Abstract
Delicate structures (such as biological samples, organic films for polymer electronics and adsorbate layers) suffer degradation under the energetic probes of traditional microscopies. Furthermore, the charged nature of these probes presents difficulties when imaging with electric or magnetic fields, or for insulating materials where the addition of a conductive coating is not desirable. Scanning helium microscopy is able to image such structures completely non-destructively by taking advantage of a neutral helium beam as a chemically, electrically and magnetically inert probe of the sample surface. Here we present scanning helium micrographs demonstrating image contrast arising from a range of mechanisms including, for the first time, chemical contrast observed from a series of metal–semiconductor interfaces. The ability of scanning helium microscopy to distinguish between materials without the risk of damage makes it ideal for investigating a wide range of systems.
Scanning helium microscopy uses neutral atoms to image traditionally challenging materials (e.g. delicate, insulating and magnetic samples) non-destructively with absolute surface sensitivity. This work reports the first observation of chemical contrast in helium microscopy via inelastic scattering.
Microscopy is an essential tool for the discovery, application and fabrication of new materials, structures and devices. Moreover, with modern fabrication taking advantage of an ever-broader library of new materials, microscopy techniques need to be applicable to a wide range of organic and inorganic samples1,2. However, there exists a range of systems that remain challenging to image, such as transparent, fragile, weakly bonded, insulating, very rough and magnetic samples3,4. Neutral helium atoms are the ideal probe of such systems owing to their low energy, lack of net charge or spin and short de Broglie wavelength. Indeed, the favourable properties of neutral helium atoms as a surface probe have already been exploited for many years in the diffraction-based technique of helium atom scattering (HAS)4,5. Scanning helium microscopy (SHeM) is a spatially resolved variant of HAS that operates analogously to a scanning electron microscope, with the electron beam replaced with a beam of neutral helium atoms6,7.
Based on extensive work with HAS, the nature of the probe–sample interaction is well understood at a qualitative level, and research into better quantitative analyses is an active topic in the field of surface scattering5,8,9,10,11,12,13. The neutral helium atoms backscatter from the outer electronic corrugation of the sample, thus giving the technique its absolute surface sensitivity and non-destructive qualities. The possible scattering pathways give rise to contrast in the collected image; a critical consideration since useful imaging depends not only on resolution, but on the contrast available. Although the field of atom optics is well established4,14,15,16,17 and SHeM is already exceeding the resolution limits of traditional optical microscopy18, the specific scattering mechanisms by which SHeM image contrast arises is a new area of research. To date, predictions of possible SHeM contrast modes have been purely speculative with no direct comparisons of experimental contrast with theory yet available4.
Here we present the first observation of chemical contrast originating from inelastic effects in neutral atom microscopy. SHeM images of different ultrathin patterned metal films on silicon substrates show strong chemical (but weak topological) contrast. Altering the mean energy of the helium beam results in a significant reduction of image contrast, thus providing an unambiguous observation of an inelastic scattering-based process. Finally, we show that current theory is not yet capable of fully explaining the observed contrast.
Results
Topological contrast
Topological contrast is the dominant mechanism in SHeM images of microscopically rough specimens. Deviations from a perfect plane scatter the helium away from the specular channel (‘diffuse scattering'), yielding Michelson contrast C:
where θ is the detector angle and δ is the angular mismatch of two scattering planes4. As such, changes in the surface morphology will influence the intensity recorded at each pixel of the image. Further topological contrast in the form of shadowing and masking is possible if either the beam or detector is completely occluded due to a surface asperity. Figure 1 shows a comparison of an optical and SHeM micrograph (taken with an instrument detailed previously7) of a section of a wing from the honey bee species Apis mellifera. The complex folds of the membrane of the wing are the almost indiscernible in the optical image due to the transparency of the membrane material and the range of sample plane heights (in Fig. 1a, the distance from the sample slide to wing top is ∼1.5 mm). However, all of these features are readily apparent in the SHeM image. The absolute surface sensitivity of the helium atoms means that only features on the top side of the wing are observable. For example, masking of the incident helium beam is visible from both the hairs on the wing surface (Fig. 1b) and where the wing rests on the substrate. Thus, SHeM produces intuitive images of biological samples with no sample preparation required and no risk of beam damage to the substrate.
Chemical contrast
While topological contrast is readily observed, the properties of the helium probe particle have been predicted to yield weaker, more exotic mechanisms4. For example, the composition and local atomic character of a sample surface should also give rise to variations in the helium reflectivity through the structure factor of the scattering centres. The mean energy and momentum of a neutral helium beam with a de Broglie wavelength of the order Ångstroms is well-matched to those of phonon-induced surface charge–density oscillations, making it capable of interacting with dynamic surface processes. Indeed, an atomic beam of helium is ideal to probe such processes since its small size and mass (as compared to heavier noble species) minimizes the lattice displacement and hence helium atoms will excite or de-excite the largest number of vibrational modes8,19. Traditionally, the Debye–Waller factor (DWF) is used to describe the variation in specular (‘in-plane') reflectivity (I/I0) via interactions with phonons. One definition of the factor has the form:
and describes an atomic beam of energy Ei, mass m, incident on a surface of atomic mass M at angle φi, temperature T and Debye temperature ΘD. D is the potential well depth of the interaction of the helium atom with the surface and k is the Boltzmann constant5. The prevalence of the vibrations can thus be seen to be highly dependent on sample composition. For example, a material with a rigid lattice or high molecular mass will divert less of the helium signal away from the specular channel (‘out-of-plane' scattering). In addition, the presence of adsorbates on the sample surface does not necessarily prevent such interactions since the surface charge density is impacted by the motion of atoms buried deep below the surface and the incident helium atoms can probe these subsurface resonances9,10. In principle, therefore, the helium–electron–phonon coupling can provide chemical contrast with which to characterize a surface even in the presence of multiple adsorbate layers.
HAS studies have demonstrated that the helium signal that is inelastically scattered is small in comparison with the elastic contribution—typically 2–3 orders of magnitude lower8,19. To achieve an unambiguous observation of any chemical contrast effects, the materials and sample geometry were carefully selected to minimize any competing contrast mechanisms. Using electron beam lithography, a 15 nm thick patterned layer of gold was laid down on a silicon (100) substrate with a native oxide layer; a 3 nm titanium layer was added first to facilitate the wetting of the silicon surface. Figure 2a shows a SHeM micrograph of the metal on silicon sample, with a topological feature provided by a dust particle, which had come to rest on part of the gold layer. As can be clearly seen in Fig. 2, the nanometre-thin metal layer is visible against the silicon substrate, but the contrast is much weaker than the topological contrast; consistent with the expected result for an inelastic process. Figure 2b–d show the image progression as the gold on silicon sample was moved back from the specular position in 500 micron increments. Although the detector acceptance angle in the current apparatus is large, the distance moved is sufficient to place the sample specular position outside the detector's acceptance cone. While the contrast between the gold and silicon diminishes with increasing sample distance from the specular position, there is no trend in the contrast between the dust particle and underlying silicon as a function of sample position. This observation indicates that while the contrast in the dust particle is purely topological in nature, the change in contrast for the gold logo cannot be due to a simple step height or mean plane topological feature. It is possible that a consistent difference in surface roughness between the gold and silicon (on a scale much smaller than the instrument spot size) could explain the observed inter-material contrast. To eliminate this possibility, atomic force microscope (AFM) studies were conducted using a Cypher scanning probe microscope with sub-nanometre resolution in both lateral and axial directions to determine the degree of surface irregularity in both materials. It was found that the gold had a root mean square (r.m.s.) roughness of 2.8 nm, while the silicon yielded a value of 1.2 nm. The level of observed roughness is orders of magnitude smaller than the beam spot size and thus it seems unlikely that this difference in roughness is the cause of the observed contrast. Furthermore, a coarser surface would (in general) be expected to cause a higher degree of diffuse scattering away from specular, and thus would appear darker in a SHeM micrograph; the inverse to the observed results in Fig. 2.
To determine whether contrast could be observed with other metal/semiconductor combinations, 40-nm-thick patterned layers of gold, chromium, nickel and platinum were laid down on the same silicon substrate with a 3 nm titanium wetting layer, in the same manner as the prior 15 nm gold sample. Figure 3 shows a composite image of the different metallic layers under SHeM. It should be noted that in each of the component images in Fig. 3, the incident helium intensities were normalized using the mean silicon signal as the reference. The SHeM images clearly show that there is distinct contrast between each metal and the silicon substrate as well as between the metal species. The mean r.m.s. roughness of the gold, nickel, platinum, chromium and silicon measured by AFM was 2.1, 0.9, 1.6, 2.6 and 1.4 nm, respectively. Given that the observed contrast does not follow the roughness trend in the AFM data, and that the same wetting layer was used for each sample, it seems unlikely that changes in diffuse scattering arising from the nanometre scale surface irregularities can explain the contrast differences in Fig. 3.
To investigate the source of the contrast further, the energy of the incident helium was varied by either warming or cooling the stagnation volume of our supersonic free-jet expansion. The 15 nm gold on silicon sample was scanned with a range of beam energies (as shown by the series of images in Fig. 4) with Michelson contrast observed to diminish with the reduction in energy. As a control study, a copper transmission electron microscope grid on a silicon wafer substrate was imaged across the same range of mean beam energies, with no trend in Michelson contrast observed. The SHeM is currently limited by the detection count rate through Poisson statistics, and thus changes in beam intensity (and hence signal:noise ratio) as a function of beam energy also influence image quality. However, simulated SHeM micrographs based on the contrast observed at room temperature and matched to the experimentally observed signal-to-noise ratio at each beam energy could not explain the reduction in image quality (Fig. 4 inset). Moreover, the simulation was only able to replicate the data set through a direct reduction in material contrast as a function of temperature.
The angular distribution of scattered helium, which controls contrast in the SHeM, is related to the local atomic scale character of the surface convolved with the beam spot and detector aperture7. The local vibrational modes influence the degree of inelastic scattering, while the average structural form factor of the scatterers controls diffuse elastic scattering. While inelastic scattering is strongly dependent on beam energy (equation 2), diffuse elastic scattering varies only weakly through the much smaller associated change in momentum of the beam. Hence, given the large reduction in contrast as a function of decreasing beam energy observed in Fig. 4, we can conclude that the dominant contrast displayed by the metal–silicon interface samples is a consequence of inelastic processes; that is, helium–electron–phonon interactions.
Discussion
Using the DWF as a starting point, one can attempt to predict the expected contrast due to lattice vibrations. Using the model presented by MacLaren et al.4, it was found that while the contrast between the nickel and platinum in Fig. 3 qualitatively matched the model, the contrast between the gold and chromium was found to be inverted from that predicted. This result is perhaps not unexpected; early work with the scattering of helium from surfaces (which attempted to verify the DWF for gas–surface interactions) produced consistent agreement with the DWF for some systems, yet for others it was wholly inadequate11,12. Attempts have been made in recent times to redefine the DWF in terms of quantum gas–surface interactions13, but it is clear that the complex vibrational behaviour of materials remains an active area of research within the field of surface scattering. It is also important to note that these observations do not rely on the specific form of the metal surface layer. The metals utilized in this study are either noble metals, or form thin, passivating oxide layers and as such were chosen to reduce the complexity of any subsequent comparisons to theory. Nevertheless, oxide or other physisorbed layers of contamination will certainly be present in our ex situ prepared samples. However, it is clear that chemical contrast from the underlying material is still evident in the micrographs in Figs 2, 3, 4. Work in this laboratory is underway to allow the addition of a low-damage sample-cleaning method to the SHeM instrument to create the ultraclean, well-ordered systems needed to examine the origins of both surface and subsurface contrast in more detail. Modelling the interaction of helium with even these ideal systems remains an open problem. Indeed, the development of SHeM provides both a new opportunity, and an impetus, for further research into the complex vibrational behaviour of materials.
Thus far, we have shown that by utilizing existing detector technology and simple atom optics, the current SHeM instrument demonstrates both topological and chemical contrast. However, looking ahead there is a further contrast mechanism that is, in principle, available to the SHeM instrument. Under normal scattering conditions, thermal helium atoms have a de Broglie wavelength comparable to the typical crystallographic dimensions. As a result, the backscattered helium atoms produce diffraction patterns characteristic of the surface corrugation potential to yield information about the surface structure. To resolve these diffraction patterns, typical HAS apparatus have an angular resolution of less than half a degree5. With a working distance of 2.8 mm, the current SHeM instrument has a beam spot size of 5.4 μm and detector acceptance angle of 14.5 degrees. The measured intensity involves a convolution of the atomic form factor spatially with the beam spot, and angularly with the acceptance angle, therefore, precluding the observation of individual diffraction peaks. Consequently, diffractive effects do not play a role in the generated contrast. Furthermore, it is well-known from prior HAS studies5 that even small amounts of disordered adsorbates lead to a reduction in the intensity of the helium diffraction peaks, and so sample cleaning would be a necessary requirement. To adapt the SHeM to achieve HAS-like angular resolution, a 3–4 order of magnitude increase in detector sensitivity is required (based on the reduction in detector aperture size), which can be realized by utilizing a solenoidal ion source20,21. Enhanced angular resolution would give the SHeM access to an imaging mode analogous to the dark field imaging of electron microscopy. The ability to resolve the local order of surfaces, even when composed of the same material, would provide critical insight into a number of current surface science investigations. An example includes the study of organic photovoltaics, where it has been shown that the crystallinity of the polymers at interfaces within the device directly affects performance22.
In summary, the experiments presented in this paper demonstrate that contrast in SHeM arises not only due to surface topology, but also the local chemical environment through inelastic interactions. Furthermore, we have clearly shown that the observed contrast does not quantitatively match present theoretical models due to the complexity of helium–electron–phonon coupling. The observation of chemical contrast in SHeM offers the prospect of new theoretical studies in this field. Most importantly, based on the technique's surface sensitivity, non-destructive inert nature, potential for nanometre resolution and range of novel contrast mechanisms, SHeM is already able to be applied to a large variety of systems as a complimentary technique to the existing surface analysis tool set.
Methods
Operation of the Mk II SHeM
Figure 5 shows a schematic diagram of the Mk II SHeM used in this study7. In the source chamber (1), a helium free-jet beam expansion (maximum pressure of 240 bar, temperature adjustable from 100 to 400 K) is created with the aid of a 10 micron nozzle, before the centreline of the expansion is selected out by a 100 micron skimmer (Beam Dynamics, Inc., model 2). The resultant helium beam enters the differential stage (2), where it is progressively apertured by the pinhole plate. In the tip of the pinhole plate a silicon nitride disc (Ted Pella part no. 21525) with a focused ion beam-milled pinhole forms the final optical element, leaving a small spot to strike the sample surface in the sample chamber (3), as shown in the inset. The helium backscattered from the sample enters the detector chamber (4) through a 1-mm diameter aperture, where it stagnates to form a stable pressure, subsequently measured by a Hiden HAL/3F PIC mass spectrometer. By rastering the sample back and forth, an image of the surface may be constructed. With the aim of the work presented here being to investigate the contrast available to the technique, a 5-μm pinhole was utilized in all images shown to keep the count rates high.
Additional information
How to cite this article: Barr, M. et al. Unlocking new contrast in a scanning helium microscope. Nat. Commun. 7:10189 doi: 10.1038/ncomms10189 (2016).
Acknowledgments
This research was supported under the Australian Research Councils Discovery Projects (Project No. DP08831308) funding scheme. Postgraduate research scholarships (M.B., A.F.) from the University of Newcastle gratefully acknowledged. We thank the Newcastle and Cavendish workshops, Donald MacLaren and Kane O'Donnell for technical support, insightful discussions and assistance. This work was performed in part at both the Materials and ACT nodes of the Australian National Fabrication Facility, which is a company established under the National Collaborative Research Infrastructure Strategy to provide nano- and microfabrication facilities for Australia's researchers.
Footnotes
Author contributions W.A., J.E., A.P.J. and P.C.D. developed the helium atom microscope concept; M.B., A.F., A.P.J., D.J.W. and P.C.D. realized the practical implementation of the instrument; M.B., A.F. and J.M. designed and constructed the instrument utilized in this paper; M.B., A.F. and J.M. made the measurements and performed the data analysis detailed in this paper; M.B., A.F., J.M. and P.C.D. wrote the manuscript, and A.F. prepared the figures. All authors read and commented on the manuscript.
References
- Glenn D. R. et al. Single-cell magnetic imaging using a quantum diamond microscope. Nat. Methods 12, 736–738 (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Shapiro D. A. et al. Chemical composition mapping with nanometre resolution by soft X-ray microscopy. Nat. Photon. 8, 765–769 (2014). [Google Scholar]
- Krivanek O. L. et al. Vibrational spectroscopy in the electron microscope. Nature 514, 209–212 (2014). [DOI] [PubMed] [Google Scholar]
- MacLaren D. A., Holst B., Riley D. J. & Allison W. Focusing elements and design considerations for a scanning helium microscope (SHeM). Surf. Rev. Lett. 10, 249–255 (2003). [Google Scholar]
- Farias D. & Rieder K. Atomic beam diffraction from solid surfaces. Rep. Prog. Phys. 61, 1575–1664 (1998). [Google Scholar]
- Barr M. et al. A design for a pinhole scanning helium microscope. Nucl. Instr. Meth. Phys. Res. B 340, 76–80 (2014). [Google Scholar]
- Fahy A., Barr M., Martens J. & Dastoor P. C. A highly contrasting scanning helium microscope. Rev. Sci. Instrum. 86, 023704 (2015). [DOI] [PubMed] [Google Scholar]
- Hulpke E. Helium Atom Scattering from Surfaces Springer (1992). [Google Scholar]
- Benedek G. et al. Theory of surface phonons at metal surfaces: recent advances. J. Phys. Condens. Matter. 22, 084020 (2010). [DOI] [PubMed] [Google Scholar]
- Benedek G. et al. Unveiling mode-selected electron-phonon interactions in metal films by helium atom scattering. Phys. Chem. Chem. Phys. 16, 7159–7172 (2014). [DOI] [PubMed] [Google Scholar]
- Comsa G. The applicability of the Debye-Waller factor to the He scattering from metal surfaces. J. Phys. C Solid State Phys. 6, 2648 (1973). [Google Scholar]
- Krishnaswamy S., Derry G., Wesner D., O'Gorman T. & Frankl D. Debye-Waller effects in atom-surface scattering. Surf. Sci. 77, 493–504 (1978). [Google Scholar]
- Gumhalter B. Single- and multi-phonon atom-surface scattering in the quantum regime. Phys. Rep. 351, 1–159 (2001). [Google Scholar]
- Koch M. et al. Imaging with neutral atoms—a new matter-wave microscope. J. Microsc. 229, 1–5 (2008). [DOI] [PubMed] [Google Scholar]
- Eder S., Reisinger T., Grebe M., Bracco G. & Holst B. Focusing of a neutral helium beam below one micron. New J. Phys. 14, 073014 (2012). [Google Scholar]
- Barredo D. et al. A quantum-stabilized mirror for atoms. Adv. Mater. 20, 3492–3497 (2008). [Google Scholar]
- Eder S. D. et al. Focusing of a neutral helium beam with a photon-sieve structure. Phys. Rev. A 91, 0436088 (2015). [Google Scholar]
- Witham P. & Sanchez E. Increased resolution in neutral atom microscopy. J. Microsc. 248, 223–227 (2012). [DOI] [PubMed] [Google Scholar]
- Doak R. B. in Atomic and Molecular Beam Methods Vol. 2, ed. Scoles G. Oxford University Press (1992). [Google Scholar]
- Martens J. et al. Design of a permanent magnet alternative for a solenoidal ion source. Nuc. Instr. Meth. Phys. Res. B 340, 85–89 (2014). [Google Scholar]
- Alderwick A., Jardine A., Hedgeland H., MacLaren D. & Allison W. Simulation and analysis of solenoidal ion sources. Rev. Sci. Instrum. 79, 123301 (2008). [DOI] [PubMed] [Google Scholar]
- Van Bavel S., Barenklau M., de With G., Hoppe H. & Loos J. P3HT/PCBM bulk heterojunction solar cells: impact of blend composition and 3D morphology on device performance. Adv. Funct. Mater. 20, 1458–1463 (2010). [Google Scholar]