Abstract
Aim
The purpose of this work is to evaluate the dosimetric parameters of a hypothetical 153Gd source for use in brachytherapy and comparison of the dosimetric parameters with those of 192Ir and 125I sources.
Materials and methods
Dose rate constant, the radial dose function and the two dimensional (2D) anisotropy function data for the hypothetical 153Gd source were obtained by simulation of the source using MCNPX code and then were compared with the corresponding data reported by Enger et al. A comprehensive comparison between this hypothetical source and a 192Ir source with similar geometry and a 125I source was performed as well.
Results
Excellent agreement was shown between the results of the two studies. Dose rate constant values for the hypothetical 153Gd, 192Ir, 125I sources are 1.173 cGyh−1 U−1, 1.044 cGyh−1 U−1, 0.925 cGyh−1 U−1, respectively. Radial dose function for the hypothetical 153Gd source has an increasing trend, while 192Ir has more uniform and 125I has more rapidly falling off radial dose functions. 2D anisotropy functions for these three sources indicate that, except at 0.5 cm distance, 192Ir and 125I have more isotropic trends as compared to the 153Gd source.
Conclusion
A more uniform radial dose function, and 2D anisotropy functions with more isotropy, a much higher specific activity are advantages of 192Ir source over 153Gd. However, a longer half-life of 153Gd source compared to the other two sources, and lower energy of the source with respect to 192Ir are advantages of using 153Gd in brachytherapy versus 192Ir source.
Keywords: Brachytherapy, Dosimetric parameters, Hypothetical 153Gd source
1. Background
Brachytherapy is an advanced cancer treatment. In brachytherapy practice radioactive seeds or sources are placed inside or in a close vicinity of the tumor, irradiating a high radiation dose to the tumor while reducing the radiation exposure to the surrounding healthy tissues. Brachytherapy is a radiation therapy modality which is accounted as localized, precise and high-technology treatment. Photon emitting sources which are currently used in brachytherapy are 60Co, 137Cs, 192Ir, 125I and 103Pd.1, 2, 3, 4 Based on the recommendation of the American Association of Physicists in Medicine (AAPM), task group No. 43 (TG-43U15), dosimetric parameters of each source should be calculated before application of the source in brachytherapy. Then, these parameters can be entered to the treatment planning system using that source. Based on this report, the dosimetric parameters should be calculated by at least two independent investigators.5 For this reason, various scientists have reported dosimetric parameters of different radioactive sources in their studies.6, 7, 8, 9, 10, 11, 12 These fundamental parameters include: air kerma strength, dose rate constant, geometry function, radial dose function and two dimensional (2D) anisotropy function. Due to the need for use in brachytherapy practice, in last few years different scientists have introduced new hypothetical brachytherapy sources and reported their dosimetric parameters. Hypothetical sources are based on the geometry of a real source but their active materials are isotopes which are not commonly used in brachytherapy or are not available commercially. In a number of previous studies, TG-43 dosimetric parameters of new hypothetical sources were calculated, compared with commercially available sources and then introduced as brachytherapy sources.2, 13, 14, 15
153Gd is one of the hypothetical sources which were introduced by Enger et al.14 According to the Oak Ridge National Laboratory (ORNL) study in 1960,16 153Gd may be a favorable brachytherapy source. This radioisotope has an intermediate energy relative to other brachytherapy sources and emits photons with energies ranging between 40 and 100 keV with half-life of 242 days. In a recent study by Adams et al.17 a novel needle, catheter and source system were presented for interstitial rotating shield brachytherapy (I-RSBT) of the prostate by 153Gd sources. Their rationale for use of a shielded source and catheter system was to reduce urethra, rectum, and bladder dose. The reason for proposal of 153Gd source rather than 125I and 192Ir sources was that the 125I source has a rapid dose fall-off in tissue and the shielding thickness for 192Ir would be large and, therefore, it could not be fitted inside catheters normally used in prostate brachytherapy. Treatment plans for I-RSBT and conventional brachytherapy treatments of a prostate cancer patient were obtained by Monte Carlo calculations. The platinum shield used could reduce the dose rate on the shielded side at 1.0 cm to 6.4% relative to the dose rate on the other (unshielded) side. This method also reduces the dose to the urethra, rectum and bladder. The treatment time for delivery of 20 Gy dose in I-RSBT with ten 62 GBq 153Gd sources was 154 min. When it is aimed to have a reasonable dosage and treatment time, multisource approach is necessary. In the study by Enger et al.14 the dosimetric parameters of a hypothetical source based on this radioisotope were calculated and reported in the form of plots. However, the results were not expressed numerically. Furthermore, the dosimetric parameters of the hypothetical 153Gd source were not compared with commercially available sources such as 192Ir. Aim is to determine the dosimetric characteristics for the hypothetical 153Gd source. The results will then be compared with those reported by Enger et al. and with the dosimetric parameters of 192Ir and 125I brachytherapy sources.
2. Materials and methods
2.1. Geometry of hypothetical 153Gd source
The source design for the hypothetical 153Gd source is similar to that of VariSource™ 192Ir source (Varian Medical Systems, Palo Alto, CA).18 This design was chosen because it was proposed in the previous study on the hypothetical 153Gd source14 and to have an appropriate comparison of the results of the two studies; it was preferred to have the same geometries. There is only one difference that the active core in the 153Gd hypothetical source was defined as a pure 153Gd radionuclide. The design and dimensions of this source are illustrated in Fig. 1. The active core of the source is in the form of a pure 153Gd cylinder (density ρ = 7.9 g cm−3) with an active length of 1.0 cm and diameter of 0.84 mm. The active core was defined in a stainless steel capsule with 11.3 mm length (including the end weld). The inner diameter of the encapsulation is 0.84 mm and its outer radius is 1.0 mm. The stainless steel encapsulation is composed of Fe (67.92%), Cr (19%), Ni (10%), Mn (2%), Si (1%) and C (0.08%) with density of 8.0 g cm−3. The length and diameter of the simulated source's guide are 5.0 mm and 0.5 mm, respectively. The energy spectrum of photons emitted by 153Gd radionuclide is tabulated in Table 1.19
Fig. 1.
A schematic diagram showing the geometry of the hypothetical 153Gd brachytherapy source (all dimensions are in millimeter).
Table 1.
The energy spectrum of 153Gd radionuclide.19
| Energy (keV) | Prevalence (%) |
|---|---|
| 5.177 | 0.374 |
| 5.816 | 0.976 |
| 5.817 | 0.1341 |
| 5.846 | 8.86 |
| 6.438 | 0.559 |
| 6.457 | 5.64 |
| 6.571 | 0.9214 |
| 6.617 | 0.0845 |
| 6.844 | 1.8512 |
| 7.484 | 0.947 |
| 7.768 | 0.173 |
| 7.795 | 0.244 |
| 14.06383 | 0.0183 |
| 21.2 | 0.0224 |
| 40.467 | 0.00953 |
| 40.902 | 35.29 |
| 41.542 | 63.516 |
| 46.905 | 6.2516 |
| 47.038 | 12.13 |
| 47.373 | 0.1848 |
| 48.249 | 4.001 |
| 48.386 | 1.597 |
| 54.1934 | 0.00842 |
| 68.2557 | 0.016223 |
| 69.673 | 2.41923 |
| 75.42213 | 0.078323 |
| 83.36717 | 0.1964 |
| 89.48595 | 0.0694 |
| 96.883 | 0.0022 |
| 97.431 | 29 |
| 103.1801 | 21.1123 |
| 118.1123 | 0.000121 |
| 166.5556 | 0.00033 |
| 172.3043 | 0.00022 |
| 172.8531 | 0.036017 |
2.2. TG-43 dosimetric parameters
Dosimetric parameters of a brachytherapy source can be determined using experimental or Monte Carlo simulation methods following the TG-43 report.5, 20 Based on this formalism, dose distribution around a brachytherapy source is determined using the following formula:
| (1) |
where SK, Λ, G(r, θ), g(r) and F(r, θ) are air kerma strength, dose rate constant, geometry function, radial dose function and 2D anisotropy function, respectively. These parameters are calculated from the following formulas:
| (2) |
| (3) |
By using a line-source approximation, for θ = 0°, the geometry function is calculated from the following equation:
| (4) |
For θ ≠ 0° the geometry function is obtained from Eq. (5):
| (5) |
| (6) |
| (7) |
2.3. Calculation of the dosimetric parameters
MCNPX code (version 2.6.0)21 was employed for stimulation of the hypothetical 153Gd source, phantom, etc. The source geometry was defined in this code and used for calculation of dosimetric parameters which were recommended by the updated task group No. 43 (TG-43U1) report of AAPM. These parameters are air kerma strength, dose rate constant, radial dose function and 2D anisotropy function. Based on Table 1, energy spectrum of the photons emitted by 153Gd was defined in source definitions in the Monte Carlo simulations of the source.
For calculation of the air kerma strength, the source was stimulated in a vacuum environment in the form of a spherical volume with 100 cm radius. In this environment, air-filled toroid tally cells in the form of rings located at 1.0–50.0 cm transverse distance from the source center were defined. These cells had 1.0 cm increments at this distance range. The thicknesses of the cylindrical rings were 0.5 cm. Energy cut off was the only variance reduction method which has been used in all input files. The energy cut off of 1 keV was defined for both electrons and photons. F6 tally was used for calculation of air kerma rate in the torus cells. This tally in MCNPX scores energy deposition averaged over a cell in terms of MeV/g. The input file for calculation of air kerma strength was run for 4.7 × 106 particles and the maximum type A Monte Carlo statistical uncertainty in the outputs was 1.2%. Finally, calculation of air kerma strength was based on Eq. (2). The input files were run on a high processing capability computer which had a central processing unit (CPU) with 12 processing cores. The characteristics of this computer are:
-
-
model of the processing core: Intel(R) Xeon(R) CPU X5675@3.07 GHz,
-
-
32 GB random access memory,
-
-
32-bit Microsoft Windows 7 operating system.
Each input file was run on a single core. On this system, the time needed for running the input file used for calculation of air kerma strength was 41 min (for scoring 4.7 × 106 particles).
To obtain dose rate constant, computation was performed by division of dose rate in water at 1.0 cm by air kerma strength. The formula of this calculation is written in Eq. (3). To obtain dose rate in water, a water phantom with 40.0 cm radius was defined and *F4 tally and related mass energy absorption coefficients of water were used. In MCNP for calculation of energy deposition, it is possible to use *F8 tally, however, to have an acceptable level of the Monte Carlo statistical uncertainty, calculation with this tally type is time consuming. Therefore, to have higher calculation performance, *F4 tally was a more suitable choice. *F4 tally in MCNPX scores energy flux averaged over a cell in terms of MeV/cm2. In the next step, when the energy flux is multiplied by the mass energy absorption coefficient, collision kerma is obtained. The latter variable can be an approximation of absorbed dose. Since the mass energy absorption coefficient depends on energy and the source is not monoenergetic, the energy spectrum emitted by the source was divided by different energy intervals. Then, *F4 tally in each energy interval was multiplied by the mass energy absorption coefficient of water in the energy interval. The mass energy absorption coefficients of water were extracted from the website of the National Institute of Standards and Technology (NIST).22 The multiplication results of energy flux by the mass energy absorption coefficient in various energy intervals were summed up.
Calculation of the radial dose function was in the same condition of dose rate constant, while the toroid tally cells, which were cylindrical rings ranging 1.0–15.0 cm distance along the transverse bisector of the source, were defined. The thickness of the toroid rings was defined as 1.0 mm for 1.0 mm to 7.0 cm distances. In the other distances (10.0–14.0 cm), the thickness was specified as 2.0 mm. *F4 tally, DE and DF cards were used for scoring dose at various radial distances. The energy cut off used was equal to 1 keV. The input file was run for 6.47 × 107 particles and the maximum type A Monte Carlo statistical uncertainty in the output results was equal to 0.5%. The calculation time for the running of the radial dose function input file on the aforementioned computer was 1408 min (to score 6.47 × 107 particle histories). Calculation of radial dose function was based on Eq. (6), using a linear-source approximation for the source with an active length of 1.0 cm.
2D anisotropy functions were calculated at distances of 0.5, 1.0, 5.0, 7.0, 10.0, 12.0 and 15.0 cm at 24 polar angles (ranging from 0° to 180°). Those data points overlapping with source or source cable were excluded in these calculations. Other details of the calculation, such as tally and energy cut off in scoring the 2D anisotropy function, was the same as that in the radial dose function. However, the thickness of the toroid rings was different for various distances (ranging from 0.04 cm for r = 0.5 and 1.0 cm; and 0.1 for other distances up to 15.0 cm). The number of photon histories in the scoring of the 2D anisotropy function was 2.5 × 108, having a maximum type A Monte Carlo statistical uncertainty of 3.6%. The processing time for the running of the 2D anisotropy function program on the described computer was 13,726 min (for scoring 2.5 × 108 particles). 2D anisotropy function was calculated from Eq. (7), as assigned by the TG-43 formalism.
Finally, the dosimetric parameters of the hypothetical 153Gd source were compared with those of this source by Enger et al. The data of Enger et al. on this source were obtained based on a personal communication (Ryan Flynn, 2014, personal communication). Furthermore, the dosimetric parameters of this source were compared with those of 192Ir and 125I sources which are two important and commercially available sources for use in brachytherapy. The dosimetric data of the VariSource 192Ir source (model classic) and BEBIG IsoSeed 125I source (model I25.S17) were extracted from the Carleton university database for brachytherapy sources for these two source models.23, 24
3. Results
Air kerma strength per GBq activity for the hypothetical 153Gd source was obtained as equal to 6.77 cGy cm2 h−1 GBq−1. Dose rate constant for this source was obtained as equal to 1.17 cGy h−1 U−1. Dose rate constant values for the VariSource 192Ir and 125I BEBIG IsoSeed sources as reported by Wang and Sloboda18 and Pantelis et al.25 studies are 1.044 cGy h−1 U−1 and 0.925 cGy h−1 U−1, respectively. It could be useful to compare the dose rate constant for the hypothetical 153Gd source obtained in this study and that in the study by Enger et al., but the value was not reported in that study. Based on our calculations, the dose rate at 1.0 cm radial distance for the 153Gd source was obtained as 7.95 cGy h−1 GBq−1. In the study by Enger et al.14 the dose rate at 1.0 cm for the hypothetical 153Gd source with 242 GBq activities having the same geometry is 13.12 Gy/h. This equals to 5.42 cGy h−1 GBq−1 dose rate at 1.0 cm.
Radial dose function values for the hypothetical 153Gd source obtained in the current study are tabulated in Table 2. Radial dose function values from the present study and the study of Enger et al., as well as the percentage differences between the values in these two studies are illustrated in Fig. 2. The maximum difference between these two data series is 2.69%. In the supplement report to the TG-43 formalism, difference within 2–5% in 2D anisotropy function was interpreted as an excellent agreement.26 For the purpose of comparison of the radial dose functions for the hypothetical 153Gd, 129Ir and 125I sources, the values were illustrated in Fig. 3.
Table 2.
Radial dose function for the hypothetical 153Gd source.
| r (cm) | g(r) |
|---|---|
| 0.1 | 0.681 |
| 0.2 | 0.745 |
| 0.3 | 0.809 |
| 0.4 | 0.857 |
| 0.5 | 0.894 |
| 0.75 | 0.958 |
| 1.0 | 1.000 |
| 1.5 | 1.062 |
| 2.0 | 1.107 |
| 2.5 | 1.134 |
| 3.0 | 1.155 |
| 3.5 | 1.162 |
| 4.0 | 1.168 |
| 4.5 | 1.166 |
| 5.0 | 1.159 |
| 6.0 | 1.130 |
| 7.0 | 1.083 |
| 10.0 | 0.926 |
| 12.0 | 0.810 |
| 15.0 | 0.642 |
Fig. 2.
(a) Radial dose function for the hypothetical 153Gd brachytherapy source obtained in the present study and by Enger et al.14 and (b) the percentage differences between two studies.
Fig. 3.
Radial dose function for the hypothetical 153Gd, 192Ir and 125I brachytherapy sources.
2D anisotropy function data for the hypothetical 153Gd source are listed in Table 3. To compare the 2D anisotropy function, values from the current study and those reported by Enger et al., the corresponding diagrams are plotted in Fig. 4(a). The percentage differences between these two studies are plotted in Fig. 4(b). The maximum difference between these two data sets is 10.87%. It should be noted that this level of difference exists only in one data point and in most of points the difference is much lower than this value. Plots of the 2D anisotropy functions for the hypothetical 153Gd, 129Ir and 125I sources are shown in Fig. 5.
Table 3.
2D anisotropy function for the hypothetical 153Gd source.
| θ (°) |
r (cm) |
||||||
|---|---|---|---|---|---|---|---|
| 0.5 | 1.0 | 5.0 | 7.0 | 10.0 | 12.0 | 15.0 | |
| 0 | – | 0.2693 | 0.5498 | 0.6356 | 0.6456 | 0.7554 | 0.7178 |
| 2 | – | – | 0.5667 | 0.6346 | 0.6960 | 0.7077 | 0.7369 |
| 5 | – | 0.2856 | 0.5744 | 0.6341 | 0.6900 | 0.7226 | 0.7337 |
| 10 | 0.7416 | 0.3472 | 0.6013 | 0.6596 | 0.7080 | 0.7327 | 0.7559 |
| 15 | – | 0.4319 | 0.6358 | 0.6891 | 0.7352 | 0.7519 | 0.7702 |
| 20 | 0.7916 | 0.5186 | 0.6794 | 0.7196 | 0.7654 | 0.7707 | 0.7969 |
| 30 | 0.8470 | 0.6693 | 0.7532 | 0.7861 | 0.8093 | 0.8262 | 0.8360 |
| 40 | 0.8948 | 0.7879 | 0.8281 | 0.8494 | 0.8685 | 0.8742 | 0.8832 |
| 50 | 0.9319 | 0.8702 | 0.8892 | 0.9036 | 0.9163 | 0.9165 | 0.9211 |
| 60 | 0.9605 | 0.9292 | 0.9384 | 0.9464 | 0.9504 | 0.9538 | 0.9533 |
| 70 | 0.9813 | 0.9702 | 0.9719 | 0.9740 | 0.9817 | 0.9813 | 0.9771 |
| 80 | 0.9957 | 0.9956 | 0.9933 | 0.9964 | 0.9981 | 0.9952 | 0.9913 |
| 90 | 1.0000 | 1.0000 | 1.0000 | 1.0000 | 1.0000 | 1.0000 | 1.0000 |
| 100 | 0.9951 | 0.9944 | 0.9913 | 0.9953 | 0.9988 | 0.9934 | 0.9962 |
| 110 | 0.9816 | 0.9690 | 0.9717 | 0.9778 | 0.9830 | 0.9796 | 0.9857 |
| 120 | 0.9618 | 0.9288 | 0.9388 | 0.9468 | 0.9430 | 0.9520 | 0.9571 |
| 130 | 0.9313 | 0.8699 | 0.8902 | 0.9041 | 0.9125 | 0.9212 | 0.9264 |
| 140 | 0.8935 | 0.7865 | 0.8263 | 0.8500 | 0.8714 | 0.8767 | 0.8836 |
| 150 | 0.8455 | 0.6738 | 0.7563 | 0.7878 | 0.8166 | 0.8280 | 0.8388 |
| 160 | 0.7919 | 0.5240 | 0.6794 | 0.7226 | 0.7593 | 0.7744 | 0.7816 |
| 170 | 0.7497 | 0.3410 | 0.6036 | 0.6579 | 0.7090 | 0.7371 | 0.7559 |
| 175 | – | 0.2679 | 0.5714 | 0.6296 | 0.6812 | 0.7100 | 0.7340 |
| 178 | – | – | 0.5603 | 0.62881 | 0.6702 | 0.7074 | 0.7255 |
| 180 | – | – | 0.5595 | 0.6606 | 0.6722 | 0.6851 | 0.8143 |
Fig. 4.
(a) 2D anisotropy function for the hypothetical 153Gd brachytherapy source obtained in the present study and by Enger et al.14 and (b) the percentage differences between two studies.
Fig. 5.
2D anisotropy function for the hypothetical 153Gd, 192Ir and 125I brachytherapy sources. Parts (a), (b), (c), and (d) are related to 0.5 cm, 1.0 cm, 5.0 cm and 10.0 cm distances, respectively.
4. Discussion and conclusion
In the present study TG-43 dosimetric parameters for a hypothetical 153Gd source were calculated and compared with previously published data and those of 192Ir and 125I brachytherapy sources. The dosimetric parameters including air kerma strength per activity, dose rate constant, radial dose function and the 2D anisotropy function obtained for the hypothetical 153Gd source can be used for this source if it would be commercially available for use in brachytherapy with the geometry which was designed in this study. The dosimetric parameters calculated herein are in agreement with the corresponding data reported by Enger et al. Therefore, with comparing the dosimetric parameters of the hypothetical 153Gd sources obtained in this study with those reported by Enger et al.,14 the simulation of the 153Gd source is validated.
A comparison of the dose rate at 1.0 cm radial distance for the 153Gd source in this study (7.95 cGy h−1 GBq−1) and the study by Enger et al.14 (5.42 cGy h−1 GBq−1) may be useful. These data have some difference. While relatively the same geometries, photon yields and energy spectra have been used in these two studies, the difference is due to the method used for dose calculation. Since in the present study the dose rate was calculated by Monte Carlo simulation, while in the study by Enger et al. an analytical method was utilized. Since the value of dose rate at 1.0 cm reported by Enger et al. is lower than the value obtained in this study and the Monte Carlo method is more precise than the analytical method, it is evident that the value reported by Enger et al., is not an overestimation of dose rate at 1.0 cm.
The results of the comparison of the radial dose function of the 153Gd source with those of 192Ir and 125I sources (Fig. 3) indicates that 192Ir has a more uniform radial dose function as compared to 153Gd source. This may be an advantage for the 192Ir source over 153Gd. Furthermore, the fall-off in the radial dose function for the 125I source is higher compared to the 153Gd and 192Ir sources. It should be noted that while the hypothetical 153Gd and VariSource 192Ir sources have the same geometries, the hypothetical 153Gd source and IsoSeed 125I source have different geometries. The differences in radial dose functions for these sources can be related to the types of the radioisotopes, photon energies of the sources, and self-absorption in the active parts. Furthermore, the difference in the geometries can also be a source of differences in the radial dose functions.
When one compares the 2D anisotropy function for various distances in Fig. 4(a), a general trend is observed with increase in distance, with an exception of 0.5 cm distance. In other words, with this general trend it is expected that the 2D anisotropy function for 0.5 cm distance should be downward to that of 1.0 cm, but a reverse effect is observed. The reason for this effect could be found by considering the definition of the 2D anisotropy function. In the definitions of the 2D anisotropy function, the dose rate for each angle at a distance is divided by the dose rate at that distance and 90°. As the point of calculations gets closer to the linear source, along the 90° angle, the dose rate decreases as it becomes dominant dose rate from the nearest point and loses the contribution from other parts of the source. Actually it can be seen from Fig. 2(a) that the radial dose functions at distances lower than 1.0 cm are higher than those at distances beyond 1 cm. Therefore, when the 2D anisotropy function of 0.5 cm is compared to 1.0 cm, there is a smaller denominator for the 0.5 cm than 1.0 cm, thus a larger 2D anisotropy function for 0.5 cm is justified.
The 2D anisotropy functions of the 153Gd, 192Ir and 125I sources (Fig. 5) imply that 153Gd has a less isotropic dose distribution in the longitudinal plane as compared to 192Ir and 125I sources. The effect can be due to the differences in photon energies, self-absorptions as well as sources’ geometries. This effect will be an advantage for the 192Ir and 125I sources over 153Gd.
153Gd emits photons with an average energy of 53.70 keV if self-shielding and encapsulation are not accounted for. The average photon energies for 192Ir and 125I sources are 360 keV and 28.37 keV, respectively. Having lower photon energy emitted by the 153Gd source as compared to 192Ir result to lower shielding requirement for the 153Gd. This will result in lower costs in room shielding constructions for the brachytherapy treatment room. On the other hand, the cost of construction of a 153Gd source is much higher than that of a 192Ir source. The half-life of 153Gd radionuclide is 242 days. The half-lives for 192Ir and 125I sources are 74 days and 59.40 days. Longer half-life of the 153Gd source minimizes the source exchange requirements. This will also reduce the times of source calibrations and commissioning costs. A comparison of the specific activities will also be useful. The specific activity of 153Gd is 5.55 TBq/g.14 The specific activity for 192Ir is 341 TBq/g. It is clear that 153Gd has a much lower specific activity than 192Ir. In summary, a more uniform radial dose function, and the 2D anisotropy function with more isotropy, having a higher specific activity are advantages of 192Ir source over 153Gd. However, lower energy of the 153Gd source with respect to 192Ir and longer half-life of 153Gd source compared to the other two sources are advantages of using 153Gd in brachytherapy versus 192Ir source.
Since it was aimed that the obtained dosimetric parameters for the 153Gd source be compared with those reported by Enger et al., the energy spectrum referenced by Enger et al. was used in the present study. However, according to a report by AAPM and European Society for Radiotherapy and Oncology (ESTRO),27 the energy spectrum which is announced by the National Nuclear Data Center (NNDC)28 is recommended to be used. For 153Gd source NNDC listed two different data sets. Rivard et al.29 have studied the influence of various photon energy spectra for three brachytherapy sources on kerma rate, dose rate and TG-43 dosimetric parameters via Monte Carlo simulations. While the differences reported for various spectra were not significant, the use of NNDC energy spectrum database was recommended in brachytherapy simulations.
As it was presented in Section 1, the rationale for the use of a shielded 153Gd source and catheter system is to reduce the dose to normal organs. However, in the study by Enger et al.,14 to have a reasonable dose rate, a relatively large (10.0 mm length and 0.84 mm diameter of active length) 153Gd source was designed; therefore, this size of source is too large to be partially shielded effectively interstitially, and inefficient due to self-attenuation. The source that is much more likely to be used clinically is that described in the study by Adams et al.17 In that study, an active diameter of 0.44 mm and active length of 10.0 mm was used for each source, while 10 such shielded sources were evaluated. Taking into account the specific activity, self-absorption, construction costs, the needed dosage and duration of an effective treatment by 153Gd source, there should be a trade-off between dosimetric effectiveness and treatment time. Reviewing the literature on 153Gd source, reveals that there are some issues with 153Gd brachytherapy that have not yet been addressed adequately. For example, no study exists on the importance of self-attenuation for 153Gd sources, and that is a very important consideration for these sources due to the relatively low photon energies emitted. The reason why this is important is that 153Gd is very expensive. As an example, the cost of a 0.9 mm diameter source relative to a 0.44 mm diameter source is much higher, then the cost to the user becomes prohibitive and, therefore, the only way to achieve a reasonable treatment time at an affordable cost is to use multiple small sources. Performing a Monte Carlo study on source dimensions with roughly the same as the commercially available sources such as GammaMed Plus, Flexisource, and VariSource and calculation of absolute dose rate at 1.0 cm distance assuming a realistic specific activity, will deliver novel results that are clinically and economically relevant. With the assumption that the overall treatment time for a typical prostate brachytherapy treatment is proportional to the dose rate at 1.0 cm for a given 153Gd source model, the number of sources and, therefore, total activity for each different design can be calculated. These issues call for further research on brachytherapy with 153Gd sources in clinical setting.
Conflict of interest
None declared.
Financial disclosure
This work was financially supported by Mashhad University of Medical Sciences (MUMS).
Acknowledgment
The authors would like to thank Mashhad University of Medical Sciences for financial support of this study.
Footnotes
This study was performed at “Medical Physics Department, Faculty of Medicine, Mashhad University of Medical Sciences, Pardis-e-Daneshgah, Vakil Abad Blvd., Mashhad, Iran”.
References
- 1.Suntharalingam N., Podgorsak E.B., Tölli H. Radiation oncology physics: a handbook for teachers and students. International Atomic Energy Agency; Vienna: 2005. Brachytherapy: physical and clinical aspects; pp. 451–484. [Google Scholar]
- 2.Enger S.A., Lundqvist H., D’Amours M. Exploring 57Co as a new isotope for brachytherapy applications. Med Phys. 2012;39(5):2342–2345. doi: 10.1118/1.3700171. [DOI] [PubMed] [Google Scholar]
- 3.Bahreyni Toossi M.T., Abdollahi M., Ghorbani M. A Monte Carlo study on dose distribution validation of GZP6 60Co stepping source. Rep Pract Oncol Radiother. 2012;18(2):112–116. doi: 10.1016/j.rpor.2012.10.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Oliveira S.M., Teixeira N.J., Fernandes L. Dosimetric effect of tissue heterogeneity for 125I prostate implants. Rep Pract Oncol Radiother. 2014;19(6):392–398. doi: 10.1016/j.rpor.2014.03.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Rivard M.J., Coursey B.M., DeWerd L.A. Update of AAPM Task Group No. 43 Report: a revised AAPM protocol for brachytherapy dose calculations. Med Phys. 2004;31(3):633–674. doi: 10.1118/1.1646040. [DOI] [PubMed] [Google Scholar]
- 6.Dini S.A., Awan S.B., Dou K. TG-43U1 parameterization of elongated RadioCoil 103Pd brachytherapy sources. J Appl Clin Med Phys. 2007;8(3):60–75. doi: 10.1120/jacmp.v8i3.2435. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Monroe J.I., Williamson J.F. Monte Carlo-aided dosimetry of the theragenics TheraSeed model 200 103Pd interstitial brachytherapy seed. Med Phys. 2002;29(4):609–621. doi: 10.1118/1.1460876. [DOI] [PubMed] [Google Scholar]
- 8.Sowards K.T. Monte Carlo dosimetric characterization of the IsoAid ADVANTAGE 103Pd brachytherapy source. J Appl Clin Med Phys. 2007;8(2):18–25. doi: 10.1120/jacmp.v8i2.2393. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Mourtada F., Mikell J., Ibbott G. Monte Carlo calculations of AAPM Task Group Report No. 43 dosimetry parameters for the 125I I-Seed AgX100 source model. Brachytherapy. 2012;11(3):237–244. doi: 10.1016/j.brachy.2011.06.002. [DOI] [PubMed] [Google Scholar]
- 10.Pirchio R., Galiano E., Saravi M. On the physical, spectral, and dosimetric characteristics of a new 125I brachytherapy source. Med Phys. 2007;34(7):2801–2806. doi: 10.1118/1.2739816. [DOI] [PubMed] [Google Scholar]
- 11.Granero D., Perez-Calatayud J., Ballester F. Monte Carlo calculation of the TG-43 dosimetric parameters of a new BEBIG Ir-192 HDR source. Radiother Oncol. 2005;76(1):79–85. doi: 10.1016/j.radonc.2005.06.016. [DOI] [PubMed] [Google Scholar]
- 12.Valicenti R.K., Kirov A.S., Meigooni A.S. Experimental validation of Monte Carlo dose calculations about a high-intensity Ir-192 source for pulsed dose-rate brachytherapy. Med Phys. 1995;22(6):821–829. doi: 10.1118/1.597484. [DOI] [PubMed] [Google Scholar]
- 13.Enger S.A., D’Amours M., Beaulieu L. Modeling a hypothetical 170Tm source for brachytherapy applications. Med Phys. 2011;38(10):5307–5310. doi: 10.1118/1.3626482. [DOI] [PubMed] [Google Scholar]
- 14.Enger S.A., Fisher D.R., Flynn R.T. Gadolinium-153 as a brachytherapy isotope. Phys Med Biol. 2013;58(4):957–964. doi: 10.1088/0031-9155/58/4/957. [DOI] [PubMed] [Google Scholar]
- 15.Ballester F., Granero D., Perez-Calatayud J. Study of encapsulated 170Tm sources for their potential use in brachytherapy. Med Phys. 2010;37(4):1629–1637. doi: 10.1118/1.3360441. [DOI] [PubMed] [Google Scholar]
- 16.Case F.N., Acree E.H., Cutshall N.H. Oak Ridge National Laboratory; Oak Ridge, TN: 1969. Production study of gadolinium-153, Report TM-2632. [Google Scholar]
- 17.Adams Q.E., Xu J., Breitbach E.K. Interstitial rotating shield brachytherapy for prostate cancer. Med Phys. 2014;41(5) doi: 10.1118/1.4870441. 051703-1–11. [DOI] [PubMed] [Google Scholar]
- 18.Wang R., Sloboda R.S. Monte Carlo dosimetry of the VariSource high dose rate 192Ir source. Med Phys. 1998;25(4):415–423. doi: 10.1118/1.598216. [DOI] [PubMed] [Google Scholar]
- 19.The Lund/LBNL nuclear data search. http://nucleardata.nuclear.lu.se/toi/nuclide.asp?iZA=640153 [accessed 19.07.14].
- 20.Nath R., Anderson L.L., Luxton G. Dosimetry of interstitial brachytherapy sources: recommendations of the AAPM Radiation Therapy Committee Task Group No. 43. Med Phys. 1995;22(2):209–234. doi: 10.1118/1.597458. [DOI] [PubMed] [Google Scholar]
- 21.Pelowitz D.B. Los Alamos National Laboratory; 2008. MCNPX user's manual, LA-CP-07-1473, Version 2.6.0. [Google Scholar]
- 22.National Institute of Standards and Technology, http://physics.nist.gov/PhysRefData/XrayMassCoef/ComTab/water.html [accessed 28.10.14].
- 23.Carleton University. http://www.physics.carleton.ca/clrp/seed_database/Ir192_HDR/VariSource_classic [accessed 19.07.14].
- 24.Carleton University. http://www.physics.carleton.ca/clrp/seed_database/I125/IsoSeed_I25.S17 [accessed 19.07.14].
- 25.Pantelis E., Papagiannis P., Anagnostopoulos G. New 125I brachytherapy source IsoSeed I25.S17plus: Monte Carlo dosimetry simulation and comparison to sources of similar design. J Contemp Brachytherapy. 2013;5(4):240–249. doi: 10.5114/jcb.2013.39631. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Rivard M.J., Butler W.M., DeWerd L.A. Supplement to the 2004 update of the AAPM Task Group No. 43 Report. Med Phys. 2007;34(6):2187–2205. doi: 10.1118/1.2736790. [DOI] [PubMed] [Google Scholar]
- 27.Perez-Calatayud J., Ballester F., Das R.K. Dose calculation for photon-emitting brachytherapy sources with average energy higher than 50 keV: report of the AAPM and ESTRO. Med Phys. 2012;39(5):2904–2929. doi: 10.1118/1.3703892. [DOI] [PubMed] [Google Scholar]
- 28.National Nuclear Data Center. http://www.nndc.bnl.gov/chart/decaysearchdirect.jsp?nuc=153Gd&unc=nds [accessed 28.10.14].
- 29.Rivard M.J., Granero D., Perez-Calatayud J. Influence of photon energy spectra from brachytherapy sources on Monte Carlo simulations of kerma and dose rates in water and air. Med Phys. 2010;37(2):869–876. doi: 10.1118/1.3298008. [DOI] [PubMed] [Google Scholar]