Patient Report and Review of Rapidly Growing Mycobacterial Infection after Cardiac Device Implantation

Varun K Phadke 1,, David S Hirsh 1, Neela D Goswami 1
PMCID: PMC4766885  PMID: 26890060

As more of these devices are implanted, such infections are likely to be more frequently reported.

Keywords: Implantable cardioverter-defibrillator, cardiac pacemaker, nontuberculous mycobacteria, rapidly growing mycobacteria, Mycobacterium fortuitum group, bacteria, tuberculosis and other mycobacteria, bloodstream infections

Abstract

Mycobacterial infections resulting from cardiac implantable electronic devices are rare, but as more devices are implanted, these organisms are increasingly emerging as causes of early-onset infections. We report a patient with an implantable cardioverter-defibrillator pocket and associated bloodstream infection caused by an organism of the Mycobacterium fortuitum group, and we review the literature regarding mycobacterial infections resulting from cardiac device implantations. Thirty-two such infections have been previously described; most (70%) were caused by rapidly growing species, of which M. fortuitum group species were predominant. When managing such infections, clinicians should consider the potential need for extended incubation of routine cultures or dedicated mycobacterial cultures for accurate diagnosis; combination antimicrobial drug therapy, even for isolates that appear to be macrolide susceptible, because of the potential for inducible resistance to this drug class; and the arrhythmogenicity of the antimicrobial drugs traditionally recommended for infections caused by these organisms.


Infection is an uncommon but potentially devastating complication of cardiac implantable electronic device (CIED) implantation (1). Staphylococcal species cause most of these infections, followed by other pyogenic bacteria (1). CIED infections caused by mycobacteria have been reported infrequently; most of these infections result from the rapidly growing nontuberculous mycobacteria. We report a patient with an infection caused by a rapidly growing mycobacterium (RGM) in the Mycobacterium fortuitum group that developed following placement of an implantable cardioverter-defibrillator (ICD). We also review the published literature of cardiac device–associated infections caused by mycobacteria, focusing particularly on RGM.

Patient Report

A 60-year-old man with chronic systolic heart failure, emphysema, and hepatitis C infection was admitted to Grady Memorial Hospital (Atlanta, Georgia, USA) with 2 days of purulent drainage from his ICD pocket site. Just before his admission, pain had developed, followed by spontaneous dehiscence (i.e., separation of the surgical incision along the suture line) of the pocket. He reported no fever or constitutional symptoms. Two months earlier, a single-chamber ICD had been inserted uneventfully in his left prepectoral region after he had an episode of ventricular fibrillation that led to cardiac arrest. At the time of device implantation, he had been receiving intravenous vancomycin and piperacillin/tazobactam for concomitant hospital-acquired pneumonia, and he was prescribed oral cephalexin for 5 days after the procedure. During outpatient follow-up visits at weeks 2 and 4 after device insertion, the patient’s wound was noted as unremarkable.

At the time of the new admission, the patient was afebrile with unremarkable vital signs. Physical examination showed mild, nontender edema over the ICD pocket; a 1-cm, shallow ulceration of the incision site; scant serous drainage; and minimal surrounding erythema. The patient had no peripheral stigmata of infective endocarditis, and the remainder of the examination was unremarkable. Laboratory studies showed a normal leukocyte count (5,500 cells/μL) with a slight (70%) neutrophil predominance. Blood cultures were collected, and he was started empirically on intravenous vancomycin for a suspected pocket infection. Transthoracic echocardiography was performed and showed no valvular vegetations. A transesophageal echocardiogram was deferred because of considerable laryngeal stenosis caused by traumatic endotracheal intubation at the time of his recent cardiac arrest.

On hospital day 2, the entire device, including the ICD generator and leads, was removed and sent for culture. Purulent fluid and necrotic tissue were noted in the pocket during explantation, and a drain was left in place at the time of wound closure (Figure, panel A). Gram stain of the pocket exudate showed beaded gram-positive rods. Within 5 days, the generator pocket tissue culture and 1 of 2 sets of blood cultures were growing an aerobic, gram-positive rod that also appeared beaded on Gram stain. Culture of the lead tip remained sterile. Because of the Gram-stain appearance of the blood culture isolate, an acid-fast stain was performed in the microbiology laboratory, and results showed the organism to be acid fast. At the request of the inpatient infectious diseases consultation service, the isolate was subcultured to mycobacterial growth media and was identified as an RGM. An HIV test result was negative.

Figure.

Figure

Photographs of the cardiac implantable electronic device pocket site for a 60-year-old man in whom infection developed at the implantation site of a cardiac implantable electronic device, Atlanta, Georgia, USA. A) Device pocket site after explantation. The wound was closed with pledged Ethibond sutures (Ethicon, Somerville, NJ, USA), and a Jackson-Pratt drain (closed-suction drainage system consisting of an internal drain connected by plastic tubing to a flexible bulb) was tunneled into the inferior aspect of the pocket. The drain was removed 24 hours postoperatively, and a small incision was left open to heal by secondary intention. B) Device pocket site 6 weeks after suture removal. Most of the incision healed well, with evidence of localized dehiscence (i.e., spontaneous partial separation of the surgical incision along the suture lines).

The patient’s antimicrobial drug regimen was empirically changed to intravenous cefoxitin, oral ciprofloxacin, and oral clarithromycin, a combination selected because of a clinical suspicion of M. fortuitum infection, surmised from the limited literature on RGM-associated CIED. Because of uncertainty about the capacity for outpatient therapeutic drug monitoring and because the patient had undergone surgical debridement and was thought to be clinically improved at the time of discharge, intravenous aminoglycoside therapy was deferred pending species identification and test results regarding antimicrobial drug susceptibility. A baseline electrocardiogram obtained before initiation of antimicrobial drugs showed a corrected QT interval (i.e., duration from start of Q wave to end of T wave in the heart’s electrical cycle) of 511 ms (reference <430 ms).

One week after device extraction, a computed tomography scan of the patient’s chest was performed, and results showed a residual, subclinical, peripherally enhancing, 2.7 × 1.0–cm collection of air and fluid over the left pectoralis muscle. Immediate device reimplantation was deferred. He was discharged on hospital day 14 with a wearable cardioverter-defibrillator to use until his physicians believed that device reimplantation was safe. The therapeutic plan at discharge was for at least 6 months of antimicrobial drug therapy, beginning with the cefoxitin, ciprofloxacin, and clarithromycin regimen and subsequently tailored to the organism and susceptibility profile when these data became available.

Five weeks after device removal, the patient returned for outpatient follow-up and reported full adherence to his prescribed antimicrobial drug regimen. On examination, most of the wound over the previous device pocket had healed, but localized dehiscence and ongoing purulent drainage was evident (Figure, panel B). He was referred to plastic surgery for ongoing management of his wound. Identification and antimicrobial drug susceptibilities of the organism were still pending, so his antimicrobial drug regimen was not changed. Unfortunately, he was lost to follow-up; 2 months later, he discontinued use of his wearable cardioverter-defibrillator and died of out-of-hospital cardiac arrest.

By use of high-performance liquid chromatography, the patient’s isolate was identified as a Mycobacterium fortuitum group organism. Further genetic testing for species-level identification was not performed. The following susceptibilities were identified: cefoxitin intermediate (MIC 64 μg/mL), ciprofloxacin susceptible (MIC <0.12 μg/mL), and clarithromycin resistant (MIC 8 μg/mL after 14 days of incubation) (Table 1).

Table 1. Antimicrobial drug susceptibility profile of patient’s Mycobacterium fortuitum group isolate*.

Antimicrobial drug MIC, μg/mL Interpretation*
Amikacin <1 Susceptible
Cefoxitin 64 Intermediate
Ciprofloxacin <0.12 Susceptible
Clarithromycin 8 Resistant†
Doxycycline >16 Resistant
Imipenem 8 Intermediate
Linezolid 2 Susceptible
Moxifloxacin <0.25 Susceptible
Tigecycline 0.12
Trimethoprim/sulfamethoxazole 1/19 Susceptible
Tobramycin >16 Resistant

*According to breakpoints defined by the Clinical and Laboratory Standards Institute (2).
†Clarithromycin MIC after 14 d of incubation.
‡No accepted breakpoints from the Clinical and Laboratory Standards Institute exist for tigecycline.

Discussion

The RGM species are ubiquitous environmental organisms that have been isolated from soil, food, natural and municipal water, various plants and animals, and hospital surfaces (3). These organisms are not believed to be permanent members of the human bacterial flora but often become transient colonizers after frequent exposure. The most commonly encountered RGM species in clinical practice are M. abscessus, M. chelonae, and M. fortuitum, but >100 species have been identified (3). These organisms are capable of growth on standard mycobacterial (e.g., Middlebrook 7H11 or Lownstein-Jensen) and routine bacteriologic (e.g., sheep’s blood and MacConkey agar) growth media. However, colonies may take >5 days to appear on standard media, exceeding the incubation time of routine cultures in many clinical microbiology laboratories. Even when growth is observed, these organisms often appear as beaded gram-positive bacilli on routine Gram stain and may be misidentified as contaminants (4). Therefore, a high index of suspicion for a potential RGM infection is needed for an accurate diagnosis.

Despite increasing recognition that these organisms can cause infections associated with prosthetic devices and surgical sites, RGM infections complicating implanted cardiac devices are still uncommon. We searched the available literature using PubMed with no starting date restrictions through March 31, 2015, and identified only 32 previously reported cases of CIED infections caused by any mycobacterial species. Including our patient, 23 (70%) of 33 reported infections were caused by an RGM species (524) (Table 2). We found 2 reports of CIED infections caused by M. avium complex (26,27) and 8 reports of infections caused by M. tuberculosis complex organisms (2833). Of the 23 RGM infections, 21 (91%) were reported in the past 10 years, a trend likely resulting from improvements in microbiologic techniques and increased recognition of these organisms as causative pathogens. Mean age of case-patients with RGM infections was 65.4 years, consistent with age trends for CIED implantation. Sixteen (70%) case-patients had infections associated with permanent pacemakers. Among 21 case-patients for which time of onset was reported, 5 (24%) infections developed >6 months after the most recent device manipulation. Although cardiac devices can become secondarily infected because of seeding from incidental bloodstream infections, the RGM species are uncommon causes of bacteremia. Instead, the source of early-onset CIED infections is more likely inoculation of the organism into the pocket at the time of the implantation procedure. This source contrasts with the probable source for the 8 reported CIED infections caused by M. tuberculosis complex. Manifesting >11 months after device manipulation, these infections more likely resulted from reactivation disease, mycobacteremia, and secondary seeding of the device.

Table 2. Clinical and demographic information for published cases of cardiac device infections due to rapidly growing mycobacteria*.

Year (ref) Age, y/
sex Organism Type Onset† Bacteremia/
lead infection‡ IE§ Macrolide resistant Device removed Antimicrobial drug therapy Outcome
Mycobacterium fortuitum group
1998 (6) 74/M M. fortuitum + M. chelonae PPM 13 d NR/NR NR NR Yes FQ + AG × 4 wk Cured
2005 (9) 62/F M. fortuitum PPM 6 mo Yes/yes Yes No Yes CLR + CIP × 4 wk, DOX + CIP × 24 wk Cured
2005 (10) 74/M M. peregrinum ICD 6 wk Yes/NR NR No Yes CLR + CIP × 6 wk Cured
2005 (8) 72/M M. fortuitum PPM 2 wk No/NR No Yes Yes CIP + AG × 2 wk, CIP × 6 mo Cured
61/M M. fortuitum ICD 17 mo No/yes No Yes Yes LVX × >1 y¶ Cured
2006 (11) 80/M M. fortuitum PPM 18 d Yes/NA No No No CLR + CIP × 6 wk Cured
2007 (13) 84/F M. fortuitum PPM 1 mo No/no No Yes Yes LVX × 3 mo Cured
2007 (15) 78/F M. fortuitum PPM 3 mo Yes/yes No NR Yes CLR + LVX + LZD × 2 wk, CLR + LVX × 6 mo Cured
2007 (16) 78/F M. fortuitum PPM <4 mo Yes/NR NR NR Yes CLR + LVX + LZD × 2 wk, CLR + LVX × 22 wk Cured
77/F M. mageritense PPM 3 wk NR/NA NR NR No FQ × 6 mo Cured
2009 (18) 15/F M. fortuitum PPM 7 wk Yes/yes No No Yes CLR + CIP × 6 mo Cured
2010 (20) 78/M M. fortuitum PPM NR Yes/NR NR NR Yes CLR + CIP × 26 wk Cured
2012 (23) 43/M M. fortuitum ICD 4 y Yes/yes Yes No Yes CLR + CIP + AG Died
2012 (22) 75/M M. peregrinum PPM 1 y Yes/yes No NR Yes CLR + CIP × mo Cured
2015#
60/M
M. fortuitum group
ICD
6 wk
Yes/no
No
Yes
Yes
CLR + CIP + FOX
Died
M. abscessus complex
1998 (5) 68/M M. abscessus PPM 19 y NR/yes NR No Yes CLR + AG + FOX × 5 wk Died
2005 (7) 53/M M. abscessus ICD 2 wk NR/NR NR Yes Yes CLR × 24 wk Cured
2007 (14)
43/F
M. massiliense
PPM
11 mo
NR/yes
NR
No
Yes
CLR × 6 mo
Cured
M. smegmatis complex
2006 (12) 86/M M. goodii PPM 16 d Yes/NR NR NR Yes Multiple, ending with MIN + AG × 2 wk Cured
2008 (17) 85/M M. goodii ICD <7 d No/NR No NR Yes TMP/SXT × 8 wk Cured
2009 (19)
23/M
M. goodii
PPM
8 d
No/NA
No
Yes
No
DOX + FQ × 6 mo
Cured
M. chelonae complex
2014 (24)
63/M
M. chelonae
PPM
NR
No/yes
Yes
NR
Yes
CLR + LVX + AG × >2 mo
Cured
Ungrouped rapidly growing species
2011 (21) 73/M M. phlei ICD 1 mo No/NR NR No Yes SXT + DOX × 12 mo Cured

*AG, aminoglycoside; CIP, ciprofloxacin; CLR, clarithromycin; DOX, doxycycline; FOX, cefoxitin; FQ, fluoroquinolone other than CIP or LVX; ICD, implantable cardioverter-defibrillator; IE, infective endocarditis; LVX, levofloxacin; LZD, linezolid; MIN, minocycline; NA, not available; NR, not reported; PPM, permanent pacemaker; ref, reference; TMP/SXT, trimethoprim/sulfamethoxazole.
†Time since most recent device manipulation.
‡Defined as positive lead culture, acid fast stain, or presumptive diagnosis on the basis of imaging or operative findings.
§Transthoracic or transesophageal echocardiographic findings as defined by the Duke criteria (25).
¶No end date for therapy was specified, but patient had at least 1 year of treatment.
#The patient described in this article.

The most commonly isolated organisms have been in the M. fortuitum group, which account for ≈50% of mycobacterial CIED infections in patients and nearly two thirds of infections caused by an RGM species. The M. fortuitum group has historically included M. fortuitum and M. peregrinum, although M. mageritense and others have also been proposed as members of this group (3). The preponderance of M. fortuitum infections among patients with cardiac device implantations mirrors trends observed for poststernotomy (34) and postaugmentation mammoplasty (35) infections caused by RGM. Although most skin and soft tissue infections caused by RGM, particularly after surgical or nonsurgical trauma, result from M. fortuitum (3), this organism is not considered a normal skin commensal. Sources of these infections are instead thought to be largely environmental (3). Nevertheless, among all skin and soft tissue infections caused by RGM, those on the chest or back seem more likely to result from M. fortuitum than from other RGM species (36).

Another trend we observed was that 11 (48%) of the 23 patients with a RGM infection had associated mycobacteremia (5 had no reported blood culture results). This finding indicates that the infection had spread beyond the device pocket to the intravascular component of the CIED system, suggesting endovascular infection. In the 13 (57%) patients for which both blood culture results and echocardiographic findings were reported, 4 (31%) had device-related endocarditis, as defined by the Duke criteria (9,18,2325). Three of these 4 patients fulfilled clinical criteria for infective endocarditis on the basis of echocardiographic findings; the fourth had no echocardiographic abnormalities but fulfilled pathologic criteria on the basis of isolation of the organism in an operative culture. This patient was the only one with valvular endocarditis among all the reports in our review (18).

Conversely, in 4 patients, including the patient described in this article, mycobacteremia was detected in the absence of echocardiographic abnormalities. The patient we describe had an unremarkable transthoracic echocardiogram, but a transesophageal echocardiogram could not be performed for definitive evaluation of CIED-related endocarditis. Overall, the low rates of valvular endocarditis or disseminated infection suggest that CIED infections caused by an RGM behave similarly to catheter-related bloodstream infections caused by these organisms (37), although severe complications of CIED infections associated with bacteremia have rarely been described (18,23).

The Clinical and Laboratory Standards Institute recommends routine broth microdilution susceptibility testing of all RGM isolates against amikacin, cefoxitin, ciprofloxacin, clarithromycin, doxycycline, imipenem, and sulfamethoxazole (or trimethoprim/sulfamethoxazole), but newer agents (e.g., linezolid, moxifloxacin, and tigecycline) also frequently show in vitro activity against these organisms (2). Of all RGMs, the M. fortuitum group is traditionally considered the most susceptible to antimicrobial drugs, with isolates frequently being susceptible to many agents tested. However, current guidelines by the American Thoracic Society and the Infectious Diseases Society of America recommend therapy with >2 active drugs for several months for optimal results. The isolate of the patient in this report showed resistance to several tested agents, including clarithromycin (Table 1). Among patients with CIED infections caused by M. fortuitum group organisms for which susceptibility data were reported, only 4 (including the patient reported in this article) had macrolide resistance. A previous review of the RGM similarly noted that most M. fortuitum clinical isolates were macrolide susceptible by in vitro methods (3).

Genetic studies published after that review revealed that most, if not all, M. fortuitum isolates also harbor an M. fortuitum rRNA methylase gene, termed erm(39), that, if active, can confer macrolide resistance (38,39). Although the clinical significance of this potential mechanism of inducible macrolide resistance is unclear, particularly in strains in which the gene is inactive at baseline, this finding has led many experts to advise caution to clinicians who consider prescribing macrolide-based regimens for serious M. fortuitum infections, even when the isolate is reported as susceptible to macrolides on the basis of broth microdilution methods (2). Nevertheless, in most reported CIED infections caused by macrolide-susceptible M. fortuitum group organisms, a macrolide and fluoroquinolone combination has been used successfully. We speculate that the patient described in this article experienced delayed wound healing resulting from inadequate activity of the empirical antimicrobial regimen against his isolate.

Given the recommended antimicrobial drug regimens for infections caused by an RGM, cardiac device infections resulting from these organisms can pose a unique therapeutic dilemma. On the one hand, the propensity to biofilm formation makes these organisms difficult to eradicate with antimicrobial drug therapy alone. Consequently, most experts advocate an extended course of antimicrobial drugs combined with device removal (37), the strategy used in all but 3 (13%) of the 23 previously described case-patients with infections caused by an RGM. On the other hand, some of the most active agents against this group of organisms belong to the macrolide or fluoroquinolone classes, which are both types of antimicrobial drugs with potential proarrhythmic effects (39,40). An antimicrobial drug combination that has the potential to precipitate arrhythmias becomes problematic in patients being considered for device removal and having a preexisting risk for conduction abnormalities (caused by an underlying conduction disease, cardiomyopathy, or concomitant proarrhythmic medications). The patient we describe was noted to have a prolonged corrected QT interval before the start of macrolide or fluoroquinolone therapy, a circumstance that made selection of an appropriate empirical antimicrobial regimen challenging. This patient highlights the importance of expedited antimicrobial drug susceptibility testing for managing these infections, including evaluation of newer antimicrobial drugs with fewer direct arrhythmogenic effects than those resulting from macrolides or fluoroquinolones.

Infections occurring after implantation of cardiac devices and caused by nontuberculous mycobacteria are uncommon, but as more devices are implanted, such infections are likely to be more frequently reported. Our patient illustrates many of the common clinical features of postimplantation CIED infections caused by RGMs, including early onset (<6 months from the most recent manipulation of the device) of disease, initial identification of the organism as a gram-positive bacillus, and isolation of a M. fortuitum group organism as the causative pathogen. In addition, this patient highlights several unique issues that warrant further investigation, such as reliability of macrolide therapy for M. fortuitum group infections and safety of long-term macrolide and fluoroquinolone use in patients with a preexisting high risk for serious arrhythmias.

Acknowledgments

The authors thank Makoto Mori for his assistance with translation of a case report in the literature.

V.K.P. is supported by the Emory Vaccinology Training Program under award no. T32AI074492 from the National Institute of Allergy and Infectious Diseases. The content of this review is solely the responsibility of the authors and does not necessarily represent the official views of the National Institute of Allergy and Infectious Diseases or the National Institutes of Health.

Biography

Dr. Phadke is an infectious diseases fellow at Emory University in Atlanta. His research interests include vaccine-preventable diseases.

Footnotes

Suggested citation for this article: Phadke VK, Hirsh DS, Goswami ND. Patient report and review of rapidly growing mycobacterial infection after cardiac device implantation. Emerg Infect Dis. 2016 Mar [date cited]. http://dx.doi.org/10.3201/eid2203.150584

References

  • 1.Uslan DZ, Sohail MR, St Sauver JL, Friedman PA, Hayes DL, Stoner SM, et al. Permanent pacemaker and implantable cardioverter defibrillator infection: a population-based study. Arch Intern Med. 2007;167:669–75. 10.1001/archinte.167.7.669 [DOI] [PubMed] [Google Scholar]
  • 2.Clinical and Laboratory Standards Institute. Susceptibility testing of mycobacteria, nocardiae, and other aerobic actinomycetes; approved standard. CLSI Document M24–A2. Wayne (PA); 2011. [PubMed] [Google Scholar]
  • 3.Brown-Elliott BA, Wallace RJ Jr. Clinical and taxonomic status of pathogenic nonpigmented or late-pigmenting rapidly growing mycobacteria. Clin Microbiol Rev. 2002;15:716–46. 10.1128/CMR.15.4.716-746.2002 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.von Graevenitz A, Punter-Streit V. Failure to recognize rapidly growing mycobacteria in a proficiency testing sample without specific request—a wider diagnostic problem? Eur J Epidemiol. 1998;14:519–20. 10.1023/A:1007463630978 [DOI] [PubMed] [Google Scholar]
  • 5.Cutay AM, Horowitz HW, Pooley RW, Van Horn K, Wormser GP. Infection of epicardial pacemaker wires due to Mycobacterium abscessus. Clin Infect Dis. 1998;26:520–1. 10.1086/517106 [DOI] [PubMed] [Google Scholar]
  • 6.Verghese S, Mullaseri A, Padmaja P, Subhadra AC, Cherian KM. Pacemaker implant site infection caused by atypical mycobacteria. Indian Heart J. 1998;50:201–2. [PubMed] [Google Scholar]
  • 7.Kessler AT, Kourtis AP. Mycobacterium abscessus as a cause of pacemaker infection. Med Sci Monit. 2004;10:CS60–2. [PubMed] [Google Scholar]
  • 8.Hemmersbach-Miller M, Cardenes-Santana MA, Conde-Martel A, Bolanos-Guerra JA, Campos-Herrero MI. Cardiac device infections due to Mycobacterium fortuitum. Can J Infect Dis Med Microbiol. 2005;16:183–5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Sharma S, Tleyjeh IM, Espinosa RE, Costello BA, Baddour LM. Pacemaker infection due to Mycobacterium fortuitum. Scand J Infect Dis. 2005;37:66–7. 10.1080/00365540410020910 [DOI] [PubMed] [Google Scholar]
  • 10.Short WR, Emery C, Bhandary M, O’Donnell JA. Misidentification of Mycobacterium peregrinum, the causal organism of a case of bacteremia and automatic implantable cardioverter defibrillator-associated infection, due to its unusual acid-fast staining characteristics. J Clin Microbiol. 2005;43:2015–7. 10.1128/JCM.43.4.2015-2017.2005 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Pastor E, Luz Andreu A, Llombart M, Chiner E. Mycobacterium fortuitum: a rare cause of pacemaker infection [in Spanish]. Enferm Infecc Microbiol Clin. 2006;24:136–7. 10.1157/13085023 [DOI] [PubMed] [Google Scholar]
  • 12.Toda H, Sato K, Iimori M, Yamazumi T, Furuta I, Satoh A, et al. A case of Mycobacterium goodii infection with isolation from blood and a pacemaker lead [in Japanese]. Kansenshogaku Zasshi. 2006;80:262–6. 10.11150/kansenshogakuzasshi1970.80.262 [DOI] [PubMed] [Google Scholar]
  • 13.Giannella M, Valerio M, Franco JA, Marin M, Bouza E, Munoz P. Pacemaker infection due to Mycobacterium fortuitum: the role of universal 16S rRNA gene PCR and sequencing. Diagn Microbiol Infect Dis. 2007;57:337–9. 10.1016/j.diagmicrobio.2006.08.010 [DOI] [PubMed] [Google Scholar]
  • 14.Simmon KE, Pounder JI, Greene JN, Walsh F, Anderson CM, Cohen S, et al. Identification of an emerging pathogen, Mycobacterium massiliense, by rpoB sequencing of clinical isolates collected in the United States. J Clin Microbiol. 2007;45:1978–80. 10.1128/JCM.00563-07 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Siu CW, Cheng LC, Woo PC, Lau CP, Tse HF. A patient with relapsing pacemaker infection due to “Gram-positive bacilli”. Int J Cardiol. 2007;114:E40–1. 10.1016/j.ijcard.2006.07.211 [DOI] [PubMed] [Google Scholar]
  • 16.Tam WO, Yew WW, Yam WC, Yuen KY, Wong PC, Tse TF. Pacemaker infections due to rapidly growing mycobacteria: further experience. Int J Tuberc Lung Dis. 2007;11:118. [PubMed] [Google Scholar]
  • 17.Chrissoheris MP, Kadakia H, Marieb M, Libertin C. Pacemaker pocket infection due to Mycobacterium goodii: case report and review of the literature. Conn Med. 2008;72:75–7. [PubMed] [Google Scholar]
  • 18.Al Soub H, Al Maslamani M, Al Khuwaiter J, El Deeb Y, Abu Khattab M. Myocardial abscess and bacteremia complicating Mycobacterium fortuitum pacemaker infection: case report and review of the literature. Pediatr Infect Dis J. 2009;28:1032–4. 10.1097/INF.0b013e3181aa6592 [DOI] [PubMed] [Google Scholar]
  • 19.Marchandin H, Battistella P, Calvet B, Darbas H, Frapier JM, Jean-Pierre H, et al. Pacemaker surgical site infection caused by Mycobacterium goodii. J Med Microbiol. 2009;58:517–20. 10.1099/jmm.0.006759-0 [DOI] [PubMed] [Google Scholar]
  • 20.van Duin D, Goldfarb J, Schmitt SK, Tomford JW, Tuohy MJ, Hall GS. Nontuberculous mycobacterial blood stream and cardiac infections in patients without HIV infection. Diagn Microbiol Infect Dis. 2010;67:286–90. 10.1016/j.diagmicrobio.2010.02.006 [DOI] [PubMed] [Google Scholar]
  • 21.Karnam S, Alla VM, Kwon J, Harbert T, Sharma A, Airey K, et al. Mycobacterium phlei, a previously unreported cause of pacemaker infection: thinking outside the box in cardiac device infections. Cardiol J. 2011;18:687–90. 10.5603/CJ.2011.0034 [DOI] [PubMed] [Google Scholar]
  • 22.Amraoui S, Texier-Maugein J, Bordachar P. PET scan in suspected but unproven pacemaker endocarditis. Arch Cardiovasc Dis. 2012;105:125–6. 10.1016/j.acvd.2011.04.011 [DOI] [PubMed] [Google Scholar]
  • 23.Sharma H, Keshavan A, Little MA, Cross J, Lipman MC, Talukdar S, et al. Fortuitous vasculitis. Ren Fail. 2012;34:378–82. 10.3109/0886022X.2011.647337 [DOI] [PubMed] [Google Scholar]
  • 24.Hooda A, Pati PK, John B, George PV, Michael JS. Disseminated Mycobacterium chelonae infection causing pacemaker lead endocarditis in an immunocompetent host. BMJ Case Rep. 2014;2014:pii: bcr201406042. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Li J, Sexton D, Mick N, Nettles R, Fowler V, Ryan T et al. Proposed modifications to the Duke criteria for the diagnosis of infective endocarditis. Clin Infect Dis. 2000;30:633–8. 10.1086/313753 [DOI] [PubMed] [Google Scholar]
  • 26.Amin M, Gross J, Andrews C, Furman S. Pacemaker infection with Mycobacterium avium complex. Pacing and clinical electrophysiology. Pacing Clin Electrophysiol. 1991;14:152–4. 10.1111/j.1540-8159.1991.tb05083.x [DOI] [PubMed] [Google Scholar]
  • 27.Katona P, Wiener I, Saxena N. Mycobacterium avium–intracellulare infection of an automatic implantable cardioverter defibrillator. Am Heart J. 1992;124:1380–1. 10.1016/0002-8703(92)90429-Y [DOI] [PubMed] [Google Scholar]
  • 28.Doherty JG, Rankin R, Kerr F. Miliary tuberculosis presenting as infection of a pacemaker pulse-generator pocket. Scott Med J. 1996;41:20–1. [DOI] [PubMed] [Google Scholar]
  • 29.Hellwig T, Ou P, Offredo C, Stephany D, Bonnet D, Sidi D. Unusual chronic pacemaker infection by Mycobacterium tuberculosis in a pediatric patient. J Thorac Cardiovasc Surg. 2005;130:937–8. 10.1016/j.jtcvs.2005.04.016 [DOI] [PubMed] [Google Scholar]
  • 30.Kestler M, Reves R, Belknap R. Pacemaker wire infection with Mycobacterium tuberculosis: a case report and literature review. Int J Tuberc Lung Dis. 2009;13:272–4. [PubMed] [Google Scholar]
  • 31.Luckie M, Zaidi A, Woodhead M, Garratt C. Mycobacterium tuberculosis causing infection of an implantable biventricular defibrillator. Indian J Tuberc. 2010;57:213–5. [PubMed] [Google Scholar]
  • 32.Stone DR, Estes NA III, Klempner MS. Mycobacterium bovis infection of an implantable defibrillator following intravesical therapy with bacille Calmette-Guerin. Clin Infect Dis. 1993;16:825–6. 10.1093/clind/16.6.825 [DOI] [PubMed] [Google Scholar]
  • 33.Kumar A, Agrawal T. A study of unusual pacemaker infection by mycobacterium tuberculosis in Indian patients. Indian Pacing Electrophysiol J. 2014;14:291–6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Wallace RJ Jr, Musser JM, Hull SI, Silcox VA, Steele LC, Forrester GD, et al. Diversity and sources of rapidly growing mycobacteria associated with infections following cardiac surgery. J Infect Dis. 1989;159:708–16. 10.1093/infdis/159.4.708 [DOI] [PubMed] [Google Scholar]
  • 35.Wallace RJ Jr, Steele LC, Labidi A, Silcox VA. Heterogeneity among isolates of rapidly growing mycobacteria responsible for infections following augmentation mammaplasty despite case clustering in Texas and other southern coastal states. J Infect Dis. 1989;160:281–8. 10.1093/infdis/160.2.281 [DOI] [PubMed] [Google Scholar]
  • 36.Uslan DZ, Kowalski TJ, Wengenack NL, Virk A, Wilson JW. Skin and soft tissue infections due to rapidly growing mycobacteria: comparison of clinical features, treatment, and susceptibility. Arch Dermatol. 2006;142:1287–92. 10.1001/archderm.142.10.1287 [DOI] [PubMed] [Google Scholar]
  • 37.El Helou G, Hachem R, Viola GM, El Zakhem A, Chaftari AM, Jiang Y, et al. Management of rapidly growing mycobacterial bacteremia in cancer patients. Clin Infect Dis. 2013;56:843–6. 10.1093/cid/cis1032 [DOI] [PubMed] [Google Scholar]
  • 38.Nash KA, Zhang Y, Brown-Elliott BA, Wallace RJ Jr. Molecular basis of intrinsic macrolide resistance in clinical isolates of Mycobacterium fortuitum. J Antimicrob Chemother. 2005;55:170–7. 10.1093/jac/dkh523 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Schembri S, Williamson PA, Short PM, Singanayagam A, Akram A, Taylor J, et al. Cardiovascular events after clarithromycin use in lower respiratory tract infections: analysis of two prospective cohort studies. BMJ. 2013;346:f1235. 10.1136/bmj.f1235 [DOI] [PubMed] [Google Scholar]
  • 40.Lapi F, Wilchesky M, Kezouh A, Benisty JI, Ernst P, Suissa S. Fluoroquinolones and the risk of serious arrhythmia: a population-based study. Clin Infect Dis. 2012;55:1457–65. 10.1093/cid/cis664 [DOI] [PubMed] [Google Scholar]

Articles from Emerging Infectious Diseases are provided here courtesy of Centers for Disease Control and Prevention

RESOURCES