Skip to main content
Journal of Clinical Pathology logoLink to Journal of Clinical Pathology
. 1971 Nov;24(8):701–704. doi: 10.1136/jcp.24.8.701

Quality control in routine haemoglobinometry

I Cavill 1
PMCID: PMC477136  PMID: 5130537

Abstract

An analysis of haemoglobin estimations in a routine laboratory shows that quality control methods based on mean values are unsatisfactory. The use of a cusum method together with control samples is a sensitive and convenient technique for the early discovery of minor deviations in results and it is not dependent on a stable population of patients.