Abstract
PURPOSE
The purpose of this study was to evaluate the effects of abutment screw length and cyclic loading on the removal torque (RTV) in external hex (EH) and internal hex (IH) implants.
MATERIALS AND METHODS
Forty screw-retained single crowns were connected to external and internal hex implants. The prepared titanium abutment screws were classified into 8 groups based on the number of threads (n = 5 per group): EH 12.5, 6.5, 3.5, 2.5 and IH 6.5, 5, 3.5, 2.5 threads. The abutment screws were tightened with 20 Ncm torque twice with 10-minute intervals. After 5 minutes, the initial RTVs of the abutment screws were measured with a digital torque gauge (MGT12). A customized jig was constructed to apply a load along the implant long axis at the central fossa of the maxillary first molar. The post-loading RTVs were measured after 16,000 cycles of mechanical loading with 50 N at a 1-Hz frequency. Statistical analysis included one-way analysis of variance and paired t-tests.
RESULTS
The post-loading RTVs were significantly lower than the initial RTVs in the EH 2.5 thread and IH 2.5 thread groups (P<.05). The initial RTVs exhibited no significant differences among the 8 groups, whereas the post-loading RTVs of the EH 6.5 and EH 3.5 thread groups were higher than those of the IH 3.5 thread group (P<.05).
CONCLUSION
Within the limitations of this study, the external hex implants with short screw lengths were more advantageous than internal hex implants with short screw lengths in torque maintenance after cyclic loading.
Keywords: Dental implant-abutment connection, Abutment screw length, Removal torque value, Cyclic loading
INTRODUCTION
Successful restorative treatment with implants is related to the mechanical stability between the abutment and the implant.1 One of the main problems that jeopardizes this stability is screw loosening, the main causes of which include excessive abutment-implant (AI) joint bending and settling effects.2 A retrospective multicenter study showed that loosening of abutment screws occurred in 9.6% of cases over a period of 8 years at 5 private clinics.3 In a recent systematic review, the incidence of screw loosening was 2.7% in external connection implants and 2.4% in internal connection implants.4
Screw preload is the critical mechanical factor that inhibits screw loosening. The preload depends mainly on the tightening torque and also on the design of the abutment screw head and thread, the screw material and lubrication, and the design of the AI connection.2 Clinically, the abutment screw is constantly subjected to extrinsic joint separating forces, such as lateral excursive contacts, cantilevered occlusal contacts, tight interproximal contacts, nonpassive fit of the restoration, and parafunctional habits.2 The joint becomes unstable when external forces exceed the screw joint preload.5 Bickford described the development of screw loosening in two stages. Continuous functional loading initially causes slippage between the threads, releasing the tension of the screw and resulting in a decreased preload. Subsequently, the preload falls below a demanding value and therefore external forces cause the screw threads to turn, preventing its function.6
Previous studies7,8 have demonstrated that the torque used to tighten the screw is greater than the torque needed to remove the screw. To achieve increased preload levels, it has been recommended to tighten the abutment screws over their suggested torque value to be within 65% of the screw's fracture strength.9
Cyclic loading is desired to simulate masticatory function to test the stability of the screw joint. The irregularities generated during manufacturing may be minimized by mechanical cycling, which causes settling, galling, and sometimes desirable adhesive wear between the two mating surfaces of abutment screw and implant.7,8,10,11,12,13,14,15
The design of the AI connection affects screw joint stability.16 Some studies showed that the mechanical stability was higher in internal connection than in external connection systems.17,18 However, Tsuge and Hagiwara7 demonstrated that there was no significant difference between external hex (EH) and internal hex (IH) implants with respect to their effects on the loosening of the abutment screw. Nevertheless, Shin et al.16 and Kim et al.19 reported that EH implants were more advantageous than IH implants with regard to torque maintenance after cyclic loading.
Generally, the abutment screw consists of a flat head seat, long stem length, and 6 threads.20 Sopwith21 reported that the numbers of engaged threads were not very significant because most of the loads were carried by the first 3 to 4 threads. Furthermore, a minimum number of threads could reduce friction. The number of threads of the abutment screws in commercially available implant systems varies from 6 to 12.5 in EH implants and is usually shorter than 7.5 in IH implants.
Recently, Kim et al.22 reported that short screws over 3.5 threads had enough fracture strength to protect the integrity of the AI joint in EH implants. Another study found that there were no significant differences in RTVs after repeated tightening and loosening with respect to the abutment screw length.23 Yeo et al.24 also showed that short abutment screws over 3.5 threads could be used to sustain joint stability in EH implants after 2000 thermocycles between 4 and 60℃.
To prevent time consumption and hand fatigue during tightening with longer abutment screws, the joint stability of EH and IH implants with short abutment screws should be studied. No studies were found in the literature that evaluated the AI joint stability in implant-supported single crowns with short abutment screws. The purpose of this study was to evaluate the effects of abutment screw length and cyclic loading on the removal torque values (RTVs) in EH and IH implants. The null hypothesis was that there was no difference in RTVs of EH and IH implants with various abutment screw lengths between before and after cyclic loading.
MATERIALS AND METHODS
Twenty EH (USII, Osstem, Seoul, Korea) and 20 IH implants (TSIII, Osstem, Seoul, Korea) were used. Abutments for screw-retained single crowns (UCLA temporary abutment, Osstem Seoul, Korea) were connected to each implant (Table 1, Fig. 1). The abutment screws were classified into 8 groups according to the length and the number of threads. To prepare different abutment screw lengths, the screws were cut with aluminum oxide cutting discs (Se-Jong) and polished with a gold polishing kit (Shofu Inc., Kyoto, Japan) (Fig. 2A, Table 2).
Table 1. Characteristic of the Experimental Implant-Abutment Assemblies.
Implant | Abutment | Abutment screw | ||
---|---|---|---|---|
External hex | Type | US II regular | UCLA Temporary | - |
Diameter | 4 mm | 4.5 mm | 1.9 mm | |
Platform | 2.7 mm external hexagon | |||
Brand code | US2S4011S | TAR200 | ASR200 | |
Material | Ti Grade 3 | Ti Grade 3 | Ti alloy | |
Internal hex | Type | TS III regular | UCLA Temporary | |
Diameter | 4.5 mm | 4.5 mm | 1.9 mm | |
Platform | 11° Morse taper/2.5 mm internal hexagon | - | ||
Brand code | TS3S4510S | GSTTA4510 | GSABSST | |
Material | Ti Grade 3 | Ti Grade 3 | Ti alloy |
Ti = Titanium
Table 2. Experimental abutment screw groups.
Groups | Number of threads | Abutment screw length |
---|---|---|
EH 12.5 | 12.5 | 5.0 mm Original screw length |
EH 6.5 | 6.5 | 2.6 mm |
EH 3.5 | 3.5 | 1.4 mm |
EH 2.5 | 2.5 | 1.0 mm |
IH 6.5 | 6.5 | 2.6 mm Original screw length |
IH 5.0 | 5.0 | 2.0 mm |
IH 3.5 | 3.5 | 1.4 mm |
IH 2.5 | 2.5 | 1.0 mm |
Abutment crowns were fabricated by self-curing acrylic resin (Tokuso Curefast, Tokuyama Dental, Tokyo, Japan) to resemble the maxillary right first molar of a dentoform. Screw-retained crowns were constructed with a bucco-lingual width of 9 mm and a mesio-distal width of 9.5 mm. To fabricate implant-supported single crowns with the same width and shape, a template of the dentoform tooth was duplicated with vinyl polysiloxane (Exafine Putty Type, GC Corp, Tokyo, Japan). The implant specimens were positioned vertically in 2.5 × 2.5 × 2.1 cm resin blocks (Orthoplast, Vertex Dental, Zeist, Netherlands) (Fig. 2B).
The abutments were fastened to the implants, while the resin blocks were fixed to the jig. Then, 20 Ncm of recommended insertion torque was applied to each abutment screw using a hand torque wrench (Biomet 3i).25 Ten minutes later, the same tightening torque was applied again to compensate for embedment relaxation (Fig. 3A).26 After 5 minutes, the initial RTV was measured with a digital torque gauge (MGT12, MARK-10 Co., New York, NY, USA) (Fig. 3B and Fig. 3C). Subsequently, to prepare the implant assembly for the loading test, the abutment screws were tightened with 20 Ncm torque twice with 10-minute intervals. The access holes were filled with gutta percha and composite resin (Filtek Z250, 3M ESPE, St. Paul, MN, USA).25
Customized jigs were constructed to apply load parallel to the implant long axis at the central fossa of the maxillary right first molar (Fig. 4). The bottom of the specimen was fixed to the jig, which was installed in a cyclic loading machine (WON CLMC-04-1, WON Engineering, Iksan, Korea). Flat surfaces were prepared for loading at the center of the access hole, which was lubricated with thin layers of grease to reduce friction. The implant assemblies were loaded with 50 N at a 1-Hz frequency for 16,000 cycles, which corresponded to approximately 3 weeks of normal masticatory function.27 After cyclic loading, the post-loading RTVs were measured with a digital torque gauge, as described in the measurement of initial RTVs.
Selected specimens were examined using a scanning electron microscope (SEM, JEOL, JSM-6360, Tokyo, Japan) to detect possible microdamage on the screw surface. Examinations of the screw threads before and after loading at ×200 magnification were analyzed.
One-way analysis of variance (ANOVA) and Tukey's (HSD) tests were performed for analysis of initial and post-loading RTVs among the 8 groups. The paired t-test was performed to compare initial RTVs with post-loading RTVs in each group. P < .05 was considered to represent a statistically significant difference.
RESULTS
The RTVs' means, standard deviations (SD), % torque maintenance (%TM) and the difference between the initial and post-loading RTVs are shown in Table 3. The mean values of both the initial and the post-loading RTVs were lower than the insertion torque in all groups (Table 3, Fig. 5). The paired t-test revealed that the post-loading RTVs were significantly lower than the initial RTVs in the EH 2.5 and IH 2.5 groups (P < .05). One-way ANOVA and Tukey's HSD test did not reveal statistically significant differences in the initial RTVs among the 8 groups, whereas the post-loading RTVs of the EH 6.5 and EH 3.5 groups were higher than those of the IH 3.5 group (P < .05).
Table 3. Initial and Post-loading RTVs (Means and Standard deviation SD in Ncm), % Torque maintenance (TM), and Difference in RTV.
Groups | Initial RTV | Post-loading RTV | Difference in RTV | ||
---|---|---|---|---|---|
Mean (SD) | % TM | Mean (SD) | % TM | Mean (SD) | |
EH12.5 | 14.11 (3.50) | 70 | 12.06 (0.57)ab | 60 | 2.05 (3.86) |
EH 6.5 | 13.43 (3.86) | 67 | 15.22 (4.05)a | 76 | - 1.79 (3.95) |
EH 3.5 | 12.17 (3.17) | 61 | 16.74 (3.66)a | 84 | - 4.56 (5.30) |
EH 2.5 | 15.39 (2.02)A | 77 | 14.61 (1.78)ab,B | 73 | 0.77 (5.56) |
IH 6.5 | 13.04 (1.21) | 65 | 13.41( 6.24)ab | 67 | - 0.37 (5.41) |
IH 5.0 | 15.18 (3.21) | 76 | 10.43 (2.61)ab | 52 | 4.74 (3.85) |
IH 3.5 | 12.67 (3.69) | 63 | 7.47 (3.15)b | 37 | 5.20 (5.55) |
IH 2.5 | 16.89 (2.80)A | 84 | 9.91 (4.42)ab,B | 50 | 6.98( 2.06) |
Means with different small letters in the same column (P < .05; Tukey test), and capital letters in the same row (P < .05; paired t-test), were significantly different.
- = Negative means increased post-loading RTV.
The SEM images showed rough spots, grooves, and irregularities on the surface of the new abutment screws. Structural changes, such as linear scratches, rounding of the thread crests, and mild burnishing on the thread flanks, were observed on the abutment screws after loading.15 However, abnormal wear or damage caused by cyclic loading was not observed on abutment screws in all of the groups (Fig. 6).
DISCUSSION
The null hypothesis of this study was rejected based on the significant differences in RTVs for before and after loading among the various abutment screw groups. RTV is a measure of the remaining preload in the abutment screw.10 Abutment screw length and cyclic loading affected RTVs in EH and IH implants. The mean initial RTVs were lower than the insertion torque of 20 Ncm in all groups, ranging from 12.17 to 16.89 Ncm in this study. The %TM of the initial RTV was approximately 61 to 84%. The reduced RTVs in comparison to the tightening torque resulted from embedment relaxation.9 When the abutment screw is subjected to external loads, micromovement occurs between the screw and the internal thread of the implant. The fine irregularities and microroughness on the screw surface, shown in SEM examination (Fig. 6A and Fig. 6C), were smoothened and flattened. Wear of the contact areas brings the two metallic surfaces closer to each other.6,10,26 It has been reported that 2 to 10% of the initial preload is lost as a result of settling.7,9,11 Tsuge and Hagiwara7 compared the initial RTVs of the titanium (Ti) alloy abutment screw tightened with 20 Ncm insertion torque in EH and IH implants (Osseotite, Biomet 3i), yielding 18.9 Ncm (94.5%) and 18.3 Ncm (91.5%), respectively.
In contrast to their study, %TM in other studies was usually lower than 90%. Haack et al.28 measured RTVs in UCLA abutment/EH implant assemblies (Implant Innovations), yielding values of 74.6% in the gold alloy screw and 74.2% in Ti alloy screw after tightening with 32 Ncm of insertion torque. Khraisat et al.10 also reported that the initial RTV decreased to 77% (24.5 Ncm) of the insertion torque in CeraOne abutment/Brånemark MK IV implant assemblies (Nobel Biocare). Delben et al.8 also reported that the initial RTVs in EH implants (Biomet 3i) decreased to 68% (23.9 Ncm) of the insertion torque (35 Ncm) in prefabricated gold UCLA abutments and to 66% (23.3 Ncm) in cast Ti UCLA abutments. The initial RTVs in this study were comparable with the results of Delben et al.; our results were 14.11 Ncm (70%) and 13.04 Ncm (65 %) in the EH 12.5 and IH 6.5 groups, respectively, with the original screw length after tightening with a 20-Ncm insertion torque.
The abutment screw threads differ among implant systems. The thread number of the original screws used in this study was 12.5 for EH implants (USII, Ossetem), the same as in the Biomet 3i implant systems; it is 7.5 for the Warantec implant system and 6 for the Nobel Biocare system. In the original design of Brånemark implants, the abutment screw had 6 threads to reduce the friction and the long stem length for the elongation.20
The length of the abutment screw did not show any significant differences in RTVs before and after cyclic loading, except for the EH 2.5 and IH 2.5 groups. These findings were in accordance with the those of Sopwith21 and Choi et al.,23 who confirmed that 3 to 4 threads in screw were adequate to maintain the integrity of the AI joint. Yeo et al.24 also demonstrated that thermal cycling had no significant effect on the RTVs of 7 groups with different abutment screw lengths in EH implants (Warantec). RTVs before and after thermal cycling ranged from 23.0 Ncm (77%) to 25.5 Ncm (85%) in the EH 3.5, 6.5, and 9.5 groups after tightening with 30-Ncm torque. The mean initial and postloading RTVs in the EH 3.5, 6.5, 12.5 groups were between 12.06 Ncm (60%) and 16.74 Ncm (84%).
The initial and post-loading RTVs in the EH 12.5 group were 14.11 and 12.06 Ncm, respectively, and the values showed no significant difference in this study. These findings were in accordance with those of Delben et al.,8 who demonstrated that the mechanical loading with 1 × 106 cycles of 50 N had no significant effect on RTV differences in 4 of the 5 groups. However, they reported that the cast Ti abutment with a resin veneer showed significantly higher RTV after cyclic loading. Tsuge and Hagiwara7 also reported that the post-loading RTVs were significantly increased relative to the initial RTVs in both implant systems, even though the initial RTV was approximately 10% lower than the insertion torque. They explained the increase in RTV was due to the improved fit from desirable adhesive wear between the thread of the Ti abutment screw and the internal thread of the implants after lateral cyclic loading. However, they concluded that the difference in the two AI connection systems had no effect on the difference of RTVs before and after loading, but the difference in two screw materials did. This may explain the increased post-loading RTVs in correlation to the initial RTVs for the EH 6.5, EH 3.5, and IH 6.5 groups in this study.
Nonetheless, Cibirka et al.1 stated that in Procera machined abutment/EH implants (Nobel Biocare), the RTV of the gold alloy screw tightened with 32 Ncm decreased to 14.40 Ncm after dynamic loading of 20 to 200 N for 5 × 106 cycles. Khraisat et al.14 also found that 1 × 106 cycles of lateral cyclic loading with 0 to 50 N at 1.25 Hz significantly lowered the RTV in CeraOne abutment/EH implants (Nobel Biocare) compared to a 0.5 × 106 loading cycle. Khraisat et al.14 concluded that long-term loading significantly affected the RTVs under a centric lateral load. Additionally, Al Jabbari et al.15 recommended applying torque to the abutment screws again after the first 6 months of service, followed by annual reapplication of torque.
The screw material, shape, size, and coating material are crucial to maintaining preload and the stability of the AI connection.18 In IH implants, the joint stability of the implant-supported single crown is achieved from the clamping force of the abutment screw and the frictional force developed by contact between the conical mating metallic surfaces of the AI connection.29 However, the IH 3.5 group exhibited significantly lower post-loading RTV in comparison with those of EH 6.5 and EH 3.5 groups in this study.
There are three studies that examined the RTVs of GSII implants, the predecessor of TSIII implants used in this study. Kim and Shin30 compared the initial and post-loading RTVs of 18.12 Ncm (60%) and 17.67 Ncm (59%), respectively, and reported no significant differences. Shin et al.16 also compared the initial and post-loading RTVs in IH and EH implants using tungsten carbide/carbon-coated titanium (WC/C Ti) alloy abutment screws in both systems. The insertion torque, initial RTV, and post-loading RTV after 1 × 105 cycles of 5-mm off-axis loading with a 10 to 150 N load at 10 Hz were 30, 26.0, and 24.6 Ncm in USII implants, and 30, 24.8, and 12.5 Ncm in GSII implants, respectively. They confirmed that IH implants had a weaker AI connection than EH implants.
Kim et al.19 reported that two-piece abutments in GSII implants exhibited 2.4 µm of settling after mechanical loading with 1 × 105 cycles at 14-Hz under 250 N. The post-loading RTV (16.92 Ncm) was significantly lower than the initial RTV (20.5 Ncm), whereas two-piece abutments in USII implants showed 0.6 µm of settling and no difference in RTVs before and after cyclic loading. They explained that the elongated abutment screw was shortened with the settling of the abutment in GSII implants after loading, which could result in a significant decrease in the post-loading RTV. However, the abutment screws in GSII implants consisted of WC/C Ti alloy, while those in USII were Ti alloy.
The post-loading RTVs ranged from 7.47 to 16.74 Ncm in all groups. The differences in RTV before and after loading ranged from 0.77 to - 4.56 Ncm in EH implants and from 6.98 to - 0.37 Ncm in IH implants. There was no significant difference in the RTVs within the IH 6.5 group before and after loading in this study. However, the post-loading RTV of the IH 3.5 group was 7.47 Ncm, and it showed 37%TM. The decrease in post-loading RTVs tended to be greater in IH relative to EH implants, and these results were in agreement with those of previous studies.16,19
Even with the reduction of RTV, no group exhibited abutment screw loosening after loading. This might imply that the remaining torque maintained screw joint stability for a longer period within the conditions of this study.
The limitation of this study was that the abutments used were not the stock abutments. Temporary abutments were inevitably selected in both AI connection systems due to the differences in screw materials: Ti alloy for USII implants and WC/C Ti alloy for TSIII implants. The short period of cyclic loading was applied to test the temporary restorations.24 Future studies should consider applying a lengthened period of cyclic loading in stock abutments with various lengths of abutment screws.
CONCLUSION
Both the mean initial and the post-loading RTVs were lower than the insertion torque in all groups. The post-loading RTVs were significantly lower than the initial RTVs in the EH 2.5 thread and IH 2.5 thread groups (P < .05). The initial RTVs exhibited no significant differences among all 8 groups, whereas the post-loading RTVs of the EH 6.5 and EH 3.5 groups were higher than those of the IH 3.5 groups (P < .05).
Within the limitations of this study, abutment screws with a minimum of 3.5 threads showed no significant difference in RTVs in IH and EH implants after cyclic loading. The EH implants with over 3.5 threads were much more advantageous than the IH implants with regard to torque maintenance after cyclic loading. In the case of EH implants, shortening of the abutment screw length would be recommended for the preservation of preload and ease of manipulation.
ACKNOWLEDGEMENTS
The authors thank Osstem for its generous support to make specimens.
References
- 1.Cibirka RM, Nelson SK, Lang BR, Rueggeberg FA. Examination of the implant-abutment interface after fatigue testing. J Prosthet Dent. 2001;85:268–275. doi: 10.1067/mpr.2001.114266. [DOI] [PubMed] [Google Scholar]
- 2.Jörnéus L, Jemt T, Carlsson L. Loads and designs of screw joints for single crowns supported by osseointegrated implants. Int J Oral Maxillofac Implants. 1992;7:353–359. [PubMed] [Google Scholar]
- 3.Bianco G, Di Raimondo R, Luongo G, Paoleschi C, Piccoli P, Piccoli C, Rangert B. Osseointegrated implant for single tooth replacement: A retrospective multicenter study on routine use in private practice. Clin Implant Dent Relat Res. 2000;2:152–158. doi: 10.1111/j.1708-8208.2000.tb00006.x. [DOI] [PubMed] [Google Scholar]
- 4.Theoharidou A, Petridis HP, Tzannas K, Garefis P. Abutment screw loosening in single-implant restorations: a systematic review. Int J Oral Maxillofac Implants. 2008;23:681–690. [PubMed] [Google Scholar]
- 5.Richter EJ. In vivo horizontal bending moments on implants. Int J Oral Maxillofac Implants. 1998;13:232–244. [PubMed] [Google Scholar]
- 6.Bickford JH. An introduction to the design and behavior of bolted joints. New York: Marcel Dekker; 1995. pp. 515–564. [Google Scholar]
- 7.Tsuge T, Hagiwara Y. Influence of lateral-oblique cyclic loading on abutment screw loosening of internal and external hexagon implants. Dent Mater J. 2009;28:373–381. doi: 10.4012/dmj.28.373. [DOI] [PubMed] [Google Scholar]
- 8.Delben JA, Gomes EA, Barão VA, Assunção WG. Evaluation of the effect of retightening and mechanical cycling on preload maintenance of retention screws. Int J Oral Maxillofac Implants. 2011;26:251–256. [PubMed] [Google Scholar]
- 9.Stüker RA, Teixeira ER, Beck JC, da Costa NP. Preload and torque removal evaluation of three different abutment screws for single standing implant restorations. J Appl Oral Sci. 2008;16:55–58. doi: 10.1590/S1678-77572008000100011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Khraisat A, Hashimoto A, Nomura S, Miyakawa O. Effect of lateral cyclic loading on abutment screw loosening of an external hexagon implant system. J Prosthet Dent. 2004;91:326–334. doi: 10.1016/j.prosdent.2004.01.001. [DOI] [PubMed] [Google Scholar]
- 11.Shigley JE. Mechanical engineering design. 3rd ed. New York: McGraw Hill; 1977. pp. 240–245. [Google Scholar]
- 12.Weiss EI, Kozak D, Gross MD. Effect of repeated closures on opening torque values in seven abutment-implant systems. J Prosthet Dent. 2000;84:194–199. doi: 10.1067/mpr.2000.108069. [DOI] [PubMed] [Google Scholar]
- 13.Delben JA, Barão VA, Dos Santos PH, Assunção WG. Influence of abutment type and esthetic veneering on preload maintenance of abutment screw of implant-supported crowns. J Prosthodont. 2014;23:134–139. doi: 10.1111/jopr.12065. [DOI] [PubMed] [Google Scholar]
- 14.Khraisat A, Abu-Hammad O, Dar-Odeh N, Al-Kayed AM. Abutment screw loosening and bending resistance of external hexagon implant system after lateral cyclic loading. Clin Implant Dent Relat Res. 2004;6:157–164. doi: 10.1111/j.1708-8208.2004.tb00216.x. [DOI] [PubMed] [Google Scholar]
- 15.Al Jabbari YS, Fournelle R, Ziebert G, Toth J, Iacopino AM. Mechanical behavior and failure analysis of prosthetic retaining screws after long-term use in vivo. Part 1: Characterization of adhesive wear and structure of retaining screws. J Prosthodont. 2008;17:168–180. doi: 10.1111/j.1532-849X.2007.00265.x. [DOI] [PubMed] [Google Scholar]
- 16.Shin HM, Huh JB, Yun MJ, Jeon YC, Chang BM, Jeong CM. Influence of the implant-abutment connection design and diameter on the screw joint stability. J Adv Prosthodont. 2014;6:126–132. doi: 10.4047/jap.2014.6.2.126. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Gracis S, Michalakis K, Vigolo P, Vult von Steyern P, Zwahlen M, Sailer I. Internal vs. external connections for abutments/reconstructions: A systematic review. Clin Oral Implants Res. 2012;23:202–216. doi: 10.1111/j.1600-0501.2012.02556.x. [DOI] [PubMed] [Google Scholar]
- 18.Feitosa PC, de Lima AP, Silva-Concílio LR, Brandt WC, Neves AC. Stability of external and internal implant connections after a fatigue test. Eur J Dent. 2013;7:267–271. doi: 10.4103/1305-7456.115407. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Kim KS, Han JS, Lim YJ. Settling of abutments into implants and changes in removal torque in five different implant-abutment connections. Part 1: Cyclic loading. Int J Oral Maxillofac Implants. 2014;29:1079–1084. doi: 10.11607/jomi.3383. [DOI] [PubMed] [Google Scholar]
- 20.Binon PP. Implants and components: entering the new millennium. Int J Oral Maxillofac Implants. 2000;15:76–94. [PubMed] [Google Scholar]
- 21.Sopwith DG. The distribution of load in screw threads. Proc Inst Mech Eng. 1948;159:373–383. [Google Scholar]
- 22.Kim BJ, Yeo IS, Lee JH, Kim SK, Heo SJ, Koak JY. The effect of screw length on fracture load and abutment strain in dental implants with external abutment connections. Int J Oral Maxillofac Implants. 2012;27:820–823. [PubMed] [Google Scholar]
- 23.Choi JH, Yang JH, Cho WP, Lee JB. The influence of abutment screw length and repeated tightening on screw loosening in dental implant. J Korean Acad Prosthodont. 2006;44:432–442. [Google Scholar]
- 24.Yeo IS, Lee JH, Kang TJ, Kim SK, Heo SJ, Koak JY, Park JM, Lee SY. The effect of abutment screw length on screw loosening in dental implants with external abutment connections after thermocycling. Int J Oral Maxillofac Implants. 2014;29:59–62. doi: 10.11607/jomi.3125. [DOI] [PubMed] [Google Scholar]
- 25.Park SD, Lee Y, Kim YL, Yu SH, Bae JM, Cho HW. Microleakage of different sealing materials in access holes of internal connection implant systems. J Prosthet Dent. 2012;108:173–180. doi: 10.1016/S0022-3913(12)60143-7. [DOI] [PubMed] [Google Scholar]
- 26.Winkler S, Ring K, Ring JD, Boberick KG. Implant screw mechanics and the settling effect: Overview. J Oral Implantol. 2003;29:242–245. doi: 10.1563/1548-1336(2003)029<0242:ISMATS>2.3.CO;2. [DOI] [PubMed] [Google Scholar]
- 27.Graf H. Bruxism. Dent Clin North Am. 1969;13:659–665. [PubMed] [Google Scholar]
- 28.Haack JE, Sakaguchi RL, Sun T, Coffey JP. Elongation and preload stress in dental implant abutment screws. Int J Oral Maxillofac Implants. 1995;10:529–536. [PubMed] [Google Scholar]
- 29.Maeda Y, Satoh T, Sogo M. In vitro differences of stress concentrations for internal and external hex implant-abutment connections: a short communication. J Oral Rehabil. 2006;33:75–78. doi: 10.1111/j.1365-2842.2006.01545.x. [DOI] [PubMed] [Google Scholar]
- 30.Kim ES, Shin SY. Influence of the implant abutment types and the dynamic loading on initial screw loosening. J Adv Prosthodont. 2013;5:21–28. doi: 10.4047/jap.2013.5.1.21. [DOI] [PMC free article] [PubMed] [Google Scholar]