Abstract
The toxicity of different rubbers and plastics used in transfusion-giving sets has been investigated by examining their effects on (a) cultures of chick embryo tissues, (b) the oxygen uptake of guinea-pig skin tissue cultures, and (c) the growth of Str. pyogenes.
The results of the laboratory tests have been compared with the incidence of thrombophlebitis after prolonged transfusions through the various materials.
It was found that where the materials inhibited the growth of Str. pyogenes they were also toxic to tissue cultures, but that some materials which were toxic to tissue cultures did not inhibit bacterial growth.
The assessments of the relative toxicity of the materials tested by the two tissue culture methods were in agreement. The skin respiration studies, however, gave more information on the early effects of the toxic materials. The relative toxicity of the materials as revealed by these tests could be correlated with the differences in the incidence of thrombophlebitis following intravenous infusions administered through giving-sets assembled with the materials tested.
It is suggested therefore that the toxicity revealed by these tests is of clinical importance, and that tissue culture toxicity tests will prove to be of value in selecting rubbers and plastics for clinical purposes.
Full text
PDF








Images in this article
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- CARTER J. F. B., MILNE E. H., WHITTET T. D. Further investigation into causes of thrombophlebitis following intravenous infusions. Lancet. 1952 Oct 4;2(6736):660–661. doi: 10.1016/s0140-6736(52)91503-1. [DOI] [PubMed] [Google Scholar]
- CRUICKSHANK C. N. Continuous observation of the respiration of skin in vitro. Exp Cell Res. 1954 Nov;7(2):374–380. doi: 10.1016/s0014-4827(54)80084-x. [DOI] [PubMed] [Google Scholar]
- FREEMAN G. G., NEW N., NOAD R. B., WILKINSON J. F. Silicone-rubber tubing in blood-transfusion work; clinical trials. Lancet. 1956 Sep 22;271(6943):621–624. doi: 10.1016/s0140-6736(56)92310-8. [DOI] [PubMed] [Google Scholar]
- HANDFIELD-JONES R. P. C., LEWIS H. B. M. Rubber tubing as a cause of infusion thrombophlebitis. Lancet. 1952 Mar 22;1(6708):585–588. doi: 10.1016/s0140-6736(52)90099-8. [DOI] [PubMed] [Google Scholar]
- JENKINS W. J., STONE B., KNOWLES G. S., TOVEY G. H., SHARPE R. A. Experiences with a disposable plastic transfusion-giving set. Lancet. 1959 Jan 17;1(7064):139–143. doi: 10.1016/s0140-6736(59)90017-0. [DOI] [PubMed] [Google Scholar]
- LAJTHA L. G. Culture of human bone marrow in vitro; the reversibility between normoblastic and megaloblastic series of cells. J Clin Pathol. 1952 Feb;5(1):67–84. doi: 10.1136/jcp.5.1.67. [DOI] [PMC free article] [PubMed] [Google Scholar]
- MACDOUGALL J. D. Toxicity studies on silicone rubber and other substances. Nature. 1953 Jul 18;172(4368):124–125. doi: 10.1038/172124c0. [DOI] [PubMed] [Google Scholar]
- OPPENHEIMER B. S., OPPENHEIMER E. T., DANISHEFSKY I., STOUT A. P., EIRICH F. R. Further studies of polymers as carcinogenic agents in animals. Cancer Res. 1955 Jun;15(5):333–340. [PubMed] [Google Scholar]
- PARKER R. C., MORGAN J. F., MORTON H. J. Toxicity of rubber stoppers for tissue cultures. Proc Soc Exp Biol Med. 1951 Mar;76(3):444–445. doi: 10.3181/00379727-76-18518. [DOI] [PubMed] [Google Scholar]
- STRUMIA M. M., COLWELL L. S., ELLENBERGER K., MAWR B. The preservation of blood for transfusion. I. The effect of plastic containers on red cells. J Lab Clin Med. 1955 Aug;46(2):225–233. [PubMed] [Google Scholar]
- TODD R. M. Complications of intravenous fluid therapy in children. Lancet. 1951 May 5;1(6662):982–985. doi: 10.1016/s0140-6736(51)92497-x. [DOI] [PubMed] [Google Scholar]
- WALL R. L., BUCKLEY N. M., DOAN C. A. An evaluation of the preservation of human blood stored in experimental plastic containers. II. In vivo studies. J Lab Clin Med. 1953 Nov;42(5):674–680. [PubMed] [Google Scholar]