Abstract
Veno-arterial extracoporeal membrane oxygenation (VA ECMO) is increasingly used for acute and refractory cardiogenic shock. Yet, in clinical practice, monitoring of cardiac loading conditions during VA ECMO can be cumbersome. To this end, we illustrate the validity and clinical applicability of a real-time cardiovascular computer simulation, which allows to integrate hemodynamics, cardiac dimensions and the corresponding degree of VA ECMO support and ventricular loading in individual patients over time.
Electronic supplementary material
The online version of this article (doi:10.1186/s12967-015-0760-1) contains supplementary material, which is available to authorized users.
Keywords: Extracorporeal life support (ECLS), Veno-arterial extracorporeal membrane oxygenation (ECMO), Real-time cardiovascular computer simulation, Cardiogenic shock, Heart failure
Background
With great interest we read the recent publication by Ostadal et al. Their experimental work underscores the clinical relevance to closely monitor left-ventricular (LV) loading conditions in cardiogenic shock supported by venoarterial extracorporeal membrane oxygenation (VA ECMO) [1]. They report on significant negative influences on intrinsic cardiac output, left ventricular performance and stroke work as a result of increasing extracorporeal blood flow intended to maximally support the circulation in cardiogenic shock. It is increasingly recognized that LV overloading during VA ECMO may cumulate in feared complications as pulmonary edema, acute lung injury after ‘bridge–bridge’ strategies and even progression to irreversible myocardial damage, all having significant impact on outcome [2–5]. Yet, in clinical practice, it can be challenging to provide adequate systemic perfusion, while equally addressing favourable LV loading conditions. The practical difficulties to optimally tailor VA ECMO may be improved by a bedside monitoring tool that allows instantaneous integration of patient-specific data on cardiac dimensions, hemodynamics, fluid loading and pharmacotherapy with the degree of circulatory VA ECMO support.
Methods and results
In order to enhance bedside clinical decision support, we introduced a real-time, patient-specific simulation model of cardiovascular dynamics [6, 7]. The model is built with 32 0-D compartments in total; four cardiac chambers, pericardium, intrathoracic space and 27 vascular compartments including systemic circulation, pulmonary circulation and coronary circulation. Pressures, volumes, flows and oxygen saturations are updated with 4000 Hz in all compartments by numerical solution of differential equations with parameters based on physical dimensions of each compartment. In addition ECMO flow is simulated by connection of a virtual centrifugal pump to vascular compartments in the model. This approach allowed us to reproduce the experimental results published by Ostadal et al. [1] using their experimental baseline characteristics on hemodynamics and cardiac dimensions as input parameters for the computer model assuming a constant systemic vascular resistance (SVR). The simulation rendered output data in agreement with the original experimental data (Fig. 1). As a next step, we translated this approach into clinical practice based on a real case, which allowed us to integrate individual patient data into the simulation model: a 35 year-old female without any medical history complained of palpitations, aspecific chest pain and progressive dyspnea 2 weeks prior to admission. After progressive clinical deterioration, she presented in acute and severe cardiogenic shock refractory to conventional management, i.e., adequate fluid loading and inotropics (i.v. Milrinone 0,1 μg/kg/min; norepinephrine 300 μg/kg/min) and was therefore urgently supported by bi-femoral (17F, 21F) VA ECMO (Rotaflow PLS, MAQUET, Cardiopulmonary AG, Hirrlingen, Germany). Diagnostic work-up revealed acute myocarditis as proven by right-ventricular endomyocardial biopsies. After initiation of immunosuppressive therapy (methylprednisolone) she was successfully weaned from VA ECMO after 3 days. Serial echocardiograms were performed in order to derive cardiac dimensions according to international standards [8]. Simultaneously, heart rate, central venous pressure, arterial blood pressure, details on pharmacological management and VA ECMO support were recorded. Clinical echocardiograms and simulated pressure–volume (PV) loops corresponding to the clinical states are shown in Fig. 2 and available as Additional files 1: Echocardiogram before ECMO and 2: Echocardiogram on ECMO. The physiological effect of only increasing ECMO flow on ventricular PV loops is shown in Fig. 3. Animations of simulations are also available as Additional files 3: Simulation of Ostedal VAECMO experiment and 4: Simulation of clinical VAECMO case.
Discussion
The experimental data by Ostadal et al. are in agreement with our clinical translation enhanced by real-time cardiovascular simulation, both demonstrating unsatisfactory LV unloading during high-flow VA ECMO (Table 1). Importantly, the simulation points to a causal direct hemodynamic effect of VA ECMO on increased LV pre- and afterload as a function of augmented extracorporeal support. Alternative mechanistic explanations as valvular regurgitation, venous bronchial circulation, coronary ischemia or vegetative influences as suggested by Ostadal et al. seem less likely as significant contributors since these phenomena are not incorporated in the current simulations generating results in agreement with the experimental data. Moreover, LV dilatation and increase of end-diastolic pressure occur in the model despite a constant SVR and a less pronounced increase in systolic arterial pressure as compared to the experiments. The steep increase in arterial pressures and stroke work found experimentally while applying high flow is likely explained not only by an increase in systemic blood flow (due to ECMO) but also by an increase in SVR not commented by the authors. This increase in SVR in the experiment may be explained by myogenic vascular autoregulation or altered anesthetic depth. The clinical relevance of our findings to strive for patient-specific tailoring towards reduced VA ECMO flow in order to optimize LV loading conditions is underscored by a report that extracorporeal support flow <60 % of the theoretical individual requirement has been associated with superior clinical outcome in VA ECMO, while accounting for both adequate systemic perfusion and LV ejection avoiding additional measures of LV venting [5].
Table 1.
Before ECMO | +ECMO only | On ECMO first day | ||||
---|---|---|---|---|---|---|
Patient | Simulation | Simulation | Patient | Simulation | ||
Heart rate | /min | 137 | 137 | 137 | 133 | 133 |
Mean arterial pressure | mmHg | 85 | 85 | 106 | 98 | 98 |
Right ventricular systolic pressure | mmHg | 49 | 49 | 46 | NA | 47 |
Central venous pressure | mmHg | 10 | 10 | 9 | 8 | 9 |
Left Ventricular end‐diastolic volume | mL | 84.6 | 84.7 | 93.3 | 101.1 | 93.7 |
Left ventricular end‐diastolic pressure | mmHg | NA | 19 | 23 | NA | 23 |
Ejection fraction | % | 27.6 | 27.8 | 19.5 | 30.9 | 23.4 |
Tricuspid regurgitation velocity | cm/s | 311 | 315 | 310 | NA | NA |
ECMO flow | L/min | 0 | 0 | 2.1 | 2.1 | 2.1 |
The middle column +ECMO only shows effects of solely increasing peripheral ECMO flow from 0 to 2.1 L/min. Simulated data concerning the first day on VA ECMO are adjusted according to clinical data (HR identical, SVR and PVR decreased proportionally to achieve identical MAP, TR removed). Simulation supports the experimental findings of increased LV loading with VA-ECMO. The larger LV and EF in the patient on the first day of VA ECMO is compatible with fluid loading and partial recovery of LV function in addition to the effects of VA-ECMO as supported by clinical data
HR heart rate, SVR systemic vascular resistance, PVR pulmonary vascular resistance, MAP mean arterial pressure, TR tricuspid regurgitation, LV left ventricle and EF ejection fraction, NA data are not available
Conclusion
Real-time cardiovascular computer modeling allows to simulate complex patient-specific clinical scenario’s as in extracorporeal support. Our findings support the notion that VA-ECMO flow should be maintained as low as possible in order to balance the simultaneous need of circulatory support and LV unloading.
Authors’ contributions
MB developed the model, performed simulations, analyzed data and contributed to drafting of the manuscript. DD collected clinical data and analyzed it in relation to simulation data and also contributed to drafting of the manuscript. Both MB and DD are clinical medical doctors working with ECMO patients. Both authors read and approved the final manuscript.
Acknowledgements
This study was supported by a Grant No. 2012-2800 from the Swedish Research Council.
Competing interests
MB is the founder and owner of the company Aplysia Medical AB, Stockholm, Sweden having the commercial rights of the simulation model and software used in this work.
Compliance with ethical guidelines
The present study is an analysis of animal experimental data presented in a study by Ostadal et al. [1] and a clinical case study. The experimental study is published in J Translational Med and approved by Czech authorities in accordance with national rules and NIH rules (NIH Publication No. 85-23, revised 1985). Written informed consent to publish anonymous history, hemodynamic and imaging data was obtained from the patient in our clinical case study, who was treated in accordance with our institutional principles without any modifications related to our research.
Additional files
Contributor Information
Michael Broomé, Phone: +46 8 517 78 004, Email: michael.broome@karolinska.se, Email: broom@kth.se.
Dirk W. Donker, Email: D.W.Donker@umcutrecht.nl
References
- 1.Ostadal P, Mlcek M, Kruger A, Hala P, Lacko S, Mates M, et al. Increasing venoarterial extracorporeal membrane oxygenation flow negatively affects left ventricular performance in a porcine model of cardiogenic shock. J Transl Med. 2015;13:266. doi: 10.1186/s12967-015-0634-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Mirabel M, Luyt CE, Leprince P, Trouillet JL, Leger P, Pavie A, et al. Outcomes, long-term quality of life, and psychologic assessment of fulminant myocarditis patients rescued by mechanical circulatory support. Crit Care Med. 2011;39(5):1029–1035. doi: 10.1097/CCM.0b013e31820ead45. [DOI] [PubMed] [Google Scholar]
- 3.Boulate D, Luyt CE, Pozzi M, Niculescu M, Combes A, Leprince P, et al. Acute lung injury after mechanical circulatory support implantation in patients on extracorporeal life support: an unrecognized problem. Eur J Cardiothorac Surg. 2013;44(3):544–549. doi: 10.1093/ejcts/ezt125. [DOI] [PubMed] [Google Scholar]
- 4.Demondion P, Fournel L, Golmard JL, Niculescu M, Pavie A, Leprince P. Predictors of 30-day mortality and outcome in cases of myocardial infarction with cardiogenic shock treated by extracorporeal life support. Eur J Cardiothorac Surg. 2014;45(1):47–54. doi: 10.1093/ejcts/ezt207. [DOI] [PubMed] [Google Scholar]
- 5.Tarzia V, Bortolussi G, Bianco R, Buratto E, Bejko J, Carrozzini M, et al. Extracorporeal life support in cardiogenic shock: impact of acute versus chronic etiology on outcome. J Thorac Cardiovasc Surg. 2015;150(2):333–340. doi: 10.1016/j.jtcvs.2015.02.043. [DOI] [PubMed] [Google Scholar]
- 6.Broome M, Maksuti E, Bjallmark A, Frenckner B, Janerot-Sjoberg B. Closed-loop real-time simulation model of hemodynamics and oxygen transport in the cardiovascular system. Biomed Eng Online. 2013;12:69. doi: 10.1186/1475-925X-12-69. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Broman M, Frenckner B, Bjallmark A, Broome M. Recirculation during veno-venous extra-corporeal membrane oxygenation—a simulation study. Int J Artif Organs. 2015;38(1):23–30. doi: 10.5301/ijao.5000373. [DOI] [PubMed] [Google Scholar]
- 8.Lang RM, Badano LP, Mor-Avi V, Afilalo J, Armstrong A, Ernande L, et al. Recommendations for cardiac chamber quantification by echocardiography in adults: an update from the American Society of Echocardiography and the European Association of Cardiovascular Imaging. J Am Soc Echocardiogr. 2015;28(1):1–39. doi: 10.1016/j.echo.2014.10.003. [DOI] [PubMed] [Google Scholar]