Abstract
Hurwitz, B. J., Zion, M. M., and Promund Obel, I. W. (1974).Thorax, 29, 678-684. Myocardial perforation by flexible Elema endocardial pacing catheters. Myocardial perforation as a complication of permanent endocardial pacing with flexible Elema catheters is described in nine patients. In eight patients pacing failed, but there were no other serious ill-effects. Diaphragmatic pacing occurred in five patients. Radiological evidence of movement, either posteriorly and/or laterally, of the electrode tip was detected in five patients. In only two of them was the electrode seen to be definitely extracardiac in position. A change in the pacing wave form aided the diagnosis in two patients. In a further two patients electrode tip electrocardiograms helped to confirm the diagnosis where, upon withdrawal of the perforating electrode, small Q waves disappeared, RS complexes enlarged, and the ST segments became more elevated, with deeply inverted T waves.
Withdrawal and repositioning of the electrode catheters was performed in seven patients. In one, a new catheter was inserted, and in another, where uninterrupted cardiac pacing was accompanied by intermittent diaphragmatic pacing, the pacing system was left unaltered. The follow-up is from six months to four and a half years with one late death apparently unrelated to pacing failure.
Full text
PDF






Images in this article
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Barold S. S., Center S. Electrographic diagnosis of perforation of the heart by pacing catheter electrode. Am J Cardiol. 1969 Aug;24(2):274–278. doi: 10.1016/0002-9149(69)90416-0. [DOI] [PubMed] [Google Scholar]
- Battle W. E., Banas J. S., Jr, Levine H. J. Papillary muscle rupture of the mitral valve complicating removal of a permanent transvenous electrode. Chest. 1973 Mar;63(3):455–457. doi: 10.1378/chest.63.3.455. [DOI] [PubMed] [Google Scholar]
- Bernstein V., Rotem C. E., Peretz D. I. Permanent pacemakers: 8-year follow-up study. Incidence and management of congestive cardiac failure and perforations. Ann Intern Med. 1971 Mar;74(3):361–369. doi: 10.7326/0003-4819-74-3-361. [DOI] [PubMed] [Google Scholar]
- Furman S., Escher D. J., Solomon N. Experiences with myocardial and transvenous implanted cardiac pacemakers. Am J Cardiol. 1969 Jan;23(1):66–72. doi: 10.1016/0002-9149(69)90242-2. [DOI] [PubMed] [Google Scholar]
- Gordon A. J., Vagueiro M. C., Barold S. S. Endocardial electrograms from pacemaker catheters. Circulation. 1968 Jul;38(1):82–89. doi: 10.1161/01.cir.38.1.82. [DOI] [PubMed] [Google Scholar]
- Goswami M., Gould L., Gomprecht R. F., Imperiale A. Perforation of the heart by flexible transvenous pacemaker. JAMA. 1971 Jun 21;216(12):2013–2014. [PubMed] [Google Scholar]
- Grant A. F., Jones D., Hunyor S. N., Coles D. M. Implantation of permanent cardiac pacemakers. Med J Aust. 1973 Aug 11;2(6):257–261. doi: 10.5694/j.1326-5377.1973.tb128820.x. [DOI] [PubMed] [Google Scholar]
- Hurwitz B. J., Obel I. W. Longstanding cardiac perforation by a permanent endocardial pacing electrode. Chest. 1973 Sep;64(3):366–368. doi: 10.1378/chest.64.3.366. [DOI] [PubMed] [Google Scholar]
- Kramer D. H., Moss A. J., Shah P. M. Mechanisms and significance of pacemaker-induced extracardiac sound. Am J Cardiol. 1970 Mar;25(3):367–371. doi: 10.1016/s0002-9149(70)80018-2. [DOI] [PubMed] [Google Scholar]
- Lagergren H., Johansson L., Schüller H., Kugelberg J., Bojs G., Alestig K., Linder E., Borst H. G., Schaudig A., Giebel O. 305 cases of permanent intravenous pacemaker treatment for Adams-Stokes syndrome. Surgery. 1966 Mar;59(3):494–497. [PubMed] [Google Scholar]
- Meyer J. A., Millar K. Perforation of the right ventricle by electrode catheters: a review and report of nine cases. Ann Surg. 1968 Dec;168(6):1048–1060. doi: 10.1097/00000658-196812000-00016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Obel I. W., Marchand P., Millar R. N. Cardiac pacemaking: experience with the first 120 patients treated in Johannesburg. S Afr Med J. 1972 Dec 23;46(51):2002–2009. [PubMed] [Google Scholar]
- Ormond R. S., Rubenfire M., Anbe D. T., Drake E. H. Radiographic demonstration of myocardial penetration by permanent endocardial pacemakers. Radiology. 1971 Jan;98(1):35–37. doi: 10.1148/98.1.35. [DOI] [PubMed] [Google Scholar]
- Schaudig A., Thurmayr R., Zenker R. Results of transvenous pacing. J Cardiovasc Surg (Torino) 1971 Jul-Aug;12(4):281–285. [PubMed] [Google Scholar]
- ZION M. M., OBEL I. W., MARCHAND P., FULLER D. N., BARLOW J. B. ELECTRICAL PACING OF THE HEART IN ATRIOVENTRICULAR BLOCK. S Afr Med J. 1964 Oct 3;38:781–786. [PubMed] [Google Scholar]