Abstract
Test strips comprising B subtilis spores or M phlei cells dried on aluminium foil (some serum-coated) were used to test sterilizing processes currently advised for safety cabinets. Exposure to formalin from a bowl, boiling excess formalin with the fan running, and ultraviolet radiation were all found to be unreliable, as was glutaraldehyde. Vaporization of 2 ml of formalin per 0·028 m3 followed by exposure for 18 hours at ambient temperatures was 100% successful, and the depolymerization of paraformaldehyde tablets with a hair drier was a simple, very effective alternative method.
Full text
PDF



Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Braymen D. T., Songer J. R. Method for quantitating formaldehyde gas in air. Appl Microbiol. 1970 Jun;19(6):1021–1022. doi: 10.1128/am.19.6.1021-1022.1970. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Collins F. M. Relative susceptibility of acid-fast and non-acid-fast bacteria to ultraviolet light. Appl Microbiol. 1971 Mar;21(3):411–413. doi: 10.1128/am.21.3.411-413.1971. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Darlow H. M. The design of microbiological safety cabinets. Chem Ind. 1967 Nov 11;45:1914–1916. [PubMed] [Google Scholar]
- Morris E. J. The practical use of ultraviolet radiation for disinfection purposes. Med Lab Technol. 1972 Jan;29(1):41–47. [PubMed] [Google Scholar]
- Taylor L. A., Barbeito M. S., Gremillion G. G. Paraformaldehyde for surface sterilization and detoxification. Appl Microbiol. 1969 Apr;17(4):614–618. doi: 10.1128/am.17.4.614-618.1969. [DOI] [PMC free article] [PubMed] [Google Scholar]