Twiddler’s syndrome in spinal cord stimulation

Rafid Al-Mahfoudh 1,2,, Yuen Chan 1, Hsu Pheen Chong 1, Jibril Osman Farah 1
PMCID: PMC4684581  PMID: 26577635

Abstract

Background

The aims are to present a case series of Twiddler’s syndrome in spinal cord stimulators with analysis of the possible mechanism of this syndrome and discuss how this phenomenon can be prevented.

Method

Data were collected retrospectively between 2007 and 2013 for all patients presenting with failure of spinal cord stimulators. The diagnostic criterion for Twiddler’s syndrome is radiological evidence of twisting of wires in the presence of failure of spinal cord stimulation.

Results

Our unit implants on average 110 spinal cord stimulators a year. Over the 5-year study period, all consecutive cases of spinal cord stimulation failure were studied. Three patients with Twiddler’s syndrome were identified. Presentation ranged from 4 to 228 weeks after implantation. Imaging revealed repeated rotations and twisting of the wires of the spinal cord stimulators leading to hardware failure.

Conclusions

To the best of our knowledge this is the first reported series of Twiddler’s syndrome with implantable pulse generators (IPGs) for spinal cord stimulation. Hardware failure is not uncommon in spinal cord stimulation. Awareness and identification of Twiddler’s syndrome may help prevent its occurrence and further revisions. This may be achieved by implanting the IPG in the lumbar region subcutaneously above the belt line. Psychological intervention may have a preventative role for those who are deemed at high risk of Twiddler’s syndrome from initial psychological screening.

Keywords: Twidder’s syndrome, Spinal cord stimulation, IPG, Failed back surgery syndrome

Introduction

Twiddler’s syndrome is rare clinical condition. It is commonly been reported in cardiac pacemakers and implantable cardioverter defribrillators [5, 7, 12, 13, 19, 24, 27, 28, 30, 32, 39]. In neurosurgical practice, Twiddler’s syndrome has been reported in deep brain stimulation but never in spinal cord stimulation (SCS) [3, 9, 16, 17, 21, 31, 34]. To the best of our knowledge, we present the first reported case series of hardware malfunction due to Twiddler’s syndrome in SCS. Bayliss et al. [5] were the first to report this phenomenon in cardiac pacemakers in 1968. In their case report, it was found that the rotation of the wires was a result of twiddling by the patient. Twiddler’s syndrome can be a conscious or subconscious manipulation of an implantable pulse generator (IPG) within its subcutaneous pocket. This ultimately leads to hardware failure, which is often the mode of presentation. There are many components of an SCS and each providing a possible point of failure. Hardware failure can be due to breakage, infection, migration [36]. In Twiddler’s syndrome in addition to the radiological evidence, there is raised impedance leading to system failure, loss of capture and a recurrence of symptoms.

Materials and methods

An average of 110 spinal cord stimulators are inserted per year in our unit. All patients presenting with failure of spinal cord stimulators between 2007 and 2013 were reviewed to identify those with evidence of Twiddler’s syndrome. Three patients with radiological evidence of Twiddler’s syndrome were identified. Informed consent was obtained from these three patients to be included in this study. Radiological evidence consisted of twisting and rotation of the wires. Their case notes, images and management were reviewed. All patients underwent a pain management program (PMP) and psychological assessment prior to considering implanting an SCS. Our department’s policy is to obtain routine postoperative radiological images for all cases of SCS as a baseline. In cases of clinical failure of the device, investigations include interrogating the device and radiological imaging to exclude disconnections, lead breaks and other causes of hardware failure.

Results

Case report 1

A 54-year-old woman with failed back surgery syndrome following a lumbar microdiscectomy had a spinal cord stimulator inserted in October 2009. A Specify 5-6-5 electrode (Medtronic, Minneapolis, MN, USA) was implanted with the connectors tunnelled subcutaneously and connected to the IPG in the right iliac fossa. There were no intraoperative or postoperative complications. The patient experienced good relief of lower limb pain with stimulation.

At the point of insertion, she weighed 78.8 kg, with a height of 157 cm (body mass index [BMI], 31.9). Her pre-PMP assessment revealed moderate levels of depression and high levels of pain-related disability. Her Beck Depression Inventory score was 18 (mild). She had a Pain Anxiety Symptoms Scale of 60 (mean, 94; SD, 39; range 0–200, where 0 = no pain anxiety and 200 = severe pain anxiety).

In October 2011, she underwent reprogramming with failure to achieve satisfactory stimulation. Her imaging revealed twisting of the connecting wires adjacent to the pulse generator (Fig. 1). As it was still partially functioning, she chose to defer revision surgery. Eventually the device failed completely. In May 2012, her IPG was repositioned with new connector leads. The connector wires were found knotted and twisted intraoperatively. The new wires were tunnelled to the right lumbar region above the belt line. No further problems have been reported on follow-up.

Fig. 1.

Fig. 1

Lateral plain abdominal X-ray showing twisting of the connecting wires adjacent to the IPG

Case report 2

A 31-year-old woman with failed back surgery syndrome following a right lumbar microdiscectomy. Her weight was 93.1 kg and height 165 cm (BMI, 34.2). In her pre-PMP assessment, she had moderate depression, moderate pain-related disability and average scores on pain distress and intensity. A percutaneous trial of spinal cord stimulation was judged positive and therefore the patient opted for permanent spinal cord stimulator insertion. An eight-electrode surgical lead was implanted with the IPG in the iliac fossa. Postoperatively there was good reduction in pain scores, analgesic usage and mobility; plain X-ray indicated satisfactory electrode placement.

Four weeks after insertion, symptoms recurred to preoperative levels. Interrogation of the system revealed high impedances suggestive of electrode failure. X-ray of the system indicated an abnormal twisting of the wiring between the implanted abdominal pulse generator and the spinal electrode. The subcutaneous pulse generator had undergone repeated rotation, twisting the connecting leads to the point of fracture (Fig. 2).

Fig. 2.

Fig. 2

Lateral plain abdominal X-ray showing multiple twists of the connecting wires adjacent to the IPG

Surgical revision of the system was performed. The X-ray findings were confirmed intraoperatively. The IPG had rotated on its axis many times (Fig. 3). A new pulse generator was implanted in her right lumbar region above the iliac crest and belt line. Postoperatively, the patient reported restoration of good pain relief.

Fig. 3.

Fig. 3

IPG with twisted wires after removal from patient

Case 3

A 50-year-old woman who initially presented with cauda equina syndrome secondary to a L5/S1 disc prolapse was later diagnosed with failed back surgery syndrome. She weighed 96.95 kg, with a height of 1.57 (BMI, 39.3). This lady had a pain anxiety and symptom scale score of 152 (mean, 94; SD, 39; range, 0–200). On the Becks depression scale, she scored 48 (severe depression).

She had a trial of spinal cord stimulation in 2008, which achieved good control of her pain symptoms. She proceeded to have a spinal cord stimulator (5-6-5 electrode) inserted permanently at the level of T10-12 with connectors tunnelled to the right iliac fossa.

She had good control of her back pain initially; however, during follow-up her pain control deteriorated. In late 2009, she complained of her IPG being mobile and catching on her clothes and it was re-sited at that point. In 2013, she presented with SCS failure. X-rays showed signs of Twiddler’s syndrome. On revision of her SCS, the whole extension wires were badly twisted and damaged. She had a replacement extension set tunnelled and connected to the main electrodes, with the IPG repositioned to the lumbar region.

In all three cases the surgical technique included anchoring the lead with the included manufacturer anchoring system. This was by suturing the anchoring system to the lumbar fascia. It was not our policy to anchor the pulse generator in its subcutaneous pocket in the primary surgical procedure. All three patients denied manipulating the IPG. In the revision procedure in all three cases, the IPG was implanted in the lumbar region (Fig. 4).

Fig. 4.

Fig. 4

Schematic diagram illustrating IPG site for revision cases

Discussion

To the best of our knowledge and following a literature research by two independent investigators, we present the first reported case series on Twiddler’s syndrome in SCS. All patients were female, aged between 31 and 54 years and all patients had varying degrees of depression and anxiety. Their BMI ranged between 31.9 and 39.3, which falls into the obese category according to the WHO classification [37]. There was no pattern to the timing of presentation, which ranged from 4 to 228 weeks after implantation. Our patients underwent validated psychological testing, which consisted of the Becks Depression Inventory [6] and the Pain Anxiety Symptoms Scale [29] prior to treatment which revealed varying degrees of depression and fear of pain.

The first SCS was implanted by Shealy et al. in 1967 [38]. Hardware failure is not an uncommon complication with SCS and it is reported to occur most commonly within the first 2 years after implantation [25, 43]. In Twiddler’s syndrome with other implantable devices, presentation is reported between 6 months and 3 years [3, 9, 16, 17, 21, 31, 34, 39].

Hardware-related complications in SCS have different aetiologies, including lead migration/fracture, electrical shorting-out with cessation of stimulation, infection, decreased stimulation and battery end of life. Such complications have been reported to be between 22 and 27.2 % is SCS [25, 41]. Hardware malfunction in Twiddler’s syndrome is due to lead coiling, which can cause displacement and lead fracture [9, 16, 20, 34]. Twisting of the wires in such a manner has not been reported with SCS previously [8, 23, 26].

Twiddler’s syndrome typically presents with loss of stimulation, resulting in a recurrence of symptoms [32, 34]. There may be pain along the course of the wires due to movement and traction of the wires. Although Twiddler’s syndrome is a rare condition, there may be an element of under-reporting of Twiddler’s syndrome due to the lack of awareness of this problem in SCS.

Twiddler’s syndrome has most commonly been described with cardiac pacemakers and defibrillators [5, 7, 12, 13, 19, 24, 27, 28, 30, 32, 39]. More recently it has been described in deep brain stimulation [3, 9, 16, 17, 21, 31, 34]. Table 1 summarises all cases of Twiddler’s syndrome reported in the literature with different implants. The aetiology of Twiddler’s syndrome is not completely understood; however, some have found an association with obese women [7, 12, 16, 24, 34]. With increased subcutaneous tissue, rotation of the device may occur during normal activity, which in turn can lead to wire coiling on the IPG. In one case series [17], the patients admitted to actively twisting the IPG. However, in most reports patients deny manipulating the IPG. Physical activity has been postulated as a causative factor in IPGs implanted in the pectoral region [7, 16, 34]. An association has been noted with neurological and psychological conditions such as dementia, obsessive-compulsive disorders, seizures, depression and anxiety [9, 10, 20, 22, 34]. All patients in this series denied manipulating their IPGs. All our patients with Twiddler’s had moderate to severe depression on preoperative psychological screening.

Table 1.

Reported cases of Twiddler’s syndrome in the literature

Author/Year Country Device involved Baseline characteristics Causes Intervention
Frizell et al. 2015 [14] USA Pacemaker/ defibrillator 53-year-old female Not stated but patient was frequently touching her incision
Denied manipulation
Lead revision
Dattilo et al. 2015 [11] Italy ICD 76-year-old male Manipulation by patient with presence of scratches around pockets Tied device with silk to pectoral muscle, repair sleeves to leads
Silva et al. 2014 [40] Portugal DBS – bilateral subthalamic nucleus stimulation 65-year-old female
Parkinson’s disease
Mild medically compensated reactive depression
Denied manipulation
No causative factor identified
Revision surgery to replace damaged leads
Raissuni et al. 2014 [35] France ICD 70-year-old female Denied manipulation
No causative factor identified
Device replaced and sutured tightly to underlying muscle
Garweg et al. 2014 [15] Belgium Dual chamber cardioverter defibrillation 72-year-old male Spontaneous
Denied manipulation
Generator fixed in pocket with ligature
Stryjewski et al. 2014 [42] Poland Pacemaker 77-year-old male
Psycho-organic syndrome
Not stated New lead inserted and pacemaker fixed to muscle with non-absorbable suture
Bali et al. 2013 [4] India Pacemaker Elderly female Over enthusiastic masseuse Repositioning leads with standard suturing technique
Trout et al. 2013 [44] USA Vagal nerve stimulator 8-year-old male
ADHD
Lifelong intractable seizures
Autistic behaviours
Global developmental arrest
Manipulation of the device by child Not stated
Liang et al. 2013 [28] USA Pacemaker 52-year-old female Not stated Lead revision and device reinforcement in its pocket to fascia or subpectoral placement
Meghetti et al. 2013 [31] Italy DBS – posteroventral globus pallidus 43-year-old female Manipulation of IPG by patient IPG positioned in a submuscular pocket and fixed to fascia
Ahmed et al. 2013 [1] UK ICD 60-year-old male Not stated Repositioning of leads and anchoring the device to the pectoral muscle
Gonzalez Bermudez et al. 2012 [18] Spain ICD 74-year-old female Denied manipulation.
Spontaneous rotation of device
Repositioning of leads and generator attached to pectoral muscle
Grapsa et al. 2013 [19] UK Pacemaker 86-year-old male
Parkinson’s syndrome
Patient manipulated pacing box due to Parkinson’s disease Reprogramming of the device
Ali et al. 2012 [2] USA ICD 44-year-old female
Obesity
Used left arm to carry heavy shopping bags (implantation of ICD in left pectoral region)
Occasionally touched and scratched skin
Denied manipulation
Lead uncoiled and repositioned. IPG repositioned.
Penn et al. 2012 [34] USA DBS – bilateral anterior nuclei of the thalamus 21-year-old female
Epilepsy, no psychiatric history
Not stated New extensions inserted and sutured. Fascial suture to anchor IPG in a polyester pouch
Pavlidis et al. 2011 [33] Greece Pacemaker 82-year-old female
Severely impaired mental status
Conscious or unconscious manipulation of the pulse generator Not stated
Astradsson et al. 2010 [3] UK DBS 65-year-old female Denied manipulation
Loose IPG in a large pocket with conscious or unconscious twiddling
Revision of leads secured underneath the IPG and fixed. Further revision and IPG secured with a prolene mesh
Bayliss et al. 1968 [5] Canada Pacemaker 79-year-old female Manipulation of device
Loose pocket
Repositioning of the leads
Bracke et al. 2005 [7] Netherlands ICD 60/59/50-year-old patients Denied manipulation
All had history of recent strenuous physical exercise, one recurred after swimming
No obesity
Repositioning of leads
Burdick et al. 2010 [9] USA DBS 79-year-old female with depression with episodes of suicidal ideation, anxiety
74-year-old male, BMI 46.7
71-year-old female, Parkinson’s disease
Denies manipulation.
Patient felt spontaneous movement of IPG
Leads and IPG anchored the fascia with silk sutures
Castillo et al. 2006 [10] USA Pacemaker 76-year-old male
Dementia
Manipulation of the device Conversion of pacemaker mode and restrictive recommendations to prevent manipulation
De Buitleir et al. 1996 [12] USA ICD 54-year-old female
Obesity
Denies manipulation but on waking in the morning notices the ICD sitting on the side Lead revision
Femenia et al. 2010 [13] Argentina Pacemaker 74-year-old female Moving arms energetically IPG anchored to pocket
Geissinger et al. 2007 [16] USA DBS 65-year-old female Patient felt IPG shifted IPG placed into a polyester pouch and secure to surrounding tissue
Gelabert-Gonzalez et al. 2010 [17] Spain DBS 68-year-old female, Parkinson’s disease
65-year-old female
Patient admitted twisting the generator. Replacement of the electrode and extension wires
Harel et al. 2008 [20] Israel Pacemaker 69-year-old female Not stated Re-implantation of electrode
Israel et al. 2008 [21] Israel DBS – bilateral thalamic surgery 65-year-old female
No psychiatric history
Patient complained of an itching sensation over incision and was advised to gently massage the incision Leads revision and IPG sutured to periosteum and fascia with silk sutures, subcutaneous pocket contracted by sutures
Jaafari et al. 2009 [22] France ICD 47-year-old male
Obsessional personality traits
Since the implantation the patient had increased anxiety and described fears of having chest pain and of dying as well as receiving shocks which led to compulsive checking behaviour of ICD Patient underwent cognitive behavioural therapy and antidepressants
Lal et al. 1990 [27] USA Pacemaker 89-year-old male Not stated Repositioned leads and re-anchored using suture sleeve and pacemaker suture to pectoralis minor muscle
Mehta et al. 1992 [30] USA ICD 45-year-old, sex not stated
Obese
Denied manipulation but reported local discomfort and excessive device mobility Replacement of leads and generator re-implanted with a Dacron patch
Nicholson et al. 2003 [32] USA Pacemaker 75-year-old male Patient had been spinning the pulse generator in its surgical pocket Pacing leads replaced and additional suture were added to secure the pacemaker body to the fascia in the surgical pocket
Sidhu et al. 2009 [39] USA Pacemaker 73-year-old female Not stated Replacement of pacemaker

ICD Implantable cardioverter-defibrillator, ADHD attention deficit hyperactivity disorder, DBS deep brain stimulator, IPG internal pulse generator

Reported measures to reduce the incidence of IPG displacement and Twiddler’s syndrome include suture sleeves [27, 30]. Others have used prolene mesh, non-absorbable sutures, anchoring the IPG to the fascia, placing the IPG in a submuscular plane and limiting the pocket size to prevent further occurrences of Twiddler’s syndrome [3, 17, 21, 31, 34]. We suggest that the relocation of the IPG to above the iliac crest allows the IPG to be anchored to the lumbar fascia. Twiddler’s syndrome maybe less likely to occur with an IPG in this position as it is less accessible and visible to the patient. This approach has the added advantage of good access to the spine lead and IPG site simultaneously in the prone position. Another measure that may help prevent Twiddler’s syndrome may be psychological intervention in those identified at risk of manipulating their IPGs [22].

Conclusions

To the best of our knowledge, this is the first reported series of Twiddler’s syndrome with IPGs for spinal cord stimulation. There are many factors that may contribute to Twiddler’s syndrome including the patient’s weight, site of implantation and psychological disorders. Based on our case series, implanting the IPG in the lumbar region subcutaneously above the iliac crest may prevent recurrence of Twiddler’s syndrome. This approach has the added advantage of easier access to the spine with tunnelling in the prone position. Psychological screening may aid in identifying those at risk of Twiddler’s syndrome.

Compliance with ethical standards

Funding

No funding was received for this research.

Conflict of interest

All authors certify that they have no affiliations with or involvement in any organization or entity with any financial interest (such as honoraria; educational grants; participation in speakers’ bureaus; membership, employment, consultancies, stock ownership, or other equity interest; and expert testimony or patent-licensing arrangements), or non-financial interest (such as personal or professional relationships, affiliations, knowledge or beliefs) in the subject matter or materials discussed in this manuscript.

Ethical approval

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.

For this retrospective study formal consent is not required.

Informed consent

Informed consent was obtained from all individual participants included in the study.

References

  • 1.Ahmed FZ, Luckie M, Goode GK. An unusual case of combined classic and reverse Twiddler’s syndrome. Can J Cardiol. 2013;29:1015.e1019–1015.e1010. doi: 10.1016/j.cjca.2012.12.006. [DOI] [PubMed] [Google Scholar]
  • 2.Ali RG, Navaravong L, Cui J, Stoenescu M. An uncommon twist on Twiddler’s syndrome. Conn Med. 2012;76:81–83. [PubMed] [Google Scholar]
  • 3.Astradsson A, Schweder PM, Joint C, Green AL, Aziz TZ. Twiddler’s syndrome in a patient with a deep brain stimulation device for generalized dystonia. J Clin Neurosci. 2011;18:970–972. doi: 10.1016/j.jocn.2010.11.012. [DOI] [PubMed] [Google Scholar]
  • 4.Bali HK, Chattree KK, Bali SK, Chauhan HK, Shukla CP. A tale of early Reel syndrome caused by an over-enthusiastic masseuse. Indian Heart J. 2013;65:703–704. doi: 10.1016/j.ihj.2013.10.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Bayliss CE, Beanlands DS, Baird RJ. The pacemaker-Twiddler’s syndrome: a new complication of implantable transvenous pacemakers. CMAJ. 1968;99:371–373. [PMC free article] [PubMed] [Google Scholar]
  • 6.Beck AT, Steer RA, Carbin MG. Psychometric properties of the Beck Depression Inventory: twenty-five years of evaluation. Clin Psychol Rev. 1988;8:77–100. doi: 10.1016/0272-7358(88)90050-5. [DOI] [Google Scholar]
  • 7.Bracke F, van Gelder B, Dijkman B, Meijer A. Lead system causing Twiddler’s syndrome in patients with an implantable cardioverter-defibrillator. J Thorac Cardiovasc Surg. 2005;129:231–232. doi: 10.1016/j.jtcvs.2004.10.008. [DOI] [PubMed] [Google Scholar]
  • 8.Burchiel KJ, Anderson VC, Brown FD, Fessler RG, Friedman WA, Pelofsky S, Weiner RL, Oakley J, Shatin D. Prospective, multicenter study of spinal cord stimulation for relief of chronic back and extremity pain. Spine. 1996;21:2786–2794. doi: 10.1097/00007632-199612010-00015. [DOI] [PubMed] [Google Scholar]
  • 9.Burdick AP, Okun MS, Haq IU, Ward HE, Bova F, Jacobson CE, Bowers D, Zeilman P, Foote KD. Prevalence of Twiddler’s syndrome as a cause of deep brain stimulation hardware failure. Stereotact Funct Neurosurg. 2010;88:353–359. doi: 10.1159/000319039. [DOI] [PubMed] [Google Scholar]
  • 10.Castillo R, Cavusoglu E. Twiddler’s syndrome: an interesting cause of pacemaker failure. Cardiology. 2006;105:119–121. doi: 10.1159/000090213. [DOI] [PubMed] [Google Scholar]
  • 11.Dattilo G, Scarano M, Casale M, Sergi M, Quattrocchi S, Parato M, Imbalzano E. An atypical manifestation of Twiddler syndrome. Int J Cardiol. 2015;186:1–2. doi: 10.1016/j.ijcard.2015.03.184. [DOI] [PubMed] [Google Scholar]
  • 12.de Buitleir M, Canver CC. Twiddler’s syndrome complicating a transvenous defibrillator lead system. Chest. 1996;109:1391–1394. doi: 10.1378/chest.109.5.1391. [DOI] [PubMed] [Google Scholar]
  • 13.Femenia F, Florentino C, Arrieta M, Arce M. Iatrogenic Twiddler’s syndrome: case report and proposed experimental model. Indian Pacing Electrophysiol J. 2010;10:517–521. [PMC free article] [PubMed] [Google Scholar]
  • 14.Frizell AW, MacVane CZ. Woman with syncope. Twiddler’s syndrome. Ann Emerg Med. 2015;66(19):22. doi: 10.1016/j.annemergmed.2014.08.013. [DOI] [PubMed] [Google Scholar]
  • 15.Garweg C, Alzand BS, Willems R. Twiddler syndrome causing an inappropriate implantable cardioverter-defibrillator shock. Eur Heart J. 2014;35:516. doi: 10.1093/eurheartj/eht275. [DOI] [PubMed] [Google Scholar]
  • 16.Geissinger G, Neal JH. Spontaneous Twiddler’s syndrome in a patient with a deep brain stimulator. Surg Neurol. 2007;68:454–456. doi: 10.1016/j.surneu.2006.10.062. [DOI] [PubMed] [Google Scholar]
  • 17.Gelabert-Gonzalez M, Relova-Quinteiro JL, Castro-Garcia A. “Twiddler syndrome” in two patients with deep brain stimulation. Acta Neurochir (Wein) 2010;152:489–491. doi: 10.1007/s00701-009-0366-6. [DOI] [PubMed] [Google Scholar]
  • 18.Gonzalez Bermudez I, Pardo Fresno M, Garcia Campo E, Arnaiz Betolaza L, Crespo Carazo N, Beiras Torrado X. Twiddler syndrome as a cause of defibrillator malfunction. Rev Port Cardiol. 2012;31:837–838. doi: 10.1016/j.repc.2012.05.012. [DOI] [PubMed] [Google Scholar]
  • 19.Grapsa J, Koa-Wing M, Fox KF. Fiddling with the pacemaker: Twiddler’s syndrome in a Parkinsonian patient. Perfusion. 2013;28:31–33. doi: 10.1177/0267659112461161. [DOI] [PubMed] [Google Scholar]
  • 20.Harel G, Georgeta E, Copperman Y. Twiddler’s syndrome: a rare cause of pacemaker failure. Isr Med Assoc J. 2008;10:160–161. [PubMed] [Google Scholar]
  • 21.Israel Z, Spivak A. A tremulous Twiddler. Stereotact Funct Neurosurg. 2008;86:297–299. doi: 10.1159/000155231. [DOI] [PubMed] [Google Scholar]
  • 22.Jaafari N, Bachollet MS, Paillot C, Amiel A, Rotge JY, Lafay N, Quentin S, Wassouf I, Camus V, Senon JL, El Hage W. Obsessive compulsive disorder in a patient with Twiddler’s syndrome. Pacing Clin Electrophysiol. 2009;32:399–402. doi: 10.1111/j.1540-8159.2008.02251.x. [DOI] [PubMed] [Google Scholar]
  • 23.Kemler MA, Barendse GA, van Kleef M, de Vet HC, Rijks CP, Furnee CA, van den Wildenberg FA. Spinal cord stimulation in patients with chronic reflex sympathetic dystrophy. N Engl J Med. 2000;343:618–624. doi: 10.1056/NEJM200008313430904. [DOI] [PubMed] [Google Scholar]
  • 24.Khalilullah M, Khanna SK, Gupta U, Padmavati S. Pacemaker Twiddler’s syndrome: a note on its mechanism. J Cardiovasc Surg (Torino) 1979;20:95–100. [PubMed] [Google Scholar]
  • 25.Kumar K, Buchser E, Linderoth B, Meglio M, Van Buyten JP. Avoiding complications from spinal cord stimulation: practical recommendations from an international panel of experts. Neuromodulation. 2007;10:24–33. doi: 10.1111/j.1525-1403.2007.00084.x. [DOI] [PubMed] [Google Scholar]
  • 26.Kumar K, Taylor RS, Jacques L, Eldabe S, Meglio M, Molet J, Thomson S, O’Callaghan J, Eisenberg E, Milbouw G, Buchser E, Fortini G, Richardson J, North RB. The effects of spinal cord stimulation in neuropathic pain are sustained: a 24-month follow-up of the prospective randomized controlled multicenter trial of the effectiveness of spinal cord stimulation. Neurosurgery. 2008;63:762–770. doi: 10.1227/01.NEU.0000325731.46702.D9. [DOI] [PubMed] [Google Scholar]
  • 27.Lal RB, Avery RD. Aggressive pacemaker Twiddler’s syndrome. Dislodgement of an active fixation ventricular pacing electrode. Chest. 1990;97:756–757. doi: 10.1378/chest.97.3.756. [DOI] [PubMed] [Google Scholar]
  • 28.Liang JJ, Fenstad ER. Twiddler’s syndrome. Lancet. 2013;382:e47. doi: 10.1016/S0140-6736(13)61419-1. [DOI] [PubMed] [Google Scholar]
  • 29.McCracken LM, Zayfert C, Gross RT. The pain anxiety symptoms scale: development and validation of a scale to measure fear of pain. Pain. 1992;50:67–73. doi: 10.1016/0304-3959(92)90113-P. [DOI] [PubMed] [Google Scholar]
  • 30.Mehta D, Lipsius M, Suri RS, Krol RB, Saksena S. Twiddler’s syndrome with the implantable cardioverter-defibrillator. Am Heart J. 1992;123:1079–1082. doi: 10.1016/0002-8703(92)90728-E. [DOI] [PubMed] [Google Scholar]
  • 31.Menghetti C, Zekaj E, Saleh C, Porta M, Servello D. How to avoid Twiddler’s syndrome in deep brain stimulation for dystonia? Neuromodulation. 2013;17:198–199. doi: 10.1111/ner.12067. [DOI] [PubMed] [Google Scholar]
  • 32.Nicholson WJ, Tuohy KA, Tilkemeier P. Twiddler’s syndrome. N Engl J Med. 2003;348:1726–1727. doi: 10.1056/NEJM200304243481722. [DOI] [PubMed] [Google Scholar]
  • 33.Pavlidis AN, Orfanidis Z, Levantakis IP, Giannakopoulos A, Manolis AJ. Twiddler’s syndrome. Acute Card Care. 2011;13:194. doi: 10.3109/17482941.2011.606475. [DOI] [PubMed] [Google Scholar]
  • 34.Penn DL, Wu C, Skidmore C, Sperling MR, Sharan AD. Twiddler’s syndrome in a patient with epilepsy treated with deep brain stimulation. Epilepsia. 2012;53:e119–e121. doi: 10.1111/j.1528-1167.2012.03489.x. [DOI] [PubMed] [Google Scholar]
  • 35.Raissuni Z, Douchet MP, Chauvin M. Where is the lead? An uncommon twist in a defibrillator Twiddler syndrome. Kardiol Pol. 2014;72:204. doi: 10.5603/KP.2014.0035. [DOI] [PubMed] [Google Scholar]
  • 36.Rosenow JM, Stanton-Hicks M, Rezai AR, Henderson JM. Failure modes of spinal cord stimulation hardware. J Neurosurg Spine. 2006;5:183–190. doi: 10.3171/spi.2006.5.3.183. [DOI] [PubMed] [Google Scholar]
  • 37.Seidell JC, Flegal KM. Assessing obesity: classification and epidemiology. Br Med Bull. 1997;53:238–252. doi: 10.1093/oxfordjournals.bmb.a011611. [DOI] [PubMed] [Google Scholar]
  • 38.Shealy CN, Mortimer JT, Reswick JB. Electrical inhibition of pain by stimulation of the dorsal columns: preliminary clinical report. Anesth Analg. 1967;46:489–491. [PubMed] [Google Scholar]
  • 39.Sidhu GS, Seifi A, Zangiabadi AH, Reinig M. Follow your leads: a case of Twiddler’s syndrome. South Med J. 2009;102:871–872. doi: 10.1097/SMJ.0b013e3181ad5d7b. [DOI] [PubMed] [Google Scholar]
  • 40.Silva PA, Chamadoira C, Costa H, Linhares P, Rosas MJ, Vaz R. Twiddler (or Not) syndrome: questioning etiology for an uncommon form of hardware malfunction in deep brain stimulation. Surg Neurol Int. 2014;5:S410–S412. doi: 10.4103/2152-7806.140201. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Spincemaille GH, Klomp HM, Steyerberg EW, van Urk H, Habbema JD. Technical data and complications of spinal cord stimulation: data from a randomized trial on critical limb ischemia. Stereotact Funct Neurosurg. 2000;74:63–72. doi: 10.1159/000056465. [DOI] [PubMed] [Google Scholar]
  • 42.Stryjewski PJ, Kuczaj A, Kulak L, Nowak J, Nessler B, Nessler J. Twiddler’s syndrome: a rare complication of pacemaker implantation. Pol Arch Med Wewn. 2014;124:209. doi: 10.20452/pamw.2196. [DOI] [PubMed] [Google Scholar]
  • 43.Taylor RS, Van Buyten JP, Buchser E. Spinal cord stimulation for chronic back and leg pain and failed back surgery syndrome: a systematic review and analysis of prognostic factors. Spine. 2005;30:152–160. doi: 10.1097/01.brs.0000152115.79236.6e. [DOI] [PubMed] [Google Scholar]
  • 44.Trout AT, Larson DB, Mangano FT, Gonsalves CH. Twiddler syndrome with a twist: a cause of vagal nerve stimulator lead fracture. Pediatr Radiol. 2013;43:1647–1651. doi: 10.1007/s00247-013-2736-8. [DOI] [PubMed] [Google Scholar]

Articles from Acta Neurochirurgica are provided here courtesy of Springer

RESOURCES