Summary
Serial block‐face scanning electron microscopy (SBEM) is becoming increasingly popular for a wide range of applications in many disciplines from biology to material sciences. This review focuses on applications for circuit reconstruction in neuroscience, which is one of the major driving forces advancing SBEM. Neuronal circuit reconstruction poses exceptional challenges to volume EM in terms of resolution, field of view, acquisition time and sample preparation. Mapping the connections between neurons in the brain is crucial for understanding information flow and information processing in the brain. However, information on the connectivity between hundreds or even thousands of neurons densely packed in neuronal microcircuits is still largely missing. Volume EM techniques such as serial section TEM, automated tape‐collecting ultramicrotome, focused ion‐beam scanning electron microscopy and SBEM (microtome serial block‐face scanning electron microscopy) are the techniques that provide sufficient resolution to resolve ultrastructural details such as synapses and provides sufficient field of view for dense reconstruction of neuronal circuits. While volume EM techniques are advancing, they are generating large data sets on the terabyte scale that require new image processing workflows and analysis tools. In this review, we present the recent advances in SBEM for circuit reconstruction in neuroscience and an overview of existing image processing and analysis pipelines.
Keywords: Microtome, registration, SBFSEM, serial block‐face scanning electron microscopy (SBEM), segmentation, three‐dimensional reconstruction, tiling
Lay Description
During the last decades, numerous technologies have been developed in order to visualize the brain at high resolution with Volume Electron Microscopy. The visualization in three dimensions and the analysis of an entire brain from an animal model where all cells are visible and can be followed and where all synapses can be examined and counted is considering as a milestone by numerous neuroscientists. One of the technique that could be used to acquire images of an entire brain at nanometre scale is the microtome‐based block‐face scanning electron microscopy developed by W. Denk and his collaborators. In this approach, a microtome integrated within the specimen chamber of a scanning electron microscope cuts away nanometres thin slice of the specimen, thus allowing the imaging of the next ‘layer’. These individual 2D images are then assembled to a 3D image of all neurons and their projections. The recent improvements of the technique due to better microscopes and better sample preparation are pushing the limits of data quality and acquisition speed. It is creating data sets requiring important storage space. At the level of analysis, new algorithms are developed to reconstruct manually or with semiautomated interfaces all the neurons present in a volume.
Volume EM techniques
Understanding the information flow in neuronal circuits of the brain requires a detailed map on the connectivity between neurons. In order to map all the connections of even a small circuit, detailed recognition of anatomical structures such as spine necks and synapses is necessary. It requires 3D ultrastructural resolution on the nanometre scale (<20 nm pixel−1; Helmstaedter, 2013). In addition, neurons are densely packed and a cubic millimetre, the typical size of a voxel in magnetic resonance imaging, contains millions of neurons and thousands of metres neurite length (Behrens et al., 2007). Therefore, microscopy techniques that rely on sparse labelling of neuronal features, such as fluorescence microscopy, are not suited for dense circuit reconstruction. The commonly used techniques for dense circuit reconstructions are volume EM techniques such as serial section TEM (ssTEM), automated tape‐collecting ultramicrotome SEM (ATUM‐SEM), serial block‐face SEM (SBEM) and focused ion beam SEM (FIB‐SEM). The former two use nondestructive manual (ssTEM) or automated (ATUM) ultrathin sectioning and slice collection with subsequent imaging. The latter two use in situ destructive on‐block sectioning inside of the SEM vacuum chamber either by a diamond knife (SBEM) or with a focused ion beam (FIB‐SEM).
In ssTEM, sections are cut by hand using an ultramicrotome and collected onto grids. Sections are imaged in a TEM, and imaging approaches such as camera array imaging (TEMCA) with typically 2–4 nm lateral resolution (Bock et al., 2011; Takemura et al., 2013; Atasoy et al., 2014) can acquire images at a rate of 5–8 megapixels s−1 (Bock et al., 2011). However, the manual sectioning process is prone to errors and is also labour‐intensive. It limits the typical section thickness to 40–50 nm and the number of consecutive sections to a few thousands, because sections can get folded or warped or even get lost. Furthermore, the subsequent registration step is more delicate due to distortions of sections occurring during cutting and imaging (Kaynig, Fischer, et al., 2010; Saalfeld et al., 2010).
ATUM‐SEM (Schalek et al., 2011) overcomes these problems partially by automated cutting and collection of sections on an electrically opaque tape. This allows to reliable cut thousands of subsequent sections as thin as 30 nm. The on‐tape sections are further processed for storage on silicon wafers and subsequent imaging in an SEM (Hayworth et al., 2014).
In SBEM [formerly called SBFSEM but renamed by W. Denk (Denk & Horstmann, 2004)], the recording chamber of an SEM is equipped with a microtome. After each cut, the block‐face is imaged with the scanning beam before the next section is shaved off. SBEM achieves field of views of >0.5 × 0.5 mm2 at a lateral resolution on the order of 6–10 nm and reliable cuts thousands of sections at section thickness 20–30 nm for neural tissue at an acquisition rate of 0.5 to 2 megapixel s−1 (Briggman et al., 2011; Helmstaedter, 2013). In FIBSEM, slices are cut using a gallium‐ion beam (Knott et al., 2008). This allows to cut sections as thin as 5 nm with a lateral resolution <5 nm and an acquisition rate of 0.1–0.5 megapixel s−1 (Knott et al., 2008; Boergens & Denk, 2013; Maco et al., 2014). However, the field of view is limited to <0.1 × 0.1 mm2.
The major advantage of the destructive on‐block methods is that they do not suffer from warping problems and section loss that can significantly affect the data quality and the subsequent data analysis. In contrast, the advantages of the nondestructive methods are that sections can be imaged multiple times (e.g. at different magnifications) and in parallel on multiple microscopes, which reduce the acquisition time significantly compared to sequential techniques such as SBEM and FIBSEM.
Currently, these techniques are to a large degree complementary. They cover different range of resolution, field of view and automation (Lichtman & Denk, 2011).
To date, only a handful of neuronal circuits have been completely reconstructed using volume EM. The entire nervous system of the nematode Caenorhabditis elegans has been acquired using serial sections and TEM, which led to the first complete connectome of an entire organism (White et al., 1986; Varshney et al., 2011). It includes all 302 neurons connected by more than 7000 synapses. Partial connectomes have been obtained from a mouse retina (Briggman et al., 2011; Helmstaedter et al., 2013) by SBEM and from mouse primary visual cortex (Bock et al., 2011) and Drosophila visual motion detection circuit (Takemura et al., 2013) by ssTEM. The retina datasets are available on http://www.knossostool.org/ and Bock et al.'s images are publicly available through the Open Connectome Project (http://www.openconnectomeproject.org/).
Development of SBEM for connectomics
SBEM was first described in 2004 (Denk & Horstmann, 2004) based on an idea introduced by Leighton (1981). The group of W. Denk inserted a completely redesigned ultramicrotome inside of an environmental scanning electron microscope with a field emission gun. In this design, SBEM allows for automated, iterative removal of thin sections from the sample block and subsequent block‐face imaging at ultrastructural resolution. The ultramicrotome comes with a movable high‐precision stage that allows dividing the field of view into a mosaic of tiled images. This allows, in principle, to acquire field of views of arbitrary size, although the edge size of a single image in an SBEM is currently limited to about 40 μm (3VBSED, Gatan Inc.).
One cubic millimetre is often described as the minimal target volume for neuronal circuit reconstruction, because it corresponds to a volume large enough to analyze a functional unit such as a cortical column in the adult mouse brain. For the dense reconstruction of all neurites and the unambiguous identification of chemical synapses in this volume, a voxel size of about 10 × 10 × 30 nm is necessary. This volume corresponds to a data set of 3.333 × 1014 voxels requiring >300 TB storage space. At a typical acquisition speed of 1–4 microseconds pixel−1, the cubic millimetre requires more than 18 years of acquisition time making this kind of project impossible. Thereby, the stage movements for the tiling alone causes an overhead of 4 years.
In order to move this kind of projects within the realm of the feasible, both, the instrumentation and the sample preparation, have been improved and optimized in the last couple of years.
At the level of instrumentation, progress has been made that now allows now to scan larger field of views with higher scan speed and lower beam voltage. The neuronal tissue used for volume EM is usually embedded in nonconductive resin. If nonconductive regions are scanned in an SBEM, electrons accumulate on the block surface and cause charging artefacts. Therefore, environmental SEMs have been used for SBEM, where a gaseous agent such as water or nitrogen (Danilatos, 1988; Danilatos, 2009; Danilatos, 2012) is introduced into the recording chamber and takes up exceeding electrons from the block‐face. This reduces any charging artefacts significantly (Denk & Horstmann, 2004), but requires a lower scanning speed and/or a higher beam voltage in order to compensate for the loss in signal due to electron scattering at the gaseous agent. Because the beam current (number of electrons) of these environmental SEMs is limited to approximately 100 pA, the scan speed/dwell time typically had to be at least around 1–5 microseconds pixel−1 in order to collect enough electrons for a sufficient signal‐to‐noise ratio (Briggman et al., 2011). Similarly, for a decent image quality, the beam voltage (speed of the beam electrons) on these instruments had to be larger than 2 kV. However, the higher the beam voltage, the deeper the electrons penetrate into the block‐face, which decreases the Z resolution and increases the minimal section thickness that can be cut (Denk & Horstmann, 2004). In order to reliably cut 25–30‐nm thin sections, a beam voltage <2 kV is necessary. Therefore, new SEMs are optimized for low voltage (1.5–2 kV) and higher beam currents (>0.5 nA) that allow faster scanning and shorter dwell times (<1 microsecond pixel−1; Titze & Denk, 2013).
Recently, new approaches have been proposed to overcome the block‐face charging problem even in high‐vacuum operation mode of an SEM. A coating device can be introduced into the SEM chamber sputtering a thin conductive layer after each cut before imaging which makes the block‐face perfectly conductive (Titze & Denk, 2013). As introduced above, the stage movements for the acquisition of a tiled field view can cause a significant overhead of more than 20% of the total acquisition time. Therefore, the groups of Winfried Denk (MPIN Munich), Moritz Helmstaedter (MPI Frankfurt) and Kevin Briggman (NIH, Circuit Dynamics and Connectivity Unit, Bethesda) developed new microtomes that allow to continuously move the sample stage along one axis while scanning, which minimizes the number of stage movements for tiled acquisition and results in a significantly reduction of acquisition time (Perkel, 2014; http://www.pbs.org/newshour/updates/watch‐ideas‐light‐fishs‐brain/).
Another promising approach to reduce the number of image tiles is the new multiSEM 505, launched by ZEISS (Marx, 2013). It allows scanning sample surfaces or sections with 61 parallel electron beams simultaneously, which increases the field of view tremendously. Combined with ATUM‐SEM or with block‐face approaches, it will allow to image at a rate of 1220 megapixels s−1 (Eberle et al., 2015).
Currently, Winfried Denk is pushing the extremes of SBEM with the goal of cutting and imaging an entire mouse brain (Mikula et al., 2012; Perkel, 2014). Besides the development of a completely new microtome capable of cutting centimetre‐sized blocks, they are also working on improving the stability of image acquisition by developing methods for automated correction of focus and stigmatism (Binding et al., 2013).
At the level of the sample preparation, several groups have put efforts in generating samples able to sustain the accelerating voltage and beam current necessary for high‐speed acquisition while keeping signal‐to‐noise ratio sufficient for subsequent analysis, such as image segmentation. The staining protocols have been adapted and optimized for SBEM (Deerinck et al., 2010; Mikula et al., 2012; Tapia et al., 2012; Starborg et al., 2013) focusing on the en bloc staining with diverse heavy metals: OTO (osmium‐thiocarbohydrazide‐osmium, rOTO (reduced Osmium), OTrO followed by en bloc uranyl acetate and/or lead citrate/asparte) in order to increase the contrast and the conductivity of the sample. The latter also reduces charging issues during acquisition. However, samples stained by multiple steps of heavy metals show reduced permeability and resin may not be able to fully penetrate (Mikula et al., 2012; Tapia et al., 2012; Starborg et al., 2013). In addition, the density of heavy metals in the sample block can compromise the cutting properties such as slice thickness and may reduce the imaging quality.
The advances on both levels, sample preparation and instrumentation, set new conditions of imaging: With higher beam currents (>1 nA) scan speed can reach 10 MHz (100 nanoseconds pixel−1) without charging artefact, while keeping the signal‐to‐noise ratio good enough for image segmentation and neurite reconstruction (Titze & Denk, 2013). At a speed of 10 MHz (0.1 microsecond pixel−1) and with continuous scanning along one axis, the millimetre cube could be acquired in about 1.5 year. This is still a significant challenge as these instruments currently cannot be run for more than 1–2 months (Briggman et al., 2011; Lichtman & Denk, 2011; Helmstaedter et al., 2013) without failure. However, this calculation is based on a single‐beam SEM. With the launch of the multi‐SEM 505, the acquisition can be parallelized on 61 beams, leading to a speed of 1220 megapixels s−1 (vs. 1 megapixel s−1 for a single‐beam SEM at 1 microsecond pixel−1 equivalent to 1 MHz). With this, the cubic millimetre could be acquired in about 2 months.
Another imaging parameter that influences the effective acquisition time significantly is the size of the acquired field of view. It is therefore advisable to restrict the acquisition to the desired region of interest. However, finding a particular region of interest in an en bloc stained sample is challenging. Possible solutions are targeted near‐infrared branding of a region of interest (Bishop et al., 2011; Maco et al., 2013; Maco et al., 2014) and staining specific structures, cells or molecules using genetic labels that accumulate electrodense molecules. (Martell et al., 2012).
Data preprocessing: alignment in 3D
To annotate/reconstruct the imaged sample in silico, image preprocessing is necessary. In particular, images need to be placed correctly in three dimensions to combine the slice data into a volume representing the specimen. Image tiles acquired from the same block‐face need to be stitched together in x and y. The overlapping region between neighbouring tiles is scanned multiple times, which increases acquisition time and in addition can compromise the cutting quality and cutting thickness due to multiple beam exposure. It is therefore advisable to minimize both, the number of image tiles as well as the amount of overlap between neighbouring image tiles. A compromise in the amount of overlap has to be found because the overlaps need to be large enough to allow for correct image registration and stitching. Because temperature changes during acquisition can cause drifts of the stage relative to the beam source on the order of several nanometres (Boergens & Denk, 2013), one cannot rely on stage coordinates alone. In addition, the actual image positions can also drift due to local changes in the electric field due to charging artefacts. In the case of SBEM, these drifts are mostly limited to translations. The necessary corrections do not need to take into account rotations or nonlinear transforms. To stitch the overlapping images pixel‐perfect, the amount of overlap between adjacent images can be calculated via cross‐correlations or by detecting and comparing matching configurations of image features, as with the scale‐invariant feature transform algorithm (Lowe, 2004). Then, the images are stitched to the resulting mosaic by overlapping the tiles and minimizing the global error at the overlaps (Fig. 1). Subsequently, the stitched mosaics need to be aligned in the z‐direction. Here, image feature‐based algorithms suppress overfitting of larger objects which are oriented nonperpendicular to the cutting plane, an artefact typical for cross‐correlation‐based alignments. Since EM images of most biological samples contain many objects at smaller scales than the diagonally cut object, their resulting image features do not have a preferential direction and outnumber the features stemming from the larger object.
Figure 1.

(A) Resin‐embedded specimen (pyramid) is sectioned from top to bottom, while each block‐face is imaged as a mosaic of overlapping tiles (B) For each imaged block‐face, the overlapping tiles taken from the same area are compared in order to determine the correct overlap (C) The mosaics of adjacent planes are compared based on image features (as in B) and shifted to form a continuous representation of the specimen (D) Tiles are stitched into a mosaic optimizing the global overlap for each section.
The workflow of stitching and aligning needs to be automated to cope with many thousands of images per data set to reduce manual labour. TrakEM2, a free open‐source software specifically designed for reconstruction of neural circuits from terabyte EM data sets is a handy tool for this work flow (Cardona et al., 2012). It uses pyramidal data organization to minimize the RAM consumption and necessary data throughput rates of mass storage.
Image analysis, annotation and segmentation
Browsing, analyzing and annotating the massive image data sets generated by SBEM is challenging. The retina stack of Briggman et al. (2011) and Helmstaedter et al. (2013) already required several hundreds of GB storage space, but as the acquisition technology of SBEM progresses, the data set sizes also increase. The acquisition of a cubic millimetre cortex at a voxel size of 10 × 10 × 30 nm would require more than 300 TB storage space after image preprocessing. Therefore, the vast size of the image data makes it impossible to load the full data set for analysis into the RAM of a normal lab computer. Therefore, several labs have developed open‐source software solutions dedicated for large‐scale 3D image data such as KNOSSOS (Helmstaedter et al., 2013), TrakEM2 (Cardona et al., 2012) and CATMAID (Saalfeld et al., 2009). These programs use demand‐driven dynamic data loading procedures, in which only the currently viewed subvolume is loaded into memory. As the user browses through the data, the corresponding subvolumes are continuously loaded in the background and therefore allow seamless navigation with minimal memory requirements. As the data sets are getting too big for local storage, both, CATMAID and KNOSSOS, feature online streaming of the data from external servers. In addition, these programs feature a wide range of manual annotation tools, including feature labelling (e.g. mitochondria, synapses, etc.), skeleton tracing of neurites and volume segmentation (Figure 2).
Figure 2.

Consolidated skeletons of six mitral cells (red) and eight interneurons (blue) projecting into the same protoglomerulus of the olfactory bulb of a larval zebra fish. The consolidated skeletons were calculated from redundant manual reconstructions of three different tracers. The SBEM data set was acquired in a period of 7 weeks and consists of more than 5000 sections of 25 nm thickness with a lateral pixel size of 10 nm. Scale bar: 10 μm.
However, manual annotation and segmentation of large image data sets is tedious, error‐prone and can be very time‐consuming. For example, the dense skeleton reconstruction and analysis of 950 neurons in the inner plexiform layer of a mouse retina required almost 30,000 human working hours (Perkel, 2014), despite 50‐fold speed‐up for manual skeletonization versus manual volume annotation (Helmstaedter et al., 2013). Therefore, new software is under development for computer‐assisted, semiautomated large‐scale annotation and segmentation (Lowe, 2004; Jain et al., 2007; Chklovskii et al., 2010; Kaynig, Fuchs, et al., 2010; Kim et al., 2014; Maco et al., 2014).
Although the currently existing automated segmentation algorithms are still far from perfect, they have been successfully combined with manual annotation or proof‐reading by humans. CATMAID is used for collaborative annotation efforts distributed over collaborating research groups (Saalfeld et al., 2009). The SBEM pioneers in the Denk lab recruited hundreds of undergraduates for skeleton tracing of neurons using KNOSSOS (Helmstaedter et al., 2013). Thereby, most neurons have been redundantly traced by multiple students in order to form a consensus and in turn reduce reconstruction error rates. Subsequently, these consensus‐skeletons have been used for automated volume segmentation of the corresponding neurons. EyeWire, a crowd sourcing online platform, has successfully used the judgment of thousands of laymen volunteers for the proof‐reading of automatically segmented neurons (Kim et al., 2014). Others have employed a small number (<10) of professional proof‐readers (Takemura et al., 2013).
Conclusion
Connectomics, the complete reconstruction of neuronal circuits in different animal models, is a major driving force for new developments in SBEM. New sample preparation protocols, microscopes, acquisition modes and microtomes enable faster acquisition of larger field of views. This in turn creates an unprecedented flood of EM data that triggers the development of new image processing and analysis workflows and software solutions. These technological advances will be and are highly beneficial for many other fields of research.
The copyright line for this article was changed on June 17, 2015 after original online publication.
References
- Atasoy, D. , Betley, J.N. , Li, W.P. , et al (2014) A genetically specified connectomics approach applied to long‐range feeding regulatory circuits. Nat. Neurosci. 17, 1830–1839. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Behrens, T.E. , Berg, H.J. , Jbabdi, S. , Rushworth, M.F. & Woolrich, M.W. (2007) Probabilistic diffusion tractography with multiple fibre orientations: what can we gain? NeuroImage. 34, 144–155. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Binding, J. , Mikula, S. & Denk, W. (2013) Low‐dosage maximum‐a‐posteriori focusing and stigmation. Microsc. Microanal., Microbeam Anal. Soc. Microsc. Soc. Canada. 19, 38–55. [DOI] [PubMed] [Google Scholar]
- Bishop, D. , Nikic, I. , Brinkoetter, M. , Knecht, S. , Potz, S. , Kerschensteiner, M. & Misgeld, T. (2011) Near‐infrared branding efficiently correlates light and electron microscopy. Nat. Methods. 8, 568–570. [DOI] [PubMed] [Google Scholar]
- Bock, D.D. , Lee, W.C. , Kerlin, A.M. , et al (2011) Network anatomy and in vivo physiology of visual cortical neurons. Nature. 471, 177–182. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Boergens, K.M. & Denk, W. (2013) Controlling FIB‐SBEM slice thickness by monitoring the transmitted ion beam. J Microsc. 252, 258–262. [DOI] [PubMed] [Google Scholar]
- Briggman, K.L. , Helmstaedter, M. & Denk, W. (2011) Wiring specificity in the direction‐selectivity circuit of the retina. Nature. 471, 183–188. [DOI] [PubMed] [Google Scholar]
- Cardona, A. , Saalfeld, S. , Schindelin, J. , et al (2012) TrakEM2 software for neural circuit reconstruction. PloS One. 7, e38011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Chklovskii, D.B. , Vitaladevuni, S. & Scheffer, L.K. (2010) Semi‐automated reconstruction of neural circuits using electron microscopy. Curr. Opin. Neurobiol. 20, 667–675. [DOI] [PubMed] [Google Scholar]
- Danilatos, G.D. (1988) Foundations of Environmental Scanning Electron Microscopy. Advances in Electronics and Electron Physics, Academic Press, 71, 109–250. [Google Scholar]
- Danilatos, G.D. (2009) Optimum beam transfer in the environmental scanning electron microscope. J. Microsc. 234, 26–37. [DOI] [PubMed] [Google Scholar]
- Danilatos, G.D. (2012) Backscattered electron detection in environmental SEM. J. Microsc. 245, 171–185. [DOI] [PubMed] [Google Scholar]
- Deerinck, T.J. , Bushong, E. , Thor, A. & Ellisman, M.H. (2010) NCMIR methods for 3D EM: a new protocol for preparation of biological specimens for serial block‐face SEM. Microscopy. 6–8. [Google Scholar]
- Denk, W. & Horstmann, H. (2004) Serial block‐face scanning electron microscopy to reconstruct three‐dimensional tissue nanostructure. PLoS Biol. 2, e329. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Eberle, A.L. , Mikula, S. , Schalek, R. , Lichtman, J.W. , Tate, M.L. , & Zeidler, D. (2015) High‐resolution, high‐throughput imaging with a multibeam scanning electron microscope. J Microsc. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hayworth, K.J. , Morgan, J.L. , Schalek, R. , Berger, D.R. , Hildebrand, D.G. & Lichtman, J.W. (2014) Imaging ATUM ultrathin section libraries with WaferMapper: a multi‐scale approach to EM reconstruction of neural circuits. Front Neural Circ. 68, 1–18. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Helmstaedter, M. (2013) Cellular‐resolution connectomics: challenges of dense neural circuit reconstruction. Nat. Methods. 10, 501–507. [DOI] [PubMed] [Google Scholar]
- Helmstaedter, M. , Briggman, K.L. , Turaga, S.C. , Jain, V. , Seung, H.S. & Denk, W. (2013) Connectomic reconstruction of the inner plexiform layer in the mouse retina. Nature. 500, 168–174. [DOI] [PubMed] [Google Scholar]
- Jain, V. , Murray, J.F. , Roth, F. , et al (2007) Supervised learning of image restoration with convolutional networks. In Proceedings of the IEEE 11th International Conference on Computer Vision(ICCV 2007) , October 14–20. pp. 1–8, Rio de Janeiro, Brazil.
- Kaynig, V. , Fischer, B. , Muller, E. & Buhmann, J.M. (2010) Fully automatic stitching and distortion correction of transmission electron microscope images. J. Struct. Biol. 171, 163–173. [DOI] [PubMed] [Google Scholar]
- Kaynig, V. , Fuchs, T.J. & Buhmann, J.M. (2010) Geometrical consistent 3D tracing of neuronal processes in ssTEM data. In Proceedings of the 13th International Conference on Medical Image Computing and Computer‐Assisted Intervention (MICCAI), September 20–24. 13, 209–216. Beijing, China. [DOI] [PubMed] [Google Scholar]
- Kim, J.S. , Greene, M.J. , Zlateski, A. , et al & EyeWirers. (2014) Space‐time wiring specificity supports direction selectivity in the retina. Nature. 509, 331–336. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Knott, G. , Marchman, H. , Wall, D. & Lich, B. (2008) Serial section scanning electron microscopy of adult brain tissue using focused ion beam milling. J. Neurosci. 28, 2959–2964. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Leighton, S.B. (1981) SEM images of block faces, cut by a miniature microtome within the SEM – a technical note. Scan. Electron. Micros. 73–76. [PubMed] [Google Scholar]
- Lichtman, J.W. & Denk, W. (2011) The big and the small: challenges of imaging the brain's circuits. Science. 334, 618–623. [DOI] [PubMed] [Google Scholar]
- Lowe, D.G. (2004) Distinctive image features from scale‐invariant keypoints. Int. J. Comput. Vision. 60, 91–110. [Google Scholar]
- Maco, B. , Holtmaat, A. , Cantoni, M. , Kreshuk, A. , Straehle, C.N. , Hamprecht, F.A. & Knott, G.W. (2013) Correlative in vivo 2 photon and focused ion beam scanning electron microscopy of cortical neurons. PloS One. 8, e57405. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Maco, B. , Cantoni, M. , Holtmaat, A. , Kreshuk, A. , Hamprecht, F.A. & Knott, G.W. (2014) Semiautomated correlative 3D electron microscopy of in vivo‐imaged axons and dendrites. Nat. Protocols. 9, 1354–1366. [DOI] [PubMed] [Google Scholar]
- Martell, J.D. , Deerinck, T.J. , Sancak, Y. , Poulos, T.L. , Mootha, V.K. , Sosinsky, G.E. , Ellisman, M.H. & Ting, A.Y. (2012) Engineered ascorbate peroxidase as a genetically encoded reporter for electron microscopy. Nat. Biotechnol. 30, 1143–1148. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Marx, V. (2013) Neurobiology: brain mapping in high resolution. Nature. 503, 147–152. [DOI] [PubMed] [Google Scholar]
- Mikula, S. , Binding, J. & Denk, W. (2012) Staining and embedding the whole mouse brain for electron microscopy. Nat. Methods. 9, 1198–1201. [DOI] [PubMed] [Google Scholar]
- Perkel, J.M. (2014) Mapping Neural Connections. Biotechniques 57, 230–236. [DOI] [PubMed] [Google Scholar]
- Saalfeld, S. , Cardona, A. , Hartenstein, V. & Tomancak, P. (2009) CATMAID: collaborative annotation toolkit for massive amounts of image data. Bioinformatics. 25, 1984–1986. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Saalfeld, S. , Cardona, A. , Hartenstein, V. & Tomancak, P. (2010) As‐rigid‐as‐possible mosaicking and serial section registration of large ssTEM datasets. Bioinformatics. 26, i57–i63. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Schalek, R. , Kasthuri, N. , Hayworth, K. , et al (2011) Development of high‐throughput, high‐resolution 3D reconstruction of large‐volume biological tissue using automated tape collection ultramicrotomy and scanning electron microscopy. Microscop. Microanal 17S2, 966–967. [Google Scholar]
- Starborg, T. , Kalson, N.S. , Lu, Y. , Mironov, A. , Cootes, T.F. , Holmes, D.F. & Kadler, K.E. (2013) Using transmission electron microscopy and 3View to determine collagen fibril size and three‐dimensional organization. Nat. Protocols. 8, 1433–1448. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Takemura, S.Y. , Bharioke, A. , Lu, Z. , et al (2013) A visual motion detection circuit suggested by Drosophila connectomics. Nature. 500, 175–181. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Tapia, J.C. , Kasthuri, N. , Hayworth, K.J. , Schalek, R. , Lichtman, J.W. , Smith, S.J. & Buchanan, J. (2012) High‐contrast en bloc staining of neuronal tissue for field emission scanning electron microscopy. Nat. Protocols. 7, 193–206. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Titze, B. & Denk, W. (2013) Automated in‐chamber specimen coating for serial block‐face electron microscopy. J. Microsc. 250, 101–110. [DOI] [PubMed] [Google Scholar]
- Varshney, L.R. , Chen, B.L. , Paniagua, E. , Hall, D.H. & Chklovskii, D.B. (2011) Structural properties of the Caenorhabditis elegans neuronal network. PLoS Comput. Biol. 7, e1001066. [DOI] [PMC free article] [PubMed] [Google Scholar]
- White, J.G. , Southgate, E. , Thomson, J.N. & Brenner, S. (1986) The structure of the nervous system of the nematode Caenorhabditis elegans. Philos. Trans. R. Soc. Lond. B Biol. Sci. 314, 1–340. [DOI] [PubMed] [Google Scholar]