Skip to main content
Springer logoLink to Springer
. 2015 Nov 6;192(2):109–117. doi: 10.1007/s00066-015-0919-y

Image-guided intensity-modulated radiotherapy of prostate cancer

Analysis of interfractional errors and acute toxicity

Bildgesteuerte intensitätsmodulierte Strahlentherapie des Prostatakarzinoms

Volker Rudat 1,, A Nour 1, M Hammoud 1, A Alaradi 1, A Mohammed 1
PMCID: PMC4718949  PMID: 26545764

Abstract

Purpose

The aim of the study was to estimate interfractional deviations in patient and prostate position, the impact of the frequency of online verification on the treatment margins, and to assess acute radiation reactions of high-dose external beam image-guided intensity-modulated radiotherapy (IG-IMRT) of localized prostate cancer.

Patients and methods

IG-IMRT was performed by daily online verification of implanted fiducial prostate markers using a megavoltage electronic portal imaging device (EPID). A total of 1011 image-guided treatment fractions from 23 consecutive unselected prostate cancer patients were analyzed. The median total dose was 79.2 Gy (range 77.4–81.0 Gy). Acute radiation reactions were assessed weekly during radiotherapy using the Common Terminology Criteria for Adverse Events (CTCAE) v.4.03.

Results

A relevant combined patient set-up and prostate motion population random error of 4–5 mm was observed. Compared to daily IGRT, image guidance every other day required an expansion of the CTV–PTV (clinical target volume–planning target volume) margin of 8.1, 6.6, and 4.1 mm in the longitudinal, vertical, and lateral directions, thereby, increasing the PTV by approximately 30–40 %. No grade 3 or 4 acute radiation reactions were observed with daily IG-IMRT.

Conclusion

A high dose with surprisingly low acute toxicity can be applied with daily IG-IMRT using implanted fiducial prostate markers. Daily image guidance is clearly superior to image guidance every other fraction concerning adequate target coverage with minimal margins.

Keywords: Prostate neoplasms, Image-guided radiotherapy, Intensity-modulated radiotherapy, Planning target volume, Fiducial markers


Several meta-analyses have shown that higher doses of radiotherapy improve the biochemical relapse-free survival of patients with organ-confined prostate cancer compared to those treated with conventional-dose radiotherapy [7, 22]. Higher radiation doses potentially increase the risk of acute and late radiation toxicity. In order to keep the risk of acute and late radiation toxicity as low as possible, the radiotherapeutic high-dose region should be as small as possible.

Early studies revealed a relevant prostate motion variability [20], evaluated the patient set-up variability without image guidance [15], and estimated the treatment margins for the combined error of both factors [17]. Image-guided radiotherapy (IGRT) and reverse planned intensity-modulated radiotherapy (IMRT) are current radiation techniques commonly used to minimize the high-dose region without compromising tumor coverage for the definitive radiotherapy of localized prostate cancer. IGRT reduces the high-dose volume by minimizing the required internal margin (IM) and set-up margin (SM), thereby, downsizing the planning target volume (PTV). IMRT reduces the high-dose volume by generating a dose distribution more conformal to the PTV compared to conventional three-dimensional conformal radiotherapy (3DCRT).

In this study, IGRT was achieved with daily online verification of implanted fiducial prostate markers using an electronic portal imaging device (EPID). The goal of the study was to assess prostate motion variability and patient set-up variability, and to estimate the safety margin to accommodate for the combined error of both factors. Furthermore, the impact of the frequency of the image guidance (every fraction versus every other fraction versus no image guidance) on the CTV–PTV margin was estimated. Acute radiation reactions were assessed weekly during radiotherapy to evaluate the tolerance to high radiation doses applied using daily IG-IMRT.

Patients and methods

Patient data and preparation for treatment planning

A total of 23 consecutive, unselected patients receiving definitive radiotherapy for localized prostate cancer between December 2013 and March 2015 were analyzed. The histopathological diagnosis was established by transrectal ultrasonography (TRUS)-guided biopsy. Usually 12 cores were taken per prostate. Prior to radiotherapy, all patients underwent implantation of three prostate gold markers (1.2 × 3.0 mm in size; Civco Medical Solutions, Coralville, IA, USA) into the prostate under TRUS guidance to enable image-guided radiotherapy. The three prostate gold markers were placed in the prostate base, mid-gland, and apex. After an interval of 3 days, the patients underwent a computed tomography (CT) scan and magnetic resonance imaging (MRI) in the supine position for radiotherapy planning. The CT simulation was performed without contrast medium using a 64-slice spiral CT scanner (Somatom Sensation 64, Siemens Healthcare, Erlangen, Germany) with a slice thickness of 3 mm. The MRI was performed using a 3 T MRI scanner (Magnetron Trio, Siemens Healthcare, Erlangen, Germany). The slice thickness was 2 mm. Five sequences (axial T1w, axial T2w, coronal T2 STIR, axial T1FS) were obtained before and three sequences (axial, sagittal, and coronal T1FS) after the application of contrast media. The MRI and CT images were electronically fused using the Auto-Register method of the syngo®-based Coherence Oncologist Workspace version 2.0.52 (Siemens Healthcare, Erlangen, Germany). The Auto-Register method uses a “(Normalized) Mutual Information” algorithm which is based on information theory (entropy). The target volumes were defined using the fused images. MRI images were used for the target volume definition in particular because the apex of the prostate can be better visualized using MRI compared to CT [13, 23]. For CT simulation and radiotherapy, patients were immobilized in supine position using a headrest, kneefix, and feetfix (Civco Medical Solutions, Coralville, IA, USA). The CT simulator and the linear accelerators were equipped with identical models of a carbon index tables and positioning devices. Patients were instructed to have a comfortably filled bladder and an empty rectum for their CT and MRI examinations and for each treatment appointment.

The study was approved by the local institutional ethical committee and conducted in accordance with the Helsinki Declaration in its current version.

Image-guided radiotherapy

At the first radiotherapy fraction, megavoltage electronic portal images were taken using an electronic portal imaging device (EPID, [16]) from orthogonal directions (0° or 180° and 270° or 90°) and from the directions of the treatment beams. At all following radiotherapy fractions megavoltage electronic portal images were taken from two orthogonal directions. Processing and analysis software was used to significantly improve the image quality of the megavoltage electronic portal images [10]. Representative pelvic bony landmarks and the three fiducial prostate markers were marked on the portal images using electronic drawing tools. The images were zoomed and electronically superposed with the reference images, the corresponding digitally reconstructed radiographs (DRR) generated by the treatment planning system (TPS). A portal imaging software was used to assess the isocenter placement error in three dimensions based on the comparison of bony landmarks or fiducial prostate markers of the portal images with the corresponding reference image (Fig. 1).

Fig. 1.

Fig. 1

Online verification of bony landmarks and implanted fiducial prostate markers using megavoltage X-rays and electronic portal imaging device (EPID). a Digital reconstructed radiograph (DRR) generated by the treatment planning system. The position of the three fiducial prostate markers is marked with crosses. b Portal image (double-exposed) obtained immediately before the radiotherapy fraction using the EPID. The three dots represent the fiducial prostate markers. c Fused images of a and b. d Calculated deviation of the compared structures of images a and b

The patient set-up error was represented by the deviation of the compared bony landmarks. The prostate motion error was represented by the deviation of the compared fiducial prostate markers after matching of the bony landmarks. The combined error of both factors (referred to as “combined error”) was represented by the deviation of the compared fiducial prostate markers. Online correction of deviation of the fiducial prostate markers was done by automatic adjustment of the treatment table in three dimensions prior to every radiotherapy application in all patients.

Inverse-planned intensity-modulated radiotherapy

The target volumes were defined and the dose prescribed according to the International Commission on Radiation Units and Measurement (ICRU) Reports 50 and 62 recommendations. Accordingly, the PTV should be surrounded by the 95 % isodose line. The contouring of the PTV and organs at risk was done according to the RTOG Consensus Contouring Guidelines “Male Pelvis Normal Tissue” and other specific recommendations [13, 23]. In very low risk and low risk patients the CTV included the prostate gland. In one patient with very high risk, the CTV included the prostate, the complete seminal vesicals, and the locoregional pelvic lymph nodes. In all other patients the prostate gland and the proximal seminal vesicals were included in the CTV. The CTV to PTV margin for the prostate gland was 5 mm and for the seminal vesicals 8 mm. The IMRT plans were generated using the TPS XIO 4.4 (CMS, Inc., St. Louis, MO, U.S.A.). The normal tissue dose volume constraints provided by the TPS were used for the IMRT plans. Tissue inhomogeneities were considered in the treatment planning optimization process, and the dose calculation algorithm used was “superposition”. An optimization with 100 iterations was applied, followed by a semiautomatic segmentation (minimum 3 cm step size). Segments with less than or equal 2 MU were expelled from the plan. A step-and-shoot technique was used with usually eight equally spaced coplanar fields. The number of segments of a typical plan was around 100 and the corresponding treatment time about 15 min. Linear accelerators (Oncor Avant Garde, Siemens Medical, Erlangen, Germany) equipped with a multileaf collimator (160 leaves) and an EPID (Optivue, Siemens Medical, Erlangen Germany) were used for the treatment.

Assessment of acute radiation reactions

Acute radiation reactions were prospectively assessed by two Radiation Oncologists using the Common Terminology Criteria for Adverse Events (CTCAE) v.4.03. The acute reactions were assessed once weekly during the course of radiotherapy and 6 weeks after radiotherapy.

Statistical analysis

Individual and population based parameters of the patient set-up variability were calculated according to the report “On target: ensuring geometric accuracy in radiotherapy” by The Royal College of Radiologists [14]. Accordingly, the individual mean patient set-up error Mindividual was defined as the mean set-up error for an individual patient. The overall population mean set-up error Mpop was defined as the overall mean for the analyzed patient group. The population systematic error Σset-up was defined as the standard deviation of the individual mean set-up error about the overall mean Mpop. The individual random (daily) positioning error σindividual was defined as the standard deviation of the set-up error around the corresponding mean individual value Mindividual. The population random error σset-up was defined as the mean of all individual random errors σindividual. The patient set-up parameters were calculated for each direction (longitudinal, vertical, and lateral).

Image-guided correction of the patient set-up and prostate motion error was performed prior to every radiotherapy fraction in all patients. In order to estimate the margins required for image guidance every other radiotherapy fraction or no image guidance during radiotherapy, the patient set-up and prostate motion error before correction of the corresponding radiotherapy fractions was used for the statistical analysis. Treatment margins were calculated using the van Herk formula [21]. Accordingly, the margin required to ensure 95 % minimum dose to the PTV for 90 % of the patients was given by

graphic file with name M1.gif 1

where Σ is the square root of the quadratic sum of the standard deviations of all contributing systematic errors, σ the square root of the quadratic sum of the standard deviations of all contributing random errors, and Inline graphicthe standard deviation describing the width of the penumbra. In our analysis Inline graphicwas used as contributing systematic error, and Inline graphic and Inline graphicas contributing random errors Inline graphic. The representative standard deviation of the penumbra width Inline graphicof our linear accelerators was 4.2 mm.

Results

A total of 1011 image-guided treatment fractions from 23 subsequent unselected prostate cancer patients were obtained for analysis. The number of radiotherapy fractions per patient ranged from 43 to 45. Patient and treatment characteristics are listed in Table 1. The patient set-up variability, prostate motion variability, and the combined error of both factors (referred to as combined error) were slightly different in the three dimensions. On average, the patient set-up variability was greater than the prostate motion variability. Most probably due to differences in the bladder and rectum filling [20, 21] and respiration [4] the prostate motion variability was greater in the longitudinal and vertical direction than in the lateral direction. All errors were compatible with a Gaussian distribution (Fig. 2). The mean combined error appeared to be constant during the course of the radiotherapy (Fig. 3). Image guidance every other fraction compared to no image guidance substantially reduced the combined error. However, considerable error remains after image guidance every other fraction. The mean combined error varied considerably between and within the patients, and depending on the direction deviations greater than 3, 5, and 10 mm were found on average in 22–29 %, 10–21 %, and 4–10 % of the radiotherapy fractions after image guidance every other fraction (Fig. 4). The patient set-up variability, prostate motion variability, and the required safety margin to accommodate for the combined error in dependence of the frequency of the image guidance are listed in Table 2. Intrafractional errors were not investigated in this study.

Table 1.

Patient and treatment characteristics

Characteristics n %
Age (years)
41–50 1 4.3
51–60 3 13.0
61–70 7 30.4
71–80 12 52.2
Body mass index
Normal weight 7 30.4
Overweight 10 43.5
Obese 6 26.1
T classification
T1 5 21.7
T2 13 56.5
T3 4 17.4
T4 1 4.3
N classification
N0 23 100.0
M classification
M0 23 100.0
Risk group (NCCN guidelines)a
Very low 2 8.7
Low 2 8.7
Intermediate 6 26.1
High 7 30.4
Very high 6 26.1
Total dose (Gy)b
Median 79.2
Min–Max 77.4–81.0
ADTc
None 4 17.4
4–6 months 6 26.1
2–3 years 13 56.5

LN Locoregional lymph nodes, ADT androgen deprivation therapy.

aGuidelines of the National Comprehensive Cancer Network.

bDose per fraction was 1.8 Gy in all patients.

cADT was given during radiotherapy.

Fig. 2.

Fig. 2

Patient set-up variability, prostate motion variability, and the combined error of both factors. The red lines represent the corresponding Gaussian distribution

Fig. 3.

Fig. 3

Combined error of the patient set-up variability and prostate motion variability during the course of radiotherapy. The squares represent the mean and the vertical lines the range of the deviation

Fig. 4.

Fig. 4

Frequency of the combined error of patient set-up variability and prostate motion variability larger than threshold. The filled symbols represent the population mean, the box the standard error, and the horizontal lines the mean of individual patients

Table 2.

Patient set-up variability, prostate motion variability, and safety margin (mm) to accommodate for the combined error of both factors using the van Herk formula

Patient set-up variability Prostate motion variability Combined error
Direction M Σ σ M Σ σ M Σ σ Safety margina
No image guidance
Longitudinal − 0.7 2.8 2.8 − 1.4 1.9 2.4 − 2.1 3.4 4.3 11.4
Vertical 0.6 2.7 3.7 − 0.4 1.7 2.4 0.1 3.2 4.8 11.5
Lateral − 0.7 2.0 3.6 − 0.1 0.7 1.5 − 0.9 2.1 4.2 8.2
Image guidance every other fraction
Longitudinal − 0.5 1.5 2.4 − 0.8 1.1 2.0 − 1.2 2.3 3.8 8.1
Vertical 0.3 1.5 2.7 − 0.1 0.9 1.9 0.2 1.7 3.6 6.6
Lateral − 0.4 1.1 2.7 − 0.1 0.4 1.1 − 0.6 1.0 3.1 4.1

M Overall population mean set-up error, Σ Population systematic error, σ Population random error.

aSafety margin to accommodate for the combined error of patient set-up accuracy and prostate motion using the van Herk formula.

In two patients a migration of one of the three fiducial prostate markers was detected. In one patient a fiducial prostate marker migration was observed between the CT simulation and the first radiotherapy fraction. In another patient a migration of a fiducial prostate marker of 4 mm was observed at radiotherapy fraction number 37. A loss of a fiducial prostate marker was not observed.

Acute radiation reactions were assessed weekly throughout the course of radiotherapy and 6 weeks after radiotherapy using CTCAE v.4.03. Despite the high radiation dose of 77.4–81.0 Gy applied no grade 3 or 4 acute radiation reactions were observed. Grade 2 acute reactions were detected in 4.3–56.5 %, and grade 1 in 13.0–78.3 % of the patients (Table 3). It should be noted that the common acute reaction “urinary frequency” is categorized in CTCAE v. 4.03 as grade 0, grade 1 (“present”), and grade 2 (“limiting instrumental activities of daily living; medical management indicated”). Grade 3 or 4 has not been defined. In contrast, the Acute Radiation Morbidity Scoring Criteria of the Radiation Therapy Oncology Group (RTOG) categorizes genitourinary morbidity into grades 0–4. This should be considered if results are compared. However, only 2 patients (13 %) of our study population developed CTCAE v. 4.03 grade 2 urinary frequency (Table 3).

Table 3.

Maximal acute radiation reactions [Common Terminology Criteria for Adverse Events (CTCAE) v.4.03]

Parameter Grade n %
Dermatitis radiation
0 9 39.1
1 13 56.5
2 1 4.3
Diarrhea
0 20 97.0
1 3 13.0
Fatigue
0 7 30.4
1 13 56.5
2 3 13.0
Gastrointestinal pain
1 10 43.5
2 13 56.5
Proctitis
0 9 39.1
1 13 56.5
2 1 4.3
Urinary frequency
0 2 8.7
1 18 78.3
2 3 13.0
Weight loss
0 23 100.0

Discussion

In our study, IGRT was achieved by daily online verification of implanted fiducial prostate markers. In order to save cost in terms of increased dose and in-room time [9], the question arises whether the frequency of image guidance can be reduced from daily to every other day without losing relevant benefit.

Our data show that the combined patient set-up and prostate motion error on average remains basically constant over the course of radiotherapy. A small number of image-guided treatment fractions at the beginning of the radiotherapy course should therefore be sufficient to significantly reduce the systematic error. However, our data also reveal a population random error of 4–5 mm in all directions, and that the random error varied grossly between and within the patients. For this reason, CTV–PTV margins derived from population-based observations would lead to unnecessarily large PTVs in many patients. The impact of the random error on the CTV–PTV margin can be significantly reduced by daily online verification of the prostate position with necessary corrections applied before delivery of treatment. Our data show that with this approach in combination with the use of IMRT high doses of 77.4–81.0 Gy can be delivered with surprisingly low acute radiation toxicity. According to the van Herk formula, reducing the frequency of image-guided fractions to every other fraction would require an expansion of the CTV–PTV margin of 4–8 mm. This additional margin would increase the PTV by approximately 30–40 % in a typical prostate cancer patient. It is likely that an increase of the PTV of this magnitude will significantly increase the risk of toxicity at the high radiation doses prescribed.

Similar results have been reported by Kupelian et al. [11]. In their study, prostate cancer patients were treated with helical tomotherapy and megavoltage computed tomography images were used for image guidance with intraprostatic metallic fiducials. In agreement with our study, imaged guidance every other day compared to daily image guidance would have increased the CTV–PTV margin by 4–7 mm using the van Herk formula. The authors concluded that high-dose external beam radiotherapy for localized prostate cancer delivered with tight treatment margins requires daily image guidance.

Clinical data concerning toxicity of IG-IMRT in daily practice are scarce. Takeda et al. [19] reported about 141 patients with localized prostate cancer treated with IG-IMRT to a total dose of 76 Gy (n = 13) and 80 Gy (n = 128). No grade 3 or 4 acute toxicities were observed. The incidence of grade 2 acute gastrointestinal (GI) and genitourinary (GU) toxicities were 1.4 and 8.5 %, respectively. The 5-year actuarial likelihood of grade 2–3 GI and GU late toxicities were 6 and 6.3 %, respectively. There was no grade 4 GI or GU late toxicity. Wortel et al. [25] assessed the acute radiation toxicity of patients treated to 78 Gy with either IG-IMRT (n = 260) or 3D-CRT (n = 215) using toxicity questionnaires distributed at baseline, prior to fraction 20 and 30, and at 90 days after treatment. IG-IMRT resulted in significantly lower overall RTOG grade 2 or higher GI toxicity (29 versus 49 %, respectively, p = 0.002) and overall GU grade 2 and higher toxicity (38 versus 48 %, respectively, p = 0.009). Michalski et al. [12] compared 491 patients treated with 3D-CRT and 257 with IMRT to a total dose of 79.2 Gy. For grade 2 and higher acute gastrointestinal/genitourinary (GI/GU) toxicity, both univariate and multivariate analyses showed a statistically significant decrease in grade 2 and higher acute collective GI/GU toxicity for IMRT. There was a trend for a clinically meaningful reduction in late grade 2 and higher GI toxicity with IMRT. Guckenberger et al. [6] analyzed 150 prostate cancer patients treated with dose-escalated, moderately hypofractionated cone-beam CT based IG-IMRT with a simultaneous integrated boost (SIB) technique. Acute genitourinary (GU) toxicity grade 1–2 was observed in 85 % of the patients. Gastrointestinal (GI) toxicity was mild with more than 80 % of the patients free from any GI toxicity during follow-up. Two patients suffered from late grade 3 GI toxicity. The rate of GU toxicity grade 2 or higher was less than 10 % at 6–12 months but increased continuously to 22.4 % at 60 months; grade 3 GU toxicity remained below 5 % during follow-up. Crehange et al. [3] evaluated the impact of PTV reduction when delivering IG-IMRT for patients with prostate cancer. The median dose prescribed to the prostate was 78 Gy (range 74–78 Gy). The incidence of grade 2 late genitourinary toxicity was 7.0 % for patients with a CTV–PTV margin of 5 mm (n = 87) and 6.6 % for patients with a CTV–PTV margin of 10 mm (n = 78; p = 1.00). The incidence of grade 2 late gastrointestinal toxicity was 1.2 and 2.6 % (p = 0.38), respectively.

The limitations of tracking the prostate position using implanted fiducial prostate markers should be mentioned. Changes of the prostate shape, rotational changes of the prostate position [5, 24] as well as intrafractional errors are not assessed with this method. Intrafractional prostate motion variability and CTV–PTV margin recommendations reported in various series have been nicely summarized by Skarsgard et al. [18]. A more sophisticated approach to minimize the PTV and to optimize the dose distribution would be “Adaptive Radiotherapy” where an adaption of the dose distribution can be achieved by daily CT-based image guidance [1, 8] in combination with gating or tracking of the target, thereby, considering inter- and intrafractional changes [2].

Conclusion

A high dose with surprisingly low acute toxicity can be applied with daily IG-IMRT using implanted fiducial prostate markers for the definitive external beam radiotherapy of localized prostate cancer. Daily image guidance is clearly superior compared to image guidance every other fraction concerning adequate target coverage with minimal margins.

Compliance with ethical guidelines

Conflict of interests

V. Rudat, A. Nour, M Hammoud, A. Alaradi, A. Mohammed state that there are no conflicts of interests.

All studies on humans described in the present manuscript were carried out with the approval of the responsible ethics committee and in accordance with national law and the Helsinki Declaration of 1975 (in its current, revised form). Informed consent was obtained from all patients included in studies

References

  • 1.Boda-Heggemann J, Lohr F, Wenz F, et al. kV cone-beam CT-based IGRT: a clinical review. Strahlenther Onkol. 2011;187:284–291. doi: 10.1007/s00066-011-2236-4. [DOI] [PubMed] [Google Scholar]
  • 2.Combs SE, Debus J, Feick G, et al. Current state of the art, multimodality research and future visions for the treatment of patients with prostate cancer: consensus results from “Challenges and Chances in Prostate Cancer Research Meeting 2013”. Radiat Oncol. 2014;9:224. doi: 10.1186/s13014-014-0224-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Crehange G, Mirjolet C, Gauthier M, et al. Clinical impact of margin reduction on late toxicity and short-term biochemical control for patients treated with daily on-line image guided IMRT for prostate cancer. Radiother Oncol. 2012;103:244–246. doi: 10.1016/j.radonc.2011.10.025. [DOI] [PubMed] [Google Scholar]
  • 4.Dinkel J, Thieke C, Plathow C, et al. Respiratory-induced prostate motion: characterization and quantification in dynamic MRI. Strahlenther Onkol. 2011;187:426–432. doi: 10.1007/s00066-011-2201-2. [DOI] [PubMed] [Google Scholar]
  • 5.Graf R, Boehmer D, Budach V, et al. Interfraction rotation of the prostate as evaluated by kilovoltage X-ray fiducial marker imaging in intensity-modulated radiotherapy of localized prostate cancer. Med Dosim. 2012;37:396–400. doi: 10.1016/j.meddos.2012.02.006. [DOI] [PubMed] [Google Scholar]
  • 6.Guckenberger M, Lawrenz I, Flentje M. Moderately hypofractionated radiotherapy for localized prostate cancer: long-term outcome using IMRT and volumetric IGRT. Strahlenther Onkol. 2014;190:48–53. doi: 10.1007/s00066-013-0443-x. [DOI] [PubMed] [Google Scholar]
  • 7.Hou Z, Li G, Bai S. High dose versus conventional dose in external beam radiotherapy of prostate cancer: a meta-analysis of long-term follow-up. J Cancer Res Clin Oncol. 2015;141:1063–1071. doi: 10.1007/s00432-014-1813-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Huttenrauch P, Witt M, Wolff D, et al. Target volume coverage and dose to organs at risk in prostate cancer patients. Dose calculation on daily cone-beam CT data sets. Strahlenther Onkol. 2014;190:310–316. doi: 10.1007/s00066-013-0483-2. [DOI] [PubMed] [Google Scholar]
  • 9.Keilholz L, Willner J, Thiel HJ, et al. Time management in radiation oncology: evaluation of time, attendance of medical staff, and resources during radiotherapy for prostate cancer: the DEGRO-QUIRO trial. Strahlenther Onkol. 2014;190:17–25. doi: 10.1007/s00066-013-0440-0. [DOI] [PubMed] [Google Scholar]
  • 10.Kirby MC, Glendinning AG. Developments in electronic portal imaging systems. Br J Radiol. 2006;79:S. doi: 10.1259/bjr/21517185. [DOI] [PubMed] [Google Scholar]
  • 11.Kupelian PA, Lee C, Langen KM, et al. Evaluation of image-guidance strategies in the treatment of localized prostate cancer. Int J Radiat Oncol Biol Phys. 2008;70:1151–1157. doi: 10.1016/j.ijrobp.2007.07.2371. [DOI] [PubMed] [Google Scholar]
  • 12.Michalski JM, Yan Y, Watkins-Bruner D, et al. Preliminary toxicity analysis of 3-dimensional conformal radiation therapy versus intensity modulated radiation therapy on the high-dose arm of the Radiation Therapy Oncology Group 0126 prostate cancer trial. Int J Radiat Oncol Biol Phys. 2013;87:932–938. doi: 10.1016/j.ijrobp.2013.07.041. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Perna L, Fiorino C, Cozzarini C, et al. Sparing the penile bulb in the radical irradiation of clinically localised prostate carcinoma: a comparison between MRI and CT prostatic apex definition in 3DCRT, Linac-IMRT and Helical Tomotherapy. Radiother Oncol. 2009;93:57–63. doi: 10.1016/j.radonc.2009.04.004. [DOI] [PubMed] [Google Scholar]
  • 14.Radiologists TRCo. On target: ensuring geometric accuracy in radiotherapy. http://www.rcr.ac.uk/docs/oncology/pdf/BFCO%2808%295_On_target.pdf. Accessed 2 Nov 2015
  • 15.Rudat V, Flentje M, Oetzel D, et al. Influence of the positioning error on 3D conformal dose distributions during fractionated radiotherapy. Radiother Oncol. 1994;33:56–63. doi: 10.1016/0167-8140(94)90086-8. [DOI] [PubMed] [Google Scholar]
  • 16.Rudat V, Hammoud M, Pillay Y, et al. Impact of the frequency of online verifications on the patient set-up accuracy and set-up margins. Radiat Oncol. 2011;6:101. doi: 10.1186/1748-717X-6-101. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Rudat V, Schraube P, Oetzel D, et al. Combined error of patient positioning variability and prostate motion uncertainty in 3D conformal radiotherapy of localized prostate cancer. Int J Radiat Oncol Biol Phys. 1996;35:1027–1034. doi: 10.1016/0360-3016(96)00204-0. [DOI] [PubMed] [Google Scholar]
  • 18.Skarsgard D, Cadman P, El-Gayed A, et al. Planning target volume margins for prostate radiotherapy using daily electronic portal imaging and implanted fiducial markers. Radiat Oncol. 2010;5:52. doi: 10.1186/1748-717X-5-52. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Takeda K, Takai Y, Narazaki K, et al. Treatment outcome of high-dose image-guided intensity-modulated radiotherapy using intra-prostate fiducial markers for localized prostate cancer at a single institute in Japan. Radiat Oncol. 2012;7:105. doi: 10.1186/1748-717X-7-105. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Ten Haken RK, Forman JD, Heimburger DK, et al. Treatment planning issues related to prostate movement in response to differential filling of the rectum and bladder. Int J Radiat Oncol Biol Phys. 1991;20:1317–1324. doi: 10.1016/0360-3016(91)90244-X. [DOI] [PubMed] [Google Scholar]
  • 21.Van Herk M, Remeijer P, Rasch C, et al. The probability of correct target dosage: dose-population histograms for deriving treatment margins in radiotherapy. Int J Radiat Oncol Biol Phys. 2000;47:1121–1135. doi: 10.1016/S0360-3016(00)00518-6. [DOI] [PubMed] [Google Scholar]
  • 22.Viani GA, Stefano EJ, Afonso SL. Higher-than-conventional radiation doses in localized prostate cancer treatment: a meta-analysis of randomized, controlled trials. Int J Radiat Oncol Biol Phys. 2009;74:1405–1418. doi: 10.1016/j.ijrobp.2008.10.091. [DOI] [PubMed] [Google Scholar]
  • 23.Villeirs GM, De Visschere PJ, Fonteyne VH, et al. Magnetic resonance imaging in diagnosis, staging and radiotherapy planning for prostate cancer. Acta Clin Belg. 2010;65:371–376. doi: 10.1179/acb.2010.65.6.001. [DOI] [PubMed] [Google Scholar]
  • 24.van de Water S, Valli L, Aluwini S, et al. Intrafraction prostate translations and rotations during hypofractionated robotic radiation surgery: dosimetric impact of correction strategies and margins. Int J Radiat Oncol Biol Phys. 2014;88:1154–1160. doi: 10.1016/j.ijrobp.2013.12.045. [DOI] [PubMed] [Google Scholar]
  • 25.Wortel RC, Incrocci L, Pos FJ, et al. Acute toxicity after image-guided intensity modulated radiation therapy compared to 3D conformal radiation therapy in prostate cancer patients. Int J Radiat Oncol Biol Phys. 2015;91:737–744. doi: 10.1016/j.ijrobp.2014.12.017. [DOI] [PubMed] [Google Scholar]

Articles from Strahlentherapie Und Onkologie are provided here courtesy of Springer

RESOURCES