Abstract
The method of Deibler, Holmes, Campbell, and Gans (1959) is a quick, accurate, and reliable method for the determination of oxygen saturation. The results should be calibrated against those obtained by the Van Slyke method, and the setting, sensitivity, and reliability of the spectrophotometer should be carefully established. This is especially necessary to meet the stringent requirements for accurate setting at the isobestic point 805 mμ. This method is superior to a method using saponin as a haemolytic agent and reading at only one wavelength.
Full text
PDF


Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- DEIBLER G. E., HOLMES M. S., CAMPBELL P. L., GANS J. Use of triton X-100 as a hemolytic agent in the spectrophotometric measurement of blood O2 saturation. J Appl Physiol. 1959 Jan;14(1):133–136. doi: 10.1152/jappl.1959.14.1.133. [DOI] [PubMed] [Google Scholar]
- WADE O. L., BISHOP J. M., CUMMING G., DONALD K. W. A method for the rapid estimation of the percentage oxygen saturation and oxygen content of blood. Br Med J. 1953 Oct 24;2(4842):902–907. doi: 10.1136/bmj.2.4842.902. [DOI] [PMC free article] [PubMed] [Google Scholar]