Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2016 Dec 1.
Published in final edited form as: J Bone Miner Res. 2015 Jul 14;30(12):2229–2238. doi: 10.1002/jbmr.2580

A Surrogate Measure of Cortical Bone Matrix Density by Long T2-Suppressed MRI

Alan C Seifert 1, Cheng Li 1, Suzanne L Wehrli 2, Felix W Wehrli 1
PMCID: PMC4683123  NIHMSID: NIHMS708089  PMID: 26085307

Abstract

Magnetic resonance has the potential to image and quantify two pools of water within bone: free water within the Haversian pore system (transverse relaxation time, T2 > 1 ms), and water hydrogen-bonded to matrix collagen (T2 ~ 300–400 µs). While total bone water concentration quantified by MRI has been shown to scale with porosity, greater insight into bone matrix density and porosity may be gained by relaxation-based separation of bound and pore water fractions. The objective of this study was to evaluate a recently developed surrogate measurement for matrix density, single adiabatic inversion recovery (SIR) zero echo-time (ZTE) MRI, in human bone.

Specimens of tibial cortical bone from 15 donors (27–97 y/o, eight female and seven male) were examined at 9.4T field strength using two methods: (1) 1H ZTE MRI, to capture total 1H signal, and (2) 1H SIR-ZTE MRI, to selectively image matrix-associated 1H signal. Total water, bone matrix, and bone mineral densities were also quantified gravimetrically, and porosity was measured by micro-CT.

ZTE apparent total water 1H concentration was 32.7±3.2 M (range: 28.5–40.3 M), and was correlated positively with porosity (R2 = 0.80) and negatively with matrix and mineral densities (R2 = 0.90 and 0.82, respectively). SIR-ZTE apparent bound water 1H concentration was 32.9±3.9 M (range: 24.4–39.8 M), and its correlations were opposite to those of apparent total water: negative with porosity (R2 = 0.73) and positive with matrix density (R2 = 0.74) and mineral density (R2 = 0.72). Porosity was strongly correlated with gravimetric matrix density (R2 = 0.91, negative) and total water density (R2 = 0.92, positive). The strong correlations of SIR-ZTE-derived apparent bound water 1H concentration with ground-truth measurements suggest that this quantitative solid-state MRI method provides a nondestructive surrogate measure of bone matrix density.

Keywords: Bone water, MRI, zero echo-time (ZTE), bone density, bone porosity

1. Introduction

Bone is composed of apatite-like mineral, collagen, and water arranged in a complex manner over several size scales(1). These three components contribute to the tissue’s mechanical properties in unique ways: mineral is responsible for bone’s compressive strength and rigidity(2), collagen matrix provides tensile strength and elasticity(3), and water serves as a medium for transport of nutrients and is responsible for bone’s viscoelasticity(4), a property which increases toughness(5).

Standard clinical assessment of bone health is most commonly performed using dual energy x-ray absorptiometry (DXA), which expresses areal bone mineral density (BMD), or mineral mass per total cross-sectional area of bone, as a T-score relative to that of a young, healthy adult(6). Though this method is in widespread use, it is insensitive to bone matrix and water content and fails to explain a significant portion of the increased fracture risk that occurs with aging(7). However, density and hydration state of bone matrix collagen are important determinants of bone quality(8), and measurements of these properties can provide complementary (and potentially superior) information on bone strength(5,911).

Over the past decade, various solid-state magnetic resonance imaging (MRI) methods have emerged that allow detection and quantification of bone water, which, due to its short transverse relaxation time (T2), is invisible to conventional MRI methods. Ultrashort echo time (UTE) MRI has been shown to visualize and quantify the short-T2 1H signal of bone water(1214), and a similar technique, zero echo-time (ZTE) imaging, has also shown promise to detect extremely short-T2 signals(1519).

Bone water exists in two major pools that are visible to solid-state MRI: water diffusing freely within the Haversian pore system(20), and water that is hydrogen-bonded to matrix collagen(2123). These two pools differ significantly in their 1H nuclear magnetic resonance (NMR) relaxation properties. Pore water has long T2 > 1 ms at 4.7T, while bound water has T2 ~ 300–400 µs(21). A separate 1H NMR signal fraction with extremely short T2 ~ 60 µs is also visible with specialized spectroscopic hardware, and has been postulated to arise from 1H nuclei in bone matrix collagen. These signal pools are diagrammed in a schematic T2 spectrum shown in Figure 1.

Figure 1.

Figure 1

Schematic T2 relaxation spectrum diagramming the three major 1H NMR signal pools in bone. Pore water has T2 > 1 ms and is broadly distributed, while bound water has T2 ~ 300–500 µs. Collagen and macromolecular signal, at T2 ~ 40–60 µs, is below the detection limit at clinical field strengths, but becomes visible using micro-imaging and spectroscopic hardware. As porosity increases, as shown in the inset µCT images of bone specimens from 27 y/o and 83 y/o female donors (dense and porous bone, respectively) collagen and bound water content decrease while pore water content increases and shifts to longer T2 values due the smaller surface-to-volume ratio of enlarged pores.

The aforementioned differences in relaxation properties between bone 1H signal components can be exploited to distinguish the matrix-associated short-T2 fraction from longer-T2 signal arising from pore water. MRI of matrix-bound water has been studied by several groups in recent years as a possible surrogate for collagen bone matrix(2434).

There are two general approaches to pore water suppression taken in prior work: bi-component effective transverse relaxation time (T2*) fitting of a free-induction decay (FID) or a series of images obtained at multiple echo times (TE)(25,27,28), and T2-selective magnetization preparation using either a long, low-amplitude radiofrequency (RF) saturation pulse(24,26,33) or an adiabatic inversion pulse followed by an inversion-recovery delay time (TI)(2932,34,35).

The major difference between T2* and T2, in general, is that T2* relaxation encompasses all mechanisms responsible for T2 relaxation, as well as an additional contribution due to magnetic field inhomogeneity on a sub-voxel scale. The large difference in volume magnetic susceptibility (χv) between water (χv = −8.9 ppm) and bone tissue (χv = −11.3 ppm(36)) and the complex arrangement of these two phases in bone leads to strong internal magnetic field gradients within the pore spaces. These gradients cause significant additional attenuation in pore water 1H signal, especially in small pores, reducing the T2* of pore water nearer to that of bound water(37), whose T2 and T2* relaxation are both dominated by dipolar relaxation. The T2 of pore water remains long, due to rephasing of isochromats by the 180° pulses.

The single adiabatic inversion-recovery (SIR) method, therefore, has the potential to outperform T2*-based methods of pore water suppression. While duration and bandwidth are inversely proportional in non-adiabatic pulses, they are less strictly linked in adiabatic RF pulses. Such a pulse can therefore simultaneously possess a long duration and broad bandwidth. The long duration allows it to saturate short-T2 signal while inverting long-T2 signal, and the broad bandwidth encompasses the broad frequency distribution of pore water within the inhomogeneous internal magnetic field environment of bone pores.

The objective of the present study was to measure zero echo time (ZTE)-derived apparent total water (aTW) and SIR-ZTE-derived apparent bound water (aBW) 1H concentrations in human cortical bone, and to evaluate the hypothesis that aBW 1H concentration parallels bone matrix density and scales inversely with porosity. These concentrations are termed ‘apparent’ in that systematic measurement errors may cause overestimation of true bone water 1H concentration. This work thus may provide a foundation upon which to build a two-part non-invasive in vivo examination of true bone tissue mineralization density (also referred to as ‘degree of mineralization of bone’(38)) consisting of bone matrix and mineral densities by 1H SIR-ZTE and 31P ZTE, respectively. Such an examination could potentially aid in the differentiation of bone-demineralizing disorders from osteoporosis.

2. Materials and Methods

2.1. Specimens and Scanners

The tissue examined consisted of cortical bone specimens taken from the tibial mid-shaft of seven male and eight female human donors, aged 27–97 years (National Disease Research Interchange, NDRI). Specimens were harvested within 30 hours of death and immediately frozen. Time between tissue harvest and this study was 8–11 years. This set ranges from young, dense bone to severely porous bone due to age-related bone loss. Donors with bone-demineralizing disorders were excluded.

A 4-mm slice was sectioned from each thawed tibia with a rotating blade at the region of maximum cortical bone thickness, 38% of the length of the tibia from the medial malleolus to the medial condyle. Then, a rectangular beam with its long axis perpendicular to the osteonal axis was cut from each slice and trimmed to fit inside a 5-mm NMR tube. Specimens ranged from 15 to 35 mm in length. The direction of the osteonal axis of the bone was indicated on the end of each beam by cutting a notch parallel to the bone’s osteonal axis, and specimens were stored individually in phosphate-buffered saline.

All scanning was performed in a 9.4T vertical-bore NMR spectrometer and micro-imaging scanner (Avance III, Bruker, Billerica, MA). For measurement of relaxation times, a broad-band inverse (BBI) probe with a one-axis z-gradient was used, and for imaging, a 20-mm quadrature birdcage probe in a three-axis microimaging gradient set was used. Bones were imaged in the presence of an intensity reference sample consisting of a 20-mm column of 10 mM MnCl2 in 90% D2O/10% H2O in a 5-mm NMR tube. This sample had a 1H concentration of 11.1 M, T2 = 530 µs, and T1 = 12.7 ms.

2.2. Measurement of Relaxation Times

Prior to imaging, all bones were scanned using a saturation recovery (SR)-prepared Carr-Purcell-Meiboom-Gill (CPMG) pulse sequence (Figure 2) to measure the T1, T2, and T2* relaxation times of the bound and pore water fractions. Knowledge of these relaxation times is necessary for subsequent conversion of raw MR image intensities to 1H concentration. The fitting of these SR-CPMG data to bi-exponential functions is illustrated in Figure 3 and described in detail in the Appendix.

Figure 2.

Figure 2

SR-CPMG pulse sequence. Saturation-recovery times (TSR) were arrayed logarithmically in 12 steps from 3 ms to 6 s, the number of refocusing pulses, N, was arrayed logarithmically from 0 to 5000 in 20 steps, and one signal acquisition was performed.

Figure 3.

Figure 3

NMR data (points) from a bone specimen from a 53-year-old male with bi-component fits (curves). Panel (a) shows a T1 fit of saturation-recovery data, (b) shows a T2 fit of CPMG echo amplitudes, and (c) shows a T2* fit of a FID. Although only one-dimensional data are shown, fits were performed using the two-dimensional methods given in the methods section (a,b: T1–T2; c: T2-T2*). Vertical axes are normalized to M0.

2.3. Zero Echo-Time Imaging

Imaging and reconstruction were performed using the standard Bruker ZTE pulse sequence (Figure 4a), and a modified form of this sequence incorporating a single adiabatic inversion preparation and delay time (SIR-ZTE, Figure 4b). ZTE was chosen due to its superior SNR performance to UTE in samples with extremely short T2*(39). Gradient amplitude was limited to 73.4 mT/m by two simultaneous requirements of the reconstruction method(40): (1) that the field of view (FOV) fully enclose all sources of signal, including the plastic support structure of the NMR probe, and (2) that the readout bandwidth be low enough (i.e., that the dwell time be long enough) so as not to lose more than ~2 readout points during the hardware-dependent transmit/receive switching dead time. Due to this finite dead time, the ZTE sequence does not actually acquire signal beginning at TE = 0; the minimum delay time is instead equal to the transmit/receive switching time of 6.4 µs. Though greater than zero, this dead time is negligible for bound and pore water signal having T2 > 300 µs.

Figure 4.

Figure 4

ZTE (a) and SIR-ZTE (b) imaging pulse sequences. ZTE parameters: 51896 projections, TR = 2 ms, 1 min 43 sec scan time. SIR-ZTE parameters: 6588 projections, TR = 200 ms, 21 min 58 sec scan time. FOV = 64 mm isotropic, resolution = 500 µm isotropic, and 1 signal acquisition for both. The pre-excitation portion of the readout gradient and the short T2* of bound and pore water 1H signal serve to effectively spoil residual transverse magnetization created by the adiabatic pulse.

First, each specimen was scanned with the reference sample using ZTE without adiabatic inversion recovery. Then, each specimen was immediately scanned again using SIR-ZTE. The adiabatic inversion pulse, with bandwidth of 5 kHz and duration of 5 ms, selectively saturates the longitudinal magnetization, Mz, of the short-T2 protons in matrix-bound water (Mz = 0) while inverting long-T2 signal arising from pore water and marrow fat (Mz < 0)(29,35,4143). The T2 response of Mz to this pulse was calculated by Bloch equation simulation. Following an appropriately chosen inversion-recovery time, the long-T2 magnetization will be nulled as it passes through zero, and the short-T2 magnetization will have recovered by longitudinal relaxation to Mz > 0. A ZTE imaging module applied at this time will thus selectively image short-T2 signal. SIR-ZTE was performed with a repetition time (TR) of 200 ms and inversion time (TI) of 50 ms, and in both of these sequences, FOV was 64 mm and resolution was 500 µm (both isotropic). Readout bandwidth in both sequences was adequate to render unnecessary any correction for signal loss due to attenuation of high spatial frequencies.

2.4. Density Quantification

To convert raw image intensity to 1H concentration, each image was corrected for longitudinal and transverse relaxation, including transverse relaxation during RF pulses. Each image was imported into Matlab (Mathworks, Natick, MA), and volumes of interest (VOIs) were drawn to fully enclose the bone and the reference sample. Then, within each VOI, the image intensity was corrected by solving Equation 3 for ρ (see Appendix), using the average relaxation times for the set of bone specimens and the measured relaxation times of the reference sample. Next, the bone and reference VOIs were refined by automatic thresholding(44), and, finally, the 1H concentration ([1H]) within the bone was calculated as the ratio of the mean corrected intensity within the bone to that of the reference, multiplied by the known 1H concentration, [1H] = 11.1 M, within the reference sample.

2.5. Micro-CT Imaging

Bone specimens were imaged using a Scanco µCT35 scanner (Scanco, Brüttisellen, Switzerland) at 18.5-µm isotropic voxel resolution. Bone boundaries were masked by 3D active snakes using ITK-SNAP(45). Pores were then segmented within these bone masks by automated thresholding, and porosity was calculated as pore volume divided by total bone volume.

2.6. Gravimetry

Bone specimens were removed from phosphate-buffered saline, blotted dry, and weighed to determine their fully hydrated mass. The bones were then placed in tared crucibles and dried at 105°C for 110 hr to remove all bound and pore water. Completion of drying was verified by observing no change in mass over a 24-hour period. The bones were again weighed, and their dry mass was recorded. Finally, the bones were incinerated at 600°C for 30 hr to burn off all organic material, and the residue was weighed. Total water mass was calculated as the difference between hydrated and dry masses, matrix mass was the difference between dry and ash masses, and mineral mass was equal to the residual ash mass. These masses were divided by total bone volume (the volume of the µCT bone boundary mask) to yield total water, matrix, and mineral densities(24,46).

3. Results

3.1. Gravimetry and µCT

All bone measurement results, both from MRI and validation methods, are given in Table 1. Mean porosity across the 15 donors was 8.96±8.61% (3.06–33.5%); all data expressed in this format are mean ± standard deviation (min-max). Volume renderings of the pore spaces of four representative bone specimens are shown in Figure 5. Gravimetric mineral density was 1118±130 mg/cc (751–1219 mg/cc), matrix density was 503.7±24.3 mg/cc (437.0–527.5 mg/cc), and total water density was 326.2±48.4 mg/cc (281.4–435.6 mg/cc) bone tissue (i.e. 32.6% v/v or 36.1 M). The mineralization mass ratio, which is the unitless ratio of gravimetric mineral density to matrix density, was 2.212±0.173 (1.719–2.367), and bone mineralization, which is the bone mineral density normalized by bone volume fraction (1 - porosity), was 1225±36 mg/cc (1130–1286 mg/cc).

Table 1.

Apparent total water (aTW) and apparent bound water (aBW) 1H concentration measurements by MRI, and reference measurements by µCT, gravimetry, and 1H NMR. Sample labels indicate age and gender.

Bone aTW [1H] (M) aBW [1H] (M) µCT Porosity (%) Mineral Density (mg/cc) Matrix Density (mg/cc) Water Density (mg/cc) T1–T2 Bound Fraction (%) Mineralization Mass Ratio Bone Mineralization (mg/cc)
27F 29.5 39.8 3.70 1197 525.2 287.4 86.6 2.280 1243
30F 28.5 38.0 3.06 1208 520.7 292.3 86.2 2.320 1246
37M 34.6 33.5 4.77 1163 494.9 316.2 78.4 2.351 1222
49M 30.2 34.2 4.08 1207 514.3 281.4 83.8 2.346 1258
53M 31.7 33.4 5.56 1159 513.1 294.6 82.9 2.258 1227
53F 29.6 35.4 3.78 1192 527.5 294.8 86.6 2.260 1239
65F 32.2 34.9 5.57 1161 513.0 297.3 81.2 2.263 1229
69M 32.8 34.7 5.18 1172 511.2 302.7 79.3 2.293 1236
74F 36.7 26.7 20.4 930.7 472.4 405.0 62.4 1.970 1169
75M 33.4 32.2 4.51 1175 510.3 298.8 80.1 2.303 1230
82F 35.5 30.6 17.3 1015 479.0 394.5 66.4 2.119 1227
83F 40.3 24.4 33.5 751.2 437.0 435.6 55.4 1.719 1130
83M 31.0 32.3 4.94 1159 520.1 317.2 77.3 2.228 1219
93M 30.7 31.8 5.18 1219 515.2 318.7 77.6 2.367 1286
97F 34.6 31.0 12.9 1058 501.2 356.2 71.4 2.110 1215

Mean 32.7 32.9 8.96 1118 503.7 326.2 77.0 2.212 1225
StDev 3.2 3.9 8.61 130 24.3 48.4 9.3 0.173 36

Figure 5.

Figure 5

Volume rendering of the pore spaces (in white) within four representative bone specimens. Note the increased number and size of pores in bone specimens from elderly female donors.

Porosity was strongly correlated with gravimetric mineral density (R2 = 0.98, negative), matrix density (R2 = 0.91, negative), and total water density (R2 = 0.92, positive). Matrix and mineral densities were also strongly positively correlated (R2 = 0.91). The mineralization mass ratio, however, was strongly positively correlated with mineral density (R2 = 0.96) and negatively correlated with porosity (R2 = 0.94).

3.2. MRI-Derived Density

ZTE-derived aTW 1H concentration was 32.7±3.2 M (28.5–40.3 M), and SIR-ZTE-derived aBW 1H concentration was 32.9±3.9 M (24.4–39.8 M). MRI quantifies the electromagnetic signal emitted by 1H nuclei within a voxel, so these measurements are properly expressed in molar concentrations of 1H nuclei, rather than mass densities in mg/cc.

1H concentration maps of aTW and aBW in four representative bone specimens are given in Figure 6. As age and porosity increase, aTW increases and aBW decreases. Note especially the region of extreme structural degradation (indicated by white arrows), with high aTW and commensurately lower aBW 1H concentration.

Figure 6.

Figure 6

Maps of apparent total water (aTW) and apparent bound water (aBW) 1H concentrations, in mol/L, in bone specimens from four representative donors. Age and gender of the donors are given within each quadrant, and the endosteal surface of each specimen faces left. Arrows indicate a region of high porosity, which has elevated total water and reduced matrix densities, and correspondingly increased aTW and decreased aBW. Non-zero background signal surrounding the bone specimen arises from extremely short-T2 signal contributed by the plastic body of the RF probe.

Apparent total water concentration was correlated positively with porosity and gravimetric water density (R2 = 0.80 and 0.79, respectively), and negatively with matrix density and mineral density (R2 = 0.90 and 0.82, respectively). Apparent bound water correlations were opposite those of aTW; aBW was correlated negatively with porosity and total water density (R2 = 0.73 and 0.76, respectively), and positively with matrix density and mineral density (R2 = 0.74 and 0.72, respectively). Scatter plots of aTW and aBW 1H concentration versus porosity, water density, and organic matrix density are given in Figure 7. All relevant correlation coefficients between parameters are given in Table 2.

Figure 7.

Figure 7

Scatter plots displaying the correlations of apparent total water (aTW, a,c,e) and apparent bound water (aBW, b,d,f) 1H concentrations versus µCT porosity (a,b), gravimetric water density (c,d), and gravimetric organic matrix density (e,f). Apparent total water 1H concentration is positively correlated with porosity and gravimetric water density and negatively with matrix density, while aBW correlations show the opposite behavior.

Table 2.

Inter-parameter correlations (R2) of measured bone properties. All correlations are statistically significant (p < 0.05).

µCT Porosity (%) Matrix Density (mg/cc) Mineral Density (mg/cc) Water Density (mg/cc) Mineralization Mass Ratio Bone Mineralization (mg/cc) T1–T2 Bound Fraction (%) aBW [1H] (M)

aTW [1H] (M) 0.80 0.90 0.82 0.79 0.69 0.71 0.86 0.77
aBW [1H] (M) 0.73 0.74 0.72 0.76 0.64 0.54 0.86
T1–T2 Bound Fraction (%) 0.90 0.89 0.88 0.96 0.79 0.63
Bone Mineralization (mg/cc) 0.73 0.70 0.84 0.62 0.84
Mineralization Mass Ratio 0.94 0.78 0.96 0.82
Water Density (mg/cc) 0.92 0.86 0.90
Mineral Density (mg/cc) 0.98 0.91
Matrix Density (mg/cc) 0.91

3.3. Relaxation Times

The short-T2 1H fraction had T1 = 480±80 ms (320–560 ms), T2 = 540±150 µs (430–980 µs), T2* = 400±50 µs (330–520 µs), and accounted for 77.0±9.3% (55.4–86.6%) of total signal by 2D bi-component T1–T2 fitting. The long-T2 fraction had T1 = 1210±300 ms (880–1910 ms), T2 = 55000±38000 µs (17000–161000 µs), T2* = 940±230 µs (600–1540 µs), and accounted for 23.0±9.3% (13.4–44.6%) of the signal.

Short-T2 fraction by 2D T1–T2 bi-component fitting was strongly negatively correlated with aTW (R2 = 0.86) and strongly positively correlated with aBW (R2 = 0.86). Short-T2 fraction was also very strongly correlated with porosity (R2 = 0.90, negative) and gravimetric matrix density (R2 = 0.89, positive).

4. Discussion

In this study, we presented non-invasive, non-destructive MRI-based measurements of apparent total and bound water 1H concentrations in human cortical bone, and compared these results to gravimetric bone density measurements and µCT porosity. The strong correlations of aBW with both gravimetric matrix density and µCT porosity support the applicability of this method as a surrogate measurement of bone matrix density. This measurement may later be combined with a previously established 31P MRI-based examination of bone mineral density(24,31) to investigate bone tissue mineralization.

Although aTW is a surrogate for bone porosity(32,47,48), it is not positively associated with the density of bone’s collagen matrix; rather, it is representative of the voids within that matrix (i.e. pore volume fraction). On the other hand, aBW is positively correlated with matrix density. It is this positive association that renders aBW a useful surrogate for matrix density, and facilitates its future pairing with a measurement of bone mineral density to create a measurement of bone mineralization; specifically, the ratio of mineral density to aBW.

Apparent total water 1H concentration (32.7±3.2 M, 28.5–40.3 M) in this work is greater than the 24 M(49) and 19.3–31.8 M(47) found in previous 1H UTE work at 3T, but the results are more consistent with the 29–41 M measured more recently by Horch et al.(29) using a 4.7T micro-imaging system. This may be due to the improved abilities of the ZTE pulse sequence and experimental (i.e. non-clinical) hardware to image the shortest-T2 components in bone. The aTW image likely contains signal not only from water, but also the extremely short-T2 signal fraction illustrated in Figure 1, which does not appear in UTE images obtained on clinical scanners.

Likewise, the aBW 1H concentration of 32.9±3.9 M (24.4–39.8 M) found in the present study using SIR-ZTE is also higher than the 12–24 M found by Horch et al.(29) or 12–23 M by Manhard et al.(30) using SIR-UTE, but these differences may also be explained by the greater ability of ZTE to capture the shortest-T2 signal present in bone. Long-range blurring of the extremely short-T2* signal arising from the plastic body of the RF probe, which appears as an increased (but constant) background signal level, and is more severe in aBW images than aTW due to lower readout gradient strength, also causes an additional systematic error in aBW values not present in aTW. The strengths of the correlations of aBW to porosity and matrix density in the present work, however, are consistent with Horch’s reported correlations versus peak stress(29) and CPMG-derived short-T2 pool fraction(29,30).

4.1. Sensitivity to Relaxation Times

All conversions of image intensity to density in this work were performed using average relaxation times for the set of 15 bone specimens, rather than individually measured relaxation times for each specimen. This ensures that the strengths of the correlations observed between the measured MRI 1H concentrations and reference measurements are translatable to eventual in vivo use, where measurement of relaxation times in each subject would not be practical. Under this constraint, the correlations of aBW with porosity and matrix density remain strong (R2 > 0.7, p < 0.00005).

If the T1 value used in Equations 36 is longer than the true T1 in the specimen, 1H concentration will be overestimated. A ±5% deviation in T1 results in ±3.0% error in calculated aTW and ±5.2% error in aBW. Variation in T2* will have very little effect on the calculated aTW: the same ±5% error in T2* results in error of approximately ±0.022%. However, due to the T2-selectivity of the adiabatic inversion pulse, a ±5% change in T2 will cause an appreciable error of ±3.8% in aBW.

In Equations 35, such errors may artificially amplify the true differences in aTW; greater porosity is associated not only with higher total water content, but also longer T1 and T2* due to decreased surface interaction and susceptibility effects in larger pores(49). Equations 46, however, suggest that these same errors may slightly reduce the sensitivity of aBW to true bound water content in the case of perfect long-T2 suppression, but this possible effect is alleviated by the fact that, while pore water relaxation properties are very strongly affected by bone porosity, bound water relaxation times are relatively constant(21). Only the stable bound water T1 and T2* values are used in quantification.

The assumption of perfect long-T2 nulling by SIR-ZTE in all specimens, however, is not realistic for a single TI applied to both dense and porous bones. In porous bone, although bound water relaxation times are stable, pore water T1 and T2 are longer than in dense bone: T1 of long-T2 pore water (T1L) by bi-component 2D T1–T2 fitting (see Equation 1) ranges from 880 to 1910 ms, and is positively correlated with porosity (R2 = 0.70, p < 0.0001). The long T2 relaxation time (T2L) ranges from 17 ms to 161 ms and is strongly positively correlated with porosity (R2 = 0.91, p < 10-7). These relaxation times vary because free water near the surface of a pore experiences an additional contribution to its transverse (1/T2) and longitudinal (1/T1) relaxation rates due to interaction with the surface, typically termed surface relaxation(20). The contribution of surface relaxation to the relaxation times of the entire pore water pool, therefore, scales with the surface to volume ratio (S/V) of the pore spaces. In very porous bones, cortical pores are large and the S/V ratio is small; however, in young, dense bone, cortical pores are very small and the S/V ratio is large, leading to an increased effect of surface relaxation and more rapid transverse and longitudinal relaxation.

In empirically choosing TI for a set of specimens such that pore water is nulled, TI is biased toward optimal nulling of pore water in more porous bones; these bones have more pore water to be nulled, so proper inversion is more important. In dense bones, with shorter T1L and T2L, pore water magnetization may be incompletely inverted during the adiabatic inversion pulse (i.e., Mz > −1) due to transverse relaxation during the pulse, and will undergo significantly faster longitudinal relaxation after inversion, thus overshooting the null point (Mz > 0) during the inversion-recovery delay. Some of this pore water magnetization in dense bone will therefore also be imaged, along with bound water magnetization, by SIR-ZTE. This additional contribution of pore water in dense bones is partially responsible for the overestimation of bound water density (which is 60–80% of total bone water(21,22)) by SIR-ZTE, causing average aBW 1H concentration (32.9±3.9 M, 24.4–39.8 M) to be slightly greater than average aTW (32.7±3.2 M, 28.5–40.3 M). This overestimation, however, does not diminish the strength of the positive correlation of aBW with bone matrix density, or the negative correlation with porosity. In fact, this phenomenon may enhance the sensitivity of aBW to changes in matrix density and porosity. The remainder of this overestimation is due to the previously discussed signal contributed by the plastic body of the RF probe. This issue could be remedied by using proton-free polymers for coil construction.

4.2. Bone Mineralization

The strong positive correlation (R2 = 0.91) between matrix and mineral densities measured by gravimetry confirms that the degree of mineralization in these bones is not the primary determinant of bulk bone mineral density; changes in mineral density are primarily a result of structural degradation rather than a deficit of mineralization. The mineralization mass ratio, which is defined as gravimetric mineral density divided by matrix density, however, was strongly negatively correlated with porosity (R2 = 0.94), consistent with the notion that bone mineralization is decreased due to rapid bone turnover in age-related bone loss.

Tissue mineralization density calculated in this work, 1225±36 mg/cc matrix (1130–1286 mg/cc matrix), was similar to previous microradiographic measurements by Boivin et al.(50) (1082±17 mg/cc). Gravimetric densities are also consistent with gravimetric measurements performed by Cao et al.(24) in rat bone.

4.3. Translatability to the Clinic

This work has established aBW 1H concentration as a surrogate for matrix density. This examination, in combination with a 31P ZTE examination of bone mineral density(31), could be developed into a non-invasive in vivo MRI assessment of bone mineral and matrix densities, and their ratio, the degree of mineralization of bone. If reduced to practice, this MRI method would allow clinicians to discriminate between age-related macroscopic bone loss and impairment of bone mineralization. Such an examination is not possible using standard x-ray-based screening methods.

This study benefitted from the enhanced performance of experimental versus clinical hardware. The 9.4T scanner used in this work is equipped with much stronger gradients than clinical scanners, narrowing the point spread function and, thus, reducing blurring. This allows for better delineation of bone margins, and even enables visualization of individual pore spaces in severely porous bones (see Figure 5, panel 83F). The use of a 20-mm RF probe and very high field strength also yields higher SNR than is achievable using clinical hardware.

Although the maximum achievable gradient strength is limited on clinical scanners, and image blurring due to signal decay during readout will be increased, this issue will be mitigated by the larger size of bones in human subjects compared to the small specimens studied in this work, and the longer T2* of both bound and pore water at lower B0. In previous work using these methods at 3T(31), point-spread function blurring was minimal. Also, at clinically relevant field strengths, such as 3T, the difference in T1 between the bound and pore water signal components is much greater (145 ms and 880 ms, respectively, at 3T, versus 480 ms and 1210 ms at 9.4T)(37), so the bound water selectivity of the SIR-ZTE method would be enhanced.

5. Conclusion

Based on the strong correlations of aBW 1H concentration with gravimetric matrix density and porosity, long T2-suppressed solid-state MRI is a promising surrogate for bone matrix density.

Acknowledgements

This work was supported by NIH R01 AR50068, NIH F31 AG042289 (A.C.S.), the Howard Hughes Medical Institute International Student Research Fellowship (C.L.), and the University of Pennsylvania Institute for Translational Medicine and Therapeutics. This work was also supported in part by the National Center for Research Resources Grant UL1RR024134, which is now the National Center for Advancing Translational Sciences Grant UL1TR000003. The content is solely the responsibility of the authors and does not necessarily represent the official views of the NIH.

We also thank Dr. Wei-Ju Tseng for assistance with µCT imaging protocol refinement, and Dr. Steven Szewczyk for facilitating the use of high-temperature furnaces for gravimetric validation.

Appendix

7.1. Measurement of Relaxation Times

Saturation recovery times (TSR) in the SR-CPMG sequence (Figure 2) were arrayed logarithmically in 12 steps from 3 ms to 6 s, and the number of refocusing pulses, N, was arrayed logarithmically from 0 to 5000 in 20 steps, resulting in TEs ranging from zero to 1 s, although the 9-µs transmit/receive dead time must be added to the beginning of each sampled echo. The duration of the refocusing RF pulse, 40 µs, is one order of magnitude less than the shortest relevant T2 relaxation time, so no corrections for incomplete refocusing are necessary. One signal acquisition was performed, and scan time for this sequence was 29 minutes. All other relevant pulse sequence parameters are given within the pulse sequence diagram in Figure 2.

Two-dimensional bi-component T1–T2 fitting was performed by fitting a sum of two exponential functions in two dimensions,

f(TSR,TE)=MS(1eTSRT1S)eTET2S+ML(1eTSRT1L)eTET2L+n, [1]

to the array of SR-CPMG echo amplitudes. Here, the magnetizations of the short- and long-T2 components are given by MS and ML, respectively. These terms are converted to short- and long-T2 fractions by dividing by their sum, i.e. MS/(MS+ML) and ML/(MS+ML). Longitudinal (T1) relaxation times are given by T1S and T1L, T2 relaxation times by T2S and T2L, and the noise level by n. This fitting process yields the sizes of each of the two pools and their characteristic T1 and T2 relaxation times.

T2* values for the pools were also obtained by fitting a similar equation,

f(TE,t)=MSeTET2SetT2S*+MLeTET2LetT2L*+n, [2]

to the two-dimensional array of FIDs beginning at the center of each CPMG echo. Here, t is the time within the FID after each echo. These data were taken from the SR-CPMG data set after the longest TSR of 6 seconds, which satisfies the condition of full longitudinal relaxation. This process yields two pools characterized by T2 and T2* relaxation times, rather than T1 and T2. Fitting was performed in two dimensions, rather than one, for improved accuracy and stability(51). All fitting was performed in Matlab (Mathworks, Natick, MA). Example data from bone from a 53-year-old male donor, along with bi-component fits, are shown in Figure 3.

7.2. Density Quantification

The steady-state signal acquired in the ZTE sequence (Figure 4a) is given in Equation 3:

S(r)ρ(r)1exp(TR/T1)1fzexp(TR/T1)fxyexp(TE/T2*), [3]

where ρ is aTW 1H concentration. Because ZTE measures total water signal, single-component exponential fitting of SR-CPMG data was performed to determine total water T1, T2, and T2* relaxation times, which were then incorporated into this equation. Also, e−TE/T2* ≈ 1 because this pulse sequence has an echo time of approximately zero. The fxy and fz terms represent the response of the transverse and longitudinal magnetization, respectively, to rectangular RF pulses in the general case, where the pulse is not infinitesimally short relative to T2*(52):

fxy=exp(τ/2T2)αsinc(α2(τ/2T2)2) [4]

and

fz=exp(τ/2T2)[cos(α2(τ/2T2)2)+(τ/2T2)sinc(α2(τ/2T2)2)], [5]

where τ is RF pulse duration and α = γB1τ is the nominal flip angle.

Quantification of aBW 1H concentration in the SIR-ZTE images is similar, except for three important differences. First, the contribution of the adiabatic inversion pulse was added to Equation 3:

S(r)ρ(r)1+(fHS1)exp(TI/T1)fHSexp(TR/T1)1fHSfzexp(TR/T1)fxyexp(TE/T2*), [6]

where fHS, the response of the longitudinal magnetization to the adiabatic inversion pulse, was calculated for the bone and reference by Bloch equation simulation based on their respective relaxation times. Also, because long-T2 signal is nulled in SIR-ZTE, the short T1, T2, and T2* relaxation times measured by bi-component fitting of SR-CPMG data were used in Equation 6 for relaxation correction, rather than the single-component relaxation times used in Equation 3. Finally, to maintain consistency, the VOIs obtained by automatic thresholding in the ZTE images were carried over for SIR-ZTE correction. Apart from these two changes, quantification of aBW is performed in the same manner as aTW.

Footnotes

Disclosures: All authors state that they have no conflicts of interest.

Study design: ACS and FWW. Study conduct: ACS, CL, and SLW. Data collection: ACS. Data analysis: ACS. Data interpretation: ACS, SLW, and FWW. Drafting manuscript: ACS. Revising manuscript content: ACS and FWW. Approving final version of manuscript: ACS, CL, SLW, and FWW. ACS takes responsibility for the integrity of the data analysis.

References

  • 1.Rho JY, Kuhn-Spearing L, Zioupos P. Mechanical properties and the hierarchical structure of bone. Medical Engineering & Physics. 1998 Mar;20(2):92–102. doi: 10.1016/s1350-4533(98)00007-1. [DOI] [PubMed] [Google Scholar]
  • 2.Glimcher MJ, Bonar LC, Grynpas MD, Landis WJ, Roufosse AH. Recent Studies of Bone Mineral: Is the Amorphous Calcium Phosphate Theory Valid? Journal of Crystal Growth. 1981;53(1):100–119. [Google Scholar]
  • 3.Mehta SS, Oz OK, Antich PP. Bone elasticity and ultrasound velocity are affected by subtle changes in the organic matrix. J Bone Miner Res. 1998 Jan;13(1):114–121. doi: 10.1359/jbmr.1998.13.1.114. [DOI] [PubMed] [Google Scholar]
  • 4.Garner E, Lakes R, Lee T, Swan C, Brand R. Viscoelastic dissipation in compact bone: implications for stress-induced fluid flow in bone. J Biomech Eng. 2000 Apr;122(2):166–172. doi: 10.1115/1.429638. [DOI] [PubMed] [Google Scholar]
  • 5.Nyman JS, Gorochow LE, Horch RA, Uppuganti S, Zein-Sabatto A, Manhard MK, Does MD. Journal of the Mechanical Behavior of Biomedical Materials. c. Vol. 22. Elsevier; 2013. Jun 1, Partial removal of pore and loosely bound water by low-energy drying decreases cortical bone toughness in young and old donors; pp. 136–145. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.World Health Organization. Geneva: World Health Organ Tech Rep Ser.; 2003. Prevention and management of osteoporosis; pp. 1–164. [PubMed] [Google Scholar]
  • 7.Kanis JA, Johnell O, Oden A, Dawson A, De Laet C, Jonsson B. Osteoporos Int. 12. Vol. 12. Springer-Verlag London Limited; 2001. Ten year probabilities of osteoporotic fractures according to BMD and diagnostic thresholds; pp. 989–995. [DOI] [PubMed] [Google Scholar]
  • 8.Paschalis EP, Mendelsohn R, Boskey AL. Infrared Assessment of Bone Quality: A Review. Clin Orthop Relat Res. 2011 Jan 6;469(8):2170–2178. doi: 10.1007/s11999-010-1751-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Unal M, Yang S, Akkus O. Bone. Vol. 67. Elsevier; 2014. Oct, Molecular spectroscopic identification of the water compartments in bone; pp. 228–236. [DOI] [PubMed] [Google Scholar]
  • 10.Samuel J, Sinha D, Zhao JC-G, Wang X. Bone. C. Vol. 59. Elsevier Inc; 2014. Feb 1, Water residing in small ultrastructural spaces plays a critical role in the mechanical behavior of bone; pp. 199–206. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Horch RA, Gochberg DF, Nyman JS, Does MD. Non-invasive Predictors of Human Cortical Bone Mechanical Properties: T2-Discriminated 1H NMR Compared with High Resolution X-ray. In: Phillips S, editor. PLoS ONE. 1. Vol. 6. 2011. Jan 21, p. e16359. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Robson MD, Gatehouse PD, Bydder M, Bydder GM. Magnetic resonance: An introduction to ultrashort TE (UTE) imaging. J Comput Assist Tomogr. 2003;27(6):825–846. doi: 10.1097/00004728-200311000-00001. [DOI] [PubMed] [Google Scholar]
  • 13.Reichert ILH, Robson MD, Gatehouse PD, He T, Chappell KE, Holmes J, Girgis S, Bydder GM. Magnetic resonance imaging of cortical bone with ultrashort TE pulse sequences. Magnetic Resonance Imaging. 2005 Jun;23(5):611–618. doi: 10.1016/j.mri.2005.02.017. [DOI] [PubMed] [Google Scholar]
  • 14.Fernández-Seara MA, Wehrli SL, Takahashi M, Wehrli FW. Water Content Measured by Proton-Deuteron Exchange NMR Predicts Bone Mineral Density and Mechanical Properties. J Bone Miner Res. 2003 Dec 16;19(2):289–296. doi: 10.1359/JBMR.0301227. [DOI] [PubMed] [Google Scholar]
  • 15.Ackerman JL, Raleigh DP, Glimcher MJ. Phosphorus-31 magnetic resonance imaging of hydroxyapatite: a model for bone imaging. Magn. Reson. Med. 1992 May;25(1):1–11. doi: 10.1002/mrm.1910250102. [DOI] [PubMed] [Google Scholar]
  • 16.Wu Y, Ackerman JL, Chesler DA, Li J, Neer RM, Wang J, Glimcher MJ. Calcified Tissue International. 6. Vol. 62. Springer-Verlag; 1998. Jun, Evaluation of bone mineral density using three-dimensional solid state phosphorus-31 NMR projection imaging; pp. 512–518. [DOI] [PubMed] [Google Scholar]
  • 17.Hafner S. Fast Imaging in Liquids and Solids with the Back-Projection Low-Angle Shot (Blast) Technique. Magnetic Resonance Imaging. 1994;12(7):1047–1051. doi: 10.1016/0730-725x(94)91236-p. [DOI] [PubMed] [Google Scholar]
  • 18.Kuethe DO, Caprihan A, Fukushima E, Waggoner RA. Imaging lungs using inert fluorinated gases. Magn. Reson. Med. 1998 Jan;39(1):85–88. doi: 10.1002/mrm.1910390114. [DOI] [PubMed] [Google Scholar]
  • 19.Madio DP, Lowe IJ. Magn. Reson. Med. 4. Vol. 34. Wiley Subscription Services, Inc., A Wiley Company; 1995. Oct, Ultra-Fast Imaging Using Low Flip Angles and Fids; pp. 525–529. [DOI] [PubMed] [Google Scholar]
  • 20.Wang XD, Ni QW. Journal of Orthopaedic Research. 2. Vol. 21. Wiley Subscription Services, Inc., A Wiley Company; 2003. Mar, Determination of cortical bone porosity and pore size distribution using a low field pulsed NMR approach; pp. 312–319. [DOI] [PubMed] [Google Scholar]
  • 21.Horch RA, Nyman JS, Gochberg DF, Dortch RD, Does MD. Characterization of 1H NMR signal in human cortical bone for magnetic resonance imaging. Magn. Reson. Med. 2010 Aug 25;64(3):680–687. doi: 10.1002/mrm.22459. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Ong HH, Wright AC, Wehrli FW. Deuterium nuclear magnetic resonance unambiguously quantifies pore and collagen-bound water in cortical bone. J Bone Miner Res. 2012 Nov 19;27(12):2573–2581. doi: 10.1002/jbmr.1709. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Ni Q, Nyman JS, Wang X, Santos ADL, Nicolella DP. Assessment of water distribution changes in human cortical bone by nuclear magnetic resonance. Meas. Sci. Technol. 2007 Jan 24;18(3):715–723. [Google Scholar]
  • 24.Cao H, Nazarian A, Ackerman JL, Snyder BD, Rosenberg AE, Nazarian RM, Hrovat MI, Dai G, Mintzopoulos D, Wu Y. Bone. 6. Vol. 46. Elsevier Inc; 2010. Jun 1, Quantitative 31P NMR spectroscopy and 1H MRI measurements of bone mineral and matrix density differentiate metabolic bone diseases in rat models; pp. 1582–1590. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Du J, Diaz E, Carl M, Bae W, Chung CB, Bydder GM. Ultrashort echo time imaging with bicomponent analysis. Magn. Reson. Med. 2011 Oct 27;67(3):645–649. doi: 10.1002/mrm.23047. [DOI] [PubMed] [Google Scholar]
  • 26.Cao H, Ackerman JL, Hrovat MI, Graham L, Glimcher MJ, Wu Y. Quantitative bone matrix density measurement by water- and fat-suppressed proton projection MRI (WASPI) with polymer calibration phantoms. Magn. Reson. Med. 2008 Dec;60(6):1433–1443. doi: 10.1002/mrm.21771. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Diaz E, Chung CB, Bae WC, Statum S, Znamirowski R, Bydder GM, Du J. Ultrashort echo time spectroscopic imaging (UTESI): an efficient method for quantifying bound and free water. NMR Biomed. 2011 Jul 15;25(1):161–168. doi: 10.1002/nbm.1728. [DOI] [PubMed] [Google Scholar]
  • 28.Biswas R, Bae W, Diaz E, Masuda K, Chung CB, Bydder GM, Du J. Bone. 3. Vol. 50. Elsevier Inc; 2012. Mar 1, Ultrashort echo time (UTE) imaging with bi-component analysis: Bound and free water evaluation of bovine cortical bone subject to sequential drying; pp. 749–755. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Horch RA, Gochberg DF, Nyman JS, Does MD. Clinically compatible MRI strategies for discriminating bound and pore water in cortical bone. Magn. Reson. Med. 2012 Jan 31;68(6):1774–1784. doi: 10.1002/mrm.24186. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Manhard MK, Horch RA, Harkins KD, Gochberg DF, Nyman JS, Does MD. Validation of quantitative bound- and pore-water imaging in cortical bone. Magn. Reson. Med. 2014 Jun;71(6):2166–2171. doi: 10.1002/mrm.24870. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Seifert AC, Li C, Rajapakse CS, Bashoor-Zadeh M, Bhagat YA, Wright AC, Zemel BS, Zavaliangos A, Wehrli FW. Bone mineral 31P and matrix-bound water densities measured by solid-state 31P and 1H MRI. NMR Biomed. 2014 Jul;27(7):739–748. doi: 10.1002/nbm.3107. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Li C, Seifert AC, Rad HS, Bhagat YA, Rajapakse CS, Sun W, Lam SCB, Wehrli FW. Cortical bone water concentration: dependence of MR imaging measures on age and pore volume fraction. Radiology. 2014 Sep;272(3):796–806. doi: 10.1148/radiol.14132585. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Wu Y, Ackerman JL, Chesler DA, Graham L, Wang Y, Glimcher MJ. Density of organic matrix of native mineralized bone measured by water- and fat-suppressed proton projection MRI. Magn. Reson. Med. 2003 Jun 13;50(1):59–68. doi: 10.1002/mrm.10512. [DOI] [PubMed] [Google Scholar]
  • 34.Du J, Carl M, Bydder M, Takahashi A, Chung CB, Bydder GM. Qualitative and quantitative ultrashort echo time (UTE) imaging of cortical bone. J. Magn. Reson. 2010 Dec;207(2):304–311. doi: 10.1016/j.jmr.2010.09.013. [DOI] [PubMed] [Google Scholar]
  • 35.Larson PEZ, Conolly SM, Pauly JM, Nishimura DG. Using adiabatic inversion pulses for long-T2 suppression in ultrashort echo time (UTE) imaging. Magn. Reson. Med. 2007;58(5):952–961. doi: 10.1002/mrm.21341. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Hopkins JA, Wehrli FW. Magnetic susceptibility measurement of insoluble solids by NMR: magnetic susceptibility of bone. Magn. Reson. Med. 1997 Apr;37(4):494–500. doi: 10.1002/mrm.1910370404. [DOI] [PubMed] [Google Scholar]
  • 37.Seifert AC, Wehrli SL, Wehrli FW. Bi-component T2* analysis of bound and pore bone water fractions fails at high field strengths. NMR Biomed. 2015 May 18;:n/a–n/a. doi: 10.1002/nbm.3305. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Boivin G, Meunier PJ. The mineralization of bone tissue: a forgotten dimension in osteoporosis research. Osteoporos Int. 2003;14(Suppl 3):S19–S24. doi: 10.1007/s00198-002-1347-2. [DOI] [PubMed] [Google Scholar]
  • 39.Seifert AC, Wright AC, Wehrli SL, Ong HH, Li C, Wehrli FW. 31P NMR relaxation of cortical bone mineral at multiple magnetic field strengths and levels of demineralization. NMR Biomed. 2013 Mar 18;26(9):1158–1166. doi: 10.1002/nbm.2930. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Kuethe DO, Caprihan A, Lowe IJ, Madio DP, Gach HM. Transforming NMR data despite missing points. Journal of Magnetic Resonance. 1999 Jul;139(1):18–25. doi: 10.1006/jmre.1999.1767. [DOI] [PubMed] [Google Scholar]
  • 41.Li C, Magland JF, Rad HS, Song HK, Wehrli FW. Comparison of optimized soft-tissue suppression schemes for ultrashort echo time MRI. Magn. Reson. Med. 2011 Dec 8;68(3):680–689. doi: 10.1002/mrm.23267. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Du J, Bydder M, Takahashi AM, Carl M, Chung CB, Bydder GM. Magnetic Resonance Imaging. 4. Vol. 29. Elsevier Inc; 2011. May 1, Short T2 contrast with three-dimensional ultrashort echo time imaging; pp. 470–482. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Norris DG, Lüdemann H, Leibfritz D. An analysis of the effects of short T2 values on the hyperbolic-secant pulse. Journal of Magnetic Resonance. 1991;92(1):94–101. [Google Scholar]
  • 44.Otsu N. Threshold Selection Method From Gray-Level Histograms. IEEE Transactions on Systems Man and Cybernetics. 1979;9(1):62–66. [Google Scholar]
  • 45.Yushkevich PA, Piven J, Hazlett HC, Smith RG, Ho S, Gee JC, Gerig G. User-guided 3D active contour segmentation of anatomical structures: Significantly improved efficiency and reliability. NeuroImage. 2006 Jul;31(3):1116–1128. doi: 10.1016/j.neuroimage.2006.01.015. [DOI] [PubMed] [Google Scholar]
  • 46.Anumula S, Magland J, Wehrli SL, Ong H, Song HK, Wehrli FW. Multi-modality study of the compositional and mechanical implications of hypomineralization in a rabbit model of osteomalacia. Bone. 2008 Feb;42(2):405–413. doi: 10.1016/j.bone.2007.10.011. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47.Techawiboonwong A, Song HK, Leonard MB, Wehrli FW. Cortical bone water: in vivo quantification with ultrashort echo-time MR imaging. Radiology. 2008 Sep;248(3):824–833. doi: 10.1148/radiol.2482071995. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Techawiboonwong A, Song HK, Wehrli FW. In vivo MRI of submillisecond T(2) species with two-dimensional and three-dimensional radial sequences and applications to the measurement of cortical bone water. NMR Biomed. 2008 Jan;21(1):59–70. doi: 10.1002/nbm.1179. [DOI] [PubMed] [Google Scholar]
  • 49.Rad HS, Lam SCB, Magland JF, Ong H, Li C, Song HK, Love J, Wehrli FW. Quantifying cortical bone water in vivo by three-dimensional ultra-short echo-time MRI. NMR Biomed. 2011 Jan 28;24(7):855–864. doi: 10.1002/nbm.1631. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Boivin G, Meunier PJ. The Degree of Mineralization of Bone Tissue Measured by Computerized Quantitative Contact Microradiography. Calcified Tissue International. 2002 May 27;70(6):503–511. doi: 10.1007/s00223-001-2048-0. [DOI] [PubMed] [Google Scholar]
  • 51.Celik H, Bouhrara M, Reiter DA, Fishbein KW, Spencer RG. Stabilization of the inverse Laplace transform of multiexponential decay through introduction of a second dimension. J. Magn. Reson. 2013 Nov;236:134–139. doi: 10.1016/j.jmr.2013.07.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52.Sussman MS, Pauly JM, Wright GA. Magn. Reson. Med. 6. Vol. 40. Wiley Subscription Services, Inc., A Wiley Company; 1998. Design of practical T2-selective RF excitation (TELEX) pulses; pp. 890–899. [DOI] [PubMed] [Google Scholar]

RESOURCES