Abstract
Background
Routine screening of prostate specific antigen (PSA) is no longer recommended because of a high rate of over-diagnosis of prostate cancer (PCa).
Objective
To evaluate the efficacy of diffusion-weighted magnetic resonance imaging (DW-MRI) for PCa detection, and to explore the clinical utility of ultrahigh b-value DW-MRI in predicting prostate biopsy outcomes.
Methodology
73 male patients were selected for the study. They underwent 3T MRI using T2WI conventional DW-MRI with b-value 1000 s/mm2, and ultrahigh b-value DW-MRI with b-values of 2000 s/mm2 and 3000 s/mm2. Two radiologists evaluated individual prostate gland images on a 5-point rating scale using PI-RADS, for the purpose of region-specific comparisons among modalities. Sensitivity, specificity, accuracy, positive predictive value (PPV), negative predictive value (NPV) and likelihood ratios (LR) were investigated for each MRI modality. The area under the receiver operating characteristic (ROC) curve (AUC) was also calculated.
Results
Results showed the improved diagnostic value of ultrahigh b-value DWI-MRI for detection of PCa when compared to other b values and conventional MRI protocols. Sensitivity values for 3000 s/mm2 in both peripheral zone (PZ) and transition zone (TZ) were significantly higher than those observed with conventional DW-MRI—Specificity values for 3000 s/mm2 in the TZ were significantly higher than other b-value images, whereas specificity values using 3000 s/mm2 in the PZ were not significantly higher than 2000 s/mm2 images. PPV and NPV between 3000 s/mm2 and the other three modalities were significantly higher for both PZ and TZ images. The PLRs and NLRs of b-value 3000 s/mm2 DW-MRI in the PZ and TZ were also recorded. ROC analysis showed greater AUCs for the b value 3000 s/mm2 DWI than for the other three modalities.
Conclusions
DW-MRI with a b-value of 3000 s/mm2 was found to be the most accurate and reliable MRI modality for PCa tumor detection and localization, particularly for TZ lesion discrimination. It may be stated that the b-value of 3000 s/mm2 is a novel, improved diagnostic biomarker with greater predictive accuracy for PCa prior to biopsy.
Introduction
Prostate cancer (PCa) is the second most commonly diagnosed cancer worldwide, accounting for about one-quarter of newly diagnosed cases in males [1,2]. The steady increase of incidence rates in PCa is largely due to widespread prostate-specific antigen (PSA) testing. However, routine screening for PSA is no longer recommended, because it is associated with a high rate of overdiagnosis [3–5]. Despite its low specificity for diagnosing PCa, PSA screening still remains the most frequently used tool for this purpose [6].The PSA test yields a positive predictive value of 25.1%, with a range of 17.0% to 57.0% [7]. Ultrasound (US)-guided systematic transrectal biopsy is the current reference standard for diagnosing prostate lesions; this invasive procedure however, requires specific expertise [8].
Diffusion-weighted magnetic resonance imaging (DW-MRI) is increasingly being used to study the abdomen and pelvis, and more specifically, the prostate [9,10]. DW-MRI is performed without administering contrast medium, and requires less time than other functional MRI techniques [11]. In clinical diagnosis, the b-value is the essential parameter that affects PCa detection capability, but the normal prostate signal intensity is often not suppressed in DWI, despite using b-values of approximately 1000 s/mm2 [12]. A higher b-value DWI would be more advantageous for highlighting the contrast between cancer signal intensity and normal tissue, because of greater diffusivity and less T2 shine-through effect. Earlier studies [13–16] indicate that ultrahigh b-value DWI improved the diagnostic accuracy for PCa detection, when compared to a standard b-value of 600–1000 s/mm2. At present, there is no consensus on the optimal b value for PCa diagnosis using 3T MRI, and the effect of an ultrahigh b-value of 3000 s/mm2 remains unclear.
The present study investigates the diagnostic accuracy of ultrahigh b-value DW-MRI at 3 T, with the aid of the Prostate Imaging Reporting and Data System: version 1(PI-RADSv1) scoring system, and US-guided systematic transrectal biopsies. The study’s objective was to explore whether ultrahigh b-value imaging could be the diagnostic biomarker in predicting prostate biopsy outcomes.
Materials and Methods
Patients
All procedures were performed in accordance with the ethical standards of the Declaration of Helsinki. The study was approved by the PLA General Hospital review board. Written informed consent was obtained from each patient prior to MRI examination. Between February 2014 and January 2015, 103 male patients who were consecutively subjected to routine screening for PCa were enrolled from a single institution. The patients’ medical history was reviewed, and they underwent a physical examination (including digital rectal examination (DRE)), and a prostate specific antigen (PSA) test. The inclusion criteria were:
Age: ≥55 years
Absence of lower urinary tract symptoms: this was indicated by the International Prostate Symptom Score (IPSS) being equal to zero
Ability and willingness to sign the form for informed consent;
Absence of contraindications to MRI examination
A physical condition that would tolerate MRI and biopsy
No evidence of PCa or any other malignancy
No history of urinary tract surgery (e.g., laser, microwave, radiofrequency ablation, stents)
No congenital malformations
An interval of three weeks between the MRI examination and biopsy
Of the 103 patients originally listed, thirty were ineligible for the investigation because of the following reasons (n indicates number of patients):
Incidence of lower urinary tract symptoms and IPSS score > 0 (n = 17)
intolerance to MRI examination (n = 1)
Refusal to undergo biopsy (n = 2)
History of previous biopsy within three months prior to MRI examination (n = 3)
Prevalence of radiotherapy, chemotherapy or other treatment of prostate disease (n = 7)
Finally, 73 patients were found to be eligible for inclusion in the present investigation. After the MRI examination, all patients underwent routine US-guided transrectal.prostate biopsies involving 12 cores. No MRI-targeted biopsies were performed.
Magnetic Resonance Imaging Protocol
MRI examinations were conducted using a 3-T MRI system (Discovery 750, GE Healthcare, Milwaukee, WI, USA). Patients were imaged in the supine position using a surface phased-array coil. Endorectal coil was not employed. Conventional MRI scan sequences included both axial fast spin-echo T1-weighted images (T1WI), and T2-weighted images (T2WI). T2WI of the entire prostate and seminal vesicles was performed in the axial, coronal, and sagittal planes by means of a fat-suppressed fast spin-echo sequence. Conventional DW-MRI with b-values of 0 and 1000 s/mm2 was also performed in the axial plane. The ultrahigh b-value DW-MRI was performed with b-values of 2000 s/mm2 and 3000 s/mm2 in the axial plane. Images were obtained by using a single-shot spin-echo planar sequence. Both axial T2WI and DW-MRI images were obtained at the same slice location (Table 1). Dynamic contrast enhanced (DCE) was excluded from the MRI protocols as DCE is not part of scanning protocols for prostate imaging in this study.
Table 1. Magnetic resonance imaging acquisition parameters.
Protocol | Sequence | TR | TE | FA | S/I | FOV | Matrix |
---|---|---|---|---|---|---|---|
T2WI | FR FSE | 4079 | 96 | 90 | 4.0/0.5 | 16 ×16 | 288 × 256 |
DW-MRI | |||||||
b = 1000 s/mm2 | SE EPI | 5000 | 73 | 90 | 4.0/0.5 | 32 ×32 | 160 × 160 |
b = 2000 s/mm2 | SE EPI | 5000 | 73 | 90 | 4.0/0.5 | 32 ×32 | 160 × 160 |
b = 3000 s/mm2 | SE EPI | 5000 | 73 | 90 | 4.0/0.5 | 32 ×32 | 160 × 160 |
Key: DW-MRI—diffusion-weighted magnetic resonance imaging; FA—flip angle; FOV—field of view; FR FSE—fast recovery fast spin echo; SE EPI—spin-echo echo-planar imaging; S/I—section thickness/intersection gap; TE = echo time; TR = repetition time
Measurement of Prostate Lesions
By imaging
Interpretation of images obtained from matched T2WI, conventional DW-MRI, and ultrahigh b-value DW-MRI of the prostate, was done by two radiologists with 3 years (L.M.) and 10 years (K.Z.) experience respectively. They were blinded to clinical data. All lesions were rated on a scale from 1 to 5, based on PI-RADS, in accordance with the European Society of Urogenital Radiology guidelines for prostate MRI (Table 2) [17].
Table 2. Evaluation of each magnetic resonance imaging sequence according to the Prostate Imaging Reporting and Data System score.
Score | T2WI for the PZ | T2WI for the TZ | DWI |
---|---|---|---|
1 | Uniform high signal intensity | Heterogeneous TZ adenoma with well-defined margins | No reduction in ADC. No increase in SI on any high b-value image (≥800) |
2 | Linear, wedge-shaped, or geographic areas of lower signal intensity, usually not well demarcated | Areas of more homogeneous low signal intensity; however well marginated, originating from the TZ/BPH | Diffuse, hypersignal intensity on ≥800 image with low ADC.No focal features (linear, triangular or geographic features are allowed) |
3 | Intermediate appearances | Intermediate appearances | Intermediate appearances |
4 | Discrete, homogeneous low signal focus/mass confined to the prostate | Ill-defined areas of more homogeneous low signal intensity | Focal areas of reduced ADC. Isointense SI on high b-value images (≥800) |
5 | Same as 4 plus extra-capsular extension/invasive behavior or mass effect on the capsule (bulging),or broad (>1.5 cm) contact with the surface | Same as 4 plus involving the anterior fibromuscular stroma or the anterior horn of the PZ, usually lenticular or water-drop shaped | Focal area/mass of reduced ADC. Hypersignal intensity on the high b-value images (≥800) |
Key: ADC—apparent diffusion coefficient; BPH—benign prostatic hyperplasia; DWI—diffusion-weighted imaging; PZ—peripheral zone; SI—signal intensity; T2WI–T2-weighted imaging; TZ—transition zone.
By biopsy
Ultrasound-guided systematic transrectal biopsies were performed by two urologists with 8 and 10 years experience, on all the sample patients in one single institution, using the same instruments and methodology. The urologists were aware of the MRI results at the time of biopsy. 12 prostate gland specimens were obtained from each patient undergoing biopsy: 6 from the peripheral zone [PZ], 4 from the transition zone [TZ], and 2 from the apex (Fig 1). Region specific comparisons among the modalities were made by dividing the prostate gland into 12 regions: bilateral PZs (including the upper, middle and lower regions), bilateral TZs (including the upper and lower regions), and bilateral apex.
Evaluation of results was carried out separately on a 5-point PI-RADS rating scale for the PZs and the TZs. The obtained results were dichotomized into numerical scores. Scores of 1 and 2 were regarded as indicating absence of cancer (normal or benign lesions), while scores of 3–5 pointed to the presence of cancer (abnormal indicative of malignant lesions) [18,19].
Statistical Analysis
Continuous variables were presented as median and range, while categorical variables were presented as percentages. Descriptive statistics for the patient, and parameters such as PSA levels, MRI, and histopathologic characteristics, were calculated for the study sample of 73 males. The 141 unique lesions identified through MRIs (99 in PZ and 42 in TZ) were compared with the help of localization and Gleason scores. The reference standard was judged by systematic evaluation of biopsy results: regions in biopsy specimens showing pathological evidence of cancer were considered as positive, whereas regions showing non-presence of cancer were considered as negative. This pattern of analysis was repeated for each prostate region including PZ and TZ. Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), negative likelihood ratios (NLRs), and positive likelihood ratios (PLRs) were calculated for each image modality. Differences between T2WI and three DW-MRI b-values (0 and 1000 s/mm2, 0 and 2000 s/mm2, and 0 and 3000 s/mm2), were analyzed using the McNemar test. Corroboration of observations made by the two readers, was achieved with the kappa coefficient. Values falling between 0.41–0.60, 0.61–0.80, and above 0.81, were respectively regarded as being in moderate, substantial or perfect agreement. Receiver operating characteristic (ROC) analysis was applied to each MRI modality. The areas under the curves (AUCs) were compared by using the Z test. Statistical analysis was performed using IBM SPSS Statistics v.19 for Windows (IBM Corp, Armonk, NY, USA). Statistical significance was defined at a p value<0.05.
Results
Patient Characteristics
Table 3 summarizes patient characteristics. PSA levels and pathological findings were calculated for the 73 sample patients. 141 unique lesions (PZ: 99; TZ: 42) were identified by performing biopsies.
Table 3. Patient characteristics.
Characteristics | Patients with cancer | Patients without cancer |
---|---|---|
Number | 54 | 19 |
Age (yrs) | 67 (55–81) | 70 (60–84) |
PSA (ng/ml) | ||
<4 | 1 (2) | 1 (5) |
4–10 | 25 (46) | 12 (63) |
10–20 | 15 (28) | 6 (32) |
>20 | 13 (24) | 0 (0) |
Gleason score | ||
3 + 3 = 6 | 16 (30) | NA |
3 + 4 = 7 | 9 (16) | NA |
4 + 3 = 7 | 13 (24) | NA |
≥8 | 16 (30) | NA |
Key: NA—not applicable; PSA—prostate specific antigen
Data is presented as median (interquartile range) for continuous variables, and numerical value (percentage) for categorical variables
Inter-Reader Agreement
Based on the PI-RADS score between the two radiologists (readers), the inter-reader agreement was found to be almost perfect for the ultrahigh b-value DW-MRI of 3000 s/mm2 (kappa values 0.81 in PZ, 0.855 in TZ). Agreement was substantial for the ultrahigh b-value DW-MRI of 2000 s/mm2 (kappa values 0.731 in PZ, 0.754 in TZ), and moderate for the conventional DW-MRI of 1000 s/mm2 (kappa values 0.554 in PZ, 0.568 in TZ) as well as T2WI (kappa values 0.523 in PZ, 0.612 in TZ).
Diagnostic Value of Magnetic Resonance Imaging
Peripheral zone lesions
141 lesions were identified in the prostate glands of the 73 sample patients. With respect to MRI analysis of 99 lesions in the peripheral zone (PZ), 65 (65.7%), 61 (61.6%), 76 (76.8%), and 84 (84.8%) were diagnosed accurately with the T2WI, conventional DW-MRI and ultrahigh b-value DW-MRI respectively. Of the 42 lesions in the transition zone (TZ), 29 (69%), 22 (52.4%), 29 (69%), and 37 (88.1%) were correctly diagnosed. Table 4 summarizes the sensitivity, specificity, PPV, NPV, PLR, and NLR for each MRI modality. The ROC curves and AUCs are displayed in Fig 2 and Table 5.
Table 4. Estimates of sensitivity, specificity, accuracy, positive predictive value and negative predictive value per modality for both peripheral zone and transition zone.
Sensitivity | Specificity | PPV | NPV | ||||||||
PZ | TZ | PZ | TZ | PZ | TZ | PZ | TZ | ||||
T2WI | 0.776 (0.655–0.865) | 0.680 (0.464–0.842) | 0.406 (0.242–0.592) | 0.706 (0.440–0.886) | 0.732 (0.612–0.827) | 0.772 (0.542–0.913) | 0.464 (0.280–0.658) | 0.600 (0.364–0.800) | |||
b1000 | 0.731 (0.607–0.829) | 0.600 (0.389–0.782) | 0.375 (0.217–0.563) | 0.412 (0.194–0.665) | 0.710 (0.587–0.810) | 0.600 (0.389–0.782) | 0.400 (0.232–0.592) | 0.412 (0.194–0.665) | |||
b2000 | 0.881 (0.773–0.943) | 0.800 (0.587–0.924) | 0.531 (0.350–0.705) | 0.529 (0.285–0.761) | 0.797 (0.685–0.878) | 0.714 (0.511–0.860) | 0.680 (0.464–0.843) | 0.643 (0.356–0.860) | |||
b3000 | 0.925 (0.827–0.972) | 0.880 (0.677–0.968) | 0.688 (0.499–0.833) | 0.882 (0.623–0.980) | 0.861 (0.755–0.928) | 0.917 (0.715–0.985) | 0.815 (0.613–0.930) | 0.833 (0.577–0.956) | |||
PLR (C) | NLR (C) | PLR (W) | NLR (W) | ||||||||
PZ | TZ | PZ | TZ | PZ | TZ | PZ | TZ | ||||
T2WI | 1.307 (0.955–1.790) | 2.31 (1.056–5.064) | 0.551 (0.323–0.941) | 0.453 (0.244–0.842) | 2.737 (1.817–4.123) | 3.400 (1.523–7.591) | 1.154 (0.754–1.766) | 0.667 (0.365–1.217) | |||
b1000 | 1.170 (0.862–1.588) | 1.020 (0.612–1.699) | 0.717 (0.436–1.778) | 0.971 (0.510–1.851) | 2.450 (1.644–3.651) | 1.500 (0.842–2.671) | 1.500 (1.015–2.216) | 1.429 (0.814–2.507) | |||
b2000 | 1.879 (1.286–2.745) | 1.700 (0.990–2.920) | 0.225 (0.111–0.455) | 0.378 (0.156–0.917) | 3.933 (2.468–6.270) | 2.500 (1.330–4.698) | 0.471 (0.256–0.865) | 0.556 (0.255–1.213) | |||
b3000 | 2.961 (1.763–4.973) | 7.480 (2.019–27.718) | 0.109 (0.045–0.259) | 0.136 (0.046–0.399) | 6.200 (3.462–11.102) | 11.000 (2.902–41.690) | 0.227 (0.101–0.510) | 0.200 (0.070–0.574)) |
Key: (C)- conventional; NLR—negative likelihood ratios; NPV—negative predictive value; PLR—positive likelihood ratio;.PPV—positive predictive value; PZ—peripheral zone; TZ—transition zone; (W)—weighted by prevalence.
Numbers are expressed as percentages. Numbers in parentheses are 95% confidence intervals.
Table 5. Modality wise areas under the curves (AUCs) for both peripheral zone and transition zone.
T2WI | b1000 | b2000 | b3000 | |||||
---|---|---|---|---|---|---|---|---|
PZ | TZ | PZ | TZ | PZ | TZ | PZ | TZ | |
AUCs | 0.591 | 0.693 | 0.553 | 0.506 | 0.698 | 0.665 | 0.806 | 0.881 |
95%CI | 0.488–0.689 | 0.532–0.826 | 0.450–0.653 | 0.347–0.663 | 0.598–0.787 | 0.503–0.803 | 0.715–0.879 | 0.744–0.960 |
In the peripheral zone, the sensitivity for ultrahigh b-value DW-MRI was 92.5% (95% confidence interval [CI]), 82.7–97.2% with a b-value of 3000 s/mm2, 88.1% (95% CI), and 77.3–94.3% for 2000 s/mm2. These values were significantly higher than those observed for T2WI (77.6% [95% CI, 65.5–86.5]; p<0.05), and conventional DW-MRI with a b-value of 1000 s/mm2 (73.1% [95% CI, 60.7–83.9]; p<0.05). The detection of lesions in the PZ was comparable with ultrahigh b-value DW-MRI at 3000 s/mm2 and 2000 s/mm2 (p = 0.249). Specificity was found to be 68.8% (95% CI, 49.9–83.3) with b-value 3000 s/mm2 images; 40.6% (95% CI, 24.2–59.2; p< 0.05) for T2WI, and 37.5% (95% CI, 21.7–56.3; p<0.05) for conventional DW-MRI at 1000 s/mm2. With respect to the PZ, these values were not significantly higher than b-value 2000 s/mm2 images (53.1% [95% CI, 35.0–70.5]; p = 0.074).
Transition zone lesions
In the TZ, sensitivity values at 3000 s/mm2 were significantly higher than those observed with conventional DW-MRI (88.0% [95% CI, 67.7–96.8] vs 60.0% [95% CI, 38.9–78.2]; p<0.05). Comparatively, the values did not reach statistical significance with images produced with b-value of 2000 s/mm2 (p = 0.480). However, the specificity value of 88.2% (95% CI, 62.3–98.0) was significantly higher statistically at 3000 s/mm2, when compared to other b-value images (p<0.05) in the TZ.
Reduction in false-positive rate
The false-positive rate of ultra-high-b-value DWI was reduced by 15.7%-31.3% in PZ, and 17.6%-47.0% in TZ. In other words, 15%–47% patients could have safely avoided biopsy by examination with b value of 3000 s/mm2 ultra-high-b-value DWI.
Other parameters
Figures for PPV obtained with b-value of 3000 s/mm2 as well as conventional DW-MRI were significantly higher, with respect to both PZ (0.861) and TZ (0.917); the same was true for NPV (PZ: 0.815; TZ: 0.833). The PLRs and NLRs of b-value 3000 s/mm2 DW-MRI were 6.2 and 0.227 respectively in the PZ, and 11 and 0.2 respectively in the TZ. ROC analysis showed greater AUCs for the b value of 3000 s/mm2 DWI, than for the other three modalities.
In an exceptional case, a male patient aged 55 years, with normal serum PSA levels, exhibited no prostate-related clinical symptoms. The ultrahigh b-value DW-MRI however, showed a suspicious lesion in the left middle region of the PZ. US-guided systematic transrectal biopsies revealed PCa with Gleason score 4 + 3 = 7 (Fig 3). Two more cases were presented as follows (Figs 4 and 5).
Discussion
Multiparametric MRI is a tool vital to PCa diagnosis, and DW-MRI has the advantages of being both free from contrast medium, and requiring less time than other MRI techniques [11, 20–22]. At present, the added value of dynamic contrast-enhanced MR imaging, which is acquired with intravenous administration of contrast agent, is not well established. Enhancement alone is not definitive for diagnosing prostate cancer, and absence of early enhancement does not exclude the possibility of being a cancerous lesion [23]. Furthermore, most published data show that the diagnostic value of DCE is modest when compared to combination of T2W and DWI [23,24].The present study explored the efficacy of ultrahigh b-value DW-MRI in predicting the outcome of prostate biopsy. The results demonstrated that for detecting prostate lesions, ultrahigh b-value DW-MRI at b-values of 2000 s/mm2 and 3000 s/mm2, were superior to conventional DW-MRI of 1000 s/mm2 imaging at 3 T. Similar studies have showed that b-value of 2000 s/mm2 DW-MRI outperformed 800 s/mm2 or 1000 s/mm2 DW-MRI at 3 T [13,14]. To the best of our knowledge, the present study is the first of its kind to show the advantage of using ultrahigh b-value (3000 s/mm2) DW-MRI for detecting PCa at 3 T.
Advances in technology have promoted the rapid growth of non-invasive methods to detect PCa. A higher b-value DW-MRI essentially provides a greater contrast between cancer and non-cancer tissue. Some earlier studies have tried to evaluate the PCa diagnostic capability of DW-MRI with an ultrahigh b-value >1000 s/mm2 [15,16]. The present study differs from that of Kim et al [15] in that the latter reported greater accuracy with 1000 s/mm2 DW-MRI in PCa localization, compared to 2000 s/mm2 at 3 T. Kim et al applied a relatively long TE (83–85 ms for 1000 s/mm2, 93–95 ms for 2000 s/mm2), whereas the present study used a shorter TE of DWI (73 ms for 1000, 2000 and 3000 s/mm2) [15]. The limitation of the Kim et al study [15] is that a relatively longer TE is indicative of more time for T2 dephasing, thereby causing a reduction of the signal-to-noise ratio (SNR). Furthermore, a study of DW-MRI, acquired with the same shorter TE for different b-values, supports the advantage of ultrahigh b-value DW-MRI [13].
Results of the present study were elicited from each prostate region. Detection of cancerous and non-cancerous lesions in the PZ, was achieved with the help of 2000 s/mm2ultrahigh b-value DW-MRI, resulting in significant improvement in specificity, PPV, and NPV. These results agree with those of Ueno et al [13]. However, these researchers [13] could not identify any significant improvement in the TZ. Cancers as pinpointing the location are difficult in this region because of the high signal intensity of normal parenchyma or benign prostate hyperplasia [16,25]. This is different from the features exhibited by cancers located in the PZ. It is important to distinguish TZ cancers accurately with imaging, in order to facilitate biopsy. Some research investigations have reported poorer diagnostic performance in the TZ for T2WI plus conventional DWI, and T2WI alone [13,14,16]. To overcome the limitations of high signal emitted by the prostate itself on DWI with b-value 1000 s/mm2, Katahira et al [16] used 2000 s/mm2 DWI that showed significantly better sensitivity, specificity, accuracy, PPV, and NPV at 1.5 T. Results of the Katahira et al study are in agreement with outcomes of the present study. However, 3 T was applied because of increased SNR using ultrahigh b-value DW-MRI at 2000 and 3000 s/mm2. The present investigation revealed that the b-value of 3000 s/mm2 markedly improved sensitivity, specificity, PPV, and NPV; moreover, there was substantial enhancement of PLR and NLR with a b-value of 2000 s/mm2. The ultrahigh b-value DW-MRI at 3000 s/mm2 had the additional advantage of detection in both the TZ and PZ, particularly in the former.
In the present study, the ultrahigh b-value DW-MRI showed outstanding diagnostic performance in the sub-group, with the 3000 s/mm2 DW-MRI alone outdoing both T2WI and conventional DW-MRI, thus implying that the ultrahigh b-value DW-MRI alone could serve as a promising biomarker for guiding the prostate biopsy decision. In the present study, a man with normal serum PSA was found to have PCa when subjected to examination with the ultrahigh b-value DW-MRI. This finding demonstrates that, especially at 3000 s/mm2, ultrahigh b-value DW-MRI seems to be the most sensitive and specific MRI modality for tumor detection and localization. Predictive values were significantly higher for both the PZ and TZ. Ultrahigh b-value DW-MRI was identified as a vital and independent predictor of PCa at the stage of prostate biopsy.
The present investigation has both advantages and limitations. Firstly, imaging was performed at 3 T with the advantage of a higher SNR, namely, improved spatial resolution and different imaging sequences [26]. Secondly, for region-specific comparisons corresponding to the results, the PPV and NPV of ultrahigh b-value DW-MRI increased respectively by 8.7–15.1% in 2000 s/mm2 and 28–41.5% in 3000 s/mm2, when compared to conventional DW-MRI in the PZ, and 9.9–30.2% and 20.5–39.5%, respectively, in the TZ.
A potential limitation of the present research effort is that all the patients were from a single center, therefore, external validation is needed through further studies. Another limitation is that lesions were determined by systematic biopsy. MRI findings do not correspond one by one to the biopsy results, because lesion-guided biopsies were not conducted. Nevertheless, attempts were made to minimize errors in the study through careful region-specific comparisons. Thirdly, we did not compare the multiparametric MRI in the diagnosis of prostate cancer. However, that was not the purpose of this study. Whether the effect of ultrahigh b-value DW-MRI could improve the grading and staging of PCa, still remains to be established through rigorous scientific evaluation.
Conclusion
It may be concluded that DW-MRI with a b-value of 3000 s/mm2is the most accurate and reliable MRI modality for PCa tumor detection and localization, particularly in the case of TZ lesion discrimination. It is superior to T2WI, and conventional DW-MRI in distinguishing patients with a high probability of PCa, when undergoing prostate biopsy. These results suggest that ultrahigh b-value DW-MRI, based on PI-RADS, helps radiologists and urologists to reach consensus on clinical decision making for prostate biopsy. It is expected to be a novel diagnostic biomarker that improves predictive accuracy for PCa. Further exploration of the value of ultrahigh b-value DW-MRI is warranted at a multicenter level.
Supporting Information
Acknowledgments
We sincerely thank Dr. Xiaojin Zhang for her technical assistance.
Data Availability
All relevant data are within the paper and its Supporting Information files.
Funding Statement
The Health Care Special Project (14BJZ02)” is The General Logistics Department of PLA. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
References
- 1.Siegel RL, Miller KD, Jemal A. Cancer statistics, 2015. CA Cancer J Clin. 2015:5–29. [DOI] [PubMed] [Google Scholar]
- 2.Malvezzi M, Bertuccio P, Levi F, La Vecchia C, Negri E. European cancer mortality predictions for the year 2014. Ann Oncol. 2014;25:1650–6. 10.1093/annonc/mdu138 [DOI] [PubMed] [Google Scholar]
- 3.Draisma G, Etzioni R, Tsodikov A, Mariotto A, Wever E, Gulati R, et al. Lead time and overdiagnosis in prostate-specific antigen screening: importance of methods and context. J Natl Cancer Inst. 2009;101:374–83. 10.1093/jnci/djp001 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Lefevre M. PSA screening: the USPSTF got it right. J Fam Pract. 2013;62:617–9. [PubMed] [Google Scholar]
- 5.Hayes JH, Barry MJ. Screening for prostate cancer with the prostate-specific antigen test: a review of current evidence. JAMA. 2014;311:1143–9. 10.1001/jama.2014.2085 [DOI] [PubMed] [Google Scholar]
- 6.Heidenreich A, Bellmunt J, Bolla M, Joniau S, Mason M, Matveev V, et al. EAU guidelines on prostate cancer. Part 1: screening, diagnosis, and treatment of clinically localised disease. Eur Urol. 2011;59:61–71. 10.1016/j.eururo.2010.10.039 [DOI] [PubMed] [Google Scholar]
- 7.Mistry K, Cable G. Meta-analysis of prostate-specific antigen and digital rectal examination as screening tests for prostate carcinoma. J Am Board Fam Pract. 2003;16:95–101. [DOI] [PubMed] [Google Scholar]
- 8.Smeenge M, de la Rosette JJ, Wijkstra H. Current status of transrectal ultrasound techniques in prostate cancer. Curr Opin Urol. 2012;22:297–302. 10.1097/MOU.0b013e3283548154 [DOI] [PubMed] [Google Scholar]
- 9.Qayyum A. Diffusion-weighted imaging in the abdomen and pelvis: concepts and applications. RadioGraphics. 2009;29:1797–1810. 10.1148/rg.296095521 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Bonekamp S, Corona-Villalobos CP, Kamel IR. Oncologic applications of diffusion-weighted MRI in the body. J Magn Reson Imaging. 2012;35:257–279. 10.1002/jmri.22786 [DOI] [PubMed] [Google Scholar]
- 11.Moore WA, Khatri G, Madhuranthakam AJ, Sims RD, Pedrosa I. Added Value of Diffusion-Weighted Acquisitions in MRI of the Abdomen and Pelvis. AJR Am J Roentgenol. 2014; 202: 995–1006. 10.2214/AJR.12.9563 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Bozkurt M, Doganay S, Kantarci M, Yalcin A, Eren S, Atamanalp SS, et al. Comparison of peritoneal tumor imaging using conventional MR imaging and diffusion-weighted MR imaging with different b values. Eur J Radiol. 2011;80:224–228. 10.1016/j.ejrad.2010.06.004 [DOI] [PubMed] [Google Scholar]
- 13.Ueno Y, Kitajima K, Sugimura K, Kawakami F, Miyake H, Obara M, et al. Ultra-high b-value diffusion-weighted MRI for the detection of prostate cancer with 3-T MRI. J Magn Reson Imaging. 2013;38:154–60. 10.1002/jmri.23953 [DOI] [PubMed] [Google Scholar]
- 14.Ohgiya Y, Suyama J, Seino N, Hashizume T, Kawahara M, Sai S, et al. Diagnostic accuracy of ultra-high-b-value 3.0-T diffusion-weighted MR.imaging for detection of prostate cancer. Clin Imaging. 2012;36:526–31. 10.1016/j.clinimag.2011.11.016 [DOI] [PubMed] [Google Scholar]
- 15.Kim CK, Park BK, Kim B. High-b-value diffusion-weighted imaging at 3 T to detect prostate cancer: comparisons between b values of 1,000 and 2,000 s/mm2. AJR Am J Roentgenol. 2010;194:W33–7. 10.2214/AJR.09.3004 [DOI] [PubMed] [Google Scholar]
- 16.Katahira K, Takahara T, Kwee TC, Oda S, Suzuki Y, Morishita S, et al. Ultra-high-b-value diffusion-weighted MR imaging for the detection of prostate cancer: evaluation in 201 cases with histopathological correlation. Eur Radiol. 2011;21:188–96. 10.1007/s00330-010-1883-7 [DOI] [PubMed] [Google Scholar]
- 17.Barentsz JO, Richenberg J, Clements R, Choyke P, Verma S, Villeirs G,et al. ESUR prostate MR guidelines 2012. Eur Radiol. 2012;22:746–57. 10.1007/s00330-011-2377-y [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Junker D, Schäfer G, Edlinger M, Kremser C, Bektic J, Horninger W, et al. Evaluation of the PI-RADS scoring system for classifying mpMRI findings in men with suspicion of prostate cancer. Biomed Res Int. 2013;2013:252939 10.1155/2013/252939 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Schimmöller L, Quentin M, Arsov C, Lanzman RS, Hiester A, Rabenalt R, et al. Inter-reader agreement of the ESUR score for prostate MRI using in-bore MRI-guided biopsies as the reference standard. Eur Radiol. 2013;23:3185–90. 10.1007/s00330-013-2922-y [DOI] [PubMed] [Google Scholar]
- 20.Hambrock T, Somford DM, Huisman HJ, van Oort IM, Witjes JA, Hulsbergen-van de Kaa CA, et al. Relationship between apparent diffusion coefficients at 3.0-T MR imaging and Gleason grade in peripheral zone prostate cancer. Radiology. 2011;259:453–461. 10.1148/radiol.11091409 [DOI] [PubMed] [Google Scholar]
- 21.Gibbs P, Liney GP, Pickles MD, Zelhof B, Rodrigues G, Turnbull LW. Correlation of ADC and T2 measurements with cell density in prostate cancer at 3.0 Tesla. Invest Radiol. 2009;44:572–576. 10.1097/RLI.0b013e3181b4c10e [DOI] [PubMed] [Google Scholar]
- 22.Le JD, Tan N, Shkolyar E, Lu DY, Kwan L, Marks LS, et al. Multifocality and prostate cancer detection by multiparametric magnetic resonance imaging: correlation with whole-mount histopathology. Eur Urol. 2015;67:569–76. 10.1016/j.eururo.2014.08.079 [DOI] [PubMed] [Google Scholar]
- 23.Weinreb JC, Barentsz JO, Choyke PL, Cornud F, Haider MA, Macura KJ, et al. PI-RADS Prostate Imaging- Reporting and Data System: 2015, Version 2. Eur Urol. 2016;69:16–40. 10.1016/j.eururo.2015.08.052 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Oto A, Kayhan A, Jiang Y, Tretiakova M, Yang C, Antic T, Dahi F, Shalhav AL,Karczmar G, Stadler WM. Prostate cancer: differentiation of central gland cancer from benign prostatic hyperplasia by using diffusion-weighted and dynamiccontrast-enhanced MR imaging. Radiology. 2010;257:715–23. 10.1148/radiol.10100021 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Rosenkrantz AB, Kim S, Campbell N, Gaing B, Deng FM, Taneja SS. Transition zone prostate cancer: revisiting the role of multiparametric MRI at 3 T. AJR Am J Roentgenol. 2015;204:W266–72. 10.2214/AJR.14.12955 [DOI] [PubMed] [Google Scholar]
- 26.Shah ZK, Elias SN, Abaza R, Zynger DL, DeRenne LA, Knopp MV, et al. Performance comparison of 1.5-T endorectal coil MRI with 3.0-T nonendorectal coil MRI in patients with prostate cancer. Acad Radiol. 2015;22:467–74. 10.1016/j.acra.2014.11.007 [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
All relevant data are within the paper and its Supporting Information files.