Abstract
The fate is described of 11 patients who had a stored pulmonary valve homograft mounted on a frame or in a Dacron jacket used for mitral valve replacement. Mitral regurgitation due to holes in the attenuated cusps occurred in nine by one year, requiring re-operation or causing death. Only one patient remains well with moderate mitral regurgitation after two years.
Since January 1967, 118 patients have had mitral valve replacement with a preserved cadaver semilunar homograft valve mounted in a jacket or frame (Murray, Roschlau, and Lougheed, 1956; Heimbecker et al., 1962). In all except 11 patients the aortic valve was used.
However, in the 11 patients in whom an inverted pulmonary valve was used (Lower, Stofer, and Shumway, 1961) the clinical course has been different. The purpose of this report is to draw attention to the problems and discourage the use of this form of mitral valve replacement.
Full text
PDF


Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Angell W. W., Wuerflein R. D., Shumway N. E. Mitral valve replacement with the fresh aortic valve homograft: experimental results and clinical application. Surgery. 1967 Oct;62(4):807–813. [PubMed] [Google Scholar]
- HEIMBECKER R. O., BAIRD R. J., LAJOS T. Z., VARGA A. T., GREENWOOD W. F. Homograft replacement of the human mitral valve. A preliminary report. Can Med Assoc J. 1962 May 5;86:805–809. [PMC free article] [PubMed] [Google Scholar]
- LOWER R. R., STOFER R. C., SHUMWAY N. E. Total excision of the mitral valve and replacement with the autologous pulmonic valve. J Thorac Cardiovasc Surg. 1961 Nov;42:696–702. [PubMed] [Google Scholar]
- MURRAY G. Homologous aortic-valve-segment transplants as surgical treatment for aortic and mitral insufficiency. Angiology. 1956 Oct;7(5):466–471. doi: 10.1177/000331975600700509. [DOI] [PubMed] [Google Scholar]
- Ross D. N. Replacement of aortic and mitral valves with a pulmonary autograft. Lancet. 1967 Nov 4;2(7523):956–958. doi: 10.1016/s0140-6736(67)90794-5. [DOI] [PubMed] [Google Scholar]
- Yacoub M. H., Kittle C. F. A new technique for replacement of the mitral valve by a semilur valve homograft. J Thorac Cardiovasc Surg. 1969 Dec;58(6):859–869. [PubMed] [Google Scholar]