Skip to main content
Thorax logoLink to Thorax
. 1995 Nov;50(11):1199–1201. doi: 10.1136/thx.50.11.1199

Accuracy of an intelligent CPAP machine with in-built diagnostic abilities in detecting apnoeas: a comparison with polysomnography.

M Gugger 1, J Mathis 1, C Bassetti 1
PMCID: PMC475094  PMID: 8553278

Abstract

BACKGROUND--In patients with sleep apnoea early diagnostic evaluation and treatment may be delayed due to limited access to full polysomnography (PSG). For "typical" patients, simplified strategies are needed. A study was performed to evaluate the accuracy of a new continuous positive airway pressure (CPAP) device with in-built diagnostic abilities (Autoset) in detecting apnoeas. METHODS--Twenty seven patients underwent full overnight polysomnography. Data with the Autoset were acquired simultaneously. Standard nasal prongs were used. Apnoeas were detected by special analysis of the flow signal. As the Autoset derives all its data from one signal, careful examination of the raw data is important to assess the quality of the flow signal. RESULTS--There was a correlation between the apnoea index (AI) assessed by the Autoset (AI-Autoset) and by polysomnography AI-PSG (r = 0.85) and between the AI-Autoset and the apnoea/hypopnoea index (AHI) during polysomnography (r = 0.87). The Autoset identified patients with an AHI-PSG of > 20 (a level of respiratory disturbance that would warrant consideration for treatment in most centres for sleep disorders) with a sensitivity of 82% and a specificity of 90%. CONCLUSIONS--The good correlation between the apnoea index measured by the Autoset and by polysomnography, and the high sensitivity in detecting patients with an AHI of > 20, may make the Autoset a valuable tool for the management of typical patients with sleep apnoea. However, very low values for nasal ventilation on the printout raises the suspicion of poor signal quality and misleading results.

Full text

PDF
1199

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Berthon-Jones M. Feasibility of a self-setting CPAP machine. Sleep. 1993 Dec;16(8 Suppl):S120–S123. doi: 10.1093/sleep/16.suppl_8.s120. [DOI] [PubMed] [Google Scholar]
  2. Chediak A. D., Lipson E., Demirozu M. C., Kiel M. The second generation of nasal continuous positive airway pressure devices. Are they created equal? Sleep. 1993 Oct;16(7):662–667. doi: 10.1093/sleep/16.7.662. [DOI] [PubMed] [Google Scholar]
  3. Condos R., Norman R. G., Krishnasamy I., Peduzzi N., Goldring R. M., Rapoport D. M. Flow limitation as a noninvasive assessment of residual upper-airway resistance during continuous positive airway pressure therapy of obstructive sleep apnea. Am J Respir Crit Care Med. 1994 Aug;150(2):475–480. doi: 10.1164/ajrccm.150.2.8049832. [DOI] [PubMed] [Google Scholar]
  4. Gould G. A., Whyte K. F., Rhind G. B., Airlie M. A., Catterall J. R., Shapiro C. M., Douglas N. J. The sleep hypopnea syndrome. Am Rev Respir Dis. 1988 Apr;137(4):895–898. doi: 10.1164/ajrccm/137.4.895. [DOI] [PubMed] [Google Scholar]
  5. Kryger M. H. Management of obstructive sleep apnea. Clin Chest Med. 1992 Sep;13(3):481–492. [PubMed] [Google Scholar]
  6. Mignot E., Nishino S., Guilleminault C., Dement W. C. Modafinil binds to the dopamine uptake carrier site with low affinity. Sleep. 1994 Aug;17(5):436–437. doi: 10.1093/sleep/17.5.436. [DOI] [PubMed] [Google Scholar]
  7. Miles L. E., Buschek G. D., McClintock D. P., Miles S. C., Narvios L. R., Wang Y. X. Development and application of automatic nasal CPAP calibration procedures for use in the unsupervised home environment. Sleep. 1993 Dec;16(8 Suppl):S118–S119. doi: 10.1093/sleep/16.suppl_8.s118. [DOI] [PubMed] [Google Scholar]
  8. Phillipson E. A. Sleep apnea--a major public health problem. N Engl J Med. 1993 Apr 29;328(17):1271–1273. doi: 10.1056/NEJM199304293281712. [DOI] [PubMed] [Google Scholar]
  9. Polo O., Berthon-Jones M., Douglas N. J., Sullivan C. E. Management of obstructive sleep apnoea/hypopnoea syndrome. Lancet. 1994 Sep 3;344(8923):656–660. doi: 10.1016/s0140-6736(94)92089-3. [DOI] [PubMed] [Google Scholar]
  10. Sullivan C. E., Issa F. G., Berthon-Jones M., Eves L. Reversal of obstructive sleep apnoea by continuous positive airway pressure applied through the nares. Lancet. 1981 Apr 18;1(8225):862–865. doi: 10.1016/s0140-6736(81)92140-1. [DOI] [PubMed] [Google Scholar]
  11. Young T., Palta M., Dempsey J., Skatrud J., Weber S., Badr S. The occurrence of sleep-disordered breathing among middle-aged adults. N Engl J Med. 1993 Apr 29;328(17):1230–1235. doi: 10.1056/NEJM199304293281704. [DOI] [PubMed] [Google Scholar]