Restricted Identification of Clinical Pathogens Categorized as Biothreats by Matrix-Assisted Laser Desorption Ionization–Time of Flight Mass Spectrometry

Peter M Keller 1,, Vera Bruderer 1, Franco Müller 1
Editor: E Munson
PMCID: PMC4767975  PMID: 26763958

LETTER

These days, many public and private clinical microbiology laboratories use matrix-assisted laser desorption ionization–time of flight mass spectrometry (MALDI-TOF MS) as a primary means of identification of bacterial isolates (1). In our laboratory, we use a MALDI Biotyper (Bruker Daltonics, Bremen, Germany) with MALDI Biotyper CA System software. This system is capable of identifying more than 90% of all routinely encountered bacterial isolates to the species level (2). In July 2015, we repeatedly isolated morphologically similar, fastidious bacterial colonies in subcultures of enrichment cultures (thioglycolate broth) of joint tissue biopsy specimens from an 84-year-old patient with a complicated clinical course following total knee endoprosthesis surgery. The isolated bacterial colonies had a fine, grayish morphology on cooked blood agar; the Gram stain showed faint Gram-negative coccoid rods. The MALDI Biotyper spectra were of good quality (based on the number of spectral peaks), but the software repeatedly reported no identification. On the basis of 16S rRNA gene sequencing and SmartGene IDNS (SmartGene, Zug, Switzerland) sequence homology searches, the bacterial isolates were finally identified as Francisella tularensis (3). Because of export restrictions, the Bruker Biotyper database does not allow the identification of potential agents of bioterrorism. Customers wishing to identify F. tularensis have to acquire a separate database termed a security-relevant database and have to obtain governmental export permission prior to installation.

Zoonotic organisms are rarely, if ever, implicated in prosthetic device infections. A single case report of F. tularensis as the cause of a chronic total-knee endoprosthesis infection is described in the literature (4). During the laboratory workup, a total of 10 laboratory personal inadvertently had significant exposure when working with primary material or cultures from this case.

Our case illustrates that commercially available MALDI-TOF MS-based identification systems, whose use is widespread in clinical microbiology laboratories for routine identification, do not allow the identification of potential agents of bioterrorism because of export restriction regulations (dual-use policy). This policy is problematic and potentially dangerous. We also note that our report is not without precedent (5). Together, these observations urge and reinforce the need for a policy change, as many bacteriology laboratories are about to lose biochemical and morphological identification methods in routine use. It is, however, important that laboratory directors have mechanisms in place to ensure the proper handling of potential biothreat organisms and confirmation of identification prior to the release of results. Relying on MALDI-TOF MS-based identification without a working knowledge base of other characteristics of the pathogen in question can result in undependable and erroneous identifications. These other characteristics may include growth characteristics on different media, colony morphology, Gram stain characteristics, motility, urease, oxidase, and results of other common rapid tests. Unrestricted, rapid identification of pathogens that are possible biothreats, such as Brucella spp., Bacillus anthracis, F. tularensis, and Yersinia pestis, is imperative in order to protect our employees from laboratory infections, to take the appropriate safety precautions, and to serve our patients by allowing diagnosis in a timely manner.

REFERENCES

  • 1.Fournier PE, Drancourt M, Colson P, Rolain JM, La Scola B, Raoult D. 2013. Modern clinical microbiology: new challenges and solutions. Nat Rev Microbiol 11:574–585. doi: 10.1038/nrmicro3068. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Schulthess B, Bloemberg GV, Zbinden R, Böttger EC, Hombach M. 2014. Evaluation of the Bruker MALDI Biotyper for identification of Gram-positive rods: development of a diagnostic algorithm for the clinical laboratory. J Clin Microbiol 52:1089–1097. doi: 10.1128/JCM.02399-13. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Bosshard PP, Zbinden R, Abels S, Boddinghaus B, Altwegg M, Böttger EC. 2006. 16S rRNA gene sequencing versus the API 20 NE system and the VITEK 2 ID-GNB card for identification of nonfermenting Gram-negative bacteria in the clinical laboratory. J Clin Microbiol 44:1359–1366. doi: 10.1128/JCM.44.4.1359-1366.2006. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Cooper CL, Van Caeseele P, Canvin J, Nicolle LE. 1999. Chronic prosthetic device infection with Francisella tularensis. Clin Infect Dis 29:1589–1591. doi: 10.1086/313550. [DOI] [PubMed] [Google Scholar]
  • 5.Dingle TC, Butler-Wu SM, Abbott AN. 2014. Accidental exposure to Burkholderia pseudomallei in the laboratory in the era of matrix-assisted laser desorption ionization–time of flight mass spectrometry. J Clin Microbiol 52:3490–3491. doi: 10.1128/JCM.01238-14. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)

RESOURCES