that these membranes expand the currently known limits of high-flux membranes. Already the MCO 1 membrane is among the best previously depicted high-flux membranes; MCO 2 expands the previous arbitrary classification beyond the currently set limits.
To deliver an appropriate renal replacement therapy is a challenge, among others, for membrane design. A blood purification membrane should mimic the filtration spectrum of the glomerular membrane, a difficult task because the glomerular membrane, as most natural membranes, shows unique selectivity and permeability. Such combination is not easily achieved with current manufacturing techniques without incorporating biological components. Additionally, the glomerular membrane suffers little to no fouling (defined as the performance detriment due to accumulation of solutes on the membrane surface) and therefore maintains its ideal performance during usage time. Renal replacement therapy is furthermore challenged by the fact that the healthy kidney accomplishes the physiological blood purification continuously Considering the limitations of synthetic membranes in renal replacement therapy, the expansion of the high-flux category is one step on the right direction that awaits confirmation in clinical application.
Conclusions
The novel MCO membranes provide for large pore sizes with appropriate pore size distribution and permeability close to that of the natural kidney. Their MWCO values suggest that, when used in hemodialysis treatments, they allow for removal of an expanded range of uremic toxins compared to conventional high-flux membranes. Formation of a protein layer on top of the synthetic membrane during hemodialysis restricts the removal of molecules above 3.5 nm in radius, optimizing removal of large uremic toxins while ensuring retention of albumin during treatment.
Methods
Devices and sample preparation
Four different types of prototype devices denoted as MCO 1 to MCO 4, which differ in membrane permeability, were investigated. As reference, a high cut-off device (Theralite) and a conventional high flux membrane (Revaclear) were also tested. All devices are manufactured by Gambro Dialysatoren GmbH, Hechingen, Germany. The membrane material is a poly(aryl ether sulfone)/polyvinylpyrrolidone blend. Further device characteristics are listed in Table 4.
Table 4. Characteristics of the devices used for membrane characterization with dextran filtration.
Device | Inner diameter /wall thickness [μm] | Surface area [m2] | UFC [mL/h*mmHg] | Membrane Type |
---|---|---|---|---|
Revaclear | 190/35 | 1.8 | 60 | High-flux |
MCO 1 | 180/35 | 1.7 | 48 | High-flux |
MCO 2 | 180/35 | 1.7 | 52 | High-flux |
MCO 3 | 180/35 | 1.7 | 49 | High-flux |
MCO 4 | 180/35 | 1.7 | 50 | High-flux |
Theralite | 215/50 | 2.1 | 52 | High Cut-off |
Membranes were characterized in minimodules, manufactured as described elsewhere6. All minimodules had a surface area of 360 cm2, nominal length of 170 mm, an effective length of approx. 120–150 mm (without PU potting) and an inner diameter of 10 mm. The internal diameter of the MCO fibers is 180 μm, and the wall thickness 35 μm, resulting in a packing density of 23.4%.
The minimodules were immersed in water for 30 min before the filtration experiments. Minimodules intended for characterization after contact with blood for simulating in vivo operation conditions were initially perfused with blood (bovine, 32% hematocrit, 60 g/L total protein, 16000 units/L heparin) for 40 min and rinsed afterwards with water for 30 to 60 min7.
Dextran sieving coefficient test
Dextran solutions were prepared and filtration experiments were carried out as previously described6. For the experiments run on MCO 4, the dextran solution included one additional fraction of 150 kDa, in order to allow for sieving coefficient calculation with similar precision as for the other MCO membranes. The sieving coefficient SC was calculated according to equation (1) as follows:
where cF is the concentration of the solute in the filtrate, cP its concentration in the permeate and cR its concentration in the retentate.
Filtration experiments were carried out under constant shear rate (γ = 750 s−1) and with ultrafiltration rate set at 20% of the blood side entrance flux QBin, calculated as:
where QBin is the flux at the blood side entrance in ml/min; n is the number of fibers in the minimodule; di is the inner diameter of the fibers in cm and γ is the constant shear rate mentioned above.
Experiments were carried out with n = 3, and results confidence could be increased with higher repetition number. Despite the low test repetition number, the standard deviations of the measurements remain below 10% of the value for all values but for the MWCO of Theralite. This is due to the lack of larger MW dextrans required for characterizing a membrane with such large pore sizes.
The chosen conditions assure a filtration regime since the Peclet-number for all the investigated membranes is well above 10 even for molecules in the 0.1 to 1 kDa range. Additionally, it should be noted that the filtration conditions are set to avoid backfiltration, contrary to the conditions typical of hemodialysis.
The obtained sieving curves were characterized by their molecular weight retention onset (MWRO) and molecular weight cut-off (MWCO). The MWCO is defined as the molecular weight at which the sieving coefficient is 0.1.; the MWRO describes the molecular weight at which the sieving coefficient is 0.96. Since membrane pore sizes are not discrete but a distribution, we have used two different approaches from membrane technology to describe the pore size distribution: (i) the effective pore size (from the MWCO) and (ii) the mean pore size from the log-normal distribution.
(i) Pore sizes correspond to molar mass as reported by Aimar11 and Venturoli4 based on the data from Granath and Kvist12
where a is the pore radius in Å and MM is the dextran molar mass in g/mol. The Stokes-Einstein radius at the MWCO corresponds to the effective membrane pore radius. The molecular weight cut-off is the molecular weight from which at least 90% of the molecules are retained by the membrane. Therefore, the hydrodynamic radius of that molecule represents the size of molecules that are retained (also at least by 90%), which is described as an effective pore size for the membranes9,13.
(ii) Pore sizes are also described by the Log-normal pore size distribution as mean pore size. Sieving curves were transformed into pore size distributions based the mentioned correlation. The distributions were evaluated as log-normal distributions and characterized by its mean and variance14.
Additional Information
How to cite this article: Boschetti-de-Fierro, A. et al. MCO Membranes: Enhanced Selectivity in High-Flux Class. Sci. Rep. 5, 18448; doi: 10.1038/srep18448 (2015).
Acknowledgments
The authors thank Werner Beck and Lars-Göran Nilsson (Gambro Dialysatoren GmbH) for fruitful discussions, as well as Anke Ermantraut, Damaris Sauter, Marina Glasebach, Marianne Schimminger and Irene Huss (Gambro Dialysatoren GmbH) for technical support. Gambro AB (including all direct and indirect subsidiaries) is now part of Baxter International Inc.
Footnotes
The authors declare competing financial interest. All authors are employees of the Gambro Dialysatoren GmbH, Hechingen (Germany). Gambro AB (including all direct and indirect subsidiaries) is now part of Baxter International Inc. Baxter is a manufacturer of dialysis devices. None of the authors has proprietary interest. All experimental information is given in great detail to exclude any bias on the results.
Author Contributions B.K. initiated and supervised the topic. A.B.F. and M.V. designed the dextran experiments and conducted the calculations. M.S. and A.B.F. designed the pore size distribution calculations. All authors discussed the results and co-wrote the manuscript.
References
- Humes H. D., Fissell W. H. & Tiranathanagul K. The future of hemodialysis membranes. Kidney Int. 69, 1115–1119 (2006). [DOI] [PubMed] [Google Scholar]
- Fujimori A. Clinical comparison of super high-flux HD and on-line HDF. Blood Purif. 35(suppl 1), 81–84 (2013). [DOI] [PubMed] [Google Scholar]
- Gondouin B. & Hutchison C. A. High cut-off dialysis membranes: current uses and future potential. Adv. Chronic Kidney Dis. 18, 180–187 (2011). [DOI] [PubMed] [Google Scholar]
- Venturoli D. & Rippe B. Ficoll and dextran vs. globular proteins as probes for testing glomerular permselectivity: effects of molecular size, shape, charge, and deformability. Am. J. Physiol. Renal Physiol. 288(4), F605 (2005). [DOI] [PubMed] [Google Scholar]
- Asgeirsson D. et al. Glomerular sieving of three neutral polysaccharides, polyethylene oxide and bikunin in rat. Effects of molecular size and conformation. Acta Physiol. 191(3), 237 (2007). [DOI] [PubMed] [Google Scholar]
- Boschetti-de-Fierro A., Voigt M., Storr M. & Krause B. Extended characterization of a new class of dialysis membranes: the high cut-off membranes. Int. J. Artif. Organs 7, 455–463 (2013). [DOI] [PubMed] [Google Scholar]
- Kunas G. A., Burke R. A., Brierton M. A. & Ofsthun N. J. The effect of blood contact and reuse on the transport properties of high-flux dialysis membranes. ASAIO J. 42, 288–294 (1996). [PubMed] [Google Scholar]
- Guest S. Hypoalbuminemia in peritoneal dialysis patients. Adv. Perit. Dial. 29, 55–60 (2013). [PubMed] [Google Scholar]
- Balafa O., Halbesma N., Struijk D. G., Dekker F. W. & Krediet R. T. Peritoneal albumin and protein losses do not predict outcome in peritoneal dialysis patients. Clin. J. Am. Soc. Nephrol. 6, 561–566 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Krieter D. H. & Canaud B. High permeability of dialysis membranes: what is the limit of albumin loss? Nephrol. Dial. Transplant. 18, 651–654 (2003). [DOI] [PubMed] [Google Scholar]
- Aimar P., Meireles M. & Sanchez V. A contribution to the translation of retention curves into pore size distributions for sieving membranes. J. Membr. Sci. 54(3), 321 (1990). [Google Scholar]
- Granath K. A. & Kvist B. E. Molecular weight distribution analysis by gel chromatography on sephadex. J. Chromatogr. A 28, 69–81 (1967). [DOI] [PubMed] [Google Scholar]
- He T., Frank M., Mulder M. H. V. & Wessling M. Preparation and characterization of nanofiltration membranes by coating polyethersulfone hollow fibers with sulfonated poly(ether ether ketone) (SPEEK). J. Membr. Sci. 307(1), 62 (2008). [Google Scholar]
- Deen W. M., Bridges C. R., Brenner B. M. & Myers B. D. Heteroporous model of glomerular size selectivity: application to normal and nephrotic humans. Am. J. Physiol. Renal Physiol. 249, F374–F389 (1985). [DOI] [PubMed] [Google Scholar]
- Axelsson J., Mahmutovic I., Rippe A. & Rippe B. Loss of size selectivity of the glomerular filtration barrier in rats following laparotomy and muscle trauma. Am. J. Physiol. Renal Physiol. 297, F577–F582 (2009). [DOI] [PubMed] [Google Scholar]
- Leypoldt J. K. & Cheung A. K. Removal of high-molecular-weight solutes during high-efficiency and high-flux haemodialysis. Nephrol. Dial. Transplant 11, 329–335 (1996). [DOI] [PubMed] [Google Scholar]
- Narhi L. O. & Arakawa T. Dissociation of recombinant tumor necrosis factor-α studied by gel permeation chromatography. Biochem. Biophys. Res. Commun. 147, 740–746 (1987). [DOI] [PubMed] [Google Scholar]
- Berggård I. & Peterson P. A. Polymeric forms of free normal κ and λ chains of human immunoglobulin. J. Biol. Chem. 244, 4299–4307 (1969). [PubMed] [Google Scholar]
- Armstrong J. K., Wenby R. B., Meiselman H. J. & Fisher T. C. The hydrodynamic radii of macromolecules and their effect on red blood cell aggregation. Biophys. J. 87, 4259–4270 (2004). [DOI] [PMC free article] [PubMed] [Google Scholar]