Abstract
Reliable measurement of total testosterone is essential for the diagnosis, treatment and prevention of a number of hormone-related diseases affecting adults and children. A mass spectrometric method for testosterone determination in human serum was carefully developed and thoroughly validated. Total testosterone from 100 μL serum is released from proteins with acidic buffer and isolated by two serial liquid–liquid extraction steps. The first extraction step isolates the lipid fractions from an acidic buffer solution using ethyl acetate and hexane. The organic phase is dried down and reconstituted in a basic buffer solution. The second extraction step removes the phospholipids and other components by hexane extraction. Liquid chromatography–isotopic dilution tandem mass spectrometry is used to quantify the total testosterone. The sample preparation is automatically conducted in a liquid-handling system with 96-deepwell plates. The method limit of detection is 9.71 pmol/L (0.280 ng/dL) and the method average percent bias is not significantly different from reference methods. The performance of this method has proven to be consistent with the method precision over a 2-year period ranging from 3.7 to 4.8% for quality control pools at the concentrations 0.527, 7.90 and 30.7 nmol/L (15.2, 228, and 886 ng/dL), respectively. This method provides consistently high accuracy and excellent precision for testosterone determination in human serum across all clinical relevant concentrations.
Keywords: Testosterone, Steroids, Hormone, LC-MS/MS
1. Introduction
Concentrations of testosterone in serum aid in the diagnosis and treatment of diseases such as hypogonadism [1,2], polycystic ovary syndrome [3], androgen deficiency in men [4–7], precocious or delayed puberty [8–10], and certain cancers [11–14]. Research findings describe associations of testosterone concentrations in blood with diabetes [15, 16], osteoporosis [17], cardiovascular diseases [15,18] and increased mortality [19,20]. Such associations have stimulated additional research studies and resulted in new clinical practice applications.
A common element to all these clinical and research applications is the need for reliable and accurate testosterone measurements. Considerable inaccuracy of testosterone assays has been described, especially at testosterone concentrations commonly observed in women and children [10,21–25]. Mass spectrometric methods were found to show improved accuracy and lower variability than immunoassays, especially at low testosterone concentrations [21,22,24]. This can be attributed in part to the higher specificity of this assay technology which is achieved by isolating the analyte from the sample matrix prior to analysis and by mass and structure-specific detection [26]. However, studies comparing routine clinical mass spectrometric assays against a reference measurement procedure showed inter-method variability in precision and accuracy [25]. Though less pronounced than the variability observed with immunoassays, the performance of many mass spectrometric assays remains inadequate for testosterone determination in all clinical and research settings. This created the need for developing new mass spectrometry-based methods for routine testing of testosterone that are suitable for measuring testosterone in men, women and children with appropriate accuracy, precision and throughput. To meet these needs, we developed a mass spectrometry-based method for the measurement of testosterone at all concentration levels in the general population.
2. Material and methods
2.1. Materials and chemicals
Testosterone traceable to Australian National Measurement Institute reference material (NMI M914) was obtained from Cerilliant (Round Rock, TX) and [2,3,4-13C3]-testosterone (13C3-testosterone) from IsoSciences (King of Prussia, PA). Ethyl acetate, hexane, 0.1% formic acid in acetonitrile, 0.1% formic acid in water, sodium acetate, and ammonium carbonate, glacial acetic acid, and ammonium hydroxide were acquired from Fisher Scientific (Suwannee, GA) and ethanol from Sigma-Aldrich (St. Louis, MO). All solvents were HPLC grade and chemicals were reagent grade. Steroids used for interference testing were obtained from Steraloids (Newport, RI), Sigma-Aldrich and Cerilliant and prepared at the concentration 6.93 nmol/L (200 ng/dL) in a solvent solution (water/acetonitrile/formic acid, 90:10:0.1, v/v/v).
Frozen human serum samples and charcoal processed (one time or six times) sera were purchased from Bioreclamation (Hicksville, NY). Three levels of quality control (QC) materials were prepared by pooling units of human sera from healthy donors (QC low: 0.527 nmol/L or 15.2 ng/dL, QC medium: 7.90 nmol/L or 228 ng/dL, QC high: 30.7 nmol/L or 886 ng/dL).
Commercially prepared, de-identified single donor human serum samples were obtained from several blood banks in the U.S. and serum specimens from a large study were used to evaluate the method. Study participants provided informed consent. The study protocol was approved by the CDC Research Ethics Review Board.
2.2. Sample preparation
Serum samples were thawed, mixed at room temperature and processed together with quality control samples, reagent blank, and calibrators. Calibrators were prepared in ethanol from 1 mg/mL certified testosterone solution with no more than 3 serial dilutions. Thirteen calibrator levels covering a range of 0.0347–34.7 nmol/L (1.00–1000 ng/dL) were processed together with each set of samples, quality control samples, and reagent blank. Samples and reagents were pipetted in 96-well plates (2 mL well volume, Seahorse Labware, Chicopee, MA) using a Hamilton Microlab STARLet Liquid Handler instrument (Reno, NV). The well plates were sealed using a PTFE coated sealing mat (ArctiSeal, Arctic White LLC, Bethlehem, PA). Serum samples (100 μL) and internal standard solution (IS, 13C3-testosterone, 100 μL, 3.43 nmol/L or 100 ng/dL in ethanol) were combined and mixed for 30 min at room temperature. Buffer solution (100 μL, 0.5 mol/L sodium acetate, pH 5.5) was added, and the sample solution was mixed for 2 h. Liquid–liquid extraction (400 μL, ethyl acetate: hexane, 60:40, v/v) was performed twice. The solvents of the combined organic layers were evaporated using a Genevac EZ-2 evaporation system (Valley Cottage, NY) and the sample extract was re-dissolved in buffer solution (150 μL, 0.2 mol/L ammonium carbonate, pH 9.8). The sample solution was extracted twice using hexane (300 μL each) and the combined organic layers were evaporated. The sample extract was reconstituted (150 μL, 0.1% formic acid in water: 0.1% formic acid in Acetonitrile, 80:20, v/v) for liquid chromatography tandem mass spectrometry (LC-MS/MS) analysis.
2.3. LC-MS/MS
Electrospray ionization LC-MS/MS analysis was carried out using an AB Sciex 5500 triple quadrupole mass spectrometry instrument (Foster City, CA) with a Shimadzu LC-10 AD VP HPLC system (Columbia, MD). Chromatographic separation was achieved at 40 °C using a reversed phase column (C18 Hypersil Gold, 50 × 3 mm, 3 μm, Fisher Scientific Suwanee, GA) protected by a guard column (10 × 3 mm, 3 μm) with the same material. The analyte was eluted from the column using a gradient with the eluent changing from 10% to 70% acetonitrile in water within 12 min. The column was washed for 3 min in 95% acetonitrile in water and equilibrated for 3 min at the initial eluent composition. Both solvents contained 0.1% formic acid. The flow rate and injection volume were 500 μL/min and 40 μL, respectively. Total testosterone was detected by selected reaction monitoring (SRM) in the positive ion mode, quantified by the ion transition m/z 289 → 97 for testosterone and 292 → 100 for IS. For confirmation of the analytes, the transitions m/z 289 → 109 for testosterone and 292 → 112 for IS were used.
The ionspray voltage and source temperature were 4.75 kV and 650 °C, respectively. Declustering, entrance and exit potentials were 96 V, 8 V and 12 V, respectively. Curtain gas, ion source gas 1, ion source gas 2, and collision gas were set to 45 psi, 35 psi, 55 psi, and 10 psi, respectively. The collision energies for quantitation ions of both testosterone and IS were 27 eV, and confirmation ions 31 eV.
2.4. Data analysis
The LC-MS/MS raw data were processed using Analyst software (version 1.5.1 or higher, AB Sciex, Foster City, CA). SAS (version 9.2) was used to define quality control limits and to evaluate analytical runs against these limits using a multi-rule quality control approach [27]. The limit of detection was determined by using an AB Sciex 5500 triple quadruple instrument according to the procedures previously described [28].
We used the average sum of squared residuals (ASSR) and the average relative sum of squared residuals (RASSR) from 20 sets of calibration runs to choose the best fitting regression model from among linear and polynomial models with no weighting, weights of 1/X, 1/X2, or 1/(Variance of Y). The ASSR was calculated by summing up the squared differences between the model predicted and observed Y values. The RASSR was calculated by dividing the ASSR by the average of the Y values. A linear calibration curve using a weight of 1/X was selected because this model had the smallest ASSR among all linear models and the second smallest RASSR among all models. Two quadratic models had slightly smaller ASSR values than the model chosen, but their RASSR values were larger than the RASSR values of the model chosen.
The measurement accuracy was calculated by analyzing 172 sera with reference values assigned by reference methods operated at CDC with the Joint Committee for Traceability in Laboratory Medicine (JCTLM) code C8RMP6 [29,30], NIST with the JCTLM code C4RMP2 [31] and the University of Ghent with the JCTLM code NRMeth-7 [32] as part of the CDC Hormone Standardization Program (Host) [29,33]. The measurement bias was assessed following CLSI protocol EP9-A2 [34] using a mean bias of ±6.4% against the reference method and a total error of ±16.7% for individual measurements as criteria [30,35]. Serum-based reference material SRM 971 at 2 concentrations was obtained from NIST, and ERM-DA345 and ERM-DA346 from LGC Standards for accuracy assessment. Correlation of our method with the reference methods using Deming regression and bias plot analysis was performed with Analyse-it (Analyse-it Software, Ltd., Leeds, United Kingdom, version 2.26).
Repeatability and method precision for low, medium and high QC pools were determined following CLSI protocol EP5-A2 [36] on 71 different days (two results on each of two runs per day) over a period of 2 years, using 3 calibrator lots and multiple operators.
Sample matrix effects (ME) were evaluated on 6 different matrices including ethanol, saline, one time charcoal processed serum, six times charcoal processed serum, male serum and female serum as previously described [30,37]. The ME was assessed using the following equation, ME% = B/A × 100, where B refers to the area count ratios of testosterone to 13C3-testosterone obtained from samples in matrix and A in matrix free solution.
The extraction efficiency was assessed in triplicate using medium QC pools and adding the IS solution before sample preparation (A), and by adding the IS solution at the end of the sample preparation (B), before injecting the sample in the LC-MS/MS analysis. The efficiency was calculated using the following equation: measured value (B)/measured value (A) × 100.
Specificity was determined with two different approaches. In the first approach, potentially interfering compounds such as structural steroid analogs were added to samples to assess co-elution, while in the second approach quantitation ion/confirmation ion (QI/CI) ratios obtained with patient samples were compared against those obtained with calibrators for each analytical run. The testosterone QI/CI ratios of calibrators and of 56 serum samples from the same batch of the above accuracy testing experiment were compared. Co-elution was assumed when the QI/CI ratio differed more than 20% [30,38,39].
Method ruggedness was assessed by comparing measurement values of the medium QC obtained with the described method against those obtained after modifying 5 main sample preparation parameters. The first parameter (P1) is the equilibrium time, during which sera are incubated with IS to achieve equilibrium between free and protein-bound IS (incubation durations tested: 20, 30 and 40 min). The second parameter is the pH of the dissociation buffer (P2), which is used to release testosterone from sex hormone binding protein and other proteins (pH levels tested: pH 5.2, 5.5 and 5.8). The third parameter is the time allowed for removing testosterone from proteins (P3, incubation times tested: 90, 120 and 150 min). The fourth parameter is the buffer concentration used to remove testosterone form proteins (P4, buffer concentrations tested: 0.3, 0.5 and 0.7 mol/L). The last parameter (P5) is the pH of buffer used in the second extraction step to remove polar lipids such as phospholipids (pH levels tested: pH 9.4, 9.8 and 10.2).
3. Results and discussion
The aim of this study was the development of an analytical method for measuring total testosterone in sera from men, women and children. Thus, this method needs to be able to measure this analyte over a wide concentration range starting at very low concentrations typically observed in children. The limits of detection (LOD) and quantitation (LOQ) [28] were determined to be 9.71 pmol/L (0.280 ng/dL) and 32.9 pmol/L (0.950 ng/dL), respectively. The linearity of this method was verified for the concentration range from 17.3 pmol/L (0.500 ng/dL) to 45.1 nmol/L (1300 ng/dL). To assure accurate and consistent calibration over such a wide concentration range, we use 13 calibrator levels and apply a weighted linear regression model. The sensitivity and measurement range of our method is sufficient for measuring testosterone levels in the general population. The signal-to-noise ratio ranged from 4.7 to 11.7 for 20 donor samples with testosterone concentration from 50 pmol/L to 140 pmol/L (0.40 to 4.10 ng/dL).
There was no undetectable testosterone concentrations in 250 individual donor samples from men and women age 6 years and older with serum testosterone concentrations ranging from 0.03 nmol/L (0.920 ng/dL) to 35.49 nmol/L (1024 ng/dL) (Table 1). A representative LC-MS/MS selected ion chromatograms of female samples were shown on Fig. 1.
Table 1.
Serum testosterone concentrations in individuals at difference age groups and sex.
| Age group | Gender | Testosteronemedian (range) |
|---|---|---|
| 6–11 years | Female (n = 50) | 0.18 (0.03–0.91) nmol/L5.20 (0.92–26.14) ng/dL |
| Male (n = 50) | 0.17 (0.06–10.77) nmol/L4.94 (1.75–310.63) ng/dL | |
| 18 years and older | Female (n = 75) | 0.80 (0.08–3.33) nmol/L23.02 (2.36–95.91) ng/dL |
| Male (n = 75) | 18.06 (1.10–35.49) nmol/L520.93 (31.66–1023.74) ng/dL |
Fig. 1.
LC-MS/MS selected ion chromatograms of the quantification transitions for testosterone (top) and IS (bottom) in a female patient serum sample (1.00 nmol/L, 29 ng/dL).
The lack of sufficient accuracy and reliability of testosterone measurements is a major concern of the clinical and public health communities [40,41]. The Centers for Disease Control and Prevention is addressing this concern through its Hormone Standardization Program that evaluates assay accuracy using predefined performance criteria [35]. Using 172 serum samples with target values assigned by the National Institute for Standards and Technology, the University of Ghent and the CDC reference method covering a concentration range from 0.145 nmol/L (4.19 ng/dL) to 34.9 nmol/L (1007 ng/dL) showed no significant difference between our method and established reference methods with Deming regression (intercept: −0.01, 95% CI: −0.05–0.04 nmol/L, slope: 0.99, 95% CI: 0.98–1.00, Fig. 2a) and with plot analysis (mean bias: −1.2%, 95%CI: −5.9–4.1%, Fig. 2b). Individual measurements on all 172 samples met the 16.7% total error criterion. The differences between target values of certified serum-based reference materials from NIST and LGC Standards are in average −1.3% (Table 2), which is within ±6.4% criterion for routine, clinical methods [30,35]. Furthermore, the mean bias observed with our method is smaller than those reported for other methods [25,42], especially at testosterone concentrations typically observed in women and children.
Fig. 2.
Deming regression (a) and bias plot (b) between described method and an established reference method using 172 individual health patient serum sample.
Table 2.
Accuracy traceability to JCTLM listed materials and NIST serum reference materials.
| Samples | Certified value | Measured value (N = 3) | %Bias |
|---|---|---|---|
| ERM DA345 | 19.7 nmol/L 570 ng/dL |
19.4 nmol/L (95% CI: 18.4–20.5) 560 ng/dL (95% CI: 530–590) |
−1.8 |
| ERM DA346 | 0.890 nmol/L 25.5 ng/dL |
0.881 nmol/L (95% CI: 0.849–0.912) 25.4 ng/dL (95% CI: 24.5–26.3) |
−0.4 |
| SRM 971 M | 22.3 nmol/L 643 ng/dL |
22.5 nmol/L (95% CI: 21.9–23.0) 649 ng/dL (95% CI: 633–664) |
0.9 |
| SRM 971 F | 0.961 nmol/L 27.8 ng/dL |
0.926 nmol/L (95% CI: 0.739–1.11) 26.7 ng/dL (95% CI: 21.3–32.1) |
−4.0 |
The consistency of measured values obtained after changing 5 sample preparation parameters suggests that this method is minimally affected by small variations in sample processing parameters (Fig. 3). This is reflected in part by the high long-term precision of this method. The method precision over 2 years covering 71 different days was 4.8%, 3.7%, and 3.8% for QC low, QC medium and QC high, respectively. The repeatability (within-run precision) was 3.7, 2.2, and 2.0% respectively. The method precision is well below the suggested maximum impression for total testosterone measurements of 5.3% [30,35] and smaller than the precision reported for other mass spectrometry methods [25,42].
Fig. 3.
Ruggedness testing (n = 7; error bar: standard deviation). *Five main method parameters — P1: equilibration time of serum and internal standard (20, 30, and 40 min); P2: pH of dissociation buffer (5.2, 5.5, 5.8); P3: dissociation time of testosterone from binding proteins (90,120, and 150 min), P4: concentration of dissociation buffer (0.3, 0.5 and 0.7 mol/L); and P5: pH of deprotonation buffer (9.4, 9.8 and 10.2).
Method sensitivity and precision can be affected by matrix effects that can cause ion suppression among other effects. The mean ME% determined in 6 different matrices is 100.2% (95% CI: 98.7–101.8%) as studied previously [30]. The matrices include ethanol, saline, 1 time charcoal processed serum, 6 times charcoal processed serum, male serum and female serum [30]. These findings suggest that measurements are minimally affected by the sample matrix, which might explain in part the high accuracy and precision obtained with this method. This method contains two liquid–liquid extraction steps. In the first step, serum lipids are separated from polar components. In the second step, polar lipids such as phospholipids are largely removed from the sample extract as evidenced in reduction in SRM (m/z 184–184) and precursor-ion scanning (pre m/z 184) signals that are indicative for these types of lipids [43,44]. Polar lipids typically accumulate on the column, deteriorate the separation and increase the ion suppression. A second extraction to remove polar lipids improved levels of precision (especially at low testosterone concentrations) and consistency of the chromatographic separation, and it prolonged column lifetime by about 5 times. In addition, we do not use protein precipitation, which is another common sample preparation approach [45]. Because testosterone is mostly bound to serum proteins, protein precipitation without complete dissociation of testosterone from binding-proteins and incomplete equilibrium of the IS with the binding proteins may result in incomplete or inconsistent analyte recovery [33]. Therefore, omitting protein precipitation seems to partially contribute to the very low measurement precision observed over 2 years. Despite multiple extraction steps, the extraction efficiency is 73% (95% CI: 70–76%). This is sufficient for quantitative measurements of testosterone at low concentrations such as those observed on women and children.
Measurement specificity has been another concern with testosterone measurements. To assess for potential interferences with known and unknown compounds we applied two approaches for interfering testing. In the first approach, the 28 structural analogs we tested showed no interfering peak in the chromatograms for the quantitation and confirmation ions (Table 3). The second approach was used to assess interferences from unknown compounds. The average QI/CI ratio calibrators was 0.98 (95% CI: 0.82–1.15) while the ratio in 56 serum samples was in average 1.12 (95% CI: 1.08–1.15). The highest QI/CI ratio observed in an individual sample was 1.24, which is well-below the suggested maximum differences of 20% difference [30,38,39]. These findings suggest that this method is either minimally affected by interfering substances or that it is not at all affected. Interferences from components in serum separator tubes (SST) have been reported [46] but were not found with our method using BD Vacutainer® SST™ (Franklin Lakes, NJ) probably because of the use of different chromatographic conditions.
Table 3.
List of steroid hormones used in the interference analysis.
| Steroids | Formula | Molecular weight | Supplier |
|---|---|---|---|
| Estrone | C18H22O2 | 270.4 | Cerilliant |
| Estradiol | C18H24O2 | 272.4 | Cerilliant |
| 5,16-Androstadien-3b-ol | C19H28O | 272.4 | Steraloids |
| 16, (5a)-Androsten-3-one | C19H28O | 272.4 | Steraloids |
| 4,16-Androstadien-3b-ol | C19H28O | 272.4 | Steraloids |
| 2,(5a)-Androsten-17-one | C19H28O | 272.4 | Steraloids |
| Estriol | C18H24O3 | 288.4 | Cerilliant |
| Dehydroepiandrasterone | C19H28O2 | 288.4 | Cerilliant |
| Epitestosterone | C19H28O2 | 288.4 | Sigma-Aldrich |
| Trans-dehydroandrosterone | C19H28O2 | 288.4 | Sigma-Aldrich |
| Trans-androsterone | C19H30O2 | 290.4 | Sigma-Aldrich |
| 5-Androstenediol | C19H30O2 | 290.4 | Steraloids |
| Androsterone | C19H30O2 | 290.4 | Cerilliant |
| Etiocholan-3a-ol-17-one | C19H30O2 | 290.4 | Sigma-Aldrich |
| Dihydrotesosterone | C19H30O2 | 290.4 | Cerilliant |
| 17a-Ethinylestradiol | C20H24O2 | 296.4 | Sigma-Aldrich |
| 19-Norethindrone | C20H26O2 | 298.4 | Sigma-Aldrich |
| D(−)-Norgestrel | C21H28O2 | 312.4 | Sigma-Aldrich |
| Progesterone | C21H30O2 | 314.5 | Cerilliant |
| 5-Pregnen-3β-ol-20-one | C21H32O2 | 316.5 | Steraloids |
| Pregnenolone | C21H32O2 | 316.5 | Sigma-Aldrich |
| 17α-Hydroxyprogesterone | C21H30O3 | 330.5 | Cerilliant |
| 17α-Hydroxypregnenolone | C21H32O3 | 332.5 | Steraloids |
| Corticosterone | C21H30O4 | 346.5 | Cerilliant |
| 11-Deoxycortisol | C21H30O4 | 346.5 | Cerilliant |
| Cortisone | C21H28O5 | 360.4 | Steraloids |
| Hydrocortisone | C21H30O5 | 362.5 | Steraloids |
| Dehydroepiandrosterone sulfate | C19H28O5S | 368.5 | Steraloids |
Some methods described in literature use derivatization procedures to enable analysis by GC-MS or to achieve higher sensitivity of LC-MS measurements [47–49]. Our method does not require derivatization, which minimizes complexity and facilitates automation. Furthermore, other methods employ ethers, such as methyl tertiary-butyl ether [24], diethyl ether [50], and diethyl ether/ethyl acetate mixtures [51] for the liquid–liquid extraction. However, these solvents are difficult to handle with automatic pipetting system and require special safety considerations. The solvents used in our method showed compatibility with automated liquid handlers and require fewer safety considerations than ethers [52–54].
The present method proofed a consistent result over a 2-year period at all concentration levels, as well as experimental variations with very minimal matrix effect and no or negligible difference for testosterone in neat and matrix. Measurement values obtained with this method are in high agreement with established reference methods [30] and meets performance criteria for accuracy used by the CDC Hormone Standardization Program [35]. This described method provides a consistently high accurate and sensitive LC-MS/MS assay for testosterone measurements in serum at all concentration levels typically observed in men as well as in women and children.
Footnotes
Disclaimer: The findings and conclusions in this report are those of the author(s) and do not necessarily represent the views of the Centers for Disease Control and Prevention.
References
- 1.Dobs AS. The role of accurate testosterone testing in the treatment and management of male hypogonadism. Steroids. 2008;73(13):1305–10. doi: 10.1016/j.steroids.2008.06.007. [DOI] [PubMed] [Google Scholar]
- 2.Goodman NF, Bledsoe MB, Cobin RH, et al. American Association of Clinical Endocrinologists medical guidelines for the clinical practice for the diagnosis and treatment of hyperandrogenic disorders. Endocr Pract. 2001;7(2):120–34. [PubMed] [Google Scholar]
- 3.Robinson S, Rodin DA, Deacon A, Wheeler MJ, Clayton RN. Which hormone tests for the diagnosis of polycystic-ovary-syndrome? Br J Obstet Gynaecol. 1992;99(3):232–8. doi: 10.1111/j.1471-0528.1992.tb14505.x. [DOI] [PubMed] [Google Scholar]
- 4.Araujo AB, Esche GR, Kupelian V, et al. Prevalence of symptomatic androgen deficiency in men. J Clin Endocrinol Metab. 2007;92(11):4241–7. doi: 10.1210/jc.2007-1245. [DOI] [PubMed] [Google Scholar]
- 5.Kazi M, Geraci SA, Koch CA. Considerations for the diagnosis and treatment of testosterone deficiency in elderly men. Am J Med. 2007;120(10):835–40. doi: 10.1016/j.amjmed.2007.02.022. [DOI] [PubMed] [Google Scholar]
- 6.Goodman NF, Cobin RH, Ginzburg SB, Katz IA, Woode DE American Association of Clinical E. American Association of Clinical Endocrinologists Medical Guidelines for Clinical Practice for the diagnosis and treatment of menopause: executive summary of recommendations. Endocr Pract. 2011;17(6):949–54. doi: 10.4158/ep.17.6.949. [DOI] [PubMed] [Google Scholar]
- 7.Petak SM, Nankin HR, Spark RF, Swerdloff RS, Rodriguez-Rigau LJ American Association of Clinical E. American Association of Clinical Endocrinologists Medical Guidelines for clinical practice for the evaluation and treatment of hypogonadism in adult male patients — 2002 update. Endocr Pract. 2002;8(6):440–56. [PubMed] [Google Scholar]
- 8.Albrecht L, Styne D. Laboratory testing of gonadal steroids in children. Pediatr Endocrinol Rev. 2007;5(Suppl 1):599–607. [PubMed] [Google Scholar]
- 9.Speiser PW. Interpretation of pediatric endocrine laboratory tests: pitfalls in steroid hormone measurements and genotyping. Pediatr Endocrinol Rev. 2007;5(Suppl 1):578–83. [PubMed] [Google Scholar]
- 10.Bhasin S, Zhang A, Coviello A, et al. The impact of assay quality and reference ranges on clinical decision making in the diagnosis of androgen disorders. Steroids. 2008;73(13):1311–7. doi: 10.1016/j.steroids.2008.07.003. [DOI] [PubMed] [Google Scholar]
- 11.Dorgan JF, Longcope C, Franz C, et al. Endogenous sex hormones and breast cancer in postmenopausal women: reanalysis of nine prospective studies. J Natl Cancer Inst. 2002;94(8):606–16. doi: 10.1093/jnci/94.8.606. [DOI] [PubMed] [Google Scholar]
- 12.Sofikerim M, Eskicorapci S, Oruc O, Ozen H. Hormonal predictors of prostate cancer. Urol Int. 2007;79(1):13–8. doi: 10.1159/000102906. [DOI] [PubMed] [Google Scholar]
- 13.Albers P, Albrecht W, Algaba F, et al. Guidelines on testicular cancer. Eur Urol. 2005;48(6):885–94. doi: 10.1016/j.eururo.2005.06.019. [DOI] [PubMed] [Google Scholar]
- 14.Albers P, Albrecht W, Algaba F, et al. EAU guidelines on testicular cancer: 2011 update. Eur Urol. 2011;60(2):304–19. doi: 10.1016/j.eururo.2011.05.038. [DOI] [PubMed] [Google Scholar]
- 15.Wang C, Jackson G, Jones TH, et al. Low testosterone associated with obesity and the metabolic syndrome contributes to sexual dysfunction and cardiovascular disease risk in men with type 2 diabetes. Diabetes Care. 2011;34(7):1669–75. doi: 10.2337/dc10-2339. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Selvin E, Feinleib M, Zhang L, et al. Androgens and diabetes in men: results from the Third National Health and Nutrition Examination Survey (NHANES III) Diabetes Care. 2007;30(2):234–8. doi: 10.2337/dc06-1579. [DOI] [PubMed] [Google Scholar]
- 17.Fink HA, Ewing SK, Ensrud KE, et al. Association of testosterone and estradiol deficiency with osteoporosis and rapid bone loss in older men. J Clin Endocrinol Metab. 2006;91(10):3908–15. doi: 10.1210/jc.2006-0173. [DOI] [PubMed] [Google Scholar]
- 18.Rosano GM, Cornoldi A, Fini M. Effects of androgens on the cardiovascular system. J Endocrinol Invest. 2005;28(3 Suppl):32–8. [PubMed] [Google Scholar]
- 19.Laughlin GA, Barrett-Connor E, Bergstrom J. Low serum testosterone and mortality in older men. J Clin Endocrinol Metab. 2008;93(1):68–75. doi: 10.1210/jc.2007-1792. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Shores MM, Matsumoto AM, Sloan KL, Kivlahan DR. Low serum testosterone and mortality in male veterans. Arch Intern Med. 2006;166(15):1660–5. doi: 10.1001/archinte.166.15.1660. [DOI] [PubMed] [Google Scholar]
- 21.Herold DA, Fitzgerald RL. Immunoassays for testosterone in women: better than a guess? Clin Chem. 2003;49(8):1250–1. doi: 10.1373/49.8.1250. [DOI] [PubMed] [Google Scholar]
- 22.Demers LM. Testosterone and estradiol assays: current and future trends. Steroids. 2008;73(13):1333–8. doi: 10.1016/j.steroids.2008.05.002. [DOI] [PubMed] [Google Scholar]
- 23.Hsing AW, Stanczyk FZ, Belanger A, et al. Reproducibility of serum sex steroid assays in men by RIA and mass spectrometry. Cancer Epidemiol Biomarkers Prev. 2007;16(5):1004–8. doi: 10.1158/1055-9965.EPI-06-0792. [DOI] [PubMed] [Google Scholar]
- 24.Moal V, Mathieu E, Reynier P, Malthiery Y, Gallois Y. Low serum testosterone assayed by liquid chromatography–tandem mass spectrometry. Comparison with five immunoassay techniques. Clin Chim Acta. 2007;386(1–2):12–9. doi: 10.1016/j.cca.2007.07.013. [DOI] [PubMed] [Google Scholar]
- 25.Vesper HW, Bhasin S, Wang C, et al. Interlaboratory comparison study of serum total testosterone [corrected] measurements performed by mass spectrometry methods. Steroids. 2009;74(6):498–503. doi: 10.1016/j.steroids.2009.01.004. [DOI] [PubMed] [Google Scholar]
- 26.Stanczyk FZ, Clarke NJ. Advantages and challenges of mass spectrometry assays for steroid hormones. J Steroid Biochem Mol Biol. 2010;121(3–5):491–5. doi: 10.1016/j.jsbmb.2010.05.001. [DOI] [PubMed] [Google Scholar]
- 27.Caudill SP, Schleicher RL, Pirkle JL. Multi-rule quality control for the age-related eye disease study. Stat Med. 2008;27(20):4094–106. doi: 10.1002/sim.3222. [DOI] [PubMed] [Google Scholar]
- 28.Taylor JK. Quality assurance of chemical measurements. CRC-Press; 1987. [Google Scholar]
- 29.CDC. 2013 http://www.cdc.gov/labstandards/hs.html. Page last updated: Thursday April 25 2013.
- 30.Botelho JC, Shacklady C, Cooper HC, et al. Isotope-dilution liquid chromatography–tandem mass spectrometry candidate reference method for total testosterone in human serum. Clin Chem. 2013;59(2):372–80. doi: 10.1373/clinchem.2012.190934. [DOI] [PubMed] [Google Scholar]
- 31.Tai SS, Xu B, Welch MJ, Phinney KW. Development and evaluation of a candidate reference measurement procedure for the determination of testosterone in human serum using isotope dilution liquid chromatography/tandem mass spectrometry. Anal Bioanal Chem. 2007;388(5–6):1087–94. doi: 10.1007/s00216-007-1355-3. [DOI] [PubMed] [Google Scholar]
- 32.Thienpont LM, Debrabandere VI, Stockl D, Deleenheer AP. Use of cyclodextrins for prepurification of progesterone and testosterone from human serum prior to determination with isotope-dilution gas-chromatography mass-spectrometry. Anal Chem. 1994;66(22):4116–9. doi: 10.1021/ac00094a041. [DOI] [PubMed] [Google Scholar]
- 33.Vesper HW, Botelho JC. Standardization of testosterone measurements in humans. J Steroid Biochem Mol Biol. 2010;121(3–5):513–9. doi: 10.1016/j.jsbmb.2010.03.032. [DOI] [PubMed] [Google Scholar]
- 34.NCCLS. NCCLS document EP9-A2. 2. 2002. Method comparison and bias estimation using patient samples; approved guideline. [Google Scholar]
- 35.Yun YM, Botelho JC, Chandler DW, et al. Performance criteria for testosterone measurements based on biological variation in adult males: recommendations from the partnership for the accurate testing of hormones. Clin Chem. 2012;58(12):1703–10. doi: 10.1373/clinchem.2012.186569. [DOI] [PubMed] [Google Scholar]
- 36.NCCLS. NCCLS document EP5-A2. 2. 2004. Evaluation of precision performance of quantitative measurement methods; approved guideline. [Google Scholar]
- 37.Matuszewski BK, Constanzer ML, Chavez-Eng CM. Strategies for the assessment of matrix effect in quantitative bioanalytical methods based on HPLC-MS/MS. Anal Chem. 2003;75(13):3019–30. doi: 10.1021/ac020361s. [DOI] [PubMed] [Google Scholar]
- 38.Li LY, Campbell DA, Bennett PK, Henion J. Acceptance criteria for ultratrace HPLC–tandem mass spectrometry: quantitative and qualitative determination of sulfonylurea herbicides in soil. Anal Chem. 1996;68(19):3397–404. doi: 10.1021/ac960375w. [DOI] [PubMed] [Google Scholar]
- 39.Rodriguez M, Orescan DB. Confirmation and quantitation of selected sulfonylurea, imidazolinone, and sulfonamide herbicides in surface water using electrospray LC/MS. Anal Chem. 1998;70(13):2710–7. doi: 10.1021/ac971128a. [DOI] [PubMed] [Google Scholar]
- 40.Rosner W, Auchus RJ, Azziz R, Sluss PM, Raff H. Position statement: utility, limitations, and pitfalls in measuring testosterone: an Endocrine Society position statement. J Clin Endocrinol Metab. 2007;92(2):405–13. doi: 10.1210/jc.2006-1864. [DOI] [PubMed] [Google Scholar]
- 41.Rosner W, Vesper H, Org ESE. Toward excellence in testosterone testing: a consensus statement. J Clin Endocrinol Metab. 2010;95(10):4542–8. doi: 10.1210/jc.2010-1314. [DOI] [PubMed] [Google Scholar]
- 42.Thienpont LM, Van Uytfanghe K, Blincko S, et al. State-of-the-art of serum testosterone measurement by isotope dilution–liquid chromatography–tandem mass spectrometry. Clin Chem. 2008;54(8):1290–7. doi: 10.1373/clinchem.2008.105841. [DOI] [PubMed] [Google Scholar]
- 43.Neville D, Houghton R, Garrett S. Efficacy of plasma phospholipid removal during sample preparation and subsequent retention under typical UHPLC conditions. Bioanalysis. 2012;4(7):795–807. doi: 10.4155/bio.12.38. [DOI] [PubMed] [Google Scholar]
- 44.Little JL, Wempe MF, Buchanan CM. Liquid chromatography–mass spectrometry/mass spectrometry method development for drug metabolism studies: examining lipid matrix ionization effects in plasma. J Chromatogr B Anal Technol Biomed Life Sci. 2006;833(2):219–30. doi: 10.1016/j.jchromb.2006.02.011. [DOI] [PubMed] [Google Scholar]
- 45.Savolainen K, Kiimamaa R, Halonen T. High-throughput analysis of testosterone in serum samples by on-line solid phase extraction liquid chromatography–tandem mass spectrometry. Clin Chem Lab Med. 2011;49(11):1845–8. doi: 10.1515/CCLM.2011.663. [DOI] [PubMed] [Google Scholar]
- 46.Shi RZ, van Rossum HH, Bowen RA. Serum testosterone quantitation by liquid chromatography–tandem mass spectrometry: interference from blood collection tubes. Clin Biochem. 2012;45(18):1706–9. doi: 10.1016/j.clinbiochem.2012.08.008. [DOI] [PubMed] [Google Scholar]
- 47.Star-Weinstock M, Williamson BL, Dey S, Pillai S, Purkayastha S. LC-ESI-MS/MS analysis of testosterone at sub-picogram levels using a novel derivatization reagent. Anal Chem. 2012;84(21):9310–7. doi: 10.1021/ac302036r. [DOI] [PubMed] [Google Scholar]
- 48.Bui HN, Struys EA, Martens F, et al. Serum testosterone levels measured by isotope dilution–liquid chromatography–tandem mass spectrometry in postmenopausal women versus those in women who underwent bilateral oophorectomy. Ann Clin Biochem. 2010;47:248–52. doi: 10.1258/acb.2010.009171. [DOI] [PubMed] [Google Scholar]
- 49.Keski-Rahkonen P, Huhtinen K, Poutanen M, Auriola S. Fast and sensitive liquid chromatography–mass spectrometry assay for seven androgenic and progestagenic steroids in human serum. J Steroid Biochem Mol Biol. 2011;127(3–5):396–404. doi: 10.1016/j.jsbmb.2011.06.006. [DOI] [PubMed] [Google Scholar]
- 50.Cawood ML, Field HP, Ford CG, et al. Testosterone measurement by isotope-dilution liquid chromatography–tandem mass spectrometry: validation of a method for routine clinical practice. Clin Chem. 2005;51(8):1472–9. doi: 10.1373/clinchem.2004.044503. [DOI] [PubMed] [Google Scholar]
- 51.Harwood DT, Handelsman DJ. Development and validation of a sensitive liquid chromatography–tandem mass spectrometry assay to simultaneously measure androgens and estrogens in serum without derivatization. Clin Chim Acta. 2009;409(1–2):78–84. doi: 10.1016/j.cca.2009.09.003. [DOI] [PubMed] [Google Scholar]
- 52.Shiraishi S, Lee PW, Leung A, Goh VH, Swerdloff RS, Wang C. Simultaneous measurement of serum testosterone and dihydrotestosterone by liquid chromatography–tandem mass spectrometry. Clin Chem. 2008;54(11):1855–63. doi: 10.1373/clinchem.2008.103846. [DOI] [PubMed] [Google Scholar]
- 53.Wang C, Shiraishi S, Leung A, et al. Validation of a testosterone and dihydrotestosterone liquid chromatography tandem mass spectrometry assay: interference and comparison with established methods. Steroids. 2008;73(13):1345–52. doi: 10.1016/j.steroids.2008.05.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54.French D. Development and validation of a serum total testosterone liquid chromatography–tandem mass spectrometry (LC-MS/MS) assay calibrated to NIST SRM 971. Clin Chim Acta. 2013;415:109–17. doi: 10.1016/j.cca.2012.10.007. [DOI] [PubMed] [Google Scholar]