Skip to main content
SAGE Open Medicine logoLink to SAGE Open Medicine
. 2014 Dec 16;2:2050312114562395. doi: 10.1177/2050312114562395

The stenting strategy of drug-eluting stents for coronary artery disease in patients on dialysis

Hiroshi Fujita 1, Kenya Nasu 2, Mitsuyasu Terashima 2, Tsuyoshi Ito 1, Tomomitsu Tani 1, Takahiko Suzuki 2, Nobuyuki Ohte 1,
PMCID: PMC4712748  PMID: 26770754

Abstract

Background:

Reports regarding the relationship between the length and diameter of implanted drug-eluting stents and clinical and angiographic outcomes in dialysis patients are limited.

Aim:

We investigated the efficiency of drug-eluting stents for coronary artery disease in patients on dialysis from the viewpoint of stent sizing.

Methods:

Sirolimus-eluting stents were implanted in 88 lesions and bare metal stents were implanted in 43 lesions. We compared stenting strategy, major adverse cardiac events, and angiographic results between sirolimus-eluting stent and bare metal stent groups.

Results:

Stent diameter was smaller and stent length was longer in the sirolimus-eluting stent group than in the bare metal stent group in our routine practices. There was no significant between-group difference in late diameter loss. Rates of angiographic restenosis and target lesion revascularization were significantly higher in the sirolimus-eluting stent group than in the bare metal stent group. Although stent length was significantly longer and stent diameter was smaller in the sirolimus-eluting stent group, sirolimus-eluting stents did not improve the subsequent clinical and angiographic results compared with bare metal stents in dialysis patients.

Conclusion:

In dialysis patients, a longer length and/or smaller diameter sirolimus-eluting stent implantation was associated with high rates of restenosis and target lesion revascularization compared with bare metal stents.

Keywords: Bare metal stent, dialysis, drug-eluting stent, major adverse cardiac events


Cardiovascular disease is the major cause of mortality in dialysis patients.1 However, the optimal method of revascularization in dialysis patients has not been fully determined. Coronary artery disease in dialysis patients frequently involves diffuse and severely calcified lesions. Therefore, the outcomes of percutaneous coronary intervention (PCI) are worse in this patient population than in nondialysis patients.2,3 Since the introduction of drug-eluting stents (DES), several studies have demonstrated that DES implantation dramatically reduces the incidences of restenosis and revascularization in PCI compared with bare metal stents (BMS).48 However, several reports indicated that the effectiveness of DES implantation in dialysis patients was limited and was not superior to BMS usage. In this retrospective observational study, we tried to clarify procedural points to be noted in the use of sirolimus-eluting stent (SES) in dialysis patients.9,10

Methods

Study patients

This retrospective study was performed at the Toyohashi Heart Center Hospital. The target population consisted of dialysis patients who had at least one lesion with significant stenosis in native coronary arteries and who underwent successful PCI with SES or BMS from March 2002 through February 2006. During this study period, all patients were treated with BMS from March 2002 through July 2004, and all patients were treated with SES from August 2004 through February 2006, following the approval of SES by the Ministry of Health, Labour and Welfare in Japan. Angiographic follow-up at 9 months and clinical follow-up at 12 months after stent implantation were performed. The internal review board of the Toyohashi Heart Center Hospital approved the study protocol. All patients gave written informed consent that they accepted to participate in this study at the same time when they consented to receive PCI procedure.

Procedure and medication

According to standard care, patients were orally pre-medicated with aspirin 100 mg once daily and ticlopidine 200 mg twice daily. An intravenous bolus of heparin was administered after sheath insertion, with a repeated bolus given as needed to maintain an activated clotting time in excess of 250 s during the procedure. After the intervention, antiplatelet therapy identical to the premedication continued for at least 12 months. Other medications (e.g. β-blockers, statins, and angiotensin-converting enzyme inhibitors) were administered as appropriate.

After a variety of procedures and devices, including plain old balloon angioplasty (POBA), cutting balloon (Boston Scientific, USA), Rotablator (Boston Scientific), and directional coronary atherectomy (DCA; Boston Scientific), were used appropriately according to plaque characteristics, SES (Cypher; Cordis, USA) or BMS was implanted at the lesion site. The SES was available in sizes from 2.5 to 3.5 mm in diameter and from 13 to 33 mm in length. BMS (Bx-Velocity (Cordis), Express (Boston Scientific), Multi-Link PENTA (Abbot Vascular, USA), or DRIVER (Medtronic, USA)) were available in sizes from 2.5 to 4.0 mm in diameter and from 8 to 32 mm in length. After stent implantation, angiographic optimization was performed with high-pressure dilatation to achieve an acceptable angiographic result of <25% residual stenosis by visual estimation.

Quantitative analysis

Off-line quantitative coronary arteriography (QCA) was conducted using the view that revealed the greatest degree of stenosis. Severity of coronary stenosis was measured using the Cardiovascular Measurement System (CMS-MEDIS Medical Imaging System, Leiden, The Netherlands). Baseline, post-procedure, and follow-up angiography were evaluated in the same view. An isolated observer who was blinded to all patients’ clinical data calculated the lesion length, reference diameter, minimal luminal diameter, and diameter stenosis. Analysis of angiographic frames was performed in end-diastole.

Patient follow-up

The primary endpoint was major adverse cardiac events (MACEs) at 12 months (i.e. death, Q- and non-Q-wave myocardial infarction (MI), and target lesion revascularization (TLR)). Death was defined as all-cause death. Q- and non-Q-wave MI were defined as an increase in creatinine kinase >2 times normal and/or creatinine kinase-MB level >20 ng/mL with and/or without new pathologic Q-wave in >2 contiguous echocardiogram leads. TLR was characterized by repeated percutaneous or surgical intervention of the treated lesions. The secondary endpoint was the rate of restenosis and late lumen loss at 9 months. Binary angiographic restenosis was defined as diameter stenosis >50% in the in-stent lesion (including the stent area and 5-mm segments proximal and distal to the stent edges). Late lumen loss was calculated as the difference in minimal lumen diameters between post-procedure and follow-up.

Statistical analysis

Statistical software (Statistical Package for the Social Sciences (SPSS) version 17.0; SPSS Inc., Chicago, IL) was used for statistical analysis. Continuous variables are shown as mean ± standard deviation (SD). Categorical variables were compared using the chi-square test, and continuous variables were compared using Student’s unpaired t-test. We considered p-values <0.05 statistically significant.

Results

From March 2002 through February 2006, PCI was performed for 132 dialysis patients with 174 de novo stenosis lesions. We studied 94 patients (131 lesions) from this population base who were implanted with SES or BMS. Patients who received PCI procedures without using stents were excluded from the study. A total of 58 patients with 88 lesions received SES from August 2004 through February 2006 (DES group), and 36 patients with 43 lesions received BMS (BMS group) from March 2002 through July 2004. The baseline patient characteristics of the two groups are presented in Table 1. The mean ages were approximately 65 and 63 years in the SES group and the BMS group, respectively. There were no significant between-group differences in baseline patient characteristics. Baseline lesion characteristics are shown in Table 2. The lesion characteristics were mostly similar between the groups, except for the American Heart Association/American College of Cardiology (AHA/ACC) classification of the lesion characteristics. The B1-type lesion was more common in the BMS group than in the SES group. Procedural data are summarized in Table 3. Pre-dilatation was performed in 93% of lesions in the SES group and in 98% of lesions in the BMS group. There was no significant difference in POBA, Rotablator, or DCA usage between the two groups. Post-procedure lesion characteristics are presented in Table 3. The number of implanted stents was higher in the SES group than in the BMS group. The mean stent length was significantly longer and the mean diameter was significantly smaller in the SES group than in the BMS group. Quantitative data are shown in Table 3, and the cumulative distribution of the minimum lumen diameter (MLD) is shown in Figure 1. The MLD after stent implantation differed significantly between the groups. Angiographic follow-up data at 9 months are shown in Table 4. Follow-up angiography was performed in 80% of patients in the SES group and in 88% patients in the BMS group. At 9 months, the MLD was significantly smaller in the SES group than in the BMS group, although there was no significant between-group difference in late lumen loss. The rate of binary restenosis was significantly higher in the SES group than in the BMS group. Clinical follow-up at 12 months was obtained in 91% of patients in the SES group and in 92% of patients in the BMS group (Table 5). The incidences of MACEs did not differ between the groups when excluding TLR.

Table 1.

Patient characteristics.

BMS DES p
Number of patients 36 58
Age, mean ± SD, years 63.12 ± 9.42 65.17 ± 9.65 0.54
Male, n (%) 25 (69) 43 (74) 0.64
Diabetes, n (%) 17 (47) 35 (60) 0.29
Hypertension, n (%) 20 (56) 43 (74) 0.07
Hyperlipidemia, n (%) 5 (14) 12 (21) 0.58
Current smoking, n (%) 4 (11) 12 (21) 0.27
Family history, n (%) 4 (11) 5 (8) 0.73
Prior CABG, n (%) 18 (50) 36 (62) 0.29
Prior MI, n (%) 3 (8) 2 (3) 0.37
1VD/2VD/3VD 6/12/18 4/22/32 0.54

BMS: bare metal stent; DES: drug-eluting stent; SD: standard deviation; CABG: coronary artery bypass grafting; MI: myocardial infarction; VD: vessel disease.

Table 2.

Lesion characteristics.

BMS DES p
Number of lesions 43 88
Treated vessel
 LMT 5 5 0.3
 LAD 12 27 0.84
 LCX 8 16 0.95
 RCA 18 40 0.71
AHA/ACC lesion type
 A 0 0 NA
 B1 5 (12) 1 (1) 0.007
 B2 21 (49) 57 (65) 0.09
 C 17 (40) 30 (34) 0.57
Angiographic calcification 36 (84) 77 (86) 0.59
Bending 10 (23) 29 (33) 0.31
Significant side branch 6 (14) 24 (27) 0.12

BMS: bare metal stent; DES: drug-eluting stent; LMT: left mine trunk; LAD: left anterior descending; LCX: left circumflex; RCA: right coronary artery; AHA/ACC: American Heart Association/American College of Cardiology.

Data are presented as number (%) of patients.

Table 3.

Procedure data.

BMS DES p
Number of lesions 43 88
Procedure before stenting
 POBA, % 84 88 0.56
 Rotablator, % 42 47 0.61
 DCA, % 2 0 0.15
Direct stent, % 2 8 0.21
Mean number of stents 1.14 1.41 0.014
Mean stent diameter, mm 3.45 ± 0.45 3.10 ± 0.37 <0.0001
Mean stent length, mm 19.79 ± 9.97 33.12 ± 18.58 <0.0001
Post-dilatation balloon, mm 3.48 ± 0.52 3.22 ± 0.39 0.0023
Post-dilatation pressure, atm 17.64 ± 4.13 19.67 ± 3.22 0.0033

BMS: bare metal stent; DES: drug-eluting stent; POBA: plain old balloon angioplasty; DCA: directional coronary atherectomy.

Data are presented as mean ± standard deviation (SD), or number (%) of patients.

Figure 1.

Figure 1.

Cumulative distribution of post-procedure and 9-month follow-up minimal lumen diameter (mm) in the two groups. The post-procedure acute gain in the BMS group was well maintained throughout the follow-up period.

BMS: bare metal stent. DMS: drug eluting stent.

Table 4.

Quantitative angiographic follow-up data.

BMS DES p
Number of lesions 43 88
Pre-procedure
 Lesion length, mm 18.32 ± 9.84 20.93 ± 11.11 0.47
 Reference diameter, mm 3.39 ± 1.03 3.09 ± 0.56 0.06
 Minimal lumen diameter, mm 1.05 ± 0.55 1.00 ± 0.50 0.62
 % diameter stenosis 68.08 ± 17.70 67.63 ± 15.03 0.89
Post-procedure
 Reference diameter, mm 3.55 ± 0.90 3.35 ± 0.59 0.16
 Minimal lumen diameter, mm 3.15 ± 0.81 2.84 ± 0.50 0.014
 % diameter stenosis 10.77 ± 10.89 13.93 ± 10.35 0.47
Follow-up at 9 months n, (%) 38 (88) 70 (80)
 Minimal lumen diameter, mm 2.16 ± 0.86 1.84 ± 0.88 0.014
 % diameter stenosis 34.29 ± 17.72 37.63 ± 27.90 0.52
 Late lumen loss, mm 0.96 ± 0.57 1.00 ± 0.94 0.83
 Binary restenosis, n (%) 7 (18) 27 (39) 0.031

BMS: bare metal stent; DES: drug-eluting stent.

Data are presented as mean ± standard deviation (SD) or number (%) of patients.

Table 5.

Clinical follow-up data.

BMS DES p
Clinical follow-up
 Number of patients 33 (92) 53 (91)
MACE overall 7 (21) 28 (53) 0.0037
 All death 1 (3) 3 (6) 0.56
 Cardiac death 1 (3) 1 (2) 0.73
 Myocardial infarction
 (Q and non-Q-wave) 0 (0) 0 (0)
 CHF 0 (0) 1 (1) 0.43
 TLR 6 (18) 26 (49) 0.004

BMS: bare metal stent; DES: drug-eluting stent; MACE: major adverse cardiac event; CHF: congestive heart disease; TLR: target lesion revascularization.

All data are presented as n (%).

Discussion

The main finding of this study was that in dialysis patients who received PCI, higher incidences of binary restenosis and TLR were observed in patients that received SES than in patients that received BMS, although DES are known to be more likely to prevent restenosis than BMS. However, there was no significant difference in late lumen loss between the SES and BMS groups. Thus, a significantly smaller minimum lesion diameter in the SES group than in the BMS group after the procedure may have brought about the unfavorable outcomes.

Dialysis patients are at high risk of developing cardiovascular disease. If patients have an established cardiovascular disease, they are likely to encounter worse outcomes and higher mortality than nondialysis patients. Cardiovascular events are the primary cause of death in dialysis patients.1 Therefore, these patients often needed invasive management of coronary artery disease in order to try to reduce the frequency of angina attacks and to maintain left ventricular function and quality of life. Most dialysis patients have severely calcified, tortuous, and diffuse lesions in their coronary arteries. Gruberg et al.3 reported that the rates of TLR and event-free survival at 1 year after procedure in dialysis patients who underwent PCI were 32.1% and 41%, respectively. Several studies reported that the long-term mortality associated with surgical revascularization was lower than that associated with revascularization by PCI in dialysis patients.2 On the other hand, coronary bypass surgery is invasive in nature and carries higher risks of peri-procedural complications and in-hospital mortality.11,12 Optimal methods for treating coronary artery disease in dialysis patients remain controversial. In randomized trials, the introduction of DES dramatically reduced the incidences of restenosis and revascularization in the management of coronary artery disease compared with the use of BMS.48 The significant suppression of neointimal hyperplasia by DES reduced late lumen loss and the rate of repeat revascularization.48,13 Thus, DES is basically considered to have much greater potential to prevent restenosis compared with BMS. This story may not be true in dialysis patients. It has been acknowledged that incomplete lesion coverage is associated with the edge restenosis of DES; therefore when DES are implanted, “healthy to healthy” stenting has been the main strategy, in comparison with the “bigger and shorter are better” theory selected in the BMS era.6,7,1416 However, in dialysis patients, the lesions are severely calcified and tortuous, so that it is sometimes necessary for the distal stent to be smaller than the proximal stent in order to match the diameter of the distal reference. Thus, post-procedure MLD was generally smaller in the DES group than in the BMS group. However, late lumen loss at the 9-month angiographic follow-up was similar between the two groups in this study, still indicating the anti-restenosis potential of SES. The rates of restenosis and TLR in the SES group were 39% and 49%, respectively, and were significantly higher than in the BMS group. We consider these findings to be related to the difference in post-procedural MLD between SES and BMS groups. The “healthy to healthy parts” stent implantation strategy for dialysis patients resulted in the use of longer length and smaller diameter stenting and provoked a smaller post-procedural MLD. Meliga et al.9 recently reported in dialysis patients that the use of DES was not associated with a reduction in TLR compared with the use of BMS in patients with similar clinical characteristics, treated vessel features, and stenting strategy. Using the extensive data from United States Renal Data System, Shroff et al.10 have also reported in dialysis patients that the probability of repeat revascularization was comparable in DES patients and in patients receiving BMS. Clinical characteristics of both groups were matched in this study; however, the study includes no data about vessel descriptions or details of the stenting procedure. Previous studies indicated that in dialysis patients, revascularization was observed with similar rates between patients who received DES and BMS. Considering this information in combination with our current findings, DES was not able to overcome the problem observed in BMS usage for coronary revascularization (i.e. longer length and/or smaller diameter stents often caused restenosis). The so-called shorter length and larger diameter stenting strategy might work to reduce the likelihood of restenosis for appropriate lesions in dialysis patients, even in the DES era.1416

Study limitations

This study has several limitations. This study was conducted in a single center and is a retrospective study. This study evaluated only a small number of patients, and angiographic and clinical follow-up were not obtained in all patients. The PCI procedures were conducted by several interventionists, and the strategy of PCI (e.g. selection of stent size and the procedures performed before and after stenting) was at the discretion of each interventionist. All measurements were performed on angiograms recorded after the administration of intracoronary nitroglycerin.

Conclusion

In dialysis patients, a longer length and/or smaller diameter SES implantation was associated with high rates of restenosis and TLR, as well as a high rate of MACEs, compared with BMS. For dialysis patients, interventionists may need to carefully consider the length and diameter of DES in order to reduce rates of angiographic restenosis and TLR.

Footnotes

Declaration of conflicting interests: The authors declared no conflicts of interest with respect to the research, authorship, and/or publication of this article.

Funding: This research received no specific grant from any funding agency in the public, commercial, or not-for-profit sectors.

References

  • 1. Foley RN, Parfrey PS, Sarnak MJ. Clinical epidemiology of cardiovascular disease in chronic renal disease. Am J Kidney Dis 1998; 32: S112–S119. [DOI] [PubMed] [Google Scholar]
  • 2. Herzog CA, Ma JZ, Collins AJ. Comparative survival of dialysis patients in the United States after coronary angioplasty, coronary artery stenting, and coronary artery bypass surgery and impact of diabetes. Circulation 2002; 106: 2207–2211. [DOI] [PubMed] [Google Scholar]
  • 3. Gruberg L, Dangas G, Mehran R, et al. Clinical outcome following percutaneous coronary interventions in patients with chronic renal failure. Catheter Cardiovasc Interv 2002; 55: 66–72. [DOI] [PubMed] [Google Scholar]
  • 4. Sousa JE, Costa MA, Abizaid A, et al. Lack of neointimal proliferation after implantation of sirolimus-coated stents in human coronary arteries: a quantitative coronary angiography and three-dimensional intravascular ultrasound study. Circulation 2001; 103: 192–195. [DOI] [PubMed] [Google Scholar]
  • 5. Morice MC, Serruys PW, Sousa JE, et al. ; Randomized Study with the Sirolimus-Coated Bx Velocity Balloon-Expandable Stent in the Treatment of Patients with de Novo Native Coronary Artery Lesions. A randomized comparison of a sirolimus-eluting stent with a standard stent for coronary revascularization. N Engl J Med 2002; 346: 1773–1780. [DOI] [PubMed] [Google Scholar]
  • 6. Moussa I, Leon MB, Baim DS, et al. Impact of sirolimus-eluting stents on outcome in diabetic patients: a SIRIUS (SIRolImUS-coated Bx Velocity balloon-expandable stent in the treatment of patients with de novo native coronary artery lesions) substudy. Circulation 2004; 109: 2273–2278. [DOI] [PubMed] [Google Scholar]
  • 7. Schampaert E, Cohen EA, Schlüter M, et al. The Canadian study of the sirolimus-eluting stent in treatment of patients with long de novo lesions in small native coronary arteries (C-SIRIUS). J Am Coll Cardiol 2004; 43: 1110–1115. [DOI] [PubMed] [Google Scholar]
  • 8. Lemos PA, Serruys PW, van Domburg RT, et al. Unrestricted utilization of sirolimus-eluting stents compared with conventional bare metal stent implantation in the “real world”: the Rapamycin-eluting Stent Evaluated At Rotterdam Cardiology Hospital (RESEARCH) Registry. Circulation 2004; 109: 190–195. [DOI] [PubMed] [Google Scholar]
  • 9. Meliga E, De Benedictis M, Gagnor A, et al. Clinical outcomes following percutaneous coronary intervention with drug-eluting stents versus bare metal stents in patients on chronic hemodialysis. J Interv Cardiol 2013; 26: 351–358. [DOI] [PubMed] [Google Scholar]
  • 10. Shroff GR, Solid CA, Herzog CA. Long-term survival and repeat coronary revascularization in dialysis patients after surgical and percutaneous coronary revascularization with drug-eluting and bare metal stents in the United States. Circulation 2013; 127: 1861–1869. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11. Franga DL, Kratz JM, Crumbley AJ, et al. Early and long-term results of coronary artery bypass grafting in dialysis patients. Ann Thorac Surg 2000; 70: 813–818. [DOI] [PubMed] [Google Scholar]
  • 12. Liu JY, Birkmeyer NJ, Sanders JH, et al. Risks of morbidity and mortality in dialysis patients undergoing coronary artery bypass surgery. Northern New England Cardiovascular Disease Study Group. Circulation 2000; 102: 2973–2977. [DOI] [PubMed] [Google Scholar]
  • 13. Suzuki T, Kopia G, Hayashi S, et al. Stent-based delivery of sirolimus reduces neointimal formation in a porcine coronary model. Circulation 2001; 104: 1188–1193. [DOI] [PubMed] [Google Scholar]
  • 14. Asano T, Kobayashi Y, Mintz GS, et al. Effect of plaque volume on subsequent vessel remodeling at edges of sirolimus-eluting stents. Am J Cardiol 2006; 98: 1041–1044. [DOI] [PubMed] [Google Scholar]
  • 15. Sonoda S, Morino Y, Ako J, et al. ; SIRIUS Investigators. Impact of final stent dimensions on long-term results following sirolimus-eluting stent implantation. J Am Coll Cardiol 2004; 43: 1959–1963. [DOI] [PubMed] [Google Scholar]
  • 16. Pan M, Suárez de Lezo J, Medina A, et al. Influence of stent treatment strategies in the long-term outcome of patients with long diffuse coronary lesions. Catheter Cardiovasc Interv 2003; 58: 293–300. [DOI] [PubMed] [Google Scholar]

Articles from SAGE Open Medicine are provided here courtesy of SAGE Publications