Increased x-ray attenuation in malignant vs. benign mediastinal nodes in an orthotopic model of lung cancer

Paul Flechsig 1, Peter Choyke 1, Clemens Kratochwil 1, Arne Warth 1, Gerald Antoch 1, Tim Holland-Letz 1, Daniel Rath 1, Viktoria Eichwald 1, Peter E Huber 1, Hans-Ulrich Kauczor 1, Uwe Haberkorn 1, Frederik L Giesel 1
PMCID: PMC4712895  PMID: 26611258

Abstract

PURPOSE

Staging of lung cancer is typically performed with fluorodeoxyglucose-positron emission tomography-computed tomography (FDG-PET/CT); however, false positive PET scans can occur due to inflammatory disease. The CT scan is used for anatomic registration and attenuation correction. Herein, we evaluated x-ray attenuation (XRA) within nodes on CT and correlated this with the presence of malignancy in an orthotopic lung cancer model in rats.

METHODS

1×106 NCI-H460 cells were injected transthoracically in six National Institutes of Health nude rats and six animals served as controls. After two weeks, animals were sacrificed; lymph nodes were extracted and scanned with a micro-CT to determine their XRA prior to histologic analysis.

RESULTS

Median CT density in malignant lymph nodes (n=20) was significantly higher than benign lymph nodes (n=12; P = 0.018). Short-axis diameter of metastatic lymph nodes was significantly different than benign nodes (3.4 mm vs. 2.4 mm; P = 0.025). Area under the curve for malignancy was higher for density-based lymph node analysis compared with size measurements (0.87 vs. 0.7).

CONCLUSION

XRA of metastatic mediastinal lymph nodes is significantly higher than benign nodes in this lung cancer model. This suggests that information on nodal density may be useful when used in combination with the results of FDG-PET in determining the likelihood of malignant adenopathy.


Lung cancer is the leading cause of cancer deaths in the United States and Europe (1). Choice of therapy and prognosis is determined by the stage at which lung cancer is detected. Mediastinal nodal involvement is a significant negative prognostic sign and portends a shorter time to progression (2). Fluorodeoxyglucose-positron emission tomography-computed tomography (FDG-PET/CT) has emerged as the leading noninvasive staging method as both primary tumor and mediastinal nodes can be assessed (2, 3). Precise mediastinal N-staging is mandatory since patients with contralateral or multiregional mediastinal lymph node metastases are often excluded from primary surgery (4). The PET component of the FDG-PET/CT examination is typically assessed by measuring the maximum standardized uptake value (SUVmax) in the primary tumor and nodes. However, the FDG-PET component is sometimes equivocal due to false positive uptake in inflammatory nodes (5). The CT component is typically assessed using Response Evaluation Criteria In Solid Tumors (RECIST 1.1) criteria in the primary and nodes, and it is based on the node’s short-axis diameter (3). However, size changes are notoriously unreliable in assessing disease status. Therefore, staging often requires additional invasive methods such as transbronchial biopsy or mediastinoscopy for histologic verification (6), especially in patients that might benefit from primary surgery.

Previous reports have suggested that nodes exhibit increases in x-ray attenuation (XRA) density when they become malignant due to replacement of the fatty nodal hilum with cancer cells (5). For instance, malignant lymph nodes obtained from patients with breast cancer showed increased density on grating-based phase-contrast x-ray tomography (710). However, this observation is not routinely incorporated into clinical interpretation of PET/CT despite the ready availability of such information. In order to further investigate potential density changes in metastatic and nonmetastatic mediastinal and hilar lymph nodes, we utilized an orthotopic lung cancer model in nude rats (11), which included ex vivo micro-CT XRA of extracted lymph nodes two weeks after transthoracic tumor cell implantation (Fig. 1). Findings were correlated with histology.

Figure 1.

Figure 1

Study design with in vivo and ex vivo measurements. Step 1: Transthoracic tumor cell transplantation in the 5th intercostal space. Step 2: Within the first two weeks after transthoracic tumor cell transplantation tumor spreads in mediastinal and hilar lymph nodes (small yellow dots). Primary tumor is seen in the right lower lobe (big yellow dot). Step 3: Micro-CT examination of extracted mediastinal lymph nodes.

Methods

Animal treatment, reagents, and study design

A total of 12 male nude rats (CR:NIH-RNU), including six animals as controls, aged four weeks were acclimatized for two weeks prior to tumor implantation. Animals were housed in sterilized cages, fed autoclaved food and water ad libitum and were assessed daily for signs of distress in compliance with international standards of humane treatment of animals in research. The study was approved by the local responsible authorities and conducted according to the guidelines of the institutional review board, the Declaration of Helsinki, and the NIH guidelines for use of laboratory animals.

A total of 1×106 NCI-H460 cells were mixed 1:1 with matrigel in a final volume of 50 μL (11, 12). Prior to implantation, animals were anesthetized using Isoflurane. Cells were transthoracically injected in the fifth intercostal space using a 28G ½″ needle (12).

Two weeks after injection, animals were sacrificed and lymph nodes were extracted. Directly after extraction, lymph nodes were examined in a micro-CT (Siemens Inveon) and thereafter placed in formalin and stained with hematoxylin and eosin (HE) for histologic analysis (Fig. 2).

Figure 2 a–c.

Figure 2 a–c.

Correlation of micro-CT (a), histogram (b), and histopathologic cross-section (c). Examples of benign (blue) and malignant (red) lymph node histology are shown. Panel (a) shows micro-CT with 3D reconstruction, in axial, coronary, and sagittal views. Panel (b) shows exemplary extracts of histograms: left histogram represents a benign lymph node (blue line), right histogram represents a malignant lymph node (red line) with median densities indicated by dotted vertical lines (HU, Hounsfield unit). Panel (c) shows corresponding histologic HE-stained cross-sections (foreground in right upper corner: 40× magnification; background: 10× magnification). The left column in panel (c) shows physiologic lymph node structures (benign), while the right column shows rat lymph node with irregular lymph node tissue and infiltrates of a solid carcinoma (arrow).

Ex vivo micro-CT

CT scanning was performed using a micro-CT scanner (Siemens Inveon MultiModality). The scanning parameters were set to 60 kV, 500 μA, 400 ms exposure time, 360° rotation with 360 rotation steps, field of view 48.64×56.25 mm and binning factor 2. Under low-medium magnification the effective pixel size was 47.5 μm. CT images were reconstructed with a down sample factor of 2 using Cobra Reconstruction Software.

Postprocessing

For computing the XRA histograms, lymph nodes were segmented by a radiologist who was blinded with regard to whether the nodes came from an experimental animal or a control, using a semi-automatic method (Fraunhofer MEVIS). Semi-automated lymph node XRA histograms were obtained by identifying a seed point in the middle of each lymph node. The semi-automated segmentation generated density thresholds that identified the contours of the node. If necessary, segmentation results were corrected manually in all three dimensions. Imaging findings were correlated to histology.

Statistical analysis

Density and short-axis diameter data were summarized using descriptive statistics (mean, standard deviation [SD], median, minimum, and maximum) and illustrated in Box-Whisker-Plots (Fig. 3). Differences were considered significant at P < 0.05 in a two-sided unpaired t test. As multiple, possibly correlated lymph nodes were investigated per animal, average densities and short-axis diameters were calculated for each animal, and only these averages were used as experimental units for the t-tests.

Figure 3 a–d.

Figure 3 a–d.

Statistical evaluation of micro-CT measurements of resected lymph nodes. Panel (a) shows density of lymph nodes in micro-CT. CT density measurements of metastatic and benign lymph nodes were significantly different. Median CT density is 125.9 HU for malignant lymph nodes and 13.9 HU for benign lymph nodes (P = 0.018). Panel (b) shows density spread of resected lymph nodes with a possible cutoff value (68.6 HU). Panel (c) shows short-axis diameter (SAD) of lymph nodes: metastatic lymph nodes demonstrate higher short-axis diameter compared with benign lymph nodes (3.35 mm vs. 2.35 mm, P = 0.025). Panel (d) shows ROC curves with AUC for lymph node density and short-axis diameter.

Diagnostic properties of the parameters were investigated in a Receiver Operating Characteristic (ROC) analysis. The optimal cutoff value was defined as the cutoff maximizing the sum of sensitivity and specificity (Youden index) in detection of malignant lymph nodes.

Statistical analysis was performed using SigmaPlot (Systat Software GmbH).

Results

All removed nodes were analyzed with ex vivo micro-CT (Fig. 2). A total number of 20 malignant lymph nodes demonstrated a median CT density of 125.9 Hounsfield Units (HU) (mean±SD, 127.8±64.5 HU; Fig. 3a). Histologically negative lymph nodes (n=12) resected from the control group, demonstrated a median CT density of 13.9 HU (mean±SD, 12.6±86.1 HU; Fig. 3a). Comparison of mean CT densities between affected and unaffected lymph nodes was statistically significant (P = 0.018).

The best possible cutoff value for the discrimination between malignant and benign lymph nodes was 68.8 HU with 17 out of 20 malignant lymph nodes having a density above 68.8 HU (85%) and 10 out of 12 benign lymph nodes measuring below 68.8 HU (83%, Fig. 3b).

Histologically proven metastatic lymph nodes had a slightly higher short-axis diameter compared with histologically negative ones. Specifically, metastatic lymph nodes demonstrated a median short-axis diameter of 3.35 mm (mean±SD, 4.2±2.0 mm; Fig. 3c), while negative nodes demonstrated a median short-axis diameter of 2.35 mm (mean±SD, 2.8±1.7 mm; P = 0.025 accounting for multiple nodes per rat; Fig. 3c).

ROC analysis revealed an area under the curve (AUC) for malignancy of 0.7 for short-axis diameter and 0.87 for density changes (Fig. 3d).

Discussion

This study shows that malignant nodes have significantly higher XRA than benign nodes in an orthotopic lung cancer model in nude rats. Density differences between malignant and benign lymph nodes were more predictive of metastatic involvement than the short-axis diameter, which was reflected in a higher diagnostic AUC (87%) for density changes than for short-axis diameter (70%).

Several previous studies have suggested a correlation between XRA and node status. For instance, Tapfer et al. (7) and Momose et al. (9) used grating-based phase-contrast x-ray tomography, and demonstrated high specificity, sensitivity, and positive and negative predictive values for this methodology (10). Jensen et al. (10) reported the visibility of lymphoid follicles in nonmetastatic lymph nodes, whereas metastatic lesions had lost their ordered morphology and had higher XRA. Even though we could not detect internal structures within lymph nodes with micro-CT, we detected statistically significant increase in lymph node XRA density, which correlated with malignant histology.

This work suggests that CT density measurements may play a role in increasing or decreasing the risk assessment for malignant nodal involvement based on FDG-PET/CT. Accurate assessment of lymph node status is crucial for therapeutic planning and prognosis (1319). Currently, PET uptake is measured by SUVmax values, while the CT-component utilizes RECIST 1.1 short axis linear measurements (20, 21). Our findings suggest that lymph node density may also provide useful diagnostic information, potentially within time-saving semi-automatic measurements (22). It is also possible that dual energy CT may enable even better discrimination of XRA than single energy CT performed in our experiment (23), and that dual energy CT could give further information on nodal status based on the measurement of iodine uptake in contrast-enhanced dual-phase dual-energy CT (24).

This work has several important limitations. Only a relatively small number (n=32) of histologically proven lymph nodes were investigated on only 12 animals. Normal nodes in rats are very difficult to identify due to their small size. Moreover, the study was only conducted in one species using one tumor model, which might not be applicable to most human tumors. Certainly the density difference between positive and negative nodes in this study is higher than would be expected in patients. To what extent this is influenced by performing these measurements ex vivo on a micro-CT unit is also unknown. Therefore, future work in this field should focus on prospective trials of CT nodal density measurements in large patient cohorts. Such large numbers of subjects could permit the identification of reasonable threshold values of density that result in clinically relevant increases in the detection of malignant nodes. This might be especially helpful in cases of FDG-equivocal lymph nodes and might therefore be of substantial importance for a better exploitation of hybrid imaging in daily routine. By adding the parameter of CT density to lymph node dimensions and FDG-PET/CT activity, more accurate characterization of node status, with fewer false positives could be achieved, thus sparing patients from unnecessary interventions prior to therapy.

In conclusion, density measurement of indeterminate lymph nodes seems to be a promising and potentially clinically valuable imaging parameter that might improve noninvasive lung cancer staging. By means of density measurements one could possibly make better use of FDG-PET/CT in order to improve therapy stratifications prior to therapy.

Main points.

  • In this orthotopic lung cancer model, x-ray attenuation in metastatic mediastinal lymph nodes is significantly higher than in benign nodes.

  • Area under the curve for malignancy was higher for density-based lymph node analysis compared with size measurement, which is the current gold standard in oncologic imaging in clinical routine.

  • Findings suggest that information on nodal density may be useful when used in combination with the results of FDG-PET in determining the likelihood of malignant adenopathy, especially in FDG-equivocal lymph nodes.