Skip to main content
Journal of Clinical Pathology logoLink to Journal of Clinical Pathology
. 1963 Sep;16(5):476–478. doi: 10.1136/jcp.16.5.476

Technical and experimental errors in the spectrophotometric determination of oxygen saturation

R G Fish 1, M Radcliffe Lee 1
PMCID: PMC480615  PMID: 14063338

Abstract

The method of Deibler, Holmes, Campbell, and Gans (1959) is a quick, accurate, and reliable method for the determination of oxygen saturation. The results should be calibrated against those obtained by the Van Slyke method, and the setting, sensitivity, and reliability of the spectrophotometer should be carefully established. This is especially necessary to meet the stringent requirements for accurate setting at the isobestic point 805 mμ. This method is superior to a method using saponin as a haemolytic agent and reading at only one wavelength.

Full text

PDF
476

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. DEIBLER G. E., HOLMES M. S., CAMPBELL P. L., GANS J. Use of triton X-100 as a hemolytic agent in the spectrophotometric measurement of blood O2 saturation. J Appl Physiol. 1959 Jan;14(1):133–136. doi: 10.1152/jappl.1959.14.1.133. [DOI] [PubMed] [Google Scholar]
  2. WADE O. L., BISHOP J. M., CUMMING G., DONALD K. W. A method for the rapid estimation of the percentage oxygen saturation and oxygen content of blood. Br Med J. 1953 Oct 24;2(4842):902–907. doi: 10.1136/bmj.2.4842.902. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Clinical Pathology are provided here courtesy of BMJ Publishing Group