In vitro assessment of the accuracy of extraoral periapical radiography in root length determination

Muhammad Rizwan Nazeer 1,, Farhan Raza Khan 1, Munawwar Rahman 1
PMCID: PMC4784151  PMID: 27011737

Abstract

Objective:

To determine the accuracy of extra oral periapical radiography in obtaining root length by comparing it with the radiographs obtained from standard intraoral approach and extended distance intraoral approach.

Materials and Methods:

It was an in vitro, comparative study conducted at the dental clinics of Aga Khan University Hospital. ERC exemption was obtained for this work, ref number 3407Sur-ERC-14. We included premolars and molars of a standard phantom head mounted with metal and radiopaque teeth. Radiation was exposed using three radiographic approaches: Standard intraoral, extended length intraoral and extraoral. Since, the unit of analysis was individual root, thus, we had a total of 24 images. The images were stored in VixWin software. The length of the roots was determined using the scale function of the measuring tool inbuilt in the software. Data were analyzed using SPSS version 19.0 and GraphPad software. Pearson correlation coefficient and Bland–Altman test was applied to determine whether the tooth length readings obtained from three different approaches were correlated. P = 0.05 was taken as statistically significant.

Results:

The correlation between standard intraoral and extended intraoral was 0.97; the correlation between standard intraoral and extraoral method was 0.82 while the correlation between extended intraoral and extraoral was 0.76. The results of Bland–Altman test showed that the average discrepancy between these methods is not large enough to be considered as significant.

Conclusions:

It appears that the extraoral radiographic method can be used in root length determination in subjects where intraoral radiography is not possible.

Keywords: Dental phobia, exaggerated gag reflex, extraoral periapical radiograph, film holder, intraoral periapical radiograph, radiographic methods

INTRODUCTION

Radiographic examination is one of the primary diagnostic tools used in dentistry to determine a disease state and to formulate an appropriate treatment plan.[1] At present, various radiographic techniques are employed for orofacial imaging. A periapical radiograph is the most frequently prescribed radiograph in dental practice. Intraorally, a periapical radiograph can be obtained either by a paralleling or a bisecting angle technique, in which paralleling method is the commonly employed approach.[2]

Certain indications for periapical radiograph include detection of dental caries, periapical pathologies, assessment of periodontal status, and root morphology before extraction, trauma to teeth and the associated structures, assessment for implant surgery as well as in endodontic procedures.[2] During root canal therapy, determination of the working length of the tooth is the most essential step in subsequent steps such as cleaning, shaping, and obturation of the root canal system cannot be performed without accurate working length estimation.[3] Failure to obtain accurate working length can cause certain procedural complications such as ledge formation, apical perforation, over or under filling[4,5] which can ultimately reduce the success rate of root canal therapy.[4]

In some patients, obtaining a correct intraoral periapical radiograph can be very difficult. These include individuals with the severe gag reflex, pediatric patients who are reluctant to intraoral film placement, patients with macroglossia or microstomia, postradiation therapy patients with trismus or patients with limited sulcus depth, etc.[6] For such patients, Newman and Friedman[7] introduced an alternative technique in which diagnostic periapical radiographs were taken by using extraoral approach. The film was placed on the cheek adjacent to the buccal surface of tooth and X-ray beam were exposed from the opposite side of the face. In 2007, a device was developed by Chen et al.,[8] which can be used successfully to obtain X-ray by extra-oral technique. Studies have been published in which diagnostic X-ray are successfully obtained using extraoral approach.[6,9,10,11,12]

Hence, the aim of the study was to evaluate the accuracy of extra oral periapical (EOPA) radiograph for determining working length by comparing it with two other sets of radiograph obtained with intraoral approaches.

MATERIALS AND METHODS

This in vitro experimental study was carried out in dental clinics of Aga Khan University Hospital Karachi, Pakistan. The Ethical Review Committee of the Institution approved the study (reference number 3407Sur-ERC-14). We carried out an in vitro experimental study for which we used phantom head mounted with radiographic training models in which premolar and molar teeth were included as part of inclusion criteria. The radiopaque metallic teeth casts were our experimental model.

Periapical radiographs of upper and lower premolar and molars were obtained from three different approaches; so we divided our images into three groups depending on the technique used.

  • Group A: Standard intraoral [Figure 1]

  • Group B: Extended distance intraoral [Figure 2]

  • Group C: Extra periapical oral radiograph [Figure 3].

Figure 1.

Figure 1

Periapical radiograph of mandibular left first molar taken from standard intraoral approach

Figure 2.

Figure 2

Periapical radiograph of mandibular left first molar taken from extended distance intraoral approach

Figure 3.

Figure 3

Periapical radiograph of mandibular left first molar taken from extraoral approach

8 teeth/side exposed with three radiographic approaches gave us a total 24 images for Group A images we used standard cone indicating device. For Group B, the same cone indicating the device was used but with an X-ray source to sensor distance of 160 mm as verified by a ruler scale. For Group C images, we have attempted to develop a customized beam aiming device as suggested by Chen et al.[8]

Armamentarium of customized beam aiming device for extraoral periapical radiograph

Following are the components of aiming device which was designed for EOPA radiographs:

  • Two locator rings for bitewing radiography (Rinn, Dentsply, York, PA)

  • Two metal supporting indicator rods

  • A bite block for horizontal bitewing radiography.

Assembling the components

The two metal indicator rods were inserted into two locator rings. The bite block was attached to one end of the indicator rod, and the sensor was firmly tied to the bite block. The two metal indicator rods were finally soldered. The reason for soldering rods instead of using rubber tube is to provide a stable and more rigid support during imaging as shown in Figure 4.

Figure 4.

Figure 4

Customized beam aiming device used for extraoral periapical radiograph

Gendex digital X-ray imaging system was used to obtain radiographs using three approaches described earlier. The voltage and ampere settings were kept constant at 70 kV and 15 mAs respectively. The duration time was kept constant at 0.8 s. Although, there is a negligible risk of harm with the use of digital radiograph but lead apron was worn by the investigator to prevent any radiation hazards.

Technique for extraoral periapical radiograph

The phantom head was connected to the dental, with its mouth opened as wide as possible, so that for extraoral images the X-ray beam can pass to the sensor in an unobstructed manner from the opposite side of the mouth. The sensor was then placed on the external surface of the cheek, directly buccal to the tooth. The X-ray cone was angled −20 to −30° for maxillary premolar and molar teeth and +20 to +30° for the mandibular premolars and molars from the horizontal plane [Figure 5].

Figure 5.

Figure 5

Extraoral periapical radiographic method for mandibular teeth using custom beam aiming device

The images were stored in VixWin software version 1.5 by Accusoft soft corporation, the USA already installed in our dental clinic setup. The length was determined using the scale function of the measuring tool inbuilt in the software.

Data were analyzed using SPSS for Windows (version 19.0, SPSS Inc. Chicago, USA). The mean and standard deviation of the root length of the three groups was computed. Pearson's correlation test was used to measure the strength and direction of association that exists between different radiographic techniques. Bland–Altman plot was used to assess whether these approaches were interchangeable. P = 0.05 was taken as statistically significant.

RESULTS

Length of the roots obtained from three different approaches were analyzed using SPSS version 19.0 and GraphPad software (Prism 6 GraphPad Inc., California, USA) Pearson's correlation coefficient was derived to measure the strength of the association between the two variable techniques. The correlation between standard intraoral and extended intraoral was significant and found to be 0.97; correlation between standard intraoral and extraoral method was again significant and turned out to be 0.82 while correlation between extended intraoral and extraoral method was 0.76 as shown in Table 1.

Table 1.

Correlation between different radiographic techniques

graphic file with name EJD-10-34-g006.jpg

Bland–Altman plot was drawn by plotting the differences between the two techniques against the averages of the two techniques. The results of Bland–Altman test showed that the average discrepancy between these methods is not large enough to be considered as significant as shown in Graphs 13.

Graph 1.

Graph 1

Bland–Altman plot comparing standard intraoral periapical technique with the extended distance intraoral technique

Graph 3.

Graph 3

Bland–Altman plot comparing extended distance intraoral technique with extra-oral periapical radiographic technique

Graph 2.

Graph 2

Bland–Altman plot comparing standard intraoral periapical technique with the extra-oral periapical technique

DISCUSSION

For successful nonsurgical root canal therapy and minimizing postoperative discomfort, it is necessary that root canal system should be thoroughly debrided. One of the most important steps in root canal preparation is the determination of precise working length. All other steps of the procedure such as cleaning, shaping, and obturation cannot be preceded without determining the correct working length.[13,14]

There are various methods of determining root canal length which includes radiographs, apex locaters, tactile sensation, and other adjunctive methods but the radiographic method is being the most common employed for measuring working length.[15,16]

The most common radiographs used for working length determination are intraoral periapical radiograph.[17] However, it is difficult to obtain ideal periapical radiograph in certain cases such as patients with trismus, exaggerated gag reflex, pediatric patients, patients maxillary, and mandibular tori or patients with dental phobia, etc.[2] and the thick and rigid sensor of digital radiograph are increasing the pool of patients.[18] For such patients, Newman and Friedman introduced an alternative technique in which diagnostic periapical radiographs were taken by using extraoral approach. The film was placed on the cheek adjacent to the buccal surface of tooth and X-ray beam were exposed from the opposite side of the face. To avoid overlap, the patient is asked to open mouth as wide as possible, and the X-ray cone is angled approximately −55° for the maxilla and for mandibular teeth the angle is − 35°.[7] In 2007, a device was developed by Chen et al., which can be used successfully to obtain X-ray by extra-oral technique. Chen et al., also advocated a lesser vertical angulation as compare to Newman and Friedman, i.e., −20 to −30 for maxillary teeth and –10 to −15 for mandibular teeth.[8]

Saberi et al. in 2012[11] modified Newman Friedman radiographic technique by giving certain anatomic landmarks while testing the techniques on phantom head models. The vertical angulation modified by Saberi et al. is −25° for the maxilla and −20° for the mandible with 10° head tilt toward the side being examined.

However in a study conducted by Reddy et al.[9] suggested −20 to −30 for maxillary teeth and +20 to +30 for the mandible. In our study, we used then angulation suggested by Reddy et al.[9] and a 10° tilt that was suggested by Saberi et al.[11] The comparisons of the aforementioned studies are summarized in Table 2.

Table 2.

Studies reporting EOPA technique

graphic file with name EJD-10-34-g010.jpg

An in vivo study,[10] evaluating the accuracy of EOPA radiograph was reported to be 94.6% and another study conducted by Sudhakar et al.,[12] it was found to be 90.7% but in our study it was found to be 82.6%. However, EOPA technique has few limitations. This technique cannot be used to obtain radiographs of anterior maxillary and mandibular region due to the curvature of arch and difficulty in the positioning of the X-ray cone. The procedure is technique sensitive and requires proper knowledge and experience.

There are few limitations of EOPA. This technique is not indicated for anterior teeth radiograph because of the arch curvature and difficulty in positioning cone for anterior teeth. The extraoral approach is technique sensitive, and it requires precise angulation and positioning so it should be attempted multiple times ex vivo in order to get knowledge and experience. Furthermore, radiographs obtained from extraoral technique have lower image resolution when compared with a standard intraoral radiograph.

CONCLUSIONS

It appears that the extraoral radiographic method can be used in root length determination in subjects where intraoral radiography is difficult or not possible. We recommended that more studies should be conducted to formulate standard guidelines for angulation of EOPA and to evaluate its accuracy. Standardization of this technique can produce better image quality and errors like overlapping of structures etc., can be avoided.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

Acknowledgment

We would like to acknowledge Dr. Khabir Ahmed (AKUH), who helped us with the statistics.

REFERENCES

  • 1.Walton RE. Endodontic radiography. In: Torabinejad M, Walton RE, editors. Endodontics: Principles and Practice. 4th ed. St. Louis: Saunders Elsevier; 2009. [Google Scholar]
  • 2.Whaites E. Essentials of Dental Radiography and Radiology. 3rd ed. Edinburgh: Churchill Livingstone; 2002. Periapical radiography; p. 92. [Google Scholar]
  • 3.Inoue N, Skinner DH. A simple and accurate way to measuring root canal length. J Endod. 1985;11:421–7. doi: 10.1016/s0099-2399(85)80079-0. [DOI] [PubMed] [Google Scholar]
  • 4.Sinai I, Seltzer S, Soltanoff W, Goldenberg A, Bender IB. Biologic aspects of endodontics. II. Periapical tissue reactions to pulp extirpation. Oral Surg Oral Med Oral Pathol. 1967;23:664–79. doi: 10.1016/0030-4220(67)90350-7. [DOI] [PubMed] [Google Scholar]
  • 5.Kim-Park MA, Baughan LW, Hartwell GR. Working length determination in palatal roots of maxillary molars. J Endod. 2003;29:58–61. doi: 10.1097/00004770-200301000-00016. [DOI] [PubMed] [Google Scholar]
  • 6.Kumar R, Khambete N, Priya E. Extraoral periapical radiography: An alternative approach to intraoral periapical radiography. Imaging Sci Dent. 2011;41:161–5. doi: 10.5624/isd.2011.41.4.161. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Newman ME, Friedman S. Extraoral radiographic technique: An alternative approach. J Endod. 2003;29:419–21. doi: 10.1097/00004770-200306000-00010. [DOI] [PubMed] [Google Scholar]
  • 8.Chen CH, Lin SH, Chiu HL, Lin YJ, Chen YK, Lin LM. An aiming device for an extraoral radiographic technique. J Endod. 2007;33:758–60. doi: 10.1016/j.joen.2006.11.021. [DOI] [PubMed] [Google Scholar]
  • 9.Reddy SS, Kaushik A, Reddy S, Agarwa K. Extraoral periapical radiography: A technique unveiled. J Indian Acad Oral Med Radiol. 2011;23:S336–9. [Google Scholar]
  • 10.Zafar MS, Javed E. Extra oral radiography: An alternative to intraoral radiography for endodontic (Root Canal System) length determination. Eur Sci J. 2013;9:51–61. [Google Scholar]
  • 11.Saberi E, Hafezi L, Farhadmolashahi N, Mokhtari M. Modified Newman and Friedman extraoral radiographic technique. Iran Endod J. 2012;7:74–8. [PMC free article] [PubMed] [Google Scholar]
  • 12.Sudhakar S, Ramaswamy P, Smitha B, Uday G. Utility of extra-oral aiming device in imaging periapical regions of posterior teeth. J Clin Diagn Res. 2014;8:51–5. doi: 10.7860/JCDR/2014/9879.5037. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Sharma M, Arora V. Determination of working length of root canal. Med J Armed Forces India. 2010;66:231–4. doi: 10.1016/S0377-1237(10)80044-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Jarad FD, Albadri S, Gamble C, Burnside G, Fox K, Ashley JR, et al. Working length determination in general dental practice: A randomised controlled trial. Br Dent J. 2011;211:595–8. doi: 10.1038/sj.bdj.2011.1052. [DOI] [PubMed] [Google Scholar]
  • 15.Farida A, Maryam E, Ali M, Ehsan M, Sajad Y, Soraya K. A comparison between conventional and digital radiography in root canal working length determination. Indian J Dent Res. 2013;24:229–33. doi: 10.4103/0970-9290.116693. [DOI] [PubMed] [Google Scholar]
  • 16.Mohan GM, Anand VS. Accuracy of different methods of working length determination in endodontics. IOSR J Dent Med Sci. 2013;12:25–38. [Google Scholar]
  • 17.Jayasinghe RD, Weerakoon BS. Quality of working length radiographs taken and used by dental students during endodontic treatment. Int J Mod Altern Med Res. 2013;1:1–4. [Google Scholar]
  • 18.Parks ET, Williamson GF. Digital radiography: An overview. J Contemp Dent Pract. 2002;3:23–39. [PubMed] [Google Scholar]

Articles from European Journal of Dentistry are provided here courtesy of Dental Investigations Society

RESOURCES