Skip to main content
Journal of Clinical Pathology logoLink to Journal of Clinical Pathology
. 1978 Jan;31(1):63–68. doi: 10.1136/jcp.31.1.63

Cross contamination of cytological smears, with automated staining machines and bulk manual staining procedures. With a specific study of the problems of the Cytotek and the Shandon Elliott staining machines.

O A Husain, J M Grainger, J Sims
PMCID: PMC476722  PMID: 75214

Abstract

Further development of an individual staining machine is to be strongly encouraged but meanwhile, using bulk stainers, frequent changing of wash fluids and staining solutions, particularly leading up to and following the haematoxylin pot, is essential to reduce the risk of cross contamination. Certain smears, such as from semen or from serous fluids where malignancy is suspected or known, must be stained on separate racks. In some laboratories it is the rule not to stain semen or serous fluids in bulk staining machines at all and this may have to become the rule everywhere until we are provided with safe individual slide stainers.

Full text

PDF
63

Images in this article

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Barr W. T., Powell D. E., Raffan J. B. Cellular contamination during automatic and manual staining of cytological smears. J Clin Pathol. 1970 Oct;23(7):604–607. doi: 10.1136/jcp.23.7.604. [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Cremer A. W. Automatic slide staining in clinical bacteriology. J Med Lab Technol. 1968 Oct;25(4):387–390. [PubMed] [Google Scholar]
  3. Drew W. L., Pedersen A. N., Roy J. J. Automated slide staining machine. Appl Microbiol. 1972 Jan;23(1):17–20. doi: 10.1128/am.23.1.17-20.1972. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Hoffmann R. Einsatz eines Objektträger-Färbeautomaten im hämatologischen Grosslaboratorium. Med Lab (Stuttg) 1972 Jul;25(7):147–151. [PubMed] [Google Scholar]
  5. Moss E. D. Automated slide staining in haematology. Can J Med Technol. 1968 Oct;30(5):169–178. [PubMed] [Google Scholar]
  6. Pickett J. P. A survey of automatic staining equipment. Am J Med Technol. 1973 Nov;39(11):483–485. [PubMed] [Google Scholar]
  7. Ryan R. W., Sedgwick A. K., Tilton R. C. Laboratory evaluation of an automatic gram staining machine. Health Lab Sci. 1973 Apr;10(2):82–87. [PubMed] [Google Scholar]
  8. Wilkins J. R., Mills S. M. Automated single-slide staining device. Appl Microbiol. 1975 Sep;30(3):485–488. doi: 10.1128/am.30.3.485-488.1975. [DOI] [PMC free article] [PubMed] [Google Scholar]