Abstract
Objective
To quantitatively compare condylar morphology using CBCT and MSCT virtual 3D surface models.
Study Design
The sample consisted of secondary data analysis of CBCT and MSCT scans obtained for clinical purposes from 74 patients treated with condylar resection and prosthetic joint replacement. 3D surface models of 146 condyles were constructed from each scan modality. Across-subject models were approximated and voxel-based registration was performed between homologous CBCT and MSCT images, making it possible to create an average CBCT and MSCT-based condylar models. SPHARM-PDM provided matching points on each correspondent model. ShapeAnalysisMANCOVA assessed statistical significant differences between observers and imaging modalities. One-sample t-test evaluated the null hypothesis that the mean differences between each CBCT and MSCT-based model were not clinically significant (<0.5mm). Tests were conducted at a significance level of P<0.05.
Results
ShapeAnalysisMANCOVA showed no statistically significant difference between the average CBCT and MSCT-based models (P>0.68). During pairwise comparison, the mean difference observed was 0.406mm, SD 0.173. One sample t-test showed that mean differences between each paired CBCT and MSCT-based models were not clinically significant (P=0.411).
Conclusion
3D surface models constructed from CBCT images are comparable to those derived from MSCT scans and may be considered reliable tools for assessing condylar morphology.
Keywords: Cone-beam computed tomography, Mandibular Condyle, Multislice spiral computed tomography
INTRODUCTION
Since the introduction of the cone-beam computed tomography (CBCT) imaging modality into dentistry in 1998,1 this exam has undergone rapid evolution and has become an increasingly important source of three-dimensional (3D) volumetric information for defining normal and abnormal anatomy of craniofacial structures.2 CBCT images have assumed a prominent role in the diagnosis of temporomandibular joint (TMJ) dysfunction, particularly for assessment of morphological changes in mandibular condyles presenting with osteoarthritis (OA).3 CBCT has been shown to render high-resolution images providing a clear visualization of the hard tissues of the TMJ,3–5 and markedly reduces radiation and cost compared to multislice spiral computed tomography (MSCT).1, 4, 6–8
CBCT scans provide isotropic voxels (i.e., equal dimension in height, width, and depth) which easily allow a multiplanar reconstruction without loss of spatial resolution.9–12 Voxel sizes in CBCT imaging range from 0.076 mm to 0.4 mm, depending on the protocol being used.9, 13 Although some authors have stated that high-resolution images (voxel size smaller than 0.2 mm) provide significantly more accurate diagnosis,14–20 others did not find differences among voxel sizes ranging from 0.125 to 0.4 mm for most clinical purposes.21–28 It has been shown that a 0.3 mm voxel size associates good diagnostic performance with lower X-ray exposure.29–31 However, further clinical studies are needed to better understand specific protocols for different diagnostic tasks.32, 33
New technologies, such as the use of CBCT-based surface models, allow for comprehensive qualitative and quantitative evaluations of the overall TMJ morphological alterations.2 3D shape correspondence analysis (SPHARM-PDM) has been described as a method to precisely locate and quantify morphological changes between healthy and pathological structures.34 This innovative method for diagnosing TMJ osteoarthritis potentially minimizes the importance of examiner’s experience, reducing intra- and inter-rater related errors, standardizes findings, and contributes to the development of new imaging markers for risk factors.5
While CBCT has been shown to provide novel 3D research data and clinically relevant diagnostic and treatment planning information,2, 5 its validity as a reliable tool compared to MSCT data is still questionable. Surgeons have often preferred MSCT data in order to produce anatomically accurate sterolitographic models of the jaws and joints through rapid prototyping technology systems.35–40 This study tested the diagnostic hypothesis that CBCT-based 3D condylar surface models are reliably comparable to MSCT. Therefore, we quantitatively compared assessments of mandibular condyle morphology using CBCT and MSCT virtual 3D surface models.
MATERIALS AND METHODS
This study is a secondary data analysis of available CBCT and MSCT scans obtained for clinical purposes from 74 patients (146 mandibular condyles) diagnosed with chronic TMJ osteoarthritis (OA) (Figure 1). Two condyles were excluded from the study due to the presence of unilateral TMJ prostheses. CBCT scans were taken as a diagnostic clinical record for detecting TMJ morphological changes in patients with clinical symptoms of TMJ OA using diagnostic criteria for temporomandibular disorders (DC/TMD).41 A 17 × 23cm extended field of view image acquisition protocol was used during a 8.9s scan, with an isotropic 0.3 mm voxel size (i-Cat® CBCT, 120 kV, 5 mA, Imaging Sciences, Hatfield, PA). The treatment plan for the patients included in this study involved condylar resection and prosthetic joint replacement of at least one of the TMJs. The precise custom-fitted fabrication of a total TMJ joint prosthesis required a MSCT scan (LightSpeed16® MultiSlice CT Scanner, 120 kV, 180 mA, 1.0 pitch, 512.512 matrix size, 0.625 mm slice thickness, pixel size of 0.390, resolution of 2.564 pixels per mm, GE Medical Systems, Milwaukee, WI). Each device was made using computer aided design/computer aided manufacturing (CAD/CAM) technology to construct a 3D stereolithographic model of the TMJ and associated bony structures. The secondary data analysis of de-identified CBCT and MSCT scans in this study was approved by the university institutional review board and is in compliance with the Helsinki Declaration.
Fig. 1.
CBCT and MSCT-based models of condyles diagnosed with chronic TMJ osteoarthritis (OA). Osteoarthritic surface changes in study included condylar flattening (A, B), erosion (C, D), osteophythosis (E, F) and severe condylar resorption (G,H).
A total of 292 three-dimensional surface models of the condyles were constructed from each CBCT (146 condyles) and MSCT scans (146 condyles). The cortical boundaries of the condylar region visible in the cross-sections of volumetric datasets were outlined using a semi-automatic segmentation procedure. Thus, after selecting the region of interest, the program automatically segmented the mandibular condyle and part of the ramus region. The operator was then able to check out slice-by-slice the effectiveness of the automatic segmentation and perform manual editing in the three planes of space. Such an approach combined the efficiency and repeatability of automatic segmentation with the sound judgment of human expertise (ITK-SNAP software v.2.4, www.itksnap.org) (Figure 2A–H). 42, 43
Fig. 2.
Construction of 3D surface models of the right condyle from CBCT and MSCT grey level images of a patient. A, B and C show MSCT multiplanar images: axial (A), coronal (B) and sagittal (C) views; D, E and F show CBCT multiplanar images: axial (D), coronal (E) and sagittal (F) views; G and H show frontal, medial, posterior and lateral views of the right condyles rendered respectively from the CBCT and MSCT grey level images; I, Semi-transparent overlays of the registered surface models; J, Color-coded maps displaying the computed differences between 4002 corresponding points in the condylar surface constructed from MSCT – CBCT models. For this condyle, the maximum difference observed was 0.48mm and the 50th, 75th, and 95th percentiles were respectively 0.21mm, 0.29mm and 0.29mm.
After generation of the 3D surface models, left condyles were mirrored in the sagittal plane to form right condyles to facilitate comparisons (Figure 3A). All CBCT-based models were consistently approximated to a chosen reference condyle in order to establish a common co-ordinate system within the three-dimensional space (Figure 3B) and make it possible to create average CBCT and MSCT-based condylar models. For a paired comparison, the homologous CBCT and MSCT images were also registered relative to each other using regional voxel-based registration, and the anisotropic voxels of the MSCT scans were automatically reformatted to 0.3 mm isotropic voxels. Thus, the grey level intensity of each voxel in the MSCT was registered to the CBCT images.
Fig. 3.
A, All left condyles were mirrored as right condyles making it possible the superimposition and construction of the average condylar morphology for both CBCT and MSCT-based models. B, Reference condylar model (yellow) with the overlay of multiple condyles approximated in the same coordinate system. C, Parameterization of 4002 correspondent surface mesh points for statistical comparisons and detailed morphological characterization.
After registration, all models were simultaneously cropped to obtain the condylar region of interest. Shape Correspondence analysis (SPHARM-PDM software, http://www.nitrc.org/projects/spharm-pdm)44 was used to generate a mesh approximation from the volumes, establishing correspondence between each of the 4002 points in the condylar surface models across all subjects from both image modalities (Figure 3C). An average 3D condylar shape was then generated for the CBCT and the MSCT groups (Linux MeshMath script, http://www.nitrc.org/projects/spharm-pdm).44
The linux MeshMath script was then used to calculate 3D point-wise subtractions between each homologous CBCT and MSCT correspondent condylar surface models, and also between the group average condylar surface models. Semi-transparent overlays between the average models were used to visually compare the two groups in the 3D Slicer software.45 The computed 4002 vector differences were displayed as corresponding signed surface distances on each comparable condylar surface (Figure 2I,J).
Intra and inter-observer errors in segmentation of the mandibular condyles were tested by two calibrated observers, using a randomly selected sample of 10 CBCTs and 10 MSCTs volumetric datasets. Systematic differences between the observers were assessed using a Hotelling T2 test in a multivariate analysis of covariance software (ShapeAnalysisMANCOVA).
The statistical framework for testing pairwise and group differences between CBCT and MSCT-based condylar models also inlcuded ShapeAnalysisMANCOVA.46 Results were analysed by using corresponding absolute distances, i.e., the magnitude of the differences between the models in each surface point. Statistical significance is graphically displayed in the average surface model using color-coded maps: highly significant differences (p < 0.01) would be color-coded in red, intermediately significant differences would vary from yellow to green (0.01 > p > 0.05) and non-significant differences would be color-coded in blue (p > 0.05). Descriptive statistics of mean, standard deviation, median, 75th and 95th percentiles, maximum and minimum differences were also calculated for each pairwise comparison. A one sample t-test was conducted to evaluate if the absolute mean differences between each CBCT and MSCT-based models were statistically significantly different than 0.5 mm. In order to exclude possible outliers, this test considered the 95th percentile data. All the tests were conducted with 5% significance level.
RESULTS
The largest intra and inter-observer differences were 0.15mm (mean 0.07mm, SD 0.02) and 0.19mm (mean 0.10mm, SD 0.03) respectively. These differences were found between the average models derived from each observer’s condyle segmentations of both CBCT and MSCT scans. The significance maps showed no statistically significant differences for any of the 4002 points in study during both intra (P > 0.88) and inter-observer reproducibility (P > 0.73) (Figure 4).
Fig. 4.
Intra and Inter-observer reproducibility results. A, B Overlay of the average models constructed by the observers for 10 CBCT and 10 MSCT images randomly selected. C, D Absolute distances map obtained after subtracting the average models constructed from the segmentations of two independent observers. E, F Significance map shows P > 0.05 for the whole condylar surface considering both intra and inter-observer reproducibility.
Descriptive statistics of the absolute differences observed between paired CBCT and MSCT-based models are summarized in Table 1 and Figure 5. For each one of the 146 pairs of condyles in this study, the mean differences observed between the 4002 correspondent points on the surface of CBCT and MSCT-based models ranged from 0.15mm to 1.12mm (mean 0.41 mm, SD 0.17, 95% CI, 0.38 to 0.44). Considering all the differences between each correspondent 4002 points in 146 condylar surface models, a total of 584.292 comparisons, the largest difference observed was 2.55 mm. The 95th percentile showed that at 95% of the condylar surface the differences observed were below 0.52mm (SD 0.25, 95% CI, 0.48 to 0.56).
Table I.
Descriptive statistics of the differences between CBCT and MSCT-based models
| N | Meana | SDa | Mina | Maxa | Percentilea
|
Sig.c | ||
|---|---|---|---|---|---|---|---|---|
| 50th (Median) | 75th | 95th | ||||||
| 146 | 0.41 | 0.17 | 0.03 | 1.01 | 0.39 | 0.52 | 0.52 | P > 0.68 |
| Range: 0.15 – 1.12 | Range: 0.07 – 0.48 | Range: 0.00b – 0.19 | Range: 0.39 – 2.55b | Range: 0.14 – 1.13 | Range: 0.19 – 1.43 | Range: 0.19 – 1.43 | ||
Values obtained from the mean differences observed for each of the 146 condyles (4.002 points)
Values refer to the minimum and maximum differences observed considering each one of the points in the surface models for all patients in this study (584.292 correspondent points)
Value obtained from the comparison between the average CBCT and MSCT-based models
Fig. 5.
Differences observed between CBCT and MSCT-based models.
Figures 6A and 6B show respectively the semi-transparencies and the absolute distances color-coded map obtained from the comparison between the average CBCT and MSCT-based models. The distances observed in the color-coded map were determined by subtracting each one of 4002 correspondent surface points on average CBCT from the average MSCT condyle. ShapeAnalysisMANCOVA showed no statistically significant difference between the groups (P > 0.68). A 3D visualization of the absence of statistically significant differences is presented in Figure 6C, where non-significant differences are color-coded in blue (P > 0.05).
Fig. 6.
A, Semi-transparences showing the superimposition of the average CBCT and MSCT-based models. B, Absolute distances obtained after subtracting the average models. C, Significance map shows P > 0.05 for the whole condylar surface.
The one sample t-test showed that the 95% of the absolute mean differences between each paired CBCT and MSCT-based models were not statistically significantly greater than 0.5 mm (P = 0.411, mean difference of 0.02mm, 95% CI, −0.02 to 0.06, as shown in Table 2).
Table II.
One-sample t test for the null hypothesis that the differences observed between CBCT and MSCT-based models were 0.5mm
| N | Mean | SD | Test value = 0.5
|
|||||
|---|---|---|---|---|---|---|---|---|
| t | df | Sig. | Mean Difference | 95% Confidence Interval of the Difference
|
||||
| Lower | Upper | |||||||
| 146 | 0.52 | 0.24 | 0.82 | 145 | 0.41 | 0.02 | −0.02 | 0.06 |
DISCUSSION
This study compared quantitative assessments of 3D mandibular condyle morphology using CBCT and MSCT scans taken for clinical purposes. CBCT has revolutionized diagnosis and treatment planning in the field of craniofacial disorders, particularly in the assessment of TMJ bony alterations. The lower radiation dose and cost compared to MSCT, provide clinicians and researchers with a valuable diagnostic tool for identifying specific changes in the morphology of the mandibular condyles with osteoarthritis.1, 4, 6–8 However, for planning surgical interventions with stereolithographic technology, a MSCT is still often acquired.35–40, 47
Previous studies have compared CBCT and MSCT images for different medical and dental applications. Those studies utilized linear measurements obtained from axial, coronal and sagittal slices, 2D measurements based on 3D rendering, or volume differences.48–53 Other investigations specifically comparing CBCT and MSCT images of the TMJ condyles were also limited to assessment of the 2D multiplanar cross-sections or to subjective evaluations.19, 33, 54–57 The current paper is the first clinical study to quantitatively compare whole 3D mandibular condyle surfaces constructed from CBCT and MSCT scans. The 3D surface models provide additional diagnostic information on size, shape, and exact location of the bone abnormality on the affected joint.2, 5, 6, 34, 58
This study did not utilize an absolute geometric ground-truth to evaluate the quality of CBCT imaging, and instead used the MSCT scans as a clinically established method of reference.54, 59–63 Therefore, the mandibular condyle models generated from CBCT scans were compared to MSCT-based models of the same structures, which were considered as the gold standard. In the present study, the models were constructed through a semi-automatic discrimination procedure, by examiners previously calibrated with other scans not included in this study. Our results showed excellent intra and inter-observer reliability of the segmentation procedures, which corroborates other authors’ findings.5, 64–66 These images were analyzed at a 0.3mm isotropic voxel size, which has been considered as an appropriate resolution for most clinical purposes.29 Moreover, the anatomical correspondence between 4002 points in each homologous CBCT and MSCT models were automatically established by voxel-based registration and quantified by SPHARM-PDM, which are observer independent tools. Thus, our results were not confounded by examiner subjectivity.
The image analysis procedures in this study utilized voxel-wise rigid registration and shape correspondence to quantify the differences between 3D surface models constructed from CBCT and MSCT. The fully automated superimposition using voxel-wise rigid registration between CBCT and MSCT scans in this study does not depend on landmarks or planes and, rather, compared the selected reference structures voxel by voxel, achieving the least grey scale density difference between the two images. Previous studies have shown that this method provides high accuracy in 3D registration.58, 67, 68 Likewise, the surface parametrization method employed in this study, the SPHARM-PDM shape analysis toolbox, 69 has been shown to provide a unique and symmetric point-to-point correspondence across all measured surfaces. In the current study, shape differences were calculated between each correspondent point on CBCT and MSCT-based models and statistical shape analysis allowed for a localized analysis of shape, via multivariate analysis of covariance (MANCOVA). After correspondence establishment using SPHARM-PDM,69, 70 alignment and scaling normalization in a shape population, the traditional statistical analysis approach consisted in testing for differences between groups at every surface location.71 This method has been validated as an accurate tool for localizing and quantifying the degree of morphological mandibular condylar changes.34
This study results showed that the mean differences between paired-comparisons of CBCT and MSCT-based models were at a sub-millimeter level, except for 3 condyles that showed mean differences greater than 1mm and smaller than 1.12mm. Taking into account all the 584.292 differences between corresponding points in this study, only 1 condyle presented with maximum point difference of 2.55mm (Figure 5). While the overall mean difference for all condyles was 0.41mm, the 95th percentile was only slightly larger, 0.52 mm (Table 1, Figure 5). Nine condyles that presented with severe condylar changes showed 95th percentile differences ranging from 1mm to 1.43mm. These morphological abnormalities possibly made the segmentation procedure more challenging and prone to larger errors.
The use of 0.3mm isotropic voxel size CBCT scans in this study has been previously justified in the literature, even though not specifically for assessments of 3D condylar dysmorphology. Patel et al. observed that the agreement between CBCT and physical truth measurements are not significantly different with the use of 0.2mm or 0.4mm voxel size31 and, Primo et al. reported no significant differences between maxillofacial prototypes produced from CBCT data with 0.25mm and 0.4mm voxel sizes, and MSCT data with 0.3 mm pixel size.28 Other specific assessments of erosive condylar changes19 have reported only findings at the 2D multiplanar views, where cross-sectional slices of CBCT images acquired using a 6-in FOV at a voxel size of 0.2 mm presented significantly better image quality than CBCT scans acquired using a 12-in FOV with a voxel size of 0.4 mm. However, those findings referred to the quality of the grey level images and not the 3D condylar surface models constructed from the scans and assessed in this study.
The accuracy of CBCT scans in detecting TMJ osteoarthritic changes has been previously tested using dry skulls with simulated bone defects or phantoms, which do not fully reproduce clinical conditions.19, 33, 54–56 The current study revealed that the mean differences between CBCT and MSCT-based models were found to be around 0.5mm, which has been considered adequate precision for most clinical applications.31, 34, 67, 72, 73 This study assessed 3D morphological changes that occur in the mandibular condyle surface, such as erosion, flattening, and osteophytosis; however, it did not evaluate internal bony alterations of degenerative arthritis, such as increased sclerosis, or the presence of subchondral cysts. Further studies are required to investigate the detection of clinically relevant subchondral changes in CBCT and MSCT.
CONCLUSION
The present findings indicate that CBCT-based models are comparable to those derived from MSCT scans in identifying some of the changes associated with OA in the mandibular condyle. 3D virtual surface models constructed from CBCT scans may be considered as reliable tools for assessment of condylar morphology.
Clinical Relevance.
CBCT-based condylar models are reliably comparable to MSCT. The CBCT images lower radiation dose and cost provide clinicians and researchers with a valuable diagnostic tool for identifying specific osteoarthritic changes in the mandibular condyles such as erosion, flattening, and osteophytosis.
Acknowledgments
Funding: Supported by the National Institute of Dental & Craniofacial Research and the National Institute of Biomedical Imaging and Bioengineering of the National Institutes of Health (award number R01DE024450) and by the Sao Paulo Research Foundation (FAPESP, Brazil, Grant 2013/22417-0).
Footnotes
Conflict of interest: The authors declare no conflict of interest.
The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health and FAPESP.
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
Contributor Information
Liliane Rosas Gomes, Email: lilianerosas@hotmail.com, PhD Student at the Department of Orthodontics, FACULDADE DE ODONTOLOGIA DE ARARAQUARA, UNESP Univ Estadual Paulista; Address: Rua Humaitá, 1680, Centro, Araraquara, SP, Brazil, 14801-903. Visiting Scholar at the Department of Orthodontics, School of Dentistry, University of Michigan. Address: 1011 N University Ave, Ann Arbor, MI, USA, 48109.
Marcelo Regis Gomes, Email: marceloregisgomes@hotmail.com, Private Practice. Address: Rua Eduardo José dos Santos, 147, s/703, Garibaldi, Salvador, BA, Brazil, 41940-455.
João Roberto Gonçalves, Email: joaogonc@foar.unesp.br, Associate Professor at the Department of Orthodontics, FACULDADE DE ODONTOLOGIA DE ARARAQUARA, UNESP Univ Estadual Paulista. Address: Rua Humaitá, 1680, Centro, Araraquara, SP, Brazil, 14801-903.
Antônio Carlos O. Ruellas, Email: antonioruellas@yahoo.com.br, Associate Professor at the Department of Orthodontics, Universidade Federal do Rio de Janeiro. CNPq Researcher. Address: Rua Prof Paulo Rocco, 325 - Cidade Universitária, Rio de Janeiro, RJ, Brazil, 21941-617.
Larry M. Wolford, Email: lwolford@drlarrywolford.com, Departments of Oral and Maxillofacial Surgery and Orthodontics Texas, A&M University Health Science Center Baylor College of Dentistry, Baylor University Medical Center. Address: 3409 Worth St. Suite 400, Dallas, TX, USA, 75246.
Beatriz Paniagua, Email: Beatriz_Paniagua@unc.edu, Research Assistant Professor at Department of Psychiatry, University of North Carolina. Address: 101 Manning Drive, Chapel Hill, NC, USA, 27599.
Erika Benavides, Email: benavid@umich.edu, Assistant Professor at the Department of Periodontics and Oral Medicine, School of Dentistry, University of Michigan. Address: 1011 North University Avenue, Ann Arbor, MI, USA 48109.
Lúcia Helena Soares Cevidanes, Email: luciacev@umich.edu, Assistant Professor at the Department of Orthodontics, School of Dentistry, University of Michigan. Address: 1011 North University Avenue, Ann Arbor, MI, USA 48109.
References
- 1.Mozzo P, Procacci C, Tacconi A, Martini PT, Andreis IA. A new volumetric CT machine for dental imaging based on the cone-beam technique: preliminary results. Eur Radiol. 1998;8(9):1558–64. doi: 10.1007/s003300050586. [DOI] [PubMed] [Google Scholar]
- 2.Kapila SD, Nervina JM. CBCT in Orthodontics: assessment of treatment outcomes and indications for its use. Dentomaxillofac Radiol. 2015;44(1) doi: 10.1259/dmfr.20140282. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Ahmad M, Hollender L, Anderson Q, et al. Research diagnostic criteria for temporomandibular disorders (RDC/TMD): development of image analysis criteria and examiner reliability for image analysis. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2009;107(6):844–60. doi: 10.1016/j.tripleo.2009.02.023. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Alexiou K, Stamatakis H, Tsiklakis K. Evaluation of the severity of temporomandibular joint osteoarthritic changes related to age using cone beam computed tomography. Dentomaxillofac Radiol. 2009;38(3):141–7. doi: 10.1259/dmfr/59263880. [DOI] [PubMed] [Google Scholar]
- 5.Cevidanes LH, Hajati AK, Paniagua B, et al. Quantification of condylar resorption in temporomandibular joint osteoarthritis. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2010;110(1):110–7. doi: 10.1016/j.tripleo.2010.01.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Cevidanes LH, Bailey LJ, Tucker GR, Jr, et al. Superimposition of 3D cone-beam CT models of orthognathic surgery patients. Dentomaxillofac Radiol. 2005;34(6):369–75. doi: 10.1259/dmfr/17102411. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Ludlow JB, Davies-Ludlow LE, Brooks SL. Dosimetry of two extraoral direct digital imaging devices: NewTom cone beam CT and Orthophos Plus DS panoramic unit. Dentomaxillofac Radiol. 2003;32(4):229–34. doi: 10.1259/dmfr/26310390. [DOI] [PubMed] [Google Scholar]
- 8.Mah J, Hatcher D. Three-dimensional craniofacial imaging. Am J Orthod Dentofacial Orthop. 2004;126(3):308–9. doi: 10.1016/j.ajodo.2004.06.024. [DOI] [PubMed] [Google Scholar]
- 9.Lukat TD, Perschbacher SE, Pharoah MJ, Lam EW. The effects of voxel size on cone beam computed tomography images of the temporomandibular joints. Oral Surg Oral Med Oral Pathol Oral Radiol. 2015;119(2):229–37. doi: 10.1016/j.oooo.2014.10.015. [DOI] [PubMed] [Google Scholar]
- 10.Thonissen P, Ermer MA, Schmelzeisen R, et al. Sensitivity and specificity of cone beam computed tomography in thin bony structures in maxillofacial surgery - A clinical trial. J Craniomaxillofac Surg. 2015;43(7):1284–8. doi: 10.1016/j.jcms.2015.05.002. [DOI] [PubMed] [Google Scholar]
- 11.Scarfe WC, Li Z, Aboelmaaty W, Scott SA, Farman AG. Maxillofacial cone beam computed tomography: essence, elements and steps to interpretation. Aust Dent J. 2012;57(Suppl 1):46–60. doi: 10.1111/j.1834-7819.2011.01657.x. [DOI] [PubMed] [Google Scholar]
- 12.Scarfe WC, Farman AG. What is Cone-Beam CT and How Does it Work? Dent Clin N Am. 2008;52:707–30. doi: 10.1016/j.cden.2008.05.005. [DOI] [PubMed] [Google Scholar]
- 13.Kamburoglu K, Eres G, Akgun C, et al. Effect of voxel size on accuracy of cone beam computed tomography-aided assessment of periodontal furcation involvement. Oral Surg Oral Med Oral Pathol Oral Radiol. 2015 doi: 10.1016/j.oooo.2015.07.030. [DOI] [PubMed] [Google Scholar]
- 14.Sun Z, Smith T, Kortam S, et al. Effect of bone thickness on alveolar bone-height measurements from cone-beam computed tomography images. Am J Orthod Dentofacial Orthop. 2011;139(2):e117–27. doi: 10.1016/j.ajodo.2010.08.016. [DOI] [PubMed] [Google Scholar]
- 15.Wenzel A, Haiter-Neto F, Frydenberg M, Kirkevang LL. Variable-resolution cone-beam computerized tomography with enhancement filtration compared with intraoral photostimulable phosphor radiography in detection of transverse root fractures in an in vitro model. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2009;108(6):939–45. doi: 10.1016/j.tripleo.2009.07.041. [DOI] [PubMed] [Google Scholar]
- 16.Melo SL, Bortoluzzi EA, Abreu M, Jr, Correa LR, Correa M. Diagnostic ability of a cone-beam computed tomography scan to assess longitudinal root fractures in prosthetically treated teeth. J Endod. 2010;36(11):1879–82. doi: 10.1016/j.joen.2010.08.025. [DOI] [PubMed] [Google Scholar]
- 17.Dalili Z, Taramsari M, Mousavi Mehr SZ, Salamat F. Diagnostic value of two modes of cone-beam computed tomography in evaluation of simulated external root resorption: an in vitro study. Imaging Sci Dent. 2012;42(1):19–24. doi: 10.5624/isd.2012.42.1.19. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Kamburoglu K, Kursun S. A comparison of the diagnostic accuracy of CBCT images of different voxel resolutions used to detect simulated small internal resorption cavities. Int Endod J. 2010;43(9):798–807. doi: 10.1111/j.1365-2591.2010.01749.x. [DOI] [PubMed] [Google Scholar]
- 19.Librizzi ZT, Tadinada AS, Valiyaparambil JV, Lurie AG, Mallya SM. Cone-beam computed tomography to detect erosions of the temporomandibular joint: Effect of field of view and voxel size on diagnostic efficacy and effective dose. Am J Orthod Dentofacial Orthop. 2011;140(1):e25–30. doi: 10.1016/j.ajodo.2011.03.012. [DOI] [PubMed] [Google Scholar]
- 20.Haiter-Neto F, Wenzel A, Gotfredsen E. Diagnostic accuracy of cone beam computed tomography scans compared with intraoral image modalities for detection of caries lesions. Dentomaxillofac Radiol. 2008;37(1):18–22. doi: 10.1259/dmfr/87103878. [DOI] [PubMed] [Google Scholar]
- 21.Ozer SY. Detection of vertical root fractures by using cone beam computed tomography with variable voxel sizes in an in vitro model. J Endod. 2011;37(1):75–9. doi: 10.1016/j.joen.2010.04.021. [DOI] [PubMed] [Google Scholar]
- 22.da Silveira PF, Vizzotto MB, Liedke GS, et al. Detection of vertical root fractures by conventional radiographic examination and cone beam computed tomography - an in vitro analysis. Dent Traumatol. 2013;29(1):41–6. doi: 10.1111/j.1600-9657.2012.01126.x. [DOI] [PubMed] [Google Scholar]
- 23.Liedke GS, da Silveira HE, da Silveira HL, Dutra V, de Figueiredo JA. Influence of voxel size in the diagnostic ability of cone beam tomography to evaluate simulated external root resorption. J Endod. 2009;35(2):233–5. doi: 10.1016/j.joen.2008.11.005. [DOI] [PubMed] [Google Scholar]
- 24.Kamburoglu K, Murat S, Yuksel SP, Cebeci AR, Paksoy CS. Occlusal caries detection by using a cone-beam CT with different voxel resolutions and a digital intraoral sensor. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2010;109(5):e63–9. doi: 10.1016/j.tripleo.2009.12.048. [DOI] [PubMed] [Google Scholar]
- 25.Patcas R, Muller L, Ullrich O, Peltomaki T. Accuracy of cone-beam computed tomography at different resolutions assessed on the bony covering of the mandibular anterior teeth. Am J Orthod Dentofacial Orthop. 2012;141(1):41–50. doi: 10.1016/j.ajodo.2011.06.034. [DOI] [PubMed] [Google Scholar]
- 26.Torres MG, Campos PS, Segundo NP, Navarro M, Crusoe-Rebello I. Accuracy of linear measurements in cone beam computed tomography with different voxel sizes. Implant Dent. 2012;21(2):150–5. doi: 10.1097/ID.0b013e31824bf93c. [DOI] [PubMed] [Google Scholar]
- 27.Sherrard JF, Rossouw PE, Benson BW, Carrillo R, Buschang PH. Accuracy and reliability of tooth and root lengths measured on cone-beam computed tomographs. Am J Orthod Dentofacial Orthop. 2010;137(4 Suppl):S100–8. doi: 10.1016/j.ajodo.2009.03.040. [DOI] [PubMed] [Google Scholar]
- 28.Primo BT, Presotto AC, de Oliveira HW, et al. Accuracy assessment of prototypes produced using multi-slice and cone-beam computed tomography. Int J Oral Maxillofac Surg. 2012;41(10):1291–5. doi: 10.1016/j.ijom.2012.04.013. [DOI] [PubMed] [Google Scholar]
- 29.Waltrick KB, Nunes de Abreu MJ, Junior, Correa M, Zastrow MD, Dutra VD. Accuracy of linear measurements and visibility of the mandibular canal of cone-beam computed tomography images with different voxel sizes: an in vitro study. J Periodontol. 2013;84(1):68–77. doi: 10.1902/jop.2012.110524. [DOI] [PubMed] [Google Scholar]
- 30.Hekmatian E, Jafari-Pozve N, Khorrami L. The effect of voxel size on the measurement of mandibular thickness in cone-beam computed tomography. Dent Res J (Isfahan) 2014;11(5):544–8. [PMC free article] [PubMed] [Google Scholar]
- 31.Patel A, Tee BC, Fields H, et al. Evaluation of cone-beam computed tomography in the diagnosis of simulated small osseous defects in the mandibular condyle. Am J Orthod Dentofacial Orthop. 2014;145(2):143–56. doi: 10.1016/j.ajodo.2013.10.014. [DOI] [PubMed] [Google Scholar]
- 32.Spin-Neto R, Gotfredsen E, Wenzel A. Impact of voxel size variation on CBCT-based diagnostic outcome in dentistry: a systematic review. J Digit Imaging. 2013;26(4):813–20. doi: 10.1007/s10278-012-9562-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Bastos LC, Campos PS, Ramos-Perez FM, Pontual Ados A, Almeida SM. Evaluation of condyle defects using different reconstruction protocols of cone-beam computed tomography. Braz Oral Res. 2013;27(6):503–9. doi: 10.1590/S1806-83242013000600010. [DOI] [PubMed] [Google Scholar]
- 34.Paniagua B, Cevidanes L, Walker D, et al. Clinical application of SPHARM-PDM to quantify temporomandibular joint osteoarthritis. Comput Med Imaging Graph. 2011;35(5):345–52. doi: 10.1016/j.compmedimag.2010.11.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Dolwick MF. Temporomandibular joint surgery for internal derangement. Dent Clin North Am. 2007;51(1):195–208. doi: 10.1016/j.cden.2006.10.003. [DOI] [PubMed] [Google Scholar]
- 36.Guarda-Nardini L, Manfredini D, Ferronato G. Temporomandibular joint total replacement prosthesis: current knowledge and considerations for the future. Int J Oral Maxillofac Surg. 2008;37(2):103–10. doi: 10.1016/j.ijom.2007.09.175. [DOI] [PubMed] [Google Scholar]
- 37.Mercuri LG, Wolford LM, Sanders B, et al. Custom CAD/CAM total temporomandibular joint reconstruction system: preliminary multicenter report. J Oral Maxillofac Surg. 1995;53(2):106– 15. doi: 10.1016/0278-2391(95)90381-x. [DOI] [PubMed] [Google Scholar]
- 38.Wolford LM. Temporomandibular joint devices: treatment factors and outcomes. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1997;83(1):143–9. doi: 10.1016/s1079-2104(97)90105-0. [DOI] [PubMed] [Google Scholar]
- 39.Wolford LM, Pitta MC, Reiche-Fischel O, Franco PF. TMJ Concepts/Techmedica custom-made TMJ total joint prosthesis: 5-year follow-up study. Int J Oral Maxillofac Surg. 2003;32(3):268– 74. doi: 10.1054/ijom.2002.0350. [DOI] [PubMed] [Google Scholar]
- 40.Wolford LM, Dingwerth DJ, Talwar RM, Pitta MC. Comparison of 2 temporomandibular joint total joint prosthesis systems. J Oral Maxillofac Surg. 2003;61(6):685–90. doi: 10.1053/joms.2003.50112. [DOI] [PubMed] [Google Scholar]
- 41.Schiffman E, Ohrbach R, Truelove E, et al. Diagnostic Criteria for Temporomandibular Disorders (DC/TMD) for Clinical and Research Applications: recommendations of the International RDC/TMD Consortium Network* and Orofacial Pain Special Interest Groupdagger. J Oral Facial Pain Headache. 2014;28(1):6–27. doi: 10.11607/jop.1151. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Itksnap. Penn Image Computing and Science Laboratory. Philapelphia: University of Pennsylvania; [Accessed April 17 2014]. < http://www.itksnap.org>. [Google Scholar]
- 43.Yushkevich PA, Piven J, Hazlett HC, et al. User-guided 3D active contour segmentation of anatomical structures: significantly improved efficiency and reliability. Neuroimage. 2006;31(3):1116–28. doi: 10.1016/j.neuroimage.2006.01.015. [DOI] [PubMed] [Google Scholar]
- 44.Neuroimaging Informatics Tools and Resources Clearinghouse (NITRC.org) Projects Spharm-PDM. Chapel Hill: Neuro Image Research and Analysis Laboratories, University of North Carolina; [Acessed July 22 2014]. < http://www.nitrc.org/projects/spharm-pdm>. [Google Scholar]
- 45.Slicer.org. 3DSlicer. Boston: Harvard Medical School; [Acessed June 08 2015]. < http://slicer.org>. [Google Scholar]
- 46.Neuroimaging Informatics Tools and Resources Clearinghouse (NITRC.org) Projects Shape Analysis Mancova. Chapel Hill: Neuro Image Research and Analysis Laboratories, University of North Carolina; [Acessed October 15 2013]. < http://www.nitrc.org/projects/shape_mancova>. [Google Scholar]
- 47.Ingawale S, Goswami T. Temporomandibular joint: disorders, treatments, and biomechanics. Ann Biomed Eng. 2009;37(5):976–96. doi: 10.1007/s10439-009-9659-4. [DOI] [PubMed] [Google Scholar]
- 48.Dahmani-Causse M, Marx M, Deguine O, et al. Morphologic examination of the temporal bone by cone beam computed tomography: comparison with multislice helical computed tomography. Eur Ann Otorhinolaryngol Head Neck Dis. 2011;128(5):230–5. doi: 10.1016/j.anorl.2011.02.016. [DOI] [PubMed] [Google Scholar]
- 49.Loubele M, Van Assche N, Carpentier K, et al. Comparative localized linear accuracy of small-field cone-beam CT and multislice CT for alveolar bone measurements. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008;105(4):512–8. doi: 10.1016/j.tripleo.2007.05.004. [DOI] [PubMed] [Google Scholar]
- 50.Medelnik J, Hertrich K, Steinhauser-Andresen S, Hirschfelder U, Hofmann E. Accuracy of anatomical landmark identification using different CBCT- and MSCT-based 3D images: an in vitro study. J Orofac Orthop. 2011;72(4):261–78. doi: 10.1007/s00056-011-0032-5. [DOI] [PubMed] [Google Scholar]
- 51.Patcas R, Markic G, Muller L, et al. Accuracy of linear intraoral measurements using cone beam CT and multidetector CT: a tale of two CTs. Dentomaxillofac Radiol. 2012;41(8):637–44. doi: 10.1259/dmfr/21152480. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52.Pellerin O, Lin M, Bhagat N, et al. Comparison of semi-automatic volumetric VX2 hepatic tumor segmentation from cone beam CT and multi-detector CT with histology in rabbit models. Acad Radiol. 2013;20(1):115–21. doi: 10.1016/j.acra.2012.07.011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Poeschl PW, Schmidt N, Guevara-Rojas G, et al. Comparison of cone-beam and conventional multislice computed tomography for image-guided dental implant planning. Clin Oral Investig. 2013;17(1):317–24. doi: 10.1007/s00784-012-0704-6. [DOI] [PubMed] [Google Scholar]
- 54.Hintze H, Wiese M, Wenzel A. Cone beam CT and conventional tomography for the detection of morphological temporomandibular joint changes. Dentomaxillofac Radiol. 2007;36(4):192–7. doi: 10.1259/dmfr/25523853. [DOI] [PubMed] [Google Scholar]
- 55.Honda K, Larheim TA, Maruhashi K, Matsumoto K, Iwai K. Osseous abnormalities of the mandibular condyle: diagnostic reliability of cone beam computed tomography compared with helical computed tomography based on an autopsy material. Dentomaxillofac Radiol. 2006;35(3):152–7. doi: 10.1259/dmfr/15831361. [DOI] [PubMed] [Google Scholar]
- 56.Katakami K, Shimoda S, Kobayashi K, Kawasaki K. Histological investigation of osseous changes of mandibular condyles with backscattered electron images. Dentomaxillofac Radiol. 2008;37(6):330–9. doi: 10.1259/dmfr/93169617. [DOI] [PubMed] [Google Scholar]
- 57.Zain-Alabdeen EH, Alsadhan RI. A comparative study of accuracy of detection of surface osseous changes in the temporomandibular joint using multidetector CT and cone beam CT. Dentomaxillofac Radiol. 2012;41(3):185–91. doi: 10.1259/dmfr/24985971. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58.Cevidanes LH, Styner MA, Proffit WR. Image analysis and superimposition of 3-dimensional cone-beam computed tomography models. Am J Orthod Dentofacial Orthop. 2006;129(5):611–8. doi: 10.1016/j.ajodo.2005.12.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59.Cara AC, Gaia BF, Perrella A, et al. Validity of single- and multislice CT for assessment of mandibular condyle lesions. Dentomaxillofac Radiol. 2007;36(1):24–7. doi: 10.1259/dmfr/54883281. [DOI] [PubMed] [Google Scholar]
- 60.Marques AP, Perrella A, Arita ES, Pereira MF, de Cavalcanti MG. Assessment of simulated mandibular condyle bone lesions by cone beam computed tomography. Braz Oral Res. 2010;24(4):467–74. doi: 10.1590/s1806-83242010000400016. [DOI] [PubMed] [Google Scholar]
- 61.Perrella A, Borsatti MA, Tortamano IP, Rocha RG, Cavalcanti MG. Validation of computed tomography protocols for simulated mandibular lesions: a comparison study. Braz Oral Res. 2007;21(2):165–9. doi: 10.1590/s1806-83242007000200012. [DOI] [PubMed] [Google Scholar]
- 62.Tsuruta A, Yamada K, Hanada K, et al. Thickness of the roof of the glenoid fossa and condylar bone change: a CT study. Dentomaxillofac Radiol. 2003;32(4):217–21. doi: 10.1259/dmfr/15476586. [DOI] [PubMed] [Google Scholar]
- 63.Yamada K, Tsuruta A, Hanada K, Hayashi T. Morphology of the articular eminence in temporomandibular joints and condylar bone change. J Oral Rehabil. 2004;31(5):438–44. doi: 10.1111/j.1365-2842.2004.01255.x. [DOI] [PubMed] [Google Scholar]
- 64.Fourie Z, Damstra J, Schepers RH, Gerrits PO, Ren Y. Segmentation process significantly influences the accuracy of 3D surface models derived from cone beam computed tomography. Eur J Radiol. 2012;81(4):e524–30. doi: 10.1016/j.ejrad.2011.06.001. [DOI] [PubMed] [Google Scholar]
- 65.Luu NS, Mandich MA, Flores-Mir C, et al. The validity, reliability, and time requirement of study model analysis using cone-beam computed tomography-generated virtual study models. Orthod Craniofac Res. 2014;17(1):14–26. doi: 10.1111/ocr.12024. [DOI] [PubMed] [Google Scholar]
- 66.Xi T, van Loon B, Fudalej P, et al. Validation of a novel semi-automated method for three-dimensional surface rendering of condyles using cone beam computed tomography data. Int J Oral Maxillofac Surg. 2013;42(8):1023–9. doi: 10.1016/j.ijom.2013.01.016. [DOI] [PubMed] [Google Scholar]
- 67.Almukhtar A, Ju X, Khambay B, McDonald J, Ayoub A. Comparison of the accuracy of voxel based registration and surface based registration for 3D assessment of surgical change following orthognathic surgery. PLoS One. 2014;9(4):e93402. doi: 10.1371/journal.pone.0093402. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 68.Nada RM, Maal TJ, Breuning KH, et al. Accuracy and reproducibility of voxel based superimposition of cone beam computed tomography models on the anterior cranial base and the zygomatic arches. PLoS One. 2011;6(2):e16520. doi: 10.1371/journal.pone.0016520. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 69.Styner M, Oguz I, Xu S, et al. Framework for the Statistical Shape Analysis of Brain Structures using SPHARM-PDM. Insight J. 2006;(1071):242–50. [PMC free article] [PubMed] [Google Scholar]
- 70.Brechbuhler C, Gerig G, Kubler O. Parameterization of closed surfaces for 3d shape description. Computer Vision, Graphics, Image Processing: Image Understanding. 1995 [Google Scholar]
- 71.Paniagua B, Styner M, Macenko M, Pantazis D, Niethammer M. Local shape analysis using MANCOVA. Insight Journal. 2009:21. [Google Scholar]
- 72.Bell A, Ayoub AF, Siebert P. Assessment of the accuracy of a three-dimensional imaging system for archiving dental study models. J Orthod. 2003;30(3):219–23. doi: 10.1093/ortho/30.3.219. [DOI] [PubMed] [Google Scholar]
- 73.Bettega G, Cinquin P, Lebeau J, Raphael B. Computer-assisted orthognathic surgery: clinical evaluation of a mandibular condyle repositioning system. J Oral Maxillofac Surg. 2002;60(1):27–34. doi: 10.1053/joms.2002.29069. [DOI] [PubMed] [Google Scholar]